Acessibilidade / Reportar erro

Reliability of biceps femoris and semitendinosus muscle architecture measurements obtained with ultrasonography

Abstract

Introduction

Currently, little attention is given to the muscle architecture reliability studies of the hamstring using a robust statistical. Our purpose was to determine the reliability of ultrasound measurements of muscle thickness, fascicle length and pennation angle of the biceps femoris and semitendinosus muscles, including heteroskedasticity and internal consistency analyses.

Methods

Two images of biceps femoris and semitendinosus at 50% of the thigh length were acquired from 21 volunteers, in two visits. The parameters were measured three times in each image, and for each muscle. The reliability was analyzed by the intraclass correlation coefficient (ICC) and Cronbach’s alpha (αCronbach). The relative standard error of the measurements (%SEM) were calculated and Bland-Altman plots were generated.

Results

All parameters presented excellent ICC for the three repeated measurements (ICC from 0.93 ‒ 0.99) and moderate to excellent reliability intraday (ICC from 0.70 ‒ 0.95) for both muscles. The present study indicates that ultrasound is a reliable tool to estimate the biceps femoris fascicle length (ICC = 0.97, αCronbach = 0.98, %SEM = 7.86) and semitendinosus (ICC = 0.90, αCronbach = 0.95, %SEM = 7.55), as well as the biceps femoris muscle thickness (ICC = 0.89, αCronbach = 0.94, %SEM = 10.23) and semitendinosus muscle thickness (ICC = 0.87, αCronbach = 0.93, %SEM = 1.35). At last, biceps femoris pennation angle (ICC = 0.93, αCronbach = 0.96 and %SEM = 4.36) and semitendinosus (ICC = 0.96, αCronbach = 0.98 and %SEM = 4.25) also had good repeatability.

Conclusion

Ultrasonography show good repeatability in estimating of muscle architecture parameters.

Keywords:
Reliability; Hamstring; Ultrasound; Muscle architecture


Introduction

Muscle architecture is defined as the arrangement of muscle fibers and plays an important role in muscle biomechanics studies (Lieber, 2010Lieber RL. Skeletal muscle anatomy. In: Lieber RL. Skeletal muscle, structure, function, and plasticity: the physiological basis of rehabilitation. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2010. p. 26-41.). Fiber length, pennation angle and muscle thickness are commonly measured as architecture parameters (Blazevich et al., 2006Blazevich AJ, Gill ND, Zhou S. Intra‐and intermuscular variation in human quadriceps femoris architecture assessed in vivo. Journal of Anatomy. 2006; 209(3):289-310. PMid:16928199. http://dx.doi.org/10.1111/j.1469-7580.2006.00619.x.
http://dx.doi.org/10.1111/j.1469-7580.20...
; Gomes et al., 2010Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
; Lima et al., 2012Lima KM, Matta TT, Oliveira LF. Reliability of the rectus femoris muscle cross‐sectional area measurements by ultrasonography. Clinical Physiology and Functional Imaging. 2012; 32(3):221-6. PMid:22487157. http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x.
http://dx.doi.org/10.1111/j.1475-097X.20...
; Martins et al., 2012Martins NS, Peixinho CC, Oliveira LF. Confiabilidade de medidas de arquitetura muscular do tríceps sural por ultrassonografia de imagem. Revista Brasileira de Cineantropometria & Desempenho Humano. 2012; 14(2):212-20.). Technological advances in imaging techniques, as magnetic resonance imaging and ultrasound, enable to estimate muscle architecture parameters in vivo (Blazevich et al., 2006Blazevich AJ, Gill ND, Zhou S. Intra‐and intermuscular variation in human quadriceps femoris architecture assessed in vivo. Journal of Anatomy. 2006; 209(3):289-310. PMid:16928199. http://dx.doi.org/10.1111/j.1469-7580.2006.00619.x.
http://dx.doi.org/10.1111/j.1469-7580.20...
; Chleboun et al., 2001Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
; Lima et al., 2012Lima KM, Matta TT, Oliveira LF. Reliability of the rectus femoris muscle cross‐sectional area measurements by ultrasonography. Clinical Physiology and Functional Imaging. 2012; 32(3):221-6. PMid:22487157. http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x.
http://dx.doi.org/10.1111/j.1475-097X.20...
). The architecture parameters can be well visualized by ultrasound at rest or during contraction (Gomes et al., 2010Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
; Martins et al., 2012Martins NS, Peixinho CC, Oliveira LF. Confiabilidade de medidas de arquitetura muscular do tríceps sural por ultrassonografia de imagem. Revista Brasileira de Cineantropometria & Desempenho Humano. 2012; 14(2):212-20.; Potier et al., 2009Potier T, Alexander C, Seynnes O. Effects of eccentric strength training on biceps femoris muscle architecture and knee joint range of movement. European Journal of Applied Physiology. 2009; 105(6):939-44. PMid:19271232. http://dx.doi.org/10.1007/s00421-008-0980-7.
http://dx.doi.org/10.1007/s00421-008-098...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
). The reliability of muscle architecture measurements assessed using ultrasound is reported in many studies with different methodologies involving repeated measures between images, sessions, and raters (Legerlotz et al., 2010Legerlotz K, Smith HK, Hing WA. Variation and reliability of ultrasonographic quantification of the architecture of the medial gastrocnemius muscle in young children. Clinical Physiology and Functional Imaging. 2010; 30(3):198-205. http://dx.doi.org/10.1111/j.1475-097X.2010.00925.x. PMid:20184623.
http://dx.doi.org/10.1111/j.1475-097X.20...
; Lima et al., 2012Lima KM, Matta TT, Oliveira LF. Reliability of the rectus femoris muscle cross‐sectional area measurements by ultrasonography. Clinical Physiology and Functional Imaging. 2012; 32(3):221-6. PMid:22487157. http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x.
http://dx.doi.org/10.1111/j.1475-097X.20...
). These studies have been important to determine a reproducibility of muscle architecture parameters which are measured very often in training and rehabilitation protocols. Concerning the lower limbs, reliability studies in vivo are described for the knee extensors (Lima et al., 2012Lima KM, Matta TT, Oliveira LF. Reliability of the rectus femoris muscle cross‐sectional area measurements by ultrasonography. Clinical Physiology and Functional Imaging. 2012; 32(3):221-6. PMid:22487157. http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x.
http://dx.doi.org/10.1111/j.1475-097X.20...
; Noorkoiv et al., 2010Noorkoiv M, Nosaka K, Blazevich AJ. Assessment of quadriceps muscle cross-sectional area by ultrasound extended-field-of-view imaging. European Journal of Applied Physiology. 2010; 109(4):631-9. PMid:20191287. http://dx.doi.org/10.1007/s00421-010-1402-1.
http://dx.doi.org/10.1007/s00421-010-140...
) and gastrocnemius muscles (Legerlotz et al., 2010Legerlotz K, Smith HK, Hing WA. Variation and reliability of ultrasonographic quantification of the architecture of the medial gastrocnemius muscle in young children. Clinical Physiology and Functional Imaging. 2010; 30(3):198-205. http://dx.doi.org/10.1111/j.1475-097X.2010.00925.x. PMid:20184623.
http://dx.doi.org/10.1111/j.1475-097X.20...
; Martins et al., 2012Martins NS, Peixinho CC, Oliveira LF. Confiabilidade de medidas de arquitetura muscular do tríceps sural por ultrassonografia de imagem. Revista Brasileira de Cineantropometria & Desempenho Humano. 2012; 14(2):212-20.). The reliability of muscle architecture measurements of the knee flexors, using ultrasound, is less examined in vivo (Chleboun et al., 2001Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
; Gomes et al., 2010Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
).

Reliability studies of knee flexors focused mainly on the biceps femoris long head muscle, and estimative of the intraclass correlation coefficient (ICC) to fascicle length, muscle thickness and pennation angle ranged around 0.87 – 0.95 (Chleboun et al., 2001Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
). For these parameters, it was found coefficient of variation (CV) below 5% at rest. Kellis et al. (2009)Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011.
http://dx.doi.org/10.1016/j.jbiomech.200...
compared the muscle thickness, fascicle length and pennation length measurements of the semitendinosus and the biceps femoris long head muscles using a caliper to validate ultrasound images of a cadaveric limb. The authors showed good reliability for all parameters (ICC > 0.79) and standard error of measurement (SEM) between 4.7 to 9.7%. Measurements of whole muscle thickness of the knee flexors reported by Gomes et al. (2010)Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
was moderate (ICC = 0.55), with 6.66% of CV and 9.6% of SEM.

Statistical estimators as SEM and ICC are commonly referred in reliability studies. As can be seen in the previous paragraph, statistical estimators as SEM presuppose absence of data heteroskedasticity, which refers to the proportional increase of the data magnitude with the random errors (Atkinson and Nevill, 1998Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Medicine. 1998; 26(4):217-38. PMid:9820922. http://dx.doi.org/10.2165/00007256-199826040-00002.
http://dx.doi.org/10.2165/00007256-19982...
). Furthermore, the internal consistency of the measurements estimates how items are correlated within group data, but has often been neglected in those studies (Hopkins, 2000Hopkins WG. Measures of reliability in sports medicine and science. Sports Medicine. 2000; 30(1):1-15. PMid:10907753. http://dx.doi.org/10.2165/00007256-200030010-00001.
http://dx.doi.org/10.2165/00007256-20003...
). Based on this, the aim of this study is to analyze the absolute test-retest reliability of the muscle thickness, fascicle length and pennation angle of the biceps femoris long head and semitendinosus muscles using ultrasound. The absence of heteroskedasticity and internal consistency were included in the statistical analysis.

Methods

Participants

Twenty-one (N = 21) healthy volunteers participated in this study. The volunteers were 6 women and 15 men aged 28.08 ± 4.61 years, body mass of 74.03 ± 13.49 kg, and height of 1.74 ± 0.08 m. They did not have any diseases in the lower limbs. The study was approved by the Ethics and Research Committee of the Clementino Fraga Filho University Hospital (n° 023/11). Informed consent was obtained from all individual participants included in the study. Volunteers were instructed not to perform any type of vigorous physical activity involving the lower limbs during the test.

Ultrasound measurements

Images acquisition were performed by the same trained examiner with an ultrasonographic system operating in B-Mode (EUB-405; Hitachi, Tokyo, Japan) with a 7.5 MHz linear array probe. With the participants at standing position, the examiner marked one point at 50% of the length of the thigh, determined by the distance between the greater trochanter and head of the fibula. After that, the subject laid with the legs relaxed in prone position for 15 minutes before image acquisition on the right leg.

The protocol was repeated in two days (interdays reliability), with a 48-hour interval between visits. Two images of biceps femoris long head and semitendinosus muscles were recorded by video capture EasyCap USB 2.0 (UAF Co, Limited, Shenzhen, China) in each day (intradays reliability). The gel was applied to ensure the acoustic coupling on the surface of the skin and a minimal compression of the probe under skin was adopted. The probe was positioned along the direction of the fascicles, where the fascicular organization between the superficial and deep aponeurosis on the muscle was better visualized.

In each of the images, muscle thickness, fascicle length and pennation angle variables were measured three times in random order using IMAGE J software (version 1.42; National Institute of Health, Bethesda, USA). Figure 1 explains the experimental design.

Figure 1
Experimental design for each participant. All volunteers (N = 21) participated of two days of the test, and two images for biceps femoris and semitendinosus were obtained (8 images per participant). In each image, pennation angle, fascicle length and muscle thickness were repeated three times.

Muscle thickness was measured as the mean distance between the superficial and deep aponeurosis at both image extremities. Fascicle length was estimated according to Blackburn and Pamukoff (2014)Blackburn J, Pamukoff D. Geometric and architectural contributions to hamstring musculotendinous stiffness. Clinical Biomechanics. 2014; 29(1):105-10. PMid:24220042. http://dx.doi.org/10.1016/j.clinbiomech.2013.10.011.
http://dx.doi.org/10.1016/j.clinbiomech....
. Parameters were calculated as the relative acute angle formed between the deep aponeurosis and muscle fascicle, as represented in Figure 2.

Figure 2
Ultrasound images of muscle thickness, fascicle length and pennation angle measurements of the biceps femoris long head (A) and semitendinosus (B).

Statistical analysis

The distribution of each variable was examined with D’Agostino-Pearson test (smallest p-value found = 0.07). ANOVA repeated measurements was used to test statistical differences among measurements of each parameter in the same image, to compare two images of the same days and, finally, all measurements obtained in two days. The relative reliability among three repeated measurements, intra and interdays was determined by ICC and ɑCronbach. Model ‘2,k’ (two-way randow factor) proposed by Shrout and Fleiss (1979)Shrout P, Fleiss J. Intraclass correlations: uses in assessing rater reliability. Psychological Bulletin. 1979; 86(2):420-8. PMid:18839484. http://dx.doi.org/10.1037/0033-2909.86.2.420.
http://dx.doi.org/10.1037/0033-2909.86.2...
was chosen to calculate ICC. After, ICC was classified according to Koo and Li (2016)Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Journal of Chiropractic Medicine. 2016; 15(2):155-63. PMid:27330520. http://dx.doi.org/10.1016/j.jcm.2016.02.012.
http://dx.doi.org/10.1016/j.jcm.2016.02....
and presented with its confidence interval. ɑCronbach was applied for estimative of internal consistency. Statistical analysis was performed using SPSS version 21 (IBM Corporation, New York, USA), with significance level of α = 0.05. The heteroskedasticity was analysed by Levene’s test and Bland-Altman plots in MATLAB (version 7.10; MathWorks, USA) to present limits of agreement of data. If heteroskedasticity was present, then Bland-Altman plots were generated after logarithmic transformation of the data. Bland-Altman plots interdays were constructed to analyze graphically the distribution of data within the 95% limits of agreement and the difference between days (bias). The relative coefficient of variation (in percent - %CV) will be estimated in the presence of heteroskedasticity, otherwise relative standard error of measurement (in percent - %SEM) will be calculated, based on the mean square error of the one-way repeated-measures ANOVA (Atkinson and Nevill, 1998Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Medicine. 1998; 26(4):217-38. PMid:9820922. http://dx.doi.org/10.2165/00007256-199826040-00002.
http://dx.doi.org/10.2165/00007256-19982...
; Hopkins, 2000Hopkins WG. Measures of reliability in sports medicine and science. Sports Medicine. 2000; 30(1):1-15. PMid:10907753. http://dx.doi.org/10.2165/00007256-200030010-00001.
http://dx.doi.org/10.2165/00007256-20003...
).

Results

Among three Measurements: Averages of fascicle length, muscle thickness and pennation angle for biceps were 10.46 ± 2.01 cm; 2.04 ± 0.36 cm and 12.52 ± 1.92°, respectively. Fascicle length averages of semitendinosus were 10.15 ± 2.18 cm; muscle thickness 2.33 ± 0.40 cm and 14.28 ± 2.14º to pennation angle. There are no significant differences among groups of the three measurements of each parameter obtained in each image acquired in two days (F = 0.003 to 3.625; p = 0.071 to 0.955). The ICC2,3 and ɑCronbach of the three measurements are presented in Table 1. All variables presented excellent ICC for both muscles, ranging from 0.93-0.99. The ɑCronbach ranged from 0.95 ‒ 0.99 for fascicle length, from 0.97 ‒ 0.99 for pennation angle, and 0.99 for muscle thickness, demonstrating excellent internal consistency.

Table 1
Reliability coefficients of the three measurements.

Intraday Reliability: There are no significant differences for the six measurements obtained for the same parameter in the two images in two days (F = 0.017 to 1.434; p = 0.245 to 0.898). The ICC2,3 of the images for all variables are presented in Table 2. The ICC for all variables of biceps femoris and semitendinosus muscles were moderate to excellent (ICC from 0.70-0.95), and there is increase confidence interval. Values of ɑCronbach also show excellent internal consistency between images.

Table 2
Intraday reliability for fascicle length, muscle thickness and pennation angle.

Interday Reliability: There are no significant differences for all the 12 measurements obtained for the same parameter in the two images along of two days (F = 0.069 to 1.013 ; p = 0.326 to 0.795). In all parameters there was no heteroskedasticity (F< 1.03; p > 0.37). Moreover, SEM was determined for each parameter. Bland-Altman plots indicate good agreement for the interday measurements within 95% limits of agreement. All parameters were closer to the zero. Results of ICC2,3, ɑCronbach and SEM are presented in Figure 3. The ICC was excellent for all parameters. Variables had good internal consistency (ɑCronbach > 0.96).

Figure 3
Bland-Altman plots of muscle architecture interdays measurements. Dashed lines represent upper and lower 95% limits of agreement. Solid lines represent the estimated bias.

Discussion

The present study aimed to examine the reliability of the muscle thickness, pennation angle and fascicle length measurements from ultrasonographic images from the biceps femoris long head and semitendinosus muscles. Our main results show that measurements of muscle architecture parameters presented good reliability.

The muscle architecture values are in accordance to the in vivo (Chleboun et al., 2001Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
; Potier et al., 2009Potier T, Alexander C, Seynnes O. Effects of eccentric strength training on biceps femoris muscle architecture and knee joint range of movement. European Journal of Applied Physiology. 2009; 105(6):939-44. PMid:19271232. http://dx.doi.org/10.1007/s00421-008-0980-7.
http://dx.doi.org/10.1007/s00421-008-098...
) and in vitro (Kellis et al., 2009Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011.
http://dx.doi.org/10.1016/j.jbiomech.200...
; 2011Kellis E, Galanis N, Natsis K, Kapetanos G. In vivo and in vitro examination of the tendinous inscription of the human semitendinosus muscle. Cells, Tissues, Organs. 2011; 195(4):365-76. PMid:21828998. http://dx.doi.org/10.1159/000327574.
http://dx.doi.org/10.1159/000327574...
; Ward et al., 2009Ward SR, Eng CM, Smallwood LH, Lieber RL. Are current measurements of lower extremity muscle architecture accurate? Clinical Orthopaedics and Related Research. 2009; 467(4):1074-82. http://dx.doi.org/10.1007/s11999-008-0594-8.
http://dx.doi.org/10.1007/s11999-008-059...
) literature. Biceps femoris fascicle length values corroborate one that reported values around 10.4 cm in vitro (Kellis et al., 2012Kellis E, Galanis N, Kapetanos G, Natsis K. Architectural differences between the hamstring muscles. Journal of Electromyography and Kinesiology. 2012; 22(4):520-6. PMid:22564790. http://dx.doi.org/10.1016/j.jelekin.2012.03.012.
http://dx.doi.org/10.1016/j.jelekin.2012...
). On other hand, fascicle length values for semitendinosus were lower than in vitro studies (13 to 19 cm) (Kellis et al., 2012Kellis E, Galanis N, Kapetanos G, Natsis K. Architectural differences between the hamstring muscles. Journal of Electromyography and Kinesiology. 2012; 22(4):520-6. PMid:22564790. http://dx.doi.org/10.1016/j.jelekin.2012.03.012.
http://dx.doi.org/10.1016/j.jelekin.2012...
; Ward et al., 2009Ward SR, Eng CM, Smallwood LH, Lieber RL. Are current measurements of lower extremity muscle architecture accurate? Clinical Orthopaedics and Related Research. 2009; 467(4):1074-82. http://dx.doi.org/10.1007/s11999-008-0594-8.
http://dx.doi.org/10.1007/s11999-008-059...
; Woodley and Mercer, 2005Woodley S, Mercer S. Hamstring muscles: architecture and innervation. Cells Tissues Organs. 2005; 179(3):125-41. http://dx.doi.org/10.1159/000085004.
http://dx.doi.org/10.1159/000085004...
). This can be explained by the large variability of muscle architecture along the whole length of the hamstring components resulting in fascicles ranging from 5.2 to 18.3 cm (Woodley and Mercer, 2005Woodley S, Mercer S. Hamstring muscles: architecture and innervation. Cells Tissues Organs. 2005; 179(3):125-41. http://dx.doi.org/10.1159/000085004.
http://dx.doi.org/10.1159/000085004...
). Additionaly, semitendinosus is divided in two compartments separated by a tendinous inscription, and there are differences between the superficial and deep regions (Woodley and Mercer, 2005Woodley S, Mercer S. Hamstring muscles: architecture and innervation. Cells Tissues Organs. 2005; 179(3):125-41. http://dx.doi.org/10.1159/000085004.
http://dx.doi.org/10.1159/000085004...
). The authors alerted that this structure has been neglected and showed that the superficial fascicles are significantly smaller than the deepest ones. We estimated the fascicle length using trigonometric extrapolation as others (Blazevich et al., 2006Blazevich AJ, Gill ND, Zhou S. Intra‐and intermuscular variation in human quadriceps femoris architecture assessed in vivo. Journal of Anatomy. 2006; 209(3):289-310. PMid:16928199. http://dx.doi.org/10.1111/j.1469-7580.2006.00619.x.
http://dx.doi.org/10.1111/j.1469-7580.20...
; Legerlotz et al., 2010Legerlotz K, Smith HK, Hing WA. Variation and reliability of ultrasonographic quantification of the architecture of the medial gastrocnemius muscle in young children. Clinical Physiology and Functional Imaging. 2010; 30(3):198-205. http://dx.doi.org/10.1111/j.1475-097X.2010.00925.x. PMid:20184623.
http://dx.doi.org/10.1111/j.1475-097X.20...
), which assumes that the fascicles are linear, ignoring the curvilinear arrangements (Noorkoiv et al., 2010Noorkoiv M, Nosaka K, Blazevich AJ. Assessment of quadriceps muscle cross-sectional area by ultrasound extended-field-of-view imaging. European Journal of Applied Physiology. 2010; 109(4):631-9. PMid:20191287. http://dx.doi.org/10.1007/s00421-010-1402-1.
http://dx.doi.org/10.1007/s00421-010-140...
). This could have resulted in underestimated semitendinosus fascicle length. As far as we know, this is the first reliability in vivo study for the semitendinosus muscle.

In this work, ICC indicates repeatability and ɑCronbach points out the internal consistency. Our results show excellent internal consistency and repeatability when considering repeated measurements in the same image for both muscles. There was a reduction of these ICC when different images at different days were analyzed followed by an increase of the confidence interval. This behavior is hardly discussed in the literature, as only the mean coefficient is reported among measurements (Gomes et al., 2010Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
). Bland-Altman plots showed small difference among measurements with bias close to zero, and almost all within limits of agreements.

Our results were similar to Timmins et al. (2015)Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
, who obtained high reproducibility (ICC > 0.93) and SEM of 4.9%, notwithstanding they estimated the fascicle length of biceps femoris by another methodology different from ours. Chleboun et al. (2001)Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
also demonstrated excellent reliability (ICC = 0.87) for biceps femoris fascicle length measurements of the in vivo. In current work, the pennation angle had excellent reliability as the data reported in others studies, ranging of 0.87 to 0.95 (Chleboun et al., 2001Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908.
http://dx.doi.org/10.1159/000047908...
; Kellis et al., 2009Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011.
http://dx.doi.org/10.1016/j.jbiomech.200...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
) and a SEM around 3.2 and 9.5% (Kellis et al., 2009Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011.
http://dx.doi.org/10.1016/j.jbiomech.200...
; Timmins et al., 2015Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
). Biceps femoris muscle thickness had excellent reliability, corroborating Timmins et al. (2015)Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507.
http://dx.doi.org/10.1249/MSS.0000000000...
that showed ICC of 0.95. Gomes et al. (2010)Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008.
http://dx.doi.org/10.1590/S1517-86922010...
found a moderate reliability (ICC = 0.55, CV = 6.66% and SEM = 4.9 mm) for knee flexors muscle thickness within interdays measurements. In this case, components of the flexors knee were singly evaluated.

Reliability data for semitendinosus ratifies the in vitro literature. Kellis et al. (2009)Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011.
http://dx.doi.org/10.1016/j.jbiomech.200...
reported for fascicle length, muscle thickness and pennation angle values to ICC of 0.77, 0.90 and 0.97, respectively, from six cadaveric limbs. The authors found SEM below 10%. In vivo, high intrarater and interrater reproducibility was reported to semitendinosus pennation angle (ICC > 0.83) (Kellis et al., 2011Kellis E, Galanis N, Natsis K, Kapetanos G. In vivo and in vitro examination of the tendinous inscription of the human semitendinosus muscle. Cells, Tissues, Organs. 2011; 195(4):365-76. PMid:21828998. http://dx.doi.org/10.1159/000327574.
http://dx.doi.org/10.1159/000327574...
). This study is the first to describe the in vivo muscle thickness and fascicle length measurements reliability for the semitendinosus muscle using ultrasound.

The present study indicates that muscle thickness, pennation angle and fascicle length for both biceps femoris long head and semitendinosus presented good reliability and repeatability in estimative of muscle architecture parameters by ultrasonography. The protocol of this study resulted in excellent ICC and ɑCronbach, and these indices can to be used by future studies approaching therapeutic interventions in muscle architecture hamstring.

Acknowledgements

The authors wish to acknowledge the financial support of the Brazilian Research agencies CAPES, FINEP, FAPERJ and CNPq.

References

  • Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Medicine. 1998; 26(4):217-38. PMid:9820922. http://dx.doi.org/10.2165/00007256-199826040-00002
    » http://dx.doi.org/10.2165/00007256-199826040-00002
  • Blackburn J, Pamukoff D. Geometric and architectural contributions to hamstring musculotendinous stiffness. Clinical Biomechanics. 2014; 29(1):105-10. PMid:24220042. http://dx.doi.org/10.1016/j.clinbiomech.2013.10.011
    » http://dx.doi.org/10.1016/j.clinbiomech.2013.10.011
  • Blazevich AJ, Gill ND, Zhou S. Intra‐and intermuscular variation in human quadriceps femoris architecture assessed in vivo. Journal of Anatomy. 2006; 209(3):289-310. PMid:16928199. http://dx.doi.org/10.1111/j.1469-7580.2006.00619.x
    » http://dx.doi.org/10.1111/j.1469-7580.2006.00619.x
  • Chleboun GS, France R, Crill MT, Braddock HK, Howell JN. In vivo measurement of fascicle length and pennation angle of the human biceps femoris muscle. Cells, Tissues, Organs. 2001; 169(4):401-9. PMid:11490120. http://dx.doi.org/10.1159/000047908
    » http://dx.doi.org/10.1159/000047908
  • Gomes PS, Meirelles CD, Leite SP, Montenegro CA. Confiabilidade da medida de espessuras musculares pela ultrassonografia. Revista Brasileira de Medicina do Esporte. 2010; 16(1):41-5. http://dx.doi.org/10.1590/S1517-86922010000100008
    » http://dx.doi.org/10.1590/S1517-86922010000100008
  • Hopkins WG. Measures of reliability in sports medicine and science. Sports Medicine. 2000; 30(1):1-15. PMid:10907753. http://dx.doi.org/10.2165/00007256-200030010-00001
    » http://dx.doi.org/10.2165/00007256-200030010-00001
  • Kellis E, Galanis N, Kapetanos G, Natsis K. Architectural differences between the hamstring muscles. Journal of Electromyography and Kinesiology. 2012; 22(4):520-6. PMid:22564790. http://dx.doi.org/10.1016/j.jelekin.2012.03.012
    » http://dx.doi.org/10.1016/j.jelekin.2012.03.012
  • Kellis E, Galanis N, Natsis K, Kapetanos G. Validity of architectural properties of the hamstring muscles: correlation of ultrasound findings with cadaveric dissection. Journal of Biomechanics. 2009; 42(15):2549-54. PMid:19646698. http://dx.doi.org/10.1016/j.jbiomech.2009.07.011
    » http://dx.doi.org/10.1016/j.jbiomech.2009.07.011
  • Kellis E, Galanis N, Natsis K, Kapetanos G. In vivo and in vitro examination of the tendinous inscription of the human semitendinosus muscle. Cells, Tissues, Organs. 2011; 195(4):365-76. PMid:21828998. http://dx.doi.org/10.1159/000327574
    » http://dx.doi.org/10.1159/000327574
  • Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Journal of Chiropractic Medicine. 2016; 15(2):155-63. PMid:27330520. http://dx.doi.org/10.1016/j.jcm.2016.02.012
    » http://dx.doi.org/10.1016/j.jcm.2016.02.012
  • Legerlotz K, Smith HK, Hing WA. Variation and reliability of ultrasonographic quantification of the architecture of the medial gastrocnemius muscle in young children. Clinical Physiology and Functional Imaging. 2010; 30(3):198-205. http://dx.doi.org/10.1111/j.1475-097X.2010.00925.x PMid:20184623.
    » http://dx.doi.org/10.1111/j.1475-097X.2010.00925.x
  • Lieber RL. Skeletal muscle anatomy. In: Lieber RL. Skeletal muscle, structure, function, and plasticity: the physiological basis of rehabilitation. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2010. p. 26-41.
  • Lima KM, Matta TT, Oliveira LF. Reliability of the rectus femoris muscle cross‐sectional area measurements by ultrasonography. Clinical Physiology and Functional Imaging. 2012; 32(3):221-6. PMid:22487157. http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x
    » http://dx.doi.org/10.1111/j.1475-097X.2011.01115.x
  • Martins NS, Peixinho CC, Oliveira LF. Confiabilidade de medidas de arquitetura muscular do tríceps sural por ultrassonografia de imagem. Revista Brasileira de Cineantropometria & Desempenho Humano. 2012; 14(2):212-20.
  • Noorkoiv M, Nosaka K, Blazevich AJ. Assessment of quadriceps muscle cross-sectional area by ultrasound extended-field-of-view imaging. European Journal of Applied Physiology. 2010; 109(4):631-9. PMid:20191287. http://dx.doi.org/10.1007/s00421-010-1402-1
    » http://dx.doi.org/10.1007/s00421-010-1402-1
  • Potier T, Alexander C, Seynnes O. Effects of eccentric strength training on biceps femoris muscle architecture and knee joint range of movement. European Journal of Applied Physiology. 2009; 105(6):939-44. PMid:19271232. http://dx.doi.org/10.1007/s00421-008-0980-7
    » http://dx.doi.org/10.1007/s00421-008-0980-7
  • Shrout P, Fleiss J. Intraclass correlations: uses in assessing rater reliability. Psychological Bulletin. 1979; 86(2):420-8. PMid:18839484. http://dx.doi.org/10.1037/0033-2909.86.2.420
    » http://dx.doi.org/10.1037/0033-2909.86.2.420
  • Timmins RG, Shield A, Williams M, Lorenzen C, Opar D. Biceps Femoris Long-Head Architecture: a reliability and retrospective injury study. Medicine and Science in Sports and Exercise. 2015; 47(5):905-13. http://dx.doi.org/10.1249/MSS.0000000000000507
    » http://dx.doi.org/10.1249/MSS.0000000000000507
  • Ward SR, Eng CM, Smallwood LH, Lieber RL. Are current measurements of lower extremity muscle architecture accurate? Clinical Orthopaedics and Related Research. 2009; 467(4):1074-82. http://dx.doi.org/10.1007/s11999-008-0594-8
    » http://dx.doi.org/10.1007/s11999-008-0594-8
  • Woodley S, Mercer S. Hamstring muscles: architecture and innervation. Cells Tissues Organs. 2005; 179(3):125-41. http://dx.doi.org/10.1159/000085004
    » http://dx.doi.org/10.1159/000085004

Publication Dates

  • Publication in this collection
    Dec 2016

History

  • Received
    22 Jan 2016
  • Accepted
    20 Jan 2017
Sociedade Brasileira de Engenharia Biomédica Centro de Tecnologia, bloco H, sala 327 - Cidade Universitária, 21941-914 Rio de Janeiro RJ Brasil, Tel./Fax: (55 21)2562-8591 - Rio de Janeiro - RJ - Brazil
E-mail: rbe@rbejournal.org