Evaluation of the tip of standardized D0 gutta percha cones of four Rotary systems, by means of an endodontic ruler

Avaliação da adaptação de cones estandardizados de guta-percha de quatro sistemas mecanizados por meio de uma régua endodôntica calibradora

ABSTRACT

Objective

: This study aimed to evaluate the D0 diameter of standardized gutta-percha cones for four mechanized systems: ProTaper Universal®, Mtwo®, Wave One® and Reciproc®.

Methods:

An endodontic calibrator ruler was used to measure the D0 diameter of 60 main cones of the above systems. Measurements were made according 3 scores as follows: gutta-percha cone fits exactly into the same hole as the endodontic ruler gauge (score 1), cone falls short of the ruler gauge whole size (score 2); or exceeding the ruler gauge hole size (score 3). The diameters D0 obtained were compared with the values reported by manufacturers. All data were analyzed by means of T Test, at 1% level of significance.

Results:

The average value measured was significantly higher than the measures established by manufacturers (p <0.001).

Conclusion

: The gutta-percha cones of ProTaper Universal® systems Mtwo®, Wave One® and Reciproc® were not standardized, except for R40 cone of the Reciproc® system (Reciproc®). The trend was for variation in the D0 Diameter towards increasing measurement values.

Indexing terms:
Dimensional measurement accuracy; Endodontics; Gutta-percha; Root canal obturation

RESUMO

Objetivo:

Avaliar o diâmetro D0 de cones estandardizados de guta-percha de quatro sistemas mecanizados:ProTaper Universal®, Mtwo®, Wave One® e Reciproc® .

Métodos:

Por meio de uma régua endodôntica calibradora, foi aferido o diâmetro D0 de sessenta cones principais dos sistemas supracitados. A aferição foi realizada de acordo com os três escores a seguir: o cone de guta-percha se adapta exatamente no mesmo calibre de orifício da régua endodôntica (escore 1), o cone fica aquém ao calibre de orifício da régua (escore 2) ou além do respectivo calibre do orifício da régua (escore 3). Os diâmetros em D0 obtidos foram comparados aos valores divulgados pelos fabricantes. Todos os dados foram tratados por meio do Teste T, ao nível de significância de 1%.

Resultados:

As médias aferidas foram significativamente maiores do que as medidas estabelecidas pelos fabricantes (p<0,001). Somente os cones R40 do sistema Reciproc® estiveram no padrão.

Conclusão:

Os cones de guta-percha dos sistemas ProTaper Universal®, Mtwo®, Wave One® e Reciproc® não apresentam padronização, à exceção do cone R40 (Reciproc® ). Houve uma tendência de variação do diâmetro D0 para uma maior medida.

Termos de indexação:
Precisão da medição dimensional; Endodontia; Guta-percha; Obturação do Canal Radicular

INTRODUCTION

One of the factors related to successful endodontic treatment is hermetic sealing of the root canal, performed during the filling stage11 Cueva-Goig R, Forner-Navarro L, Llena-Puy MC. Microscopic assessment of the sealing ability of three endodontic filling techniques. J Clin Exp Dent. 2016;8(1):27-31. doi: 10.4317/jced.52847
https://doi.org/10.4317/jced.52847...
. The correct choice of the main gutta percha cone is essential to obtain a close fit to the dentin walls and interlocking it at the working length. Correct adaptation of the gutta percha cone to the apical region is fundamentally known to depend on perfect coincidence of its D0 diameter with that of the instrument used for apical preparation22 Moura AAM, Carvalho CF, Santos M, Davidowicz H. Avaliação in vitro através da análise radiográfica entre cones de guta-percha estandardizados e calibrados adaptados à região apical. Rev Inst Ciênc Saúde. 1994;12(1):7-11..

Diverse studies analyzing the diameter and morphology of gutta percha cones have warned about the lack of standardization in the manufacturing process33 Davidowicz H, Moura AAM, Strefezza F. Avaliação do diâmetro "D0" de cones de guta-percha estandardizados e calibrados comparados com limas tipo "K" através do microscópio comparador. Rev ABO Nac. 1994;2(3):181-5.

4 Ferreira JJ, Rhodes JS, Pittford TR. The efficacy of gutta-percha removal using profiles. Int Endod J. 2001;(34):267-74. doi: 10.1046/j.1365-2591.2001.00379.x
https://doi.org/10.1046/j.1365-2591.2001...

5 Moule AJ, Kellaway R, Clarkson R, Rowell J, Macfarlane R, Lewis D, et al. Variability of master gutta-percha cones. Aust Endod J. 2002;28(1):38-43. doi: 10.1111/j.1747-4477.2002.tb00365.x
https://doi.org/10.1111/j.1747-4477.2002...
-66 Borges AH, Dorileo MCO, Pedro FLM, Segundo AS, Volpato LER, Semenoff TADV, et al. Avaliação da padronização dos cones de guta-percha de diferentes conicidades. Rev Odontol Bras Central. 2011;20(55):313-6.. This reality harms the objectives of standardizing materials, and consequently, makes clinical practice difficult.

Due to the difficulties with handling gutta percha, it is believed that there is practically no uniformity of their calibers. For example, according to Mayne et al.77 Mayne JR, Shapiro S, Abramson II. An evaluation of standardized gutta-percha points. Part I - Reliability and validity of standardization. Oral Surg Oral Med Oral Pathol. 1971;31(2):250-7. doi: 10.1016/0030-4220(71)90080-6
https://doi.org/10.1016/0030-4220(71)900...
and Goldberg et al.88 Goldberg F, Gurfinkel J, Spielberg C. Microscopic study of standardized gutta-percha points. Oral Surg Oral Med Oral Pathol. 1979;47(3):275-6. doi: 10.1016/0030-4220(79)90153-1
https://doi.org/10.1016/0030-4220(79)901...
, among the contents of a box of No.40 gutta percha cones, variations in their diameter range from Number 35 to 45, demonstrating imprecision in the entire sample analyzed.

Parallel to this condition, new materials used for root canal filling have appeared, bearing in mind the evolution of various mechanized root preparation systems. Filling techniques with a single cone have been recommended99 Hörsted-Bindslev P, Andersen MA, Jensen MF, Nilsson JH, Wenzel A. Quality of molar root canal fillings performed with the lateral compaction and the single-cone technique. J Endod. 2007;33(4):468-71. doi: 10.1016/j.joen.2006.12.016
https://doi.org/10.1016/j.joen.2006.12.0...
-1010 Hegde V, Arora S. Sealing ability of three hydrophilic single-cone obturation systems: an in vitro glucose leakage study. Contemp Clin Dent. 2015;6(1):S86-9. doi: 10.4103/0976-237x.152953
https://doi.org/10.4103/0976-237x.152953...
. Therefore, the rotary system and its respective gutta percha cone must be carefully measured.

This being the case, the aim of this study was to use an endodontic calibration ruler to evaluate the fit in D0 of the gutta percha cones of the ProTaper Universal®, Mtwo®, WaveOne® and Reciproc® systems.

METHODS

In total, 60 gutta percha cones of the following systems were used: Protaper Universal® - F1, F2 and F3 (Dentsply/Maillefer Instruments S.A., Ballaigues, Switzerland), Mtwo® - 25.06, 30.05, 35.04, 40.04 and 25.07 (VDW® Endodontic Synergy, Munich, Germany), Wave One® - Small, Primary and Large (Dentsply/Maillefer Instruments S.A., Ballaigues, Switzerland) and Reciproc® - R25, R40 and R50 (VDW® Endodontic Synergy, Munich, Germany). To detect a difference in the means of the measurements observed in the groups, attaining a statistical power of 90%, and a 1% level of significance, the need for 60 gutta percha cones for each of the systems was calculated. The boxes of cones of each of the systems belonged to the same manufacturing lot.

To verify the fit of the cones in D0, an endodontic calibrating ruler (Dentsply/Maillefer Instruments S.A., Ballaigues, Switzerland) was used. The diameter of the holes in the ruler were previously measured and considered correctly calibrated, by mean of Image Tool® software .

Three distinct situations were taken into consideration during verification of adaptation of the cones. In the first situation, the gutta percha cone fitted exactly into the same caliber hole as that of the endodontic ruler, generating score 1 (Figure 1).

Figure 1
D0 of gutta-percha cone fitted into hole of the Maillefer® endodontic calibrating ruler.

In the second situation, the gutta percha cone was smaller than the hole of the calibrating rule, and was tested in the previous hole in the sequence. This act was repeated until the cone fitted perfectly, thereby generating Score 2.

Whereas, in the third situation, the gutta percha cone exceeded the size of the hole of the calibrating rule, and was tested in the subsequent hole in the sequence. This act was repeated until the cone fitted perfectly, thereby generating Score 3 (Figure 2).

Figure 2
D0 of gutta-percha cone exceeding the hole diameter equivalent of the Maillefer® endodontic calibrating ruler.

A single, previously trained operator verified the fit of the gutta percha cones with the aid of a magnifying glass providing 4x magnification (Bio-Art Equipamentos Odontológicos Ltda., São Carlos, São Paulo, Brazil).

The data obtained in analysis of the cones were statistically treated by means of the paired T test at a 1% level of significance. For analysis the program SPSS version 22.0 (SPSS Inc, Chicago, IL) was used.

RESULTS

Table 1 presents the results obtained for the gutta percha cones in relation to the endodontic calibrating ruler.

Table 1
Table with means and standard deviations relative to cross-matching the sizes of the gutta percha cones with the holes of the endodontic calibrating ruler.

The means shown in Table 1 denoted values significantly higher than those of the measurements established by the manufacturers (p<0.001). Only cones R40 of the Reciproc® system maintained the standard.

DISCUSSION

The lack of uniformity and standardization of gutta percha cones are fundamental factors for perfect root canal system sealing1111 Valois CR, Silva LP, Azevedo RB, Costa Junior ED. Atomic force microscopy study of gutta-percha cone topography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2004;98(2):250-5. doi: 10.1016/s1079210404001854
https://doi.org/10.1016/s107921040400185...

12 Silva EJ, Rocha AC, Ferreira C, Herrera DR, Coutinho-Filho TS. Assessment of the tip surface of gutta-percha cones after different cutting methods. Acta Odontol Latinoam. 2011;24(3):236-9.
-1313 Celikten B, F Uzuntas C, I Orhan A, Tufenkci P, Misirli M, O Demiralp K, Orhan K. Micro-CT assessment of the sealing ability of three root canal filling techniques. J Oral Sci. 2015;57(4):361-6. doi: 10.2334/josnusd.57.361
https://doi.org/10.2334/josnusd.57.361...
.

The choice of analyzing position D0 of the gutta percha cones was because this is a critical region, responsible for its fit into the apical limit of the prepared root canal. The mistaken choice of gutta percha cone, such as for example, when it exceeds the limit of the working length established by odontometry, may frequently lead to endodontic treatment failure.

The authors opted to verify the fit of gutta percha cones by means of an endodontic calibrating ruler (Gutta Percha Gauge), because it is a resource the orthodontist frequently uses to verify or correct the diameter of cones during clinical attendance, because it is a method that is easy to handle and to acquire.

Although the literature reports other resources for checking the diameter of gutta percha cones, such as the use of a microscope or profilometer, the use of the calibrating ruler when compared with other previously mentioned methodologies, has shown similar results33 Davidowicz H, Moura AAM, Strefezza F. Avaliação do diâmetro "D0" de cones de guta-percha estandardizados e calibrados comparados com limas tipo "K" através do microscópio comparador. Rev ABO Nac. 1994;2(3):181-5.,88 Goldberg F, Gurfinkel J, Spielberg C. Microscopic study of standardized gutta-percha points. Oral Surg Oral Med Oral Pathol. 1979;47(3):275-6. doi: 10.1016/0030-4220(79)90153-1
https://doi.org/10.1016/0030-4220(79)901...
,1414 Lopes HP. Diâmetro e conicidades dos cones de guta-percha acessórios tipo M, FM e F. RBO. 1999;56(5):252-4.-1515 Camões ICG, Reis PEG, Freitas LF, Gomes CC, Pinto SS. Avaliação do calibre apical de cones de guta-percha de três marcas comerciais. Pesq Bras Odontoped Clin Integr. 2006;6(2):111-6..

During analysis of the results, the authors found that the majority of gutta percha cones, whether with reference to continuous rotary or reciprocating instruments, did not follow any standardization relative to their diameter at D0.

Specification No. 28 of the ADA has admitted that there is a discrepancy of up to 0.02 mm between the standard measurements. This tolerance is perfectly understandable when considering the complexity involved in manufacturing such delicate products. However, when dealing with a plastic material with a multiplicity of compositions, we need to question whether the gutta percha could undergo dimensional changes derived from factors such as, for example, storage conditions and ambient temperature from the time of their measurement up to their acquisition.

The data obtained in the study of Capar et al.19 are in agreement with those presented in our study. The authors also verified various degrees of imprecision in the gutta-percha cones of the Protaper Next®, Twisted File Adaptive®, OneShape®, Protaper Universal®, WaveOne® and Reciproc® systems.

According to some authors, this imprecision in the D0 diameter of the cones reinforces the importance of using thermoplasticized gutta-percha techniques, with the purpose of correcting defects or faults that may arise during the stage of performing root canal filling1616 Bombana AC, Facchini MEB, Moura AAM. Avaliaçäo dimensional de cones de guta-percha estandardizados de diferentes procedências. Rev Inst Ciênc Saúde. 1989;7(1):5-14.

17 Figueiredo JAP. Assessment of the diameter of gutta percha cones of different manufacturers. Braz Endod J. 2000; 4(1): 23-8.
-1818 Cunningham KP, Walker MP, Kulild JC, Lask JT. Variability of the diameter and taper of size #30, 0.04 gutta-percha.

In view of these findings, we must place great value on the clinical tactile sensitivity of the operator who checks locking of the main cone. The professional who imagines using the main cone adequately, as being related only to the final instrument, is a candidate for leaving empty spaces or some filled only with endodontic cement in the apical region.

CONCLUSION

According to the results obtained, it was possible to conclude that the gutta percha cones of the ProTaper Universal®, Mtwo®, Wave One® and Reciproc® systems showed no standardization, with the exception of cone R40 (Reciproc®). The trend was for variation in the D0 Diameter towards increasing measurement values.

REFERENCES

  • 1
    Cueva-Goig R, Forner-Navarro L, Llena-Puy MC. Microscopic assessment of the sealing ability of three endodontic filling techniques. J Clin Exp Dent. 2016;8(1):27-31. doi: 10.4317/jced.52847
    » https://doi.org/10.4317/jced.52847
  • 2
    Moura AAM, Carvalho CF, Santos M, Davidowicz H. Avaliação in vitro através da análise radiográfica entre cones de guta-percha estandardizados e calibrados adaptados à região apical. Rev Inst Ciênc Saúde. 1994;12(1):7-11.
  • 3
    Davidowicz H, Moura AAM, Strefezza F. Avaliação do diâmetro "D0" de cones de guta-percha estandardizados e calibrados comparados com limas tipo "K" através do microscópio comparador. Rev ABO Nac. 1994;2(3):181-5.
  • 4
    Ferreira JJ, Rhodes JS, Pittford TR. The efficacy of gutta-percha removal using profiles. Int Endod J. 2001;(34):267-74. doi: 10.1046/j.1365-2591.2001.00379.x
    » https://doi.org/10.1046/j.1365-2591.2001.00379.x
  • 5
    Moule AJ, Kellaway R, Clarkson R, Rowell J, Macfarlane R, Lewis D, et al. Variability of master gutta-percha cones. Aust Endod J. 2002;28(1):38-43. doi: 10.1111/j.1747-4477.2002.tb00365.x
    » https://doi.org/10.1111/j.1747-4477.2002.tb00365.x
  • 6
    Borges AH, Dorileo MCO, Pedro FLM, Segundo AS, Volpato LER, Semenoff TADV, et al. Avaliação da padronização dos cones de guta-percha de diferentes conicidades. Rev Odontol Bras Central. 2011;20(55):313-6.
  • 7
    Mayne JR, Shapiro S, Abramson II. An evaluation of standardized gutta-percha points. Part I - Reliability and validity of standardization. Oral Surg Oral Med Oral Pathol. 1971;31(2):250-7. doi: 10.1016/0030-4220(71)90080-6
    » https://doi.org/10.1016/0030-4220(71)90080-6
  • 8
    Goldberg F, Gurfinkel J, Spielberg C. Microscopic study of standardized gutta-percha points. Oral Surg Oral Med Oral Pathol. 1979;47(3):275-6. doi: 10.1016/0030-4220(79)90153-1
    » https://doi.org/10.1016/0030-4220(79)90153-1
  • 9
    Hörsted-Bindslev P, Andersen MA, Jensen MF, Nilsson JH, Wenzel A. Quality of molar root canal fillings performed with the lateral compaction and the single-cone technique. J Endod. 2007;33(4):468-71. doi: 10.1016/j.joen.2006.12.016
    » https://doi.org/10.1016/j.joen.2006.12.016
  • 10
    Hegde V, Arora S. Sealing ability of three hydrophilic single-cone obturation systems: an in vitro glucose leakage study. Contemp Clin Dent. 2015;6(1):S86-9. doi: 10.4103/0976-237x.152953
    » https://doi.org/10.4103/0976-237x.152953
  • 11
    Valois CR, Silva LP, Azevedo RB, Costa Junior ED. Atomic force microscopy study of gutta-percha cone topography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2004;98(2):250-5. doi: 10.1016/s1079210404001854
    » https://doi.org/10.1016/s1079210404001854
  • 12
    Silva EJ, Rocha AC, Ferreira C, Herrera DR, Coutinho-Filho TS. Assessment of the tip surface of gutta-percha cones after different cutting methods. Acta Odontol Latinoam. 2011;24(3):236-9.
  • 13
    Celikten B, F Uzuntas C, I Orhan A, Tufenkci P, Misirli M, O Demiralp K, Orhan K. Micro-CT assessment of the sealing ability of three root canal filling techniques. J Oral Sci. 2015;57(4):361-6. doi: 10.2334/josnusd.57.361
    » https://doi.org/10.2334/josnusd.57.361
  • 14
    Lopes HP. Diâmetro e conicidades dos cones de guta-percha acessórios tipo M, FM e F. RBO. 1999;56(5):252-4.
  • 15
    Camões ICG, Reis PEG, Freitas LF, Gomes CC, Pinto SS. Avaliação do calibre apical de cones de guta-percha de três marcas comerciais. Pesq Bras Odontoped Clin Integr. 2006;6(2):111-6.
  • 16
    Bombana AC, Facchini MEB, Moura AAM. Avaliaçäo dimensional de cones de guta-percha estandardizados de diferentes procedências. Rev Inst Ciênc Saúde. 1989;7(1):5-14.
  • 17
    Figueiredo JAP. Assessment of the diameter of gutta percha cones of different manufacturers. Braz Endod J. 2000; 4(1): 23-8.
  • 18
    Cunningham KP, Walker MP, Kulild JC, Lask JT. Variability of the diameter and taper of size #30, 0.04 gutta-percha

Publication Dates

  • Publication in this collection
    Dec 2017

History

  • Received
    06 Feb 2017
  • Reviewed
    24 Apr 2017
  • Accepted
    31 July 2017
Faculdade São Leopoldo Mandic R. José Rocha Junqueira, 13, 13045-755 Campinas/SP Brasil, Tel.: (55 19) 3211-3689 - Campinas - SP - Brazil
E-mail: contato@revistargo.com.br