Evaluation of gloves as a water bag coupling agent for therapeutic ultrasound

Abstract

Introduction

Therapeutic ultrasound (TUS) is a widespread modality in physiotherapy, and the water bag technique is a coupling method employed in the presence of anatomical irregularities in the treatment area. The aim of the present study is to evaluate the acoustic attenuation of the water bag and its effectiveness as a TUS coupling agent.

Methods

The rated output powers (ROPs) of the TUS equipment were evaluated based on IEC 61689. Then, a radiation force balance was used to measure ROP with and without a water bag (latex and nitrile gloves filled with deionized water) between a TUS transducer and the cone-shaped target of the balance. Each experiment was performed five times for each nominal power (0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, and 7.0 W) and in the following configurations: without the water bag (A), with nitrile gloves and with (B) and without (C) a height controller, and latex gloves with (D) and without (E) height controller. ROPs obtained in different media were compared.

Results

The highest relative error of ROP was 16.72% for 0.5 W. Although the power values of the equipment were within the range recommended by IEC, there was a significant difference between the ROP values measured with A and with B, C and D.

Conclusion

As intensity differences below 0.5 W/cm2 are considered clinically not relevant, conditions A, B, C, D, or E can be used interchangeably.

Keywords
Therapeutic ultrasound; Water bag technique; Acoustic power

Introduction

Ultrasound (US) is a mechanical wave with frequency above the human hearing range (20 Hz to 20 kHz) that transmits energy from one point to another through the vibration of molecules in a medium (Shung, 2006Shung KK. Diagnostic ultrasound: imaging and blood flow measurements. Boca Raton: CRC Press; 2006.).

Therapeutic ultrasound (TUS) is a technological modality that has been widely used in physical therapy worldwide. One of the therapeutic goals of TUS is to raise the temperature of the treated tissue to the range 40-45 °C to promote pain relief, improve signs of inflammation, increase cell membrane permeability, and enhance the extensibility of collagen.

Moreover, TUS improves blood supply and reduces muscle spasms (Cameron, 2012Cameron MH. Physical agents in rehabilitation: from research to practice. 4th ed. St. Louis: Saunders Elsevier; 2012.; Delforge, 2002Delforge G. Musculoskeletal trauma: implications for sports injury management. Champaign: Human Kinetics; 2002.; Nanda, 2008Nanda BK. Electrotherapy simplified. New Delhi: Jaypee Brothers Medical; 2008.; Prentice, 2005Prentice WE. Therapeutic modalities in rehabilitation. 3rd ed. Nova York: McGraw-Hill Medical; 2005.; Robertson et al., 2006Robertson V, Ward A, Low J, Reed A. Electrotherapy explained: principles and practice. 4th ed. Oxford: Butterworth-Heinemann Elsevier; 2006.; Watson, 2008aWatson T. Electrotherapy: evidence-based practice. 12th ed. Philadelphia: Churchill Livingstone Elsevier; 2008a.). The literature describes three techniques for the coupling of TUS to the treatment site: the direct-contact method, in which the transducer is coupled to the patient skin using clinical water-based gel; the submersion technique, in which the region receiving US irradiation and the US transducer are separated inside a water tank, with the water acting as the coupling medium; and the water bag method, in which a bag filled with water (or coupling gel) is placed between the skin and the US transducer. The last two techniques have the advantage of being suitable for anatomically uneven surfaces (Cameron, 2012Cameron MH. Physical agents in rehabilitation: from research to practice. 4th ed. St. Louis: Saunders Elsevier; 2012.; Kenyon and Kenyon, 2009Kenyon K, Kenyon J. The physiotherapist’s pocketbook: essential facts at your fingertips. 2nd ed. London: Churchill Livingstone Elsevier; 2009.; Robertson et al., 2006Robertson V, Ward A, Low J, Reed A. Electrotherapy explained: principles and practice. 4th ed. Oxford: Butterworth-Heinemann Elsevier; 2006.; Starkey, 2013Starkey C. Therapeutic modalities. 4th ed. Philadelphia: F.A. Davis; 2013.).

In clinical practice, gloves filled with water usually substitute the water bag as a coupling agent because they are easily accessible and have very low cost. Although TUS with water glove coupling is a common physiotherapy technique, there is a lack of scientific studies addressing its influence on the desired TUS performance. However, the beneficial physiological effects of TUS are known to depend on several parameters, including the selected dosage and the parameters of the US equipment, such as the US frequency and intensity (Baker et al., 2001Baker KG, Robertson VJ, Duck FA. A review of therapeutic ultrasound: biophysical effects. Physical Therapy. 2001; 81(7):1351-8. PMid:11444998.; Kollmann et al., 2005Kollmann C, Vacariu G, Schuhfried O, Fialka-Moser V, Bergmann H. Variations in the output power and surface heating effects of transducers in therapeutic ultrasound. Archives of Physical Medicine and Rehabilitation. 2005; 86(7):1318-24. PMid:16003657. http://dx.doi.org/10.1016/j.apmr.2005.02.001.
http://dx.doi.org/10.1016/j.apmr.2005.02...
). The acoustic energy deposited in a particular area of tissue depends on the calibration of the US equipment, the method of application (direct-contact, submersion and water bag), and the coupling agents (such as different types of hydrogel) between the US transducer and the insonated tissue (Artho et al., 2002Artho PA, Thyne JG, Warring BP, Willis CD, Brismée JM, Latman NS. A calibration study of therapeutic ultrasound units. Physical Therapy. 2002; 82(3):257-63. PMid:11869154.; Pye, 1996Pye S. Ultrasound therapy equipment: does it perform? Physiotherapy. 1996; 82(1):39-44. http://dx.doi.org/10.1016/S0031-9406(05)66996-9.
http://dx.doi.org/10.1016/S0031-9406(05)...
). Additionally, Klucinec et al. (2000a)Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Effectiveness of wound care products in the transmission of acoustic energy. Physical Therapy. 2000a; 80(5):469-76. PMid:10792857. quantified the sound energy transmitted through four types of hydrogel (coupling medium) manufactured by different companies and observed a considerable difference in the transmissivity of the US wave depending on the used hydrogel, with the transmitted voltage ranging between 42.8 ± 5.9% and 77.2 ± 4.6%. To maximize the energy transfer from an ultrasonic wave emitted by a transducer to the irradiated biological tissue, the acoustic impedance of the coupling medium must optimize the matching between the acoustic impedances of the transducer and the irradiated medium. When no coupling medium is used, the largest portion of the acoustic energy can be reflected on the transducer-air interface owing to the large impedance mismatch between these mediums. Nevertheless, Poltawski and Watson (2007)Poltawski L, Watson T. Relative transmissivity of ultrasound coupling agents commonly used by therapists in the UK. Ultrasound in Medicine & Biology. 2007; 33(1):120-8. PMid:17189054. http://dx.doi.org/10.1016/j.ultrasmedbio.2006.07.026.
http://dx.doi.org/10.1016/j.ultrasmedbio...
reported that depending on the composition, the absorption and scattering coefficients of the coupling medium may induce a loss of energy of the acoustic beam that reaches the tissue. Besides, the impedance of the coupling medium itself can reduce the wave energy transmitted to the tissue. Another study reported that the transmitted power is affected by the temperature of the coupling agents (Oshikoya et al., 2000Oshikoya CA, Shultz SJ, Mistry D, Perrin DH, Arnold BL, Gansneder BM. Effect of coupling medium temperature on rate of intramuscular temperature rise using continuous ultrasound. Journal of Athletic Training. 2000; 35(4):417-21. PMid:16558655.) because of changes in the coupling medium density, which is related to the acoustic impedance. Guirro et al. (2001)Guirro R, Cancelieri AS, Sant’Anna IL. Avaliação dos meios intermediários utilizados na aplicação do ultra-som terapêutico. Rev Bras Fisiot. 2001; 5(2):49-52. evaluated the transmissivity of a balloon and a latex glove filled with hydro-soluble gel to be 2% and 18%, respectively, using degassed water as a reference.

Klucinec et al. (2000b)Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Transmissivity of coupling agents used to deliver ultrasound through indirect methods. The Journal of Orthopaedic and Sports Physical Therapy. 2000b; 30(5):263-6. PMid:10817413. http://dx.doi.org/10.2519/jospt.2000.30.5.263.
http://dx.doi.org/10.2519/jospt.2000.30....
stated that the literature on the transmissivity of gel pads, bladder techniques, and water bath immersion is inconclusive and these topics require further investigation. Therefore, this study aims to evaluate the efficacy of TUS with the water bag technique by comparing the measurement of US power with and without the use of water-filled latex and nitrile gloves, which is an arrangement commonly used by physical therapy professionals.

Methods

The following equipment was used in the experiments: a radiation force balance (UPM-DT 1AV; Ohmic Instruments Co., St. Charles, Mo, USA) with a display resolution of ± 2 mW, one pair of latex gloves size M (Lagrotta Azzurra, São Paulo, SP, Brazil), one pair of nitrile gloves size M (Lagrotta Azzurra, São Paulo, SP, Brazil), ultrapure and degassed water, and a physical therapy US device with a circular-plane piston transducer (Seven Pro 977; Quark Medical Products, Piracicaba, SP, Brazil). According to the manufacturer of this device, the effective radiating area (ERA) of the transducer was 3.8 cm2, the beam non-uniformity ratio was lower than 6, and the ceramic with 2.48 cm in diameter. The thickness of the latex and nitrile gloves was 12 µm. The experimental setup is illustrated in Figure 1A.

Figure 1
(A) Experimental setup; (B) Latex and nitrile gloves filled with water. PVC ring filled with water and covered with (C) latex and (D) nitrile gloves.

The radiation force balance was composed of a 45° air-backed cone target immersed in a tank with rubber-coated inner surface. A positioning clamp was employed to keep the transducer immersed in ultrapure and degassed water above the conical target. As the physical therapy US device was configured to operate in continuous mode, the ultrasonic transducer generated sinusoidal continuous waves with 1 MHz frequency. This wave propagated through the water and was reflected from the conical target, exerting an ultrasonic force on this target. The vertical component of this force was registered as an increase in the apparent mass of the target. This measurement was performed by the support assembly that connected the target to an electronic digital balance. Using this approach, an apparent mass of 1 g corresponded to a power of 14.65 W. After being reflected by the target, the ultrasonic energy was absorbed by the rubber lining of the tank.

After preparing the experimental setup as shown in Figure 1A, the ultrasonic equipment remained switched off for 5 min before starting the experiment, as this is the time recommended by the manufacturer for the radiation force balance to stabilize. Then, a body mass of 1 g, which was provided by the manufacturer, was employed to calibrate the equipment.

Ultrapure and degassed water was used in the experiments because in common water, the concentration of diluted gases can promote different degrees of US cavitation, which can influence the US therapy performance (Wells, 1977Wells PNT. Biomedical ultrasonics. London: Academic Press; 1977.). Purified water was obtained from an Elga Purelab Option Q purification system (Elga Labwater, High Wycombe, Bucks, UK). To minimize cavitation, a vacuum chamber (model 166; Primar Equipment and Services Ltd., Itu, SP, Brazil) was used to remove diluted gases from the water, as recommended by International Electrotechnical Commission (IEC) 61689 (International…, 2007International Electrotechnical Commission. IEC 61689:2007. Ultrasonics: physiotherapy systems: field specifications and methods of measurement in the frequency range 0.5 MHz to 5 MHz. Geneva: IEC; 2007.). The water temperature was 25.0 ± 0.4 °C.

Experimental procedure

The first experimental setup, without the gloves, was used to determine the power emitted by the TUS equipment for all the applied nominal power values (calibration curve). The measurement served as a reference for the power measured when gloves were introduced in the experiments.

The second experimental setup included a glove filled with deionized water to evaluate its influence on the measured power. This condition simulates the water bag technique.

In an effort to perform measurements with a more controlled glove thickness, the measurement was repeated with the same TUS power, this time using a PVC cylinder (12.8 mm in height, 32.65 mm internal and 44.00 mm external diameters) filled with deionized water and covered with two gloves. A PVC ring was used to ensure that there were no wrinkles on the glove surface and that the distance between the two glove membranes corresponded to the height of the ring. The PVC ring alone, filled with water, and covered with the latex and nitrile gloves are shown in Figure 1B-D.

Each experiment was repeated five times at each nominal power of the equipment (0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, and 7.0 W) without the gloves (reference) and with the two gloves of different materials (latex and nitrile), with and without the height controller (PVC cylinder). All experiments were performed with the same equipment at the same time of day (afternoon) and by one operator, at an ambient temperature of 25 ± 2 °C. Attention was paid to prevent the formation of air bubbles inside the glove when closing it to avoid the decrease of the output power.

Evaluation of output power of the TUS equipment

Using the reference experimental setup (without gloves), the output power was verified to be within the range recommended by IEC 61689 (2007) for each nominal value (± 20%). The error between the measured and the nominal power (expressed as percentage) was calculated using Equation 1:

E r r o r = | ( P M e a s u r e d P N o min a l P N o min a l ) 100 | (1)

Statistical analysis

Statistical analyses were performed with the software BioEstat 5.0 (Ayres et al., 2007Ayres M, Ayres Júnior M, Ayres DL, Santos AS. BioEstat 5.0: aplicações estatísticas nas áreas das ciências biológicas e médicas. 5th ed. Belém: Sociedade Civil Mamirauá/CNPq; 2007.). The Shapiro-Wilk test was used to analyze the normality of the data from the measured power values for each experimental condition. The next step was to examine the ratio between the highest and the lowest standard deviations of the measured power for all nominal values and experimental conditions. If the data do not have a normal distribution and/or the ratio obtained was greater than 2, a nonparametric test (Kruskal-Wallis test) was performed to verify if the output power variations for the different experimental conditions were statistically significant (Moore, 2006Moore DS. The basic practice of statistics. 4th ed. New York: W. H. Freeman & Co LTD; 2006.). When the Kruskal-Wallis test results exhibited significant differences, the Student-Newman-Keuls multiple comparisons post-test was performed. All results were considered statistically significant if p-value <0.05.

Results

The mean and standard deviation of the output power were determined for the five different experimental conditions, which were categorized as follows: A: deionized water without the presence of gloves; B, C: nitrile gloves filled with deionized water with (B) and without (C) height control by the PVC ring; D, E: latex gloves filled with deionized water with (D) and without (E) height control by the PCV ring. The results are shown in Table 1.

Table 1
Nominal power, mean, and standard deviation of the measured power for A (deionized water without glove), B (nitrile glove filled with deionized water and height control by a PCV ring), C (nitrile glove filled with deionized water), D (latex glove filled with deionized water and height control by a PCV ring), and E (latex glove filled with deionized water).

Table 2 shows the relative error (%) of rated output power estimated in condition A with reference to the nominal power.

Table 2
The nominal power (W) and the relative error (%) of measured power for A (deionized water without glove).

The Shapiro-Wilk test was applied to the 40 distribution groups (5 techniques and 8 acoustic power values) of the measured power and determined that four of them did not have a normal distribution (0.5 and 3.0 W for configuration A and 2.0 and 5.0 W for D). Additionally, the ratio between the highest and lowest standard deviations was evaluated to be higher than 2 in four cases (for 1.0, 4.0, 6.0, and 7.0 W). Then, the Kruskal-Wallis test was used to compare the power measurements obtained for all conditions and nominal powers (the null hypothesis of this test is that the means of the A, B, C, D, and E groups are the same). All p-value results calculated using the Kruskal-Wallis test were below 0.01. Therefore, the multiple-comparison Student-Newman-Keuls test was applied. The difference between the average measured power and the results of the multiple-comparison test are presented in Table 3. The superscript indicates that p-value < 0.05 for the compared pair of conditions.

Table 3
Difference between the average measured power and the Student-Newman-Keuls post hoc test results for all conditions (A, B, C, D, and E). The superscript ✓ indicates that p<0.05

Discussion

Ultrasonic transmission from the transducer to the skin depends on the presence of a coupling medium between the transducer surface and the tissue. When the tissue surface is irregular, the transducer contact may be compromised.

On the other hand, a coupling medium with a high absorption coefficient, with air bubbles, or agents that alter the energy transmission is not recommended. Furthermore, the presence of a coupling agent facilitates the matching of the acoustic impedances of the tissue and transducer (Watson, 2008bWatson T. Ultrasound in contemporary physiotherapy practice. Ultrasonics. 2008b; 48(4):321-9. PMid:18466945. http://dx.doi.org/10.1016/j.ultras.2008.02.004.
http://dx.doi.org/10.1016/j.ultras.2008....
).

The water bag technique is an indirect-contact and low-cost US coupling approach. Its use is very common in everyday clinical practice, especially for the treatment of anatomically uneven surfaces (Guirro and Guirro, 2002Guirro E, Guirro R. Fisioterapia dermato-funcional: fundamentos, recursos e patologias. 3rd ed. São Paulo: Manole; 2002.; Guirro et al., 2001Guirro R, Cancelieri AS, Sant’Anna IL. Avaliação dos meios intermediários utilizados na aplicação do ultra-som terapêutico. Rev Bras Fisiot. 2001; 5(2):49-52.; Starkey, 2013Starkey C. Therapeutic modalities. 4th ed. Philadelphia: F.A. Davis; 2013.). However, scientific studies investigating this technique are scarce (Guirro et al., 2001Guirro R, Cancelieri AS, Sant’Anna IL. Avaliação dos meios intermediários utilizados na aplicação do ultra-som terapêutico. Rev Bras Fisiot. 2001; 5(2):49-52.; Klucinec et al., 2000bKlucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Transmissivity of coupling agents used to deliver ultrasound through indirect methods. The Journal of Orthopaedic and Sports Physical Therapy. 2000b; 30(5):263-6. PMid:10817413. http://dx.doi.org/10.2519/jospt.2000.30.5.263.
http://dx.doi.org/10.2519/jospt.2000.30....
).

The study conducted by Poltawski and Watson (2007)Poltawski L, Watson T. Relative transmissivity of ultrasound coupling agents commonly used by therapists in the UK. Ultrasound in Medicine & Biology. 2007; 33(1):120-8. PMid:17189054. http://dx.doi.org/10.1016/j.ultrasmedbio.2006.07.026.
http://dx.doi.org/10.1016/j.ultrasmedbio...
on the performance of different coupling agents (gels) on US transmission showed differences in the energy transmission of ultrasonic waves that had no clinical significance.

Klucinec et al. (2000b)Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Transmissivity of coupling agents used to deliver ultrasound through indirect methods. The Journal of Orthopaedic and Sports Physical Therapy. 2000b; 30(5):263-6. PMid:10817413. http://dx.doi.org/10.2519/jospt.2000.30.5.263.
http://dx.doi.org/10.2519/jospt.2000.30....
demonstrated a reduction in the acoustic energy transfer when degassed and tap water in latex gloves, gel in a latex glove, and a gel-filled condom were used as coupling media. The same effect was observed in this work for conditions B, C, D, and E for all nominal powers except 0.5 W (see Table 1). Generally, the measured powers in the control configuration (A) did not show significant differences from those obtained in E, with the exception of the nominal power of 0.5 W. The mean difference between group pairs exhibited a general tendency to increase as the nominal power increased.

Guirro et al. (2001)Guirro R, Cancelieri AS, Sant’Anna IL. Avaliação dos meios intermediários utilizados na aplicação do ultra-som terapêutico. Rev Bras Fisiot. 2001; 5(2):49-52. indicated that the coupling means (balloon and latex glove filled with hydro-soluble gel) employed in their experiments presented a high attenuation coefficient, i.e., had low transmissivity. Nevertheless, the present study shows (Table 1) that the highest difference between the mean power values of A and those measured in configurations containing a glove (B, C, D, or E) was below 15.0%, irrespective of the nominal power. In addition, the smallest mean power difference was obtained between A and E (<3.0%), regardless of the nominal power.

Comparison of conditions A, B and C

The significant difference between the power values measured at different experimental conditions (A-B and/or A-C) that was observed in some situations (nominal power 4-7 W) is noteworthy because, in some cases, the averages of the compared pairs differed by more than 0.45 W. This was observed for the nominal powers 4.0 W (A-B), 5.0 W (A-B and A-C), 6.0 W (A-B and A-C), and 7.0 W (A-B and A-C). When comparing the measurements obtained in B and C configurations, the highest power difference was 0.16 W for a nominal power of 4 W, which is not clinically relevant (Itakura et al., 2012Itakura DA, Magas V, Neves EB, Nohama P. Alteração da temperatura nos tecidos biológicos com a aplicação do ultrassom terapêutico: uma revisão. Fisioterapia em Movimento. 2012; 25(4):857-68. http://dx.doi.org/10.1590/S0103-51502012000400019.
http://dx.doi.org/10.1590/S0103-51502012...
).

In summary, the experimental results exhibited no clinical difference for the nitrile gloves with or without controlled height.

Comparison of conditions A, D and E

The results of the pairwise comparison test showed that only the output power measured with the experimental condition pair A-E was not significantly different (except for the nominal power of 0.5 W). Nevertheless, the difference between the averages was below 0.19 W, which is clinically irrelevant. The corresponding difference for the pair A-D was higher than 0.5 W for nominal powers of 6 and 7 W. When comparing the measurements obtained in D and E, the difference between the averages reached 0.36 W for nominal power 6 W, which is not clinically relevant. In other words, for latex gloves, there no clinically relevant difference was observed between the measured powers, with or without height control.

Comparison of conditions B, C, D and E

No significant difference was observed for the experimental conditions B and D (except for nominal powers 4.0 and 5.0 W). Nevertheless, the mean power differences were not clinically relevant. On the other hand, significant difference was observed for the experimental conditions C and E for power values 6 and 7 W.

General comments

The reduction of the output power and the differences observed in the study can be interpreted by phenomena such as cavitation and standing waves and by considering the intrinsic factors of the coupling media, such as the thickness of the coupling agent layer, the positioning of the glove membrane, the target size, and beam divergence (Casarotto et al., 2004Casarotto RA, Adamowski JC, Fallopa F, Bacanelli F. Coupling agents in therapeutic ultrasound: acoustic and thermal behavior. Archives of Physical Medicine and Rehabilitation. 2004; 85(1):162-5. PMid:14970985. http://dx.doi.org/10.1016/S0003-9993(03)00293-4.
http://dx.doi.org/10.1016/S0003-9993(03)...
; Poltawski and Watson, 2007Poltawski L, Watson T. Relative transmissivity of ultrasound coupling agents commonly used by therapists in the UK. Ultrasound in Medicine & Biology. 2007; 33(1):120-8. PMid:17189054. http://dx.doi.org/10.1016/j.ultrasmedbio.2006.07.026.
http://dx.doi.org/10.1016/j.ultrasmedbio...
; Shaw and Hodnett, 2008Shaw A, Hodnett M. Calibration and measurement issue for therapeutic ultrasound. Ultrasonics. 2008; 48(4):234-52. PMid:18234261. http://dx.doi.org/10.1016/j.ultras.2007.10.010.
http://dx.doi.org/10.1016/j.ultras.2007....
; Wells, 1977Wells PNT. Biomedical ultrasonics. London: Academic Press; 1977.).

Moreover, in the configuration containing a glove filled with deionized water and height control by the PVC ring (B and D), reverberation can occur in the presence of two parallel membranes, causing destructive interference of the wave, which can reduce the power of the wave in the region of interest.

Another factor that can reduce US power is the distance between the treatment region and the TUS transducer. Therefore, a power measurement was conducted to compare a glove filled with deionized water without controlled height (h ≈ 31.0 mm) and one with controlled height (h = 12.8 mm). However, the mentioned factor has not been proven by experiments, for the same kind of gloves. Only in one case (B-C for nominal power of 2.0 W) the average of rated output power was significantly different. Nevertheless, the difference between the averages of D-E is greater than that of B-C. According to recommendations (Low and Reed, 2000Low J, Reed A. Electrotherapy explained: principles and practice. 3rd ed. Oxford: Butterworth Heinemann; 2000.; Wells, 1977Wells PNT. Biomedical ultrasonics. London: Academic Press; 1977.), all experiments were performed with the 45° air-backed cone target positioned in the near field (a2/λ=1.242/0.1510.3cm) because the treatment area is recommended to be placed in this region.

Although the relative error of the rated output power is within the range recommended by IEC 61689 (International…, 2007), the observed variation of the measured acoustic power can be also attributed to the bath immersion technique. Klucinec et al. (2000b)Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Transmissivity of coupling agents used to deliver ultrasound through indirect methods. The Journal of Orthopaedic and Sports Physical Therapy. 2000b; 30(5):263-6. PMid:10817413. http://dx.doi.org/10.2519/jospt.2000.30.5.263.
http://dx.doi.org/10.2519/jospt.2000.30....
showed that transmissivity is reduced when this technique is employed.

Similarly to latex, nitrile is a type of rubber. Because of the different chemical compositions of these materials, their attenuation coefficients also differ. For this reason, the rated output powers measured using latex and nitrile gloves have statistically different values.

As household water is known to contain several diluted gases, which facilitates the occurrence of cavitation (Kenyon and Kenyon, 2009Kenyon K, Kenyon J. The physiotherapist’s pocketbook: essential facts at your fingertips. 2nd ed. London: Churchill Livingstone Elsevier; 2009.), in this study, only deionized and degassed water was used to minimize this phenomenon. Deionized water was placed in the vacuum chamber to extract the diluted gases and thus avoid the formation of bubbles inside the glove. Although this does not realistically reflect the practices followed in clinics, it serves as a reference. If air bubbles form in deionized water, then, the use of tap water would induce cavitation. To exemplify this phenomenon, a glove was filled with tap water and irradiated with a TUS frequency of 1 MHz and power of 7 W for 5 min using a clinical gel as the coupling medium between the transducer and the glove. Even small cavitation can form microbubbles, which can assemble to create a bubble (see Figure 2). Microbubbles (wavelength lower than 1.5 mm) generate scattering of the US wave, and bubbles induce reflection of the incident wave. In both cases, there is a reduction in the power of the transmitted wave.

Figure 2
Formation of air bubble cluster, indicated by the arrow, on a latex glove filled with common tap water. The coupling gel was removed for a better view of the bubbles.

In the experimental condition shown in Figure 2, the bubble formation caused severe attenuation, leading to acoustic power measurements below the nominal power on several occasions.

Consequently, as deionized and degassed water are not commonly available in clinics, the use of the water bag technique in clinical practices can lead to ineffective therapy. When tap water is employed, cavitation occurs to the point of clear bubble formation, considerably decreasing the measured power to values below the detection limit of the radiation force balance. Otherwise, since the transducer does not stay still, but the physician moves it around the body region, there must exist another biophysical effect, which needs future research. A possible explanation may be that ultrasonic beam will not be totally reflected during the irradiation.

An interesting effect that has not been sufficiently studied is the increase in the average power difference of the compared techniques with increasing nominal power, which was observed in the values of Table 3.

The highest difference between the rated output powers (0.78 W) was obtained for the pair A-B and nominal power 7 W. From the clinical point view, the important parameter is intensity (power/ERA). Therefore, we can estimate the intensity by using the nominal value of the ERA of the ultrasonic transducer, 3.8 cm2, and the highest difference between the intensities is 0.21 W/cm2 (0.78 W/3.8 cm2). In our study, the ERA was not measured; as it is usually smaller than the nominal ERA, we assumed the extreme case in which the real ERA was 50% lower than the nominal one. In this case, the highest difference between the intensities was 0.41 W/cm2 (0.78 W/1.9 cm2), which is lower than 0.5 W/cm2. From the clinical point of view, intensity differences below 0.5 W/cm2 are not considered relevant; thus, conditions A, B, C, D, and E are equivalent.

In summary, the experimental results showed that the water bag technique with a latex glove filled with deionized and degassed water (E) can be used as a coupling agent. This is because the statistical results indicated that the power values measured with this technique and the control (A) were not statistically significant and the difference between the averages of the power for these methods was below 0.19 W. In addition, this technique is a viable option to treat irregular anatomic regions, because deionized and degassed water is used in the process. The use of common tap water can increase cavitation and US attenuation, which can compromise the success of the therapy. Additionally, it is advisable to avoid the formation of air bubbles inside the glove when closing it, as this could cause greater attenuation of the ultrasonic wave beam, reducing the effective power and treatment effectiveness. On the other hand, as intensity differences lower than 0.5 W/cm2 are considered clinically irrelevant, conditions A, B, C, D, and E can be used interchangeably.

Oshikoya et al. (2000)Oshikoya CA, Shultz SJ, Mistry D, Perrin DH, Arnold BL, Gansneder BM. Effect of coupling medium temperature on rate of intramuscular temperature rise using continuous ultrasound. Journal of Athletic Training. 2000; 35(4):417-21. PMid:16558655. observed that the temperature of the coupling medium may reduce the measured power, which may influence the treatment effectiveness. However, throughout the present experiment, both the ambient and the water temperature were 25.0 ± 0.4 °C and the measured power could not be affected by the temperature of the coupling medium.

Further studies must be performed on this topic employing TUS equipment of different manufacturers.

Acknowledgements

The authors thank FAPERJ, CNPq, and CAPES for financial support.

  • How to cite this article: Lima LS, Oliveira DP, Costa-Júnior JFS, Pinto PA, Omena TP, Costa RM, von Krüger MA, Pereira WCA. Evaluation of gloves as a water bag coupling agent for therapeutic ultrasound. Res Biomed Eng. 2017. DOI: 10.1590/2446-4740.01816.

References

  • Artho PA, Thyne JG, Warring BP, Willis CD, Brismée JM, Latman NS. A calibration study of therapeutic ultrasound units. Physical Therapy. 2002; 82(3):257-63. PMid:11869154.
  • Ayres M, Ayres Júnior M, Ayres DL, Santos AS. BioEstat 5.0: aplicações estatísticas nas áreas das ciências biológicas e médicas. 5th ed. Belém: Sociedade Civil Mamirauá/CNPq; 2007.
  • Baker KG, Robertson VJ, Duck FA. A review of therapeutic ultrasound: biophysical effects. Physical Therapy. 2001; 81(7):1351-8. PMid:11444998.
  • Cameron MH. Physical agents in rehabilitation: from research to practice. 4th ed. St. Louis: Saunders Elsevier; 2012.
  • Casarotto RA, Adamowski JC, Fallopa F, Bacanelli F. Coupling agents in therapeutic ultrasound: acoustic and thermal behavior. Archives of Physical Medicine and Rehabilitation. 2004; 85(1):162-5. PMid:14970985. http://dx.doi.org/10.1016/S0003-9993(03)00293-4
    » http://dx.doi.org/10.1016/S0003-9993(03)00293-4
  • Delforge G. Musculoskeletal trauma: implications for sports injury management. Champaign: Human Kinetics; 2002.
  • Guirro E, Guirro R. Fisioterapia dermato-funcional: fundamentos, recursos e patologias. 3rd ed. São Paulo: Manole; 2002.
  • Guirro R, Cancelieri AS, Sant’Anna IL. Avaliação dos meios intermediários utilizados na aplicação do ultra-som terapêutico. Rev Bras Fisiot. 2001; 5(2):49-52.
  • International Electrotechnical Commission. IEC 61689:2007. Ultrasonics: physiotherapy systems: field specifications and methods of measurement in the frequency range 0.5 MHz to 5 MHz. Geneva: IEC; 2007.
  • Itakura DA, Magas V, Neves EB, Nohama P. Alteração da temperatura nos tecidos biológicos com a aplicação do ultrassom terapêutico: uma revisão. Fisioterapia em Movimento. 2012; 25(4):857-68. http://dx.doi.org/10.1590/S0103-51502012000400019
    » http://dx.doi.org/10.1590/S0103-51502012000400019
  • Kenyon K, Kenyon J. The physiotherapist’s pocketbook: essential facts at your fingertips. 2nd ed. London: Churchill Livingstone Elsevier; 2009.
  • Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Effectiveness of wound care products in the transmission of acoustic energy. Physical Therapy. 2000a; 80(5):469-76. PMid:10792857.
  • Klucinec B, Scheidler M, Denegar C, Domholdt E, Burgess S. Transmissivity of coupling agents used to deliver ultrasound through indirect methods. The Journal of Orthopaedic and Sports Physical Therapy. 2000b; 30(5):263-6. PMid:10817413. http://dx.doi.org/10.2519/jospt.2000.30.5.263
    » http://dx.doi.org/10.2519/jospt.2000.30.5.263
  • Kollmann C, Vacariu G, Schuhfried O, Fialka-Moser V, Bergmann H. Variations in the output power and surface heating effects of transducers in therapeutic ultrasound. Archives of Physical Medicine and Rehabilitation. 2005; 86(7):1318-24. PMid:16003657. http://dx.doi.org/10.1016/j.apmr.2005.02.001
    » http://dx.doi.org/10.1016/j.apmr.2005.02.001
  • Low J, Reed A. Electrotherapy explained: principles and practice. 3rd ed. Oxford: Butterworth Heinemann; 2000.
  • Moore DS. The basic practice of statistics. 4th ed. New York: W. H. Freeman & Co LTD; 2006.
  • Nanda BK. Electrotherapy simplified. New Delhi: Jaypee Brothers Medical; 2008.
  • Oshikoya CA, Shultz SJ, Mistry D, Perrin DH, Arnold BL, Gansneder BM. Effect of coupling medium temperature on rate of intramuscular temperature rise using continuous ultrasound. Journal of Athletic Training. 2000; 35(4):417-21. PMid:16558655.
  • Poltawski L, Watson T. Relative transmissivity of ultrasound coupling agents commonly used by therapists in the UK. Ultrasound in Medicine & Biology. 2007; 33(1):120-8. PMid:17189054. http://dx.doi.org/10.1016/j.ultrasmedbio.2006.07.026
    » http://dx.doi.org/10.1016/j.ultrasmedbio.2006.07.026
  • Prentice WE. Therapeutic modalities in rehabilitation. 3rd ed. Nova York: McGraw-Hill Medical; 2005.
  • Pye S. Ultrasound therapy equipment: does it perform? Physiotherapy. 1996; 82(1):39-44. http://dx.doi.org/10.1016/S0031-9406(05)66996-9
    » http://dx.doi.org/10.1016/S0031-9406(05)66996-9
  • Robertson V, Ward A, Low J, Reed A. Electrotherapy explained: principles and practice. 4th ed. Oxford: Butterworth-Heinemann Elsevier; 2006.
  • Shaw A, Hodnett M. Calibration and measurement issue for therapeutic ultrasound. Ultrasonics. 2008; 48(4):234-52. PMid:18234261. http://dx.doi.org/10.1016/j.ultras.2007.10.010
    » http://dx.doi.org/10.1016/j.ultras.2007.10.010
  • Shung KK. Diagnostic ultrasound: imaging and blood flow measurements. Boca Raton: CRC Press; 2006.
  • Starkey C. Therapeutic modalities. 4th ed. Philadelphia: F.A. Davis; 2013.
  • Watson T. Electrotherapy: evidence-based practice. 12th ed. Philadelphia: Churchill Livingstone Elsevier; 2008a.
  • Watson T. Ultrasound in contemporary physiotherapy practice. Ultrasonics. 2008b; 48(4):321-9. PMid:18466945. http://dx.doi.org/10.1016/j.ultras.2008.02.004
    » http://dx.doi.org/10.1016/j.ultras.2008.02.004
  • Wells PNT. Biomedical ultrasonics. London: Academic Press; 1977.