Endotension: rupture of abdominal aortic aneurysm

Endotensão: ruptura de aneurisma de aorta abdominal

Abstracts

Aortic endovascular exclusion technique called 'chimney' consists of placing stents through abdominal aortic visceral branches and a prosthesis that excludes the thoraco-abdominal aneurysm. Stents and an aortic endoprosthesis are placed in the renal arteries. This method is primarily used when open surgery is too risky. The mechanism that provides aneurysm sac increase without the visible presence of endoleaks has not been fully elucidated. The expansion of the aneurysm sac, due to endotension, is difficult to diagnose, even with the use of advanced imaging tests. Its diagnosis is made by exclusion. We present a case of a late complication in a high-risk patient after a 'chimney' endovascular procedure. Following the surgery, the patient presented a ruptured aneurysm sac without a visible endoleak. A second intervention was not feasible due to the high risk of occluding all of the branches, and complicated by previous 'chimney'. Endotension is a possible cause of aneurysm rupture and death.

aortic rupture; endoleak; postoperative complications


A técnica de exclusão endovascular conhecida como 'chaminé' consiste na colocação de stent em ramos viscerais e de endoprótese excluindo o aneurisma toracoabdominal. São colocados stents revestidos nas artérias renais e uma endoprótese aórtica, que o método utilizado quando a cirurgia aberta tem risco muito alto. O mecanismo que causa a expansão aneurismática sem a presença detectável de vazamento pelos métodos de imagem não está completamente esclarecido. A expansão do saco aneurismático por endotensão é de difícil diagnóstico, mesmo com o uso de técnicas de imagem avançadas, como tomografia computadorizada e eco-Doppler, sendo o diagnóstico por exclusão. Apresenta-se um caso de complicação tardia após o tratamento endovascular pela técnica da 'chaminé'. Após a cirurgia, o paciente apresentou ruptura sem endoleak visível. Outro procedimento endovascular foi impossibilitado pela técnica da 'chaminé', que dificulta novos procedimentos e há alto risco de oclusão dos ramos. Endotensão é causa de ruptura e óbito.

ruptura aórtica; endoleak; complicações pós-operatórias


CASE REPORT

Endotension: rupture of abdominal aortic aneurysm

Endotensão: ruptura de aneurisma de aorta abdominal

Alexandre Campos Moraes AmatoI; Flávio Amim AbrahamII; Henrique Dini KraideII; Leandro Teixeira RochaII; Ricardo Virginio dos SantosI

IProfessor of the Vascular Surgery Discipline at Unisa - São Paulo (SP), Brazil

IIMedical students attending the sixth year at Unisa - São Paulo (SP), Brazil

Correspondence Correspondence: Alexandre Campos Moraes Amato Avenida Juriti, 144 CEP 05612-010 - São Paulo (SP), Brazil E-mail: dr.alexandre@amato.com.br

ABSTRACT

Aortic endovascular exclusion technique called 'chimney' consists of placing stents through abdominal aortic visceral branches and a prosthesis that excludes the thoraco-abdominal aneurysm. Stents and an aortic endoprosthesis are placed in the renal arteries. This method is primarily used when open surgery is too risky. The mechanism that provides aneurysm sac increase without the visible presence of endoleaks has not been fully elucidated. The expansion of the aneurysm sac, due to endotension, is difficult to diagnose, even with the use of advanced imaging tests. Its diagnosis is made by exclusion. We present a case of a late complication in a high-risk patient after a 'chimney' endovascular procedure. Following the surgery, the patient presented a ruptured aneurysm sac without a visible endoleak. A second intervention was not feasible due to the high risk of occluding all of the branches, and complicated by previous 'chimney'. Endotension is a possible cause of aneurysm rupture and death.

Keywords: aortic rupture; endoleak; postoperative complications.

RESUMO

A técnica de exclusão endovascular conhecida como 'chaminé' consiste na colocação de stent em ramos viscerais e de endoprótese excluindo o aneurisma toracoabdominal. São colocados stents revestidos nas artérias renais e uma endoprótese aórtica, que o método utilizado quando a cirurgia aberta tem risco muito alto. O mecanismo que causa a expansão aneurismática sem a presença detectável de vazamento pelos métodos de imagem não está completamente esclarecido. A expansão do saco aneurismático por endotensão é de difícil diagnóstico, mesmo com o uso de técnicas de imagem avançadas, como tomografia computadorizada e eco-Doppler, sendo o diagnóstico por exclusão. Apresenta-se um caso de complicação tardia após o tratamento endovascular pela técnica da 'chaminé'. Após a cirurgia, o paciente apresentou ruptura sem endoleak visível. Outro procedimento endovascular foi impossibilitado pela técnica da 'chaminé', que dificulta novos procedimentos e há alto risco de oclusão dos ramos. Endotensão é causa de ruptura e óbito.

Palavras-chave: ruptura aórtica; endoleak; complicações pós-operatórias.

Introduction

With the development of endovascular surgery, a new treatment option for high-risk patients who require conventional aneurysm surgery, but are not suited for standard endovascular procedures, is being considered. The new technique (called 'chimney') is the percutaneous transluminal placement of a stent through the aortic visceral and renal branches1. A prosthesis is placed to exclude the thoraco-abdominal aneurysm. This method is mainly used when the aneurysm affects the renal and visceral vessels, and open surgery is considered too risky and not feasible.

One of the complications of an endovascular aneurysm repair (EVAR) is the endoleak, which maintains the flow within the excluded aneurysm sac and pressurises it. These leaks are divided into five kinds, according to their mechanisms and locations. The type-V endoleak, also called endotension, is defined by the persistent or recurrent pressurisation of the aneurysm sac through the parietal thrombi. What differentiates an endoleak from endotension is that, in the latter, pressure is transmitted without the presence of blood flow within the aneurysm2-4.

The prevalence of endoleaks after an EVAR is highly variable. In the literature, it has been reported to be between 2.4 and 45.5%5. The clinical importance of endoleaks is directly related to an increased risk of aneurysm rupture3, although endotension-causing ruptures have rarely been reported3,6.

Case report

The patient was a 66-year-old male, who was hypertensive, obese, with chronic obstructive pulmonary disease and dyslipidaemic. A computed tomography scan was conducted in October, 2006, and a type-IV thoraco-abdominal aortic aneurysm was diagnosed. Since then, the aneurysm has showed continuous growth in subsequent bi-annual examinations without visible endoleak.

In July, 2008, a computed tomography angiography (CTA) revealed that the aneurysm dilatation in the distal abdominal aorta was juxtarenal and extended up to the bifurcation. The centre of the predominating concentric parietal thrombus had a maximum proximal diameter of 5.2 x 4.1 cm. The CTA presented a lower calibre in the left renal artery, with an ipsilateral renal reduction (Figure 1).


The patient underwent an endovascular aneurysm exclusion (Powerlink bifurcated stent, Endologix, Inc., Irvine, CA, the USA) with an endovascular revascularization of the renal arteries, celiac trunk, and superior mesenteric artery by the chimney technique (Figure 2).


Two years later, the patient presented with increasing lower abdominal pain, which radiated to the lumbar region and had evolved over three weeks. The patient was conscious and eupneic and showed no signs of shock. The abdomen was slightly distended, produced bowel sounds, and was somewhat painful to palpation. The posterior tibial pulse was bilaterally present, and the left pedal pulse was absent.

The patient was admitted into the Intensive Care Unit (ICU), where he was held under tight blood-pressure control, and a CTA was performed, as an endoleak was suspected from previous procedure. Aneurysm growth was observed (Figure 3).


The aneurysm had a maximum diameter of approximately 13x12 cm. The dilation and tortuosity of the entire abdominal aorta reflected the largest diameter in its infra-renal portion. In the distal region on the same side and near the bifurcation of the iliac artery, irregular areas (suggestive of ulcers) were observed. A ruptured aneurysm wall without a contrast leak was observed on the left anterior side.

Conservative medical management and observation were performed, as the patient was deemed to have a high anaesthetic risk given his age and comorbidities. Angiography was requested to confirm the absence of endoleaks2. Therefore, the patient was kept in the ICU with a normo/hypotensive blood pressure and rigorous clinical support. He worsened clinically with no signs of initial hypovolemic shock. However, he died three days after his hospital admission, due to sudden hypovolemic shock. Angiography was not performed for it required moving the patient to another facility2.

Discussion

The chimney technique essentially consists of deploying a covered stent parallel to the main aortic stent-graft. It protrudes similarly to a chimney and preserves the flow of the vital side branches, which are covered by the aortic stent graft. The chimney technique enables the use of a standard off-the-shelf endoprosthesis to treat a lesion, with an inadequate fixation zone and it provides an alternative to a fenestrated stent graft in urgent cases and in aneurysms with challenging neck morphologies. This technique also enables the rebuilding of the side branches of the aorta that might have been involuntarily covered during the endovascular repair1,7.

Several theories have been proposed in an attempt to elucidate the mechanisms of expansion of the aneurysm sac, without the presence of a detectable endoleak. These include increased porosity of the prosthesis5, direct transmission of pressure from the stent to the lumen of the aneurysm sac4,8, and a low flow endoleak which is undetectable by the imaging methods2,9. Mennander et al. found five patients (3.1%) with endotension within 160 patients who underwent endovascular aneurysm repair, and endoleaks were not detected. Three suffered aneurysm sac ruptures during the study period, but none of them presented clinical evidence of major bleeding9.

The best explanation for these cases may be the association of more than one of the aforementioned theories. Of these, the exudate is the most accepted one that may explain the cases of endotension. However, in this case, it cannot be exclusively attributed to the exudate, since there are no studies demonstrating the effectiveness of the chimney seal between the endoprosthesis (Figure 2).

The continuous sweating through the endoprosthesis may result in the rupture of the sac without any evident hemodynamic changes. But also, the sac rupture may cause endoprosthesis migration due to sudden morphologic changes of the aneurismal sac and lack of satisfactory anchoring of the stent-graft. Patients with aneurysm sac enlargements after EVAR may be difficult to diagnose. Therefore, the first step is to check for the presence of endoleaks. Multi-slice CT, high-resolution Doppler echocardiography, and angiography were the most widely used imaging modalities at the beginning of the investigation. Endotension has also been measured by indirect methods using arterial punctures6,10.

In the present case, the endoprosthesis expanded polytetrafluoroethylene (ePTFE) graft material had low porosity. We cannot exclude the possibility that the increased pressure in the aneurysm sac was developed due to the passage of exudates through the prosthesis material or the sealing difficulty associated with the chimney technique.

Kougias et al. have recently reported a less-invasive technique, which uses stent-based reinforcement with the deployment of new graft components to re-line the endograft and to minimize the porosity-related endotension. This technique is applicable to patients with a porous endoprosthesis6. However, it does not apply to this case because stent reinforcement would increase the main radial graft strength and possibly occlude the parallel visceral branches.

Conclusions

The development of aneurysm sac expansion due to endotension is difficult to diagnose, even with the use of advanced imaging tests, such as multi-slice CT, Doppler echocardiography, and angiography, as they may not be sufficiently sensitive to detect small endoleaks. By definition in endotension, there must be no leaks. This technical limitation of imaging may falsely categorize a patient as having endotension when the actual reason for the expansion is a very small, but persistent, endoleak. The chimney technique may be associated with an apparently technically successful procedure because images do not show any endoleaks between stents. There may be small endoleaks that are responsible for future sac expansion and rupture.

Endotension diagnosis is currently made by exclusion and therefore should encourage the development of new techniques, which would allow the detection of small pressure changes in the aneurysm sac. Aneurysm sac pressure measured by an implantable remote sensor may help to elucidate endotension causes11.

The chimney technique is not a standard procedure and should only be performed in few selected cases, because it can compromise the ability to perform further procedures to treat possible complications. The sealing of the chimney technique is not guaranteed, as our current imaging techniques are not sufficiently sensitive to detect small endoleaks2. Even without any detectable endoleaks, and the aneurysm sac may grow due to endotension or undetectable endoleaks, this aneurysm sac may rupture and cause death. In the event of an aneurysm sac rupture, after a chimney procedure, no endovascular procedure is safe, and open surgery that was once avoided now has a higher risk and is characterized by even greater technical difficulties3,5,8-10.

Submitted on: 04.15.11.

Accepted on: 01.11.12.

Financial support: none.

Conflict of interest: nothing to declare.

Author's contributions

Conception and design: ACMA, RVS, FAA, HDK, LTR

Analysis and interpretation: ACMA, RVS

Data collection: ACMA, RVS, FAA, HDK, LTR

Writing the article: ACMA, FAA, HDK, LTR

Critical revision of the article: ACMA, RVS

Final approval of the article*: ACMA, RVS, FAA, HDK, LTR

Statistical analysis: N/A

Overall responsibility: ACMA

*All authors have read and approved the final version submitted to J Vasc Bras.

Study carried out at the Vascular Surgery Discipline at the Santo Amaro University (Unisa) - Santo Amaro (SP), Brazil.

  • 1. Ohrlander T, Sonesson B, Ivancev K, et al. The chimney graft: A technique for preserving or rescuing aortic branch vessels in stent-graft sealing zones. J Endovasc Ther. 2008;15:427-32.
  • 2. Chagas Neto FAD, Barreto ARF, Reis HFD, et al. A importância do diagnóstico por imagem na classificação dos endoleaks como complicação do tratamento endovascular de aneurismas aórticos. Radiol Bras. 2010;43:289-94.
  • 3. Görich J, Rilinger N, Sokiranski R, et al. Leakages after endovascular repair of aortic aneurysms: classification based on findings at CT, angiography, and radiography. Radiology. 1999;213:767-72.
  • 4. White GH. What are the causes of endotension? J Endovasc Ther. 2001;8:454-6.
  • 5. Williams GM. The management of massive ultrafiltration distending the aneurysm sac after abdominal aortic aneurysm repair with a polytetrafluoroethylene aortobiiliac graft. J Vasc Surg. 1998;28:551-5.
  • 6. Kougias P, Lin PH, Dardik A, et al. Successful treatment of endotension and aneurysm sac enlargement with endovascular stent graft reinforcement. J Vasc Surg. 2007;46:124-7.
  • 7. Schlösser FJ, Gusberg RJ, Dardik A, et al. Aneurysm rupture after EVAR: can the ultimate failure be predicted? Eur J Vasc Endovasc Surg. 2009;37:15-22.
  • 8. Meier GH, Parker FM, Godziachvili V, et al. Endotension after endovascular aneurysm repair: The ancure experience. J Vasc Surg. 2001;34:421-6.
  • 9. Mennander A, Pimenoff G, Heikkinen M, et al. Nonoperative approach to endotension. J Vasc Surg. 2005;42:194-9.
  • 10. Risberg B, Delle M, Lönn L, et al. Management of aneurysm sac hygroma. J Endovasc Ther. 2004;11:191-5.
  • 11. Hoppe H, Segall JA, Liem TK et al. Aortic aneurysm sac pressure measurements after endovascular repair using an implantable remote sensor: Initial experience and short-term follow-up. Eur Radiol. 2008;18:957-65.
  • Correspondence:

    Alexandre Campos Moraes Amato
    Avenida Juriti, 144
    CEP 05612-010 - São Paulo (SP), Brazil
    E-mail:
  • Publication Dates

    • Publication in this collection
      17 July 2012
    • Date of issue
      June 2012

    History

    • Received
      15 Apr 2011
    • Accepted
      01 Nov 2012
    Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV) Rua Estela, 515, bloco E, conj. 21, Vila Mariana, CEP04011-002 - São Paulo, SP, Tel.: (11) 5084.3482 / 5084.2853 - Porto Alegre - RS - Brazil
    E-mail: secretaria@sbacv.org.br