Technical note: ultrasonography evaluation of aortic aneurysms treated with endoprosthesis

Abstracts

Aortic aneurysms treated with endovascular techniques may develop endoleaks. In addition to anatomic location, these endoleaks should be classified based on the type of blood flow. We describe the types of blood flow into the endoleaks as detected by Doppler ultrasonography: in-and-out flow as a "pseudoaneurysm", channel flow through the aneurysm, and direct flow into the aneurysm. Endoleaks may be intermittent and actually be endoseepage, detected by aneurysmal growth without an apparent inflow. Ultrasound virtual histology is suggested to detect endoseepage. Relative quantity and location of blood/liquid in the aneurysm can be detected by evaluating pixel brightness of the ultrasonographic B-mode image. Such anatomofunctional information allows for quantitative follow-up and emphasizes the risk of an endoleak with direct flow into the aneurysm.

aortic aneurysm; endovascular procedures; endoleak; ultrasonography


Aneurismas da aorta tratados com endopróteses podem desenvolver endoleaks. Esses endoleaks devem ser classificados com base nos tipos de fluxo, além da localização anatômica. Enfatizamos tipos de fluxo dos endoleaks detectados pela ultrassonografia com Doppler: fluxo entra-e-sai como "pseudoaneurisma", fluxo de canal e fluxo direto. Ao se notar a possibilidade de endoleaks intermitentes, salientamos o conceito de endoseepage detectáveis por crescimento de dimensões sem fluxo entrante aparente. Sugerimos o uso da histologia virtual ultrassonográfica para detecção de infiltrações. Quantidade relativa e localização de sangue ou líquido dentro do saco aneurismático pode ser detectado pela avaliação dos níveis de brilho da ultrassonografia modo B. Essa informação permite um seguimento compreensivo do aneurisma tratado com endoprótese e salienta o risco maior deum endoleak de fluxo direto entrante no aneurisma.

aneurisma aórtico; procedimentos endovasculares; endoleak; ultrassonografia


INNOVATION

Technical note: ultrasonography evaluation of aortic aneurysms treated with endoprosthesis* * All authors have read and approved the final version submitted to J Vasc Bras. Study carried out at the Angiolab Curitiba - Curitiba (PR), Brazil.

Sergio Xavier Salles CunhaI

I Consultant of non-invasive techniques for vascular diagnosis at Angiolab Curitiba - Curitiba (PR), Brazil; e no Angiloab Vitória - Vitória (ES), Brazil.

Correspondence to Sergio Xavier Salles CunhaRua Ernesto Zwarg, 215 - SatéliteCEP 11740-000 - Itanhaém (SP), BrasilE-mail: sallescunha@yahoo.com

ABSTRACT

Aortic aneurysms treated with endovascular techniques may develop endoleaks. In addition to anatomic location, these endoleaks should be classified based on the type of blood flow. We describe the types of blood flow into the endoleaks as detected by Doppler ultrasonography: in-and-out flow as a "pseudoaneurysm", channel flow through the aneurysm, and direct flow into the aneurysm. Endoleaks may be intermittent and actually be endoseepage, detected by aneurysmal growth without an apparent inflow. Ultrasound virtual histology is suggested to detect endoseepage. Relative quantity and location of blood/liquid in the aneurysm can be detected by evaluating pixel brightness of the ultrasonographic B-mode image. Such anatomofunctional information allows for quantitative follow-up and emphasizes the risk of an endoleak with direct flow into the aneurysm.

Keywords: aortic aneurysm; endovascular procedures; endoleak; ultrasonography.

Introduction

Aortic aneurysms treated with endoprosthesis may develop endoleaks1,2. An endoleak is a leak of blood flow, or prosthesis, or a branch of the aorta into the aneurysm. They have been classified according to their anatomical location3. This technical note emphasizes the hemodynamics of endoleaks' flow, describes the concepts of endoseepage and intermittent endoleak, and summarizes the technique of intravascular ultrasound virtual histology (IVUS-VH) for the detection of blood or fluids, acute, subacute or fibrotic thrombi inside the aneurysm.

Endoleak flow types

Endoleaks have been identified with at least three types of flow:

  1. similar to that of a pseudoaneurysm;

  2. channel flow through the aneurysm;

  3. direct flow into the aneurysm.

Pseudoaneurysm flow

Figure 1 shows the flow of an endoleak into a pseudoaneurysm inside an aortic aneurysm treated with endoprosthesis. The flow into the pseudoaneurysm is characterized as in- and outflow. This type of endoleak does not seem to contribute with increase in the diameter or volume of the treated aneurysm. Occasionally, the pseudoaneurysm may thrombose and occlude. However, thrombosis of the pseudoaneurysm does not guarantee the closure of the endoleak. The flow can change direction, creating a new pseudoaneurysm in other area of the aneurysm. Periodic monitoring of this type of endoleak is recommended, even if the first pseudoaneurysm has been closed. This type of endoleak may be intermittent.


Channel flow

Figure 2 shows a blood flow channel inside the aortic aneurysm treated with endoprosthesis. This channel may have varied sources of blood flow, such as the endoprosthesis itself, or a branch of the aorta or the iliac arteries. The drainage of this channel also varies, but it is generally found at a branch of the aneurysm. This channel is apparently restricted by a chronic thrombosed tissue, that is, a blood thrombus in various stages of fibrosis. The blood leakage from the channel into the aneurysm depends on the type of tissue around this false lumen. Apparently, this type of endoleak does not dramatically affect the diameter or volume of the treated aneurysm. However, it is this type of endoleak is thought to be more dangerous than the endoleak similar to that of a pseudoaneurysm. A closer follow-up is recommended in such cases.


Direct flow

Figure 3 shows a direct flow of endoleak into the aneurysm sac. This type of endoleak is classified as high-risk. The volume of blood drained into the aneurysm may quickly rupture the aneurysm wall. This aneurysm ruptured two hours after the ultrasound and, as a consequence, the 93-year-old patient, who was not candidate for revascularization, died. The vascular surgeon should be informed immediately in case of endoleak direct flow shown at ultrasonography.


Endoseepage

Diameter measurements illustrate cases of endoseepage. The continuing growth of 2 cm in diameter per year of an aneurysm is equivalent to 0.05 mm per day or 0.002 mm per hour. It is not well established whether endoseepage can be identified by any modern flow measurement technique. The concept of endoseepage was then created for aneurysms that grow in size without an apparent source of blood flow.

Intermittent endoleak

Some studies suggest that Doppler flow ultrasound is more accurate than radiographic techniques used to detect endoleaks, while other studies show the other way around4-6. In general, ultrasonography has detected more endoleaks than computed tomography scan; the low positive predictive value of ultrasonography can be analyzed as CT scan false-negative result.

Another explanation for these discrepancies is that these techniques were not performed simultaneously, and that the endoleak could be intermittent. An intermittent endoleak could explain the contradictory results, the change in direction of the pseudoaneurysm-like endoleaks in follow-up, or the growth of the aneurysm dimensions without an apparent source of endoleak. Monitoring of the dimensions, diameter and even of the volume parallel to the flow is recommended.

Ultrasound-virtual histology

Figure 4 shows two Ultrasound-virtual histology images of aneurysms treated with endoprosthesis. This technique has artificial color brightness levels of the B-mode ultrasound imaging7-10. The image shows regions with echoes similar to blood, lipids, muscle, fiber and calcium. In the case of aortic aneurysms, it would be interpreted as blood or fluid, recent thrombus or thrombi in various stages of fibrosis. Based on intravenous studies, the acute thrombus would generate images similar to lipids (green stain in Figure 4); the subacute thrombus would tend to appear in bright blue8,9.


The virtual histology technique creates a range of brightness in the B-mode ultrasound imaging. Usually, these images have 256 levels of brightness. In these applications, the blood (or what appears in black) is zero in the scale. The adventitia (or what is in white at the prosthesis) is 200 in the scale. Blood, lipids, muscle, fiber, and calcium have echoes in the intervals 0-4, 8-26, 41-76, 112-196 e 211-2557.

Evaluation of endoleaks, endoseepage and dimensions is recommended in periodic follow-up of aneurysms with hypoechoic regions defined by B-mode virtual histology imaging.

Conclusions

Simple analysis of ultrasound B-mode imaging or the complex processing of the image using the technique of Ultrasound-virtual histology can identify hypoechoic regions inside the aneurysm treated with endoprosthesis. Duplex and Doppler ultrasound imaging can show the location and type of endoleak flow. The results of these imaging techniques can be used to determine the follow-up of the aneurysm or the need for emergency care of an aneurysm with direct-flow endoleak.

References

Correspondence

Submitted on: 09.06.11.

Accepted on: 02.23.12.

inancial support: none.

Conflict of interest: nothing to declare.

Authors' contributions

Conception and design: SXSC

Analysis and interpretation: SXSC

Data collection: SXSC

Writing the article: SXSC

Critical revision of the article: SXSC

Final approval of the article*: SXSC

Statistical analysis: SXSC

Overall responsibility: SXSC

  • 1. Silvestre JMS, Motta F, Sardinha WE, et al. Tratamento endovascular do aneurisma da aorta abdominal infrarrenal em pacientes com anatomia favorável para o procedimento - experiência inicial em um serviço universitário. J Vas Bras. 2011;10(1):31-9
  • 2. Mendonça CT, Moreira RCR, Timi JRR, et al. Comparação entre os tratamentos aberto e endovascular dos aneurismas da aorta abdominal em pacientes de alto risco cirúrgico. J Vasc Bras. 2005;4(3):232-42.
  • 3. Cao P, De Rango P, Verzini F, et al. Endoleak after endovascular aortic repair: classification, diagnosis and management following endovascular thoracic and abdominal aortic repair. J Cardiovasc Surg (Torino). 2010;51(1):53-69.
  • 4. Schmieder GC, Stout CL, Stokes GK, et al. Endoleak after endovascular aneurysm repair: duplex ultrasound imaging is better than computed tomography at determining the need for intervention. J Vasc Surg. 2009;50(5):1012-7.
  • 5. Raman KG, Missig-Carroll N, Richardson T, et al. Color-flow duplex ultrasound scan versus computed tomographic scan in the surveillance of endovascular aneurysm repair. J Vasc Surg. 2003;38(4):645-51.
  • 6. Manning BJ, O'Neill SM, Haider SN, et al. Duplex ultrasound in aneurysm surveillance following endovascular aneurysm repair: a comparison with computed tomography aortography. J Vasc Surg. 2009;49(1):60-5.
  • 7. Lal BK, Hobson RW 2nd, Pappas PJ, et al. Pixel distribution analysis of B-mode ultrasound scan images predicts histologic features of atherosclerotic carotid plaques. J Vasc Surg. 2002;35(6):1210-7.
  • 8. Cassou-Birckholz MF, Engelhorn CA, Salles-Cunha SX, et al. Assessment of deep venous thrombosis by grayscale median analysis of ultrasound images. Ultrasound Q. 2011;27(1):55-61.
  • 9. Menezes FH, Silveira SAF, Salles-Cunha SX. Pixel characterization for development of ultrasound-based virtual histology of deep venous thrombosis. Apresentado no Congresso Anual da Society of Vascular Ultrasound (SVU), Chicago, IL, 15-18 de Junho, 2011, e-published na svunet.org (members only), Presentation Abstract 109, p. 3
  • 10. Engelhorn AL, Engelhorn CA, Salles-Cunha SX. Initial evaluation of virtual histology ultrasonographic techniques applied to a case of renal transplant. Apresentado no Congresso Anual da Society of Vascular Ultrasound (SVU), Chicago, IL, USA, 15-18 de Junho, 2011, e-published na svunet.org (members only) Poster Abstract 412, p.20
  • Sergio Xavier Salles Cunha
    Rua Ernesto Zwarg, 215 - Satélite
    CEP 11740-000 - Itanhaém (SP), Brasil
    E-mail:
  • *
    All authors have read and approved the final version submitted to J Vasc Bras.
    Study carried out at the Angiolab Curitiba - Curitiba (PR), Brazil.
  • Publication Dates

    • Publication in this collection
      17 July 2012
    • Date of issue
      June 2012

    History

    • Received
      06 Sept 2011
    • Accepted
      23 Feb 2012
    Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV) Rua Estela, 515, bloco E, conj. 21, Vila Mariana, CEP04011-002 - São Paulo, SP, Tel.: (11) 5084.3482 / 5084.2853 - Porto Alegre - RS - Brazil
    E-mail: secretaria@sbacv.org.br