Fracture strength of orthodontic mini-implants

ABSTRACT

Objective:

This study aimed at evaluating the design and dimensions of five different brands of orthodontic mini-implants, as well as their influence on torsional fracture strength.

Methods:

Fifty mini-implants were divided into five groups corresponding to different manufactures (DEN, RMO, CON, NEO, SIN). Twenty-five mini-implants were subjected to fracture test by torsion in the neck and the tip, through arbors attached to a Universal Mechanical Testing Machine. The other 25 mini-implants were subjected to insertion torque test into blocks of pork ribs using a torquimeter and contra-angle handpiece mounted in a surgical motor. The shape of the active tip of the mini-implants was evaluated under microscopy. The non-parametric Friedman test and Snedecor’s F in analysis of variance (ANOVA) were used to evaluate the differences between groups.

Results:

The fracture torque of the neck ranged from 23.45 N.cm (DEN) to 34.82 N.cm (SIN), and of the tip ranged from 9.35 N.cm (CON) to 24.36 N.cm (NEO). Insertion torque values ranged from 6.6 N.cm (RMO) to 10.2 N.cm (NEO). The characteristics that most influenced the results were outer diameter, inner diameter, the ratio between internal and external diameters, and the existence of milling in the apical region of the mini-implant.

Conclusions:

The fracture torques were different for both the neck and the tip of the five types evaluated. NEO and SIN mini-implants showed the highest resistance to fracture of the neck and tip. The fracture torques of both tip and neck were higher than the torque required to insert mini-implants.

Keywords:
Orthodontics; Skeletal anchorage; TADs; Mini-implant; Dental materials.

RESUMO

Objetivo:

o objetivo do presente estudo foi avaliar o desenho e as medidas de cinco marcas diferentes de mini-implantes ortodônticos, e sua influência na resistência à fratura em torção.

Métodos:

cinquenta mini-implantes foram divididos em cinco grupos, correspondentes a diferentes fabricantes (DEN, Dentaurum; RMO, Rocky Mountain Orthodontics; CON, Conexão; NEO, Neodent; SIN, Sistema de Implantes Nacional). Vinte e cinco mini-implantes foram submetidos ao teste de fratura por torção no pescoço e na ponta, com mandris fixados a uma máquina universal de testes mecânicos. Os outros 25 mini-implantes foram submetidos ao teste de torque de inserção em blocos de costelas suínas, utilizando-se um torquímetro e um contra-ângulo montado em motor cirúrgico. O formato da ponta ativa dos mini-implantes foi avaliado por microscopia. O teste não-paramétrico de Friedman e o teste F de Snedecor na análise de variância (ANOVA) foram utilizados para avaliar as diferenças entre os grupos.

Resultados:

o torque de fratura do pescoço variou de 23,45 N.cm (DEN) a 34,82 N.cm (SIN); e o da ponta, entre 9,35 N.cm (CON) e 24,36 N.cm (NEO). O valor do torque de inserção variou de 6,6 N.cm (RMO) a 10,2 N.cm (NEO). As características que mais influenciaram os resultados foram: diâmetro externo, diâmetro interno, razão entre o diâmetro interno e o externo, e a presença de fresagem na região apical do mini-implante.

Conclusões:

os torques de fratura foram diferentes entre os cinco tipos avaliados, tanto no pescoço quanto na ponta. Os mini-implantes NEO e SIN foram os mais resistentes à fratura do pescoço e da ponta. Os torques de fratura tanto na ponta quanto no pescoço foram maiores do que o torque necessário para a inserção dos mini-implantes.

Palavras-chave:
Ortodontia; Ancoragem óssea; Mini-implante; Materiais dentários.

INTRODUCTION

Mini-implants have been an effective anchorage method,11 Brandão LBC, Mucha JN. Grau de aceitação de mini-implantes por pacientes em tratamento ortodôntico - estudo preliminar. Rev Dental Press Ortod Ortop Facial. 2008 Set-Out;13(5):118-27. being very well tolerated by patients,22 Elias CN, Guimarães GS, Muller CA. Torque de inserção e de remoção de miniparafusos ortodônticos. Rev Bras Implant. 2005;11(3):5-8. creating new possibilities for orthodontic treatment that requires minimal cooperation and maximum aesthetics results, particularly in adults.33 Mah J, Bergstrand F. Temporary anchorage devices: a status report. J Clin Orthod. 2005 Mar;39(3):132-6. Unlike osseointegrated dental implants made of commercially pure titanium, mini-implants are manufactured with the alloy Ti6Al4V (ASTM grade 5).44 Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56.,55 Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44. This alloy has higher mechanical strength than pure titanium66 Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006 May;67(3):162-74. and it is best suited to the small diameter of mini-implants. It also presents bioactive characteristics inferior to pure titanium, thus facilitating removal because it promotes lower osseointegration.44 Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56.,55 Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44.

The mini-implant installation is simple, and it can be inserted in various sites because of its reduced size,11 Brandão LBC, Mucha JN. Grau de aceitação de mini-implantes por pacientes em tratamento ortodôntico - estudo preliminar. Rev Dental Press Ortod Ortop Facial. 2008 Set-Out;13(5):118-27.,77 Kuroda S, Sugawara Y, Tamamura N, Takano-Yamamoto T. Anterior open bite with temporomandibular disorder treated with titanium screw anchorage: evaluation of morphological and functional improvement. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):550-60.,88 Morarend C, Qian F, Marshall SD, Southard KA, Grosland NM, Morgan TA, et al. Effect of screw diameter on orthodontic skeletal anchorage. Am J Orthod Dentofacial Orthop. 2009 Aug;136(2):224-9. including those between the roots of teeth using mono or bicortical anchor,99 Iijima M, Muguruma T, Brantley WA, Okayama M, Yuasa T, Mizoguchi I. Torsional properties and microstructures of miniscrew implants. Am J Orthod Dentofacial Orthop. 2008 Sept;134(3):333.e1-6. alone or connected by an interchangeable structure.1010 Itsuki Y, Imamura E, Sugawara J. Temporary anchorage device with interchangeable superstructure for mandibular tooth movement. J World Fed Orthod. 2013;2(1):e19-29. On the other hand, the small size of mini-implants empowers an increase of fracture during insertion, deformation or fracture during its removal.44 Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56. Bicortical anchorage adds a new challenge to the fracture resistance of orthodontic mini-implants especially in the tip region, which has a smaller diameter. Another risk of fracture could occur when mini-implants are placed in the external oblique ridge area of the mandible because of the thickness and hardness of the cortical bone in this area.1010 Itsuki Y, Imamura E, Sugawara J. Temporary anchorage device with interchangeable superstructure for mandibular tooth movement. J World Fed Orthod. 2013;2(1):e19-29.

Mini-implants can be found in the market with different shapes, designs, diameters, lengths, degree of titanium alloy purity and surface treatments.55 Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44. However, mini-implants with similar dimensions have different design that may influence the fracture resistance and are relied upon by the various manufacturers to improve clinical performance.66 Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006 May;67(3):162-74.,1111 Kitahara-Ceia FMF, Assad-Loss TF, Mucha JN, Elias CN. Morphological evaluation of five active types of six types of orthodontic mini-implants. Dental Press J Orthod. 2013 Mar-Apr;18(2):36-41. The fracture resistance of the mini-implants is a decisive aspect to assist the clinician in choosing the most appropriate and safe device.

Therefore, the main objective of this study was to evaluate the design, to measure the insertion torque and to quantify the maximum fracture torque resistance of five different orthodontic mini-implants.

MATERIAL AND METHODS

Fifty orthodontic self-perforating mini-implants from five different manufacturers (DEN: Dentaurum, Ispringen, Baden-Württemberg, Germany; RMO: Rocky Mountain Orthodontics, Seoul, South Korea; CON: Conexão, Arujá, São Paulo, Brazil; NEO: Neodent, Curitiba, Paraná, Brazil; and SIN: Sistema de Implantes Nacional, São Paulo, São Paulo, Brazil) similar in size dimension, were divided into five groups, as shown in Tables 1 and 2.

Table 1
Groups code, milling in the apical region, trade name, manufacturer, source, chemical composition, nominal diameter, length and transmucosal profile of mini-implants used in the study.

Table 2
Values of linear measurements (µm) and angle (degrees) of mini-implants.

The surface morphology and chemical composition of the mini-implants in each group was analyzed by scanning electron microscope (SEM; Jeol JSM-5800LV) at 20kV and observed at 500x magnification and 50 µm away, by energy dispersive spectroscopy (EDS). Figure 1 illustrates SEM microphotographs of the mini-implants for each group.

Figure 1
Microphotograph illustrating the superficial aspects and superficial finish quality of five types of mini-implants used (DEN, RMO, CON, NEO, SIN).

Mini-implants design evaluation

Pictures of the mini-implants were obtained under optical microscope (Zeiss Stemi 2000-C, Zeiss) observed at 1.6 x magnification. The pictures of the surfaces were analyzed with Axio Vision program (Zeiss) which were made of linear and angular measurements, as shown in Figures 2 and 3.

Figure 2
Linear measurements made in pictures of the mini-implants in optical microscope with an increase of 1.6 x. Pitch of the threads (a), total length or active tip of the mini-implant (b), the external diameter of mini-implant (c), steps for calculating the taper of the mini-implant (d), and internal diameter of the mini-implant (d’).

Figure 3
Linear and angular measurements made in pictures of the mini-implants in optical microscope with an magnification of 1.6 x. Angle of the screw thread (e), the free length of thread (f), and the length of the bottom fillet of the thread (g) of the screw thread pitch (h).

Mini-implant insertion torque

Six frozen pig ribs, from the same animal, were prepared to be used as material for insertion of the mini-implants. After slaughter, the ribs were cleaned to remove tissue and stored in normal saline at 4°C for 24 hours. Then, they were cut transversely to obtain 60 bone blocks of 2 x 2 cm and stored again in physiological saline at 4°C until assay insertion occurred between the first and the third day.

Each pork rib bone block was randomly picked among the 60 bone blocks and was attached to an adjustable piece of metal, for size and shape standardization. This piece of metal was attached to the digital torque meter (Lutron TQ-8800 torque meter) connected to a computer and pre set to prevent any movement during the test.

Tests on insertion of the mini-implant were carried out without performing any preliminary drilling, using the mini-implant coupled to the switch for short contra angle for each specific group of mini-implants, fitted to a hand piece with reduced speed (20:1) at 50 rpm (Anthogyr Instruments) with MC -101 surgical drill, and Omega II Dentscler motor. Each mini-implant was inserted in the block located in the central portion of pork ribs perpendicular to the cortical bone (Fig 4).

Figure 4
Mini-implant position in pork rib (A) and cortical bone thickness of pork rib block used in the insertion test of mini-implants (B).

The same calibrated operator performed the whole assay. Initial tests were undertaken to verify the procedure and to maintain the strength standardized in relation to the long axis of the mini-implant. The insertion torques were recorded continuously by the Lutron 101 program, version V0011TW (Lutron Electronic Enterprise).

Maximum and minimum insertion torque values were considered as the largest and the smallest values, respectively. Each bone block, properly identified, was taken to the Optical Zeiss Stemi 2000-C Microscope (Zeiss) with 1.6x and 5x magnification for measurement of cortical thickness of the bone at the site of insertion of the mini-implant. The images were captured in the cortical computer and sent to the Axio Vision program (Zeiss) where their thicknesses were measured.

MINI-IMPLANT FRACTURE TORQUE

Five mini-implants in each group (DEN: Dentaurum, RMO: Rocky Mountain Orthodontics, CON: Conexão, NEO: Neodent; SIN: Sistema de Implantes Nacional) were submitted to fracture torsion testing. The maximum torque fracture was determined in the lower third of the active tip (tip) and the upper third of the active tip (neck). For this test, a device with two mandrels coupled to universal mechanical testing machine (EMIC) with a 500N load cell was used.

In the fracture test, the key for short contra-angle surgical kit, specific for each group coupled to the head of the mini-implant, was attached to the mandrel, and the other mandrel was attached to the other end of the mini-implant. The left mandrel rotated by traction of a polymer strand attached to the shaft of the testing device and the load cell, thereby applying a torque on the mini-implant, once the right mandrel was fixed.

In the first mechanical testing, the fracture torque of the mini-implant tip was measured. In the second fracture testing, the maximum fracture torque of the region close to the neck was determined.

In order to calculate the fracture torque (N.cm) the maximum force was obtained from records of the Universal Mechanical Testing Machine (EMIC), based on the following formula: T = F x 0.4, where T = Torque (N.cm), F = Force (N), and 0.4 cm is the radius of the shaft on which the polymer strand was wound.

STATISTICAL ANALYSIS

The numerical results were presented as means and standard deviations. The non-parametric Friedman test was used to assess torque values of fracture of the tip and the neck of the mini-implant. Snedecor’s F in analysis of variance (ANOVA) was used to evaluate the insertion torque and the thickness of cortical bone. The data obtained from all measurements were processed with SPSS software, version 13.0 (IBM, Chicago, Illinois, USA). The level of significance for all statistical tests was predetermined at 1%.

RESULTS

Table 2 presents the results of mini-implants design and the measures corresponding to the length of tip, outer diameter, inner diameter, and the number of threads (Figs 2 and 3).The bone material used for the test was the pork rib, which has been used in other studies, as the thickness of cortical bone is similar to that found in the human jaw, around 0.5 to 1.0 mm.1212 Kim JH, Ahn SJ, Chang YI. Histomorfometric and mechanical analyses of the drill free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2005 Aug;128(2):190-4. In this study the cortical thickness of the pork rib blocks showed a thickness average from 1.30 to 1.52 mm and no significant difference among groups was determined (Table 3). Therefore, it can be assumed that the bone material used to insert the mini-implants did not influence the results. The average and standard deviations of the torque during insertion in the five groups and the average values and standard deviations for the thickness of the cortical bone blocks used as material for insertion were calculated (Table 3).

Table 3
Means and standard deviations (SD) of the insertion torque (N.cm) and cortical bone thickness (mm), Snedecor’s F test in ANOVA.

The torque values of fracture on the neck region were higher than the torque values to obtain a fracture in the region of the tip in all groups (Table 4).

Table 4
Means and standard deviations (SD) of fracture torque (N.cm) in the neck and tip of the five groups, tested with Snedecor’s F in ANOVA and XR22 Elias CN, Guimarães GS, Muller CA. Torque de inserção e de remoção de miniparafusos ortodônticos. Rev Bras Implant. 2005;11(3):5-8. non-parametric “Friedman” for comparison between groups (p < 0.01).

DISCUSSION

The mini-implants diameter showed a variation from 1,630.72 µm (NEO) to 1,482.95 µm (CON). The CON group, besides having the smallest diameter, also showed a lower torque value of fracture of the tip (9.35 N.cm) and the second lowest value of fracture torque on the neck (25.7 N.cm). A similar result was observed in mini-implants with a diameter less than 1.5 mm, which were more susceptible to fracture.1313 Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8.,1414 Nova MFP, Carvalho FR, Elias CN, Artese F. Avaliação do torque de inserção, remoção e fratura de diferentes mini-implantes ortodônticos. Rev Dent Press Ortod Ortop Facial. 2008 Set-Oct;13(5):76-87.,1515 Hong SB, Kusnoto B, Kim EJ, BeGole EA, Hwang HS, Lim HJ. Prognostic factors associated with the success rates of posterior orthodontic miniscrew implants: a subgroup meta-analysis. Korean J Orthod. 2016 Mar;46(2):111-26.

The torque values of fracture on the neck and the tip region could indicate that the diameter of the fracture site is an important variable in fracture torque, because the small-diameter mini-implant is an important risk factor for fractures, especially in insertion or removal time.11 Brandão LBC, Mucha JN. Grau de aceitação de mini-implantes por pacientes em tratamento ortodôntico - estudo preliminar. Rev Dental Press Ortod Ortop Facial. 2008 Set-Out;13(5):118-27.,44 Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56.,88 Morarend C, Qian F, Marshall SD, Southard KA, Grosland NM, Morgan TA, et al. Effect of screw diameter on orthodontic skeletal anchorage. Am J Orthod Dentofacial Orthop. 2009 Aug;136(2):224-9.,1414 Nova MFP, Carvalho FR, Elias CN, Artese F. Avaliação do torque de inserção, remoção e fratura de diferentes mini-implantes ortodônticos. Rev Dent Press Ortod Ortop Facial. 2008 Set-Oct;13(5):76-87.,1515 Hong SB, Kusnoto B, Kim EJ, BeGole EA, Hwang HS, Lim HJ. Prognostic factors associated with the success rates of posterior orthodontic miniscrew implants: a subgroup meta-analysis. Korean J Orthod. 2016 Mar;46(2):111-26.,1616 Melsen B, Verna C. Miniscrew implants: the Aarhus anchorage system. Semin Orthod. 2005 Mar;11(1):24-31.,1717 Barros SE, Janson G, Chiqueto K, Garib DG, Janson M. Effect of mini-implant diameter on fracture risk and self-drilling efficacy. Am J Orthod Dentofacial Orthop. 2011 Oct;140(4):e181-92. Although mini-implants with a reduced diameter have increased risk of fracture, the choice of mini-implants with a diameter that is too large can lead to bone necrosis through micro fractures in bone, and dental structures risks.1515 Hong SB, Kusnoto B, Kim EJ, BeGole EA, Hwang HS, Lim HJ. Prognostic factors associated with the success rates of posterior orthodontic miniscrew implants: a subgroup meta-analysis. Korean J Orthod. 2016 Mar;46(2):111-26.,1616 Melsen B, Verna C. Miniscrew implants: the Aarhus anchorage system. Semin Orthod. 2005 Mar;11(1):24-31.

In relation to the tip region, the lowest torque value of fracture was found in the CON group (9.35 N.cm), which was statistically significant (p <0.01) compared with the NEO (24.36 N.cm), SIN (22.88 N.cm), and RMO (21.95 N.cm) groups. These differences may be a consequence of the great difference among them in shape, in ratio of external to internal diameter and in the number of threads, and consequently, in the step of threads.1818 Smith A, Hosein YK, Dunning CE, Tassi A. Fracture resistance of commonly used self-drilling orthodontic mini-implants. Angle Orthod. 2015 Jan;85(1):26-32.

CON group presented the second lowest value of fracture torque in the neck (25.7 N.cm), and it is statistically significant (p < 0.01) when compared with SIN (34.82 N.cm), which showed highest value of fracture torque. Additionally, when the internal diameter of mini-implant between these two groups was compared, great variation was found. The CON group exhibited the lowest mean (772.75 µm), while the SIN group presented the highest (1164.77 µm). The inner diameter of the mini-implant is an important characteristic of susceptibility to fracture.1919 Squeff LR, Simonson MBA, Elias CN, Nojima LI. Caracterização de miniimplantes utilizados na ancoragem ortodôntica. Rev Dent Press Ortod Ortop Facial. 2008 Set-Out;13(5):49-56.

Furthermore, the mini-implants in the CON group also showed the smallest external diameter (1482.95 µm), and a greater number of threads.77 Kuroda S, Sugawara Y, Tamamura N, Takano-Yamamoto T. Anterior open bite with temporomandibular disorder treated with titanium screw anchorage: evaluation of morphological and functional improvement. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):550-60.

The taper of CON group was smaller, which means it was a cylinder mini-implant. Mini-conical implants would be most suitable because it could have the thickness thinner and a diameter more resistant immediately below the point of loading application.2020 Freire JN, Silva NR, Gil JN, Magini RS, Coelho PG. Histomorphologic and histomophometric evaluation of immediately and early loaded miniimplants for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 June;131(6):704.e1-9. However, fractures occur less frequently during application of orthodontic force and more often during insertion or removal of the mini-implants.88 Morarend C, Qian F, Marshall SD, Southard KA, Grosland NM, Morgan TA, et al. Effect of screw diameter on orthodontic skeletal anchorage. Am J Orthod Dentofacial Orthop. 2009 Aug;136(2):224-9. In the present study, the tip region was the most susceptible area to fracture.

The results showed that SIN and NEO exhibited the highest mean fracture resistance as referred elsewhere.2121 Consolaro A, Sant´Anna E, Francischone Jr CE, Consolaro MFM, Barbosa BA. Mini-implantes: pontos consensuais e questionamentos sobre o seu uso clínico. Rev Dent Press Ortod Ortop Facial. 2008 Set-Out;13(5):20-9. The lower fracture torque on the neck was found in DEN (23.45 N.cm). In the tip region, the DEN presented the second lowest value of fracture torque (10.56 N.cm). Possibly this was due to differences in relation to the steps of thread and taper. The DEN group had means values of thread steps and taper larger than the other groups, contradicting the fact that fracture strength can be increased with the conical design.2222 Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res. 2006 Feb;17(1):109-14.

The lowest torque fracture values was shown in the tip region of CON and DEN groups and were also the groups that had the apical milling. The presence of this lateral groove reduces the internal diameter of the narrower region of the mini-implants which probably increased the fracture susceptibility at the tip, which would justify the lower values of fracture torque found in these groups.

Mini-implants of these five groups were made with the Ti6Al4V alloy, as perceived by energy dispersive spectroscopy. However, the strength of the titanium alloy depends also on the microstructure, which is influenced by the composition, heating treatment, and thermo-mechanical processing of the mini-implants.66 Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006 May;67(3):162-74. These variables must be evaluated in further studies.

The insertion torques ranged from 6.6 N.cm (RMO) to 10.2 N.cm (NEO). All groups had values that were within the recommended boundary2424 Herman RJ, Currier GF, Miyakec A. Mini-implant anchorage for maxillary canine retraction: A pilot study. Am J Orthod Dentofacial Orthop. 2006 Aug;130(2):228-35.,2828 Papageorgiou SN, Zogakis IP, Papadopoulos MA. Failure rates and associated risk factors of orthodontic miniscrew implants: a meta-analysis. Am J Orthod Dentofacial Orthop. 2012 Nov;142(5):577-95. between 5 and 10 N.cm, and the values of insertion torque could reach 15N.cm without major problems. All fracture torque resistance were higher than the insertion torque values in both the fracture tip region and the neck in all groups, showing that all mini-implants tested can be considered safe from the risk of fracture in clinical use. The insertion values showed significant differences among NEO (10.2 N.cm), CON (7.4 N.cm) and RMO (6.6 N.cm). The difference in insertion torque among the groups was possibly due to the difference in design among the mini-implants. The mini-implants designs showed a large difference between them in relation to the number of threads of screws, and consequently the pitch of the thread.

The NEO group had the second highest number of threads (n = 9) and was the group with the highest insertion torque (10.20 N.cm). As the retention mechanism is based on fitting mechanical structure in cortical and not necessarily the concept of osseointegration,44 Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56.,55 Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44. the shape and length of the screw threads are fundamental to primary stability.2020 Freire JN, Silva NR, Gil JN, Magini RS, Coelho PG. Histomorphologic and histomophometric evaluation of immediately and early loaded miniimplants for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 June;131(6):704.e1-9.,2222 Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res. 2006 Feb;17(1):109-14.,2525 Morais LS, Serra GG, Muller CA, Andrade LR, Palermo EFA, Elias CN, et al. Titanium alloy mini-implants for orthodontic anchorage: Immediate loading and metal ion release. Acta Biomater. 2007 May;3(3):331-9. Greater number of threads and closeness between them increase the mechanical stability and resistance in the insertion in the bone.55 Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44. However, there were significant differences between NEO and CON groups, which had the largest number of threads,77 Kuroda S, Sugawara Y, Tamamura N, Takano-Yamamoto T. Anterior open bite with temporomandibular disorder treated with titanium screw anchorage: evaluation of morphological and functional improvement. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):550-60. and the second lowest value of insertion torque (7.4 N.cm). This result was probably due to the difference in diameter between these two groups. The diameter is significantly associated with stability.1616 Melsen B, Verna C. Miniscrew implants: the Aarhus anchorage system. Semin Orthod. 2005 Mar;11(1):24-31. NEO and CON groups showed the highest (1,630.72 µm) and lowest (1,482.95 µm) outer diameter, respectively, among all groups. The outer diameter appears to be an important feature of variation of the insertion torque.2626 Wilmes B, Ottenstreuer S, Su Y, Drescher D. Impact of implant design on primary stability of orthodontic mini-implants. J Orofac Orthop. 2008 Jan;69(1):42-50. NEO group, with the largest diameter (1,630.72 µm), had the highest value of insertion torque (10.20 N.cm), and RMO and CON groups, with smaller diameters, showed the lowest values.

During the insertion, control features should be used such as micro-motor with controlled torque,1313 Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8. manual wrench11 Brandão LBC, Mucha JN. Grau de aceitação de mini-implantes por pacientes em tratamento ortodôntico - estudo preliminar. Rev Dental Press Ortod Ortop Facial. 2008 Set-Out;13(5):118-27. or dynanometer2727 Kyung HM, Park HS, Bae SM, Sung JH, Kim IB. Development of orthodontic microimplants for intraoral anchorage. J Clin Orthod. 2003 June;37(6):321-8. to prevent this torque from approaching or reaching the fracture torque of mini-implants. Fracture toughness of mini-implants varies according to the manufacturer and type of mini-implant, so the operator should be aware of the characteristics of mini-implants that influence the torque values of fracture before choosing one to be used, and the maximum torque that can safely be used clinically for the insertion.1818 Smith A, Hosein YK, Dunning CE, Tassi A. Fracture resistance of commonly used self-drilling orthodontic mini-implants. Angle Orthod. 2015 Jan;85(1):26-32.

CONCLUSIONS

Mini-implants of different brands have different design and morphology, and fracture torques resistance is determined by: outside diameter, internal diameter, ratio of inner and outer diameter, and milling in the apical region.

Among the mini-implants trademark evaluated, NEO and SIN mini-implants have showed the highest resistance to fracture of the neck and tip.

The insertion values torques was always lower than the fracture torque resistance of both the tip and the neck in all groups tested.

REFERENCES

  • 1
    Brandão LBC, Mucha JN. Grau de aceitação de mini-implantes por pacientes em tratamento ortodôntico - estudo preliminar. Rev Dental Press Ortod Ortop Facial. 2008 Set-Out;13(5):118-27.
  • 2
    Elias CN, Guimarães GS, Muller CA. Torque de inserção e de remoção de miniparafusos ortodônticos. Rev Bras Implant. 2005;11(3):5-8.
  • 3
    Mah J, Bergstrand F. Temporary anchorage devices: a status report. J Clin Orthod. 2005 Mar;39(3):132-6.
  • 4
    Araújo TM, Nascimento MHA, Bezerra F, Sobral M. C. Ancoragem esquelética em Ortodontia com mini-implantes. Rev Dental Press Ortod Ortop Facial. 2006 Jul-Ago;11(4):126-56.
  • 5
    Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):537-44.
  • 6
    Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006 May;67(3):162-74.
  • 7
    Kuroda S, Sugawara Y, Tamamura N, Takano-Yamamoto T. Anterior open bite with temporomandibular disorder treated with titanium screw anchorage: evaluation of morphological and functional improvement. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4):550-60.
  • 8
    Morarend C, Qian F, Marshall SD, Southard KA, Grosland NM, Morgan TA, et al. Effect of screw diameter on orthodontic skeletal anchorage. Am J Orthod Dentofacial Orthop. 2009 Aug;136(2):224-9.
  • 9
    Iijima M, Muguruma T, Brantley WA, Okayama M, Yuasa T, Mizoguchi I. Torsional properties and microstructures of miniscrew implants. Am J Orthod Dentofacial Orthop. 2008 Sept;134(3):333.e1-6.
  • 10
    Itsuki Y, Imamura E, Sugawara J. Temporary anchorage device with interchangeable superstructure for mandibular tooth movement. J World Fed Orthod. 2013;2(1):e19-29.
  • 11
    Kitahara-Ceia FMF, Assad-Loss TF, Mucha JN, Elias CN. Morphological evaluation of five active types of six types of orthodontic mini-implants. Dental Press J Orthod. 2013 Mar-Apr;18(2):36-41.
  • 12
    Kim JH, Ahn SJ, Chang YI. Histomorfometric and mechanical analyses of the drill free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2005 Aug;128(2):190-4.
  • 13
    Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8.
  • 14
    Nova MFP, Carvalho FR, Elias CN, Artese F. Avaliação do torque de inserção, remoção e fratura de diferentes mini-implantes ortodônticos. Rev Dent Press Ortod Ortop Facial. 2008 Set-Oct;13(5):76-87.
  • 15
    Hong SB, Kusnoto B, Kim EJ, BeGole EA, Hwang HS, Lim HJ. Prognostic factors associated with the success rates of posterior orthodontic miniscrew implants: a subgroup meta-analysis. Korean J Orthod. 2016 Mar;46(2):111-26.
  • 16
    Melsen B, Verna C. Miniscrew implants: the Aarhus anchorage system. Semin Orthod. 2005 Mar;11(1):24-31.
  • 17
    Barros SE, Janson G, Chiqueto K, Garib DG, Janson M. Effect of mini-implant diameter on fracture risk and self-drilling efficacy. Am J Orthod Dentofacial Orthop. 2011 Oct;140(4):e181-92.
  • 18
    Smith A, Hosein YK, Dunning CE, Tassi A. Fracture resistance of commonly used self-drilling orthodontic mini-implants. Angle Orthod. 2015 Jan;85(1):26-32.
  • 19
    Squeff LR, Simonson MBA, Elias CN, Nojima LI. Caracterização de miniimplantes utilizados na ancoragem ortodôntica. Rev Dent Press Ortod Ortop Facial. 2008 Set-Out;13(5):49-56.
  • 20
    Freire JN, Silva NR, Gil JN, Magini RS, Coelho PG. Histomorphologic and histomophometric evaluation of immediately and early loaded miniimplants for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 June;131(6):704.e1-9.
  • 21
    Consolaro A, Sant´Anna E, Francischone Jr CE, Consolaro MFM, Barbosa BA. Mini-implantes: pontos consensuais e questionamentos sobre o seu uso clínico. Rev Dent Press Ortod Ortop Facial. 2008 Set-Out;13(5):20-9.
  • 22
    Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res. 2006 Feb;17(1):109-14.
  • 23
    Pithon MM, Santos RL, Derech CD, Elias CN, Ruellas ACO, Nojima LI. In vitro evaluation of torsional strength of orthodontic mini-implants. Braz J Oral Sci. 2008 Apr-June;7(25):1563-5.
  • 24
    Herman RJ, Currier GF, Miyakec A. Mini-implant anchorage for maxillary canine retraction: A pilot study. Am J Orthod Dentofacial Orthop. 2006 Aug;130(2):228-35.
  • 25
    Morais LS, Serra GG, Muller CA, Andrade LR, Palermo EFA, Elias CN, et al. Titanium alloy mini-implants for orthodontic anchorage: Immediate loading and metal ion release. Acta Biomater. 2007 May;3(3):331-9.
  • 26
    Wilmes B, Ottenstreuer S, Su Y, Drescher D. Impact of implant design on primary stability of orthodontic mini-implants. J Orofac Orthop. 2008 Jan;69(1):42-50.
  • 27
    Kyung HM, Park HS, Bae SM, Sung JH, Kim IB. Development of orthodontic microimplants for intraoral anchorage. J Clin Orthod. 2003 June;37(6):321-8.
  • 28
    Papageorgiou SN, Zogakis IP, Papadopoulos MA. Failure rates and associated risk factors of orthodontic miniscrew implants: a meta-analysis. Am J Orthod Dentofacial Orthop. 2012 Nov;142(5):577-95.
  • »
    The authors report no commercial, proprietary or financial interest in the products or companies described in this article.