Furosemide in water matrix: HPLC-UV method development and degradation studies

Furosemida em matriz de água: desenvolvimento de um método HPLC-UV e estudos de degradação

Abstract

This study developed a method for furosemide quantification through high performance liquid chromatographic technique. Special attention was given to solute loss and storage stability due to furosemide’s low solubility and photosensitivity, respectively. The performance of Nylon and PVDF filters was tested in a 2 mg.L-1 furosemide solution. PVDF filters showed better recovery capacity and therefore are more suitable for furosemide filtration. Over eight days, three different storage conditions were studied to access furosemide degradation susceptibility: (i) exposure to light at room temperature, (ii) storage at room temperature without exposure to light, and (iii) storage at 4ºC without exposure to light. The study demonstrated that after 48 h under natural light exposure furosemide was completely degraded. Furosemide solution stored in the dark was stable. Storage temperature did not seem to affect furosemide concentration. The study shows that the selection of more suitable filter and storage conditions for furosemide determination is crucial to avoid underestimation errors.

Keywords:
filter retention; pharmaceutical; photodegradation

Resumo

O presente estudo desenvolve um método para a quantificação de furosemida através de cromatografia de alta resolução. Devido à baixa solubilidade do composto e à sua fotossensibilidade, neste estudo avaliou-se as possíveis perdas do composto ao longo do método de quantificação e a sua estabilidade durante o armazenamento ao longo de oito dias. Para tal, foram testados dois filtros (Nylon e PVDF) com o mesmo diâmetro e tamanho de poro para uma concentração de furosemida de 2 mg.L-1. A degradação da furosemida foi estudada para três condições diferentes de armazenamento: com exposição solar à temperatura ambiente, sem exposição solar à temperatura ambiente e sem exposição solar a 4°C. O estudo demonstrou que o composto furosemida atinge a degradação completa com exposição solar após 48h. A solução de furosemida permaneceu estável ao longo da experiência quando protegida da luz. A temperatura de armazenamento não parece afetar a concentração da solução de furosemida. O presente estudo mostrou que para análises da furosemida, a escolha do filtro adequado e as condições de armazenamento devem ser consideradas para evitar erros de subestimação.

Palavras-chave:
fármaco; foto degradação; retenção do filtro

1. INTRODUCTION

The presence of pharmaceutical active compounds (PhACs) in different water bodies is well documented in several studies (Böger et al., 2018BÖGER, B.; AMARAL, B.; SILVEIRA, P. L.; WAGNER, R.; PERALTA-ZAMOR, P. G.; GOMES, E. C. Determination of carbamazepine and diazepam by SPE-HPLC-DAD in Belém river water, Curitiba-PR/Brazil. Revista Ambiente & Água, v. 13, n. 2, 2018. https://dx.doi.org/10.4136/ambi-agua.2196
https://dx.doi.org/10.4136/ambi-agua.219...
; Kuster et al., 2008KUSTER, M.; LÓPEZ, J.; DE ALDA, M.; HERNANDO, M. D.; PETROVIC, M.; MARTIN-ALONSO, J.; BARCELÓ, D. Analysis and occurrence of pharmaceuticals, estrogens, progestogens and polar pesticides in sewage treatment plant effluents, river water and drinking water in the Llobregat river basin (Barcelona, Spain). Journal of Hydrology, v. 358, p. 112-123, 2008. https://doi.org/10.1016/j.jhydrol.2008.05.030
https://doi.org/10.1016/j.jhydrol.2008.0...
; Paíga and Delerue-Matos, 2016PAÍGA, P.; DELERUE-MATOS, C. Determination of pharmaceuticals in groundwater collected in five cemeteries’ areas (Portugal). Science of the Total Environment, v. 569-570, p. 16-22, 2016. https://dx.doi.org/10.1016/j.scitotenv.2016.06.090
https://dx.doi.org/10.1016/j.scitotenv.2...
; Pereira et al., 2017PEREIRA, A. M. P. T.; SILVA, L. J. G.; LARANJEIRO, C. S. M.; MEISEL, L. M.; LINO, C. M.; PENA, A. Human pharmaceuticals in Portuguese rivers: The impact of water scarcity in the environmental risk. Science of the Total Environment, v. 609, p. 1182-1191, 2017. https://dx.doi.org/10.1016/j.scitotenv.2017.07.200
https://dx.doi.org/10.1016/j.scitotenv.2...
). After ingestion, pharmaceuticals are only partially absorbed and the remainder is excreted as the parent compound or their metabolites via urine or feces. Conventional wastewater treatment processes are not completely efficient in removing PhACs (Ternes, 1998TERNES, T. A. Occurrence of drugs in german sewage treatment plants and rivers. Water Research, v. 32, n. 11, p. 3245-3260, 1998. https://dx.doi.org/10.1016/S0043-1354(98)00099-2
https://dx.doi.org/10.1016/S0043-1354(98...
; Martín et al., 2012MARTÍN, J.; CAMACH-MUÑOZ, D.; SANTOS, J. L.; APARICIO, I.; ALONSO, E. Occurrence of pharmaceutical compounds in wastewater and sludge from wastewater treatment plants: Removal and ecotoxicological impact of wastewater discharges and sludge disposal. Journal of Hazardous Materials, v. 239-240, p. 40-47, 2012. https://dx.doi.org/10.1016/j.jhazmat.2012.04.068
https://dx.doi.org/10.1016/j.jhazmat.201...
). Therefore, the development of methods to correctly quantify these compounds in specific matrices are required, and several options already exist. Environmental studies commonly determine the presence of PhACs in samples using methods for the determination of a broad variety of compounds (Cruz-Morató et al., 2014CRUZ-MORATÓ, C.; LUCAS, D.; LLORCA, M.; RODRIGUEZ-MOZAZ, S.; GORGA, M.; PETROVIC, M.; BARCELÓ, D.; VICENT, T.; SARRÀ, M.; MARCO-URREA, E. Hospital wastewater treatment by fungal bioreactor: Removal efficiency for pharmaceuticals and endocrine disruptor compounds. Science of the Total Environment, v. 493, p. 365-376, 2014. https://dx.doi.org/10.1016/j.scitotenv.2014.05.117
https://dx.doi.org/10.1016/j.scitotenv.2...
). Although this approach brings benefits such as reductions in time and cost, some inherent restrictions to the accuracy of the determination of each compound is associated, such as analyte loss due to membrane filter adsorption. HPLC-UV analyses requires samples to be filtered to avoid clogging the equipment and capillaries and to extend the column’s lifespan (Carlson and Thompson, 2000CARLSON, M.; THOMPSON, R. D. Analyte Loss Due to Membrane Filter Adsorption as Determined by High-Performance Liquid Chromatography. Journal of Chromatographic Science, v. 38, p. 78-83, 2000. https://dx.doi.org/10.1093/chromsci/38.2.77
https://dx.doi.org/10.1093/chromsci/38.2...
). Moreover, compound stability studies are also necessary to prevent associated quantification errors, particularly along analyses procedures. For some pollutants, the knowledge of all interferences in the quantification process are still not fully known and require further investigation.

In Europe, anti-hypertensive substances, recently included in the cardiovascular system pharmacotherapeutic group, have the highest consumption rates (OECD, 2016OECD. Demographic references: General demographics. 2016. Available online at Available online at https://stats.oecd.org/Index.aspx?DataSetCode=HEALTH_DEMR . Access: Feb. 14th 2018.
https://stats.oecd.org/Index.aspx?DataSe...
). Included in this group, furosemide (FUR) is a loop diuretic pharmaceutical prescribed to treat cases of hypertension and edema (Bosch et al., 2008BOSCH, M. E.; SANCHEZ, A. J. R.; ROJAS, F. S.; OJEDA, C. B. Recent developments in analytical determination of furosemide. Journal of Pharmaceutical and Biomedical Analysis, v. 48, p. 519-532, 2008. https://dx.doi.org/10.1016/j.jpba.2008.07.003
https://dx.doi.org/10.1016/j.jpba.2008.0...
). In Portugal in 2014, FUR occupied the 15th place in the top 100 active substances ranking with the highest number of packages sold in the National Health System (INFARMED, 2014INFARMED. Estatística do Medicamento. 2014. Available at: Available at: https://www.infarmed.pt/documents/15786/1229727/Estat%C3%ADstica+do+Medicamento+2014/988074f4-4f89-4a7c-9055-844cb88e93fd?version=1.2 Access: April 18th, 2018.
https://www.infarmed.pt/documents/15786/...
). According to Prandota and Witkowska (1976)PRANDOTA, J.; WITKOWSKA, M. Pharmacokinełics and mełabolism of furosemide in man. European Journal of Drug Metabolism and Pharmacokinetics, v. 4, p. 177-181, 1976. https://dx.doi.org/10.1007/BF03189275
https://dx.doi.org/10.1007/BF03189275...
, after human consumption up to 70% of furosemide is absorbed. However, 90% of the non-absorbed drug is excreted as the parent compound (Zuccato et al., 2005ZUCCATO, E.; CASTIGLIONI, S.; FANELLI, R. Identification of pharmaceuticals for human use contaminating the Italian aquatic environment. Journal of Hazardous Materials, v. 122, p. 205-209, 2005. https://dx.doi.org/10.1016/j.jhazmat.2005.03.001
https://dx.doi.org/10.1016/j.jhazmat.200...
). Therefore, in countries such as Portugal, FUR is expected to be found in wastewater treatment plants (WWTPs) (Santos et al., 2013SANTOS, L. H. M. L. M.; GOS, M.; RODRIGUEZ-MOZAZ, S.; DELERUE-MATOS, C.; PENA, A.; BARCELÓ, D.; MONTENEGRO, M. C. B. S. M. Contribution of hospital effluents to the load of pharmaceuticals in urban wastewaters: Identification of ecologically relevant pharmaceuticals. Science of the Total Environment, v. 461-462, p. 302-316, 2013. https://dx.doi.org/10.1016/j.scitotenv.2013.04.077
https://dx.doi.org/10.1016/j.scitotenv.2...
). According to the literature, the presence of FUR in WWTP influents and effluents can widely fluctuate from mg.L-1 to ng.L-1 (Jelic et al., 2011JELIC, A.; MERITXELL, G.; GINEBREDA, A.; CÉSPEDES-SÁNCHEZ, R.; VENTURA, F.; PETROVIC, M.; BARCELO, D. Occurrence, partition and removal of pharmaceuticals in sewage water and sludge during wastewater treatment. Water Research, v. 45, n. 3, p. 1165-1176, 2011. https://dx.doi.org/10.1016/j.watres.2010.11.010
https://dx.doi.org/10.1016/j.watres.2010...
; Salgado et al., 2010SALGADO, R.; NORONHA, J. P.; OEHMEN, A.; CARVALHO, G.; REIS, M. A. M. Analysis of 65 pharmaceuticals and personal care products in 5 wastewater treatment plants in Portugal using a simplified analytical methodology. Water Science and Technology, v. 62, n. 12, p. 2862-2871, 2010. https://dx.doi.org/10.2166/wst.2010.985
https://dx.doi.org/10.2166/wst.2010.985...
; Santos et al., 2013SANTOS, L. H. M. L. M.; GOS, M.; RODRIGUEZ-MOZAZ, S.; DELERUE-MATOS, C.; PENA, A.; BARCELÓ, D.; MONTENEGRO, M. C. B. S. M. Contribution of hospital effluents to the load of pharmaceuticals in urban wastewaters: Identification of ecologically relevant pharmaceuticals. Science of the Total Environment, v. 461-462, p. 302-316, 2013. https://dx.doi.org/10.1016/j.scitotenv.2013.04.077
https://dx.doi.org/10.1016/j.scitotenv.2...
), while in surface water bodies it usually occurs in low concentrations of µg.L-1 to ng.L-1 (Silva et al., 2011SILVA, B. F.; JELIC, A.; LÓPEZ-SERNA, R.; MOZETO, A. A.; PETROVIC, M.; BARCELÓ, D. Occurrence and distribution of pharmaceuticals in surface water, suspended solids and sediments of the Ebro river basin, Spain. Chemosphere, v. 88, n. 8, p. 1331-1339, 2011. https://dx.doi.org/10.1016/j.chemosphere.2011.07.051
https://dx.doi.org/10.1016/j.chemosphere...
; Valcárcel, 2011VALCÁRCEL, Y.; GOLZÁLEZ ALONSO, S.; RODRÍGUEZ-GIL, J. L.; ROMO MAROTO, R.; GIL, A.; CATALÁ, M. Analysis of the presence of cardiovascular and analgesic/anti-inflammatory/antipyretic pharmaceuticals in river- and drinking-water of the Madrid Region in Spain. Chemosphere, v. 82, p. 1062-1071, 2011. https://dx.doi.org/10.1016/j.chemosphere.2010.10.041
https://dx.doi.org/10.1016/j.chemosphere...
; Matamoros et al., 2012MATAMOROS, V.; ARIAS, C. A.; NGUYEN, L. X.; SALVADÓ, V.; BRIX, H. Occurrence and behavior of emerging contaminants in surface water and a restored wetland. Chemosphere, v. 88, p. 1083-1089, 2012. https://dx.doi.org/10.1016/j.chemosphere.2012.04.048
https://dx.doi.org/10.1016/j.chemosphere...
).

Although several studies show the presence of FUR in different environmental samples, its concentration varies widely, inducing uncertainty in the PhAC relevance in terms of environment contamination and particularly in the accurate efficiency assessment of the WWTPs on removing this contaminant. Therefore, developing a quantification method for FUR that accounts for the different interferences that can exist along the analyses is vital. Moreover, the use of simple matrices such as ultrapure water for preliminary tests can provide more clear results, especially for compounds such as FUR that are hydrophobic and have low solubilities (Table 1), and thus are more prone to be lost during simple sample preparation procedures, such as filtration. Saturation of the filter adsorption sites generally can minimize these losses (Pillai et al., 2016PILLAI, S.A.; CHOBISA, D.; URIMI, D.; RAVINDRA, N. Filters and Filtration: A Review of Mechanisms That Impact Cost, Product Quality and Patient Safety. Journal of Pharmaceutical Sciences and Research, v. 8, n. 5, p. 271-278, 2016.). However, sometimes the sample volume is not enough for pre-wetting the filters, and thus the selection of the most suitable type of filter can be a key aspect.

Table 1.
Structure, physical and chemical properties of furosemide.

Another relevant factor is that the literature on methodologies for FUR quantification as a single target is only common in the scope of human or animal health, or for pharmaceutical dosage control (Bosch et al., 2008BOSCH, M. E.; SANCHEZ, A. J. R.; ROJAS, F. S.; OJEDA, C. B. Recent developments in analytical determination of furosemide. Journal of Pharmaceutical and Biomedical Analysis, v. 48, p. 519-532, 2008. https://dx.doi.org/10.1016/j.jpba.2008.07.003
https://dx.doi.org/10.1016/j.jpba.2008.0...
). Hence, most of the studies on FUR determination were performed in matrices of human and animal urine or plasma. The complexity of such matrices can mask other interferences along the analytical procedure.

This study sought to contribute to the definition of an appropriate methodology for FUR determination in a water matrix by HPLC-UV and to better understand the stability of the compound along the analysis-preparation procedure.

2. MATERIALS AND METHODS

2.1. Chemicals and materials

Analytical grade furosemide (99.8% purity) was purchased from Sigma-Aldrich (Lisbon, Portugal); see Table 1 for its main characteristics. Phosphoric acid (>85% purity), HPLC-grade solvents acetonitrile and methanol were obtained from Enzymatic, S.A. (Loures, Portugal). Ultra-pure water was prepared from a Millipore Milli Q system. Both Whatman 13 Puradisc syringe filters (nylon and PVDF - Polyvinylidene fluoride) were supplied by Enzymatic, S.A. (Loures, Portugal).

2.2. Standards solutions

Furosemide stock standard solution containing 10 mg.L-1 was prepared. Furosemide was dissolved in 1 ml of methanol and filled to the 1 L mark with ultra-pure water. An ultrasonic bath was used to help the dissolution of furosemide. For furosemide quantification, standard solutions were prepared within the range of 0.1-4 mg.L-1. All solutions were stored in the dark and covered with aluminium foil to avoid photodegradation. The 10 mg.L-1 FUR solution was scanned between the 190 and 350 nm UV region. The wavelength of 233 nm corresponding to the maximum absorption peak has been therefore selected to carry on the experiment. Electric conductivity and pH were also assessed.

2.3. Quantification and analytical method validation

Furosemide quantification was carried out in a HPLC - Beckman Coulter System Gold, with a Solvent Module 126 and a Diode Array Detector 168, using 32 Karat Software, version 8.0, with a variable wavelength detector (190-800 nm) at 233 nm and with a 20 µL volume injector loop. Samples were determined with a flow rate of 1 mL.min-1. A reversed-phase analytical column Zorbax Eclipse XDB-C8 (4.6x150 mm; 5 µm) was used. The composition of the mobile phase used in furosemide analyses was previously selected for an isocratic elution mode using trial mixtures of acetonitrile and ultra-pure water acidified with phosphoric acid (0.1% v/v) in different proportions (50:50, 55:45, 60:40, 65:35 and 70:30). The mixture composed of 60% acetonitrile and 40% ultra-pure water (acidified with phosphoric acid) was selected, as the corresponding retention time for furosemide was below 4 min., thereby allowing the quantification of the pharmaceutical in a reasonably short time. Samples were injected (three replicates) with an automatic injector manufactured by Spark Holland BV - MIDAS, at room temperature (16-20˚C).

This analytical method was validated according to the International Conference on Harmonization (ICH) guidelines for validation of analytical procedures (United States, 1996UNITED STATES. Food and Drug Administration. ICH Q1B: Photostability testing of new drug substances and products. Washington, 1996.), based on the following parameters: linearity, sensitivity, precision (intra-day) and accuracy.

A calibration curve was constructed using a set of furosemide standard solutions with concentrations of 0.1-4 mg.L-1. Three replicate injections were made for each standard solution. The regression equation and the value of the correlation coefficient were calculated using linear regression analysis. The sensitivity of the analytical method was estimated in terms of the limits of detection (LOD) and of quantification (LOQ). LOD and LOQ were defined based on signal-to-noise ratios of 3:1 and 10:1, respectively. Moreover, for the assessment of the analytical method’s accuracy and precision (intra-day), standard solutions at seven different concentrations (from 0.25 mg.L-1 to 4 mg.L-1) were injected at three different times in the same day. The precision of the proposed method was obtained by calculating the relative standard deviation (RSD) values of the peak areas for the three different injections with acceptance criteria of not more than 2% (United States, 1996UNITED STATES. Food and Drug Administration. ICH Q1B: Photostability testing of new drug substances and products. Washington, 1996.). The accuracy of the quantification method was assessed through the percentage ratio between the measured concentrations of the furosemide standards and their nominal concentrations.

2.4. Procedure

Different syringe filters were tested to study the FUR recovery percentage throughout the process. Nylon and PVDF (Polyvinylidene fluoride), both with 0.45 µm pore size as recommended for HPLC analysis, were selected. The filter size (13 mm) was chosen in accordance with the low sample volume required for further analysis. Through HPLC analysis, filter recovery capacity was tested for 2 mg.L-1 of FUR solution, filtering a volume of 2 mL. Additionally, the recovery capacity was further validated comparing a standard curve, between 0.25 and 2 mg.L-1 of FUR, with and without filtration.

For FUR degradation studies, an experiment over 8 days was performed under three different storage conditions. At room temperature (12-18˚C), three replicates of a FUR solution with approximately 2 mg.L-1 were exposed to daylight (L), and the other three were kept in the dark (D) to test photodegradation sensitivity. The effect of storage temperature was tested with another three replicates that were kept at 4˚C. Samples were taken from all the standard solutions after 0, 24, 48, 72 and 192 hours of exposure and analysed through HPLC.

3. RESULTS AND DISCUSSION

3.1. Furosemide quantification

3.1.1. Furosemide recovery

Figure 1 compares the extent of FUR retention by two different filters, Nylon and PVDF, for the same solution. Nylon filters presented 75% more FUR adsorption than PVDF filters. Hence, the PVDF filter seems more efficient for FUR solution filtration. Several factors can affect adsorption onto membranes, from characteristics of the compounds to the properties of the filters. Both filters have hydrophilic membranes, the same pore size (0.45µm) and the same effective filter area (1.3 cm2). Nylon filters, compared with PVDF, bind proteins and are more suitable for filtering samples with a high pH. In turn, PVDF filters are highly inert and have low protein binding (GE Healthcare, 2018GE HEALTHCARE. Digizuite. 2018. Available at: Available at: https://cdn.gelifesciences.com/dmm3bwsv3/AssetStream.aspx?mediaformatid=10061&destinationid=10016&assetid=16239 . Access: April 18th, 2018.
https://cdn.gelifesciences.com/dmm3bwsv3...
). Since the standard samples have a pH of 5.5±0.17, both filters were appropriate. Hence, the results can be an outcome of the inert feature of the PVDF filters. To further validate the PVDF filters’ suitability, a comparison of a standard curve with and without filtration was performed (see Figure 2). No significant losses were observed. Therefore, the PVDF filter was considered appropriate to carry on experiments with FUR.

Figure 1.
FUR retention (2 mg.L-1) in Nylon and PVDF filters.

Figure 2.
Standard curve of FUR for samples without filtration and filtered through PVDF filters.

3.1.2. Method validation

An optimal peak shape and retention time of FUR was obtained for the selected chromatographic conditions (Figure 3). The method employed also showed a linear pattern for the tested concentration range (0.1-4.0 mg.L-1) with a correlation r2=0.9999 (Figure 4). Limit of quantification (LOQ) was 0.144 mg.L-1 and limit of detection (LOD) was 0.048 mg.L-1 (Equations 1 and 2, respectively). The accuracy of the method was estimated to be within 93.7-101.8%, whereas in regard to the intra-day precision, RSD of measurements were below 2%. Therefore, the method can be considered to provide a good sensitivity, precision and accuracy.

L O Q = 10 x σ s l o p e o f c a l i b r a t i o n c u r v e (1)

L O D = 3.3 x σ s l o p e o f c a l i b r a t i o n c u r v e (2)

Figure 3.
Chromatogram of FUR in ultrapure water matrix.

Figure 4.
Linearity of FUR in an ultrapure water matrix.

3.2. Degradation studies

Figure 5 shows FUR stability over 8 days under different storage conditions. The FUR solutions kept from light exposure maintained their concentration throughout the experiment. Storage temperature does not seem to affect FUR concentration. Moreover, since FUR quantification is performed at room temperature, FUR-solution storage should benefit if similar temperatures are maintained to avoid potential precipitation. Figure 5 shows that during the first 24h of exposure there is a slight increase in concentration in 4D solutions. The minor increase is due to the initial lower values of one of the replicates. Poor initial solubility of the replicate could explain the lower value at 0 hours of exposure, compared to the 24-hour sample.

Figure 5.
FUR conservation over 8 days for three different conditions: L - Light exposure at room temperature; D - Kept in the dark at room temperature; and, 4D - Kept in the dark at 4˚C.

FUR's sensitivity to light is well-documented, and the appearance of photodegradation products as a result of the exposure to light is to be expected (Greca et al., 2004GRECA, M. D.; IESCE, M. R.; PREVITERA, L.; RUBINO, M.; TEMUSSI, F. A new photoproduct of the drug furosemide in aqueous media. Environmental Chemistry Letters, v. 2, p. 155-158, 2004. https://dx.doi.org/10.1007/s10311-004-0080-9
https://dx.doi.org/10.1007/s10311-004-00...
). In fact, Figure 6a shows the presence of two new peaks (2 and 3) along with the peak of FUR (1). Substance (3) appears in all samples after 24h and disappears before 48h of exposure; therefore, light exposure and temperature are not the key factors to explain this phenomenon. Moreover, substance (3) seems to suffer further degradation since it disappears after 48h.

In the first 72h of exposure, the decrease of FUR is followed by an increase of the concentration of substance (2). Afterwards, substance (2) decreases to concentrations below detection levels. Greca et al., (2004)GRECA, M. D.; IESCE, M. R.; PREVITERA, L.; RUBINO, M.; TEMUSSI, F. A new photoproduct of the drug furosemide in aqueous media. Environmental Chemistry Letters, v. 2, p. 155-158, 2004. https://dx.doi.org/10.1007/s10311-004-0080-9
https://dx.doi.org/10.1007/s10311-004-00...
tested the photodegradation of furosemide in different water matrices through a 150-W solar simulator over 36h. In all water matrices, including a simple distilled water medium, only one new photoproduct was found. The same result occurred when the distilled water solution was exposed to natural solar radiation. Therefore, the main photoproduct mentioned above, substance (2), which is visible in the solution exposed to light, can be the same as the one found in the previously referred to work. FUR, in aqueous solutions exposed to UV light, hydrolyses into 2-amino-4-chloro-5 sulfamoylanthranilic acid (CSA) and furfuryl alcohol that will be further degraded into levulinic acid (Ruiz‐Angel et al., 2006RUIZ‐ANGEL, M. J.; BERTHOD, A.; CARDA‐BROCH, S.; GARCÍA‐ÁLVAREZ‐COQUE, M. C. Analytical Techniques for Furosemide Determination. Separation & Purification Reviews, v. 35, n. 2, p. 39-58, 2006. https://dx.doi.org/10.1080/15422110600671726
https://dx.doi.org/10.1080/1542211060067...
). Therefore, substance (2) could be the FUR photoproduct CSA, but further liquid chromatography mass spectrometry analysis should be undertaken for substance identification (Figure 7).

Figure 6.
Chromatogram of furosemide in ultrapure water under: a) light exposure/room temperature; b) darkness/room temperature; c) darkness/4˚C. (Bundgaard et al., 1988BUNDGAARD, H.; NØRGAARD, T.; MØRKNIELSEN, N. Photodegradation and hydrolysis of furosemide and furosemide esters in aqueous solutions. International Journal of Pharmaceutics, v. 42, n. 1-3, p. 217-224, 1988. https://dx.doi.org/10.1016/0378-5173(88)90178-0
https://dx.doi.org/10.1016/0378-5173(88)...
) Continue.
Figure 6.
Continue.
Figure 6.
Continue.

Figure 7.
a) Furosemide photodegradation over eight days; b) First-order kinetic of furosemide photodegradation.

4. CONCLUSIONS

The developed method allows fast and accurate determination of FUR in water solution by HPLC-UV. In the 0.1-4 mg.L-1 concentration range, the quantification of FUR follows a linear regression. Due to its low solubility and sorption susceptibility, FUR standard solutions should be filtered through PVDF filters or others with a similar inert feature. Stability studies validated the well-known photo sensitivity characteristic of FUR. Total photodegradation of a FUR solution was achieved after 48h of exposure to natural light. Therefore, it is recommended that FUR standards solutions should be stored in the dark at room temperature to guarantee concentration conservation. The study emphasizes that when targeting FUR in low volume environmental samples, special attention should be given to the chosen filter, as well as to storage conditions to avoid underestimation errors.

5. ACKNOWLEDGEMENTS

The present study was supported by the FCT - Fundação para a Ciência e a Tecnologia (doctoral grant SFRH/BD/52511/2014), inserted in the doctoral program FCT-FLUVIO and River Restoration and Management (Reference: PD/00424/2012). A special thanks to researcher Antonio Eduardo Leitão for access to HPLC-UV equipment.

6. REFERENCES

  • BERTHOD, A.; CARDA-BROCH, S.; GARCIA-ALVAREZ-COQUE, M. C. Hydrophobicity of Ionizable Compounds. A theoretical study and measurements of diuretic octanol-water partition coefficients by countercurrent chromatography. Analytical Chemistry, v. 71, p. 879-888, 1999. https://dx.doi.org/10.1021/ac9810563
    » https://dx.doi.org/10.1021/ac9810563
  • BÖGER, B.; AMARAL, B.; SILVEIRA, P. L.; WAGNER, R.; PERALTA-ZAMOR, P. G.; GOMES, E. C. Determination of carbamazepine and diazepam by SPE-HPLC-DAD in Belém river water, Curitiba-PR/Brazil. Revista Ambiente & Água, v. 13, n. 2, 2018. https://dx.doi.org/10.4136/ambi-agua.2196
    » https://dx.doi.org/10.4136/ambi-agua.2196
  • BOSCH, M. E.; SANCHEZ, A. J. R.; ROJAS, F. S.; OJEDA, C. B. Recent developments in analytical determination of furosemide. Journal of Pharmaceutical and Biomedical Analysis, v. 48, p. 519-532, 2008. https://dx.doi.org/10.1016/j.jpba.2008.07.003
    » https://dx.doi.org/10.1016/j.jpba.2008.07.003
  • BUNDGAARD, H.; NØRGAARD, T.; MØRKNIELSEN, N. Photodegradation and hydrolysis of furosemide and furosemide esters in aqueous solutions. International Journal of Pharmaceutics, v. 42, n. 1-3, p. 217-224, 1988. https://dx.doi.org/10.1016/0378-5173(88)90178-0
    » https://dx.doi.org/10.1016/0378-5173(88)90178-0
  • CARLSON, M.; THOMPSON, R. D. Analyte Loss Due to Membrane Filter Adsorption as Determined by High-Performance Liquid Chromatography. Journal of Chromatographic Science, v. 38, p. 78-83, 2000. https://dx.doi.org/10.1093/chromsci/38.2.77
    » https://dx.doi.org/10.1093/chromsci/38.2.77
  • CRUZ-MORATÓ, C.; LUCAS, D.; LLORCA, M.; RODRIGUEZ-MOZAZ, S.; GORGA, M.; PETROVIC, M.; BARCELÓ, D.; VICENT, T.; SARRÀ, M.; MARCO-URREA, E. Hospital wastewater treatment by fungal bioreactor: Removal efficiency for pharmaceuticals and endocrine disruptor compounds. Science of the Total Environment, v. 493, p. 365-376, 2014. https://dx.doi.org/10.1016/j.scitotenv.2014.05.117
    » https://dx.doi.org/10.1016/j.scitotenv.2014.05.117
  • GE HEALTHCARE. Digizuite. 2018. Available at: Available at: https://cdn.gelifesciences.com/dmm3bwsv3/AssetStream.aspx?mediaformatid=10061&destinationid=10016&assetid=16239 Access: April 18th, 2018.
    » https://cdn.gelifesciences.com/dmm3bwsv3/AssetStream.aspx?mediaformatid=10061&destinationid=10016&assetid=16239
  • GRECA, M. D.; IESCE, M. R.; PREVITERA, L.; RUBINO, M.; TEMUSSI, F. A new photoproduct of the drug furosemide in aqueous media. Environmental Chemistry Letters, v. 2, p. 155-158, 2004. https://dx.doi.org/10.1007/s10311-004-0080-9
    » https://dx.doi.org/10.1007/s10311-004-0080-9
  • INFARMED. Estatística do Medicamento. 2014. Available at: Available at: https://www.infarmed.pt/documents/15786/1229727/Estat%C3%ADstica+do+Medicamento+2014/988074f4-4f89-4a7c-9055-844cb88e93fd?version=1.2 Access: April 18th, 2018.
    » https://www.infarmed.pt/documents/15786/1229727/Estat%C3%ADstica+do+Medicamento+2014/988074f4-4f89-4a7c-9055-844cb88e93fd?version=1.2
  • JELIC, A.; MERITXELL, G.; GINEBREDA, A.; CÉSPEDES-SÁNCHEZ, R.; VENTURA, F.; PETROVIC, M.; BARCELO, D. Occurrence, partition and removal of pharmaceuticals in sewage water and sludge during wastewater treatment. Water Research, v. 45, n. 3, p. 1165-1176, 2011. https://dx.doi.org/10.1016/j.watres.2010.11.010
    » https://dx.doi.org/10.1016/j.watres.2010.11.010
  • KUSTER, M.; LÓPEZ, J.; DE ALDA, M.; HERNANDO, M. D.; PETROVIC, M.; MARTIN-ALONSO, J.; BARCELÓ, D. Analysis and occurrence of pharmaceuticals, estrogens, progestogens and polar pesticides in sewage treatment plant effluents, river water and drinking water in the Llobregat river basin (Barcelona, Spain). Journal of Hydrology, v. 358, p. 112-123, 2008. https://doi.org/10.1016/j.jhydrol.2008.05.030
    » https://doi.org/10.1016/j.jhydrol.2008.05.030
  • MARTÍN, J.; CAMACH-MUÑOZ, D.; SANTOS, J. L.; APARICIO, I.; ALONSO, E. Occurrence of pharmaceutical compounds in wastewater and sludge from wastewater treatment plants: Removal and ecotoxicological impact of wastewater discharges and sludge disposal. Journal of Hazardous Materials, v. 239-240, p. 40-47, 2012. https://dx.doi.org/10.1016/j.jhazmat.2012.04.068
    » https://dx.doi.org/10.1016/j.jhazmat.2012.04.068
  • MATAMOROS, V.; ARIAS, C. A.; NGUYEN, L. X.; SALVADÓ, V.; BRIX, H. Occurrence and behavior of emerging contaminants in surface water and a restored wetland. Chemosphere, v. 88, p. 1083-1089, 2012. https://dx.doi.org/10.1016/j.chemosphere.2012.04.048
    » https://dx.doi.org/10.1016/j.chemosphere.2012.04.048
  • O'NEIL, M. J. (ed.). The Merck Index - An Encyclopedia of Chemicals, Drugs, and Biologicals. 13th Ed. Whitehouse Station: Merck and Co., 2001.
  • OECD. Demographic references: General demographics. 2016. Available online at Available online at https://stats.oecd.org/Index.aspx?DataSetCode=HEALTH_DEMR Access: Feb. 14th 2018.
    » https://stats.oecd.org/Index.aspx?DataSetCode=HEALTH_DEMR
  • PAÍGA, P.; DELERUE-MATOS, C. Determination of pharmaceuticals in groundwater collected in five cemeteries’ areas (Portugal). Science of the Total Environment, v. 569-570, p. 16-22, 2016. https://dx.doi.org/10.1016/j.scitotenv.2016.06.090
    » https://dx.doi.org/10.1016/j.scitotenv.2016.06.090
  • PEREIRA, A. M. P. T.; SILVA, L. J. G.; LARANJEIRO, C. S. M.; MEISEL, L. M.; LINO, C. M.; PENA, A. Human pharmaceuticals in Portuguese rivers: The impact of water scarcity in the environmental risk. Science of the Total Environment, v. 609, p. 1182-1191, 2017. https://dx.doi.org/10.1016/j.scitotenv.2017.07.200
    » https://dx.doi.org/10.1016/j.scitotenv.2017.07.200
  • PILLAI, S.A.; CHOBISA, D.; URIMI, D.; RAVINDRA, N. Filters and Filtration: A Review of Mechanisms That Impact Cost, Product Quality and Patient Safety. Journal of Pharmaceutical Sciences and Research, v. 8, n. 5, p. 271-278, 2016.
  • PRANDOTA, J.; WITKOWSKA, M. Pharmacokinełics and mełabolism of furosemide in man. European Journal of Drug Metabolism and Pharmacokinetics, v. 4, p. 177-181, 1976. https://dx.doi.org/10.1007/BF03189275
    » https://dx.doi.org/10.1007/BF03189275
  • RUIZ‐ANGEL, M. J.; BERTHOD, A.; CARDA‐BROCH, S.; GARCÍA‐ÁLVAREZ‐COQUE, M. C. Analytical Techniques for Furosemide Determination. Separation & Purification Reviews, v. 35, n. 2, p. 39-58, 2006. https://dx.doi.org/10.1080/15422110600671726
    » https://dx.doi.org/10.1080/15422110600671726
  • SALGADO, R.; NORONHA, J. P.; OEHMEN, A.; CARVALHO, G.; REIS, M. A. M. Analysis of 65 pharmaceuticals and personal care products in 5 wastewater treatment plants in Portugal using a simplified analytical methodology. Water Science and Technology, v. 62, n. 12, p. 2862-2871, 2010. https://dx.doi.org/10.2166/wst.2010.985
    » https://dx.doi.org/10.2166/wst.2010.985
  • SANGSTER, J. LOGKOW Databank. Montreal Quebec: Sangster Res. Lab., 2001.
  • SANTOS, L. H. M. L. M.; GOS, M.; RODRIGUEZ-MOZAZ, S.; DELERUE-MATOS, C.; PENA, A.; BARCELÓ, D.; MONTENEGRO, M. C. B. S. M. Contribution of hospital effluents to the load of pharmaceuticals in urban wastewaters: Identification of ecologically relevant pharmaceuticals. Science of the Total Environment, v. 461-462, p. 302-316, 2013. https://dx.doi.org/10.1016/j.scitotenv.2013.04.077
    » https://dx.doi.org/10.1016/j.scitotenv.2013.04.077
  • SILVA, B. F.; JELIC, A.; LÓPEZ-SERNA, R.; MOZETO, A. A.; PETROVIC, M.; BARCELÓ, D. Occurrence and distribution of pharmaceuticals in surface water, suspended solids and sediments of the Ebro river basin, Spain. Chemosphere, v. 88, n. 8, p. 1331-1339, 2011. https://dx.doi.org/10.1016/j.chemosphere.2011.07.051
    » https://dx.doi.org/10.1016/j.chemosphere.2011.07.051
  • TERNES, T. A. Occurrence of drugs in german sewage treatment plants and rivers. Water Research, v. 32, n. 11, p. 3245-3260, 1998. https://dx.doi.org/10.1016/S0043-1354(98)00099-2
    » https://dx.doi.org/10.1016/S0043-1354(98)00099-2
  • UNITED STATES. Food and Drug Administration. ICH Q1B: Photostability testing of new drug substances and products. Washington, 1996.
  • VALCÁRCEL, Y.; GOLZÁLEZ ALONSO, S.; RODRÍGUEZ-GIL, J. L.; ROMO MAROTO, R.; GIL, A.; CATALÁ, M. Analysis of the presence of cardiovascular and analgesic/anti-inflammatory/antipyretic pharmaceuticals in river- and drinking-water of the Madrid Region in Spain. Chemosphere, v. 82, p. 1062-1071, 2011. https://dx.doi.org/10.1016/j.chemosphere.2010.10.041
    » https://dx.doi.org/10.1016/j.chemosphere.2010.10.041
  • YALKOWSKY, S. H.; DANNENFELSER, R. M. Aquasol Database of Aqueous Solubility. 5th ed. Tucson: College of Pharmacy, University of Arizona, 1992.
  • ZUCCATO, E.; CASTIGLIONI, S.; FANELLI, R. Identification of pharmaceuticals for human use contaminating the Italian aquatic environment. Journal of Hazardous Materials, v. 122, p. 205-209, 2005. https://dx.doi.org/10.1016/j.jhazmat.2005.03.001
    » https://dx.doi.org/10.1016/j.jhazmat.2005.03.001