Influence of assessment site in measuring transcutaneous bilirubin

Influência do local de avaliação na mensuração da bilirrubina transcutânea

Abstracts

Objective

: To investigate the influence of the site of measurement of transcutaneous bilirubin (forehead or sternum) in reproducibility of results as compared to plasma bilirubin.

Methods

: A cohort study including 58 term newborns with no hemolytic disease. Transcutaneous measurements were performed on the forehead (halfway between the headline and the glabella, from the left toward the right side, making consecutive determinations, one-centimeter apart) and the sternum (five measurements, from the suprasternal notch to the xiphoid process with consecutive determinations, one-centimeter apart) using Bilicheck® (SpectRx Inc, Norcross, Georgia, USA). The correlation and agreement between both methods and plasma bilirubin were calculated.

Results

: There was a strong linear correlation between both determinations of serum bilirubin at the forehead and sternum (r=0.704; p<0.01 and r=0.653; p<0.01, respectively). There was correspondence of the mean values of transcutaneous bilirubin measured on the sternum (9.9±2.2mg/dL) compared to plasma levels (10.2±1.7mg/dL), but both differ from the values measured on the forehead (8.6±2.0mg/dL), p<0.05.

Conclusion

: In newborn term infants with no hemolytic disease, measuring of transcutaneous bilirubin on the sternum had higher accuracy as compared to serum bilirubin measurement on the forehead.

Bilirubin; Jaundice, neonatal/diagnosis; Infant, newborn


Objetivo

: Verificar a influência do local da mensuração da bilirrubina transcutânea (fronte ou esterno) na reprodutibilidade dos resultados, quando comparada à bilirrubina plasmática.

Métodos

: Estudo tipo coorte incluindo 58 recém-nascidos a termo sem doença hemolítica. Foram realizadas determinações transcutâneas na fronte (a meia distância entre a raiz do couro cabeludo e a glabela, iniciando do lado esquerdo em direção ao direito, realizando determinações consecutivas, separadas por 1cm) e no esterno (cinco mensurações iniciando da fúrcula até o apêndice xifoide, com determinações consecutivas, separadas por 1cm), utilizando o equipamento Bilicheck® (SpectRx Inc, Norcross, Geórgia, Estados Unidos). Foram calculadas as correlações e a concordância entre ambos os métodos, e a bilirrubina plasmática.

Resultados

: Foi encontrada forte correlação linear tanto entre a determinação na fronte quanto no esterno, com nível sérico de bilirrubina (r=0,704; p<0,01 e r=0,653; p<0,01, respectivamente). Houve correspondência dos valores médios de bilirrubina transcutânea aferidos no esterno (9,9±2,2mg/dL) com os valores plasmáticos (10,2±1,7mg/dL), porém ambos diferiram dos valores medidos na fronte (8,6±2,0mg/dL), com p<0,05.

Conclusão

: Em recém-nascidos a termo sem doença hemolítica, a mensuração da bilirrubina transcutânea realizada no esterno apresenta maior acurácia em relação à mensuração na fronte, quando comparada à bilirrubina sérica.

Bilirrubina; Icterícia neonatal/diagnóstico; Recém-nascido


INTRODUCTION

Most newborn (NB) infants develop jaundice in the first week of life. It occurs in up to 92% of term and late preterm NB infants.(11. Keren R, Tremont K, Luan X, Cnaan A. Visual assessment of jaundice in term and late preterm infants. Arch Dis Child Fetal Neonatal Ed. 2009;94(5):F317-22.) The occurrence of high serum bilirubin levels for prolonged time may permanently damage structures of the Central Nervous System, such as the globus pallidus, subthalamic nuclei, hippocampus, and oculomotor nucleus, among others, leading to kernicterus.(22. Ip S, Chung M, Kulig J, O’Brien R, Sege R, Glicken S, et al. American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. An evidence-based review of important issues concerning neonatal hyperbilirubinemia. Pediatrics. 2004;114(1):e130-53.)

The indication of phototherapy for treating neonatal jaundice will depend on serum bilirubin levels, presence of blood incompatibility, weight, and chronological and gestational ages, in addition to associated comorbidities.(33. Facchini FP, Mezzacappa MA, Rosa IR, Mezzacappa Filho F, Netto AA, Marba ST. Acompanhamento da icterícia neonatal em recém-nascidos de termo e prematuros tardios. J Pediatr (Rio J). 2007;83(4):313-8.) Hence the American Academy of Pediatrics recommends that every NB have its bilirubin level measured before hospital discharge and this measurement should be repeated on the first days after discharge.(44. American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics. 2004;114(1):297-316.)

Invasive bilirubin dosing demands drawing blood, with many inconveniences, such as technical difficulties in venous puncture, delay to obtain results, discomfort caused by pain(55. Facchini FP. Avaliação da variabilidade da dosagem transcutânea de bilirrubina pelo Bilicheck em neonatos. Rev Paul Pediatria. 2006;24(2):149-54.,66. Boo NY, Ishak S. Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer. J Paediatr Child Health. 2007;43(4):297-302.), and parental stress(77. Elias LS, Guinsburg R, Peres CA, Balda RC, Santos AM. Disagreement between parents and health professionals regarding pain intensity in critically ill neonates. J Pediatr (Rio J). 2008;84(1):35-40.), so it is important to minimize not only the amount of blood the NB loses in blood draws, but also to reduce to number of draws to the minimum possible.(88. Ebbesen F, Rasmussen LM, Wimberley PD. A new transcutaneous bilirubinometer, Bilicheck, used in the neonatal intensive care unit and the maternity ward. Acta Pediatr. 2002;91(2):203-11.)

In that sense, in the early 1980’s non-invasive (transcutaneous) techniques were developed to measure bilirubin and minimize the inconvenience of blood draws. The first equipment developed only correlated the intensity of yellow skin color with bilirubinemia, suffering the interference of many factors, like the amount of melanin, hemoglobin and connective tissue.(55. Facchini FP. Avaliação da variabilidade da dosagem transcutânea de bilirrubina pelo Bilicheck em neonatos. Rev Paul Pediatria. 2006;24(2):149-54.)

In the past few years, a new generation of devices for the transcutaneous measurement of bilirubin has been produced(66. Boo NY, Ishak S. Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer. J Paediatr Child Health. 2007;43(4):297-302.). They differ from previous models for being based on microspectrometry, which enables determining the optical density of bilirubin, hemoglobin and melanin in the subcutaneous layer of the NB infant skin. Excluding the factors that interfere in the determination of bilirubin leads to measuring its optical density in the subcutaneous capillaries and tissues with greater accuracy,(66. Boo NY, Ishak S. Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer. J Paediatr Child Health. 2007;43(4):297-302.) thus enabling the replacement of plasma measurements for transcutaneous measurements.(99. Lodha R, Dorari AK, Jatana V, Paul VK. Non-invasive estimation of total serum bilirrubin by multi-wavelength spectral reflectance in neonates. Indian Pediatr. 2000;37(7):771-5.) This technique has been studied in our context, and a good correlation between transcutaneous and serum bilirubin(1010. Bertini G, Rubaltelli FF. Non invasive bilirubinometry in neonatal jaundice. Semin. Neonatol. 2002;7(2):129-33.) has been found even in a multiracial population.(1111. Raimondi F, Lama S, Landolfo F, Sellitto M, Borrelli AC, Maffucci R, et al. Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population. BMC Pediatr. 2012;12:70.) Recently, a meta-analysis gathered the results of 21 studies comparing transcutaneous bilirubin to serum bilirubin in preterm infants, confirming the accuracy of this technique also in this NB population.(1212. Nagar G, Vandermeer B, Campbell S, Kumar M. Reliability of Transcutaneous Bilirubin Devices in Preterm Infants: A Systematic Review. Pediatrics. 2013;132(5):871-81. Review.)

On the other hand, bilirubin measurement may suffer the influence of the site of measurement: forehead or sternum.(1313. El-Beshbishi SN, Shattuck KE, Mohammad AA, Petersen JR. Hyperbilirubinemia and Transcutaneous Bilirubinometry. Clin Chem. 2009;55(7):1280-7.) The literature presents conflicting results, and it was demonstrated in term NB that transcutaneous measurements on the forehead and sternum are equivalent.(1010. Bertini G, Rubaltelli FF. Non invasive bilirubinometry in neonatal jaundice. Semin. Neonatol. 2002;7(2):129-33.,1414. Leite MG, Granato V, Facchini FP, Marba ST. Comparação entre a dosagem transcutânea e plasmática de bilirrubina. J Pediatr (Rio J). 2007; 83(3):283-6.

15. Poland RL, Hartenberger C, McHenry H, Hsi A. Comparison of skin sites for estimating serum total bilirubin in in-patients and out- patients: chest is superior to brow. J Perinatol. 2004;24(9):541-3.
-1616. Povaluk P, Shwetz EA, Kliemann R. Comparative study between plasma and transcutaneous bilirubin measurements in newborns. Rev Paul Pediatr. 2011;29(1):6-12.) Moreover, sternal measurement results in higher bilirubin levels than on the forehead.(1717. Rubaltelli F, Gourley G, Loskamp N. Transcutaneous Bilirubin Measurement: A multicenter evaluation of a new device. Pediatrics. 2001;107(6):1264-71.

18. Maisels MJ, Ostrea EM Jr, Touch S, Clune SE, Cepeda E, Kring E, et al. Evaluation of a new transcutaneous bilirubinometer. Pediatrics. 2004;113(6):1628-35.
-1919. Wong CM, van Dijk PJ, Laing IA. A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields. Arch Dis Child Fetal Neonatal Ed. 2002;87 (2):F137-40.)

OBJECTIVE

To verify the influence of the measurement site of transcutaneous bilirubin, forehead or sternum, in the reproducibility of results, as compared to plasma bilirubin.

METHODS

After approval by the Research Ethics Committee (REF CEP/Einstein 08/896), a prospective cohort study was conducted including healthy term infants born at the Hospital Israelita Albert Einstein, a private tertiary-care hospital in the city of São Paulo (SP), in the period between April to September 2009. NB with gestational age ≥37 weeks with <72 hours of life were included. Newborn infants with hemolytic disease, with skin abnormalities or previous phototherapy treatment were excluded. Jaundice due to hemolytic disease was defined as that with early onset (in the first 48 hours of life), or with laboratory values that were incompatible with physiological jaundice (occurrence of reticulocytosis, positive Coombs or eluate tests). After obtaining the informed consent from parents or legal guardian, NB infants with clinical indication of plasma bilirubin measurement, according to the routine assessment by the neonatologist, had bilirubin measured transcutaneously too.

Immediately after the collection of plasma bilirubin, always the same researcher measured transcutaneous bilirubin on the forehead and on the sternum, using the Bilicheck® device (SpectRx Inc, Norcross, Georgia, USA). The device was calibrated before each measurement, according to the manufacturer’s instructions to ensure measurement accuracy.(2020. Children’s Medical Ventures. Respironics. BiliChek® Noninvasive bilirubin analyzer: suggested usage protocol [Internet]. 2006 [cited 2014 Feb 7]. Available from: http://www.olusummedikal.com/bili/33.pdf
http://www.olusummedikal.com/bili/33.pdf...
) For each measurement, the device was positioned on the infant’s skin, and five individual measurements in different points led to one result. On the forehead, the five measurements were taken halfway the hairline and glabella, starting on the left towards right side, one-centimeter apart. On the sternum, five measurements were taken, starting on the suprasternal notch to the xiphoid process, with consecutive one-centimeter apart determinations.

The statistical analysis was performed through the calculation of Pearson’s correlation coefficient for both transcutaneous measurements compared to plasma measurements, and then Bland-Altman charts were plotted to assess agreement, and, finally, simple variance analysis (one-way ANOVA) to compare means, with Student-Newman-Keuls as a discriminatory post-test. The calculated size of the sample was 50 NB, considering the difference between means to be detected as 1.0mg/dL, with an expected standard deviation of 2.4mg/dL(1717. Rubaltelli F, Gourley G, Loskamp N. Transcutaneous Bilirubin Measurement: A multicenter evaluation of a new device. Pediatrics. 2001;107(6):1264-71.), 0.80 power test, and 0.05 significance level. For the statistical analysis, Sigma Stat software, version 2.0 was used.

RESULTS

A total of 58 NB were studied, with birth weight of 3221±402g (mean±standard deviation) gestational age of 38.4±1.2 weeks, one-minute Apgar score of 9.0±0.3 and five-minute Apgar score of 10.0±0.0. All NB infants were Caucasian, and measurements were taken with 1.8±0.9 days of life. A total of 94.8% of the NB included were born from C-sections and 5.2% from vaginal deliveries. A good linear correlation was observed both between transcutaneous bilirubin measurements on the forehead and serum levels (r=0.704; p<0.01), as well as between transcutaneous measurements on the sternum and serum levels (r=0.653; p<0.01) (Figures 1 and 2). Differences in the results between transcutaneous measurements on the forehead and sternum and total serum bilirubin are shown in figures 3A and 3B. The mean difference between transcutaneous measurements from the sternum and plasma bilirubin was 0.3mg/dL, number below that found in the difference between measurements taken on the forehead and corresponding plasma bilirubin levels (1.6mg/dL). The comparison between the means of bilirubin values found on the forehead, sternum, as well as the corresponding serum bilirubin is shown in figure 4. There was a correspondence of the values measured on the sternum with plasma values, but both differed from the values measured on the forehead (p<0.05).

Figure 1
. Linear correlation between transcutaneous measurement on the forehead and serum total bilirubin levels

Figure 2
. Linear correlation between transcutaneous measurement on the sternum and serum total bilirubin level

Figure 3
. Bland-Altman charts plotted with the differences in the results between transcutaneous measurements on the sternum and (A), and transcutaneous measurements taken on the forehead and plasma bilirubin (B)

Figure 4
. Bilirubin levels measured on the forehead, sternum and plasma. Values as mean±standard deviation (DP)

DISCUSSION

This study had the objective of verifying in term NB with no hemolytic disease the influence of transcutaneous bilirubin measurement site in accuracy of results. The principal contribution of this study was to provide, with data obtained in our environment, information about an issue that the international literature shows non-homogeneous, and sometimes conflicting, results.

Accuracy in the measurement of transcutaneous bilirubin in relation to serum bilirubin has been recently demonstrated in a meta-analysis gathering data from 3527 patients published in 21 studies. In 16 of those studies, measurements were taken on the forehead, in 10 on the sternum, and in 3 on the abdomen.(1212. Nagar G, Vandermeer B, Campbell S, Kumar M. Reliability of Transcutaneous Bilirubin Devices in Preterm Infants: A Systematic Review. Pediatrics. 2013;132(5):871-81. Review.) Our data demonstrate that measurements taken on the sternum have a good correlation with serum measurements, unlike measurements taken on the forehead. Possibly as a result of the continuous exposure to room light, bilirubin measurements in areas that are not covered by clothes, like the face, may present lower bilirubin values. Many authors tried to relate the measurement site of transcutaneous bilirubin (forehead, sternum, dorsum, knee, or foot) with accuracy of results,(1313. El-Beshbishi SN, Shattuck KE, Mohammad AA, Petersen JR. Hyperbilirubinemia and Transcutaneous Bilirubinometry. Clin Chem. 2009;55(7):1280-7.) and the measurements taken on the forehead and sternum presented the best correlations with serum bilirubin.(66. Boo NY, Ishak S. Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer. J Paediatr Child Health. 2007;43(4):297-302.,1717. Rubaltelli F, Gourley G, Loskamp N. Transcutaneous Bilirubin Measurement: A multicenter evaluation of a new device. Pediatrics. 2001;107(6):1264-71.,2121. Tan KL, Dong F. Transcutaneous bilirubinometry during and after phototherapy. Acta Paediatr. 2003;92(3):327-31.,2222. Randeberg LL, Roll EB, Nilsen LT, Christensen T, Svaasand LO. In vivo spectroscopy of jaundiced newborn skin reveals more than a bilirubin index. Acta Paediatr. 2005;94(1):65-71.) Like the results of this study, Maisels et al. found a better correlation with serum bilirubin when transcutaneous measurements were taken on the sternum (r=0.953) as compared to measurements on the forehead (r=0,914).(1818. Maisels MJ, Ostrea EM Jr, Touch S, Clune SE, Cepeda E, Kring E, et al. Evaluation of a new transcutaneous bilirubinometer. Pediatrics. 2004;113(6):1628-35.) Similarly to our results, a revision published in 2009, including 13 studies addressing the influence of measurement site on the results of transcutaneous bilirubin, concluded that the sternum presents good correlation with serum bilirubin. However, in six studies of this revision, no differences were observed between measurements taken on the forehead and sternum, and, in two studies, measurements taken on the forehead were more reliable than those taken on the sternum.(1313. El-Beshbishi SN, Shattuck KE, Mohammad AA, Petersen JR. Hyperbilirubinemia and Transcutaneous Bilirubinometry. Clin Chem. 2009;55(7):1280-7.)

In another study, transcutaneous measurement of bilirubin taken on the forehead suffered the influence of crying, and the lowest values were found in the NB who were crying at the time of measurement.(2323. Kazmierczak S, Bhutani V, Gourley G, Kerr S, Lo S, Robertson A, et al. Transcutaneous bilirubin testing. In: Nichols JH, editors. Evidence-Based Practice for Point-of-Care Testing - The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines. Washington; 2006. p.5-12.) In our study, NB infants were not crying upon measurement.

In disagreement with this study, Bertini and Rubaltelli demonstrated that the precision of transcutaneous measurements, when taken on the forehead and sternum, are comparable, but sternum measurements are, in average, 0.8 to 0.9mg/dL higher.(1010. Bertini G, Rubaltelli FF. Non invasive bilirubinometry in neonatal jaundice. Semin. Neonatol. 2002;7(2):129-33.) Likewise, the average of transcutaneous bilirubin measurements taken on the trunk was demonstrated to be 0.4mg/dL higher than serum measurements, whereas measurements on the forehead were 0.3mg/dL (5mol/L) smaller than serum measurements. The authors concluded that plasma measurements and transcutaneous measurements showed approximate values, but after hospital discharge, forehead measurements underestimated values by 5%. The authors recommend using trunk measurements for bilirubinemia.(1515. Poland RL, Hartenberger C, McHenry H, Hsi A. Comparison of skin sites for estimating serum total bilirubin in in-patients and out- patients: chest is superior to brow. J Perinatol. 2004;24(9):541-3.)

Similar results to this study were found in a group of 345 NB, in which a better correlation between blood bilirubin and transcutaneous measurements taken on the sternum than on the forehead were found.(1919. Wong CM, van Dijk PJ, Laing IA. A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields. Arch Dis Child Fetal Neonatal Ed. 2002;87 (2):F137-40.) Even though accuracy is similar for measurements taken on the frontal region or on the sternum, correlation is greater on the latter, possibly due to the head’s exposure to room light.(2424. Quintas C, Silva A. Icterícia neonatal [Internet]. In: Consensos Nacionais em Neonatologia. Seção de Neonatologia. Coimbra: Sociedade Portuguesa de Pediatria[citado 2013 Nov 12]. Disponível em: http://www.spp.pt/UserFiles/File/Consensos_ Nacionais_Neonatologia_ 2004/Ictericia_Neonatal.pdf
http://www.spp.pt/UserFiles/File/Consens...
)

Disagreeing with the findings of this study, in our environment, a group of 44 NB, with average gestational age of 35.1±3.4 weeks and average birth weight of 2151±889g, 73% of them Caucasian, was analyzed between the second and third days of life. The authors did not find differences between serum bilirubin levels and transcutaneous bilirubin levels measured on covered areas of the forehead and sternum 24 hours after the beginning of phototherapy.(1616. Povaluk P, Shwetz EA, Kliemann R. Comparative study between plasma and transcutaneous bilirubin measurements in newborns. Rev Paul Pediatr. 2011;29(1):6-12.)

This study presents some weaknesses, including its relatively small sample formed exclusively by term Caucasian NB with no hemolytic disease. The inclusion of only Caucasian NB infants was unintentional, at the time of patient recruitment, which resulted in a sample that is different from that of the Brazilian population. But this fact does not compromise significantly our conclusions. It is important, however, to highlight that the influence of the measurement site of transcutaneous bilirubin (forehead or sternum) on serum bilirubin should be also assessed in preterm infants of other races and with hemolytic disease, before generalizing these conclusions to these groups.

CONCLUSION

In Caucasian term NB without hemolytic disease, transcutaneous bilirubin measurement taken on the sternum presents greater accuracy than forehead measurements, when compared to serum bilirubin.

REFERENCES

  • 1
    Keren R, Tremont K, Luan X, Cnaan A. Visual assessment of jaundice in term and late preterm infants. Arch Dis Child Fetal Neonatal Ed. 2009;94(5):F317-22.
  • 2
    Ip S, Chung M, Kulig J, O’Brien R, Sege R, Glicken S, et al. American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. An evidence-based review of important issues concerning neonatal hyperbilirubinemia. Pediatrics. 2004;114(1):e130-53.
  • 3
    Facchini FP, Mezzacappa MA, Rosa IR, Mezzacappa Filho F, Netto AA, Marba ST. Acompanhamento da icterícia neonatal em recém-nascidos de termo e prematuros tardios. J Pediatr (Rio J). 2007;83(4):313-8.
  • 4
    American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics. 2004;114(1):297-316.
  • 5
    Facchini FP. Avaliação da variabilidade da dosagem transcutânea de bilirrubina pelo Bilicheck em neonatos. Rev Paul Pediatria. 2006;24(2):149-54.
  • 6
    Boo NY, Ishak S. Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer. J Paediatr Child Health. 2007;43(4):297-302.
  • 7
    Elias LS, Guinsburg R, Peres CA, Balda RC, Santos AM. Disagreement between parents and health professionals regarding pain intensity in critically ill neonates. J Pediatr (Rio J). 2008;84(1):35-40.
  • 8
    Ebbesen F, Rasmussen LM, Wimberley PD. A new transcutaneous bilirubinometer, Bilicheck, used in the neonatal intensive care unit and the maternity ward. Acta Pediatr. 2002;91(2):203-11.
  • 9
    Lodha R, Dorari AK, Jatana V, Paul VK. Non-invasive estimation of total serum bilirrubin by multi-wavelength spectral reflectance in neonates. Indian Pediatr. 2000;37(7):771-5.
  • 10
    Bertini G, Rubaltelli FF. Non invasive bilirubinometry in neonatal jaundice. Semin. Neonatol. 2002;7(2):129-33.
  • 11
    Raimondi F, Lama S, Landolfo F, Sellitto M, Borrelli AC, Maffucci R, et al. Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population. BMC Pediatr. 2012;12:70.
  • 12
    Nagar G, Vandermeer B, Campbell S, Kumar M. Reliability of Transcutaneous Bilirubin Devices in Preterm Infants: A Systematic Review. Pediatrics. 2013;132(5):871-81. Review.
  • 13
    El-Beshbishi SN, Shattuck KE, Mohammad AA, Petersen JR. Hyperbilirubinemia and Transcutaneous Bilirubinometry. Clin Chem. 2009;55(7):1280-7.
  • 14
    Leite MG, Granato V, Facchini FP, Marba ST. Comparação entre a dosagem transcutânea e plasmática de bilirrubina. J Pediatr (Rio J). 2007; 83(3):283-6.
  • 15
    Poland RL, Hartenberger C, McHenry H, Hsi A. Comparison of skin sites for estimating serum total bilirubin in in-patients and out- patients: chest is superior to brow. J Perinatol. 2004;24(9):541-3.
  • 16
    Povaluk P, Shwetz EA, Kliemann R. Comparative study between plasma and transcutaneous bilirubin measurements in newborns. Rev Paul Pediatr. 2011;29(1):6-12.
  • 17
    Rubaltelli F, Gourley G, Loskamp N. Transcutaneous Bilirubin Measurement: A multicenter evaluation of a new device. Pediatrics. 2001;107(6):1264-71.
  • 18
    Maisels MJ, Ostrea EM Jr, Touch S, Clune SE, Cepeda E, Kring E, et al. Evaluation of a new transcutaneous bilirubinometer. Pediatrics. 2004;113(6):1628-35.
  • 19
    Wong CM, van Dijk PJ, Laing IA. A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields. Arch Dis Child Fetal Neonatal Ed. 2002;87 (2):F137-40.
  • 20
    Children’s Medical Ventures. Respironics. BiliChek® Noninvasive bilirubin analyzer: suggested usage protocol [Internet]. 2006 [cited 2014 Feb 7]. Available from: http://www.olusummedikal.com/bili/33.pdf
    » http://www.olusummedikal.com/bili/33.pdf
  • 21
    Tan KL, Dong F. Transcutaneous bilirubinometry during and after phototherapy. Acta Paediatr. 2003;92(3):327-31.
  • 22
    Randeberg LL, Roll EB, Nilsen LT, Christensen T, Svaasand LO. In vivo spectroscopy of jaundiced newborn skin reveals more than a bilirubin index. Acta Paediatr. 2005;94(1):65-71.
  • 23
    Kazmierczak S, Bhutani V, Gourley G, Kerr S, Lo S, Robertson A, et al. Transcutaneous bilirubin testing. In: Nichols JH, editors. Evidence-Based Practice for Point-of-Care Testing - The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines. Washington; 2006. p.5-12.
  • 24
    Quintas C, Silva A. Icterícia neonatal [Internet]. In: Consensos Nacionais em Neonatologia. Seção de Neonatologia. Coimbra: Sociedade Portuguesa de Pediatria[citado 2013 Nov 12]. Disponível em: http://www.spp.pt/UserFiles/File/Consensos_ Nacionais_Neonatologia_ 2004/Ictericia_Neonatal.pdf
    » http://www.spp.pt/UserFiles/File/Consensos_ Nacionais_Neonatologia_ 2004/Ictericia_Neonatal.pdf

Publication Dates

  • Publication in this collection
    Jan-Mar 2014

History

  • Received
    4 Feb 2013
  • Accepted
    2 Dec 2013
Instituto Israelita de Ensino e Pesquisa Albert Einstein Avenida Albert Einstein, 627/701 , 05651-901 São Paulo - SP, Tel.: (55 11) 2151 0904 - São Paulo - SP - Brazil
E-mail: revista@einstein.br