Effect of nickel-titanium alloys on root canal preparation and on mechanical properties of rotary instruments

Abstract

The aim of this study was to evaluate the quality of curved root canal preparation, torsional fatigue, and cyclic fatigue of rotary systems manufactured with different NiTi alloys. Ninety single-rooted canals with curvatures of 15° to 30o were scanned and divided into three groups according to the rotary system used: BT-Race (BTR) – 10.06, 35.00, 35.04; SequenceRotaryFile (SRF) – 15.04, 25.06, 35.04; and ProDesignLogic (PDL) – 25.01, 25.06, 35.05. Each system was used on three specimens. The teeth were prepared, scanned, and analyzed to assess increase in volume, transportation, and centering ability of the root canal. Torsional fatigue of glide path instruments (BTR 10.06, SRF 15.04 and PDL 25.01) and cyclic fatigue of the finishing instrument (BTR 35.04, SRF 35.04 and PDL 35.05) were obtained by analyzing completely new instruments (n = 10) and instruments after they had been used three times (n = 10). After the torsional and cyclic fatigue tests, the fractured surface of the new and used instruments were examined by scanning electron microscopy. Increase in volume, canal transportation, and centering ability showed no significant differences among the groups (p > 0.05). The torsional test showed that SRF 15.04 produced the highest torque values for both new and used instruments, followed by PDL 25.01 and BTR 10.06 (p < 0.05). PDL 25.01, both new and used, exhibited higher values of angular deflection followed by SRF 15.04 and BTR 10.06 (p < 0.05). As regards cyclic fatigue, use of PDL 35.05, both new and used, required a longer time and larger number of cycles than did SRF 35.04 and BTR 35.04 (p < 0.05). Clinical use affected the torsional fatigue of BTR; however, cyclic fatigue was not significantly affected (p < 0.05). All rotary systems were able to prepare the curved canals satisfactorily and were used safely on the three specimens. Relative to torsional fatigue, SRF 15.04 exhibited a higher torque, and PDL 25.01, higher angular deflection. BTR 10.06 was the most affected by clinical use. PDL 35.05 showed greater resistance to cyclic fatigue.

Endodontics; Dental Instruments; Root Canal Preparation; X-Ray Microtomography

Introduction

In recent years, several modifications in alloys and in the design of nickel-titanium (NiTi) instruments have been made to improve their performance, with special reference to instruments used in curved canals. These modifications prevent canal transportation and instrument fractures, which could compromise the success of endodontic treatment.11.De-Deus G, Silva EJ, Vieira VT, Belladonna FG, Elias CN, Plotino G, et al. Blue Thermomechanical Treatment Optimizes Fatigue Resistance and Flexibility of the Reciproc Files. J Endod. 2017 Mar;43(3):462-6. https://doi.org/10.1016/j.joen.2016.10.039
https://doi.org/10.1016/j.joen.2016.10.0...
, 22.Marks Duarte P, Silva PB, Alcalde MP, Vivan RR, Rosa RA, Duarte MA, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018 Sep;44(9):1425-9. https://doi.org/10.1016/j.joen.2018.06.002
https://doi.org/10.1016/j.joen.2018.06.0...

During root canal preparation, procedures are performed in successive steps, with the use of instruments for different purposes. One of these purposes is to create a glide path, a trajectory that leads from the canal entrance to the apical foramen.33.Arias A, Singh R, Peters OA. Differences in torsional performance of single- and multiple-instrument rotary systems for glide path preparation. Odontology. 2016 May;104(2):192-8. https://doi.org/10.1007/s10266-015-0199-0
https://doi.org/10.1007/s10266-015-0199-...
, 44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
Nowadays, several rotary systems have specific files for this step. However, because they are used at the beginning of the procedure, these instruments are more prone to torsional stresses.44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
, 55.De-Deus G, Belladonna FG, Souza EM, Alves VO, Silva EJ, Rodrigues E, et al. Scouting Ability of 4 Pathfinding Instruments in Moderately Curved Molar Canals. J Endod. 2016 Oct;42(10):1540-4. https://doi.org/10.1016/j.joen.2016.07.001
https://doi.org/10.1016/j.joen.2016.07.0...
Torsional stress occurs when the tip of the instrument becomes locked into the dentin wall as the shaft continues to rotate until fracture occurs.66.Goo HJ, Kwak SW, Ha JH, Pedullà E, Kim HC. Mechanical Properties of Various Heat-treated Nickel-titanium Rotary Instruments. J Endod. 2017 Nov;43(11):1872-7. https://doi.org/10.1016/j.joen.2017.05.025
https://doi.org/10.1016/j.joen.2017.05.0...
, 77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...

After glide path preparation, another important step in endodontic treatment begins: root canal enlargement. This step becomes even more challenging when the tooth has a curved canal, because at the maximum bending point of the instrument, it is subjected to tensile/compressive forces. If these forces, which are cumulative, exceed the maximum value the instrument can withstand, the instrument will fracture because of cyclic fatigue.66.Goo HJ, Kwak SW, Ha JH, Pedullà E, Kim HC. Mechanical Properties of Various Heat-treated Nickel-titanium Rotary Instruments. J Endod. 2017 Nov;43(11):1872-7. https://doi.org/10.1016/j.joen.2017.05.025
https://doi.org/10.1016/j.joen.2017.05.0...
, 77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
Considering the foregoing factors, several instruments have been introduced in the market, such as BT-RaCe (FKG Dentaire, La Chaux-de-Fonds, Switzerland). This is a rotary system with a new – booster tip – design which, according to the manufacturer, facilitates advancement of the instrument and reduces the risks of canal transportation, thereby improving centering ability. Furthermore, this system is made of a conventional NiTi alloy that has been subjected to electrochemical and thermal surface treatment. The BT-RaCe system has a glide path instrument with size 10.06 and two instruments for canal preparation with size 35.00 and size 35.04.88.Bürklein S, Mathey D, Schäfer E. Shaping ability of ProTaper NEXT and BT-RaCe nickel-titanium instruments in severely curved root canals. Int Endod J. 2015 Aug;48(8):774-81. https://doi.org/10.1111/iej.12375
https://doi.org/10.1111/iej.12375...

9.Brasil SC, Marceliano-Alves MF, Marques ML, Grillo JP, Lacerda MF, Alves FR, et al. Canal Transportation, Unprepared Areas, and Dentin Removal after Preparation with BT-RaCe and ProTaper Next Systems. J Endod. 2017 Oct;43(10):1683-7. https://doi.org/10.1016/j.joen.2017.04.012
https://doi.org/10.1016/j.joen.2017.04.0...
- 1010.Krokidis A, Bonfanti C, Cerutti A, Barabanti N, Zinelis S, Panopoulos P. Comparative analysis of SAF, Protaper Next and BT-Race in eliminating Enterococcus faecalis from long oval canals: an ex vivo study. Aust Endod J. 2017 Dec;43(3):110-4. https://doi.org/10.1111/aej.12177
https://doi.org/10.1111/aej.12177...

Another rotary system, the Sequence Rotary File (MKLife, Porto Alegre, Brazil), has been recently introduced in the market, offering heat-treated instruments manufactured with Blue technology. With a cross section and an inactive guide, this system presents certain similarity to Vortex Blue. This system consists of a glide path with file size 15.04 and three shaping instruments with size 20.06, size 25.06, and size 35.04. Except for instrument 15.04, which has a quadrangular cross section, the other instruments have a convex triangular cross section. However, to the best of our our knowledge, there are no studies in the literature evaluating this system.

ProDesign Logic (Easy Dental Equipment, Belo Horizonte, MG, Brazil), a continuous rotary system, is manufactured of NiTi with controlled memory (CM). In the United States, the system is sold as Bassi Logic. This system has a glide path instrument with size 25.01, quadrangular section, and instruments for canal preparation with size 25.06 and size 35.05, with an S-shaped cross section.1111.Coelho BS, Amaral RO, Leonardi DP, Marques-da-Silva B, Silva-Sousa YT, Carvalho FM, et al. Performance of Three single instrument systems in the preparation of long oval canals. Braz Dent J. 2016 Mar-Apr;27(2):217-22. https://doi.org/10.1590/0103-6440201302449
https://doi.org/10.1590/0103-64402013024...
, 1212.Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
https://doi.org/10.1111/iej.12881...

Several studies have investigated new instruments made available by different systems11.De-Deus G, Silva EJ, Vieira VT, Belladonna FG, Elias CN, Plotino G, et al. Blue Thermomechanical Treatment Optimizes Fatigue Resistance and Flexibility of the Reciproc Files. J Endod. 2017 Mar;43(3):462-6. https://doi.org/10.1016/j.joen.2016.10.039
https://doi.org/10.1016/j.joen.2016.10.0...

2.Marks Duarte P, Silva PB, Alcalde MP, Vivan RR, Rosa RA, Duarte MA, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018 Sep;44(9):1425-9. https://doi.org/10.1016/j.joen.2018.06.002
https://doi.org/10.1016/j.joen.2018.06.0...

3.Arias A, Singh R, Peters OA. Differences in torsional performance of single- and multiple-instrument rotary systems for glide path preparation. Odontology. 2016 May;104(2):192-8. https://doi.org/10.1007/s10266-015-0199-0
https://doi.org/10.1007/s10266-015-0199-...

4.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
- 55.De-Deus G, Belladonna FG, Souza EM, Alves VO, Silva EJ, Rodrigues E, et al. Scouting Ability of 4 Pathfinding Instruments in Moderately Curved Molar Canals. J Endod. 2016 Oct;42(10):1540-4. https://doi.org/10.1016/j.joen.2016.07.001
https://doi.org/10.1016/j.joen.2016.07.0...
, 77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
without evaluating the interference of dentin, exposure to the irrigation solution, and steam sterilization procedures accomplished in an autoclave. In addition, the literature shows the possibility of reusing instruments,1313.Vieira EP, Nakagawa RK, Buono VT, Bahia MG. Torsional behaviour of rotary NiTi ProTaper Universal instruments after multiple clinical use. Int Endod J. 2009 Oct;42(10):947-53. https://doi.org/10.1111/j.1365-2591.2009.01602.x
https://doi.org/10.1111/j.1365-2591.2009...
, 1414.Bueno CS, Oliveira DP, Pelegrine RA, Fontana CE, Rocha DG, Bueno CE. Fracture incidence of waveone and reciproc files during root canal preparation of up to 3 posterior teeth: a prospective clinical study. J Endod. 2017 May;43(5):705-8. https://doi.org/10.1016/j.joen.2016.12.024
https://doi.org/10.1016/j.joen.2016.12.0...
however, the extent to which clinical use affects NiTi instruments and their mechanical properties is not known.

Cyclic fatigue is typically evaluated by testing smaller diameter instruments.77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
Nonetheless, the literature has shown improved characteristics of biomechanical preparation when wider enlargement of the root canals is performed.1515.Boutsioukis C, Gogos C, Verhaagen B, Versluis M, Kastrinakis E, Van der Sluis LW. The effect of apical preparation size on irrigant flow in root canals evaluated using an unsteady Computational Fluid Dynamics model. Int Endod J. 2010 Oct;43(10):874-81. https://doi.org/10.1111/j.1365-2591.2010.01761.x
https://doi.org/10.1111/j.1365-2591.2010...

16.Rodrigues RC, Zandi H, Kristoffersen AK, Enersen M, Mdala I, Ørstavik D, et al. Influence of the apical preparation size and the irrigant type on bacterial reduction in root canal-treated teeth with apical periodontitis. J Endod. 2017 Jul;43(7):1058-63. https://doi.org/10.1016/j.joen.2017.02.004
https://doi.org/10.1016/j.joen.2017.02.0...
- 1717.Pérez AR, Alves FR, Marceliano-Alves MF, Provenzano JC, Gonçalves LS, Neves AA, et al. Effects of increased apical enlargement on the amount of unprepared areas and coronal dentine removal: a micro-computed tomography study. Int Endod J. 2018 Jun;51(6):684-90. https://doi.org/10.1111/iej.12873
https://doi.org/10.1111/iej.12873...
Therefore, the aim of this study was based on the importance of elucidating which system would be safer for use in the instrumentation of curved canals. Specifically considering the scarcity of evidence on the above-mentioned aspect of safety, the aim was to evaluate the quality of preparation performed with SRF, BT-Race, and PDL systems, and the extent of cyclic fatigue of instruments with larger diameters. Consequently, the quality of the preparation of curved canals, torsional fatigue of glide path instruments, and cyclic fatigue of finishing instruments of continuous rotary systems with different types of NiTi alloys were assessed. The null hypotheses tested were as follows: a) there would be no difference in quality of preparation among the different systems, b) there would be no difference in torsional fatigue strength of the different glide path instruments, c) there would be no difference in cyclic fatigue strength among the finishing instruments of the tested systems, and d) there would be no influence of simulated clinical use on the mechanical properties of instruments.

Methodology

The present study was approved by the Human Research Ethics Committee (process number: 88418518.4.0000.5417).

The sample size was calculated using the G*Power v3.1 for Mac (Heinrich Heine, Universität Düsseldorf, Germany), and the data were statistically analyzed by the Wilcoxon-Mann Whitney test. The data obtained from a previous study of the cyclic and torsional tests77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
was used to establish the effect size of the present study (= 1.2). An alpha type error of 0.05, a beta power of 0.80, and a ratio N2/N1 of 1 were also stipulated. A total of eight samples per group were indicated as the ideal size required for finding significant differences. Nevertheless, 20% of the total number of instruments were added to compensate for possible atypical values that might lead to sample loss.

Tooth preparation

In total, 90 extracted human mandibular premolars, stored in a 0.1% thymol solution, were selected. The following inclusion criteria were used: type I canal configuration according to Vertucci’s classification,1818.Vertucci FJ. Root canal anatomy of the human permanent teeth. Oral Surg Oral Med Oral Pathol. 1984 Nov;58(5):589-99. https://doi.org/10.1016/0030-4220(84)90085-9
https://doi.org/10.1016/0030-4220(84)900...
an angle of curvature between 15° and 30° according to Schneider’s method,1919.Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971 Aug;32(2):271-5. https://doi.org/10.1016/0030-4220(71)90230-1
https://doi.org/10.1016/0030-4220(71)902...
without calcifications, root fractures or separated instruments, and complete apex formation.2020.Fernandes PO, Freire LG, Iglecias EF, Vieira BR, Zuolo ML, Gavini G. Assessment of mechanical root canal preparation with centric reciprocating or eccentric rotary kinematics: a micro-computed tomographic study. J Endod. 2020 Sep;46(9):1309-16. https://doi.org/10.1016/j.joen.2020.06.005
https://doi.org/10.1016/j.joen.2020.06.0...
Specimens with an anatomical diameter of 0.20 mm were selected according to measurements of mandibular premolar diameters investigated in previous studies.2121.Hassanien EE, Hashem A, Chalfin H. Histomorphometric study of the root apex of mandibular premolar teeth: an attempt to correlate working length measured with electronic and radiograph methods to various anatomic positions in the apical portion of the canal. J Endod. 2008 Apr;34(4):408-12. https://doi.org/10.1016/j.joen.2007.12.013
https://doi.org/10.1016/j.joen.2007.12.0...
, 2222.Wolf TG, Kim P, Campus G, Stiebritz M, Siegrist M, Briseño-Marroquín B. 3-dimensional analysis and systematic review of root canal morphology and physiological foramen geometry of 109 mandibular first premolars by micro-computed tomography in a mixed Swiss-German population. J Endod. 2020 Jun;46(6):801-9. https://doi.org/10.1016/j.joen.2020.03.002
https://doi.org/10.1016/j.joen.2020.03.0...
The teeth were standardized and randomized to obtain the groups in terms of degree of curvature, assessed by digital radiography. The samples were then scanned by using computed microtomography (SkyScan 1174v2; Bruker-micro-CT, Ethiolles, Belgium). A silicone mold was made for each tooth to ensure that it would be scanned in the same position, so as not to interfere in the subsequent analysis. The parameters used were a 0.5-mm aluminum filter, 50 kV, 800 mA, and a voxel size of 19.6 μm, rotation angle of 0.7, and total rotation of 360º, thus producing images with a resolution of 1304 x 1024 pixels. Each scan resulted in images that were reconstructed using NRecon software v1.6.4.8 (Bruker-micro-CT). After the segmentation and binarization processes, three-dimensional models were reconstructed with CTAn v.1.12 software (Bruker-microCT Kontich, Belgium). CTVol v.2.2.1 and Data Viewer software programs (Bruker-micro CT Kontich, Belgium) were used for visualizing and evaluating the internal anatomy. The baseline volume of both the entire canal and the apical third were standardized based on micro-CT assessment.

After creating the endodontic access cavity, the canals of all teeth were explored with size 06, 08, and 10 K-files. The working length (WL), 1 mm short of the apical foramen, was established with a 10 K-file (Dentsply Sirona, Ballaigues, Switzerland) when its tip was visualized in the foramen under a stereo microscope (Stemi 2000C; Carl Zeiss, Jena, Germany). Subsequently, the teeth were subjected to biomechanical root canal preparation with the following rotary systems (n=30):

Group 1: BT-Race (BTR) – The canal was irrigated with 2 mL of 2.5% sodium hypochlorite (NaOCl) and then instrumented with the BTR system according to the sequence proposed by the manufacturer. BTR 10.06, 35.00, and 35.04 instruments were coupled to an endodontic electric motor (VDW GmbH, Munich, Germany) and used in continuous rotation at a speed of 800 rpm and torque of 1.5 N. For each instrument, three in-and-out movements were performed in the apical direction. After these movements, the instrument was cleaned with gauze and the canal was irrigated with 2.5% NaOCl. Instrumentation was performed repeatedly, as described earlier, until the WL was reached. During root canal preparation, the canals were irrigated with 20 mL of 2.5% NaOCl using the syringe-needle irrigation technique. In addition, a 30-G side-vented needle was used at a distance of 3 mm from the WL. After root canal shaping, the instruments were ultrasonically cleaned for 5 min and sterilized at 121° C, at a pressure of 30 psi, for 20 min. Each system was used on three teeth and the procedures were repeated.

Group 2: Sequence Rotary File (SRF) – The procedure was similar to that described earlier, but SRF 15.04, 25.06, and 35.04 instruments were used in continuous rotation, with the glide path 15.04 instrument at a speed of 350 rpm and torque of 1 N, and the shaping instruments 25.06 and 35.04, at a speed of 500 rpm and torque of 2 N, driven by a VDW endodontic electric motor in the Doctors mode, according to the manufacturer’s instructions.

Group 3: ProDesign Logic (PDL) – A procedure similar to the previous two was performed using PDL instruments 25.01, 25.06, and 35.05 in continuous rotation, with glide path 25.01 instrument at a speed of 350 rpm and torque of 1 N and shaping instruments 25.06 and 35.05 at a speed of 950 rpm and torque of 4 N, driven by a VDW endodontic electric motor in the Doctors mode, according to the manufacturer’s instructions.

Upon the completion of biomechanical root canal preparation, the canals were irrigated with 5 mL of 17% EDTA (Biodynamics, Ibiporã, Paraná, Brazil) for 3 min to remove the smear layer and then rinsed with 5 mL of saline solution, dried with #35 absorbent paper points, and subjected to computerized microtomography (micro-CT) using the same parameters used initially.

Volume measurement

The pre-instrumentation and post-instrumentation images were superimposed using the 3D recording function of DataViewer v.1.5.1 (Bruker micro-CT) software. The images recorded were processed in CTAN software v.1.14.4 (Bruker micro-CT) to calculate the apical volume, comprising the last four apical millimeters, and the total volume, which consisted of the volume of the canal that extended from the root apex to 10 mm in the cervical direction.

Evaluation of root canal transportation

For the analysis of root canal transportation, axial sections corresponding to distances of 1, 2, 3, and 4 mm from the apex, which indicated the region of the curvature, were selected. Transportation was calculated in millimeters using the formula ([X1-X2] - [Y1-Y2]) as described by Gambill et al.,2323.Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod. 1996 Jul;22(7):369-75. https://doi.org/10.1016/S0099-2399(96)80221-4
https://doi.org/10.1016/S0099-2399(96)80...
where X1 is the shortest distance between the inner wall of the root curvature and the lumen of the uninstrumented canal; X2 is the shortest distance between the inner wall of the root curvature and the lumen of the canal after being instrumented; Y1 is the shortest distance between the outer wall of the root curvature and the lumen of the uninstrumented canal; and Y2 is the shortest distance between the outer wall of the root curvature and the lumen of the canal after being instrumented ( Figure 1 ). According to this formula, a result of 0 (zero) indicated no root canal transportation; a negative result indicated transportation to the outer region of the curvature; and a positive result indicated transportation to the inner region of the curvature.

Figure 1
Representative images of the matching of axial reconstructions with the method used to detect root canal transportation in the last four millimeters of the apical portion: a) Axial sections indicating the thickness of the preoperative root canal dentin (X1 and Y1) and b) thickness of the dentin where root canal transportation (X2 and Y2) occurred in the postoperative period.

Evaluation of centering ability

The measurements of X1, Y1, X2, and Y2 at the apical third were used to evaluate centering ability ( Figure 1 ), according to the equation suggested by Gambill et al.2323.Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod. 1996 Jul;22(7):369-75. https://doi.org/10.1016/S0099-2399(96)80221-4
https://doi.org/10.1016/S0099-2399(96)80...
: (X1-X2)/(Y1-Y2) or (Y1-Y2)/(X1-X2). The lowest value was considered the numerator. A result of 1 indicated perfect centering ability.

Torsional fatigue test

During biomechanical preparation, the same system was used on three different teeth, and after being used three times, the glide path files were always subjected to a torsional fatigue test.

Torsion testing was based on ISO 3630-1 (1992) specification, for which a specific machine was used, as described in detail in other studies.2424.Bahia MG, Melo MC, Buono VT. Influence of simulated clinical use on the torsional behavior of nickel-titanium rotary endodontic instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 May;101(5):675-80. https://doi.org/10.1016/j.tripleo.2005.04.019
https://doi.org/10.1016/j.tripleo.2005.0...
, 2525.Alcalde MP, Tanomaru-Filho M, Bramante CM, Duarte MA, Guerreiro-Tanomaru JM, Camilo-Pinto J, et al. Cyclic and Torsional Fatigue Resistance of Reciprocating Single Files Manufactured by Different Nickel-titanium Alloys. J Endod. 2017 Jul;43(7):1186-91. https://doi.org/10.1016/j.joen.2017.03.008
https://doi.org/10.1016/j.joen.2017.03.0...
The test was performed only with the glide path files of the systems used in this study (BTR 10.06, SRF 15.04, and PDL 25.01), of both new instruments (n = 10) and those used three times for biomechanical preparation (n = 10). The test established the average torque and maximum angular deflection values before instrument fracture. The torque values were evaluated by measuring the force exerted on a small load cell by a lever arm connected to the axis of torsion. The angle of rotation was measured and controlled by a resistive angle transducer connected to a process controller. To perform the test, the instrument handle was removed at the point where it was attached to the shaft. The end of the shaft was attached to a mandrel connected to a reversible gear motor. Three millimeters of the instrument tip was attached to another mandrel with brass jaws to prevent slippage ( Figure 2A ).

Figure 2
Experimental setup showing the (A) torsional fatigue test device and (B) cyclic fatigue test device.

The clockwise rotation speed was set to 2 rpm. The continuous recording of torque and angular rotation was monitored and the final torsional force and angular rotation (o) were provided by a specifically designed computer program (Analogica, Belo Horizonte, MG, Brazil) and recorded.2525.Alcalde MP, Tanomaru-Filho M, Bramante CM, Duarte MA, Guerreiro-Tanomaru JM, Camilo-Pinto J, et al. Cyclic and Torsional Fatigue Resistance of Reciprocating Single Files Manufactured by Different Nickel-titanium Alloys. J Endod. 2017 Jul;43(7):1186-91. https://doi.org/10.1016/j.joen.2017.03.008
https://doi.org/10.1016/j.joen.2017.03.0...

Cyclic fatigue test

The cyclic fatigue test was conducted with the finishing instruments (BTR 35.04, SRF 35.04, and PDL 35.05) of each system used in the biomechanical preparation, and the methodology was based on that of the test described and used by Marks Duarte et al.22.Marks Duarte P, Silva PB, Alcalde MP, Vivan RR, Rosa RA, Duarte MA, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018 Sep;44(9):1425-9. https://doi.org/10.1016/j.joen.2018.06.002
https://doi.org/10.1016/j.joen.2018.06.0...
and Klymus et al.2626.Klymus ME, Alcalde MP, Vivan RR, Só MV, Vasconselos BC, Duarte MA. Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2019 Jul;23(7):3047-52. https://doi.org/10.1007/s00784-018-2718-1
https://doi.org/10.1007/s00784-018-2718-...
An angle of 30° and a radius of 5 mm were used in the test ( Figure 2B ). The cyclic fatigue test was performed with brand-new instruments (n = 10) and also with the instruments that had been used three times in the biomechanical preparation (n = 10). Time (in seconds) until instrument fracture occurred was measured with a digital stopwatch. A video was recorded simultaneously to establish the exact moment when the fracture occurred. In addition, the number of cycles until instrument fracture was calculated.

SEM evaluation

After the torsional and cyclic fatigue tests, the instruments were ultrasonically cleaned and the fractured surfaces were examined by scanning electron microscopy (JSM-TLLOA; JEOL, Tokyo, Japan) to evaluate topographic features. The photomicrographs were taken at 50X, 200X, and 1000x magnification.

Statistical analysis

Statistical analysis was performed using GraphPad Prism 5 software (La Jolla, USA). The data were subjected to the D’Agostino-Pearson test to verify if they were normally distributed. For volume, canal transportation, and centering ability, the data were not normally distributed; thus, the non-parametric Kruskal-Wallis and Dunn’s tests were used. For the torsional and cyclic fatigue tests, the data distribution was normal, and they were subjected to ANOVA and Tukey’s parametric tests. A significance level of 5% was considered for all tests.

Results

The median, minima and maxima of baseline and final volume, and percent increase in volume are presented in Table 1 . Baseline volume and percent increase in volume showed no significant difference in apical and total root canal volume (p > 0.05) among the groups. In the within-group comparison, however, there was a significant difference in all groups between baseline and final volume, both for apical and total volume (p < 0.05).

Table 1
Median, minimum and maximum values for volume, in mm3, and percent volume increase after the use of different rotary systems.

The median, minima and maxima of canal transportation, and centering ability are shown in Table 2 . There was no significant difference among the groups for any of the analyses (p > 0.05). All groups exhibited little canal transportation and satisfactory centering ability.

Table 2
Mean, minimum and maximum values, in mm3, for canal transportation and centering ability after the use of different rotary systems.

The torsional strength (maximum load torque and angle of rotation) of the instruments is shown in Table 3 . SRF 15.04, new, and used instruments had higher torsional strength than did PDL 25.01 and BTR 10.06, respectively (p < 0.05). In relation to the angle of rotation of both new and used instruments, PDL showed higher values, followed by SRF and BTR, respectively (p < 0.05). The within-group comparison between new and used instruments showed that BTR significantly reduced torque strength (p < 0.05), while SRF and PDL were not affected by clinical use (p > 0.05). In relation to the angle of rotation, all instruments were significantly affected by clinical use (p < 0.05).

Table 3
Mean values for torque (N.cm) and angle of rotation (o) of new instruments and used ones (after three uses).

Cyclic fatigue resistance of the instruments is shown in Table 4 . Among the new instruments, relative to both the time and the number of cycles until fracture, PDL 35.05 had the highest resistance to fatigue, followed by SRF 35.04 and BTR 35.04, respectively (p < 0.05). Among the used instruments, PDL and SRF exhibited a longer time to fracture than did BTR (p < 0.05). Nonetheless, when the number of cycles was analyzed, PDL had higher cyclic fatigue resistance than did SRF and BTR, respectively (p < 0.05). The within-group comparison between new and used instruments showed that no instrument was significantly affected by clinical use (p > 0.05).

Table 4
Mean values for time (in seconds) and number of cycles (NCF) of new instruments and used ones (after three uses).

Scanning electron microscopy of the fractured surface showed typical features of torsional failure for all brands ( Figure 3 ) and typical features of cyclic fatigue ( Figure 4 ) for new and used instruments. In the side view ( Figure 3 ), deformation of the spiral flutes of the instruments could be noted. The instruments had concentric abrasion and fibrous dimple marks at the center of rotation for torsional failure ( Figure 3 ). In Figure 4 , all the instruments displayed fractured surfaces with microvoids, which morphologically characterize ductile fractures.

Figure 3
Scanning electron microscopy images of new (A, B, and C; G, H, and I; M, N, and O) and used (D, E, and F; J, K, and L; P, Q, and R) instruments of BT-Race, Sequence Rotary File, and ProDesign Logic systems, respectively, after torsional fatigue testing.

Figure 4
Scanning electron microscopy images of new (A, C, and E) and used (B, D, and F) instruments of BT-Race, Sequence Rotary File, and ProDesign Logic systems, respectively, after cyclic fatigue testing.

Discussion

This study evaluated the quality of preparation of curved canals, torsional fatigue, and cyclic fatigue of new and used instruments of three rotary systems with different types of NiTi alloys. In view of the results obtained, the first null hypothesis was accepted, and the others rejected, given that there was no interference in the quality of the preparation between the systems; however, there were differences in the torsional and cyclic fatigue resistance tests, in addition to interference of the simulated clinical use in the properties of glide path instruments.

The study specimens consisted of single-rooted teeth with curvature between 15º and 30º. Despite the difficulty in pairing samples of natural teeth, they were standardized in terms of degree of curvature and assessed by digital radiography. The baseline volume of both the entire canal and the apical third were standardized by micro-CT.

Curved canals pose a great challenge during mechanical preparation because errors such as zipping, step, excessive wear, and instrument fractures can occur, worsening the prognosis of endodontic treatment.2727.Liu W, Wu B. Root canal surface strain and canal center transportation induced by 3 different nickel-titanium rotary instrument systems. J Endod. 2016 Feb;42(2):299-303. https://doi.org/10.1016/j.joen.2015.10.023
https://doi.org/10.1016/j.joen.2015.10.0...
, 2828.Yammine S, Jabbour E, Nahas P, Majzoub Z. Foramen changes following over instrumentation of curved canals with three engine-driven instruments: an in vitro study. Iran Endod J. 2017;12(4):454-61. https://doi.org/10.22037/iej.v12i4.16268
https://doi.org/10.22037/iej.v12i4.16268...
In addition, there is still no consensus about the ideal diameter of apical preparation; however, several studies have demonstrated numerous benefits when this preparation was larger than that obtained with a #25 instrument.1616.Rodrigues RC, Zandi H, Kristoffersen AK, Enersen M, Mdala I, Ørstavik D, et al. Influence of the apical preparation size and the irrigant type on bacterial reduction in root canal-treated teeth with apical periodontitis. J Endod. 2017 Jul;43(7):1058-63. https://doi.org/10.1016/j.joen.2017.02.004
https://doi.org/10.1016/j.joen.2017.02.0...
, 1717.Pérez AR, Alves FR, Marceliano-Alves MF, Provenzano JC, Gonçalves LS, Neves AA, et al. Effects of increased apical enlargement on the amount of unprepared areas and coronal dentine removal: a micro-computed tomography study. Int Endod J. 2018 Jun;51(6):684-90. https://doi.org/10.1111/iej.12873
https://doi.org/10.1111/iej.12873...
, 2929.Siqueira JF Jr, Pérez AR, Marceliano-Alves MF, Provenzano JC, Silva SG, Pires FR, et al. What happens to unprepared root canal walls: a correlative analysis using micro-computed tomography and histology/scanning electron microscopy. Int Endod J. 2018 May;51(5):501-8. https://doi.org/10.1111/iej.12753
https://doi.org/10.1111/iej.12753...
Therefore, three rotary systems were investigated with the use of a size 35 finishing instrument. Probably, because of standardization of the final diameter and similar taper (BTR 0.04, SRF 0.04, and PDL 0.05), there was no significant difference among the groups regarding percent volume increase, even though the final instruments had different types of cross sections and were made of NiTi alloys (BTR – triangular section and conventional NiTi, SRF – triangular convex section and NiTi Blue, and PDL – S-shaped section and NiTi CM).

During biomechanical preparation of curved canals, preservation of their path is an important factor.1212.Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
https://doi.org/10.1111/iej.12881...
, 2828.Yammine S, Jabbour E, Nahas P, Majzoub Z. Foramen changes following over instrumentation of curved canals with three engine-driven instruments: an in vitro study. Iran Endod J. 2017;12(4):454-61. https://doi.org/10.22037/iej.v12i4.16268
https://doi.org/10.22037/iej.v12i4.16268...
Therefore, we analyzed canal transportation and centering ability of the systems in the apical region (last 4 mm), which corresponds to the region of the curvature. Relative to canal transportation, the three groups exhibited values close to 0, i.e., lower than 0.3 – a threshold value that has no negative impact on the prognosis of endodontic treatment.1212.Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
https://doi.org/10.1111/iej.12881...
Regarding centering ability, based on the formula proposed by Gambill et al.,2323.Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod. 1996 Jul;22(7):369-75. https://doi.org/10.1016/S0099-2399(96)80221-4
https://doi.org/10.1016/S0099-2399(96)80...
it is known that the closer to 1, the better the centering ability. Although no significant differences were noted, PDL obtained values closer to 1, followed by SRF and BTR, respectively. This could be explained mainly by the types of alloys used in the systems. A CM alloy is used for PDL, and a Blue alloy is used for SRF. BTR is made of a conventional NiTi alloy, and several studies have demonstrated that heat-treated alloys have greater flexibility and better preservation of the canal path than do conventional alloys.1212.Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
https://doi.org/10.1111/iej.12881...
, 3030.Duque JA, Vivan RR, Cavenago BC, Amoroso-Silva PA, Bernardes RA, Vasconcelos BC, et al. Influence of NiTi alloy on the root canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems. J Appl Oral Sci. 2017 Jan-Feb;25(1):27-33. https://doi.org/10.1590/1678-77572016-0230
https://doi.org/10.1590/1678-77572016-02...
, 3131.Frota MM, Bernardes RA, Vivan RR, Vivacqua-Gomes N, Duarte MA, Vasconcelos BC. Debris extrusion and foraminal deformation produced by reciprocating instruments made of thermally treated NiTi wires. J Appl Oral Sci. 2018 Jan;26(0):e20170215. https://doi.org/10.1590/1678-7757-2017-0215
https://doi.org/10.1590/1678-7757-2017-0...

Because these systems are relatively new in the market, few studies have analyzed the quality of preparation. The majority of these studies have revealed results for the behavior of BTR that were similar to those found in the present study.88.Bürklein S, Mathey D, Schäfer E. Shaping ability of ProTaper NEXT and BT-RaCe nickel-titanium instruments in severely curved root canals. Int Endod J. 2015 Aug;48(8):774-81. https://doi.org/10.1111/iej.12375
https://doi.org/10.1111/iej.12375...
, 99.Brasil SC, Marceliano-Alves MF, Marques ML, Grillo JP, Lacerda MF, Alves FR, et al. Canal Transportation, Unprepared Areas, and Dentin Removal after Preparation with BT-RaCe and ProTaper Next Systems. J Endod. 2017 Oct;43(10):1683-7. https://doi.org/10.1016/j.joen.2017.04.012
https://doi.org/10.1016/j.joen.2017.04.0...
Pinheiro et al.1212.Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
https://doi.org/10.1111/iej.12881...
observed that despite no statistical difference, PDL had better centering ability than that observed for different heat-treated rotary systems, which is in line with the results of this study.

During biomechanical preparation of curved canals, another important factor is creation of the glide path, as it can improve the quality of preparation and reduce the risk of fracture of shaping instruments.44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
, 55.De-Deus G, Belladonna FG, Souza EM, Alves VO, Silva EJ, Rodrigues E, et al. Scouting Ability of 4 Pathfinding Instruments in Moderately Curved Molar Canals. J Endod. 2016 Oct;42(10):1540-4. https://doi.org/10.1016/j.joen.2016.07.001
https://doi.org/10.1016/j.joen.2016.07.0...
, 3232.Hartmann RC, Peters OA, de Figueiredo JA, Rossi-Fedele G. Association of manual or engine-driven glide path preparation with canal centring and apical transportation: a systematic review. Int Endod J. 2018 Nov;51(11):1239-52. https://doi.org/10.1111/iej.12943
https://doi.org/10.1111/iej.12943...
, 3333.Santos CB, Simões-Carvalho M, Perez R, Vieira VT, Antunes HS, Cavalcante DF, et al. Torsional fatigue resistance of r-pilot and waveone gold glider NiTi glide path reciprocating systems. Int Endod J. 2019 Jun;52(6):874-9. https://doi.org/10.1111/iej.13068
https://doi.org/10.1111/iej.13068...
Therefore, the majority of rotary systems have a specific instrument for this purpose. Generally, glide path instruments show high variability among systems in relation to tip diameter, taper, cross section, and type of NiTi alloy. These factors can contribute to different behaviors of the instruments.44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
In the present study, glide path instruments were used in three teeth with curved canals and then subjected to the torsional fatigue test in order to analyze maximum torque (N.cm) and maximum angular deflection (o). Concomitantly, brand-new instruments of exactly the same type were also subjected to the same test using a previously validated methodology.77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
, 3333.Santos CB, Simões-Carvalho M, Perez R, Vieira VT, Antunes HS, Cavalcante DF, et al. Torsional fatigue resistance of r-pilot and waveone gold glider NiTi glide path reciprocating systems. Int Endod J. 2019 Jun;52(6):874-9. https://doi.org/10.1111/iej.13068
https://doi.org/10.1111/iej.13068...

Analysis of maximum torque of the new and used instruments showed that SRF 15.04 produced higher values than did PDL 25.01 and BTR 10.06. These findings could be explained by the fact that SRF has a quadrangular cross section and a larger tip diameter than does BTR, which has a triangular cross section, with a smaller amount of metal mass in its initial millimeters. Analysis of maximum torque of the new and used instruments showed that SRF 15.04 produced higher values than did PDL 25.01 and BTR 10.06. These findings could be explained by the fact that SRF has a quadrangular cross section and a larger tip diameter than does BTR, which has a triangular cross section, with a smaller amount of metal mass in its initial millimeters. The fact that torsional fatigue did not change in SRF and PDL after clinical use may be related to the heat treatment to which both were subjected. As the alloy used in the manufacture of BTR does not undergo heat treatment, its torsional fatigue resistance decreased, corroborating the findings of a previous study that analyzed instruments made of conventional NiTi alloys.1313.Vieira EP, Nakagawa RK, Buono VT, Bahia MG. Torsional behaviour of rotary NiTi ProTaper Universal instruments after multiple clinical use. Int Endod J. 2009 Oct;42(10):947-53. https://doi.org/10.1111/j.1365-2591.2009.01602.x
https://doi.org/10.1111/j.1365-2591.2009...

In regards to angular deflection, in the evaluation of both new and used instruments, PDL 25.01 obtained significantly better results than did SRF 15.04 and BTR 10.06, respectively. This variable is directly related to the type of heat treatment and taper of this instrument, so the more flexible the instrument, the greater the angle of deflection.44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
, 3636.Nishijo M, Ebihara A, Tokita D, Doi H, Hanawa T, Okiji T. Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires. Dent Mater J. 2018 Jul;37(4):549-54. https://doi.org/10.4012/dmj.2017-276
https://doi.org/10.4012/dmj.2017-276...
There are no studies in the literature evaluating the torsional fatigue of BTR and SRF systems for comparison of the results obtained. Only one study evaluated PDL 25.01 and obtained results similar to those of the present study.44.Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
https://doi.org/10.1111/iej.12879...
In addition, it should be noted that the angular deflection of the three instruments was affected by clinical use. This may have occurred because the instruments used in the preparation of the curved canals were not only subjected to torsional fatigue, but also to cyclic fatigue, reducing angular deflection.

After creating the glide path, the root canal is shaped by the instruments. When these instruments work in a curved canal, they undergo cyclic fatigue because of tensile and compressive forces at the maximum point of curvature.22.Marks Duarte P, Silva PB, Alcalde MP, Vivan RR, Rosa RA, Duarte MA, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018 Sep;44(9):1425-9. https://doi.org/10.1016/j.joen.2018.06.002
https://doi.org/10.1016/j.joen.2018.06.0...
, 3737.Silva EJ, Vieira VT, Hecksher F, Oliveira MRS, Antunes HS, Moreira EJ. Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2018 Sep;22(7):2633-8. https://doi.org/10.1007/s00784-018-2362-9
https://doi.org/10.1007/s00784-018-2362-...
The new and used finishing instruments (size 35) were subjected to the cyclic fatigue test in the present investigation to verify the extent of their safe use in canals with severe curvature. In the analysis of the time necessary until fracture of the instruments, both new and used, PDL 35.05 demanded significantly longer time than did SRF 35.04 and BTR 35.04, respectively. The same results were also observed for the number of cycles, as this varies according to the speed used. In the literature, there is a paucity of studies related to the mechanical properties of the systems tested. In the study of de Menezes et al.,3838.Menezes SE, Batista SM, Lira JO, Monteiro GQM. Cyclic Fatigue resistance of waveone gold, prodesign r and prodesign logic files in curved canals in vitro. Iran Endod J. 2017;12(4):468-73. https://doi.org/10.22037/iej.v12i4.17494
https://doi.org/10.22037/iej.v12i4.17494...
better behavior was also found for PDL, even when it was compared with heat-treated Gold instruments. These results might have been obtained owing to the instrument design (S-shaped cross section), but they were mainly attributable to CM heat treatment of PDL, which provided greater flexibility and resistance to cyclic fatigue.77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
, 2525.Alcalde MP, Tanomaru-Filho M, Bramante CM, Duarte MA, Guerreiro-Tanomaru JM, Camilo-Pinto J, et al. Cyclic and Torsional Fatigue Resistance of Reciprocating Single Files Manufactured by Different Nickel-titanium Alloys. J Endod. 2017 Jul;43(7):1186-91. https://doi.org/10.1016/j.joen.2017.03.008
https://doi.org/10.1016/j.joen.2017.03.0...
, 3838.Menezes SE, Batista SM, Lira JO, Monteiro GQM. Cyclic Fatigue resistance of waveone gold, prodesign r and prodesign logic files in curved canals in vitro. Iran Endod J. 2017;12(4):468-73. https://doi.org/10.22037/iej.v12i4.17494
https://doi.org/10.22037/iej.v12i4.17494...
With respect to the intermediate results obtained for SRF, which has a Blue heat treatment, an analogy can be made with the findings of Alcalde et al.77.Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
https://doi.org/10.1007/s00784-017-2295-...
and Silva et al.,3737.Silva EJ, Vieira VT, Hecksher F, Oliveira MRS, Antunes HS, Moreira EJ. Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2018 Sep;22(7):2633-8. https://doi.org/10.1007/s00784-018-2362-9
https://doi.org/10.1007/s00784-018-2362-...
in which the instrument with Blue heat treatment obtained good results; nevertheless, they were worse than those of the instrument with CM heat treatment. Concerning BTR, made of conventional NiTi alloy, several studies have shown this type of alloy has less resistance to cyclic fatigue than do NiTi instruments with some type of heat treatment.3939.Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod. 2010 Jul;36(7):1205-9. https://doi.org/10.1016/j.joen.2010.02.015
https://doi.org/10.1016/j.joen.2010.02.0...

40.Kaval ME, Capar ID, Ertas H. Evaluation of the cyclic fatigue and torsional resistance of novel nickel-titanium rotary files with various alloy properties. J Endod. 2016 Dec;42(12):1840-3. https://doi.org/10.1016/j.joen.2016.07.015
https://doi.org/10.1016/j.joen.2016.07.0...
- 4141.Silva EJ, Rodrigues C, Vieira VT, Belladonna FG, De-Deus G, Lopes HP. Bending resistance and cyclic fatigue of a new heat-treated reciprocating instrument. Scanning. 2016 Nov;38(6):837-41. https://doi.org/10.1002/sca.21333
https://doi.org/10.1002/sca.21333...
Another factor that could influence cyclic fatigue resistance is the speed at which each instrument was used. However, one study revealed no influence of speed when the instrument design and type of alloy of which it was made were considered to be the most important factors.3939.Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod. 2010 Jul;36(7):1205-9. https://doi.org/10.1016/j.joen.2010.02.015
https://doi.org/10.1016/j.joen.2010.02.0...
Note that none of the finishing instruments tested in the present study were significantly affected by clinical use. This may have occurred because the test was performed with the finishing instrument, which does not undergo as much stress as the first type used for shaping.

Conclusion

Despite the limitations of this study, mainly with regard to obtaining a similar sample, it could be concluded that BTR, SRF, and PDL rotary instruments were safe and produced a similar quality of shaping in canals with moderate curvature. Yet, different behaviors were observed in the mechanical properties of the instruments. For torsional fatigue, SRF 15.04 exhibited the highest torque values, PDL 25.01 had the highest angular deflection values, and BTR 10.06 was the most frequently affected by simulated clinical use. PDL 35.05 showed higher resistance to cyclic fatigue, followed by SRF 35.04, and the finishing instruments were not affected by simulated clinical use.

Acknowledgments

The authors declare that they have no conflicts of interest. The authors thank FAPESP for supporting this research (Fapesp Grant No. 2016/19956-5). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee. For this type of study, formal consent is not required.

References

  • 1
    De-Deus G, Silva EJ, Vieira VT, Belladonna FG, Elias CN, Plotino G, et al. Blue Thermomechanical Treatment Optimizes Fatigue Resistance and Flexibility of the Reciproc Files. J Endod. 2017 Mar;43(3):462-6. https://doi.org/10.1016/j.joen.2016.10.039
    » https://doi.org/10.1016/j.joen.2016.10.039
  • 2
    Marks Duarte P, Silva PB, Alcalde MP, Vivan RR, Rosa RA, Duarte MA, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018 Sep;44(9):1425-9. https://doi.org/10.1016/j.joen.2018.06.002
    » https://doi.org/10.1016/j.joen.2018.06.002
  • 3
    Arias A, Singh R, Peters OA. Differences in torsional performance of single- and multiple-instrument rotary systems for glide path preparation. Odontology. 2016 May;104(2):192-8. https://doi.org/10.1007/s10266-015-0199-0
    » https://doi.org/10.1007/s10266-015-0199-0
  • 4
    Alcalde MP, Duarte MA, Bramante CM, Tanomaru-Filho M, Vasconcelos BC, Só MV, et al. Torsional fatigue resistance of pathfinding instruments manufactured from several nickel-titanium alloys. Int Endod J. 2018 Jun;51(6):697-704. https://doi.org/10.1111/iej.12879
    » https://doi.org/10.1111/iej.12879
  • 5
    De-Deus G, Belladonna FG, Souza EM, Alves VO, Silva EJ, Rodrigues E, et al. Scouting Ability of 4 Pathfinding Instruments in Moderately Curved Molar Canals. J Endod. 2016 Oct;42(10):1540-4. https://doi.org/10.1016/j.joen.2016.07.001
    » https://doi.org/10.1016/j.joen.2016.07.001
  • 6
    Goo HJ, Kwak SW, Ha JH, Pedullà E, Kim HC. Mechanical Properties of Various Heat-treated Nickel-titanium Rotary Instruments. J Endod. 2017 Nov;43(11):1872-7. https://doi.org/10.1016/j.joen.2017.05.025
    » https://doi.org/10.1016/j.joen.2017.05.025
  • 7
    Alcalde MP, Duarte MA, Bramante CM, Vasconselos BC, Tanomaru-Filho M, Guerreiro-Tanomaru JM, et al. Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments. Clin Oral Investig. 2018 May;22(4):1865-71. https://doi.org/10.1007/s00784-017-2295-8
    » https://doi.org/10.1007/s00784-017-2295-8
  • 8
    Bürklein S, Mathey D, Schäfer E. Shaping ability of ProTaper NEXT and BT-RaCe nickel-titanium instruments in severely curved root canals. Int Endod J. 2015 Aug;48(8):774-81. https://doi.org/10.1111/iej.12375
    » https://doi.org/10.1111/iej.12375
  • 9
    Brasil SC, Marceliano-Alves MF, Marques ML, Grillo JP, Lacerda MF, Alves FR, et al. Canal Transportation, Unprepared Areas, and Dentin Removal after Preparation with BT-RaCe and ProTaper Next Systems. J Endod. 2017 Oct;43(10):1683-7. https://doi.org/10.1016/j.joen.2017.04.012
    » https://doi.org/10.1016/j.joen.2017.04.012
  • 10
    Krokidis A, Bonfanti C, Cerutti A, Barabanti N, Zinelis S, Panopoulos P. Comparative analysis of SAF, Protaper Next and BT-Race in eliminating Enterococcus faecalis from long oval canals: an ex vivo study. Aust Endod J. 2017 Dec;43(3):110-4. https://doi.org/10.1111/aej.12177
    » https://doi.org/10.1111/aej.12177
  • 11
    Coelho BS, Amaral RO, Leonardi DP, Marques-da-Silva B, Silva-Sousa YT, Carvalho FM, et al. Performance of Three single instrument systems in the preparation of long oval canals. Braz Dent J. 2016 Mar-Apr;27(2):217-22. https://doi.org/10.1590/0103-6440201302449
    » https://doi.org/10.1590/0103-6440201302449
  • 12
    Pinheiro SR, Alcalde MP, Vivacqua-Gomes N, Bramante CM, Vivan RR, Duarte MA, et al. Evaluation of apical transportation and centring ability of five thermally treated NiTi rotary systems. Int Endod J. 2018 Jun;51(6):705-13. https://doi.org/10.1111/iej.12881
    » https://doi.org/10.1111/iej.12881
  • 13
    Vieira EP, Nakagawa RK, Buono VT, Bahia MG. Torsional behaviour of rotary NiTi ProTaper Universal instruments after multiple clinical use. Int Endod J. 2009 Oct;42(10):947-53. https://doi.org/10.1111/j.1365-2591.2009.01602.x
    » https://doi.org/10.1111/j.1365-2591.2009.01602.x
  • 14
    Bueno CS, Oliveira DP, Pelegrine RA, Fontana CE, Rocha DG, Bueno CE. Fracture incidence of waveone and reciproc files during root canal preparation of up to 3 posterior teeth: a prospective clinical study. J Endod. 2017 May;43(5):705-8. https://doi.org/10.1016/j.joen.2016.12.024
    » https://doi.org/10.1016/j.joen.2016.12.024
  • 15
    Boutsioukis C, Gogos C, Verhaagen B, Versluis M, Kastrinakis E, Van der Sluis LW. The effect of apical preparation size on irrigant flow in root canals evaluated using an unsteady Computational Fluid Dynamics model. Int Endod J. 2010 Oct;43(10):874-81. https://doi.org/10.1111/j.1365-2591.2010.01761.x
    » https://doi.org/10.1111/j.1365-2591.2010.01761.x
  • 16
    Rodrigues RC, Zandi H, Kristoffersen AK, Enersen M, Mdala I, Ørstavik D, et al. Influence of the apical preparation size and the irrigant type on bacterial reduction in root canal-treated teeth with apical periodontitis. J Endod. 2017 Jul;43(7):1058-63. https://doi.org/10.1016/j.joen.2017.02.004
    » https://doi.org/10.1016/j.joen.2017.02.004
  • 17
    Pérez AR, Alves FR, Marceliano-Alves MF, Provenzano JC, Gonçalves LS, Neves AA, et al. Effects of increased apical enlargement on the amount of unprepared areas and coronal dentine removal: a micro-computed tomography study. Int Endod J. 2018 Jun;51(6):684-90. https://doi.org/10.1111/iej.12873
    » https://doi.org/10.1111/iej.12873
  • 18
    Vertucci FJ. Root canal anatomy of the human permanent teeth. Oral Surg Oral Med Oral Pathol. 1984 Nov;58(5):589-99. https://doi.org/10.1016/0030-4220(84)90085-9
    » https://doi.org/10.1016/0030-4220(84)90085-9
  • 19
    Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971 Aug;32(2):271-5. https://doi.org/10.1016/0030-4220(71)90230-1
    » https://doi.org/10.1016/0030-4220(71)90230-1
  • 20
    Fernandes PO, Freire LG, Iglecias EF, Vieira BR, Zuolo ML, Gavini G. Assessment of mechanical root canal preparation with centric reciprocating or eccentric rotary kinematics: a micro-computed tomographic study. J Endod. 2020 Sep;46(9):1309-16. https://doi.org/10.1016/j.joen.2020.06.005
    » https://doi.org/10.1016/j.joen.2020.06.005
  • 21
    Hassanien EE, Hashem A, Chalfin H. Histomorphometric study of the root apex of mandibular premolar teeth: an attempt to correlate working length measured with electronic and radiograph methods to various anatomic positions in the apical portion of the canal. J Endod. 2008 Apr;34(4):408-12. https://doi.org/10.1016/j.joen.2007.12.013
    » https://doi.org/10.1016/j.joen.2007.12.013
  • 22
    Wolf TG, Kim P, Campus G, Stiebritz M, Siegrist M, Briseño-Marroquín B. 3-dimensional analysis and systematic review of root canal morphology and physiological foramen geometry of 109 mandibular first premolars by micro-computed tomography in a mixed Swiss-German population. J Endod. 2020 Jun;46(6):801-9. https://doi.org/10.1016/j.joen.2020.03.002
    » https://doi.org/10.1016/j.joen.2020.03.002
  • 23
    Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod. 1996 Jul;22(7):369-75. https://doi.org/10.1016/S0099-2399(96)80221-4
    » https://doi.org/10.1016/S0099-2399(96)80221-4
  • 24
    Bahia MG, Melo MC, Buono VT. Influence of simulated clinical use on the torsional behavior of nickel-titanium rotary endodontic instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 May;101(5):675-80. https://doi.org/10.1016/j.tripleo.2005.04.019
    » https://doi.org/10.1016/j.tripleo.2005.04.019
  • 25
    Alcalde MP, Tanomaru-Filho M, Bramante CM, Duarte MA, Guerreiro-Tanomaru JM, Camilo-Pinto J, et al. Cyclic and Torsional Fatigue Resistance of Reciprocating Single Files Manufactured by Different Nickel-titanium Alloys. J Endod. 2017 Jul;43(7):1186-91. https://doi.org/10.1016/j.joen.2017.03.008
    » https://doi.org/10.1016/j.joen.2017.03.008
  • 26
    Klymus ME, Alcalde MP, Vivan RR, Só MV, Vasconselos BC, Duarte MA. Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2019 Jul;23(7):3047-52. https://doi.org/10.1007/s00784-018-2718-1
    » https://doi.org/10.1007/s00784-018-2718-1
  • 27
    Liu W, Wu B. Root canal surface strain and canal center transportation induced by 3 different nickel-titanium rotary instrument systems. J Endod. 2016 Feb;42(2):299-303. https://doi.org/10.1016/j.joen.2015.10.023
    » https://doi.org/10.1016/j.joen.2015.10.023
  • 28
    Yammine S, Jabbour E, Nahas P, Majzoub Z. Foramen changes following over instrumentation of curved canals with three engine-driven instruments: an in vitro study. Iran Endod J. 2017;12(4):454-61. https://doi.org/10.22037/iej.v12i4.16268
    » https://doi.org/10.22037/iej.v12i4.16268
  • 29
    Siqueira JF Jr, Pérez AR, Marceliano-Alves MF, Provenzano JC, Silva SG, Pires FR, et al. What happens to unprepared root canal walls: a correlative analysis using micro-computed tomography and histology/scanning electron microscopy. Int Endod J. 2018 May;51(5):501-8. https://doi.org/10.1111/iej.12753
    » https://doi.org/10.1111/iej.12753
  • 30
    Duque JA, Vivan RR, Cavenago BC, Amoroso-Silva PA, Bernardes RA, Vasconcelos BC, et al. Influence of NiTi alloy on the root canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems. J Appl Oral Sci. 2017 Jan-Feb;25(1):27-33. https://doi.org/10.1590/1678-77572016-0230
    » https://doi.org/10.1590/1678-77572016-0230
  • 31
    Frota MM, Bernardes RA, Vivan RR, Vivacqua-Gomes N, Duarte MA, Vasconcelos BC. Debris extrusion and foraminal deformation produced by reciprocating instruments made of thermally treated NiTi wires. J Appl Oral Sci. 2018 Jan;26(0):e20170215. https://doi.org/10.1590/1678-7757-2017-0215
    » https://doi.org/10.1590/1678-7757-2017-0215
  • 32
    Hartmann RC, Peters OA, de Figueiredo JA, Rossi-Fedele G. Association of manual or engine-driven glide path preparation with canal centring and apical transportation: a systematic review. Int Endod J. 2018 Nov;51(11):1239-52. https://doi.org/10.1111/iej.12943
    » https://doi.org/10.1111/iej.12943
  • 33
    Santos CB, Simões-Carvalho M, Perez R, Vieira VT, Antunes HS, Cavalcante DF, et al. Torsional fatigue resistance of r-pilot and waveone gold glider NiTi glide path reciprocating systems. Int Endod J. 2019 Jun;52(6):874-9. https://doi.org/10.1111/iej.13068
    » https://doi.org/10.1111/iej.13068
  • 34
    Acosta EC, Resende PD, Peixoto IF, Pereira ÉS, Buono VT, Bahia MG. Influence of cyclic flexural deformation on the torsional resistance of controlled memory and conventional nickel-titanium instruments. J Endod. 2017 Apr;43(4):613-8. https://doi.org/10.1016/j.joen.2016.11.007
    » https://doi.org/10.1016/j.joen.2016.11.007
  • 35
    Ha JH, Kwak SW, Kim SK, Sigurdsson A, Kim HC. Effect from rotational speed on torsional resistance of the nickel-titanium Instruments. J Endod. 2017 Mar;43(3):443-6. https://doi.org/10.1016/j.joen.2016.10.032
    » https://doi.org/10.1016/j.joen.2016.10.032
  • 36
    Nishijo M, Ebihara A, Tokita D, Doi H, Hanawa T, Okiji T. Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires. Dent Mater J. 2018 Jul;37(4):549-54. https://doi.org/10.4012/dmj.2017-276
    » https://doi.org/10.4012/dmj.2017-276
  • 37
    Silva EJ, Vieira VT, Hecksher F, Oliveira MRS, Antunes HS, Moreira EJ. Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments. Clin Oral Investig. 2018 Sep;22(7):2633-8. https://doi.org/10.1007/s00784-018-2362-9
    » https://doi.org/10.1007/s00784-018-2362-9
  • 38
    Menezes SE, Batista SM, Lira JO, Monteiro GQM. Cyclic Fatigue resistance of waveone gold, prodesign r and prodesign logic files in curved canals in vitro. Iran Endod J. 2017;12(4):468-73. https://doi.org/10.22037/iej.v12i4.17494
    » https://doi.org/10.22037/iej.v12i4.17494
  • 39
    Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod. 2010 Jul;36(7):1205-9. https://doi.org/10.1016/j.joen.2010.02.015
    » https://doi.org/10.1016/j.joen.2010.02.015
  • 40
    Kaval ME, Capar ID, Ertas H. Evaluation of the cyclic fatigue and torsional resistance of novel nickel-titanium rotary files with various alloy properties. J Endod. 2016 Dec;42(12):1840-3. https://doi.org/10.1016/j.joen.2016.07.015
    » https://doi.org/10.1016/j.joen.2016.07.015
  • 41
    Silva EJ, Rodrigues C, Vieira VT, Belladonna FG, De-Deus G, Lopes HP. Bending resistance and cyclic fatigue of a new heat-treated reciprocating instrument. Scanning. 2016 Nov;38(6):837-41. https://doi.org/10.1002/sca.21333
    » https://doi.org/10.1002/sca.21333

Publication Dates

  • Publication in this collection
    10 June 2022
  • Date of issue
    2022

History

  • Received
    5 Feb 2020
  • Accepted
    24 Feb 2022