INTRODUCTION
Sagittal alignment of the spine is a concept based on the understanding and treatment of various spine pathologies.1-8 The correct alignment of the bony structures and their articulations is fundamental to the proper functioning of the musculoskeletal system.9 Furthermore, their complex interactions with the nervous system and subsequent muscle recruitment are essential to ergonomic balance and the maintenance of upright posture with minimal energetic expenditure, thus allowing the deliberate movement of the human body.10-12
Currently, the parameters used in the radiological evaluation of overall sagittal balance include the morphology and position of the pelvis, which act as regulators of sagittal alignment,9 among other spinal parameters. The spino-pelvic parameters are key components in the evaluation and treatment of sagittal deformities9 and also of degenerative pathologies of the spine, as correct sagittal alignment directly impacts the quality of life of these patients.13,14 The current standard for acquiring the measurements uses manual measuring of these parameters with a goniometer, a process that requires a dermatograph pencil for drawing lines on the radiographic film, as well as the goniometer itself, which the clinician many times does not have on hand.15
Recently, with the inclusion of technological tools in medical practice, these technologies can be used to assist physicians in clinical practice. The growing interaction between technology and clinical practice has enabled both time savings and the maintenance of communication between health professionals and their patients even outside the workplace.16 And the measurement of spino-pelvic parameters can be accomplished through digital means, using programs and applications available on different platforms, such as computers and smartphones, for example.
The current generation of smartphones has incorporated an accelerometer based on microelectromechanical systems (MEMS) that is sensitive to the acceleration and the inclination of the apparatus with precision of 1/10 of a degree, providing much greater accuracy in measuring spino-pelvic parameters as compared to measurements obtained from drawing lines on radiographic film.17,18 However, today the manual measurement of spino-pelvic parameters remains the standard method for acquiring these parameters.
Thus, this study aims to verify the equivalence and intra- and inter-evaluator reproducibility of an application called CobbMeter15 (CobbMeter, Altavi, Luxembourg) in the clinical practice of spine surgeons, comparing it to the current standard of measurement using a goniometer and dermatograph pencil, in order to confirm whether it is possible for a greater number of medical service providers to trust using a tool capable of providing reliable measurements, with a better cost-benefit ratio and greater convenience in clinical practice.
METHOD
For the development of this study, 20 panoramic lateral view x-ray films of the spine in neutral standing position, in which it was possible to identify the vertebral plateaus and femur heads (with at least 50% overlap), were analyzed. Patients with prior hip arthroplasty or who did not meet the inclusion criteria were excluded. All the patients included in the study agreed voluntarily to participate in the study and signed the Informed Consent Form (ICF), and the study was authorized by the Institutional Review Board (CAAE 45924615.8.0000.0086).
After collecting the 20 imaging exams, the author de-identified them and assigned a reference code to each. In the first phase, all the exams were presented to three different evaluators: two experienced physicians specializing in the spine and one researcher with extensive experience in the area of the spine, following the model proposed by Ritter et al.19 Measurements of the lumbar lordosis (LL), pelvic tilt (PT), and pelvic incidence (PI) were taken as described by Berthonnaud et al.20 and using the Portuguese nomenclature standardized by Pratali et al.21 The positioning of the smartphone was standardized for all the measurements and is illustrated in Figure 1. The LL was obtained by measuring the angle between the upper plateau of the L1 vertebra and the upper plateau of the S1 vertebra. The PT was acquired by measuring the angle obtained by the line formed between the midpoint of the segment that joins the centers of the femur heads and the midpoint of the upper plateau of vertebra S1 in relation to the vertical line that originates in the center, at the midpoint of the segment that joins the centers of the femur heads. The PI corresponds to the angle obtained from the line formed between the midpoint of the segment that joins the centers of the femur heads and the midpoint of the upper vertebral plate of the S1 vertebra in relation to the line perpendicular to the sacral plateau.

Figure 1 (A) Measurement of lumbar lordosis using the upper plateaus of vertebrae L1 and S1 and the CobbMeter application. (B) Measurement of pelvic tilt using the CobbMeter application. (C) Measurement of pelvic incidence using the CobbMeter application.
First, the data was collected using the CobbMeter application (CobbMeter CE, Regis Rigal, Version 2.5.2), running on a single Apple iPhone device (Apple Inc., Cupertino, USA) and then, after shuffling the order of the exams, the data was collected using a goniometer and a dermatograph pencil. In the second phase, a week following the first exposure, the same exams were presented at random to the evaluators who measured them again, using first the CobbMeter application and second a goniometer and dermatograph pencil, in the same form as in the first evaluation. Finally, in the third phase, 14 days following the first exposure, the same methodology was used to collect the measurements.
The data collected were plotted in a spreadsheet and complied for subsequent statistical analysis by the author. The compiled data were analyzed using the SPSS 20.0 program (IBM Corporation, Armonk, NY, USA). The intra-class correlation coefficients (ICC) were computed using the two-way effect model to test intra- and inter-evaluator consistency. The results of the ICC were interpreted according to the following scale: 0 to 0.24, absent or poor consistency; 0.25 to 0.49, low consistency; 0.50 to 0.69, moderate consistency; 0.70 to 0.89, good consistency; and 0.90 to 1.0, excellent consistency.22
RESULTS
For the reproducibility analysis, the intra-evaluator ICC, using the application, yielded a Kappa (K) for the evaluation of pelvic incidence (PI) of 0.975 (p<0.001). For pelvic tilt (PT), the K found was 0.981 (p<0.001) and for the analysis of lumbar lordosis (LL) the K 0.987 (p<0.001).
Also in the reproducibility evaluation, the inter-evaluator assessment, using the application, yielded a K of 0.917 (p<0.001) for the PI evaluation. For PT, the K found was 0.930 (p<0.001) and the K measured in the analysis of LL was 0.951 (p<0.001).
For the assessment of equivalence of the methods used in the evaluation of the spino-pelvic parameters, comparing the application to the standard goniometer and dermatograph pencil method, the K values found for the measurement of PI, PT, and LL were 0.873 (p<0.001), 0.939 (p<0.001), and 0.914 (p<0.001), respectively.
DISCUSSION
The technological advancement and popularization of smartphones, incorporating microelectromechanical systems (MEMS) technology, offered the user a new tool in the measurement of angles that, together with new software and applications, has had a great impact on several academic and professional areas, in our case highlighting greater efficiency and practicality in the evaluation and medical diagnosis of various spinal deformites.18 Nevertheless, there must be scientific proof that certifies the applicability and reproducibility of the technology, even though clinical decisions are being made for patients using the new technology.
The objective of this cross-sectional, non-interventional, observational study was to verify the intra- and inter-evaluator equivalence and reproducibility of the CobbMeter smartphone application in clinical practice, particularly in spine surgery, so that health professionals can rely on the use of the application to measure spino-pelvic parameters and, thus, as it becomes a more practiced methodology, user confidence will facilitate and disseminate the use of such radiographic measurements in clinical practice related to spine pathologies, directly impacting the improvement of patient care.
Thus, three evaluators conducted a set of observations of three different pelvic parameters in 20 patients using both the application and a goniometer and dermatograph pencil at three distinct instances, for a total of 1080 measurements. The study design was efficient and generated very high ICC values, showing high reliability between the methods and the evaluators. Therefore, the use of the cellular application insures measurements that are comparable and correlated to those acquired using the traditional measurement acquisition method proposed by Cobb.23
Other studies have already studied the use of smartphone applications to measure vertebral parameters, whether coronal24,25 (Cobb angles in scoliosis) or sagittal15 (kyphosis). Unlike earlier studies, this work measured spino-pelvic sagittal angles, which involve not only vertebral plateau lines, but also imaginary lines drawn between two anatomical points (such as for example the line that joins the center of the femur heads and the center of the S1 plateau). Although these lines were not pre-drawn on an x-ray film of the spine, we saw that no pencil was needed and by simply positioning the smartphone to connect the points, it is feasible to obtain the correct measurement of the desired angle. The benefit of using the cellular application to perform radiological measurements is that it uses a device that is always on hand, unlike a dermatograph pencil and goniometer, adding yet another cell phone function besides telephone calls, email management, agenda organization, among others. Add to this the fact that it is not necessary to mark up the x-ray film itself. It is worth noting that, while proposing a change in the method of data acquisition, the objective of obtaining these measurements remains the same, which is to provide a better understanding of the pathology evaluated, giving the physician anatomical and physiological bases to propose the best therapy and achieve successful treatment. Thus, the application should be used to measure other spino-pelvic parameters in clinical practice, but other studies should be conducted to evaluate the performance of the application in each of these clinical situations.
CONCLUSIONS
The reliability analysis in this study showed a high correlation between the data obtained with the application and those obtained using the traditional method and both high intra- and inter-evaluator correlations. These results show that the tested application is equivalent to the standard method and its use is reproducible. Thus, we concluded that this tool could be used in clinical practice without issues to evaluate the sagittal alignment of the spine.