Analysis of the reproducibility of the gray values and noise of a direct digital radiography system

Abstract

The aim of this study was to evaluate the reproducibility of the gray values and noise of a direct digital radiography system (Visualix eHD) for various exposure times and analyzed regions. To obtain radiographic images in a standardized manner, the digital sensor of the system and a stepwedge were positioned in a phantom at a focus-film distance of 30 cm in a dental device at 70 kV, 7 mA and 2.2 mm filtration. Ten consecutive repetitions of X-ray imaging were performed at each exposure time (0.05, 0.07, 0.09 and 0.13 s). Gray values were analyzed using ImageJ software in five regions of interest (ROIs): alveolar bone (AB), soft tissue (ST) and three steps of the stepwedge (S1, S2 and S3). The results showed that both the variability of the gray values and the noise were statistically greater (p < 0.05) in the most radiolucent region (ST). Only the noise was affected by the exposure time. In conclusion, the reproducibility of the gray values and the noise of the Visualix eHD system can vary in specific areas with different radiolucency.

Radiography, Dental, Digital; Reproducibility of Results; Diagnostic Imaging


Introduction

The technique of digital radiography is similar to that of conventional (analogic) radiology; however, a computer is substituted for the chemical processing phase. This substitution is advantageous because it eliminates technical errors inherent in chemical processing. Other advantages of digital systems include fast processing and the ability to use the same sensor multiple times.1Grondahl HG. Digital radiology in dental diagnosis: a critical view. Dentomaxillofac Radiol. 1992 Nov;21(4):198-202.,2Vandre RH, Webber RL. Future trends in dental radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995 Oct;80(4):471-8.,3Versteeg CH, Sanderink GC, van der Stelt PF. Efficacy of digital intra-oral radiography in clinical dentistry. J Dent. 1997 May-Jul;25(3-4):215-24.,4Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An in vivo comparison of working length determination of two frequency-based electronic apex locators. Int Endod J. 2009 Nov;42(11):1026-31.,5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.,6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4. The image visualization programs used in digital radiography systems have various post-processing resources, enabling the user to alter brightness and contrast, rotate and amplify or reduce the image, invert the gray scale, apply pseudocolors to different gray tones, perform linear and angular measurements and quantify gray values in a region of interest (ROI).5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.,6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4.,9Nelvig P, Wing K, Welander U. Sens-A-Ray. A new system for direct digital intraoral radiography. Oral Surg Oral Med Oral Pathol. 1992 Dec;74(6):818-23.,1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.,1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8.

Damante et al.1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.developed a protocol for radiographic diagnosis and follow-up studies of simple bone cysts based on gray values. The authors evaluated ten patients and concluded that it is possible to apply a non-interventional follow-up protocol that avoids unnecessary surgery of these lesions. Ferreira Junior et al.1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8. also performed a study using gray values for the diagnosis of simple bone cysts and odontogenic keratocysts. However, gray values can vary, depending on the quality of the system used and on the exposure time.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4. As a result, these situations can compromise the diagnosis and follow-up studies of the lesions.

Various studies have evaluated the physical and clinical outcomes associated with digital systems under different conditions.1Grondahl HG. Digital radiology in dental diagnosis: a critical view. Dentomaxillofac Radiol. 1992 Nov;21(4):198-202.,3Versteeg CH, Sanderink GC, van der Stelt PF. Efficacy of digital intra-oral radiography in clinical dentistry. J Dent. 1997 May-Jul;25(3-4):215-24.,4Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An in vivo comparison of working length determination of two frequency-based electronic apex locators. Int Endod J. 2009 Nov;42(11):1026-31.,5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.,6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,9Nelvig P, Wing K, Welander U. Sens-A-Ray. A new system for direct digital intraoral radiography. Oral Surg Oral Med Oral Pathol. 1992 Dec;74(6):818-23.,1212 Attaelmanan AG, Borg E, Grondahl HG. Signal-to-noise ratios of 6 intraoral digital sensors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 May;91(5):611-5.,1313 Borg E, Attaelmanan A, Grondahl HG. Subjective image quality of solid-state and photostimulable phosphor systems for digital intra-oral radiography. Dentomaxillofac Radiol. 2000 Mar;29(2):70-5.,1414 Hayakawa Y, Farman AG, Kelly MS, Kuroyanagi K. Intraoral radiographic storage phosphor image mean pixel values and signal-to-noise ratio: effects of calibration. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998 Nov;86(5):601-5.,1515 Molteni R. Direct digital dental x-ray imaging with visualix/VIXA. Oral Surg Oral Med Oral Pathol. 1993 Aug;76(2):235-43.,1616 Rasimick BJ, Shah RP, Musikant BL, Deutsch AS. Radiopacity of endodontic materials on film and a digital sensor. J Endod. 2007 Sep;33(9):1098-101.,1717 Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93. However, the current literature does not include studies of the reproducibility of gray values and noise associated with the Visualix eHD system when using gray values analysis to diagnose and track lesions.8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4.,1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.,1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8.,1212 Attaelmanan AG, Borg E, Grondahl HG. Signal-to-noise ratios of 6 intraoral digital sensors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 May;91(5):611-5. Therefore, the objective of this study was to evaluate the reproducibility of the gray values and noise associated with the Visualix eHD system when varying the exposure times and the regions analyzed.

Methodology

For this study, the charge-coupled device (CCD) sensor (universal size) of the Visualix eHD system (Gendex Dental Systems, Lake Zurich, USA) was used, which had dimensions of 37.5 × 25.5 mm, a pixel size of 19.5 µm and a spatial resolution of 25.6 lp/mm. The CCD sensor was positioned in a phantom with a soft tissue simulator (acrylic), such that radiographic imaging could be performed in a standardized manner, without altering the position of the images. The dental X-ray machine (Kaycor X-707, Yoshida Dental MFG Co. Ltd., Tokyo, Japan) was operated with 70 kV, 7 mA, 2.2 mm filtration and a focus-film distance of 30 cm. A stepwedge with three steps, corresponding to 0.2, 0.4 and 0.6 mm of thickness, was coupled to the sensor. Ten consecutive radiographic images were acquired without altering the position of the sensor for each exposure time selected (0.05, 0.07, 0.09 and 0.13 s) (Figure 1).

Figure 1
Images of the different groups.

Radiographic images were obtained, processed and stored in TIFF format (with no compression) and opened in the ImageJ program (National Institutes of Health, Washington, USA) on a computer with a 15-inch monitor (Satellite, Toshiba, Tokyo, Japan). For each image, five areas corresponding to ROIs were opened in the same square format (102 x 102 pixels), including three steps of the stepwedge (S1, S2 and S3), one for soft tissue (ST) and one for alveolar bone (AB) (Figure 2). All of the areas were saved so that their sizes and formats would be maintained for all measurements. The mean gray values of the ROIs were determined using the Analyse/Measure tool of the ImageJ program.5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.,6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4.,1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.,1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8.

Figure 2
Definitions of the regions of interest (ROIs).

To analyze the reproducibility of the gray values, the following formula was used: (highest difference/mean) × 100. The highest difference corresponds to the mean gray value minus the lowest gray value or the highest gray value minus the mean gray value in the ROI on each image.5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.The mean, highest and lowest gray values were obtained after ten consecutive scannings of the sensor for each of the four exposure times and for each of the five regions. The noise was calculated in the ROI of the images. The formula applied to calculate the noise was: (standard deviation [SD]/gray value) × 100.8 The means of the standard deviation and gray value for each ROI were obtained from ten consecutive radiographic images.

The reproducibility of the gray values and noise for the various exposure times and for each region analyzed were compared using analysis of variance (ANOVA). All statistical analyses were performed using Statistica for Windows 6.0 (Statistica for Windows, StatSoft Inc., Tulsa, USA). Statistically significant differences were defined as those with p < 0.05.

Results

Table 1 shows the results of comparisons of the gray values for various exposure times and ROIs. There were no significant differences between mean gray values evaluated for analyzed ROIs and exposure times (p > 0.05).

Table 1
Comparison of the gray values in the various exposure times and ROIs (ANOVA).

There were no significant differences in the reproducibility of the gray values for each ROI at the different exposure times (p > 0.05). However, there was a significant difference in the reproducibility between the soft tissue ROI and the other four ROIs (p = 0.000) at all four exposure times (Table 2).

Table 2
Comparison of the reproducibility of the gray values in the various exposure times and ROIs.

Different letters in the vertical direction represent statistically significant differences (Tukey’s test).

There were significant differences in the noise values between the different exposure times and ROIs (p < 0.05) (Table 3).

Table 3
Comparison of the noise in the various exposure times and ROIs.

Different uppercase letters in the horizontal direction represent statistically significant differences (Tukey’s test).

Different lowercase letters in the vertical direction represent statistically significant differences (Tukey’s test).

Discussion

No studies to date have physically analyzed the reproducibility of the gray values and noise associated with the Visualix eHD system. The results of this study demonstrate the reproducibility of the gray values and noise for the system in five regions of radiographic imaging when the exposure time was varied.

Our study did not detect a statistically significant difference between the gray values with different exposure times, perhaps because there was relatively little variation in the exposure times selected (ranging from 0.05 to 0.13) or because the phantom with the soft tissue simulator attenuated the radiation.5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13. According to Pietrobelli et al.1818 Pietrobelli A, Formica C, Wang Z, Heymsfield SB. Dual-energy X-ray absorptiometry body composition model: review of physical concepts. Am J Physiol. 1996 Dec;271(6 Pt 1):E941-51. attenuation occurs when the photon signal crosses an element, chemical component or solution, spontaneously reducing the intensity through atomic interactions and leading to the absorption and dissipation of photons. Thus, when an X-ray beam penetrates an analyzed region, the presence of soft and hard tissues in this trajectory influences the gray values. In this study, the highest gray values were observed in the most radiopaque regions (AB and S3), and the lowest values were observed in the most radiolucent regions (ST and S1)5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.,8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4.,1616 Rasimick BJ, Shah RP, Musikant BL, Deutsch AS. Radiopacity of endodontic materials on film and a digital sensor. J Endod. 2007 Sep;33(9):1098-101.,1717 Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93. (Table 1).

No significant differences were found between the reproducibility of the gray values (p ≥ 0.185) for various exposure times (0.05, 0.07, 0.09 and 0.13 s), because the gray values were not influenced by the exposure times. However, the reproducibility of the gray values did vary between the soft tissue ROI and the other four ROIs (p = 0.000). This result demonstrated that the variability of the gray shades after consecutive repetitions of the same ROI was greater in more radiolucent ROIs (Table 2). The higher mean variability of the gray shades for the Visualix eHD system in this study was 4.09%, which is lower than the results reported for the Digora system, i.e., 75% by Freitaset al.5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13. and 27.95% by Teixeira et al.1717 Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93.

The lack of reproducibility of gray values of radiographic systems could be deleterious when limitations are unknown, because the system could potentially attribute differences in gray values as corresponding to a true change in the clinical condition of a patient. This situation could compromise the diagnosis and follow-up studies when a quantitative analysis of gray values is used,5Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13. with clinical implications. For example, when a carious lesion is relatively large, the area analyzed will be more radiolucent, and there will be greater variation in gray values. In contrast, when gray values analysis is used to evaluate the evolution of bone lesions,1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.,1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8. the area analyzed will be more radiopaque, leading to lower variability of the gray values and more accurate analysis using the Visualix eHD system. This finding is supported by the studies of Damante et al.1010 Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6. and Ferreira Junior et al.1111 Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8.which proposed a protocol for the radiographic diagnosis and follow-up of simple bone cysts and odontogenic keratocysts based on gray values of the images.

A key factor for successful radiographic imaging is that the X-ray signal detected by the sensor must overcome the inherent noise of the digital radiographic system.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4. Noise is the undesirable fluctuation of pixel intensities, which can compromise the image studied,1212 Attaelmanan AG, Borg E, Grondahl HG. Signal-to-noise ratios of 6 intraoral digital sensors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 May;91(5):611-5.,1717 Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93. thereby affecting the grey-scale intensity and, consequently, the gray values.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,1717 Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93. Noise levels can vary, depending on the quality of the system used6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.,8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4. and the exposure time.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4. Moreover, noise levels can be evaluated using various physical parameters, such as the standard deviation.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.,7Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.

In this study, we identified differences in noise levels for various exposure times and ROIs. The noise decreased with increasing exposure times for the three areas of the stepwedge (Table 3). This finding differs from a study by Wenzel,6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4. which found an increase in noise with an increase in exposure time from 0.20 to 0.60 s. The reason for this difference is based on the dose of radiation needed to saturate the sensor used in each of these studies.6Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4. In our study, the exposure times were smaller (ranging from 0.05 to 0.13 s), whereas the study by Wenzel’s had longer exposure times (ranging from 0.20 to 0.60 s). On the other hand, our study found that the noise was statistically significantly greater when the area analyzed was more radiolucent (i.e., soft tissue). These findings are in accordance with the results reported by Rubira-Bullen et al.8Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4. for digital radiographic images. Thus, as observed in the analysis of the reproducibility of the gray values in this study, as the noise of the system used increases, the interference for the quantitative analysis of the gray values also increases, especially in radiolucent lesions, such as dental caries.

This study did not assess the stability of the output power of the dental X-ray machine used. However, any instability of the machine was not sufficient to alter the reproducibility of the gray values or the noise.

Studies that evaluate the physical analysis of the radiographic systems are important to show the limitations and improve the diagnosis of lesions and follow-up studies.

Conclusions

Based on the methods used, the reproducibility of the gray values and noise of the Visualix eHD system can vary in specific areas with different radiolucency.

References

  • 1
    Grondahl HG. Digital radiology in dental diagnosis: a critical view. Dentomaxillofac Radiol. 1992 Nov;21(4):198-202.
  • 2
    Vandre RH, Webber RL. Future trends in dental radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995 Oct;80(4):471-8.
  • 3
    Versteeg CH, Sanderink GC, van der Stelt PF. Efficacy of digital intra-oral radiography in clinical dentistry. J Dent. 1997 May-Jul;25(3-4):215-24.
  • 4
    Pascon EA, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani MA. An in vivo comparison of working length determination of two frequency-based electronic apex locators. Int Endod J. 2009 Nov;42(11):1026-31.
  • 5
    Freitas P, Yaedu RY, Rubira-Bullen IR, Escarpinati M, Vieira MC, Schiabel H, et al. Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings. Braz Oral Res. 2006 Jul-Sep;20(3):207-13.
  • 6
    Wenzel A. Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography. Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):70-4.
  • 7
    Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: a comparison between two subtraction programs. Dentomaxillofac Radiol. 2005 Nov;34(6):357-61.
  • 8
    Rubira-Bullen IR, Escarpinati MC, Schiabel H, Vieira MA, Rubira CM, Lauris JR. Evaluating noise in digitized radiographic images by means of histogram. J Appl Oral Sci. 2006 Dec;14(6):410-4.
  • 9
    Nelvig P, Wing K, Welander U. Sens-A-Ray. A new system for direct digital intraoral radiography. Oral Surg Oral Med Oral Pathol. 1992 Dec;74(6):818-23.
  • 10
    Damante JH, Guerra ENS, Ferreira Junior O. Spontaneous resolution of simple bone cysts. Dentomaxillofac Radiol. 2002 May;31(3):182-6.
  • 11
    Ferreira Junior O, Damante JH, Lauris JR. Simple bone cyst versus odontogenic keratocyst: differential diagnosis by digitized panoramic radiography. Dentomaxillofac Radiol. 2004 Nov;33(6):373-8.
  • 12
    Attaelmanan AG, Borg E, Grondahl HG. Signal-to-noise ratios of 6 intraoral digital sensors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 May;91(5):611-5.
  • 13
    Borg E, Attaelmanan A, Grondahl HG. Subjective image quality of solid-state and photostimulable phosphor systems for digital intra-oral radiography. Dentomaxillofac Radiol. 2000 Mar;29(2):70-5.
  • 14
    Hayakawa Y, Farman AG, Kelly MS, Kuroyanagi K. Intraoral radiographic storage phosphor image mean pixel values and signal-to-noise ratio: effects of calibration. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998 Nov;86(5):601-5.
  • 15
    Molteni R. Direct digital dental x-ray imaging with visualix/VIXA. Oral Surg Oral Med Oral Pathol. 1993 Aug;76(2):235-43.
  • 16
    Rasimick BJ, Shah RP, Musikant BL, Deutsch AS. Radiopacity of endodontic materials on film and a digital sensor. J Endod. 2007 Sep;33(9):1098-101.
  • 17
    Teixeira RC, Rubira-Bullen IRF, Rubira CMF, Lauris JR. Avaliação da reprodutibilidade do valor de pixel no sistema Digora. Rev ABRO. 2007;8(2):86-93.
  • 18
    Pietrobelli A, Formica C, Wang Z, Heymsfield SB. Dual-energy X-ray absorptiometry body composition model: review of physical concepts. Am J Physiol. 1996 Dec;271(6 Pt 1):E941-51.

Publication Dates

  • Publication in this collection
    2015

History

  • Received
    09 Oct 2014
  • Accepted
    21 Jan 2015
  • Reviewed
    22 Apr 2015
Sociedade Brasileira de Pesquisa Odontológica - SBPqO Av. Prof. Lineu Prestes, 2227, 05508-000 São Paulo SP - Brazil, Tel. (55 11) 3044-2393/(55 11) 9-7557-1244 - São Paulo - SP - Brazil
E-mail: office.bor@ingroup.srv.br