Evaluation of zebu nellore cattle blood samples using the Cell-Dyn 3500 hematology analyzer

Avaliação de amostras de sangue de gado zebuíno nelore usando o analisador hematológico Cell-Dyn 3500

Abstracts

The Cell-dyn 3500 is a multiparameter flow cytometer, which may analyze samples from several species performing several simultaneous analyses. It is able to perform white blood cells, red blood cells and platelet counts, besides differential leukocyte counts, packed cell volume and hemoglobin determination. Cell-Dyn 3500 performs total leukocyte count both optically and by impedance. The equipment may choose one or other method, based on the reliability of the results. Erythrocyte and platelet counts are determined by impedance. Leukocyte differentiation is based on an optical principle, using separation in multiangular polarized light. The objective of this study was to compare the results of complete blood count of Zebu Nellore heifers from Cell-dyn 3500, with those obtained from a semi-automated cell counter (Celm CC 510) and the manual technique. Blood samples were collected from the jugular vein in 5 mL EDTA vacuum tubes from 58 Nellore heifers, at 24 months of age. Samples were processed in parallel in the three different techniques. Results were analyzed using paired t test, Pearson's correlation and the Bland-Altmann method. There was a strong correlation for all parameters analyzed by Cell-Dyn 3500, manual method and semi-automated cell counter, except for basophils and monocytes counts. These results confirm that this analyzer is reliable for blood samples analysis of zebu cattle.

CC-510; Cell-Dyn 3500; hematology; Nellore


O Cell-Dyn 3500 é um citômetro de fluxo multi-parâmetros que realiza várias análises simultâneas em diversas espécies de animais. Ele realiza a contagem total de leucócitos, de hemácias e de plaquetas, além do diferencial de leucócitos, o volume globular e a determinação de hemoglobina. O Cell-Dyn 3500 realiza a contagem total de leucócitos, tanto pelo método óptico quanto pela impedância. O equipamento pode escolher um ou outro método, baseado na confiabilidade dos resultados. A contagem de hemáceas e plaquetas é determinada pelo método de impedância. A diferenciação de leucócitos é baseada em um princípio óptico de análise, utilizando-se separação em luz polarizada multiangular. O objetivo deste trabalho foi comparar os resultados de hemogramas, realizados pelo Cell-Dyn 3500, com aqueles realizados por um contador celular semi-automático (Celm CC-510) e pela técnica manual. Amostras de sangue foram colhidas da veia jugular, em tubos de 5,0 ml à vácuo com EDTA, de 58 novilhas Nelore de 24 meses de idade. As amostras foram processadas paralelamente nas três técnicas. Os resultados foram analisados pelo teste t pareado, correlação de Pearson e pelo método de Bland-Altmann. Houve forte correlação entre os resultados hematológicos obtidos pelo Cell-Dyn 3500, pelo método manual e pelo contador celular semi-automático (Celm CC-510), exceto para a contagem de basófilos e monócitos. Esses resultados confirmam que o equipamento é confiável para a análise de amostras de sangue de gado zebuíno.

CC-150; Cell-Dyn 3500; hematologia; Nelore


  • 1
    Vives-Corrons J.L., Besson I., Jou, J.M., Gutierrez G. Evaluation of the Abbott Cell-Dyn 3500 hematology analyzer in university hospital. American Journal of Clinical Pathology. 1996;105:553-559.
  • 2
    Gaunt S.D., Prescott-Mathews J.S., King W.W., Scholl D.T. Clinical hematology practices at veterinary teaching hospitals and private diagnostic laboratories. Veterinary Clinical Pathology. 1995;24:64-67.
  • 3
    Kieffer J, Winkler G, Van Hove L, Walsh A., Thomann P., Wyss S., Eggenberger E. and Lutz H. Evaluation of the Cell-Dyn 3500 haematology instrument for the analysis of the mouse and rat blood. Comparative Clinical Pathology. 1999;9:92-106.
  • 4
    Hennesy M, Buckley T.C., Leadon D.P, Scott C.S. Automated analysis of blood samples from thoroughbred horses with the Abbot cell DYN 3500 (CD3500). Comparative Clinical Pathology. 1998;8:150-158.
  • 5
    Lilliehook I, Larsson B. Evaluation of Cell-Dyn 3500, a haematological instrument adapted for testing animal samples. Svensk-Veterinartidning. 1998;50:643-648.
  • 6
    Bleul U., Sobiraj A., Bostedt H. Evaluation of the Cell-Dyn 3500 Haematology Analyser for Bovine Blood. Comparative Clinical Pathology. 2002;11:201-210.
  • 7
    Lilliehook I, Wall H, Tauson R, Tvedten H. Differential leukocyte counts determined in chicken blood using the Cell-Dyn 3500. Veterinary Clinical Pathology. 2004;33:133-138.
  • 8
    Weiser, M G. Hematologic Techniques. Veterinary Clinics of North America (Small animal Practice). 1981;11:189-208.
  • 9
    Low E M, Hill HB, Searcy RL. Simple method for detection of abnormal plasma fibrinogen levels. Technical bulletin of the Registry of Medical Technologists. 1967;37:72-74.
  • 10
    Bland J.M, Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8476:307-310.
  • 11
    Chow E.Y, Leung K.K. Evaluating the Cell-Dyn 3500 haematology analyzer in an acute general hospital. Clinical and Laboratory Haematology. 1996;18:187-193.
  • 12
    Sanzari M, De Toni S, D'osualdo A, Rossetti M., Floriani F., Plebani M. Complete analytical and diagnostic performance of the Abbott Cell Dyn 3500. Panminerva Medicine. 1998;40:116-125.
  • 13
    Buttarello M., Gadotti M., Lorenz C., Toffalori E., Ceschini N., Valentini A, Rizzotti P. Evaluation of four automated hematology analysers. American Journal of Clinical Pathology. 1992; 97:345-352.
  • 14
    Al-Ismail SA, Bond K, Carter AB, Grant D., Machin SJ., Patterson K.G., Pearman K.J., Pollard Y.C., Popeck, M. X. Two-centre evaluation of the Abbott CELL-DYN 3500 blood counter. Clinical and Laboratory Haematology. 1992;17:11-21.
  • 15
    Knoll I.S, Rowell S.L. Clinical Hematology. In: Roskin RE, Meyer DJ. Clinic analysis control, reference values, and system selection. Veterinary Clinics of North American Small Animal Practice. 1996; 26:981-1002.
  • 16
    Weiser M.G. Modifications and evaluation of a multichannel blood cell counting system for blood analysis in veterinary hematology. Journal of the American Veterinary Medical Association, 1987; 190:411-423.

Publication Dates

  • Publication in this collection
    22 Jan 2015
  • Date of issue
    Dec 2014

History

  • Received
    26 Sept 2007
  • Accepted
    30 Sept 2014
Universidade Federal de Goiás Universidade Federal de Goiás, Escola de Veterinária e Zootecnia, Campus II, Caixa Postal 131, CEP: 74001-970, Tel.: (55 62) 3521-1568, Fax: (55 62) 3521-1566 - Goiânia - GO - Brazil
E-mail: revistacab@gmail.com