Acessibilidade / Reportar erro

Hepatic vein migration of a totally implantable venous access port-a-cath for chemotherapy in a breast carcinoma patient: case report

Abstract

A totally implantable venous access port (TIVAP) is used for chemotherapy administration. Venous port migration to the systemic circulation occurs in less than 1% of complications. The aim of this study is to describe a case of TIVAP migration to the hepatic vein. A 44-year-old female patient with breast cancer was prescribed neoadjuvant chemotherapy. A port-a-cath was surgically implanted for chemotherapy. During the port puncture procedure, blood returned normally when aspirated. When the port was first accessed and flushed with saline solution, swelling was observed at the port site and blood could no longer be aspirated. A chest radiography showed catheter embolization in the region of the hepatic vein. The catheter was retrieved using a snare technique (without complications) and the patient was discharged the next day. The care team should be alert to possible TIIVAP malfunction.

Keywords:
complications; catheters; neoadjuvant chemotherapy; hepatic veins; breast neoplasms; case reports

Resumo

O cateter totalmente implantável (CTI) é utilizado na administração da quimioterapia. Em menos de 1% dos casos de complicação, pode ocorrer migração do CTI para quimioterapia para a circulação sistêmica. O objetivo deste estudo foi descrever um caso de migração do CTI para a veia hepática. Uma paciente do sexo feminino, de 44 anos de idade, teve diagnóstico de câncer de mama com indicação de quimioterapia neoadjuvante. Realizou-se a implantação de cateter port-a-cath. Durante o procedimento de punção do cateter, houve retorno normal de sangue, e foi realizada infusão de soro fisiológico. Em seguida, houve um aumento de volume no local do port e não retorno de sangue à aspiração. A radiografia de tórax mostrou embolização do cateter em topografia hepática. Retirou-se o cateter pela técnica do laço (sem complicações), e a paciente recebeu alta no dia seguinte. Possíveis alterações no funcionamento do CTI devem chamar atenção da equipe responsável.

Palavras-chave:
complicações; cateteres; quimioterapia neoadjuvante; veias hepáticas; neoplasias da mama; relatos de caso

INTRODUCTION

Totally implantable venous access ports (TIVAP) are commonly indicated for patients who need long-duration chemotherapy for cancer treatment. Migration of a chemotherapy TIVAP catheter into the systemic circulation is a rare event, accounting for less than 1% of all complications related to these devices. Embolization of fragments may occur in the atrium, ventricle, pulmonary artery, hepatic veins, or vena cava, among other sites.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...
,22 Fonseca IY, Krutman M, Nishinari K, et al. Brachial insertion of fully implantable venous catheters for chemotherapy: complications and quality of life assessment in 35 patients. Einstein (Sao Paulo). 2016;14(4):473-9. http://dx.doi.org/10.1590/s1679-45082016ao3606. PMid:28076593.
http://dx.doi.org/10.1590/s1679-45082016...

The objective of this study is to describe a case of migration of a chemotherapy TIVAP catheter to the hepatic vein.

CASE DESCRIPTION

A 44-year-old female patient was referred to the service with a breast nodule. Results of anatomopathological and immunohistochemical tests of a biopsy of the nodule indicated that it was an invasive breast carcinoma of no special type (IBCNST), G2, nuclear grade 2, with estrogen receptors (ER)+: 80%, progesterone receptors (PR)+: 40%, human epidermal growth factor receptor 2 (HER2) negative: 0-1, and Ki67: 30%.

The patient was prescribed neoadjuvant chemotherapy, and a port-a-cath catheter was placed by phlebotomy with access via the right cephalic vein. The position of the catheter in the superior vena cava was confirmed intraoperatively with radioscopy and the catheter was heparinized. The procedure was uneventful, with no complications, and the patient was discharged with no complaints. When the time came to administer the first chemotherapy cycle, the port was accessed and blood returned as normal, proceeding to infusion of saline, at which point swelling was observed at the port site. The chemotherapy procedure was aborted and chest X-rays (Figures 1A and 1B) showed embolization of the catheter in the region of the hepatic vein. The patient did not have any symptoms.

Figure 1
(A) and (B) Chest X-ray showing migration of the chemotherapy totally implantable venous access catheter to the hepatic vein. Explanatory Note: Arrows indicate that totally implantable venous catheter for chemotherapy migrated to hepatic vein.

The patient was admitted and transported to the surgical center for removal of the catheter using a loop snare technique, which was conducted with percutaneous venous access obtained via the left femoral vein. An introducer was inserted to enable manipulation of the materials needed. A catheter with the appropriate curvature to reach the site of the foreign body was advanced over a guidewire and placed adjacent to one of the extremities of the migrated catheter. After the guidewire had been placed, the first catheter was removed and a foreign body retrieval snare catheter was inserted over the same guidewire, positioning it exactly at the extremity of the migrated catheter. At this point, the intravenous foreign body (TIVAP) was captured with the snare (Figure 2A) and withdrawn via the introducer (Figure 2B).33 Motta Leal JM Fo, Carnevale FC, Nasser F, et al. Endovascular techniques and procedures, methods for removal of intravascular foreign bodies. Rev Bras Cir Cardiovasc. 2010;25(2):202-8. PMid:20802912. The procedure was conducted with no complications and the patient was discharged with no intercurrent conditions on the first postoperative day. The following month, she underwent placement of a new port-a-cath to proceed with neoadjuvant chemotherapy (Figure 3). This procedure was conducted with no complications. Currently, at around 25-months’ follow-up (December 31, 2019 to January 28, 2022), the patient is in good health and there is no evidence of any complications resulting from migration of the TIVAP or from the retrieval procedure.

Figure 2
Illustration of the snare technique employed to retrieve the chemotherapy totally implantable venous access catheter migrated to the hepatic vein (using access via the left femoral vein). (A) Capturing the catheter; (B) Catheter removed.33 Motta Leal JM Fo, Carnevale FC, Nasser F, et al. Endovascular techniques and procedures, methods for removal of intravascular foreign bodies. Rev Bras Cir Cardiovasc. 2010;25(2):202-8. PMid:20802912.
Figure 3
Chest X-ray taken after the procedure to remove the chemotherapy totally implantable venous access catheter that had migrated to the hepatic vein, showing the new port-a-cath fitted without complications. Explanatory Note: Arrows indicate that the new venous access port was implanted without complications; The posteroanterior (PA) chest view refers to the X-ray beams passing through the patient from posterior to anterior, with better evaluation of the structures closer to the ventral region.

This study is part of a larger project with cancer patients seen at a private oncology service in the town of Teresina (Piauí [PI]), Brazil. The project was approved by the research ethics committee at the Universidade Estadual do Piauí, Teresina (PI), Brazil – decision number: 4.311.835 (CAAE: 30154720.0.0000.5209). The study complies with all applicable Brazilian (National Health Council resolution 466/12) and international research ethical principles. The patient signed a free and informed consent form.

DISCUSSION

The TIVAP is a long-stay device that facilitates infusion of medications, solutions, and chemotherapy agents, blood transfusions, and blood draws for tests. One of its most common indications is for cancer patients who require long duration chemotherapy.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...
,44 Danski MTR, de Amorim SD, de Oliveira GLR, Pedrolo E, da Silva SR, Greboge TML. Infectious complications associated with the totally implantable central venous catheter. J Nurs UFPE on line. 2017;11(12):5049-58. https://doi.org/10.5205/1981-8963-v11i12a25104p5049-5058-2017.
https://doi.org/10.5205/1981-8963-v11i12...
This type of catheter reduces the risk of leakage of irritant chemotherapy substances, offers continuous access, and avoids the discomfort of multiple vein punctures.44 Danski MTR, de Amorim SD, de Oliveira GLR, Pedrolo E, da Silva SR, Greboge TML. Infectious complications associated with the totally implantable central venous catheter. J Nurs UFPE on line. 2017;11(12):5049-58. https://doi.org/10.5205/1981-8963-v11i12a25104p5049-5058-2017.
https://doi.org/10.5205/1981-8963-v11i12...
,55 Cakır F, Geze S, Ozturk MH, Dınc H. Percutaneous endovascular removal of intracardiac migrated port A catheter in a child with acute lymphoblastic leukemia. Braz J Anesthesiol. 2014;64(4):275-7. http://dx.doi.org/10.1016/j.bjane.2012.11.001.
http://dx.doi.org/10.1016/j.bjane.2012.1...

It is important to emphasize that, despite the rarity of their occurrence, using a TIVAP involves the potential for complications, such as infection, embolization, catheter occlusion, venous perforation, atrial perforation, arrhythmia, phlebitis, leakage, migration, and catheter fracture, among others.55 Cakır F, Geze S, Ozturk MH, Dınc H. Percutaneous endovascular removal of intracardiac migrated port A catheter in a child with acute lymphoblastic leukemia. Braz J Anesthesiol. 2014;64(4):275-7. http://dx.doi.org/10.1016/j.bjane.2012.11.001.
http://dx.doi.org/10.1016/j.bjane.2012.1...

6 Xu H, Chen R, Jiang C, et al. Implanting totally implantable venous access ports in the upper arm is feasible and safe for patients with early breast cancer. J Vasc Access. 2020;21(5):609-14. http://dx.doi.org/10.1177/1129729819894461. PMid:31841061.
http://dx.doi.org/10.1177/11297298198944...

7 Skelton WP 4th, Franke AJ, Welniak S, et al. Investigation of complications following port insertion in a cancer patient population: a retrospective analysis. Clin Med Insights Oncol. 2019;13:1179554919844770. http://dx.doi.org/10.1177/1179554919844770. PMid:31040735.
http://dx.doi.org/10.1177/11795549198447...
-88 Tumay LV, Guner OS. Availability of totally implantable venous access devices in cancer patients is high in the long term: a seven-year follow-up study. Support Care Cancer. 2021;29(7):3531-8. http://dx.doi.org/10.1007/s00520-020-05871-6. PMid:33155092.
http://dx.doi.org/10.1007/s00520-020-058...
These complications can be classified as early complications (within 7 days of implantation) or late complications (after this period).44 Danski MTR, de Amorim SD, de Oliveira GLR, Pedrolo E, da Silva SR, Greboge TML. Infectious complications associated with the totally implantable central venous catheter. J Nurs UFPE on line. 2017;11(12):5049-58. https://doi.org/10.5205/1981-8963-v11i12a25104p5049-5058-2017.
https://doi.org/10.5205/1981-8963-v11i12...

Voog et al.99 Voog E, Campion L, du Rusquec P, et al. Totally implantable venous access ports: a prospective long-term study of early and late complications in adult patients with cancer. Support Care Cancer. 2018;26(1):81-9. http://dx.doi.org/10.1007/s00520-017-3816-3. PMid:28756472.
http://dx.doi.org/10.1007/s00520-017-381...
published a prospective, single-center, observational study with 483 patients who had undergone implantation of at least one TIVAP between January 1, 2006 and December 31, 2006, with a total of 493 TIVAPs fitted, equating to a total of 367,359 catheter-days. These patients were followed until removal of the device, death, or end of follow-up on December 31, 2013. Over a mean follow-up of 18 months (1–94 months), the study reported 87 complications, described in Table 1 (0.237 complications/1,000 catheter-days). The three most common complications were infections (37 cases), thrombotic events (17 cases), and extravasations (9 cases).

Table 1
Possible complications caused by totally implantable venous access ports (TIVAP) observed in a study by Voog et al.99 Voog E, Campion L, du Rusquec P, et al. Totally implantable venous access ports: a prospective long-term study of early and late complications in adult patients with cancer. Support Care Cancer. 2018;26(1):81-9. http://dx.doi.org/10.1007/s00520-017-3816-3. PMid:28756472.
http://dx.doi.org/10.1007/s00520-017-381...
with 483 patients with TIVAP fitted.

In the present study, the patient suffered an early complication involving embolization of the catheter – a rare event that is reported in just 1% of TIVAP complications.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...
,33 Motta Leal JM Fo, Carnevale FC, Nasser F, et al. Endovascular techniques and procedures, methods for removal of intravascular foreign bodies. Rev Bras Cir Cardiovasc. 2010;25(2):202-8. PMid:20802912. Migration may involve the entire length of the catheter or part of it, in cases of catheter fracture. In the present case, the entire catheter had migrated; and the catheter had therefore become disconnected from the port, which could occur because of excessive pressure during infusion of drugs or because of poor surgical technique. In contrast, fracture is generally a late complication and may be linked with catheters placed by puncture, resulting from compression of the catheter where it passes between the clavicle and the first rib. The rate of TIVAP fracture is lower when the catheter is fitted using dissection, whether of the cephalic vein or the external jugular.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...

Embolization may occur in the atrium, ventricle, pulmonary artery, hepatic veins, or vena cava and, because of circulation of fragments, the major clinical concerns are occurrence of cardiac perforation, arrhythmia, sepsis, and pulmonary emboli. Despite this, patients with embolization have scant symptomology and are generally asymptomatic, so diagnostic suspicion is aroused when it is impossible to perform infusion or reflux of blood is absent.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...
In the case described here, the condition was suspected immediately before administration of chemotherapy, when swelling around the port site was observed and no blood could be aspirated.

Death from embolization by chemotherapy TIVAP is rare, and diagnosis is achieved by radiography showing the foreign body. Treatment must always be provided promptly, to avoid major complications, and the endovascular technique is standard treatment, as performed in the patient described.11 Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717. PMid:29930688.
http://dx.doi.org/10.1590/1677-5449.0077...
In the past, surgery was the only treatment option for fractured and migrated catheters. However, percutaneous removal of migrated catheters is linked with lower morbidity and mortality than surgical procedures.55 Cakır F, Geze S, Ozturk MH, Dınc H. Percutaneous endovascular removal of intracardiac migrated port A catheter in a child with acute lymphoblastic leukemia. Braz J Anesthesiol. 2014;64(4):275-7. http://dx.doi.org/10.1016/j.bjane.2012.11.001.
http://dx.doi.org/10.1016/j.bjane.2012.1...

CONCLUSIONS

Migration of TIVAP fragments to the hepatic vein is an extremely rare and potentially lethal complication. Treatment must therefore be provided as early as possible and the standard treatment employs the endovascular access technique because of its low rate of complications. Despite the rarity of events such as these, the chemotherapy team must always be alert to any type of difficulty with drawing blood or administrating liquid via the catheter.

  • How to cite: Silveira ACMRL, Fernandes PSF, Fontinele DRS, Costa REAR, Araújo JEP, Sousa Junior WO, Vieira SC. Hepatic vein migration of a totally implantable venous access port-a-cath for chemotherapy in a breast carcinoma patient: case report. J Vasc Bras. 2022;21:e20210189. https://doi.org/10.1590/1677-5449.202101892
  • Financial support: None.
  • The study was carried out at Oncocenter, Teresina, PI, Brazil.

REFERÊNCIAS

  • 1
    Garcez BBD, Eulálio WMN Fo, Vieira SC. Embolização para veia cava inferior de cateter totalmente implantável para quimioterapia. J Vasc Bras. 2018;17(1):89-92. http://dx.doi.org/10.1590/1677-5449.007717 PMid:29930688.
    » http://dx.doi.org/10.1590/1677-5449.007717
  • 2
    Fonseca IY, Krutman M, Nishinari K, et al. Brachial insertion of fully implantable venous catheters for chemotherapy: complications and quality of life assessment in 35 patients. Einstein (Sao Paulo). 2016;14(4):473-9. http://dx.doi.org/10.1590/s1679-45082016ao3606 PMid:28076593.
    » http://dx.doi.org/10.1590/s1679-45082016ao3606
  • 3
    Motta Leal JM Fo, Carnevale FC, Nasser F, et al. Endovascular techniques and procedures, methods for removal of intravascular foreign bodies. Rev Bras Cir Cardiovasc. 2010;25(2):202-8. PMid:20802912.
  • 4
    Danski MTR, de Amorim SD, de Oliveira GLR, Pedrolo E, da Silva SR, Greboge TML. Infectious complications associated with the totally implantable central venous catheter. J Nurs UFPE on line. 2017;11(12):5049-58. https://doi.org/10.5205/1981-8963-v11i12a25104p5049-5058-2017
    » https://doi.org/10.5205/1981-8963-v11i12a25104p5049-5058-2017
  • 5
    Cakır F, Geze S, Ozturk MH, Dınc H. Percutaneous endovascular removal of intracardiac migrated port A catheter in a child with acute lymphoblastic leukemia. Braz J Anesthesiol. 2014;64(4):275-7. http://dx.doi.org/10.1016/j.bjane.2012.11.001
    » http://dx.doi.org/10.1016/j.bjane.2012.11.001
  • 6
    Xu H, Chen R, Jiang C, et al. Implanting totally implantable venous access ports in the upper arm is feasible and safe for patients with early breast cancer. J Vasc Access. 2020;21(5):609-14. http://dx.doi.org/10.1177/1129729819894461 PMid:31841061.
    » http://dx.doi.org/10.1177/1129729819894461
  • 7
    Skelton WP 4th, Franke AJ, Welniak S, et al. Investigation of complications following port insertion in a cancer patient population: a retrospective analysis. Clin Med Insights Oncol. 2019;13:1179554919844770. http://dx.doi.org/10.1177/1179554919844770 PMid:31040735.
    » http://dx.doi.org/10.1177/1179554919844770
  • 8
    Tumay LV, Guner OS. Availability of totally implantable venous access devices in cancer patients is high in the long term: a seven-year follow-up study. Support Care Cancer. 2021;29(7):3531-8. http://dx.doi.org/10.1007/s00520-020-05871-6 PMid:33155092.
    » http://dx.doi.org/10.1007/s00520-020-05871-6
  • 9
    Voog E, Campion L, du Rusquec P, et al. Totally implantable venous access ports: a prospective long-term study of early and late complications in adult patients with cancer. Support Care Cancer. 2018;26(1):81-9. http://dx.doi.org/10.1007/s00520-017-3816-3 PMid:28756472.
    » http://dx.doi.org/10.1007/s00520-017-3816-3

Publication Dates

  • Publication in this collection
    09 May 2022
  • Date of issue
    2022

History

  • Received
    27 Oct 2021
  • Accepted
    07 Mar 2022
Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV) Rua Estela, 515, bloco E, conj. 21, Vila Mariana, CEP04011-002 - São Paulo, SP, Tel.: (11) 5084.3482 / 5084.2853 - Porto Alegre - RS - Brazil
E-mail: secretaria@sbacv.org.br