Acessibilidade / Reportar erro

Effect of the number of abutments on biomechanics of Branemark prosthesis with straight and tilted distal implants

Abstract

OBJECTIVE: This study aimed to evaluate the bending moments, and compressive and tensile forces in implant-supported prostheses with three, four or five abutments. MATERIAL AND METHODS: Ten Pd-Ag frameworks were tested over two master models with: 1) parallel vertical implants, and 2) tilted distal implants. Strain gauges were fixed on the abutments of each master model to measure the deformation when a static load of 50 N was applied on the cantilever (15 mm). The deformation values were measured when the metallic frameworks were tested over three, four or five abutments, and transformed into force and bending moment values. Data were analyzed by ANOVA and Tukey's test for multiple comparisons at 5% level of significance. RESULTS: Abutment #1 (adjacent to the cantilever) had the highest values of force and sagittal bending moment for all tests with three, four or five abutments. Independently from the number of abutments, axial force in abutment #1 was higher in the vertical model than in the tilted model. Total moment was higher with three abutments than with four or five abutments. Independently from the inclination of implants, the mean force with four or five abutments was lower than that with three abutments. CONCLUSION: The results suggest that in the set-ups with four or five abutments tilted distal implants reduced axial force and did not increase bending moments.

Biomechanics; Implant-supported prostheses; Number of abutments; Tilted implants; Strain gauges


ORIGINAL ARTICLES

Effect of the number of abutments on biomechanics of Branemark prosthesis with straight and tilted distal implants

Marcos Michelon NaconecyI; Tomás GeremiaII; André CervieriIII; Eduardo Rolim TeixeiraIV; Rosemary Sadami ShinkaiIV

IDDS, MSc, PhD, Research Fellow, Graduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul, Porto Alegre, RS, Brazil

IIDDS, MSc, Assistant Professor, Department of Prosthodontics, Pontifical Catholic University of Rio Grande do Sul, Porto Alegre, RS, Brazil

IIIEng., MSc, Ph.D, Associate Professor, Department of Engineering, Lutheran University of Brazil, Canoas, RS, Brazil

IVDDS, MSc, PhD, Associate Professor, Department of Prosthodontics, Pontifical Catholic University of Rio Grande do Sul, Porto Alegre, RS, Brazil

Corresponding address Corresponding address: Marcos M. Naconcecy Rua Mariate, 288, cj. 606 - Porto Alegre, RS - 90430-180 - Brazil Phone: +55-51-3222-1105 e-mail: marcosnaconecy@terra.com.br

ABSTRACT

OBJECTIVE: This study aimed to evaluate the bending moments, and compressive and tensile forces in implant-supported prostheses with three, four or five abutments.

MATERIAL AND METHODS: Ten Pd-Ag frameworks were tested over two master models with: 1) parallel vertical implants, and 2) tilted distal implants. Strain gauges were fixed on the abutments of each master model to measure the deformation when a static load of 50 N was applied on the cantilever (15 mm). The deformation values were measured when the metallic frameworks were tested over three, four or five abutments, and transformed into force and bending moment values. Data were analyzed by ANOVA and Tukey's test for multiple comparisons at 5% level of significance.

RESULTS: Abutment #1 (adjacent to the cantilever) had the highest values of force and sagittal bending moment for all tests with three, four or five abutments. Independently from the number of abutments, axial force in abutment #1 was higher in the vertical model than in the tilted model. Total moment was higher with three abutments than with four or five abutments. Independently from the inclination of implants, the mean force with four or five abutments was lower than that with three abutments.

CONCLUSION: The results suggest that in the set-ups with four or five abutments tilted distal implants reduced axial force and did not increase bending moments.

Key words: Biomechanics. Implant-supported prostheses. Number of abutments. Tilted implants. Strain gauges.

INTRODUCTION

Studies on the biomechanics of different designs of implant-supported prostheses may help maximizing the clinical outcomes of implant treatment. Five to six implants and distal cantilever were traditionally indicated to rehabilitate the edentulous mandible and maxilla by means of fixed implant-supported prosthesis6. More recently, clinical reports have shown short- and medium-term success using less implants combined or not with inclination of distal implants4,7,19-21. Nevertheless, it still is controversial how many implants would be necessary to support a fixed implant-supported prosthesis with greater predictability. Also, no experimental data are currently available showing biomechanical gain with combination of implant inclination and reduction of the number of implants.

The technique of implant inclination was introduced for selected cases of multiple implants in the edentulous maxilla and mandible. In the mandible, the procedure can be used when the mental foramen is positioned low in relation to the alveolar ridge to reduce the cantilever extension and increase the polygonal area of prosthesis support19,23. In maxilla with bone atrophy and presence of large sinuses, longer inclined implants can be placed in areas of high bone density, with emergency at the first molar region. Without using tilted implants, these regions would receive shorter implants or would need maxillary sinus floor augmentation or bone grafting, increasing the treatment complexity, time, and costs1,5,8.

Experimental strain levels transmitted to tilted implants and surrounding bone and the deformation of prosthetic components still are unclear. The axial and non-axial forces generated during oral function may result from sommatory, synergistic or antagonic effects of implant inclination and number and distribution of implants in the arch. Strain gauges studies12,13,15 as well as mathematical23, photoelastic25 and finite element models2,3 have been used to explain the biomechanical behavior of implant-supported prostheses simulating the variation of number or inclination of implants, yet no experimental study has evaluated the combined effect of these variables. Therefore, this study used strain gauges to assess the effect of the number of abutments and inclination of distal implants on axial forces and bending moments in implant-supported prostheses.

MATERIAL AND METHODS

Two trapezoid epoxy resin bases were used to fabricate the master model with vertical (straight) implants and the master model with posterior tilted implants. The arch (curve of 134.30º and radium of 17.61 mm) of a mandible model (ETH 0301-10 Nobel Biocare, Gothenburg, Sweden) was transposed to the epoxy bases for the perforation of the implant sites. For the model with vertical implants, the central implant site was marked at the sagittal line; five perforations (4 mm-diameter, 17 mm-length) were made parallel and 1 cm apart from each other. For the model with tilted implants, the three central perforations were made vertical, and the two posterior perforations were tilted using an index with a 27-degree inclination plane. Ten 4.0 x 15-mm screw-type implants (OSS 415, 3i Implant Innovations, Palm Beach Gardens, FL, USA) were fixed into the perforations with fluid epoxy resin. After 12 h, ten 7-mm standard abutments (AB700, 3i Implant Innovations, Palm Beach Gardens, FL, USA) were attached with 20 Ncm torque.

Over each master model, five bars (rectangular section, 3-mm width, 4-mm height, and 20-mm of cantilever at the left side) were waxed up 1 mm above the epoxy base. The cantilever was placed on the left side beginning at the emergency point of the posterior implant. The wax patterns were sectioned into five segments and cast in a Pd-Ag alloy (Porson 4, Degussa, Dusseldorf, Germany) according to standard procedures. After finishing, the bar segments were laser-welded (EV LASER 900, Bergamo, Italy). The dimensions of the metallic bars were verified using digital calipers, and passivity of fit of the welded framework was checked by tightening one screw at time. The loading point on the cantilever was standardized at a 15-mm distance from the posterior emergency of the distal implant (Figure 1). With a milling machine, a concave notch was made with half-depth of a round tungsten bur of 2-mm diameter. This notch matched the load applicator tip (2-mm diameter) of a customized mechanical device used to deliver the 50 N static load during the tests.


Three strain gauges (KFG02-120C1-11N15C2, Kyowa Electronic Instruments Co Ltd, Tokyo, Japan) were attached to the abutment surface, 120º apart, in the following geometric position: one anterior, one posterior to the right, and one posterior to the left. The 0.2-mm measuring grid was placed 1 mm above the implant platform in parallel with the axes of the cylinders. One strain gauge formed one channel for reading deformation (1/4 of a Wheatstone bridge). Therefore, 15 reading channels were built for each master-model (three per abutment). Each strain gauge was connected to two cables leading the signals to a 15-channel strain gauge conditioner (MGC Plus, HBM Inc, Berlin, Germany). The analogical signal of electric resistance variation was converted into a digital signal via a 12-byte resolution converter (MGC Plus, HBM Inc, Berlin, Germany). These signals were software-processed (MGC Plus, HBM Inc, Berlin, Germany), and channel signals originally measured in millivolts were converted into microstrain units (µm/m).

Each framework was screwed (GS300; 3i Implant Innovations, Palm Beach Gardens, FL, USA) onto the respective master model with a 10 Ncm torque (DEC 600-1 Ossecare Drilling Equipment, and DIA 189-0, Nobel Biocare AB, Gothenburg, Sweden). The abutments were numbered clockwise (#1 to #5; abutment #1 was adjacent to the cantilever), and the tightening sequence was 2, 4, 3, 1, and 518. A new set of screws was used for each framework to avoid screw fatigue. After the strain gauges were calibrated to zero, a 50 N static load was applied on the cantilever generating a graph of deformation. The point of signal stabilization was selected, and the deformation values were extracted. This test procedure was performed for all five frameworks supported by five abutments, then repeated with the frameworks supported by four or three implants/abutments. For the four-implant configuration, the central abutment (abutment #3) was removed. For the three-implant configuration, abutments #2 and #4 were removed (Figure 2).


The readings of the strain gauges (deformation in microstrain unit) were transformed into normal axial force and bending moments around the X- and Y-axis using the calibration method and equations described by Duyck, et al.12,13 All abutments were individually calibrated with a 50 N static load to the implant/abutment axis. This calibration was performed by loading a custom-cast disc fixed to each abutment in five standardized positions so that the axial force (in relation to the abutment axis) and bending moments (sagittal and lateral) were computed separately. For axial force, a positive signal was conventionally adopted for compressive force and a negative signal for tensile force; all calculations were performed using the absolute values. Data were analyzed by Analysis of Variance for random blocks design using the Proc Mixed tool of the software SAS 9.1 - Type 3 Tests of Fixed Effects, followed by pairwise comparisons Tukey's tests. A significance level of 0.05 was used for all tests.

RESULTS

Figures 3, 4, and 5 display the mean axial force, sagittal bending moment, and lateral bending moment in each abutment for the models with three, four, and five abutments, respectively. For the model with three abutments, both main effects were significant for axial force (Model: P=0.006; Abutment: P<0.001), only the factor Abutment was significant for sagittal bending moment (P<0.001), and there was a significant interaction between Model and Abutment for lateral bending moment (P<0.001). For the model with four abutments, a significant interaction between Model and Abutment was found for axial force (P=0.029) and lateral bending moment (P=0.018); for sagittal bending moment, both main effects were significant (Model: P=0.018; Abutment: P<0.001). Regarding the model with five abutments there was a significant interaction between Model and Abutment for axial force (P<0.001), but only the factor Abutment was significant for sagittal and lateral bending moments (P<0.001).



Figure 6 shows the results for abutment #1. For axial force, interaction between Model and Number of abutments (P=0.623) and the main factor Number of abutments (P=0.759) were not significant. Only the factor Model was significant (P=0.001): the straight model had higher mean force than the tilted model. Sagittal and lateral bending moment means were not statistically different.


DISCUSSION

This study showed that four or five abutments provided better distribution of forces and bending moments compared with the configuration with three abutments. Overall, the inclination of distal implants reduced the axial force and bending moments independently from the number of abutments. In relation to the direction of axial forces, compressive forces were measured in the distal abutments and tensile forces in the most anterior abutments in the arch in accordance with the "hinging effect" proposed by Duyck, et al.13.

The magnitude of forces measured with four or five abutments was similar and lower than that with three abutments. Davis, Zarb and Chao11 also observed the highest deformation of frameworks supported by only two abutments, but the distribution of axial forces and bending moments were similar when four or five abutments were used. Conversely, Duyck, et al.13 found in vivo lower forces with five implants in comparison with the arrangements with four or three implants. Some of these discrepancies may be explained by the difference of interimplant distances that alter the geometric distribution of abutments in the arch and the length of framework segments. The magnitude of resulting forces also depends on the deformation of the entire system, which may be influenced by the elastic deformation of the framework17 and by the material used to fabricate the prosthetic screws, which influence preload and torque values24.

In the present study, the values of bending moment were lower than those of axial force independently from the number of abutments, but higher bending moments were usually observed with fewer abutments. Conversely, Glantz, et al.15 recorded higher values of bending moments than axial forces during maximal biting in vivo, which may be explained by the distribution of occlusal contacts and resulting non-axial forces.

The results observed for the framework supported by three abutments suggest that, although the central abutment provided a longer resistance arm (15 mm) than the configuration with four abutments (11.35 mm), the quadrilateral polygon resulted in better distribution of forces. This was more evident for the non-axial forces represented by the sagittal and lateral bending moments. Other biomechanical studies using an analytical mathematical model23 and finite elements analysis2 demonstrated that a spread-out arrangement of implants in the arch is more significant than the number of implants per se for the distribution of masticatory forces.

The tilted implants/abutments were calibrated with a 50 N load axial to the implant/abutment axis and non-perpendicular to the metallic bar. Therefore, this study measured the axial force in relation to the implant/abutment axis to evaluate the effect of implant inclination. The inclination of distal implants reduced the axial forces in all abutments and also in abutment #1 when four or five abutments were used. This inclination allowed simultaneous reduction of the cantilever length at the connection abutment-framework and increase of the prosthesis support area. Using a mathematical model, Skalak23 predicted that the inclination of 30 degrees of the distal implant considerably reduces the forces in a configuration with three implants. In mandibular prostheses, the fulcrum of rotation for the distal implant depends on the anatomical position of the mental foramen in relation to the alveolar ridge. The more apical the foramen, the more apical the fulcrum. In this experiment, the distal implants were inclined having the implant platform as the fulcrum of rotation, and the implant was not displaced distally any further. However, considering the connection framework-abutment, the use of 7-mm abutments reduced the cantilever length from 15 to 12.16 mm. This reduction of the cantilever length is inherent to the implant tilting. This in vitro study aimed to assess the effect of tilting the distal implant per se on the axial forces and bending moments of the system as the tilted implant platforms did not emerge more distally in relation to the vertical model cast. This was designed to allow the evaluation of the isolated variable inclination without the combined effect of further distalizing the implant platform. The results in abutment #1 were of particular interest and showed that not only the axial force is lower on the tilted abutment but also the sagittal bending moment did not increase. Because the present study did not vary the position of the distal platform, this finding can be attributed to the variable implant inclination. If it had chosen to position the tilted implant more distally, it would not be possible to isolate the effect of implant inclination.

The main strength of the paper is that this is the first experimental work to prove that tilting the distal implants may offer a biomechanical advantage over vertical implants when cantilever is needed. Previous works used an analytical method or finite element analysis, which simulate experimental conditions to some extent and require several simplifications of the geometric models and mathematical approach. However, this study has some limitations because the experiment simulated a specific design of a fixed implant-supported prosthesis for the edentulous mandible. Previous studies10,16,22 have highlighted that the load transference from implants to bone depends on the type and place of loading, bone-implant interface, implant geometry and surface, framework alloy, density of cancellous bone, and abutment length. The literature also reports variation of deformation, forces and/or bending moments in abutments and implants due to implant brand9, framework alloy22, cantilever extension17, and occlusal contacts14.

Another limitation of the present study is inherent to the strain gauge size and placement, which do not allow measurement of the forces and bending moments directly in implants and bone interface. Furthermore, the absolute values of forces and bending moments are valid only for the present study set-up. The absolute values of forces and bending moments in abutments cannot be directly generalized to implants because of the joints and gaps among prosthetic components, screws, and implants. However, theoretically one can expect having similar vectors and biomechanical behavior in tilted implants and abutments as long as the abutment and the implant are aligned in the same longitudinal axis.

CONCLUSIONS

The present results suggest that tilted distal implants reduce axial force and do not increase bending moments when four or five abutments are used. Further controlled clinical studies are necessary to evaluate the combined effect of the number of abutments and inclination of distal implants on short- and long-term success in the daily practice.

ACKNOWLEDGEMENTS

We would like to thank Mr. Cedenir Albani, CDT, and Mr. Cristiano Friederichs, CDT, from the PortoDent Dental Laboratory and Mr Eubirajara Medeiros, Eng, MSc, from the PROMM Surgical Materials Ltd., for their technical assistance, and Dr. Telmo Strohaecker, Eng, PhD, from the Federal University of Rio Grande do Sul, coordinator of the laboratory where the experimental data were collected. This research was partially supported by the Brazilian Ministry of Education and Culture/CAPES.

Received: August 15, 2008

Modification: July 21, 2009

Accepted: October 22, 2009

  • 1- Aparicio C, Perales P, Rangert B. Tilted implants as an alternative to maxillary sinus grafting: a clinical, radiologic, and periotest study. Clin Implant Dent Relat Res. 2001;3:39-49.
  • 2- Bellini CM, Romeo D, Galbusera F, Taschieri S, Raimondi MT, Zampelis A, et al. Comparison of tilted versus nontilted implant-supported prosthetic designs for the restoration of the edentuous mandible: a biomechanical study. Int J Oral Maxillofac Implants. 2009;24:511-7.
  • 3- Bevilacqua M, Tealdo T, Pera F, Menini M, Mossolov A, Drago C, et al. Three-dimensional finite element analysis of load transmission using different implant inclinations and cantilever lengths. Int J Prosthodont. 2008;21:539-42.
  • 4- Branemark P-I, Engstrand P, Ohrnell L-O, Grondahl K, Nilsson P, Hagberg K, et al. Branemark Novum: a new treatment concept for rehabilitation of the edentulous mandible. Preliminary results from a prospective clinical follow-up study. Clin Implant Dent Relat Res. 1999;1:2-16.
  • 5- Branemark P-I, Grondahl K, Ohrnell LO, Nilsson P, Petruson B, Svensson B, et al. Zygoma fixture in the management of advanced atrophy of the maxilla: technique and long-term results. Scand J Plast Reconstr Surg Hand Surg. 2004;38:70-85.
  • 6- Branemark P-I, Hansson BO, Adell R, Breine U, Lindstrom J, Hallen O, et al. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg Suppl. 1977;16:1-132.
  • 7- Branemark P-I, Svensson B, Van Steenberghe D. Ten-year survival rates of fixed prostheses on four or six implants ad modum Branemark in full edentulism. Clin Oral Implants Res. 1995;6:227-31.
  • 8- Calandriello R, Tomatis M. Simplified treatment of the atrophic posterior maxilla via immediate/early function and tilted implants: a prospective 1-year clinical study. Clin Implant Dent Relat Res. 2005;7(Suppl 1):S1-12.
  • 9- Çehreli M, Duyck J, De Cooman M, Puers R, Naert I. Implant design and interface force transfer. A photoelastic and strain-gauge analysis. Clin Oral Implants Res. 2004;15:249-57.
  • 10- Correia A, Piloto P, Campos JC; Vaz M. Finite element analysis of the mechanical behavior of a partially edentulous mandible as a function of cancellous bone density. Rev Odonto ciênc. 2009;24:22-7.
  • 11- Davis DM, Zarb GA, Chao YL. Studies on frameworks for osseointegrated prostheses: Part 1. The effect of varying the number of supporting abutments. Int J Oral Maxillofac Implants. 1988;3:197-201.
  • 12- Duyck J, Van Oosterwyck H, De Cooman M, Puers R, Vander Sloten J, Naert I. Three-dimensional force measurements on oral implants: a methodological study. J Oral Rehabil. 2000;27:744-53.
  • 13- Duyck J, Van Oosterwyck H, Vander Sloten J, De Cooman M, Puers R, Naert I. Magnitude and distribution of occlusal forces on oral implants supporting fixed prostheses: an in vivo study. Clin Oral Implants Res. 2000;11:465-75.
  • 14- Falk H, Laurell L, Lundgren D. Occlusal interferences and cantilever joint stress in implant-supported prostheses occluding with complete dentures. Int J Oral Maxillofac Implants. 1990;5:70-7.
  • 15- Glantz PO, Rangert B, Svensson A, Stafford GD, Arnvidarson B, Randow K, et al. On clinical loading of osseointegrated implants. A methodological and clinical study. Clin Oral Implants Res. 1993;4:99-105.
  • 16- Greco GD, Jansen WC, Landre J Jr, Seraidarian PI. Stress analysis on the free-end distal extension of an implant-supported mandibular complete denture. Braz Oral Res. 2009;23:182-9.
  • 17- Jacques LB, Moura MS, Suedam V, Souza EA, Rubo JH. Effect of cantilever length and framework alloy on the stress distribution of mandibular-cantilevered implant-supported prostheses. Clin Oral Implants Res. 2009;20:737-41.
  • 18- Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Branemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants. 1991;6:270-6.
  • 19- Krekmanov L, Kahn M, Rangert B, Lindstrom H. Tilting of posterior mandibular and maxillary implants for improved prosthesis support. Int J Oral Maxillofac Implants. 2000;15:405-14.
  • 20- Malo P, Rangert B, Nobre M. "All-on-Four" immediate-function concept with Branemark System implants for completely edentulous mandibles: a retrospective clinical study. Clin Implant Dent Relat Res. 2003;5(Suppl 1):2-9.
  • 21- Malo P, Rangert B, Nobre M. All-on-4 immediate-function concept with Branemark System implants for completely edentulous maxillae: a 1-year retrospective clinical study. Clin Implant Dent Relat Res. 2005;7(Suppl 1):S88-S94.
  • 22- Rubo JH, Souza EA. Finite element analysis of stress in bone adjacent to dental implants. J Oral Implantol. 2008;34:248-55.
  • 23- Skalak R. Biomechanical considerations in osseointegrated prostheses. J Prosthet Dent. 1983;49:843-8.
  • 24- Stüker RA, Teixeira ER, Beck JC, Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci. 2008;16:55-8.
  • 25- Ueda C, Markarian RA, Sendyk CL, Lagana DC. Photoelastic analysis of stress distribution on parallel and angled implants after installation of fixed prostheses. Braz Oral Res. 2004;18:45-52.
  • Corresponding address:

    Marcos M. Naconcecy
    Rua Mariate, 288, cj. 606 - Porto Alegre, RS - 90430-180 - Brazil
    Phone: +55-51-3222-1105
    e-mail:
  • Publication Dates

    • Publication in this collection
      04 May 2010
    • Date of issue
      Apr 2010

    History

    • Received
      15 Aug 2008
    • Accepted
      22 Oct 2009
    • Reviewed
      21 July 2009
    Faculdade De Odontologia De Bauru - USP Serviço de Biblioteca e Documentação FOB-USP, Al. Dr. Octávio Pinheiro Brisolla 9-75, 17012-901 Bauru SP Brasil, Tel.: +55 14 32358373 - Bauru - SP - Brazil
    E-mail: jaos@usp.br