Correlation between ultrasound velocity and densitometry in fresh and demineralized cortical bone

Abstract

OBJECTIVE:

To compare ultrasound propagation velocity with densitometry in the diaphyseal compact cortical bone of whole sheep metatarsals.

METHODS:

The transverse ultrasound velocity and bone mineral density of 5-cm-long diaphyseal bone segments were first measured. The bone segments were then divided into four groups of 15 segments each and demineralized in an aqueous 0.5 N hydrochloric acid solution for 6, 12, 24 or 36 hours. All measurements were repeated after demineralization for each time duration and the values measured before and after demineralization were compared.

RESULTS:

Ultrasound velocity and bone mineral density decreased with demineralization time, and most differences in the pre- and post-demineralization values within each group and between groups were significant: A moderate correlation coefficient (r=0.75956) together with a moderate agreement was determined between both post-demineralization parameters, detected by the Bland-Altman method.

CONCLUSION:

We conclude that both ultrasound velocity and bone mineral density decrease as a result of demineralization, thus indicating that bone mineral content is of great importance for maintaining the acoustic parameters of cortical bone, as observed for cancellous bone. Ultrasound velocity can be used to evaluate both compact cortical bone quality and bone mineral density.

Bone; Demineralization; Ultrasound Velocity; Bone Mineral Density


INTRODUCTION

Bone tissue is a composite material that primarily consists of collagen fibers and an inorganic matrix. The mechanical properties of bone vary according to site and direction (anisotropy) and also depend on the relative proportion of each component at different sites. The mineral component of bone accounts for approximately 65% of its dry mass, most of which is hydroxyapatite [Ca10(PO4)6(OH)2]. As hydroxyapatite is a very prevalent component in bone, it plays an important role in ultrasound transmission through the bone, which has been well established in the literature.

Dual-energy x-ray absorptiometry (DEXA), commonly known as densitometry, is the tool most frequently used to assess bone mass. DEXA is based on a low radiation dose (1-6 µSv) and can rapidly obtain measurements (∼2 min), although the method is size-dependent and provides an areal density (g/cm2) rather than a volumetric density (g/cm3) (11. Adams JE. Dual-energy X-ray absorptiometry. In: Guglielmi G, editor. Osteoporosis and Bone Densitometry Measurements, Medical Radiology. Diagnostic Imaging. Berlin Heidelberg: Springer-Verlag; 2013, p 101-122, http://dx.doi.org/10.1007/978-3-642-27884-6.
http://dx.doi.org/10.1007/978-3-642-2788...
). Nevertheless, it is currently the gold standard for diagnosing osteoporosis and assessing fracture risk.

The quantitative ultrasonic properties (ultrasound absorption, velocity and attenuation) of any biological tissue are primarily determined by protein constituents at the macromolecular level, particularly collagen. The ultrasound velocity is significantly higher in collagen-rich tissues (22. Goss SA, Frizzell LA, Dunn F. Ultrasonic absorption and attenuation in mammalian tissues. Ultrasound Med Biol. 1979;5(2):181-6, http://dx.doi.org/10.1016/0301-5629(79)90086-3.
http://dx.doi.org/10.1016/0301-5629(79)9...
,33. Goss SA, Frizzell LA, Dunn F. Dependence of the ultrasonic properties of biological tissue on constituent proteins. J Acous Soc Am. 1980;67(3):1041-4, http://dx.doi.org/10.1121/1.384071.
http://dx.doi.org/10.1121/1.384071...
,44. Goss SA, Dunn F. Ultrasonic propagation properties of collagen. Phys Med Biol. 1980;25(5):827-37, http://dx.doi.org/10.1088/0031-9155/25/5/001.
http://dx.doi.org/10.1088/0031-9155/25/5...
), the ultrasonic properties of which are also affected by the direction ultrasound is applied (angular dependence) (55. Hoffmeister BK, Verdonk ED, Wickline SA, Miller JG. Effect of collagen on the anisotropy of quasi-longitudinal mode ultrasonic velocity in fibrous soft tissues: a comparison of fixed tendon and fixed myocardium. J Acoust Soc Am. 1994;96(4):1957-64, http://dx.doi.org/10.1121/1.410139.
http://dx.doi.org/10.1121/1.410139...
).

Ultrasound transmission and propagation vary according to the local bone composition and the direction of wave insertion, although quantitative ultrasound, or ultrasonometry, can still provide a very distinct measure of bone anisotropy, thus providing an indirect measure of bone quality and mechanical properties that is somewhat comparable to densitometry (66. Turner CH, Eich M. Ultrasonic velocity as a predictor of strength in bovine cancellous bone. Calcif Tissue Int. 1991;49(2):116-9, http://dx.doi.org/10.1007/BF02565132.
http://dx.doi.org/10.1007/BF02565132...
,77. Glüer CC, Wu CY, Genant HK. Broadband ultrasound attenuation signals depend on trabecular orientation: an in vitro study. Osteoporos Int. 1993;3(4):185-91, http://dx.doi.org/10.1007/BF01623674.
http://dx.doi.org/10.1007/BF01623674...
,88. Markel MD, Chao EY. Noninvasive monitoring techniques for quantitative description of callus mineral content and mechanical properties. Clin Orthop Relat Res. 1993;293:37-45.,99. Njeh CF, Kuo CW, Langton CM, Atrah HI, Boivin CM. Prediction of human femoral bone strength using ultrasound velocity and BMD: an in vitro study. Osteoporos Int. 1997;7(5):471-7, http://dx.doi.org/10.1007/s001980050035.
http://dx.doi.org/10.1007/s001980050035...
,1010. Sievänen H, Cheng S, Ollikainen S, Uusi-Rasi K. Ultrasound velocity and cortical bone characteristics in vivo. Osteoporos Int. 2001;12(5):399-405, http://dx.doi.org/10.1007/s001980170109.
http://dx.doi.org/10.1007/s001980170109...
). The correlation between ultrasound velocity (UV) and attenuation and volumetric density (Archimedes principle) was established for cancellous bone approximately 25 years ago, and UV was found to be the most sensitive parameter (1111. Tavakoli MB, Evans JA. Dependence of the velocity and attenuation of ultrasound in bone on the mineral content. Phys Med Biol. 1991;36(11):1529-37, http://dx.doi.org/10.1088/0031-9155/36/11/012.
http://dx.doi.org/10.1088/0031-9155/36/1...
). In a more recent and sophisticated study using ultrasonometry associated with Fourier analysis, UV, elasticity and density were compared in cancellous bone specimens, and good correlation was found between UV and density but not between UV and elasticity (1212. De Terlizzi F, Battista S, Cavani F, Canà V, Cadossi R. Influence of bone tissue density and elasticity on ultrasound propagation: an in vitro study. J Bone Miner Res. 2000;15(12):2458-66, http://dx.doi.org/10.1359/jbmr.2000.15.12.2458.
http://dx.doi.org/10.1359/jbmr.2000.15.1...
).

The ability of ultrasound to predict the mechanical properties of bone with varying levels of mineral content was confirmed by comparing bone elastic moduli (biomechanical essays) with acoustic impedance results, and a strong positive correlation was demonstrated between the two parameters. This indicated that the acoustic impedance was a strong predictive factor of the elastic modulus and can help predict the mechanical properties of bone tissue as a function of mineral content (1313. Kotha SP, DePaula CA, Mann AB, Guzelsu N. High frequency ultrasound prediction of mechanical properties of cortical bone with varying amount of mineral content. Ultrasound Med Biol. 2008;34(4):630-7, http://dx.doi.org/10.1016/j.ultrasmedbio.2007.09.020.
http://dx.doi.org/10.1016/j.ultrasmedbio...
).

The correlation between UV and bone mineral density (BMD) has also been demonstrated in clinical studies, which are more attractive to clinicians who frequently interact with patients. Some clinical studies have been conducted to compare UV with BMD as measured by DEXA at different sites of normal (1414. Foldes AJ, Rimon A, Keinan DD, Popovtzer MM. Quantitative ultrasound of the tibia: a novel approach for assessment of bone status. Bone. 1995;17(4):363-7, http://dx.doi.org/10.1016/S8756-3282(95)00244-8.
http://dx.doi.org/10.1016/S8756-3282(95)...
,1515. Ng DC, Sundram FX. Bone mineral density--correlation between quantitative ultrasound characteristics and dual energy X-ray absorptiometry. Ann Acad Med Singapore. 1998;27(4):524-6.) and pagetic bone (1616. Pande KC, Bernard J, McCloskey V, de Takats D, Kanis JA. Ultrasound velocity and dual-energy X-ray absorptiometry in normal and pagetic bone. Bone. 2000;26(5):525-8, http://dx.doi.org/10.1016/S8756-3282(00)00247-7.
http://dx.doi.org/10.1016/S8756-3282(00)...
). These studies have shown a positive correlation between UV and BMD, therefore suggesting that UV is an accurate method for assessing osteoporosis.

The mechanical properties, BMD and UV, and broadband ultrasound attenuation (BUA) were also compared in human and bovine bone, leading to the conclusion that BMD and UV were the most accurate parameters for predicting the mechanical properties of high density trabecular bone (1717. Töyräs J, Nieminen MT, Kröger H, Jurvelin JS. Bone mineral density, ultrasound velocity, and broadband attenuation predict mechanical properties of trabecular bone differently. Bone. 2002;31(4):503-7, http://dx.doi.org/10.1016/S8756-3282(02)00843-8.
http://dx.doi.org/10.1016/S8756-3282(02)...
,1818. Funke M, Kopka L, Vosshenrich R, Fischer U, Ueberschaer A, Oestmann JW, et al. Broadband ultrasound attenuation in the diagnosis of osteoporosis: correlation with osteodensitometry and fracture. Radiology. 1995;194(1):77-81, http://dx.doi.org/10.1148/radiology.194.1.7997586.
http://dx.doi.org/10.1148/radiology.194....
).

According to the above-mentioned references, most research in the field of bone ultrasonometry has considered trabecular bone, whereas few studies have considered compact cortical diaphyseal bone, and hardly any studies have evaluated whole cylindrical diaphyseal bone. We hypothesized that the ultrasonic behavior of cortical bone would be somewhat similar to that of trabecular bone and that a reduction in BMD would therefore imply a reduction in the ultrasonic properties of bone, even whole diaphyseal bone. To test this hypothesis, we measured UV and BMD before and after demineralization of the whole metatarsal bone of sheep. Both UV and BMD were measured using the appropriate specific equipment.

MATERIAL AND METHODS

This experiment was approved by the Ethics Committee on the Experimental Use of Animals of the authors’ institution (Protocol #035/2010). Sixty fresh sheep metatarsals (Santa Ignez breed, male, 15 months of age on average, for human consumption) were obtained from a local authorized slaughterhouse.

All bones were completely freed from any soft tissue, including the periosteum, by careful dissection and stripping, and divided into three segments of equal length (∼5 cm). Only the middle segment was used for the analyses because we were particularly interested in cortical bone (Figure 1). The bone segments were initially identified, weighed, radiographed and divided into four groups (G6, G12, G24 and G36) of 15 specimens each according to demineralization time. They were then submitted to pre-demineralization UV and BMD analyses, and all parts were thereafter decalcified. Conventional radiographs were recorded and UV and BMD measurements were repeated after demineralization and pre- and post-demineralization images and figures were compared.

Figure 1
Photograph of a 6-cm-long diaphyseal bone segment that was obtained.

Bone demineralization procedure

The bone segments were individually measured for volume via water column dislocation (1919. Shah KM, Goh JC, Karunanithy R, Low SL, Das De S, Bose K. Effect of decalcification on bone mineral content and bending strength of feline femur. Calcif Tissue Int. 1995;56(1):78-82, http://dx.doi.org/10.1007/BF00298748.
http://dx.doi.org/10.1007/BF00298748...
) and then placed in identified glass flasks containing 20 times the segment’s volume of an aqueous 0.5 N hydrochloric acid solution. The segments were left to stand for 6, 12, 24 or 36 hours according to the group. After the demineralization period, the bone segments were thoroughly washed with deionized distilled water and again submitted to UV and BMD measurements and radiography (Figure 2).

Figure 2
Radiographic appearance of the bone segments after the demineralization process: before demineralization (a) and after 6 (b), 12 (c), 24 (d) and 36 (e) hours. Differences are not apparent, except for some thinning of the cortex and rarefaction of the medullary cancellous bone.

Ultrasonometry

Transverse underwater UV was measured with the bone segments immersed in an acoustic tank that was adapted with two unfocused ultrasound transducers (2 mm-thick PZT-5 disc, 12 mm in diameter, 1 MHz frequency), one on each of the tank in the geometric center of the walls and separated by a 25 mm distance but precisely aligned with each other along the central longitudinal axis (Figure 3). To record measurements, the two transducers, one for US wave emission and the other for reception, were connected to an ultrasound generator-receiver amplifier source (Biotecnosis do Brasil Ltda., Model US01, Ribeirão Preto, São Paulo, Brazil, http://www.biotecnosis.com) that could generate high power (up to 300 V) narrow (1 µs) well-defined ultrasonic pulses. The source was linked to a computer loaded with software to automatically calculate the UV based on the time of flight of the first arrived signal (FAS), as observed on a digital storage oscilloscope (Agilent Technologies, Inc., model DSO3062A, Shangai, China), together with the emitted ultrasound signal.

Figure 3
The diaphyseal segment inside the acoustic tank (arrow) positioned between the transmitting (a) and receiving (b) ultrasound transducers.

The FAS was defined as the first 5% positive deflection above baseline preceding the signal corresponding to the transmission mean (water) (Figure 4). The equipment was calibrated after every five measurements with a 20-mm-thick Teflon disc (constant 1274 m/s UV, at 35°C, 0.3% variation) to check for regularity and accuracy. The bone segments were positioned lengthwise in the acoustic tank supported by a Teflon stand on each end, transverse to the transducers and with the midpoint of the bone segment precisely aligned with the central axis of both transducers but closer to the emitting transducer than to the receiving transducer. The ultrasound emission frequency was 1 MHz, the pulse duration was ±1 μs, the rise time was ±0.1 ns, and the repetition time was 1 s. UV was calculated according to the formula:

(1)
Figure 4
The emitted US wave (short arrow) and the first arrived US signal (FAS, long arrow), as observed using an oscilloscope.

where Vs=velocity through the specimen; Vr=velocity through the reference propagation medium (water); τr=time for reference propagation medium alone (water); τs=time for reference propagation medium and specimen; and d=distance (coronal diameter of each individual bone segment, corresponding to the distance travelled by the ultrasound waves through the bone).

Five consecutive measurements were made at 5-minute intervals, the lowest and highest values were discarded, and the mean of the three remaining values was used for statistical comparisons.

Densitometry

BMD was measured by DEXA in each individual segment using a clinical densitometer (Hologic Inc., model QDR4500 Elite, Bedford, MA, USA, http://www.hologic.com). Each tibial segment was positioned lengthwise directly on the table without any other material and precisely centered below the beam with the aid of the laser collimator. Three consecutive measurements were made, and the mean value was calculated and used for statistical comparisons. The results are reported as g/cm2.

Although a regularly checked densitometer currently used for clinical purposes was employed, the precision error was calculated for the untouched bone segments (before demineralization) according to the procedure recommended by the International Society for Clinical Densitometry (ISCD) and adopted by the densitometry facility of our institution. The calculated precision error was 1.9% with a least significant change (LSC) of 5.2%.

Statistical analysis

Two-way analysis of variance (ANOVA) was used to compare the individual results of UV and BMD before and after demineralization (2020. Montgomery DC. Design and Analysis of Experiments, 5th edition. New York, John Wiley & Sons, 2000.). Data referring to UV and BMD before and after demineralization were compared using a mixed effects linear model, adjusted by the Proc Mixed procedure of SAS v.9 software at a 5% significance level (p≤0.05). A raw comparison between average values was also undertaken for each situation and the Pearson correlation coefficient was proposed to quantify the correlation between UV and BMD. Bland-Altman graphs were used to evaluate the agreement between the two previous methods (2121. Altman DG, Bland JM. Measurement in medicine: the analysis of method comparison studies. Statistician. 1983; 32:307-17, http://dx.doi.org/10.2307/2987937.
http://dx.doi.org/10.2307/2987937...
).

RESULTS

The radiographic appearance of the specimens did not substantially change with the time of demineralization, but some thinning of the cortical layer, dark spots and areas of more pronounced demineralization were visible (Figure 2).

Ultrasound velocity

The pre-demineralization UV, as measured separately for each group (G6, G12, G24 and G36), ranged from an average of 2992 m/s to 3260 m/s, and the differences among groups were not significant. Post-demineralization UV progressively decreased with time from G6 (2840 m/s) to G36 (2327 m/s) to approximately 92.43% of the pre-demineralization value in G6, 87.39% in G12, 78.96% in G24 and 78% in G36. Significant differences (p<0.0001) were found between pre- and post-demineralization intra-group values for all groups (Table 1, Figure 5). The differences among the post-demineralization values showed variable significance for comparisons made between pairs of groups (G6xG12: p=0.0027; G6xG24: p<0.0001; G6xG36: p<0.0001; G12xG24: p=0.0037; G12xG36: p<0.0001; G24xG36: p<0.0001).

Table 1
UV (m/s) data (mean±SD, range, median) before and after demineralization according to demineralization time (hours).

Figure 5
Box-plot graph of UV before (light grey boxes) and after (dark grey boxes) demineralization, according to demineralization time.

Bone mineral density

The mean pre-demineralization BMD ranged from 0.65 to 0.69 g/cm2 with no significant difference among groups. The post-demineralization BMD progressively decreased from 0.54 g/cm2 in G6 to 0.44 g/cm2 in G36, corresponding to 83.32%, 82.96%, 67.37% and 65.37% of the intra-group pre-demineralization values for G6, G12, G24 and G36, respectively, with significant differences (p<0.0001) between the pre- and post-demineralization values in all groups (Table 2, Figure 6). A significant difference was also observed for most comparisons of the post-demineralization values between groups (G6xG12: p=0.0245; G6xG24: p<0.0001; G6xG36: p<0.0001; G12xG24: p<0.0001; G12xG36: p<0.0001), but not for and G24xG36 (p=0.1381).

Table 2
BMD (g/cm2) data (mean±SD, range, median) before and after demineralization according to demineralization time (hours).

Figure 6
Box-plot graph of the bone mineral density (BMD) before (light grey boxes) and after (dark grey boxes) demineralization, according to demineralization time.

The correlation between UV and BMD progressively increased with the post-demineralization period (G6: r=0.38308, p=0.0002; G12: r=0.64451, p<0.0001; G24: r=0.88733, p<0.0001; G36: r=0.89270, p<0.0001), with a relatively high overall correlation of r=0.75956 (p<0.0001) (Figure 7).

Figure 7
Graph of the individual UV x BMD dispersion before and after demineralization. The concentration of the dots around the average line is evident after demineralization.

UV and BMD figures were then transformed into percent values of baseline, taking the pre-demineralization values as the baseline. The differences and averages of the obtained values were then applied to a Bland-Altman graph to evaluate the agreement between the two methods. The Bland-Altman graph allows for visualization of how much each difference deviates from zero (bias), the dispersion of the differences around the average (error) and the tendency of the distribution. A perfect agreement occurs when the mean (bias) is zero and the dispersion between the upper (ULA) and lower (LLA) limits of agreement is very small or as close as possible to the mean line; data above ULA and below LLA are not in agreement. In the present case, a moderate agreement was observed because the mean was well above the zero line [9.44] and the average dots were dispersed around the mean line, although concentrated between the ULA and LLA lines. A single case was above ULA, while four cases were below LLA (Figure 8). The dot distribution also showed that lower percent differences between UV and BMD corresponded to averages closer to 100 and that this tendency of the comparison likely indicates that the agreement between the two methods is easier to detect in normal untouched bones.

Figure 8
Bland-Altman graph showing the distribution of the individual average points dispersed around the mean line but concentrated between the ULA and LLA lines; a single case was above the ULA, while four were below the LLA.

DISCUSSION

Composite materials consist of two or more materials that highly differ in character but the combination of which usually results in a third material with superior physical and mechanical properties. This is the case for bone, in which the combination of a protein matrix (collagen) and a mineral component (mostly hydroxyapatite) results in a far stronger tissue that is also flexible. However, due to the nature of its synthesis, bone is likely to show more variation in measurable physical and mechanical properties than typical engineering composites. Such variability is caused by several factors, such as age, gender, anatomical location, general health status, local disease and so on, all of which may affect the composition and structure of bone, particularly its collagen and mineral contents. Additionally, bone is an anisotropic material (with physical and mechanical properties that vary according to axle), and therefore, the measured values of a given property commonly fall within a relatively wide range; the average of the measured values is simply used as a reference value.

On clinical grounds, bone quality can be evaluated by a variety of methods, including some well-established methods, such as X-ray based methods, and some experimental methods that are still under investigation. DEXA includes exposing the patient to two low-dose x-ray beams with different energy level peaks. One peak is absorbed by the soft tissue, the energy of one peak which is then subtracted from the total, thus leaving the amount corresponding to the bone. DEXA is the current standard method for measuring BMD, which is altered in several diseases, including osteopenia and osteoporosis. Although it is a noninvasive and fast method, it involves the use of ionizing radiation, which is coupled with potential cumulative hazardous effects for patients who need repetitive examinations or also undergo other forms of examination that involve radiation.

A second alternative is ultrasonometry, or quantitative ultrasound. Although bone is an organic composite material, it has physical and mechanical properties comparable to those of man-made engineering materials (metals, plastic and so on); therefore, bone is often examined in a similar manner. Accordingly, since the 1980s, quantitative ultrasound has been investigated as a means to measure bone quality and anisotropy in an attempt to quantify osteoporosis and to estimate the risk of fracture, with the use of specific equipment (1010. Sievänen H, Cheng S, Ollikainen S, Uusi-Rasi K. Ultrasound velocity and cortical bone characteristics in vivo. Osteoporos Int. 2001;12(5):399-405, http://dx.doi.org/10.1007/s001980170109.
http://dx.doi.org/10.1007/s001980170109...
,2222. Rosenthall L. Selective supplementation of calcaneal ultrasound densitometry with dual-energy x-ray absorptiometry of the spine and femur for population screening. Can Assoc Radiol J. 1997;48(1):38-41.,2323. Wüster C, Heilmann P, Pereira-Lima J, Schlegel J, Anstätt K, Soballa T. Quantitative ultrasonometry (QUS) for the evaluation of osteoporosis risk: reference data for various measurement sites, limitations and application possibilities. Exp Clin Endocrinol Diabetes. 1998;106(4):277-88, http://dx.doi.org/10.1055/s-0029-1211986.
http://dx.doi.org/10.1055/s-0029-1211986...
,2424. Oliveira PP, Marinheiro LP, Wender MC, Mendes JB, Roisenberg F. Quantitative ultrasound and risk of fractures in elderly women. Rev Assoc Med Bras. 2011;57(6):651-6, http://dx.doi.org/10.1590/S0104-42302011000600011.
http://dx.doi.org/10.1590/S0104-42302011...
). The use of the so-called transmission ultrasonometry for this purpose is based on the observation that physical properties interfere with the propagation of ultrasound waves while passing through a bone segment. However, ultrasonometry does not provide a measure of areal (g/cm2) bone density, such as DEXA does, but rather the UV and attenuation that can be measured while the ultrasound waves pass through a bone segment. UV and attenuation vary according to structure, density, elasticity and other physical and mechanical properties of bone (99. Njeh CF, Kuo CW, Langton CM, Atrah HI, Boivin CM. Prediction of human femoral bone strength using ultrasound velocity and BMD: an in vitro study. Osteoporos Int. 1997;7(5):471-7, http://dx.doi.org/10.1007/s001980050035.
http://dx.doi.org/10.1007/s001980050035...
). Therefore, theoretically, both could be used to provide an indirect measure of those properties. It has already been demonstrated, for instance, that ultrasonometry can help diagnose the healing status of fractures and can be used as an ancillary method for that purpose, particularly when the precise situation cannot be assured by conventional methods (radiographs, CT) beyond any doubt (2525. Saulgozis J, Pontaga I, Lowet G, Van der Perre G. The effect of fracture and fracture fixation on ultrasonic velocity and attenuation. Physiol Meas. 1996;17(3):201-11, http://dx.doi.org/10.1088/0967-3334/17/3/006.
http://dx.doi.org/10.1088/0967-3334/17/3...
,2626. Njeh CF, Kearton JR, Hans D, Boivin CM. The use of quantitative ultrasound to monitor fracture healing: a feasibility study using phantoms. Med Eng Phys. 1998;20(10):781-6, http://dx.doi.org/10.1016/S1350-4533(99)00014-4.
http://dx.doi.org/10.1016/S1350-4533(99)...
,2727. Barbieri G, Barbieri CH, de Matos PS, Pelá CA, Mazzer N. Ultrasonometric evaluation of bone healing: Experimental study using a model of diaphyseal transverse osteotomy of sheep tibiae. Ultrasound Med Biol. 2006;32(6):875-82, http://dx.doi.org/10.1016/j.ultrasmedbio.2006.02.1413.
http://dx.doi.org/10.1016/j.ultrasmedbio...
,2828. Barbieri G, Barbieri CH, Mazzer N, Pelá CA. Ultrasound propagation velocity and broadband attenuation can help evaluate the healing process of an experimental fracture. J Orthop Res. 2011;29(3):444-51, http://dx.doi.org/10.1002/jor.21258.
http://dx.doi.org/10.1002/jor.21258...
,2929. Barbieri G, Mazzer N, Ribeiro EA, Nogueira-Barbosa MH, Barbieri CH. A comparative analysis between ultrasonometry and computer-aided tomography to evaluate bone healing. J Orthop Res. 2012;30(7):1076-82, http://dx.doi.org/10.1002/jor.22039.
http://dx.doi.org/10.1002/jor.22039...
,3030. Barbieri G, Barbieri CH. A comparative in vivo ultrasonometric evaluation of normal and delayed fracture healing in sheep tibiae. Clinics. 2014;69(9):634-40, http://dx.doi.org/10.6061/clinics/2014(09)10.
http://dx.doi.org/10.6061/clinics/2014(0...
,3131. Protopappas VC, Vavva MG, Fotiadis DI, Malizos KN. Ultrasonic monitoring of bone fracture healing. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55(6):1243-55, http://dx.doi.org/10.1109/TUFFC.2008.787.
http://dx.doi.org/10.1109/TUFFC.2008.787...
).

Although regularly used by many professionals to evaluate bone mass and density, ultrasonometry seems to be still an experimental method, despite the long existence of many commercially available devices (1515. Ng DC, Sundram FX. Bone mineral density--correlation between quantitative ultrasound characteristics and dual energy X-ray absorptiometry. Ann Acad Med Singapore. 1998;27(4):524-6.). We believe that new clinical and/or experimental studies would be helpful in clarifying doubts regarding the ultrasonometry method and technique, particularly for cortical bone, which is less prone to both collagen and mineral content variability due to its compact structure. The aim of the present study was to determine the correlation between UV, as measured by specific equipment, and BMD, as measured by DEXA, in an experimental bench-top setup using whole cortical diaphyseal bone segments before and after demineralization. The ultrasound equipment used had already been exhaustively tested in many investigations of cortical bone, and the demineralization method was also routinely used in our laboratory.

Accordingly, we used the underwater transverse ultrasonometry modality exactly as used in our previous investigations. In this method, both emitting and receiving transducers are focused on the region of interest of the analyzed bone segment and the ultrasonic waves cross the entire bone from side to side, thus providing a more complete evaluation of the entire bone thickness and likely a more accurate measure of the UV. Actually, in the longitudinal modality, the ultrasound waves tend to run superficially, thus only evaluating the subperiosteal layer of the entry cortex depending on the US frequency/wavelength. It is well known that, for a normal (90o) ultrasound incidence, the ultrasound waves travel superficially if the wavelength is smaller than the thickness of the cortex, thus providing no information pertaining to the deep parts of the specimen. A more complete evaluation can be performed when the wavelength is greater than the thickness of the cortex, a situation in which the ultrasound waves travel through the entire cortex thickness (3232. Prevrhal S, Fuerst T, Fan B, Njeh C, Hans D, Uffmann M, Srivastav S, Genant HK. Quantitative ultrasound of the tibia depends on both cortical density and thickness. Osteoporos Int. 2001;12(1):28-34, http://dx.doi.org/10.1007/s001980170154.
http://dx.doi.org/10.1007/s001980170154...
). The 1 MHz frequency emission used here produced ultrasound waves of 1.5 mm wavelength, which is approximately the thickness of the cortex of the analyzed bones, therefore indicating that the emitted ultrasound waves were capable of travelling through the entire thickness.

However, the cylindrical shape of the bone shaft imposes a ring-fashion pattern to the wave propagation by which the waves running along both the anterior and posterior cortices exit on the opposite side of the emission together with a few waves that manage to pass directly through the bone marrow at a lower UV (3333. Moilanen P. Ultrasonic guided waves in bone. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55(6):1277-86, http://dx.doi.org/10.1109/TUFFC.2008.790.
http://dx.doi.org/10.1109/TUFFC.2008.790...
).

The demineralization intervals (6, 12, 24 and 36) selected in our experimental design were short because we intended to check the UV capability to detect small differences between intervals. The degree of demineralization was verified by DEXA examination at all time points (6, 12, 24 and 36 hours) of the investigation to ensure that the demineralization method was as effective as we believed and that the results of the UV and BMD measurements obtained could be compared among groups without restriction. Accordingly, the mean post-demineralization UV and BMD values progressively and significantly decreased in a linear manner with demineralization time. UV reached approximately 74% of the average pre-demineralization value and 82% of the 6-hour post-demineralization (G6) value at the 36-hour demineralization (G36) time point; BMD followed that behavior, reaching 65% and 81.5%, respectively, for the same comparisons. The intra- and inter-group differences were significant for most comparisons for both separate UV and BMD data and also between UV and BMD, indicating a moderately positive (r=0.75956) correlation coefficient and a favorable agreement between the two methods according to the Bland-Altman graph. These findings indicate that the methods are highly similar and that either method can be used to assess the loss of mineral content in cortical bone.

In conclusion, the above results clearly confirm the dependence of ultrasound wave propagation on BMD at the level of demineralization obtained during the period of study. The UV values adequately correlated and agreed with the BMD values; therefore, UV can be measured to evaluate cortical bone quality in bones affected by the loss of mineral content. However, we suggest that in clinical practice, the UV method should be limited to superficial bones such as the tibia, ulna and clavicle, as currently available equipment does not permit the examination of deeper bones surrounded by thick muscle layers.

ACKNOWLEDGMENTS

The authors thank Fundação de Amparo è Pesquisa do Estado de São Paulo (FAPESP*) for financial support (Grant 09/52797-4).

REFERENCES

  • 1
    Adams JE. Dual-energy X-ray absorptiometry. In: Guglielmi G, editor. Osteoporosis and Bone Densitometry Measurements, Medical Radiology. Diagnostic Imaging. Berlin Heidelberg: Springer-Verlag; 2013, p 101-122, http://dx.doi.org/10.1007/978-3-642-27884-6
    » http://dx.doi.org/10.1007/978-3-642-27884-6
  • 2
    Goss SA, Frizzell LA, Dunn F. Ultrasonic absorption and attenuation in mammalian tissues. Ultrasound Med Biol. 1979;5(2):181-6, http://dx.doi.org/10.1016/0301-5629(79)90086-3
    » http://dx.doi.org/10.1016/0301-5629(79)90086-3
  • 3
    Goss SA, Frizzell LA, Dunn F. Dependence of the ultrasonic properties of biological tissue on constituent proteins. J Acous Soc Am. 1980;67(3):1041-4, http://dx.doi.org/10.1121/1.384071
    » http://dx.doi.org/10.1121/1.384071
  • 4
    Goss SA, Dunn F. Ultrasonic propagation properties of collagen. Phys Med Biol. 1980;25(5):827-37, http://dx.doi.org/10.1088/0031-9155/25/5/001
    » http://dx.doi.org/10.1088/0031-9155/25/5/001
  • 5
    Hoffmeister BK, Verdonk ED, Wickline SA, Miller JG. Effect of collagen on the anisotropy of quasi-longitudinal mode ultrasonic velocity in fibrous soft tissues: a comparison of fixed tendon and fixed myocardium. J Acoust Soc Am. 1994;96(4):1957-64, http://dx.doi.org/10.1121/1.410139
    » http://dx.doi.org/10.1121/1.410139
  • 6
    Turner CH, Eich M. Ultrasonic velocity as a predictor of strength in bovine cancellous bone. Calcif Tissue Int. 1991;49(2):116-9, http://dx.doi.org/10.1007/BF02565132
    » http://dx.doi.org/10.1007/BF02565132
  • 7
    Glüer CC, Wu CY, Genant HK. Broadband ultrasound attenuation signals depend on trabecular orientation: an in vitro study. Osteoporos Int. 1993;3(4):185-91, http://dx.doi.org/10.1007/BF01623674
    » http://dx.doi.org/10.1007/BF01623674
  • 8
    Markel MD, Chao EY. Noninvasive monitoring techniques for quantitative description of callus mineral content and mechanical properties. Clin Orthop Relat Res. 1993;293:37-45.
  • 9
    Njeh CF, Kuo CW, Langton CM, Atrah HI, Boivin CM. Prediction of human femoral bone strength using ultrasound velocity and BMD: an in vitro study. Osteoporos Int. 1997;7(5):471-7, http://dx.doi.org/10.1007/s001980050035
    » http://dx.doi.org/10.1007/s001980050035
  • 10
    Sievänen H, Cheng S, Ollikainen S, Uusi-Rasi K. Ultrasound velocity and cortical bone characteristics in vivo. Osteoporos Int. 2001;12(5):399-405, http://dx.doi.org/10.1007/s001980170109
    » http://dx.doi.org/10.1007/s001980170109
  • 11
    Tavakoli MB, Evans JA. Dependence of the velocity and attenuation of ultrasound in bone on the mineral content. Phys Med Biol. 1991;36(11):1529-37, http://dx.doi.org/10.1088/0031-9155/36/11/012
    » http://dx.doi.org/10.1088/0031-9155/36/11/012
  • 12
    De Terlizzi F, Battista S, Cavani F, Canà V, Cadossi R. Influence of bone tissue density and elasticity on ultrasound propagation: an in vitro study. J Bone Miner Res. 2000;15(12):2458-66, http://dx.doi.org/10.1359/jbmr.2000.15.12.2458
    » http://dx.doi.org/10.1359/jbmr.2000.15.12.2458
  • 13
    Kotha SP, DePaula CA, Mann AB, Guzelsu N. High frequency ultrasound prediction of mechanical properties of cortical bone with varying amount of mineral content. Ultrasound Med Biol. 2008;34(4):630-7, http://dx.doi.org/10.1016/j.ultrasmedbio.2007.09.020
    » http://dx.doi.org/10.1016/j.ultrasmedbio.2007.09.020
  • 14
    Foldes AJ, Rimon A, Keinan DD, Popovtzer MM. Quantitative ultrasound of the tibia: a novel approach for assessment of bone status. Bone. 1995;17(4):363-7, http://dx.doi.org/10.1016/S8756-3282(95)00244-8
    » http://dx.doi.org/10.1016/S8756-3282(95)00244-8
  • 15
    Ng DC, Sundram FX. Bone mineral density--correlation between quantitative ultrasound characteristics and dual energy X-ray absorptiometry. Ann Acad Med Singapore. 1998;27(4):524-6.
  • 16
    Pande KC, Bernard J, McCloskey V, de Takats D, Kanis JA. Ultrasound velocity and dual-energy X-ray absorptiometry in normal and pagetic bone. Bone. 2000;26(5):525-8, http://dx.doi.org/10.1016/S8756-3282(00)00247-7
    » http://dx.doi.org/10.1016/S8756-3282(00)00247-7
  • 17
    Töyräs J, Nieminen MT, Kröger H, Jurvelin JS. Bone mineral density, ultrasound velocity, and broadband attenuation predict mechanical properties of trabecular bone differently. Bone. 2002;31(4):503-7, http://dx.doi.org/10.1016/S8756-3282(02)00843-8
    » http://dx.doi.org/10.1016/S8756-3282(02)00843-8
  • 18
    Funke M, Kopka L, Vosshenrich R, Fischer U, Ueberschaer A, Oestmann JW, et al. Broadband ultrasound attenuation in the diagnosis of osteoporosis: correlation with osteodensitometry and fracture. Radiology. 1995;194(1):77-81, http://dx.doi.org/10.1148/radiology.194.1.7997586
    » http://dx.doi.org/10.1148/radiology.194.1.7997586
  • 19
    Shah KM, Goh JC, Karunanithy R, Low SL, Das De S, Bose K. Effect of decalcification on bone mineral content and bending strength of feline femur. Calcif Tissue Int. 1995;56(1):78-82, http://dx.doi.org/10.1007/BF00298748
    » http://dx.doi.org/10.1007/BF00298748
  • 20
    Montgomery DC. Design and Analysis of Experiments, 5th edition. New York, John Wiley & Sons, 2000.
  • 21
    Altman DG, Bland JM. Measurement in medicine: the analysis of method comparison studies. Statistician. 1983; 32:307-17, http://dx.doi.org/10.2307/2987937
    » http://dx.doi.org/10.2307/2987937
  • 22
    Rosenthall L. Selective supplementation of calcaneal ultrasound densitometry with dual-energy x-ray absorptiometry of the spine and femur for population screening. Can Assoc Radiol J. 1997;48(1):38-41.
  • 23
    Wüster C, Heilmann P, Pereira-Lima J, Schlegel J, Anstätt K, Soballa T. Quantitative ultrasonometry (QUS) for the evaluation of osteoporosis risk: reference data for various measurement sites, limitations and application possibilities. Exp Clin Endocrinol Diabetes. 1998;106(4):277-88, http://dx.doi.org/10.1055/s-0029-1211986
    » http://dx.doi.org/10.1055/s-0029-1211986
  • 24
    Oliveira PP, Marinheiro LP, Wender MC, Mendes JB, Roisenberg F. Quantitative ultrasound and risk of fractures in elderly women. Rev Assoc Med Bras. 2011;57(6):651-6, http://dx.doi.org/10.1590/S0104-42302011000600011
    » http://dx.doi.org/10.1590/S0104-42302011000600011
  • 25
    Saulgozis J, Pontaga I, Lowet G, Van der Perre G. The effect of fracture and fracture fixation on ultrasonic velocity and attenuation. Physiol Meas. 1996;17(3):201-11, http://dx.doi.org/10.1088/0967-3334/17/3/006
    » http://dx.doi.org/10.1088/0967-3334/17/3/006
  • 26
    Njeh CF, Kearton JR, Hans D, Boivin CM. The use of quantitative ultrasound to monitor fracture healing: a feasibility study using phantoms. Med Eng Phys. 1998;20(10):781-6, http://dx.doi.org/10.1016/S1350-4533(99)00014-4
    » http://dx.doi.org/10.1016/S1350-4533(99)00014-4
  • 27
    Barbieri G, Barbieri CH, de Matos PS, Pelá CA, Mazzer N. Ultrasonometric evaluation of bone healing: Experimental study using a model of diaphyseal transverse osteotomy of sheep tibiae. Ultrasound Med Biol. 2006;32(6):875-82, http://dx.doi.org/10.1016/j.ultrasmedbio.2006.02.1413
    » http://dx.doi.org/10.1016/j.ultrasmedbio.2006.02.1413
  • 28
    Barbieri G, Barbieri CH, Mazzer N, Pelá CA. Ultrasound propagation velocity and broadband attenuation can help evaluate the healing process of an experimental fracture. J Orthop Res. 2011;29(3):444-51, http://dx.doi.org/10.1002/jor.21258
    » http://dx.doi.org/10.1002/jor.21258
  • 29
    Barbieri G, Mazzer N, Ribeiro EA, Nogueira-Barbosa MH, Barbieri CH. A comparative analysis between ultrasonometry and computer-aided tomography to evaluate bone healing. J Orthop Res. 2012;30(7):1076-82, http://dx.doi.org/10.1002/jor.22039
    » http://dx.doi.org/10.1002/jor.22039
  • 30
    Barbieri G, Barbieri CH. A comparative in vivo ultrasonometric evaluation of normal and delayed fracture healing in sheep tibiae. Clinics. 2014;69(9):634-40, http://dx.doi.org/10.6061/clinics/2014(09)10
    » http://dx.doi.org/10.6061/clinics/2014(09)10
  • 31
    Protopappas VC, Vavva MG, Fotiadis DI, Malizos KN. Ultrasonic monitoring of bone fracture healing. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55(6):1243-55, http://dx.doi.org/10.1109/TUFFC.2008.787
    » http://dx.doi.org/10.1109/TUFFC.2008.787
  • 32
    Prevrhal S, Fuerst T, Fan B, Njeh C, Hans D, Uffmann M, Srivastav S, Genant HK. Quantitative ultrasound of the tibia depends on both cortical density and thickness. Osteoporos Int. 2001;12(1):28-34, http://dx.doi.org/10.1007/s001980170154
    » http://dx.doi.org/10.1007/s001980170154
  • 33
    Moilanen P. Ultrasonic guided waves in bone. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55(6):1277-86, http://dx.doi.org/10.1109/TUFFC.2008.790
    » http://dx.doi.org/10.1109/TUFFC.2008.790

Publication Dates

  • Publication in this collection
    Nov 2016

History

  • Received
    5 Apr 2016
  • Reviewed
    10 May 2016
  • Accepted
    5 Aug 2016
Creative Common - by 4.0
This is an Open Access article distributed under the terms of the Creative Commons License (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium or format, provided the original work is properly cited.
Faculdade de Medicina / USP Rua Dr Ovídio Pires de Campos, 225 - 6 and., 05403-010 São Paulo SP - Brazil, Tel.: (55 11) 2661-6235 - São Paulo - SP - Brazil
E-mail: clinics@hc.fm.usp.br