Acessibilidade / Reportar erro

Safety and efficacy of biolimus-eluting stent with biodegradable polymer: insights from EINSTEIN (Evaluation of Next-generation drug-eluting STEnt IN patients with coronary artery disease) Registry

Abstracts

OBJECTIVE: To evaluate the incidence of major adverse cardiac events (cardiac death, or acute myocardial infarct, or target vessel revascularization) at one year in "real world" patients. METHODS: The EINSTEIN registry is an observational, prospective, single center study that consecutively included 103 patients (152 lesions) treated with the BiomatrixTM stent, a biolimus A9-eluting stent with biodegradable polymer. RESULTS: The mean age was 65.0±12.4 years; male gender represented 83.5% of the patients; and 37.9% of them were diabetic. At one-year, major adverse cardiac events occurred in 11.7% of the patients, including 2.9% of cardiac death, 4.9% of with non ST- segment elevation acute myocardial infarction, and 3.9% of target vessel revascularization. Stent thrombosis occurred in only 1% (1 patient) at one-yearfollow-up. CONCLUSION: The present Registry suggests that new generation biolimus A9 drug-eluting stents are safe and effective in a "real world", all-comers patients, showing low rates of major cardiac adverse events on long-term follow-up.

Biocompatible events; Pharmacological stents; Drugdelivery systems; Polymers; Sirolimus; Coronary artery disease


OBJETIVO: Avaliar a incidência de eventos adversos cardíacos maiores (morte cardíaca, infarto agudo do miocárdio ou necessidade de nova revascularização do vaso-alvo) em 1 ano em pacientes do "mundo real". MÉTODOS: O registro EINSTEIN é um estudo observacional prospectivo, unicêntrico, que incluiu consecutivamente 103 pacientes (152 lesões) tratados com o stent BiomatrixTM, eluidor de biolimus A9 e polímero biodegradável. RESULTADOS: A média das idades foi de 65,0±12,4 anos; o gênero masculino representou 83,5% dos pacientes; e 37,9% deles eram diabéticos. Em 1 ano, eventos adversos cardíacos maiores ocorreram em 11,7% dos pacientes, incluindo 2,9% de morte cardíaca, 4,9% de infarto agudo do miocárdio sem supradesnivelamento do segmento ST e 3,9% de revascularização do vaso-alvo. Trombose de stent esteve presente em apenas 1% (1 paciente) após o seguimento clínico de 1 ano. CONCLUSÃO: O presente registro sugere que os novos stents farmacológicos eluidores de biolimus A9 com polímero biodegradável são seguros e eficazes em pacientes da prática clínica diária, com baixas taxas de eventos adversos cardíacos maiores no longo prazo.

Materiais biocompatíveis; Stents farmacológicos; Sistemas de liberação de medicamentos; Polímeros; Sirolimo; Doença da artéria coronariana


ORIGINAL ARTICLE

Safety and efficacy of biolimus-eluting stent with biodegradable polymer: insights from EINSTEIN (Evaluation of Next-generation drug-eluting STEnt IN patients with coronary artery disease) registry

Cristiano Freitas de Souza; Anwar Mohamed El Mouallem; Fábio Sândoli de Brito Júnior; Alexandre Antônio Cunha Abizaid; Breno Oliveira Almeida; Amanda Gonçalves Almeida; Teresa Cristina Dias Cunha Nascimento; Marco Antonio Perin; Adriano Caixeta

Hospital Israelita Albert Einstein, Sao Paulo, SP, Brazil

Corresponding author

ABSTRACT

OBJECTIVE: To evaluate the incidence of major adverse cardiac events (cardiac death, or acute myocardial infarct, or target vessel revascularization) at one year in "real world" patients.

METHODS: The EINSTEIN registry is an observational, prospective, single center study that consecutively included 103 patients (152 lesions) treated with the BiomatrixTM stent, a biolimus A9-eluting stent with biodegradable polymer.

RESULTS: The mean age was 65.0±12.4 years; male gender represented 83.5% of the patients; and 37.9% of them were diabetic. At one-year, major adverse cardiac events occurred in 11.7% of the patients, including 2.9% of cardiac death, 4.9% of with non ST- segment elevation acute myocardial infarction, and 3.9% of target vessel revascularization. Stent thrombosis occurred in only 1% (1 patient) at one-yearfollow-up.

CONCLUSION: The present Registry suggests that new generation biolimus A9 drug-eluting stents are safe and effective in a "real world", all-comers patients, showing low rates of major cardiac adverse events on long-term follow-up.

Keywords: Biocompatible events; Pharmacological stents; Drug-delivery systems; Polymers/chemical synthesis; Sirolimus/analogs & derivatives; Coronary artery disease/therapy

INTRODUCTION

Since its incorporation into medical practice a decade ago, drug-eluting stents (DES) have been used in various clinical and anatomical scenarios due to their reduction in restenosis rates. Compared to bare-metal stents, first generation DES were more effective in decreasing neointimal hyperplasia, which consequently led to a reduction in late luminal loss and binary restenosis(1-3). Subsequently, however, concerns have been raised about the increased late and very late thrombosis with first generation DES(4-6). Although the mechanism of thrombotic events is not yet fully elucidated, hypersensitivity reaction to the permanent/non degradable polymer is one of the potential explanation in this complex equation(7,8). Thus, a novel generation of DES such as BiomatrixTM (Biosensors International, Morges, Switzerland) has been emerged with the aim to diminish stent thrombosis as well as other safety complications related to first generation DES. One of the BiomatrixTM main features is the presence of a biodegradable polymer comprised of polylactic acid. This polymer is applied only on its abluminal surface which, after a total drug release in 6 to 9 months, is metabolized into carbon acid and water. BiomatrixTM is eluted with e biolimus A9 drug, an immunosuppression and anti-proliferative compound of the same family as sirolimus, but with solubility ten times higher. In the randomized LEADERS study, the BiomatrixTM stent proved to be safer than the first-generation DES CypherTM(9). However, the long-term clinical results with BiomatrixTM stent in a real world, all- comers patients are scarce.

OBJECTIVE

The primary endpoint of the EINSTEIN (Evaluation of Next-generation drug-eluting STEnt IN patients with coronary artery disease) registry is to evaluate the incidence of major adverse cardiac events (MACE). MACE was defined as the composite of cardiac death, acute myocardial infarct, or target vessel revascularization at 30 days, 6 months, 12 months, and annually up to 3 years follow-up. The secondary objectives were (1) procedure success, defined as the technical success with residual stenosis <30%; (2) occurrence of in-hospital MACE; (3) individual events of death, acute myocardial infarct, target vessel revascularization, and stent thrombosis at 30 days, 6 months, and 12 months.

METHODS

Patients and study design

The EINSTEIN registry was a prospective and single-center registry that enrolled consecutive and non-selected patients underwent percutaneuos coronary intervention (PCI) with BiomatrixTM stent implantation. From October 2008 to December 2010, a total of 103 patients with 152 coronary lesions were treated with BiomatrixTM stent at the Department of Interventional Cardiology at Hospital Israelita Albert Einstein (HIAE), in São Paulo. Among them, 101 (98.0%) completed one-year follow-up, representing the subject of analysis of this manuscript. Clinical follow-up was prespecified at 30 days, 6 months, 1 year, and yearly thereafter for a total 3 years in all patients. In order to demonstrate the daily clinical practice usefulness of the BiomatrixTM stent, the inclusion criteria were permissive, enrolling all-comers patients >18 years underwent elective or emergency PCI. Since it is an observational and retrospective analysis of our database, all signed informed Consent Forms were waived by the institutional Ethics Committee. The protocol was approved by the Ethics Committee of HIAE under the Nº. 11/1530 (CAAE: 0038.0.028.000-11).

Coronary intervention procedure and anti-aggregation protocol

The antiplatelet protocol consisted dual antiplatelet therapy with aspirin and clopidogrel or prasugrel prior stenting. After the procedure, aspirin 100mg/day and indefinitely, clopidogrel 75mg/day or prasugrel 10mg/day were recommended for at least 12 months. The number of stents implanted and techniques used were at physician's discretion.

Definitions

Cardiac death was defined as any death not attributable to a non-cardiac cause. Myocardial infarction was defined as when at least one of the following criteria was present: elevation of CPK-MB over three times the upper limit of normal; electrocardiogram with a new pathological Q-wave (>0.04 seconds) in at least two contiguous leadings with positive CPK-MB. Target vessel revascularization was defined as any attempt or successful PCI or coronary artery bypass grafting (CABG) percutaneous of the target-vessel. Stent thrombosis was classified according to the Academic Research Consortium (ARC) definition as definite, probable or possible. As per protocol, definite stent thrombosis was defined as a thrombus detected by angiography or total occlusion within the stented vessel at the time of clinically driven angiography for ischemia; or by pathological confirmation of an acute thrombosis in a patient with acute coronary syndrome. Probable stent thrombosis was defined as any death not attributed to a noncardiac cause or any myocardial infarction in the target vessel territory in the absence of documental angiographic stent patency. Possible stent thrombosis was defined as unexplained death after 30 days(10).

Clinical follow-up information was collected by telephone contact or medical visit at 30 days, 6 months, 12 months, and then annually thereafter.

Collection and management of data

Preparation and implementation of the registry were under the responsibility of the Department of Cardiovascular Intervention at HIAE.

Statistical analysis

In the descriptive analysis, categorical variables are presented as counts and percentages. Continuous variables are expressed as mean±standard deviation. One-year outcome are summarized as Kaplan-Meier curves and compared by log-rank statistical test. A p value of <0.05 was considered statistical significant. All statistical analysis were performed by the Statistical Package for the Social Science (SPSS) 13.0 software.

RESULTS

Clinical and procedure characteristics

The baseline clinical characteristics are shown in table 1. Mean age was 65.0±12.4 years, with a predominance of male gender (83.5%). Among risk factors for coronary artery disease (CAD), we highlight the high prevalence of diabetes mellitus (37.9%), systemic arterial hypertension (75.7%), and hypercholesterolemia (70.9%). The most prevalent clinical presentation was stable angina/silent ischemia (47.6%), followed by unstable angina (23.3%). Primary PCI for acute myocardial infarction was performed in 11.7% of the patients.

Procedural and angiographic characteristics are shown in tables 2 e 3. Most patients treated were multivessell (63%), including one third with three-vessel disease. Treatment included de novo lesions, intra-stent restenosis (12.7%), and treatment of bifurcation lesions (16.4%).

Clinical follow-up

Table 4 displays the major clinical endpoints during hospitalization, at 30 days, and12 months. Clinical follow-up was completed in 98% of the patients.

At 12 months, MACE occurred in 12 patients, 6 all-cause deaths (3 of cardiac death and 3 of non-cardiac death), 5 cases of non-STEMI, and 4 patients has TVR. However, it is important to point out the low incidence of stent thrombosis in the present study, occurring in only 1 patient; which was an acute stent thrombosis <24 hours after stenting. There were no cases of late or very late thrombosis.

Figure 1 shows Kaplan-Meyer curves in the overall population. Of note, albeit with no statistical significance, there was an trend toward between patients with acute coronary syndrome versus those with stable angina (87.0% versus 91.8%; p=0.43) (Figure 2). In addition, there was no difference in terms of event-free survival between the groups of patients with single versus multivessel disease (89.5% versus 89.2%; p=0.97) (Figure 3).




DISCUSSION

The results of the present manuscript corroborate and add new data to those previously published on the use of the BiomatrixTM biolimus A9-eluting (9,11-17) stent. The main findings of this registry are: in a "real world", all comers patients – including those with clinical (diabetes) and anatomical (multivessel, bifurcation, in-stent restenosis, and STEMI) complexity – the use of BiomatrixTM stent was safe and effective at long-term follow-up.

The presence of a durable polymer with a lack of biocompatibility in first-generation DES was associated with inflammation and vascular hypersensitivity reaction delay in reendothelization, and most importantly, with late and very late stent thrombosis and death. Biolimus is a synthetic drug analog of sirolimus, with a ten-fold higher lipophilicity than sirolimus, but with similar antiproliferative profile. The polylactic acid polymer is applied only on the abluminal surface, and after total release of the drug, it is completely dissolved into carbonic gas and water over a period of 6 to 9 months. These pharmacodynamics and bioengineering advantages make it, in thesis, a safer platform than the durable polymers used in first generation DES. The phase II study, STEALTH(11) by Grube et al., was the first to evaluate the safety and efficacy of the BiomatrixTM stent. In a randomized and double blind study, enrolling 120 patients showed an extremely favorable results for the BiomatrixTM arm (intra-stent loss late luminal of 0.26mm) when compared with the control arm (0.74mm). The event-free survival at 6 months was similar between the two groups. These results have encouraged the development of a new clinical trial, LEADERS(9), which randomized, 1,707 patients in ten centers in Europe. In this study, the safety and efficay of the BiomatrixTM (BES group) were compared to those of a first-generation sirolimus-eluting stent with a durable polymer (Cypher SELECTTM, Cordis, Miami Lakes, FL, USA). At 9 months, the rates of MACE (cardiac death, AMI, or TVR) in the BES group was non-inferior to those treated sirolimus-eluting stent (SES). Additionally, BES was also not angiographicaly inferior to SES (in-stent stenosis of 20.9% versus 23.3%, respectivaly). This early clinical outcome reported similar rates of acute and subacute definitive stent thrombosis (1.6% for BES versus 1.6% for SES), as well as late and very late stent thrombosis (0.2% for BES versus 0.5% for SES). Our present analysis showed rates of ischemic events similar to the LEADERS study, although our data included more diabetic patients (1/3 versus 25% in the LEADERS), more STEMI, and small vessels (2.4mm versus 2.6 mm in the LEADERS). Remarkably, we observed only one case of early stent thrombosis (related to the procedure) and no cases of late or very late stent thrombosis. Thus, despite the small sample size in this registry, the present study suggests that this novel BiomatrixTM platform with a biodegradable polymer is safe even in higher-risk patients. Nevertheless, due to an incremental rates of stent thrombosis over time with the first-generation of DES, longer term follow-up is necessary to confirm the non-inferiority and/or superiority of the DES with biodegradable polymer Indeed, 4-year follow-up of the LEADERS(12) study, showed once again non-inferiority of the biolimus-eluting stent relative to the primary outcome for BES (18.7%) versus SES (22.6%). Additionally, there was a 38% relative risk reduction of definite stent thrombosis between one and four years of follow-up in the BES group compared with the SES group. BiomatrixTM safety superiority compared with first-generation DES may be explained by more strut stent coverage on long-term follow-up, as demonstrated by optical coherence tomography(13).

Importantly, the BiomatrixTM stent also had its performance and safety evaluated in specific subgroups of patients. When patients were divided according to the Syntax Score (SX) into 3 groups (low SX if <8, intermediate SX if >8 and <16, and high SX if >16), the following rates of MACE were noted after two years follow-up: 9.4% for low SX, 12.0% for intermediate SX, and 18.4% for high SX (p<0.01). The rate of cardiac death in patients with high SX was significantly higher than the other two groups (7% versus 2.4% versus 1.8%). Nevertheless, in a subgroup of patients with greater anatomical complexity, those patients treated with BES had significantly lower rates of cardiac death when compared to the group treated with SES(14). Conversely, in a subgroup of small vessels (reference diameter <2.75mm), there was no significant difference in the rates of MACE and in the rates of TVR between the 2 groups (15). Finally, when analyzing STEMI patients, there was a significant reduction in MACE in patients treated with BES compared to those with SES (8.1% versus 19.3%)(16).

In a study by Soares Junior et al.(17), assessing long-term follow-up in 100 patients (164 lesions) treated with BiomatrixTM stent (mean clinical follow-up of 243 days), the authors reported similar results to our analysis. In their study, the primary endpoint occurred in 9% of the patients (cardiac death in 4%, non-fatal AMI in 2%, and TVR in 3%), and the rate of stent thrombosis was 1%. Likewise, no case of late or very late thrombosis has been reported in this cohort of patients.

Recently, the COMPARE II(18) study showed the safety and efficacy of biolimus-eluting stent with the biodegradable polymer (NoboriTM, Terumo Corporation, Tokyo, Japan) compared with new-generation everolimus-eluting stent with durable polymer (Xience VTM, Abbot Vascular, Santa Clara, CA, USA). The positive safety endpoints of the biolimus-eluting stent of this large-scale and randomized trial open a new perspective for dual antiplatelet therapy <1 year, even including higher-risk patients.

The present study has the following limitations. It is an observational study, performed in a single center, without active control group. The present study sample size is small and the results should be considered exploratory only. Further studies are warranted to drawn definite analysis. The treatment of patients biolimus-eluting stent was at physicians' discretion and, therefore, it is subject to selection. Finally, a longer clinical follow-up (>1 year) is required since the minority of patient had follow-up over one year. Therefore, the present data on safety and efficacy of biolimus-eluting stent cannot be extrapolated to those patients with longer follow-up period.

CONCLUSIONS

Novel generation biolimus A9-eluting stent with biodegradable polymer is safe and effective in a real world, all-comers patients including those with high clinical and anatomical complexity, presenting low ischemic event rates on long-term-follow-up.

ACKNOWLEDGEMENTS

This study was partially funded by Bioassist Medical LTDA, for quantitative coronary analysis.

REFERENCES

  • 1. Schwietz T, Ehrlich JR, De Rosa S, Fichtlscherer S, Schachinger V, Baier G, et al. Prognostic impact of using drug-eluting-stents on outcome and strategy in multivessel PCI: Data from the Frankfurt MV-PCI registry. J Cardiol. 2013;61(1):38-43.
  • 2. Bangalore S, Kumar S, Fusaro M, Amoroso N, Attubato MJ, Feit F, et al. Short- and long-term outcomes with drug-eluting and bare-metal coronary stents: a mixed-treatment comparison analysis of 117 762 patient-years of follow-up from randomized trials. Circulation. 2012;125(23):2873-91.
  • 3. Yeh RW, Normand SL, Wolf RE, Jones PG, Ho KK, Cohen DJ, et al. Predicting the restenosis benefit of drug-eluting versus bare metal stents in percutaneous coronary intervention. Circulation. 2011;124(14):1557-64.
  • 4. Camenzind E, Wijns W, Mauri L, Kurowski V, Parikh K, Gao JP, Boersma E, Vranckx P, McFadden E, Serruys PW, O'Neil WW, Jorissen B, Van Leeuwen F, Steg PG; PROTECT Steering Committee and Investigators. Stent thrombosis and major clinical events at 3 years after zotarolimuseluting or sirolimus-eluting coronary stent implantation: a randomised, multicentre, open-label, controlled trial. Lancet. 2012;380(9851):1396-405.
  • 5. Karjalainen PP, Nammas W, Ylitalo A. Late stent thrombosis of a secondgeneration drug-eluting stent. J Invasive Cardiol. 2012;24(10):E225-7.
  • 6. Räber L, Magro M, Stefanini GG, Kalesan B, van Domburg RT, Onuma Y, et al. Very late coronary stent thrombosis of a newer-generation everolimuseluting stent compared with early-generation drug-eluting stents: a prospective cohort study. Circulation. 2012;125(9):1110-21.
  • 7. Chen JP, Hou D, Pendyala L, Goudevenos JA, Kounis NG. Drug-eluting stent thrombosis: the Kounis hypersensitivity-associated acute coronary syndrome revisited. JACC Cardiovasc Interv. 2009;2(7):583-93.
  • 8. Siddiqi OK, Faxon DP. Very late stent thrombosis: current concepts. Curr Opin Cardiol. 2012;27(6):634-41.
  • 9. Windecker S, Serruys PW, Wandel S, Buszman P, Trznadel S, Linke A, et al. Biolimus-eluting stent with biodegradable polymer versus sirolimus-eluting stent with durable polymer for coronary revascularisation (LEADERS): a randomized non-inferiority trial. Lancet. 2008;372(9644):1163-73.
  • 10. Cutlip DE, Windecker S, Mehran R, Boam A, Cohen DJ, van Es GA, et al. Academy Research Consortium. Clinical end points in coronary stent trials. A case for standardized definitions. Circulation. 2007;115(17):2344-51.
  • 11. Grube E, Hauptmann KE, Buellesfeld L, Lim V, Abizaid A. Six-month results of a randomized study to evaluate safety and efficacy of a Biolimus A9 eluting stent with a biodegradable polymer coating. EuroIntervention. 2005;1(1):53-7.
  • 12. Stefanini GG, Kalesan B, Serruys PW, Heg D, Buszman P, Linke A, et al. Long-term clinical outcomes of biodegradable polymer biolimus-eluting stents versus durable polymer sirolimus-eluting stents in patients with coronary artery disease (LEADERS): 4 year follow-up of a randomised noninferiority trial. Lancet. 2011;378(9807):1940-8.
  • 13. Barlis P, Regar E, Serruys PW, Dimopoulos K, van der Giessen WJ, van Geuns RJ, et al. An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-study. Eur Heart J. 2010;31(2):165-76.
  • 14. Wykrzykowska JJ, Garg S, Onuma Y, de Vries T, Morel MA, van Es GA, et al. Implantation of the biodegradable polymer biolimus-eluting stent in patients with high SYNTAX score is associated with decreased cardiac mortality compared to a permanent polymer sirolimus-eluting stent: two year follow-up results from the "all-comers" LEADERS trial. EuroIntervention. 2011;7(5):605-13.
  • 15. Wykrzykowska JJ, Serruys PW, Onuma Y, de Vries T, van Es GA, Buszman P, et al. Impact of vessel size on angiographic and clinical outcomes of revascularization with biolimus-eluting stent with biodegradable polymer and sirolimus-eluting stent with durable polymer the LEADERS trial substudy. JACC CardiovascInterv. 2009;2(9):861-70.
  • 16. Raber L, Kelbaek H, Ostojic M, Baumbach A, Heg D, Tuller D, et al. Effect of biolimus-eluting stents with biodegradable polymer vs bare-metal stents on cardiovascular events among patients with acute myocardial infarction: the COMFORTABLE AMI randomized trial. JAMA. 2012;308(8)777-87.
  • 17. Soares Junior RS, Majeski JD, Carnieto NM, Zuliani MF, Lapa GA, Bavaresco SA, et al. RESULTADOS da intervenção coronária percutânea com o stent BiomatrixTM em pacientes não-selecionados. Rev Bras Cardiol Invasiva. 2011;19(1):40-6.
  • 18. Smits P, van Boven AJ, Valdes M, Serra A, Goy JJ, Voudris V, et al. TCT573 Randomized comparison of Biolimus-eluting (Nobori) and Everolimuseluting (Xience/Promus) stents in patients with multivessel coronary artery disease: 12-month follow-up data from COMPARE II study. J Am Coll Cardiol. 2012;60(17_S).
  • Autor correspondente:

    Adriano Caixeta
    Hospital Israelita Albert Einstein, Cardiologia Intervencionista
    Avenida Albert Einstein, 627/701 − Morumbi
    CEP: 05652-900 − São Paulo, SP, Brasil
    Tel.: (11) 2151-0434
    E-mail:
  • Publication Dates

    • Publication in this collection
      15 Oct 2013
    • Date of issue
      Sept 2013

    History

    • Received
      18 Jan 2013
    • Accepted
      13 Aug 2013
    Instituto Israelita de Ensino e Pesquisa Albert Einstein Avenida Albert Einstein, 627/701 , 05651-901 São Paulo - SP, Tel.: (55 11) 2151 0904 - São Paulo - SP - Brazil
    E-mail: revista@einstein.br