vol.47 issue1Method development and validation for naratriptan determination in human plasma by HPLC with tandem mass spectrometry detection, and its application to bioequivalence studyTherapeutic monoclonal antibodies: scFv patents as a marker of a new class of potential biopharmaceuticals author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

Share


Brazilian Journal of Pharmaceutical Sciences

On-line version ISSN 2175-9790

Braz. J. Pharm. Sci. vol.47 no.1 São Paulo Jan./Mar. 2011

http://dx.doi.org/10.1590/S1984-82502011000100004 

ARTICLE

 

Validation of a spectrophotometric method to determine ciprofibrate content in tablets

 

 

Guilherme Nobre Lima do NascimentoI; Daniel Lerner da RosaI; Hisao NishijoII; Tales Alexandre Aversi-FerreiraI, II, *

IBiochemistry and Neurosciences Laboratory, LABINE, Federal University of Goiás
IISystem Emotional Science - Graduate School of Medicine and Pharmaceutical Sciences, University of Toyama

 

 


ABSTRACT

Ciprofibrate is a drug indicated in cases of hypertriglyceridemia and mixed hyperlipidemia, but no monographs are available in official compendia for the analysis of this substance in tablets. The objective of this work was to develop and validate a spectrophotometric method for routine analysis of ciprofibrate in tablets. In this study, commercial and standard ciprofibrate were used, as well as placebo in absolute ethanol, analyzed by UV spectrophotometer. All tests followed the rules of Resolution RE-899, 2003. The results showed that the developed and validated method offers low cost, easy implementation, precision and accuracy, and may be included in the routine of quality control laboratories.

Uniterms: Ciprofibrate/analysis. Spectrophotometric method/medicines analysis. Medicines/quality control.


RESUMO

O ciprofibrato é um fármaco indicado em casos de hipertrigliceridemia e hiperlipidemia mista, mas não há monografias em compêndios oficiais para a análise desta substância em comprimidos. O objetivo deste trabalho é desenvolver e validar um método espectrofotométrico para análise de rotina de ciprofibrato em comprimidos. Neste estudo foram empregados ciprofibrato comercial, padrão e placebo em etanol absoluto, analisadas por espectrofotometria UV. Todos os testes seguiram as regras da Resolução RE- 899, 2003. Os resultados mostraram que o método desenvolvido e validado apresenta baixo custo, fácil implementação, precisão e exatidão e pode ser incluído em rotina de laboratórios de controle de qualidade.

Unitermos: Ciprofibrato/análise. Método espectrofotométrico/análise de medicamentos. Medicamentos/ controle de qualidade.


 

 

INTRODUCTION

Dyslipidemia is a chronic degenerative disease characterized by abnormal levels of lipids in the blood (Molly, Brunzel, 2004) which affects a large proportion of the population in different ways according to ethnicity, socioeconomic status and other factors (Bertolami et al., 1993; Eizerik, Manfroi, 2008; Souza et al., 2003).

In 1998, it was found that 38% of men and 42% of women in the Brazilian population had total cholesterol levels above 200 mg/dL (DBD, 2007), and a high incidence was also found in other countries including India and the United States of America (Davidson, Yannicelli, 2004). Dyslipidemia is one of the factors predisposing patients to a number of diseases such as atherosclerosis (Luo et al., 2009) and cardiovascular problems (Lotufo, 1996).

Fibrates are a group of drugs derived from fibric acid (Xavier, 2005), referred to as drugs of choice for cases of hypertriglyceridemia as an adjuvant of statins (Schultz, 2006) and for conditions including mixed dyslipidemia in patients with type-2 diabetes mellitus (DBD, 2007; Xavier, 2005), especially for patients resistant or intolerant to statins (Abourbih, 2009).

Ciprofibrate is chemically known as 2-[4-(2,2-dichlorocyclopropyl) phenoxy]-2-methylpropanoic acid (Figure 1), having a molecular formula C13H14Cl2O3 and molecular weight of 289.2 (Merck Index, 2001). It is a white crystalline or lightly yellowish powder, with a melting point of about 115.0 °C that is practically insoluble in water, freely soluble in anhydrous ethanol, and soluble in toluene (BP, 2008).

 

 

In Brazil, ciprofibrate is marketed as Oroxadin (Sanifi-Aventis), Lipless (Biolab Sanus) and Ciprolip (Uci-Farma), all containing 100 mg per tablet in their presentation for oral administration.

According to Xavier (2005), fibrates are absorbed through the gastrointestinal tract, and glucuronidated compounds by the liver, and are excreted by the kidneys, peaking at between 2 and 8 h. Fibrates have an elimination half-life of between 2 and 80 h depending on the doses used, and act through the stimulation of nuclear receptors activated by the proliferation of alpha-peroxisomes, leading to increased production and action of lipoprotein lipase, reduction of apolipoprotein CIII; decrease in the synthesis of VLDL and increase in HDL-c levels.

The concern about the quality of drugs sold increases every year (La Roca et al., 2007), not only in commercial terms, but also legal and ethical aspects, since the health of patients depends on the quality and effectiveness of these drugs (Linsbinski et al., 2008; Ribani et al., 2004). In this regard, various regulatory agencies around the world are demanding validation methods for the registration of new drugs to ensure the quality of the drugs marketed (Brasil, 2003; Grillo et al., 2009; ICH, 1996; Santana et al., 2007; Valentini et al., 2004).

There is a great interest in developing rapid and efficient analytical methods that provide precise and accurate parameters for the quantitative analysis of drugs, important for routine analysis during quality control and development of new drug forms (Ruela et al., 2009), with spectrophotometric analysis being the most likely candidate.

To date, no validated methods are available in official compendia for the analysis of ciprofibrate tablets. This fact, together with the importance of this drug in the treatment of a chronic disease that affects millions of people around the world, justified the conducting of a validation test for this drug.

The objective of this study was to develop and validate a spectrophotometric method for routine analysis of quality control of ciprofibrate tablets, to be used by industry and pharmacies.

 

MATERIAL AND METHODS

Equipment and reagents

The following devices were used in this study: UV / Vis spectrophotometer label Agilent model G1103A, adjusted for detection at 233nm, using a quartz cuvette with an optical path of 1.0 cm, and UV/HPLC grade methanol, labels JTBaker and Tedia.

Samples

Commercial Ciprofibrate samples of 100mg/cp and its CRS ciprofibrate placebo (Chemical Reference Substances), previously standardized with the power of 99.8%, were used.

Preparation of solutions

Standard Solution

To reach the adopted working concentration of 0.01 mg/mL, the equivalent of 50.0 mg of CRS ciprofibrate was weighed and transferred to a volumetric flask of 50.0 mL, dissolved, and this volume completed with UV/HPLC grade methanol. From this solution, 1.0 mL was transferred to a volumetric flask of 100.0 mL, completed to volume with UV/HPLC grade methanol, and then homogenized.

Sample Solution

The average weight of 20 tablets of commercial sample was determined for the ciprofibrate solution sample. Once sprayed, the equivalent of 100.0 mg of ciprofibrate (equivalent to an average weight) was weighed and transferred to volumetric flask of 100.0 mL, dissolved, and the volume completed with methanol. The solution was filtered through filter paper for the retention of insoluble particles, and 1.0 mL was then transferred to a volumetric flask of 100.0 mL, filled with methanol, and then homogenized, in order to obtain the theoretical working concentration of 0.01 mg/mL.

Method validation

The validation process was performed according to the present guidelines contained in Resolution RE-899 of 29 May 2003 (Brasil, 2003).

Specificity/selectivity

Commercial ciprofibrate sample solution, standard and placebo sample at the working concentration (0.01 mg/mL) were determined through analysis, using UV/HPLC grade methanol as the diluent.

Linearity

A standard calibration curve was constructed by reading, in duplicate, 5 standard solutions at concentrations corresponding to 60%, 80%, 100%, 120% and 140% compared to the theoretical value of the working concentration adopted, with values 0.006 mg/mL, 0.008 mg/mL, 0.01 mg/mL, 0.012 mg/mL and 0.014 mg/mL, to verify the correspondence between absorbance and the concentrations of solutions, expressed by the correlation coefficient obtained by the least squares method.

Precision

Precision was evaluated through the repeatability and intermediate precision methods. Repeatability was determined by preparation, under the same conditions, of 6 analytical solutions of the sample at working concentration, by two different analysts. Intermediate accuracy was assessed in the same manner, on different days by different analysts. Relative standard deviation (RSD) of up to 5% was considered as acceptable, according to resolution 899/2003 (Brasil, 2003).

Accuracy

Accuracy was assessed based on the degree of recovery, as percentage of placebo solutions fortified with known amounts of CRS ciprofibrate, in triplicate, at concentrations of 80%, 100% and 120% of the working concentration.

Robustness

The variation among solvent manufacturers was evaluated with sample and standard solutions at working concentration (0.01 mg/mL), with two different solvent brands (A and B), concomitant with the stability time of sample and standard solutions, by making periodic readings at 0, 2, 4, 6, 8, 10, 12, 15, 20 and 24 h of standard and sample solutions at working concentration (0.01 mg/mL).

 

RESULTS

Specificity/selectivity

The method used was specific and selective for ciprofibrate through comparative analysis of placebo, standard and samples, prepared under the same conditions. There was no significant interference from other components (placebo) at a wavelength of 233nm (Figure 2).

 


 

Linearity

The correlation coefficient (r) of the calibration curve was 0.99, according to RE 899/2003 (Brasil, 2003) minimum acceptable criterion (Figure 3).

 

 

Precision

Considering data obtained from the six analytical solutions prepared, the method demonstrated repeatability with a relative standard deviation (RSD%) of up to 2.09% among samples prepared on the same day under the same conditions, within the maximum permitted by applicable law of 5% RE 899/2003 (Brasil, 2003). The method also showed intermediate precision with RSD% of 1.86% among results of six analytical solutions prepared by two different analysts on different days.

Accuracy

The results for accuracy, at concentrations of 80%, 100% and 120%, showed a mean recovery percentage of between 99.47% and 100.84% (Table I), compared to the ideal recovery of 98.0 to 102.0%.

Robustness

For the same sample, both standard solution and ciprofibrate sample solutions analyzed using different solvent brands (methanol), A and B, revealed significant variations in the robustness test results (Table II and Table III).

 

 

 

 

The stability of analytical standard and sample solutions after 2, 4, 6, 8, 10, 12, 15, 20 and 24 hours from the initial analysis was also evaluated. The ciprofibrate sample and standard solutions seemed to be stable for 24 hours after the initial analysis, and a maximum variation of 1.78% in 6 hours was found (Table IV and Table V).

 

 

 

 

Statics analysis of linearity

The correlation coefficient (r) for linearity was according to RE 899/2003 (Brasil, 2003), established at 0.99 as a minimum criterion. Therefore, statistical analysis of linearity can be done to verify the robustness of this figure (Figure 4, Table VI). From the statistical data seen in Table VI, it is possible to affirm that these data present strong linearity (Figure 4, Table VI).

 

 

DISCUSSION

Methods to determine the content of drugs should never fail because, according to Honda and Magalhães (2001), possible failures may lead to erroneous decisions subsequently being taken based on these results.

The selectivity and specific features of the proposed method is the primary principle of a high-accuracy procedure, as demonstrated for the method in this study, since no significant interference was observed from other components (placebo) at a wavelength of 233 nm (Figure 2).

In Figure 2a, for λ = 233 nm, the absorbance is approximately 0.45 AU, with coordinates located at the peak of a typical curve obtained in spectroscopy, while in Figure 2b, for λ = 233 nm, the absorbance is less than 0.01, with coordinates located on a decreasing curve after the methanol absorbance peak. Based on these results, it was confirmed that the method to determine ciprofibrate content in tablets via spectrophotometer is both specific and selective.

The linearity of a method must be obtained by the correlation coefficient (r) of the calibration curve. According to RE 899/2003 (Brasil, 2003), the minimum acceptable criterion is 0.99,a level verified by statistical analysis (Figure 4, Table VI). According to Pimentel and Barros Neto (1996), the least squares method is widely used to verify the proximity of points obtained in relation to the standard line, since it provides unbiased results with minimum variance.

Figure 3 shows the value of R = 0.99978 for the equation y = 45.62x + 0006 in the calibration curve of the CRS ciprofibrate through absorption spectrophotometry in the UV region at 233nm. According to Braga and Poppi (2004) and to findings of the present analysis, it can be concluded that the linear model efficiently fits the data obtained.

The degree of accuracy of a method is demonstrated by the correlation among measurements made for the same homogeneous sample under repeatability conditions, prepared on the same day under the same conditions, and by different analysts (Braga, Poppi, 2004).

For the spectrophotometric determination method used in this work, the repeatability assessment had a maximum standard deviation of 2.09% for samples (6) prepared on the same day. This value lies within the limits allowed under current legislation (Brasil, 2003), stipulating that this value should not exceed 5%, while intermediate precision evaluation should have a maximum standard deviation of 1.86% for samples (6) analyzed by different analysts. Therefore, the method to determine ciprofibrate content is highly accurate under the conditions of this experiment.

Moreover, a method must be exact. The RE 899/2003 (Brasil, 2003) stipulates that the ideal recovery degree for concentrations of 80%, 100% and 120% should range from a minimum of 98.0% to a maximum of 102.0%. The data in Table I show a recovery degree for these concentrations of between a minimum of 99.47% and maximum of 100.84% (Table I). In statistical terms, the data obtained in the present study are consistent and within the limits required by law.

In statistical terms, the least squares estimator is the minimum linear variance of the response variable (Pimentel, Barros Neto, 1996), and therefore the values obtained for the detection and quantification method were within the parameters of safety for the linearity of data.

The results obtained by quantification of the analytical solution of the same sample, analyzed with different brands (methanol), A and B, revealed significant variations in the results, indicating high and satisfactory robustness of the method for small variations (Table II and Table III).

Regarding the stability of standard and sample analytical solutions, a maximum variation of 1.78% was detected in 6 hours, thus not exceeding the maximum acceptable limit adopted of 2% according to RE 899/2003 (Brasil, 2003) (Table IV and Table 5).

 

CONCLUSIONS

The method to determine ciprofibrate content in tablets through spectrophotometry in the ultraviolet region was validated according to resolution 899/2003. The results showed the method to be simple, inexpensive, easy to apply, precise and accurate for the determination of ciprofibrate. Moreover, the method provides the reliability required for an analytical method, and also the practicality required for the routine of quality control laboratories.

 

REFERENCES

ABOURBIH, S.; FILION, K.B.; JOSEPH, L.; SCHIFFRIN, E.L.; RINFRET, S.; POIRIER, P.; PILOTE, L.; GENEST, J.J.; EISENBERG, M.J. Effect of fibrates on lipid profiles and cardiovascular outcomes: a systematic review. Am. J. Med., v.122, n.10, p.962e1-962e8, 2009.         [ Links ]

BRASIL, Agência Nacional de Vigilância Sanitária (ANVISA). R. E, nº 899 de 29 de maio de 2003 – Guia para validação de métodos qualitativos e bioanalíticos. Available at: <http://e-legis.anvisa.gov.br/leisref/public/showAct.php?mode=PRINT_VERSION&id=15132>. Accessed on: 05 dez. 2008.         [ Links ]

BERTOLAMI, M.C.; FALUDI, A.A.; LATORRE, M.R.D.O.; ZAIDAM FILHO; T. Perfil lipídico de funcionários de indústria metalúrgica e sua relação com outros fatores de risco. Arq. Bras. Cardiol., v.60, n.5, p.293-299, 1993.         [ Links ]

BRAGA, J.W.B.; POPPI, R.J. Validation of models of multivariate calibration: an application in the determination of polymorphic purity of carbamazepine by near infrared spectroscopy. Quim. Nova, v.27, n.6, p.1004-1011, 2004.         [ Links ]

BRITISH PHARMACOPOEIA. Ciprofibrate, 2008. Available at: <http://www.pharmacopoeia.co.uk/>. Accessed on: 05 dez. 2008.         [ Links ]

DAVIDSON, M.H.; YANNICELLI, D. New concepts in dyslipidemia in the metabolic syndrome and diabetes. Metab. Syndr. Relat. Disord., v.4, p.299-314, 2004.         [ Links ]

DIRETRIZES brasileiras sobre dislipidemias e diretriz de prevenção da aterosclerose do departamento de aterosclerose da sociedade brasileira de cardiologia. DBD. IV Arq. Bras. Cardiol., v.88, supl.1, p. 02-19, 2007.         [ Links ]

EIZERIK, D.P.; MANFROI, W.C. Eficácia da atenção farmacêutica em dislipidemia: revisão sistemática. Rev. HCPA, v.28, n.1, p.37-40, 2008.         [ Links ]

GRILLO, R.; MELO, N.F.S.; ARAÚJO, D.R.; PAULA, E.; DIAS FILHO, N.L.; ROSA, A.H.; FRACETO, L.F. Validation of an HPLC method for quantitative determination of Benzocaine in PHBV-microparticles and PLA-nanoparticles. Lat. Am. J. Pharm., v.28, n.3, p.393-399, 2009.         [ Links ]

HONDA, A.M.; MAGALHÃES, J.F. Validação do método espectrofotométrico da pentoxifilina em comprimidos revestidos de 400mg. Rev. Bras. Farm., v.37, n.2, p.165- 170, 2001.         [ Links ]

INTERNATIONAL CONFERENCE ON HARMONIZATION. ICH. Validation of Analytical Procedures: Text and Methodology Q2(R1). 1996. Available at: <http://www.ich.org/LOB/media/MEDIA417.pdf>. Accessed on: 05 dez. 2008.         [ Links ]

LA ROCA, M.F.; SOARES SOBRINHO, J.L.; NUNES, L.C.C.; ROLIM NETO, P.J. Desenvolvimento e validação de método analítico: passo importante na produção de medicamentos. Rev. Bras. Farm., v.88, n.4, p.177-180, 2007.         [ Links ]

LINSBINSKI, L.M.; MUSIS, C.R.; MACHADO, S.R.P. Avaliação da equivalência farmacêutica de comprimidos de captopril. Rev. Bras. Farm., v.89, n.3, p.214-219, 2008.         [ Links ]

LOTUFO, P.A. Epidemiologia das doenças cardíacas no Brasil: histórico, situação atual e proposta de modelo teórico. Rev. Soc. Cardiol., v.6, n.5, p.541-547, 1996.         [ Links ]

LUO, Y.; CHEN, G.; LI, B.; JI, B.; XIAO, Z.; YI, G.; TIAN, F. Dietary intervention with AHP, a functional formula diet, improves both serum and hepatic lipids profile in dyslipidemia mice. J. Food Sci., v.74, n.6, p.189-195, 2009.         [ Links ]

MERCK INDEX. Ciprofibrate. 13.ed. New Jersey: Merck & CO. Inc., 2001. p.402.         [ Links ]

MOLLY, C.C.; BRUNZEL J.D. Abdominal obesity and dyslipidemia in the metabolic syndrome: importance of type 2 diabetes and familial combined hyperlipidemia in coronary artery disease risk. J. Clin. Endocrinol. Metab., v.89, n.6, p.2601-2607, 2004.         [ Links ]

PIMENTEL, M.F.; BARROS NETO, B. Calibração: uma revisão para químicos analíticos. Quim. Nova, v.19, n.3, p.268-277, 1996.         [ Links ]

RIBANI, M.; BOTTOLI, C.B.G.; COLLINS, C.H.; JARDIM, I.C.S.F.; MELO, L.F.C. Validação em métodos cromatográficos e eletroforéticos. Quim. Nova, v.27, n.5, p.771-780, 2004.         [ Links ]

RUELA, A.L.M.; ARAÚJO, M.B.; PEREIRA, G.R. Desenvolvimento e validação de um método analítico rápido por cromatografia líquida de alta eficiência para determinação de nimesulida em estudos de liberação in vitro. Quim. Nova, v.32, n.1, p.165-168, 2009.         [ Links ]

SANTANA, A.K.M.; NUNES, L.C.C.; MEDEIROS, F.P.M.; SILVA, M.J.; LAVRA, Z.M.M.; ROLIM-NETO, P.J. Otimização e validação do método analítico volumétrico para quantificação do carbonato de cálcio. Rev. Ciênc. Farm. Básica Apl., v.28, n.2, p.177-183, 2007.         [ Links ]

SCHULTZ, I. Tratamento das dislipidemias – como e quando indicar a combinação de medicamentos hipolipemiantes. Arq. Bras. Endocrinol. Metab., v.50, n.2, p.344-359, 2006.         [ Links ]

SOUZA, L.J.; FILHO, J.T.D.S.; SOUZA, T.F.; REIS, A.F.F.; NETO, C.G.; BASTOS, D.A.; CÔRTES, V.A.; CHALITA, F.E.B.; TEIXEIRA, C.L. Prevalência de dislipidemia e fatores de risco em Capos dos Goytacazes – RJ. Arq. Bras. Cardiol., v.81, n.3, p.249-256, 2003.         [ Links ]

SWARTZ, M.E.; KRULL, I.S. Validação de métodos cromatográficos. Pharm. Technol., v.2, n.3, p.12-20, 1998.         [ Links ]

VALENTINI, S.R.; SOMMER, W.A.; MATIOLI, G. Validação de métodos analíticos na quantificação de comprimidos de Captopril - comparação de metodologias para um programa de garantia de qualidade. Acta. Sci. Health. Sci., v.26, n.2, p.357-364, 2004.         [ Links ]

XAVIER, H.T. Farmacologia do fibratos. Arq. Bras. Cardiol., v.85, supl.5, p.15-16, 2005.         [ Links ]

 

 

Received for publication on 19th February 2010
Accepted for publication 27th October 2010

 

 