Use of CentriMag for refractory cardiogenic shock in a puerperal woman: case report

ABSTRACT

CONTEXT:

Heart failure in Brazil is a major public health problem and, even with advances in treatment, it still presents high morbidity and mortality. As a treatment option, mechanical circulatory assist devices (MCADs) have greatly increased in importance over the last decade.

CASE REPORT:

This report concerns a case of refractory cardiogenic shock due to acute myocarditis in a 35-year-old puerperal female patient who presented with retrosternal pain, fatigue and dyspnea. At the hospital, she was diagnosed with myocarditis. There was no improvement in perfusion even after receiving dobutamine, intra-aortic balloon passage (IAB) and venoarterial extracorporeal membrane oxygenation (VA-ECMO). Therefore, it was decided to implant a MCAD (CentriMag). During hospitalization, recovery from the bi-ventricular dysfunction was achieved. The CentriMag device was removed 10 days after it had been implanted, and the patient was discharged after another 8 days. The myocarditis was proven to be due to the Coxsackie virus.

CONCLUSIONS:

The decision to implant a MCAD should be individualized, as patient profiles do not always match the indications in the guidelines and protocols. In this study, clinical discussion of the case among the medical and multi-professional teams was essential in order to be able to successfully reverse the patient’s severe clinical condition without sequelae, through using a CentriMag implant.

KEY WORDS (MeSH terms):
Heart failure; Heart-assist devices; Extracorporeal membrane oxygenation; Myocarditis

AUTHORS’ KEY WORDS:
CentriMag; ECMO; Puerperal; Refractory cardiogenic shock

INTRODUCTION

Worldwide, there were approximately 26 million people with heart failure in 2014. This therefore represents a major public health problem.11. Ambrosy AP, Fonarow GC, Butler J, et al. The Global Health and Economic Burden of Hospitalizations for Heart Failure. J Am Coll Cardiol. 2014;63(12):1123-33. PMID: 24491689; http://dx.doi.org/10.1016/j.jacc.2013.11.053.
https://doi.org/http://dx.doi.org/10.101...
In Brazil, the scenario is not different and, even with advances in treatment, heart failure still presents high morbidity and mortality. The incidence of heart failure in Brazil is 199 cases per 100,000 person-years, and the one-year mortality rate is 24.5% (95% confidence interval, CI, 19.4%-30.0%).22. Ciapponi A, Alcaraz A, Calderón M, et al. Burden of Heart Failure in Latin America: A Systematic Review and Meta-analysis. Rev Esp Cardiol (Engl Ed). 2016;69(11):1051-60. PMID: 27553287; http://dx.doi.org/10.1016/j.rec.2016.04.054.
https://doi.org/http://dx.doi.org/10.101...
Recent data from the Department of Information Technology of the Brazilian National Health System (DATASUS) have shown that in Brazil the number of heart failure deaths was 27,461 just in 2017.33. Ministério da Saúde. DATASUS - Ocorrência de óbitos por Categoria CID-10 I50 segundo Ano do Óbito - 2017. Available from: Available from: http://tabnet.datasus.gov.br/cgi/tabcgi.exe?sim/cnv/obt10uf.def . Accessed in 2021 (Feb 18).
http://tabnet.datasus.gov.br/cgi/tabcgi....

As a treatment option, mechanical circulatory assist devices (MCADs) have greatly increased in importance over the last decade. Although there are solid guidelines for indication of MCAD implantation,44. Ayub-Ferreira S, Souza Neto J, Almeida D, et al. Diretriz de Assistência Circulatória Mecânica da Sociedade Brasileira de Cardiologia. Arq Bras Cardiol. 2016;107(2 Suppl 2):1-33. PMID: 27627754; http://dx.doi.org/10.5935/abc.20160128
https://doi.org/http://dx.doi.org/10.593...
some particular cases still need to be studied, such as cases of refractory cardiogenic shock due to myocarditis. The objective of this case report was to present a case of refractory cardiogenic shock due to acute myocarditis in a young puerperal woman.

CASE REPORT

This present case report was approved on April 13, 2020 (#1599; CAAE 30403220.7.0000.5461). A 35-year-old married female patient who was in the early puerperal period (childbirth in April 2019) sought emergency assistance with flu symptoms that she had had for approximately six days. At the emergency room, she was diagnosed with tonsillitis and was treated with azithromycin and prednisone. In the absence of symptom improvement, she presented retrosternal pain, fatigue and dyspnea. In a new medical evaluation, markers for myocardial necrosis were examined, with positive results, suggesting a diagnostic hypothesis of myopericarditis.

After admission to the coronary unit, she presented signs of low cardiac output with precordial pain, nausea and peripheral perfusion and was then referred to the advanced heart failure unit. She used vasoactive drugs (VAD), but without improvement in perfusion and clinical presentation. After her case has been discussed by the team, it was decided to perform intra-aortic balloon (IAB) passage. Because of progression of dysfunction and worsening of her general condition, it was decided to install peripheral venoarterial extracorporeal membrane oxygenation (ECMO). The patient maintained the signs of low output and low flow in ECMO, in addition to poor perfusion in the right lower limb after cannulation. It was therefore decided to replace the ECMO with a ventricular assist system (VAS) (CentriMag; Levitronix LLC, Waltham, MA, United States), on July 1, 2019. The procedure was performed by means of median sternotomy, with decannulation of the ECMO and myocardial biopsy.

During the hospitalization with the device, she received corticotherapy and immunotherapy lasting five days. During this time, she required correction of a pseudoaneurysm in the right femoral artery. In addition, she underwent decompression fasciotomy in a right anterior tibial store because of presentation of compartmental syndrome. She evolved with progressive hemodynamic stability and, on July 8, 2019, she presented aphasia of momentary expression, shown by transcranial doppler microembolization in the left middle cerebral artery. Her anticoagulation was adjusted, but the microembolization was maintained.

On July 11, 2019 (10 days after implantation), she was decannulated from the CentriMag. An intracavitary left ventricle (LV) thrombus from the device (Figure 1) was seen, which evolved with progressive weaning from VAD and corticotherapy, with clear recovery from the bi-ventricular dysfunction (ejection fraction, EF: 17% to 58%) and recovery of strength in the right forefoot. Analysis on the myocardial biopsy confirmed that the cause of the myocarditis was positivity for the Coxsackie virus (viral load of Erythroparvovirus and HHV6 type 6B).

Figure 1.
Left ventricular intracavitary thrombus of the device.

Eight days after implant removal (July 19, 2019), the patient was discharged from the hospital with preserved bi-ventricular function (Figure 2). Currently, the patient is in outpatient follow-up without complaints, presenting good quality of life and preserved heart function (EF 62%).

Figure 2.
Bi-dimensional echocardiogram with N-flow mapping after explantation of CentriMag device, showing preserved bi-ventricular function (July 16, 2019).

DISCUSSION

In this case report, we present a peculiar case of a 35-year-old puerperal woman who progressed with significant worsening of cardiac function within a few hours, even with passage of an IAB and ECMO. The decision to install a CentriMag device in this patient was crucial for enabling myocardial recovery and significant improvement of cardiac output. Reports involving use of CentriMag to treat heart failure in the postpartum period remain rare. Table 1 shows the results from a systematic search for similar studies using the PubMed and EMBASE databases.

Table 1.
Systematic review of the literature

Temporary mechanical circulatory assist devices (MCADs) are important for re-establishing the hemodynamic condition and should be indicated individually. In addition, they serve as an aid for decision-making up to the point of defining the approach to be taken in cases where immediate hemodynamic support is required (as a bridge to decision); or for recovery of ventricular function in cases of acute myocardial infarction (as a bridge to recovery); or as hemodynamic support and clinical stabilization of patients in a severe condition who are in a transplant queue (as a bridge to transplantation).44. Ayub-Ferreira S, Souza Neto J, Almeida D, et al. Diretriz de Assistência Circulatória Mecânica da Sociedade Brasileira de Cardiologia. Arq Bras Cardiol. 2016;107(2 Suppl 2):1-33. PMID: 27627754; http://dx.doi.org/10.5935/abc.20160128
https://doi.org/http://dx.doi.org/10.593...

It is known that IAB is widely used as the first option for treating heart failure,55. Peura JL, Colvin-Adams M, Francis GS, et al. Recommendations for the use of mechanical circulatory support: Device strategies and patient selection: a scientific statement from the American Heart Association. Circulation. 2012;126(22):2648-67. PMID: 23109468; https://doi.org/10.1161/CIR.0b013e3182769a54.
https://doi.org/https://doi.org/10.1161/...
but in some cases in which refractory cardiogenic shock occurs, ECMO is an excellent and rapid option. CentriMag provides a temporary option for ventricular function support until the myocardium recovers,66. Fiorelli AI, Oliveira Júnior JL, Coelho GHB, Rocha DC. Assistência circulatória mecânica: porquê e quando [Mechanical circulatory support: why and when]. Rev Med (São Paulo). 2008;87(1):1-15. Available from: Available from: https://www.revistas.usp.br/revistadc/article/view/59055/62042 . Accessed in 2021 (Feb 18).
https://www.revistas.usp.br/revistadc/ar...
thus preventing low output from leading the heart to irreversible cellular conditions.

Faced with the clinical picture of worsening of our patient’s cardiac function, the decision to implant CentriMag was made in a matter of hours. This was in accordance with the recommendations of the Interagency Registry for Mechanically Assisted Circulatory Support Classification (INTERMACS),77. Kirklin JK, Naftel DC, Kormos RL, et al. Third INTERMACS Annual Report: the evolution of destination therapy in the United States. J Heart Lung Transplant. 2011;30(2):115-23. PMID: 21215378; https://doi.org/10.1016/j.healun.2010.12.001.
https://doi.org/https://doi.org/10.1016/...
in which use of MCADs is strongly indicated for patients with the profiles INTERMACS 1 (severe cardiogenic shock) and INTERMACS 2 (progressive decline in renal, hepatic, nutritional and lactatemia function, despite use of inotropes).

CONCLUSIONS

The decision to implant a MCAD should be individualized, as patient profiles do not always match the indications in the guidelines and protocols. In this study, clinical discussion of the case among the medical and multi-professional teams was essential in order to be able to successfully reverse the patient’s severe clinical condition without sequelae, through using a CentriMag implant.

Acknowledgement:

All the authors are grateful for the patient’s agreement and the action of the professional team who was present at the assistance

REFERENCES

  • 1
    Ambrosy AP, Fonarow GC, Butler J, et al. The Global Health and Economic Burden of Hospitalizations for Heart Failure. J Am Coll Cardiol. 2014;63(12):1123-33. PMID: 24491689; http://dx.doi.org/10.1016/j.jacc.2013.11.053.
    » https://doi.org/http://dx.doi.org/10.1016/j.jacc.2013.11.053
  • 2
    Ciapponi A, Alcaraz A, Calderón M, et al. Burden of Heart Failure in Latin America: A Systematic Review and Meta-analysis. Rev Esp Cardiol (Engl Ed). 2016;69(11):1051-60. PMID: 27553287; http://dx.doi.org/10.1016/j.rec.2016.04.054.
    » https://doi.org/http://dx.doi.org/10.1016/j.rec.2016.04.054
  • 3
    Ministério da Saúde. DATASUS - Ocorrência de óbitos por Categoria CID-10 I50 segundo Ano do Óbito - 2017. Available from: Available from: http://tabnet.datasus.gov.br/cgi/tabcgi.exe?sim/cnv/obt10uf.def Accessed in 2021 (Feb 18).
    » http://tabnet.datasus.gov.br/cgi/tabcgi.exe?sim/cnv/obt10uf.def
  • 4
    Ayub-Ferreira S, Souza Neto J, Almeida D, et al. Diretriz de Assistência Circulatória Mecânica da Sociedade Brasileira de Cardiologia. Arq Bras Cardiol. 2016;107(2 Suppl 2):1-33. PMID: 27627754; http://dx.doi.org/10.5935/abc.20160128
    » https://doi.org/http://dx.doi.org/10.5935/abc.20160128
  • 5
    Peura JL, Colvin-Adams M, Francis GS, et al. Recommendations for the use of mechanical circulatory support: Device strategies and patient selection: a scientific statement from the American Heart Association. Circulation. 2012;126(22):2648-67. PMID: 23109468; https://doi.org/10.1161/CIR.0b013e3182769a54.
    » https://doi.org/https://doi.org/10.1161/CIR.0b013e3182769a54
  • 6
    Fiorelli AI, Oliveira Júnior JL, Coelho GHB, Rocha DC. Assistência circulatória mecânica: porquê e quando [Mechanical circulatory support: why and when]. Rev Med (São Paulo). 2008;87(1):1-15. Available from: Available from: https://www.revistas.usp.br/revistadc/article/view/59055/62042 Accessed in 2021 (Feb 18).
    » https://www.revistas.usp.br/revistadc/article/view/59055/62042
  • 7
    Kirklin JK, Naftel DC, Kormos RL, et al. Third INTERMACS Annual Report: the evolution of destination therapy in the United States. J Heart Lung Transplant. 2011;30(2):115-23. PMID: 21215378; https://doi.org/10.1016/j.healun.2010.12.001.
    » https://doi.org/https://doi.org/10.1016/j.healun.2010.12.001
  • 1
    Hospital Sírio Libanês, São Paulo (SP), Brazil
  • Sources of funding: None

Publication Dates

  • Publication in this collection
    10 May 2021
  • Date of issue
    Jul-Aug 2021

History

  • Reviewed
    21 Dec 2020
  • Received
    21 Dec 2020
  • Accepted
    18 Feb 2021
Associação Paulista de Medicina - APM APM / Publicações Científicas, Av. Brigadeiro Luís Antonio, 278 - 7º and., 01318-901 São Paulo SP - Brazil, Tel.: +55 11 3188-4310 / 3188-4311, Fax: +55 11 3188-4255 - São Paulo - SP - Brazil
E-mail: revistas@apm.org.br