Discrepancies in degree of conversion measurements by FTIR

Abstract

The purpose of this study was to evaluate the influence of the internal standard peak on the measured degree of conversion (DC) for methacrylate-based resins analyzed by Fourier-transform infrared spectroscopy (FTIR). The influence of different baseline measurement methods was also determined. Different blends were prepared, mixing BisGMA/BisEMA and BisGMA/TEGDMA at proportions of 0% to 100%, in weight. Camphoroquinone and ethyl-4-dimethylamino-benzoate were used to permit the light activation of blends. Degree of conversion was evaluated using FTIR equipped with an attenuated total reflectance (ATR) device. Samples were dispensed onto ATR crystal and light activated for 40 s. The DC was calculated by relating the height of the peak 1637 cm-1 to the intensity of different internal standard peaks (1715, 1608, or 1582 cm-1), measured using two different baseline methods, before and after polymerization. Data were compared by ANOVA at 5% significance. The relationship between DC and monomer ratio was obtained by regression analysis. Double-bond conversion ranged from 32.75% to 78.50% for BisGMA/BisEMA blends, and from 32.75% to 76.22% for BisGMA/TEGDMA blends. For the BisGMA/BisEMA blends, the DC showed a linear association with the composition of the comonomer blends, independent of the internal standard peak and baseline method used. In contrast, the trends in DC for BisGMA/TEGDMA blends were different for each method of measurement. The internal standard peaks and measurement baseline should be taken into account when using FTIR to calculate the DC of methacrylate-based resins, especially comonomer blends containing a high degree of monomers that lack aromatic rings.

Polymerization; Bisphenol A-Glycidyl Methacrylate; Dental Materials


Introduction

The degree of conversion (DC) of polymerizable resins has been widely investigated in the dental literature.11. Porto IC, Soares LE, Martin AA, Cavalli V, Liporoni PC. Influence of the photoinitiator system and light photoactivation units on the degree of conversion of dental composites. Braz Oral Res. 2010 Oct-Dec;24(4):475-81.

2. Collares FM, Ogliari FA, Zanchi CH, Petzhold CL, Piva E, Samuel SM. Influence of 2-hydroxyethyl methacrylate concentration on polymer network of adhesive resin. J Adhes Dent. 2011 Apr;13(2):125-9.

3. Zhang Y, Wang Y. Distinct photopolymerization efficacy on dentin of self-etch adhesives. J Dent Res. 2012 Aug;91(8):795-9.

4. Wu N, Zhang Y, Wang Y. Photo-polymerization efficiency of self-etch dental adhesives composed of camphorquinone or trimethylbenzoyl-diphenyl-phosphine oxide. Int J Adhes Adhes. 2013 Sep;45:53-8.
-55. Marovic D, Panduric V, Tarle Z, Ristic M, Sariri K, Demoli N, et al. Degree of conversion and microhardness of dental composite resin materials. J Mol Struct. 2013 Jul;1044(24):299-302. A high degree of double-bond conversion of methacrylate-based resins yields favorable mechanical properties66. Ferracane JL. Hygroscopic and hydrolytic effects in dental polymer networks. Dent Mater. 2006 Mar;22(3):211-22.,77. Palin WM, Fleming GJ, Burke FJ, Marquis PM, Randall RC. Monomer conversion versus flexure strength of a novel dental composite. J Dent. 2003 Jul;31(5):341-51. and reduced degradation of the polymer.66. Ferracane JL. Hygroscopic and hydrolytic effects in dental polymer networks. Dent Mater. 2006 Mar;22(3):211-22. Despite the high DC achieved in dental resins, complete conversion of aliphatic carbon-carbon double bonds is typically not achieved. The unconverted double bonds remaining in the resin arise from at least two situations, namely, unreacted monomers or pendant double bonds at the extremities of polymer chains. Unreacted monomers in the polymer network may irritate surrounding tissues.88. Durner J, Obermaier J, Draenert M, Ilie N. Correlation of the degree of conversion with the amount of elutable substances in nano-hybrid dental composites. Dent Mater. 2012 Nov;28(11):1146-53.

9. Mantellini MG, Botero TM, Yaman P, Dennison JB, Hanks CT, Nor JE. Adhesive resin induces apoptosis and cell-cycle arrest of pulp cells. J Dent Res. 2003 Aug;82(8):592-6.
-1010. Pashley DH, Zhang Y, Agee KA, Rouse CJ, Carvalho RM, Russell CM. Permeability of demineralized dentin to HEMA. Dent Mater. 2000 Jan;16(1):7-14. Furthermore, low DC may render the polymer more prone to oxidation and hydrolytic degradation over time,66. Ferracane JL. Hygroscopic and hydrolytic effects in dental polymer networks. Dent Mater. 2006 Mar;22(3):211-22. decreasing the success of restorative procedures.

One of the most common methods to determine the extent of double-bond conversion is Fourier-transform infrared spectroscopy (FTIR),22. Collares FM, Ogliari FA, Zanchi CH, Petzhold CL, Piva E, Samuel SM. Influence of 2-hydroxyethyl methacrylate concentration on polymer network of adhesive resin. J Adhes Dent. 2011 Apr;13(2):125-9.,1111. Rueggeberg FA, Hashinger DT, Fairhurst CW. Calibration of FTIR conversion analysis of contemporary dental resin composites. Dent Mater. 1990 Oct;6(4):241-9. which can detect the stretching vibrations of carbon-carbon double bonds involved in polymerization.1111. Rueggeberg FA, Hashinger DT, Fairhurst CW. Calibration of FTIR conversion analysis of contemporary dental resin composites. Dent Mater. 1990 Oct;6(4):241-9. The method typically utilizes the height ratio of the peaks corresponding to aliphatic (1640 cm-1) and aromatic (1610 cm-1) double bonds22. Collares FM, Ogliari FA, Zanchi CH, Petzhold CL, Piva E, Samuel SM. Influence of 2-hydroxyethyl methacrylate concentration on polymer network of adhesive resin. J Adhes Dent. 2011 Apr;13(2):125-9. to determine the DC. However, not all dental methacrylate-based polymers contain aromatic rings. Moreover, currently available dental polymer blends present different monomer compositions compared to each other (i.e., BisGMA, TEGDMA, BisEMA and UDMA) and ratios.1212. Jia W, Jin S, inventors; Pentron Clinical Technologies, LLC., assignee. Dental resin materials, method of manufacture, and uses thereof. United States patent US 6,787,629 B2. 2004 Sep 7 .

13. Salz U, Zimmermann J, Poppe D, Rheinberger V, Tiller J, Waschinski C, inventors; Ivoclar Vivadent AG, assignee. Dental materials based on radically polymerizable macromers with antimicrobial effect. United States patent US 7,553,881 B2. 2009 Jun 30.

14. Rao PS, Aasen SM, Abuelyaman AS, Mitra SB, Craig BD, Shukla BA, Plaut DJ, Russel VA, inventors; 3M Innovative Properties Company, assignee. Adhesive composition for hard tissue. United States patent US 8,076,389 B2. 2011 Dec 13.
-1515. Jin X, inventor; Dentsply International Inc., assignee. Method and antibacterial/antimicrobial compositions in dental compositions. United States patent US 2012/0328553 A1. 2012 Dec 27. Because FTIR measurements of DC use the aromatic carbon-carbon double bond peak as internal standard, variations in comonomer blend composition and ratio can influence the results. FTIR spectra of methacrylate-based dental materials also present other peaks related to carbon bond stretching, at 1715 and 1580 cm-1. Therefore, it is of great interest to investigate whether these peaks may be used as internal standards in FTIR to determine the DC.

The purpose of this study was to evaluate the influence of the internal standard peak on the measured degree of conversion (DC) for methacrylate-based resins analyzed by FTIR. Two statistical null hypotheses were tested:

  1. 1

    that modifying the internal standards does not influence the DC of comonomer blends, and

  2. 2

    that the baseline used to measure the peak height does not influence the DC results

Methodology

Materials

Bisphenol-A glycol dimethacrylate (BisGMA) was blended with monomers of either ethoxylated bisphenol-A glycol dimethacrylate 6 (BisEMA) or triethylene glycol dimethacrylate (TEGDMA). Concentrations of BisEMA and TEGDMA in BisGMA were varied from 0% to 100% by weight. To each sample was added 1% mol of camphoroquinone, as a photosensitizer, and 1% mol of ethyl-4-dimethylamino-benzoate, as a reducing agent. All reagents were obtained from Esstech Inc., Essington, USA.

Degree of conversion analysis

The DC was evaluated by FTIR (Vertex 70, Bruker Optics, Ettlingen, Germany). The spectrometer was coupled to a horizontal attenuated total reflectance (ATR) device consisting of a diamond crystal of 2 mm in diameter (Platinum ATR-QL, Bruker Optics, Ettlingen, Germany). A mid-infrared detector (MIR ATR A225 Diamont, Bruker Optics, Ettlingen, Germany) was used to capture the reflected IR. A support device was used to affix the light-curing unit at a distance of 5 mm from the sample. A drop of each sample (n = 3), more than 1 mm thick and sufficiently wide to cover the entire surface of the diamond crystal, was directly dispensed onto the ATR crystal and photoactivated for 40 s by a light-emitting diode (Radii, SDI, Bayswater, Australia). The irradiance value was 1200 mW/cm², confirmed with a digital power meter (Ophir Optronics, North Andover, USA). Sixteen co-addition scans were made at 10 kHz velocity and 4 cm-1 resolution, with an instrument aperture of 6 mm. Absorbance spectra were obtained before and immediately after light polymerization using Opus software (Opus 6.5, Bruker Optics, Ettlingen, Germany), with Blackman-Harris 3-Term apodization over the range of 4000 to 400 cm-1. Spectra were transferred to IrfanView 4.25 software (Irfan Skiljan, Wiener Neustadt, Austria), and different baselines were traced onto them, according to Figure 1. The peak height was measured from the top of the peak to the baseline.

Figure 1
Schematic representation of the baseline methods proposed to determine the ratios of the absorbance peaks corresponding to aliphatic (1637 cm−1), aromatic (1608 and 1582 cm−1) or carbonyl (1715 cm−1) bonds. The peak heights were measured in relation to the baseline, determined as follows. (A) Baseline was traced by connecting the troughs of each peak. (B) Baseline was obtained by connecting the 1637 cm−1 initial trough and the 1608 cm−1 final trough.

The DC for each blend, polymerized and unpolymerized, was determined by the ratio of the absorbance peak corresponding to the aliphatic carbon-carbon double bond (1637 cm-1 peak height) with that of the internal standard (1715, 1608, and 1582 cm-1 peak height), according to methods described in a previous study.1111. Rueggeberg FA, Hashinger DT, Fairhurst CW. Calibration of FTIR conversion analysis of contemporary dental resin composites. Dent Mater. 1990 Oct;6(4):241-9.

Statistical analysis

The DC data of each comonomer blend was analyzed by one-way ANOVA (internal standard) and post-hoc Tukey test. The relationship between DC and comonomer blend ratio or internal standard was analyzed by linear regression, yielding a correlation of determination (R²). All analyses were conducted at a statistical significance level of 5%.

Results

Tables 1 and 2 show the DC of the polymer blends, measured by different methods. The baseline tracing (Figure 1) influenced the reported DC, as illustrated in Figures 1A and 1B, where the height measurement of the 1608 cm-1 internal standard peak was performed according to two different baseline tracing methods (1608 cm-1a and 1608 cm-1b). Double-bond conversion ranged from 32.75% to 78.50% for BisGMA/BisEMA blends, and from 32.75% to 76.22% for BisGMA/TEGDMA blends. Use of 1608 cm-1a as an internal standard yielded the highest reported values of DC for almost all analyses.

Table 1
Mean (standard deviation) DC (%) for BisGMA/ BisEMA blends, measured using different internal standard peaks.
Table 2
Mean (standard deviation) DC (%) for BisGMA/ TEGDMA blends, measured using different internal standard peaks.

BisGMA/BisEMA blends

Trends in DC for BisGMA/BisEMA blends were adequately fitted by linear regression (R² > 0.9; p < 0.001). Increasing the BisGMA/BisEMA ratio decreased the DC (Figure 2). Regardless of the internal standard peak or measurement baseline, the DC showed a strong association with the composition of the comonomer blend.

Figure 2
Linear regression analysis to obtain the relationship between DC and proportions of BisGMA/BisEMA blends.

BisGMA/TEGDMA blends

BisGMA/TEGDMA blends could not be characterized by linear regression, with the exception of the experiment using the 1608 cm-1a standard peak and baseline method (R² = 0.803; p < 0.001). This set of experiments was better fit by Gaussian 3-parameter regression (p1715 cm-1 < 0.001; p1608 cm-1b = 0.004; p1582 cm-1 = 0.041). Increasing the ratio of BisGMA/TEGDMA to 50/50 promoted an increase in DC. At BisGMA/TEGDMA ratios greater than 50/50, a decrease in DC was observed (Figure 3).

Figure 3
Regression analysis to obtain the relationship between DC and composition of BisGMA/TEDGMA blends. The fitting equations for each plot are presented in the legend.

Discussion

This study aimed to evaluate how the measurement of DC via FTIR is influenced by the internal standard peak and baseline method. Comonomer blends with or without aromatic rings were tested. The DC of the comonomer blends varied from 32.75% to 78.50% as function of monomer composition and ratio. These results indicate that, contrary to our two null hypotheses, the internal standard and the baseline do, in some cases, affect measurements of DC.

The DC of BisGMA/BisEMA blends varied from 32.75% to 78.50%, presenting an indirect linear relationship to BisGMA concentration for all internal standards. The decrease in DC at larger BisGMA/BisEMA ratios can be attributed to differences in viscosity. Increasing the content of BisGMA can decrease the mobility of monomers into the polymerizing bulk.1616. Andrzejewska E. Photopolymerization kinetics of multifunctional monomers. Prog Polym Sci. 2001 May;26(4):605-65. The results for these samples indicate that the blends composed of monomers with aromatic rings used in this study were not influenced by differences in the internal standard peak or baseline method.

The DC of BisGMA/TEGDMA blends ranged from 32.75% to 76.22%, presenting an indirect linear relationship to BisGMA concentration only for the 1608 cm-1a internal standard. Trends in DC for BisGMA/TEGDMA blends differed when the internal standard peak was varied from 1715 cm-1 to 1608 cm-1b and 1582 cm-1. In general, the DC exhibited an initial increase as the BisGMA content was increased to 50 wt%, followed by a decrease as the BisGMA content was increased further. The results for BisGMA/TEGDMA show that the internal standards and baseline do influence the measured DC of comonomer blends containing monomers without aromatic rings (i.e., TEGDMA).

Our results are in accordance with those of a previous study that showed an increase in DC upon decreasing the BisGMA content in a BisGMA/TEGDMA blend.1717. Lovell LG, Newman SM, Bowman CN. The effects of light intensity, temperature, and comonomer composition on the polymerization behavior of dimethacrylate dental resins. J Dent Res. 1999 Aug;78(8):1469-76. Thus, despite the large number of studies that determine DC via FTIR with the aromatic carbon-carbon peak (1608 cm-1) as the internal standard, caution must be taken when using this method to find the DC of blends that contain monomers without aromatic rings. In the FTIR spectra of resins comprising a high content of monomers without aromatic carbon-carbon double bonds, the peaks at 1608 and 1582 cm-1 have low or nonexistent intensity, decreasing the reliability and reproducibility of DC measurements on these samples. However, most commercial methacrylate-based dental materials do comprise monomers with aromatic rings; such monomers produce a stiff central core in the final polymer, increasing its mechanical strength.

Ideally, the DC of TEGDMA-rich blends (containing > 80% TEGDMA by weight) should be determined by other internal standard peaks besides 1608 cm-1 due to the low intensity of this peak in the FTIR spectrum of such blends. For these blends, the 1715 cm-1 peak, corresponding to the carbonyl C=O bond, showed reliable results. The lower DC values measured using the 1715 cm-1 internal standard in comparison to the 1608 cm-1a standard can be explained by the cyclization that occurs during TEGDMA polymerization. The TEGDMA molecule has a short and flexible chain that permits a methacrylate radical from one extremity to react with the methacrylate of another extremity, reducing the chain mobility and hindering the copolymerization of TEGDMA with BisGMA. This reaction forms an inhomogeneous polymer that presents a low DC.1818. Elliott JE, Lovell LG, Bowman CN. Primary cyclization in the polymerization of bis-GMA and TEGDMA: a modeling approach to understanding the cure of dental resins. Dent Mater. 2001 May;17(3):221-9.

Unexpectedly, the 1608 cm-1b internal standard yielded reasonable results for the DC of pure TEGDMA. This result should be viewed with caution, because this peak, theoretically, should not be present on the TEGDMA IR spectrum.

In general, for both blend compositions, the measured DC values followed the following order: 1608 cm-1a > 1582 cm-1 > 1608 cm-1b > 1715 cm-1. Using the 1608 cm-1 peak as internal standard, and determining its height by baseline method "a", appears to be the best way to evaluate DC, yielding results that are comparable to those of other studies.1919. Amirouche-Korichi A, Mouzali M, Watts DC. Effects of monomer ratios and highly radiopaque fillers on degree of conversion and shrinkage-strain of dental resin composites. Dent Mater. 2009 Nov;25(11):1411-8.,2020. Goncalves F, Azevedo CL, Ferracane JL, Braga RR. BisGMA/TEGDMA ratio and filler content effects on shrinkage stress. Dent Mater. 2011 Jun;27(6):520-6. Previous studies showed a direct linear proportional relationship between TEGDMA concentration and DC in dimethacrylate blends.2121. Sideridou I, Tserki V, Papanastasiou G. Effect of chemical structure on degree of conversion in light-cured dimethacrylate-based dental resins. Biomaterials. 2002 Apr;23(8):1819-29.

Conclusion

The internal standard peak and baseline measurement affect the measured DC of methacrylate-based resins and its relationship to the comonomer blend composition, as determined by FTIR spectroscopy. The best internal standard for DC measurement is the 1608 cm-1 peak, whose height is measured by baseline method "a", for the comonomer blends that were used in this study.

Fernando Freitas Portela gratefully acknowledges scholarship support from CAPES (Coordenação de Aperfeiçoamento de Pessoal de Nível Superior).

References

  • 1
    Porto IC, Soares LE, Martin AA, Cavalli V, Liporoni PC. Influence of the photoinitiator system and light photoactivation units on the degree of conversion of dental composites. Braz Oral Res. 2010 Oct-Dec;24(4):475-81.
  • 2
    Collares FM, Ogliari FA, Zanchi CH, Petzhold CL, Piva E, Samuel SM. Influence of 2-hydroxyethyl methacrylate concentration on polymer network of adhesive resin. J Adhes Dent. 2011 Apr;13(2):125-9.
  • 3
    Zhang Y, Wang Y. Distinct photopolymerization efficacy on dentin of self-etch adhesives. J Dent Res. 2012 Aug;91(8):795-9.
  • 4
    Wu N, Zhang Y, Wang Y. Photo-polymerization efficiency of self-etch dental adhesives composed of camphorquinone or trimethylbenzoyl-diphenyl-phosphine oxide. Int J Adhes Adhes. 2013 Sep;45:53-8.
  • 5
    Marovic D, Panduric V, Tarle Z, Ristic M, Sariri K, Demoli N, et al. Degree of conversion and microhardness of dental composite resin materials. J Mol Struct. 2013 Jul;1044(24):299-302.
  • 6
    Ferracane JL. Hygroscopic and hydrolytic effects in dental polymer networks. Dent Mater. 2006 Mar;22(3):211-22.
  • 7
    Palin WM, Fleming GJ, Burke FJ, Marquis PM, Randall RC. Monomer conversion versus flexure strength of a novel dental composite. J Dent. 2003 Jul;31(5):341-51.
  • 8
    Durner J, Obermaier J, Draenert M, Ilie N. Correlation of the degree of conversion with the amount of elutable substances in nano-hybrid dental composites. Dent Mater. 2012 Nov;28(11):1146-53.
  • 9
    Mantellini MG, Botero TM, Yaman P, Dennison JB, Hanks CT, Nor JE. Adhesive resin induces apoptosis and cell-cycle arrest of pulp cells. J Dent Res. 2003 Aug;82(8):592-6.
  • 10
    Pashley DH, Zhang Y, Agee KA, Rouse CJ, Carvalho RM, Russell CM. Permeability of demineralized dentin to HEMA. Dent Mater. 2000 Jan;16(1):7-14.
  • 11
    Rueggeberg FA, Hashinger DT, Fairhurst CW. Calibration of FTIR conversion analysis of contemporary dental resin composites. Dent Mater. 1990 Oct;6(4):241-9.
  • 12
    Jia W, Jin S, inventors; Pentron Clinical Technologies, LLC., assignee. Dental resin materials, method of manufacture, and uses thereof. United States patent US 6,787,629 B2. 2004 Sep 7 .
  • 13
    Salz U, Zimmermann J, Poppe D, Rheinberger V, Tiller J, Waschinski C, inventors; Ivoclar Vivadent AG, assignee. Dental materials based on radically polymerizable macromers with antimicrobial effect. United States patent US 7,553,881 B2. 2009 Jun 30.
  • 14
    Rao PS, Aasen SM, Abuelyaman AS, Mitra SB, Craig BD, Shukla BA, Plaut DJ, Russel VA, inventors; 3M Innovative Properties Company, assignee. Adhesive composition for hard tissue. United States patent US 8,076,389 B2. 2011 Dec 13.
  • 15
    Jin X, inventor; Dentsply International Inc., assignee. Method and antibacterial/antimicrobial compositions in dental compositions. United States patent US 2012/0328553 A1. 2012 Dec 27.
  • 16
    Andrzejewska E. Photopolymerization kinetics of multifunctional monomers. Prog Polym Sci. 2001 May;26(4):605-65.
  • 17
    Lovell LG, Newman SM, Bowman CN. The effects of light intensity, temperature, and comonomer composition on the polymerization behavior of dimethacrylate dental resins. J Dent Res. 1999 Aug;78(8):1469-76.
  • 18
    Elliott JE, Lovell LG, Bowman CN. Primary cyclization in the polymerization of bis-GMA and TEGDMA: a modeling approach to understanding the cure of dental resins. Dent Mater. 2001 May;17(3):221-9.
  • 19
    Amirouche-Korichi A, Mouzali M, Watts DC. Effects of monomer ratios and highly radiopaque fillers on degree of conversion and shrinkage-strain of dental resin composites. Dent Mater. 2009 Nov;25(11):1411-8.
  • 20
    Goncalves F, Azevedo CL, Ferracane JL, Braga RR. BisGMA/TEGDMA ratio and filler content effects on shrinkage stress. Dent Mater. 2011 Jun;27(6):520-6.
  • 21
    Sideridou I, Tserki V, Papanastasiou G. Effect of chemical structure on degree of conversion in light-cured dimethacrylate-based dental resins. Biomaterials. 2002 Apr;23(8):1819-29.

Publication Dates

  • Publication in this collection
    Jan-Feb 2014

History

  • Received
    02 May 2013
  • Accepted
    12 Aug 2013
  • Reviewed
    27 Aug 2013
Sociedade Brasileira de Pesquisa Odontológica - SBPqO Av. Prof. Lineu Prestes, 2227, 05508-000 São Paulo SP - Brazil, Tel. (55 11) 3044-2393/(55 11) 9-7557-1244 - São Paulo - SP - Brazil
E-mail: office.bor@ingroup.srv.br