Insertion angle of orthodontic mini-implants and their biomechanical performance: finite element analysis

Ângulo de inserção de mini-implantes ortodônticos e seu desempenho biomecânico: análise de elementos finitos

Abstract

Objective

The aim of this study was to assess the stresses and strains generated after the application of two types of forces (traction of 200 gf and torsion of 20 N.cm) in two types of orthodontic mini-implants inserted at different (45° and 90° to the cortical bone) angles.

Material and method

three-dimensional models of two brands of mini-implant (SIN – Sao Paulo, Brazil, and RMO – South Korea) were exported and analyzed by finite element analysis (FEA). Analyses were performed on simulations of cortical bone, cancellous bone and the screw.

Result

FEA analysis showed that RMO mini-implants had greater elastic deformation when subjected to tensile and torsional forces when compared with SIN mini-implants. For both trademarks and insertion angles tested, there was greater cortical bone deformation, but with the greatest strain located on the mini-implant. Tension on the mini-implant was located in its transmucosal profile region.

Conclusion

When comparing the two brands of mini-implants by FEA, it is fair to conclude that that the larger number of threads and their greater angle of inclination resulted in less resistance to deformation and induced a higher level of tension in the mini-implant and cortical bone when subjected to forces, especially when inserted at an angle of 45º to the cortical bone.

Descriptors:
Orthodontic anchorage procedures; orthodontic; finite element analysis

Resumo

Objetivo

O objetivo deste estudo foi avaliar as tensões e deformações de duas marcas comerciais de mini-implantes ortodônticos geradas após a aplicação de dois tipos de forças (de tração de 200 gf e torção de 20 N.cm) inseridos em duas angulações (45° e 90° em relação ao osso cortical).

Material e método

Modelos tridimensionais das duas marcas de mini-implantes (SIN - Sao Paulo, Brasil, e RMO – Coréia do Sul) foram construídos e analisados por análise de elementos finitos (FEA). As análises foram realizadas em simulações no osso cortical, osso esponjoso e no parafuso.

Resultado

A análise FEA mostrou que os mini-implantes da marca RMO apresentaram maior deformação elástica quando submetidos à tração e as forças de torção quando comparado aos mini-implantes da marca SIN. Em ambas as marcas testadas, e para os diferentes ângulos de inserção, houve uma maior deformação do osso cortical, com maior tensão localizado no mini-implante. A tensão no mini-implante foi localizado na região do perfil transmucoso.

Conclusão

Ao comparar as análises de elementos finitos das duas marcas comerciais de mini-implantes, concluiu-se que um maior número de roscas e maior inclinação resultam em menor resistência à deformação e induzem uma maior tensão no osso cortical quando submetidos à forças de torção e tração, especialmente quando inserido em um ângulo de 45º com o osso cortical.

Descritores:
Procedimentos de ancoragem ortodôntica; ortodontia; análise de elementos finitos

INTRODUCTION

The use of orthodontic mini-implants (OMI) has changed many concepts in orthodontics11. Liu TC, Chang CH, Wong TY, Liu JK. Finite element analysis of miniscrew implants used for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2012 Apr;141(4):468-76. http://dx.doi.org/10.1016/j.ajodo.2011.11.012. PMid:22464529.
http://dx.doi.org/10.1016/j.ajodo.2011.1...
,22. Crismani AG, Bertl MH, Celar AG, Bantleon HP, Burstone CJ. Miniscrews in orthodontic treatment: review and analysis of published clinical trials. Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):108-13. http://dx.doi.org/10.1016/j.ajodo.2008.01.027. PMid:20122438.
http://dx.doi.org/10.1016/j.ajodo.2008.0...
. The possibility of moving teeth without causing unwanted side-effects has allowed the treatment of more complex cases, and added predictability of the outcome of treatment plans11. Liu TC, Chang CH, Wong TY, Liu JK. Finite element analysis of miniscrew implants used for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2012 Apr;141(4):468-76. http://dx.doi.org/10.1016/j.ajodo.2011.11.012. PMid:22464529.
http://dx.doi.org/10.1016/j.ajodo.2011.1...
,22. Crismani AG, Bertl MH, Celar AG, Bantleon HP, Burstone CJ. Miniscrews in orthodontic treatment: review and analysis of published clinical trials. Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):108-13. http://dx.doi.org/10.1016/j.ajodo.2008.01.027. PMid:20122438.
http://dx.doi.org/10.1016/j.ajodo.2008.0...
. To reduce the failure rate of OMIs (10-30%), several biological and biomechanical factors associated with their performance have been studied33. Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8. http://dx.doi.org/10.1016/S0889-5406(03)00565-1. PMid:14560266.
http://dx.doi.org/10.1016/S0889-5406(03)...

4. Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T. Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S68-73. http://dx.doi.org/10.1016/j.ajodo.2006.06.017. PMid:17448389.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
-55. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants. 2004 Jan-Feb;19(1):100-6. PMid:14982362.. From the literature, it can be speculated that the design of the OMI, as well as the inclination angle in which it was inserted, are factors of major relevance33. Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8. http://dx.doi.org/10.1016/S0889-5406(03)00565-1. PMid:14560266.
http://dx.doi.org/10.1016/S0889-5406(03)...

4. Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T. Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S68-73. http://dx.doi.org/10.1016/j.ajodo.2006.06.017. PMid:17448389.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
-55. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants. 2004 Jan-Feb;19(1):100-6. PMid:14982362..

One of the methodologies that can be used to assess the biomechanical impact of possible changes in the design and in the insertion angle of OMIs is the finite element analysis (FEA). FEA is a methodology that relies on the “discretization” of large structures to smaller elements of known size, which are net-connected and capable of being mathematically interpreted66. Williams KR, Edmundson JT. Orthodontic tooth movement analysed by the Finite Element Method. Biomaterials. 1984 Nov;5(6):347-51. http://dx.doi.org/10.1016/0142-9612(84)90033-4. PMid:6525394.
http://dx.doi.org/10.1016/0142-9612(84)9...

7. Lotti RS, Machado AW, Mazzieiro ET, Landre Júnior J. Aplicabilidade científica do método dos elementos finitos. Rev Dent Press Ortodon Ortop Facial. 2006 Mar-Apr;11(2):35-43. http://dx.doi.org/10.1590/S1415-54192006000200006.
http://dx.doi.org/10.1590/S1415-54192006...
-88. Middleton J, Jones ML, Wilson AN. Three-dimensional analysis of orthodontic tooth movement. J Biomed Eng. 1990 July;12(4):319-27. http://dx.doi.org/10.1016/0141-5425(90)90007-A. PMid:2395358.
http://dx.doi.org/10.1016/0141-5425(90)9...
. The main advantage of using FEA to evaluate the biological performance of an OMI is that this methodology is based on a virtual environment, saves time, costs, and avoids the use of animals66. Williams KR, Edmundson JT. Orthodontic tooth movement analysed by the Finite Element Method. Biomaterials. 1984 Nov;5(6):347-51. http://dx.doi.org/10.1016/0142-9612(84)90033-4. PMid:6525394.
http://dx.doi.org/10.1016/0142-9612(84)9...

7. Lotti RS, Machado AW, Mazzieiro ET, Landre Júnior J. Aplicabilidade científica do método dos elementos finitos. Rev Dent Press Ortodon Ortop Facial. 2006 Mar-Apr;11(2):35-43. http://dx.doi.org/10.1590/S1415-54192006000200006.
http://dx.doi.org/10.1590/S1415-54192006...

8. Middleton J, Jones ML, Wilson AN. Three-dimensional analysis of orthodontic tooth movement. J Biomed Eng. 1990 July;12(4):319-27. http://dx.doi.org/10.1016/0141-5425(90)90007-A. PMid:2395358.
http://dx.doi.org/10.1016/0141-5425(90)9...
-99. Rubin C, Krishnamurthy N, Capilouto E, Yi H. Stress analysis of the human tooth using a three-dimensional finite element model. J Dent Res. 1983 Feb;62(2):82-6. http://dx.doi.org/10.1177/00220345830620021701. PMid:6571871.
http://dx.doi.org/10.1177/00220345830620...
. The literature has shown a positive correlation between the biomechanical characteristics of OMIs and their in vivoperformance1010. Chatzigianni A, Keilig L, Duschner H, Götz H, Eliades T, Bourauel C. Comparative analysis of numerical and experimental data of orthodontic mini-implants. Eur J Orthod. 2011 Oct;33(5):468-75. http://dx.doi.org/10.1093/ejo/cjr097. PMid:21852288.
http://dx.doi.org/10.1093/ejo/cjr097...
.

Previous studies using FEA have shown the quality of trabecular bone in the region of OMI insertion is not critical for its stability, however, a layer of cortical bone at least 1 mm thick is1111. Yu W, Park HS, Kyung HM, Kwon OW. Dynamic simulation of the self-tapping insertion process of orthodontic microimplants into cortical bone with a 3-dimensional finite element method. Am J Orthod Dentofacial Orthop. 2012 Dec;142(6):834-41. http://dx.doi.org/10.1016/j.ajodo.2012.08.016. PMid:23195369.
http://dx.doi.org/10.1016/j.ajodo.2012.0...
. On the other hand, the thinner the cortical bone, the larger the tension the OMI will cause to the bone, increasing the risk of implant failure1212. Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect of cortical bone thickness on the stability of orthodontic mini-implants and on the stress distribution in surrounding bone. Int J Oral Maxillofac Surg. 2009 Jan;38(1):13-8. http://dx.doi.org/10.1016/j.ijom.2008.09.006. PMid:18963818.
http://dx.doi.org/10.1016/j.ijom.2008.09...

13. Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F. Optimal palatal configuration for miniscrew applications. Angle Orthod. 2010 Jan;80(1):145-52. http://dx.doi.org/10.2319/122908-662.1. PMid:19852654.
http://dx.doi.org/10.2319/122908-662.1...
-1414. Holberg C, Winterhalder P, Rudzki-Janson I, Wichelhaus A. Finite element analysis of mono- and bicortical mini-implant stability. Eur J Orthod. 2014 Oct;36(5):550-6. http://dx.doi.org/10.1093/ejo/cjt023. PMid:23598610.
http://dx.doi.org/10.1093/ejo/cjt023...
. It has been shown that, Even the thick cortical bone layer of the palate region has been shown to be incapable of supporting forces greater than 480 gf1313. Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F. Optimal palatal configuration for miniscrew applications. Angle Orthod. 2010 Jan;80(1):145-52. http://dx.doi.org/10.2319/122908-662.1. PMid:19852654.
http://dx.doi.org/10.2319/122908-662.1...
. The shape and geometry (macrostructure) of an OMI are directly connected with the amount of tension it will induce into the bone1515. Singh S, Mogra S, Shetty VS, Shetty S, Philip P. Three-dimensional finite element analysis of strength, stability, and stress distribution in orthodontic anchorage: a conical, self-drilling miniscrew implant system. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):327-36. http://dx.doi.org/10.1016/j.ajodo.2011.07.022. PMid:22381493.
http://dx.doi.org/10.1016/j.ajodo.2011.0...

16. Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, et al. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):279-88. http://dx.doi.org/10.1016/j.ajodo.2011.09.008. PMid:22381488.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
-1717. Luiz NE, Jacomini A Fo, Lima LAC, Ciuccio RL, Soares MAD, Coutinho LL. Optimization design of orthodontic screws aiming at increasing the mechanical strength. Innov. Implant. J., Biomater. Esthet. 2010 Aug;5(2): 30-34.. The region in which the highest tensions to the cortical bone are detected is the first thread of the OMI1818. Marcé-Nogué J, Walter A, Gil L, Puigdollers A. Finite element comparison of 10 orthodontic microscrews with different cortical bone parameters. Int J Oral Maxillofac Implants. 2013 July-Aug;28(4):e177-89. http://dx.doi.org/10.11607/jomi.2447. PMid:23869375.
http://dx.doi.org/10.11607/jomi.2447...
,1919. Motoyoshi M, Inaba M, Ueno S, Shimizu N. Mechanical anisotropy of orthodontic mini-implants. Int J Oral Maxillofac Surg. 2009 Sept;38(9):972-7. http://dx.doi.org/10.1016/j.ijom.2009.05.009. PMid:19559569.
http://dx.doi.org/10.1016/j.ijom.2009.05...
. Furthermore, the angle at which the OMI is inserted might alter this parameter33. Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8. http://dx.doi.org/10.1016/S0889-5406(03)00565-1. PMid:14560266.
http://dx.doi.org/10.1016/S0889-5406(03)...
.

To provide a better understanding of the parameters related to OMI success, the aim of this study was to assess the tensions and deformations (stresses and strains) generated after the application of two types of forces (traction and torsion) in two types of OMIs inserted at different angles.

MATERIAL AND METHOD

Two brands of OMIs were tested: SIN (Implant System, São Paulo/SP, Brazil) and RMO (Rocky Mountain Orthodontics, Seoul, South Korea). They were both self-drilling, measured 1.6 mm in diameter and 6 mm long. Further details of implants tested are shown in Table 1.

Table 1
Implant characteristics

Tridimensional models of the implants were constructed using dedicated software (Autodesk Inc., San Raphael, CA - USA), based on reverse engineering: the implants were measured under a stereomicroscope, and the measurements established were used to construct their 3D models. Two simplified bone blocks were also constructed using the same software, and they simulated of two layers of bone: one of cortical bone (1.5 mm thick), and one of trabecular bone (18.5 mm thick). The interface between the two layers of bone was considered a continuous line. Tridimensional models were then exported and analyzed by a single trained operator using FEA dedicated software (ANSYS Workbench 14.5, Swanson Analysis Systems, Canonsburg, PA - USA). All materials were considered as homogeneous and isotropic, linearly elastic, and the biomechanical properties attributed to them are shown in Table 22020. Meijer HJ, Starmans FJ, Steen WH, Bosman F. A three-dimensional, finite-element analysis of bone around dental implants in an edentulous human mandible. Arch Oral Biol. 1993 June;38(6):491-6. http://dx.doi.org/10.1016/0003-9969(93)90185-O. PMid:8343071.
http://dx.doi.org/10.1016/0003-9969(93)9...

21. Menicucci G, Mossolov A, Mozzati M, Lorenzetti M, Preti G. Tooth-implant connection: some biomechanical aspects based on finite element analyses. Clin Oral Implants Res. 2002 June;13(3):334-41. http://dx.doi.org/10.1034/j.1600-0501.2002.130315.x. PMid:12010166.
http://dx.doi.org/10.1034/j.1600-0501.20...

22. American Society for Metals. Metals Handbook. 10th ed. Metals Park: ASM; 1990. Vol. 2, Properties and selection: nonferrous alloys and special-purpose materials.

23. American Society for Metals. Metals Handbook. 9th ed. Metals Park: ASM; 1980. Vol. 3, Properties and selection: stainless steels, tool materials and special-purpose metals.
-2424. Holt JM, Mindlin H, Ho CY, editors. Structural alloys handbook. West Lafayette: CINDAS/Purdue University; 1996.. Assessments were made considering the implant already inserted and integrated into the bone1212. Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect of cortical bone thickness on the stability of orthodontic mini-implants and on the stress distribution in surrounding bone. Int J Oral Maxillofac Surg. 2009 Jan;38(1):13-8. http://dx.doi.org/10.1016/j.ijom.2008.09.006. PMid:18963818.
http://dx.doi.org/10.1016/j.ijom.2008.09...
,1313. Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F. Optimal palatal configuration for miniscrew applications. Angle Orthod. 2010 Jan;80(1):145-52. http://dx.doi.org/10.2319/122908-662.1. PMid:19852654.
http://dx.doi.org/10.2319/122908-662.1...
,1616. Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, et al. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):279-88. http://dx.doi.org/10.1016/j.ajodo.2011.09.008. PMid:22381488.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
,1818. Marcé-Nogué J, Walter A, Gil L, Puigdollers A. Finite element comparison of 10 orthodontic microscrews with different cortical bone parameters. Int J Oral Maxillofac Implants. 2013 July-Aug;28(4):e177-89. http://dx.doi.org/10.11607/jomi.2447. PMid:23869375.
http://dx.doi.org/10.11607/jomi.2447...
,2525. Lin TS, Tsai FD, Chen CY, Lin LW. Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis. Am J Orthod Dentofacial Orthop. 2013 Feb;143(2):182-9. http://dx.doi.org/10.1016/j.ajodo.2012.09.012. PMid:23374924.
http://dx.doi.org/10.1016/j.ajodo.2012.0...
.

Table 2
Tested materials properties

OMIs were tested under traction (200 gf) and torsion (20 N.cm) forces (commonly used for orthodontic movement and insertion of the mini-implant respectively), in addition to considering insertion in two directions: at angles of 90º and 45º to the cortical bone. Traction forces were applied in the opposite direction to OMI inclination. Assessments were made at the level of cortical bone, trabecular bone, and OMI, separately.

RESULT

Tables 3 and 4 show the tensions and deformations found for the conditions tested. For all tested setups, the highest tension was seen in the OMI, and the largest deformation was seen in the cortical bone (as shown in Figures 1, 2, 3 and 4). The worst results were observed when the inclination of 45 degrees was tested. Deformation was usually larger for RMO implants. Tension was higher for SIN implants after the use of traction forces, while it was higher for RMO implants after the use of torsion forces.

Table 3
Maximum tension (MPa) and deformation (x10-5, mm/mm) values assessed following the use of traction forces
Table 4
Maximum tension (MPa) and deformation (x10-3, mm/mm) values assessed following the use of torsion forces
Figure 1
Maximum deformation (mm/mm) of the OMI when inserted in the bone block, under a 200 gf traction force. a SIN 900, b SIN 450, c RMO 900, d RMO 450.
Figure 2
Maximum tension (MPa) induced on the OMI when inserted in the bone block, under a 200 gf traction force. a SIN 900, b SIN 450, c RMO 900, d RMO 450.
Figure 3
Maximum deformation (mm/mm) of the OMI when inserted in the bone block, under a 20 N.cm torsion force. a SIN 900, b SIN 450, c RMO 900, d RMO 450.
Figure 4
Maximum tension (MPa) induced on the OMI when inserted in the bone block, under a 20 N.cm torsion force. a SIN 900, b SIN 450, c RMO 900, d RMO 450.

DISCUSSION

The aim of this study was to assess the tensions and deformations (stresses and strains) generated after the application of two types of forces (traction and torsion) in two types of OMIs inserted at different angles, by means of FEA. Not only should the limitations of the selected method of analysis (the homogeneity of the considered bone blocks and the absence of biological-triggered response being the most relevant) be borne in mind, but also its well-known importance in precisely identifying and quantifying stresses and strains in pre-defined (and validated) models1515. Singh S, Mogra S, Shetty VS, Shetty S, Philip P. Three-dimensional finite element analysis of strength, stability, and stress distribution in orthodontic anchorage: a conical, self-drilling miniscrew implant system. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):327-36. http://dx.doi.org/10.1016/j.ajodo.2011.07.022. PMid:22381493.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
,2424. Holt JM, Mindlin H, Ho CY, editors. Structural alloys handbook. West Lafayette: CINDAS/Purdue University; 1996.. The loading protocol was identical for the two brands of OMI tested, in order to minimize the risk of bias. Furthermore, it could be speculated that the differences found are mainly related to the macroscopic features (design) of the tested OMIs, since both were fabricated of the same metal alloy (Ti6Al4Va).

When the brands tested were compared, RMO was shown to undergo greater deformation than SIN, even when subjected to tension forces. Therefore, it seems reasonable to assume that a smaller distance between threads, rather than their number, was responsible for the reduced resistance to elastic deformation of the screw. This speculation is based on the fact - according to the manufacturer's description – that the two implants tested were made of the same alloy, and were equally evaluated (e.g. size and conditions). These findings corroborate those previously published on the importance of the shape of the screw in the effectiveness of the technique1818. Marcé-Nogué J, Walter A, Gil L, Puigdollers A. Finite element comparison of 10 orthodontic microscrews with different cortical bone parameters. Int J Oral Maxillofac Implants. 2013 July-Aug;28(4):e177-89. http://dx.doi.org/10.11607/jomi.2447. PMid:23869375.
http://dx.doi.org/10.11607/jomi.2447...
.

With regard to stress distribution on the bone, tension in the cortical bone was far higher than that found in cancellous bone, showing the cortical bone to be responsible for the actual interlocking with the bone and the OMI. These data have been confirmed in literature, in which the quality of cortical bone has been shown to influence the primary stability of the mini-implant11. Liu TC, Chang CH, Wong TY, Liu JK. Finite element analysis of miniscrew implants used for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2012 Apr;141(4):468-76. http://dx.doi.org/10.1016/j.ajodo.2011.11.012. PMid:22464529.
http://dx.doi.org/10.1016/j.ajodo.2011.1...
. Therefore, the smaller the cortical bone thickness, the higher the tension induced into it, which may lead to resorption (and consequently failure) of the OMI1212. Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect of cortical bone thickness on the stability of orthodontic mini-implants and on the stress distribution in surrounding bone. Int J Oral Maxillofac Surg. 2009 Jan;38(1):13-8. http://dx.doi.org/10.1016/j.ijom.2008.09.006. PMid:18963818.
http://dx.doi.org/10.1016/j.ijom.2008.09...

13. Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F. Optimal palatal configuration for miniscrew applications. Angle Orthod. 2010 Jan;80(1):145-52. http://dx.doi.org/10.2319/122908-662.1. PMid:19852654.
http://dx.doi.org/10.2319/122908-662.1...

14. Holberg C, Winterhalder P, Rudzki-Janson I, Wichelhaus A. Finite element analysis of mono- and bicortical mini-implant stability. Eur J Orthod. 2014 Oct;36(5):550-6. http://dx.doi.org/10.1093/ejo/cjt023. PMid:23598610.
http://dx.doi.org/10.1093/ejo/cjt023...

15. Singh S, Mogra S, Shetty VS, Shetty S, Philip P. Three-dimensional finite element analysis of strength, stability, and stress distribution in orthodontic anchorage: a conical, self-drilling miniscrew implant system. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):327-36. http://dx.doi.org/10.1016/j.ajodo.2011.07.022. PMid:22381493.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
-1616. Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, et al. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):279-88. http://dx.doi.org/10.1016/j.ajodo.2011.09.008. PMid:22381488.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
. Although there was an increased amount of deformation in the cortical bone in all the situations tested, there was a greater distribution of elastic deformation when cancellous bone tensile strength was considered. It was evident that When the OMI was subjected to orthodontic forces, there was evident elastic deformation in the bone around the entire screw, which was more concentrated in the cervical region. SIN showed higher strain values ​​in cancellous bone, and the greater distribution in the body and apex of the screw could be a result of the type of thread tip used in this brand.

The results of this study showed that irrespective of force (tension or torsion), the OMI brand, or the inclination of the screw, there was always a greater strain on cortical bone and increased tension in the OMI. It could be speculated that the OMI should be stronger, especially in the transmucosal profile, the region that showed the highest level of stress concentration. Manufacturers would be prudent if they were to increase the resistance of the screw in this region, thus avoiding the occurrence of fractures during insertion. Other FEA-based studies have also observed an accumulation of stress in the same region of OMIs1515. Singh S, Mogra S, Shetty VS, Shetty S, Philip P. Three-dimensional finite element analysis of strength, stability, and stress distribution in orthodontic anchorage: a conical, self-drilling miniscrew implant system. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):327-36. http://dx.doi.org/10.1016/j.ajodo.2011.07.022. PMid:22381493.
http://dx.doi.org/10.1016/j.ajodo.2011.0...
,1717. Luiz NE, Jacomini A Fo, Lima LAC, Ciuccio RL, Soares MAD, Coutinho LL. Optimization design of orthodontic screws aiming at increasing the mechanical strength. Innov. Implant. J., Biomater. Esthet. 2010 Aug;5(2): 30-34..

Insertion at an angle of 45 degrees increased the tension and bone strain for both brands of OMI tested, and in both loading protocols tested. This was evidently caused by the larger contact area between the cortical bone and the OMI. This is similar to findings previously shown in the literature with regard to increase in the number of exposed threads of the OMI (as caused by the angled OMI) resulting in increased bone tension around the OMI2525. Lin TS, Tsai FD, Chen CY, Lin LW. Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis. Am J Orthod Dentofacial Orthop. 2013 Feb;143(2):182-9. http://dx.doi.org/10.1016/j.ajodo.2012.09.012. PMid:23374924.
http://dx.doi.org/10.1016/j.ajodo.2012.0...
. The authors also suggested that the traction force applied to an angled implant produced greater tension due to the increased lever arm which is formed, since a larger portion of the screw is outside of the bone2525. Lin TS, Tsai FD, Chen CY, Lin LW. Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis. Am J Orthod Dentofacial Orthop. 2013 Feb;143(2):182-9. http://dx.doi.org/10.1016/j.ajodo.2012.09.012. PMid:23374924.
http://dx.doi.org/10.1016/j.ajodo.2012.0...
.

A few studies in literature have suggested that OMIs should not be inserted perpendicular to the cortical bone2828. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51. http://dx.doi.org/10.1016/j.ajodo.2006.04.027. PMid:17448385.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
,2929. Büchter A, Wiechmann D, Koerdt S, Wiesmann HP, Piffko J, Meyer U. Load-related implant reaction of mini-implants used for orthodontic anchorage. Clin Oral Implants Res. 2005 Aug;16(4):473-9. http://dx.doi.org/10.1111/j.1600-0501.2005.01149.x. PMid:16117773.
http://dx.doi.org/10.1111/j.1600-0501.20...
. This would avoid them coming into contact with the roots of teeth, in addition to increasing the insertion torque and primary stability of the OMI2828. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51. http://dx.doi.org/10.1016/j.ajodo.2006.04.027. PMid:17448385.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
,2929. Büchter A, Wiechmann D, Koerdt S, Wiesmann HP, Piffko J, Meyer U. Load-related implant reaction of mini-implants used for orthodontic anchorage. Clin Oral Implants Res. 2005 Aug;16(4):473-9. http://dx.doi.org/10.1111/j.1600-0501.2005.01149.x. PMid:16117773.
http://dx.doi.org/10.1111/j.1600-0501.20...
. Nevertheless, a previous study has shown no significant difference in the success rate of OMIs inserted at an angle to the bone, when compared with OMIs inserted perpendicular to the bone2525. Lin TS, Tsai FD, Chen CY, Lin LW. Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis. Am J Orthod Dentofacial Orthop. 2013 Feb;143(2):182-9. http://dx.doi.org/10.1016/j.ajodo.2012.09.012. PMid:23374924.
http://dx.doi.org/10.1016/j.ajodo.2012.0...
. The major problem associated with angulated insertion is that it does not allow complete insertion of the OMI threads into the bone, leading to a larger lever arm formed off-the-bone, negatively contributing to primary and secondary stability, as well as increasing the possibility of forming niches where food may accumulate2828. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51. http://dx.doi.org/10.1016/j.ajodo.2006.04.027. PMid:17448385.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
. Based on FEA analyses of human hemi-mandibles (ex vivo) it has been shown that the greater the insertion angle of OMIs, the greater the tension induced in the cortical bone, and the lower the tensile strength when subjected to forces parallel to the occlusal plane2828. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51. http://dx.doi.org/10.1016/j.ajodo.2006.04.027. PMid:17448385.
http://dx.doi.org/10.1016/j.ajodo.2006.0...
. The distance between the cortical bone and the application of force varied from 2 to 4 mm, for insertion angles of 90 and 30 degrees, respectively. These findings corroborate those previously reported in literature showing that that the greater the distance between the bone and the application of force, the higher the OMI failure rate2929. Büchter A, Wiechmann D, Koerdt S, Wiesmann HP, Piffko J, Meyer U. Load-related implant reaction of mini-implants used for orthodontic anchorage. Clin Oral Implants Res. 2005 Aug;16(4):473-9. http://dx.doi.org/10.1111/j.1600-0501.2005.01149.x. PMid:16117773.
http://dx.doi.org/10.1111/j.1600-0501.20...
. A recent study also showed that best results for OMI were achieved when an insertion angle of 90 degrees was used3030. Lee J, Kim JY, Choi YJ, Kim KH, Chung CJ. Effects of placement angle and direction of orthopedic force application on the stability of orthodontic miniscrews. Angle Orthod. 2013 July;83(4):667-73. http://dx.doi.org/10.2319/090112-703.1. PMid:23241005.
http://dx.doi.org/10.2319/090112-703.1...
.

Divergences among the results of this and previous studies in the literature suggest the need of further in vitro and in vivo studies, to allow the formulation of a definitive rationale on the topic. However, from the mechanical point of view, a higher level of tension between the bone and the OMI would lead to a higher insertion torque (and therefore to better primary stability); the definition of a threshold, from which bone-tissue injuries would be induced, leading to implant failure should be considered2626. Chen Y, Kyung HM, Gao L, Yu WJ, Bae EJ, Kim SM. Mechanical properties of self-drilling orthodontic micro-implants with different diameters. Angle Orthod. 2010 Sept;80(5):821-7. http://dx.doi.org/10.2319/103009-607.1. PMid:20578851.
http://dx.doi.org/10.2319/103009-607.1...
,2727. Suzuki EY, Suzuki B. Placement and removal torque values of orthodontic miniscrew implants. Am J Orthod Dentofacial Orthop. 2011 May;139(5):669-78. http://dx.doi.org/10.1016/j.ajodo.2010.11.017. PMid:21536211.
http://dx.doi.org/10.1016/j.ajodo.2010.1...
.

When the FEA of the two brands of mini-implants were compared, it is fair to conclude that the larger number of threads and greater inclination of the threads resulted in less resistance to deformation and induced a higher level of tension in the mini-implant and cortical bone when subject to forces, especially when mini-implants were inserted at an angle of 45º to the cortical bone.

REFERENCES

  • 1
    Liu TC, Chang CH, Wong TY, Liu JK. Finite element analysis of miniscrew implants used for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2012 Apr;141(4):468-76. http://dx.doi.org/10.1016/j.ajodo.2011.11.012. PMid:22464529.
    » http://dx.doi.org/10.1016/j.ajodo.2011.11.012
  • 2
    Crismani AG, Bertl MH, Celar AG, Bantleon HP, Burstone CJ. Miniscrews in orthodontic treatment: review and analysis of published clinical trials. Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):108-13. http://dx.doi.org/10.1016/j.ajodo.2008.01.027. PMid:20122438.
    » http://dx.doi.org/10.1016/j.ajodo.2008.01.027
  • 3
    Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8. http://dx.doi.org/10.1016/S0889-5406(03)00565-1. PMid:14560266.
    » http://dx.doi.org/10.1016/S0889-5406(03)00565-1
  • 4
    Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T. Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S68-73. http://dx.doi.org/10.1016/j.ajodo.2006.06.017. PMid:17448389.
    » http://dx.doi.org/10.1016/j.ajodo.2006.06.017
  • 5
    Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants. 2004 Jan-Feb;19(1):100-6. PMid:14982362.
  • 6
    Williams KR, Edmundson JT. Orthodontic tooth movement analysed by the Finite Element Method. Biomaterials. 1984 Nov;5(6):347-51. http://dx.doi.org/10.1016/0142-9612(84)90033-4. PMid:6525394.
    » http://dx.doi.org/10.1016/0142-9612(84)90033-4
  • 7
    Lotti RS, Machado AW, Mazzieiro ET, Landre Júnior J. Aplicabilidade científica do método dos elementos finitos. Rev Dent Press Ortodon Ortop Facial. 2006 Mar-Apr;11(2):35-43. http://dx.doi.org/10.1590/S1415-54192006000200006.
    » http://dx.doi.org/10.1590/S1415-54192006000200006
  • 8
    Middleton J, Jones ML, Wilson AN. Three-dimensional analysis of orthodontic tooth movement. J Biomed Eng. 1990 July;12(4):319-27. http://dx.doi.org/10.1016/0141-5425(90)90007-A. PMid:2395358.
    » http://dx.doi.org/10.1016/0141-5425(90)90007-A
  • 9
    Rubin C, Krishnamurthy N, Capilouto E, Yi H. Stress analysis of the human tooth using a three-dimensional finite element model. J Dent Res. 1983 Feb;62(2):82-6. http://dx.doi.org/10.1177/00220345830620021701. PMid:6571871.
    » http://dx.doi.org/10.1177/00220345830620021701
  • 10
    Chatzigianni A, Keilig L, Duschner H, Götz H, Eliades T, Bourauel C. Comparative analysis of numerical and experimental data of orthodontic mini-implants. Eur J Orthod. 2011 Oct;33(5):468-75. http://dx.doi.org/10.1093/ejo/cjr097. PMid:21852288.
    » http://dx.doi.org/10.1093/ejo/cjr097
  • 11
    Yu W, Park HS, Kyung HM, Kwon OW. Dynamic simulation of the self-tapping insertion process of orthodontic microimplants into cortical bone with a 3-dimensional finite element method. Am J Orthod Dentofacial Orthop. 2012 Dec;142(6):834-41. http://dx.doi.org/10.1016/j.ajodo.2012.08.016. PMid:23195369.
    » http://dx.doi.org/10.1016/j.ajodo.2012.08.016
  • 12
    Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect of cortical bone thickness on the stability of orthodontic mini-implants and on the stress distribution in surrounding bone. Int J Oral Maxillofac Surg. 2009 Jan;38(1):13-8. http://dx.doi.org/10.1016/j.ijom.2008.09.006. PMid:18963818.
    » http://dx.doi.org/10.1016/j.ijom.2008.09.006
  • 13
    Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F. Optimal palatal configuration for miniscrew applications. Angle Orthod. 2010 Jan;80(1):145-52. http://dx.doi.org/10.2319/122908-662.1. PMid:19852654.
    » http://dx.doi.org/10.2319/122908-662.1
  • 14
    Holberg C, Winterhalder P, Rudzki-Janson I, Wichelhaus A. Finite element analysis of mono- and bicortical mini-implant stability. Eur J Orthod. 2014 Oct;36(5):550-6. http://dx.doi.org/10.1093/ejo/cjt023. PMid:23598610.
    » http://dx.doi.org/10.1093/ejo/cjt023
  • 15
    Singh S, Mogra S, Shetty VS, Shetty S, Philip P. Three-dimensional finite element analysis of strength, stability, and stress distribution in orthodontic anchorage: a conical, self-drilling miniscrew implant system. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):327-36. http://dx.doi.org/10.1016/j.ajodo.2011.07.022. PMid:22381493.
    » http://dx.doi.org/10.1016/j.ajodo.2011.07.022
  • 16
    Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, et al. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop. 2012 Mar;141(3):279-88. http://dx.doi.org/10.1016/j.ajodo.2011.09.008. PMid:22381488.
    » http://dx.doi.org/10.1016/j.ajodo.2011.09.008
  • 17
    Luiz NE, Jacomini A Fo, Lima LAC, Ciuccio RL, Soares MAD, Coutinho LL. Optimization design of orthodontic screws aiming at increasing the mechanical strength. Innov. Implant. J., Biomater. Esthet. 2010 Aug;5(2): 30-34.
  • 18
    Marcé-Nogué J, Walter A, Gil L, Puigdollers A. Finite element comparison of 10 orthodontic microscrews with different cortical bone parameters. Int J Oral Maxillofac Implants. 2013 July-Aug;28(4):e177-89. http://dx.doi.org/10.11607/jomi.2447. PMid:23869375.
    » http://dx.doi.org/10.11607/jomi.2447
  • 19
    Motoyoshi M, Inaba M, Ueno S, Shimizu N. Mechanical anisotropy of orthodontic mini-implants. Int J Oral Maxillofac Surg. 2009 Sept;38(9):972-7. http://dx.doi.org/10.1016/j.ijom.2009.05.009. PMid:19559569.
    » http://dx.doi.org/10.1016/j.ijom.2009.05.009
  • 20
    Meijer HJ, Starmans FJ, Steen WH, Bosman F. A three-dimensional, finite-element analysis of bone around dental implants in an edentulous human mandible. Arch Oral Biol. 1993 June;38(6):491-6. http://dx.doi.org/10.1016/0003-9969(93)90185-O. PMid:8343071.
    » http://dx.doi.org/10.1016/0003-9969(93)90185-O
  • 21
    Menicucci G, Mossolov A, Mozzati M, Lorenzetti M, Preti G. Tooth-implant connection: some biomechanical aspects based on finite element analyses. Clin Oral Implants Res. 2002 June;13(3):334-41. http://dx.doi.org/10.1034/j.1600-0501.2002.130315.x. PMid:12010166.
    » http://dx.doi.org/10.1034/j.1600-0501.2002.130315.x
  • 22
    American Society for Metals. Metals Handbook. 10th ed. Metals Park: ASM; 1990. Vol. 2, Properties and selection: nonferrous alloys and special-purpose materials.
  • 23
    American Society for Metals. Metals Handbook. 9th ed. Metals Park: ASM; 1980. Vol. 3, Properties and selection: stainless steels, tool materials and special-purpose metals.
  • 24
    Holt JM, Mindlin H, Ho CY, editors. Structural alloys handbook. West Lafayette: CINDAS/Purdue University; 1996.
  • 25
    Lin TS, Tsai FD, Chen CY, Lin LW. Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis. Am J Orthod Dentofacial Orthop. 2013 Feb;143(2):182-9. http://dx.doi.org/10.1016/j.ajodo.2012.09.012. PMid:23374924.
    » http://dx.doi.org/10.1016/j.ajodo.2012.09.012
  • 26
    Chen Y, Kyung HM, Gao L, Yu WJ, Bae EJ, Kim SM. Mechanical properties of self-drilling orthodontic micro-implants with different diameters. Angle Orthod. 2010 Sept;80(5):821-7. http://dx.doi.org/10.2319/103009-607.1. PMid:20578851.
    » http://dx.doi.org/10.2319/103009-607.1
  • 27
    Suzuki EY, Suzuki B. Placement and removal torque values of orthodontic miniscrew implants. Am J Orthod Dentofacial Orthop. 2011 May;139(5):669-78. http://dx.doi.org/10.1016/j.ajodo.2010.11.017. PMid:21536211.
    » http://dx.doi.org/10.1016/j.ajodo.2010.11.017
  • 28
    Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51. http://dx.doi.org/10.1016/j.ajodo.2006.04.027. PMid:17448385.
    » http://dx.doi.org/10.1016/j.ajodo.2006.04.027
  • 29
    Büchter A, Wiechmann D, Koerdt S, Wiesmann HP, Piffko J, Meyer U. Load-related implant reaction of mini-implants used for orthodontic anchorage. Clin Oral Implants Res. 2005 Aug;16(4):473-9. http://dx.doi.org/10.1111/j.1600-0501.2005.01149.x. PMid:16117773.
    » http://dx.doi.org/10.1111/j.1600-0501.2005.01149.x
  • 30
    Lee J, Kim JY, Choi YJ, Kim KH, Chung CJ. Effects of placement angle and direction of orthopedic force application on the stability of orthodontic miniscrews. Angle Orthod. 2013 July;83(4):667-73. http://dx.doi.org/10.2319/090112-703.1. PMid:23241005.
    » http://dx.doi.org/10.2319/090112-703.1

Publication Dates

  • Publication in this collection
    06 Oct 2015
  • Date of issue
    Sep-Oct 2015

History

  • Received
    15 Dec 2014
  • Accepted
    25 Mar 2015
Universidade Estadual Paulista Júlio de Mesquita Filho Rua Humaitá, 1680 - Caixa Postal 331, 14801-903 Araraquara,São Paulo,SP, Tel.: (55 16) 3301-6376, Fax: (55 16) 3301-6433 - Araraquara - SP - Brazil
E-mail: adriana@foar.unesp.br