A wavelet-based method for power-line interference removal in ECG signals

Abstract

Introduction

The analysis of electrocardiogram (ECG) signals allows the experts to diagnosis several cardiac disorders. However, the accuracy of such diagnostic depends on the signals quality. In this paper it is proposed a simple method for power-line interference (PLI) removal based on the wavelet decomposition, without the use of thresholding techniques.

Methods

This method consists in identifying the ECG and noise frequency range for further zeroing wavelet detail coefficients in the subbands with no ECG coefficients in the frequency content. Afterward, the enhanced ECG signal is obtained by the inverse discrete wavelet transform (IDWT). In order to choose the wavelet function, several experiments were performed with synthetic signals with worse Signal-to-Noise Ratio (SNR).

Results

Considering the relative error metrics and runtime, the best wavelet function for denoising was Symlet 8. Twenty synthetic ECG signals with different features and eight real ECG signals, obtained in the Physionet Challenge 2011, were used in the experiments. Results show the advantage of the proposed method against thresholding and notch filter techniques, considering classical metrics of assessment. The proposed method performed better for 75% of the synthetic signals and for 100% of the real signals considering most of the evaluation measures, when compared with a thresholding technique. In comparison with the notch filter, the proposed method is better for all signals.

Conclusion

The proposed method can be used for PLI removal in ECG signals with superior performance than thresholding and notch filter techniques. Also, it can be applied for high frequencies denoising even without a priori frequencies knowledge.

Keywords
Power-line interference; Denoising ECG signals; Wavelet decomposition

Introduction

ECG signals represent the heart electrical activity and they are obtained through electrodes placed in specific regions of the human body. There are five main waves characterizing ECG signals: P, Q, R, S, and T. Each wave or complex has an exclusive significance. The combination of Q, R and S waves forms which is known as QRS complex. The P wave and the QRS complex represent the atrial and ventricular depolarization, respectively. On the other hand, the T wave characterizes the ventricular repolarization. The spectrum of the QRS complex is located in the ECG frequency bands whose typical frequency components range from 10 Hz to 25 Hz ( Köhler et al., 2002 Köhler BU, Hennig C, Orglmeister R. The principles of software QRS detection. IEEE Eng Med Biol Mag. 2002; 21(1):42-57. http://dx.doi.org/10.1109/51.993193. PMid:11935987.
http://dx.doi.org/10.1109/51.993193 ...
) and its detection is still subject of many studies. Pan and Tompkins (1985) Pan J, Tompkins WJ. A real-time QRS detection algorithm. IEEE Trans Biomed Eng. 1985; 32(3):230-6. http://dx.doi.org/10.1109/TBME.1985.325532. PMid:3997178.
http://dx.doi.org/10.1109/TBME.1985.325...
developed an objective QRS detection algorithm using a band-pass filter from 5 Hz to 12 Hz. Zidelmal et al. (2012) Zidelmal Z, Amirou A, Adnane M, Belouchrani A. QRS detection based on wavelet coefficients. Comput Methods Programs Biomed. 2012; 107(3):490-6. http://dx.doi.org/10.1016/j.cmpb.2011.12.004. PMid:22296976.
http://dx.doi.org/10.1016/j.cmpb.2011.1...
computed the power spectrum for four QRS types concluding that their energies are concentrated in the range from 5 Hz to 22 Hz. Such spectral information is used to QRS complexes detection in Challenge 2011 (Training Set A) database ( Oliveira et al., 2015 Oliveira BR, Duarte MAQ, Vieira Filho J. Detecção de complexos QRS do ECG pela decomposição em valores singulares em multirresolução. In: Anais da IX ENAMA; 2015 Nov 4-6; Cascavel, Brasil; São Carlos: SBMAC; 2015. p. 143-4. ). In addition, it is important to remark that ECG signals are non-stationary, non-symmetric in relation to the x -axis and are originally impulsive signals ( Łęski and Henzel, 2005 Łęski JM, Henzel N. ECG baseline wander and powerline interference reduction using nonlinear filter bank. Signal Process. 2005; 85(4):781-93. http://dx.doi.org/10.1016/j.sigpro.2004.12.001.
http://dx.doi.org/10.1016/j.sigpro.2004...
).

The specific morphology of the ECG signals allows identifying various cardiac diseases. However, for an accurate analysis, signals should have high Signal-to-Noise Ratio (SNR) ( Łęski and Henzel, 2005 Łęski JM, Henzel N. ECG baseline wander and powerline interference reduction using nonlinear filter bank. Signal Process. 2005; 85(4):781-93. http://dx.doi.org/10.1016/j.sigpro.2004.12.001.
http://dx.doi.org/10.1016/j.sigpro.2004...
). Low SNR can difficult the analysis performed by experts or computational applications, since it changes the signal waveform. Typical noise present in ECG signals are due to power-line interference in a frequency band varying from 50 Hz to 60 Hz ( Łęski and Henzel, 2005 Łęski JM, Henzel N. ECG baseline wander and powerline interference reduction using nonlinear filter bank. Signal Process. 2005; 85(4):781-93. http://dx.doi.org/10.1016/j.sigpro.2004.12.001.
http://dx.doi.org/10.1016/j.sigpro.2004...
; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
; Rahman et al., 2010 Rahman MZU, Shaik RA, Reddy DVRK. Baseline wander and power line interference elimination from cardiac signals using error nonlinearity LMS algorithm. In: International Conference on Systems in Medicine and Biology (ICSMB). 2010 Dec 16-18; Kharagpur, India; New Jersey: IEEE; 2010. p. 217-20. ) depending on the country. It occurs due to interferences of electrical equipment as X-ray, air conditioners, elevators ( Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
), and also due to the differences in electrode impedances ( Bahoura and Ezzaidi, 2010 Bahoura M, Ezzaidi H. FPGA-implementation of wavelet-based denoising technique to remove power-line interference from ECG signal. Inf Technol Appl Biomed (ITAB). In: Proceedings of the 10th IEEE International Conference; 2010 Nov 3-5; Corfu, Greece. New Jersey: IEEE; 2010. p. 1-4. ).

Several researchers have proposed denoising approaches to enhance ECG signals and preserve their original characteristics. Some noise reduction techniques are based on digital filters, wavelet transform and adaptive filtering ( AlMahamdy and Riley, 2014 AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
http://dx.doi.org/10.1016/j.procs.2014....
); singular value decomposition ( Bandarabadi and Karami-Mollaei, 2010 Bandarabadi AAJGM, Karami-Mollaei MR. ECG denoising using singular value decomposition. Aust J Basic Appl Sci. 2010; 4(7):2109-13. ); independent component analysis ( Phegade and Mukherji, 2013 Phegade M, Mukherji P. ICA based ECG signal denoising. In: Proceedings of the International Conference on Advances in Computing, Communications and Informatics (ICACCI); 2013 Aug 22-25; Mysore, India; New Jersey: IEEE; 2013. p. 1675-80. ) and S-transform ( Das and Ari, 2013 Das M, Ari S. Analysis of ECG signal denoising method based on s-transform. IRBM. 2013; 34(6):362-70. http://dx.doi.org/10.1016/j.irbm.2013.07.012.
http://dx.doi.org/10.1016/j.irbm.2013.0...
). Among the algorithms for PLI removal there are digital processing methods based on: fuzzy thresholding ( Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
); nonlinear filter bank ( Łęski and Henzel, 2005 Łęski JM, Henzel N. ECG baseline wander and powerline interference reduction using nonlinear filter bank. Signal Process. 2005; 85(4):781-93. http://dx.doi.org/10.1016/j.sigpro.2004.12.001.
http://dx.doi.org/10.1016/j.sigpro.2004...
); Fast Fourier Transform and adaptive nonlinear noise estimator ( Shirbani and Setarehdan, 2013 Shirbani F, Setarehdan SK. ECG power line interference removal using combination of FFT and adaptive non-linear noise estimator. In: Proceedings of the 21st Iranian Conference on Electrical Engineering (ICEE). 2013 May 14-16. Mashhad, Iran. New Jersey: IEEE; 2013. p. 1-5. http://dx.doi.org/10.1109/IranianCEE.2013.6599622.
http://dx.doi.org/10.1109/IranianCEE.20...
); Empirical Mode Decomposition ( Agrawal and Gupta, 2013 Agrawal S, Gupta A. Fractal and EMD based removal of baseline wander and powerline interference from ECG signals. Comput Biol Med. 2013; 43(11):1889-99. http://dx.doi.org/10.1016/j.compbiomed.2013.07.030. PMid:24209934.
http://dx.doi.org/10.1016/j.compbiomed....
); neural networks ( Mateo et al., 2008 Mateo J, Sanchez C, Tortes A, Cervigon R, Rieta JJ. Neural network based canceller for powerline interference in ECG signals. In: Proceedings of the 35th Annual Computers in Cardiology Conference (CinC); 2008 Sep 14-17; Bolongna, Italy. New Jersey: IEEE; 2008. p. 1073-76. http://dx.doi.org/10.1109/CIC.2008.4749231.
http://dx.doi.org/10.1109/CIC.2008.4749...
) and wavelet transform ( Agrawal and Gupta, 2013 Agrawal S, Gupta A. Fractal and EMD based removal of baseline wander and powerline interference from ECG signals. Comput Biol Med. 2013; 43(11):1889-99. http://dx.doi.org/10.1016/j.compbiomed.2013.07.030. PMid:24209934.
http://dx.doi.org/10.1016/j.compbiomed....
; Garg et al., 2011 Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
; Poornachandra and Kumaravel, 2008 Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
; Rahman et al., 2010 Rahman MZU, Shaik RA, Reddy DVRK. Baseline wander and power line interference elimination from cardiac signals using error nonlinearity LMS algorithm. In: International Conference on Systems in Medicine and Biology (ICSMB). 2010 Dec 16-18; Kharagpur, India; New Jersey: IEEE; 2010. p. 217-20. ).

Wavelet analysis has been successfully used for ECG signal denoising because it deals well with non-stationary signals and also presents better resolution in time-frequency domain than Fourier analysis ( Rahman et al., 2010 Rahman MZU, Shaik RA, Reddy DVRK. Baseline wander and power line interference elimination from cardiac signals using error nonlinearity LMS algorithm. In: International Conference on Systems in Medicine and Biology (ICSMB). 2010 Dec 16-18; Kharagpur, India; New Jersey: IEEE; 2010. p. 217-20. ). In the comparative study presented by AlMahamdy and Riley (2014) AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
http://dx.doi.org/10.1016/j.procs.2014....
, the wavelet transform produced better results in most of the experiments. Chouakri et al. (2006) Chouakri SAS, Bereksi-Reguig AF, Fokapu O. ECG signal smoothing based on combining wavelet denoising levels. Asian J Inf Technol. 2006; 5(6):666-77. compared the performance of Butterworth filters and the multilevel wavelet transform, concluding that improved results were achieved by the wavelet technique. Usually, wavelet-based methods for ECG denoising use thresholding techniques with some additional processing ( Agante and Sa, 1999 Agante PM, Sa JPM. ECG noise filtering using wavelets with soft-thresholding methods. Comput Cardiol. 1999; 26:535-8. ; AlMahamdy and Riley, 2014 AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
http://dx.doi.org/10.1016/j.procs.2014....
; Awal et al., 2014 Awal MA, Mostafa SS, Ahmad M, Rashid MA. An adaptive level dependent wavelet thresholding for ECG denoising. Biocybern Biol Eng. 2014; 34(4):238-49. ; Bahoura and Ezzaidi, 2010 Bahoura M, Ezzaidi H. FPGA-implementation of wavelet-based denoising technique to remove power-line interference from ECG signal. Inf Technol Appl Biomed (ITAB). In: Proceedings of the 10th IEEE International Conference; 2010 Nov 3-5; Corfu, Greece. New Jersey: IEEE; 2010. p. 1-4. ; Chouakri et al., 2006 Chouakri SAS, Bereksi-Reguig AF, Fokapu O. ECG signal smoothing based on combining wavelet denoising levels. Asian J Inf Technol. 2006; 5(6):666-77. ; Garg et al., 2011 Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
; Germán-Salló, 2010 Germán-Salló Z. Nonlinear filtering in ECG signal denoising. Acta Univ Sapientiae Elec Mech Eng. 2010; 2:136-45. ; Karthikeyan et al., 2012 Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306. ; Li et al., 2009 Li S, Liu G, Lin Z. Comparisons of wavelet packet, lifting wavelet and stationary wavelet transform for de-noising ECG. In: Proceedings of the 2nd IEEE International Conference on Computer Science and Information Technology (ICCSIT 2009); 2009 Aug 8-11; Beijing, China; New Jersey: IEEE; 2009. p. 491-4. ; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
; Poornachandra and Kumaravel, 2008 Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
; Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
). Patil and Chavan (2012) Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
compared the PLI removal for different wavelet basis using hard and soft shrinkage functions. They conclude that hard thresholding achieves better SNR scores than soft thresholding, and the best wavelet basis depends on the analyzed signal.

Garg et al. (2011) Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
worked on optimal wavelet-based algorithm for ECG denoising, analyzing SNR for several wavelet families, decomposition level and threshold selection method. In order to calculate the threshold, four rules were used: min-max, rigorous sure, universal and heuristic sure. The best configuration was achieved with the Symlet wavelet with ten vanishing moments and five decomposition levels, hard shrinkage function and heuristic sure rule or rigorous sure thresholding. Poornachandra and Kumaravel (2008) Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
proposed to use of hyper shrinkage function in the subbands that contained the PLI noise at some decomposition level. The obtained results were better when compared to those of the state-of-the-art algorithms.

In this paper, it is proposed to use discrete wavelet transform (DWT) to decompose an ECG signal degraded by high power-line interference. The goal is to have the ECG signal represented by the approximation coefficients and the noise by the detail coefficients. The basic idea is to inspect the ECG and noise frequency range in each subband of the wavelet filter bank. The wavelet scale whose frequency range exceeds the maximum frequency of the ECG signals is set to zero. Then, the IDWT is applied in order to obtain a better quality signal without the use of thresholding techniques. It is common that such techniques do not completely eliminate noise, generating residual noise that can still distorts the QRS waveform. Therefore, this original ECG analysis methodology eliminates the need of thresholding function and is based solely on wavelet filter bank and the characteristics of PLI. Comparisons with a thresholding technique and a classical digital filter were carried out to demonstrate the effectiveness of the proposed method. Furthermore, the proposed method presents low computational load, reduces the residual noise and can be easily implemented.

Methods

A noisy ECG, s(t) , contaminated by PLI can be represented as follows:

s(t)=x(t)+αcos(2πfrt+φ) , (1)

where t is time and fr=50 Hz or fr= 60 Hz is the PLI frequency. The unknown parameters are φ [0, 2π] , PLI signal phase and α > 0 , the amplitude. The clean ECG is represented by x(t) .

Dynamical model for generating synthetic ECG signals

In order to analyze ECG denoising algorithms performances, many metrics based on comparisons between the estimated signal and the original one have been proposed. Real ECG signals may have other noise besides PLI. To focus only on this kind of noise, synthetic ECG signals are used in some experiments. The mathematical model for generating such signals was given by McSharry et al. (2003) McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
http://dx.doi.org/10.1109/TBME.2003.808...
. They proposed a dynamical model that generates a trajectory in the three-dimensional state-space given by three coupled ordinary differential equations (ODE). The displacement around the “attracting limit cycle of unit radius” describes the ECG signal for each RR-interval ( McSharry et al., 2003 McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
http://dx.doi.org/10.1109/TBME.2003.808...
). ECG characteristic waves are events with fixed angles in relation to the unit circle, given by θi , for i {P, Q, R, S, T} . The ODE model ( McSharry et al., 2003 McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
http://dx.doi.org/10.1109/TBME.2003.808...
) is given by y˙=(1x2+y2)y+ρx , x˙=(1x2+y2)xρx and z˙=iaiΔθiexp(Δθi22bi2)[z0.15sin(2πf2)] , where θ=atan2(x, y) and atan2(x, y) is the element-by-element arctangent between x and y arrays, Δθi=(θθi) , ρ is the angular trajectory velocity and f2 is the respiratory frequency for modeling the baseline wander. z˙ is the generated ECG itself.

This dynamical model yields realistic signals when compared to real ECG signals ( McSharry et al., 2003 McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
http://dx.doi.org/10.1109/TBME.2003.808...
). By setting parameters ai, bi, θi and f2 , different morphologies can be generated for ECG characteristic waves. Note that z˙ is a superposition of sinusoidal and Gaussian functions. Therefore, parameters ai and bi affect the amplitude and the width of the simulated ECG waves, respectively.

Some specific features of the heart rate can also be set, such as mean and standard deviation, besides spectral properties. Generated waveforms are similar to the 12-lead ECG lead I. Although it is possible to produce multi-lead signals ( McSharry et al., 2003 McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
http://dx.doi.org/10.1109/TBME.2003.808...
), in this work they were not considered.

In order to set parameters of the proposed method and perform experiments for validation, twenty synthetic signals were generated using the described model. Model parameters and features of the generated ECG signals are summarized in Table 1 and Table 2 . All signals were sampled at fs = 500 Hz, in an interval of 60 seconds (30001 samples). Parameters were set for the dynamical model in order to obtain ECG signals with different heart rate, SNR and PLI values. By using different settings, it was possible to obtain ECG waveforms with different lengths, amplitudes and fiducial points. Noise was added to each synthetic signal according to equation (1) , considering the parameters described in Table 1 and Table 2 . Such parameters simulate different PLI noises which can be found in a real ambulatory ( Huhta and Webster, 1973 Huhta JC, Webster JG. 60-Hz Interference in Electrocardiogram. IEEE Trans Biomed Eng. 1973; 20(2):91-101. http://dx.doi.org/10.1109/TBME.1973.324169. PMid:4688314.
http://dx.doi.org/10.1109/TBME.1973.324...
), unlike the most state-of-the-art methods which focus on pure sinusoidal noise. So, in addition to the fundamental frequency, harmonics interferences were also considered ( Costa and Tavares, 2009 Costa MH, Tavares MC. Removing harmonic power line interference from biopotential signals in low cost acquisition systems. Comput Biol Med. 2009; 39(6):519-26. http://dx.doi.org/10.1016/j.compbiomed.2009.03.004. PMid:19376509.
http://dx.doi.org/10.1016/j.compbiomed....
).

Table 1
ECG synthetic signals configuration using the model described in this section. Parameters: fr is PLI frequency; φ and α are noise phase and noise amplitude, respectively; other parameters refer to the dynamical model.
Table 2
ECG synthetic signals configuration using the model described in this section. Parameters: fr is PLI frequency; φ and α are noise phase and noise amplitude, respectively; other parameters refer to the dynamical model.

Real ECG signals

In order to validate the proposed method using real ECG signals that have been originally corrupted by PLI, the Challenge 2011 (Training Set A) database from Physionet was chosen ( Goldberger et al., 2000 Goldberger AL, Amaral LAN, Glass L, Hausdorff JM, Ivanov PC, Mark RG, Mietus JE, Moody GB, Peng CK, Stanley HE. PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation. 2000; 101(23):E215-20. https://doi.org/10.1161/01.CIR.101.23.e215. PMid:10851218.
https://doi.org/10.1161/01.CIR.101.23.e...
). Their records were sampled at 500 Hz with 16 -bit resolution, during 10 seconds, for standard 12-lead (leads I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6) and their parameters are summarized in Table 3 . Note from Table 3 that Power Spectral Density (PSD) column is the sum of the power spectrum density only for frequencies over 25 Hz, and it is expressed as PSD×104 decibels (dB).

Table 3
Parameters of the real ECG signals.

Discrete wavelet transform

The wavelet analysis has been applied to various problems in biomedical engineering including noise removal in ECG signals ( Agante and Sa, 1999 Agante PM, Sa JPM. ECG noise filtering using wavelets with soft-thresholding methods. Comput Cardiol. 1999; 26:535-8. ; AlMahamdy and Riley, 2014 AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
http://dx.doi.org/10.1016/j.procs.2014....
; Awal et al., 2014 Awal MA, Mostafa SS, Ahmad M, Rashid MA. An adaptive level dependent wavelet thresholding for ECG denoising. Biocybern Biol Eng. 2014; 34(4):238-49. ; Bahoura and Ezzaidi, 2010 Bahoura M, Ezzaidi H. FPGA-implementation of wavelet-based denoising technique to remove power-line interference from ECG signal. Inf Technol Appl Biomed (ITAB). In: Proceedings of the 10th IEEE International Conference; 2010 Nov 3-5; Corfu, Greece. New Jersey: IEEE; 2010. p. 1-4. ; Chouakri et al., 2006 Chouakri SAS, Bereksi-Reguig AF, Fokapu O. ECG signal smoothing based on combining wavelet denoising levels. Asian J Inf Technol. 2006; 5(6):666-77. ; Garg et al., 2011 Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
; Germán-Salló, 2010 Germán-Salló Z. Nonlinear filtering in ECG signal denoising. Acta Univ Sapientiae Elec Mech Eng. 2010; 2:136-45. ; Karthikeyan et al., 2012 Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306. ; Li et al., 2009 Li S, Liu G, Lin Z. Comparisons of wavelet packet, lifting wavelet and stationary wavelet transform for de-noising ECG. In: Proceedings of the 2nd IEEE International Conference on Computer Science and Information Technology (ICCSIT 2009); 2009 Aug 8-11; Beijing, China; New Jersey: IEEE; 2009. p. 491-4. ; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
; Poornachandra and Kumaravel, 2008 Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
; Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
). Due to its better time-frequency resolution, it overcomes other classical methods, such as short time Fourier Transform, for instance ( Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
). One of the advantages when using wavelets is the computational efficiency of Mallat’s pyramidal algorithm ( Mallat, 1989 Mallat SG. A theory for multiresolution signal decomposition: the wavelet representation. IEEE Trans Pattern Anal Mach Intell. 1989; 11(7):674-93. http://dx.doi.org/10.1109/34.192463.
http://dx.doi.org/10.1109/34.192463 ...
). This algorithm is indeed a two-channel filter bank that splits the input signal in low and high frequencies by using quadrature mirror filters. The filters can be described through the wavelet ψ(t) and the scaling ϕ(t) basis functions ( Mallat, 1989 Mallat SG. A theory for multiresolution signal decomposition: the wavelet representation. IEEE Trans Pattern Anal Mach Intell. 1989; 11(7):674-93. http://dx.doi.org/10.1109/34.192463.
http://dx.doi.org/10.1109/34.192463 ...
):

ψj,n(t) = 2j/2ψ(t2jn2j) , (2)
ϕj,n(t)= 2j/2ϕ(t2jn2j) , (3)
h[n]=21/2ϕ(t/2),ϕ(tn) , (4)
g[n]=21/2ψ(t/2),ϕ(tn) , (5)

where g[n]=(1)1nh[1n] , j=1,2,,J and n integer. Such basis functions satisfy the conditions |ϕ(t)|2dt=1 ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ) and ψ(t)dt=0 ( Daubechies, 1992 Daubechies I. Ten lectures on wavelets. Philadelphia: SIAM; 1992. ). For a discrete analysis, wavelets are constructed by discretizing a “mother” function, and scaling it by 2j ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ), according to Equations (2) and (3) . In this way, a signal x(t) is decomposed as follows ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ):

d j [ n ] = x ( t ) , ψ j , n ( t ) (6)

and

aj[n]=x(t),ϕj,n(t) , (7)

where aj[n] and dj[n] are, respectively, the j -th approximation and detail coefficients at scale 2j , and j is the decomposition level. In a general way, given two sequences f1[n] and f2[n] , their inner product is defined as f1,f2=f1(t)f2¯(t)dt , where f2¯(t) is the complex conjugate of f2(t) . These coefficients can be computed in a fast way by a cascade algorithm, using discrete convolutions and subsamplings, by means of aj+1[p]=aj*h¯[2p] and dj+1[p]=aj[n]*g¯[2p] ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ), where x¯[n]=x[n], p is an integer and * is the convolution operation. This operation means coefficients filtering in a lower resolution, i.e., j . In this way, h¯[2p] filters the higher frequencies and g¯[2p] lets them pass. Therefore, an orthogonal representation for a signal x(t) is composed by the largest scale approximation coefficients plus detail coefficients at the scales j ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ), as follows:

x(t)=naJ[n]ϕJ,n(t)+jJndj[n]ψj,n(t) . (8)

Therefore, the wavelet decomposition output is a smooth signal representing the original one in a coarse way. In addition, the details are obtained when moving from a lower to a higher scale. Note that the smooth signal and details represent the similarity between the scaling and wavelet functions, according to Equations (7) and (6) , respectively. For an ECG signal, the approximation coefficients represent its smoothed version. On the other hand, detail coefficients capture abrupt changes, such as high-frequency noises. In order to reconstruct the signal x(t) the following equation is required ( Mallat, 2009 Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009. ): aj[p]=aj+1*h[p]+dj+1*g[p] , where y[n]=0 if n=2p+1 or y[n]=y[p] if n=2p

In the analysis step, the output wavelet filter bank frequency spectrum is divided into two octave bands. In each new decomposition level, the low-frequency spectrum is again divided into two new octave bands, at the ideal cut-off frequencies, and so on, resulting in a logarithmical set of bandwidth ( Germán-Salló, 2010 Germán-Salló Z. Nonlinear filtering in ECG signal denoising. Acta Univ Sapientiae Elec Mech Eng. 2010; 2:136-45. ). Therefore, if fs is the sampling frequency, the frequency contents for approximation and detail coefficients, in the j -th decomposition level, are in the interval [0, fs/2j+1] and [fs/2j+1, fs/2j] , respectively. In practice, the ideal cut-off frequencies are not realizable ( Peng et al., 2009 Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
http://dx.doi.org/10.1016/j.ymssp.2008....
). Therefore, the intervals are not exactly those mentioned before. Abrupt changes in the frequency intervals do not occur, but the filters frequency responses magnitude decreases gradually, tending to a constant value. Thus, leakage energy affects the frequency content for each DWT decomposition subband. Approximation coefficients are scattered higher than fs/2j+1 frequencies. On the other hand, detail coefficients are lower than fs/2j+1 frequencies and higher than fs/2j . Hence, there is a band overlap. Therefore, DWT frequency behavior impacts the decomposition level and the wavelet function choices. As an example, Figure 1 shows the frequency content for the quadrature mirror analysis filters given by Equations (2) and (3) for a 500 Hz sampling rate.

Figure 1
Frequency response for the quadrature mirror analysis filters in Equations (2) (black line) and (3) (red line) for Symlet 8-tap wavelet in three decomposition levels and 500 Hz sampling rate. (a) Level one. (b) Level two. (c) Level three.

Thresholding techniques

Classical methods for ECG denoising based on thresholding techniques present good performances ( AlMahamdy and Riley, 2014 AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
http://dx.doi.org/10.1016/j.procs.2014....
; Garg et al., 2011 Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
; Poornachandra and Kumaravel, 2008 Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
). Basically, in such methods, the goal is to estimate the signal x(t) from a contaminated signal s(t) = x(t) + r(t) , where r(t) is the additive noise ( Donoho and Johnstone, 1994 Donoho DL, Johnstone IM. Ideal spatial adaptation by wavelet shrinkage. Biometrika. 1994; 81(3):425-55. http://dx.doi.org/10.1093/biomet/81.3.425.
http://dx.doi.org/10.1093/biomet/81.3.4...
). For this purpose, the DWT is applied to s(t) and then dj[n]=s(t),ψj,n(t) in the wavelet domain is obtained. For hard shrinkage function, the absolute value of dj[n] is compared to a threshold λ . If |dj[n]| λ , the corresponding value associated to the index n is set to 0 . Otherwise, it is preserved.

In order to implement the wavelet shrinkage method, it has considered the Symlet 8 wavelet with three DWT decomposition levels and universal threshold (given by λ =2 log(N) , where N is the signal length), combined with hard thresholding function. It is noteworthy that this is the best configuration for this method ( Garg et al., 2011 Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
http://dx.doi.org/10.1109/IICPE.2011.57...
; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
). The choice of the wavelet function order and the number of decomposition levels are justified since they are the same used by the proposed method.

Notch filter

McManus et al. (1993) McManus CD, Neubert K, Cramer E. Characterization and Elimination of AC Noise in Electrocardiograms: A Comparison of Digital Filtering Methods. Comput Biomed Res. 1993; 26(1):48-67. http://dx.doi.org/10.1006/cbmr.1993.1003. PMid:8444027.
http://dx.doi.org/10.1006/cbmr.1993.100...
present four categories of digital filters for PLI removal: low-pass, notch-rejection, adaptive and global. For the implementation of the narrow-band-rejection filter (notch) it is considered a recursive filtering that includes a two-pole and two-zero filter. The filter output is given by y(k)=x(k)+a1x(k1)+a2x(k2)+a3y(k1)+a4y(k2) where x(k) is the input ECG signal and the coefficients are a1=2(1πBC/fs)cos(θ), a2=(1πBC/fs)2, a3=2(1πB/fs)cos(θ) and a4=(1πB/fs)2 , in which θ=2πfnom/fs , and B, C, fnom, fs are -3 dB response bandwidth (in Hz), center-frequency response, PLI frequency and sampling rate, respectively ( McManus et al., 1993 McManus CD, Neubert K, Cramer E. Characterization and Elimination of AC Noise in Electrocardiograms: A Comparison of Digital Filtering Methods. Comput Biomed Res. 1993; 26(1):48-67. http://dx.doi.org/10.1006/cbmr.1993.1003. PMid:8444027.
http://dx.doi.org/10.1006/cbmr.1993.100...
, apud Lynn, 1971 Lynn PA. Recursive digital filters for biological signals. Med Biol Eng. 1971; 9(1):37-43. http://dx.doi.org/10.1007/BF02474403. PMid:5580486.
http://dx.doi.org/10.1007/BF02474403 ...
).

In order to compare the proposed method with a classical approach, the recursive notch filter was selected. Lynn (1971) Lynn PA. Recursive digital filters for biological signals. Med Biol Eng. 1971; 9(1):37-43. http://dx.doi.org/10.1007/BF02474403. PMid:5580486.
http://dx.doi.org/10.1007/BF02474403 ...
apud McManus et al., (1993) McManus CD, Neubert K, Cramer E. Characterization and Elimination of AC Noise in Electrocardiograms: A Comparison of Digital Filtering Methods. Comput Biomed Res. 1993; 26(1):48-67. http://dx.doi.org/10.1006/cbmr.1993.1003. PMid:8444027.
http://dx.doi.org/10.1006/cbmr.1993.100...
, set B10 Hz, C0.01 and fs=1 kHz. Such parameters do not assure the best results for the signals tested in this work. Therefore, they were empirically set as B=5 Hz, C=0.005 and fs=500 Hz, assuring best results.

Considering that the objective of this work is to introduce a new method that overcomes the thresholding techniques, the results were compared to the ones obtained by a classical approach.

Evaluation metrics

In the literature, many objective measures are proposed to assess denoising techniques. One of them is the SNR, given by SNR=10log10(σx2/σy2) dB where σx2 and σy2 are, respectively, signal and noise variances. For a signal x(t) , the SNR improvement is defined by Awal et al. (2014) Awal MA, Mostafa SS, Ahmad M, Rashid MA. An adaptive level dependent wavelet thresholding for ECG denoising. Biocybern Biol Eng. 2014; 34(4):238-49. as SNRimp=10log10[|s(t)x(t)|2/|x^(t)x(t)|2] dB, where x^(t) , s(t) and x(t) represents the denoised signal, noisy signal and original ECG signal, respectively. Another measure, associated with mean square error is the root mean square error ( RMSE ) expressed by RMSE=1/N [x^(t)x(t)]2 , where N is the signal length. The relative error in the signal estimation can be written as ε=1/N |x^(t)x(t)| . A statistical measure that allows the linear association between the predicted signal and the original one is the correlation coefficient ( Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
): r=[x(t)μx][x^(t)μx^]/[x(t)μx]2[x^(t)μx^]2 , where μx and μx^ are the expected values for x(t) and x^(t) , respectively. This measure varies from 1 to 1 and the zero means no linear relationship.

The measures presented before are appropriate for synthetic ECG signals, but not for real signals, since, in that case, there is no prior access to noiseless ECG signals samples. In this way, two metrics, proposed by Li et al. (2014) Li Q, Rajagopalan C, Clifford GD. A machine learning approach to multi-level ECG signal quality classification. Comput Methods Programs Biomed. 2014; 117(3):435-47. http://dx.doi.org/10.1016/j.cmpb.2014.09.002. PMid:25306242.
http://dx.doi.org/10.1016/j.cmpb.2014.0...
, are used: the relative QRS complexes energy, given by eSQI=Eri/Ea , and the relative standard deviation, given by stdSQI=(σri)/2Mσai , where Ea is the energy of the whole signal, Eri is the energy in each QRS complex, M is the total number of QRS complexes, σri is the standard deviation of each QRS complex and σai is the standard deviation around the i -th QRS complex (from: R 0.2s to R+ 0.2s ; where R is the location of each R peak).

Statistical analysis

In order to evaluate whether the differences among the means in the experimental results are merely due to some random samples in the population, it is used the Kruskal-Wallis test. In this statistical test, ranks are used instead of the original observations. Firstly, all observations are ranked together and then the sum of the ranks is computed for each sample by means of the equation: H=12ni(ni+1)i=1NRi2ni3(ni+1) , where N is the number of samples, ni is the number of the observations and Ri is the sum of ranks in the i -th sample ( Kruskal and Wallis, 1952 Kruskal WH, Wallis WA. Use of ranks in one-criterion variance analysis. J Am Stat Assoc. 1952; 260(17):583-621. http://dx.doi.org/10.1080/01621459.1952.10483441.
http://dx.doi.org/10.1080/01621459.1952...
). H corresponds to some value in the χ2 distribution. Thus, the probability to get a highest or equal value than H is given by Pr[χN12H] , that is named p-value , with N1 degrees of freedom. When H is large and p-valueα , then the null hypothesis is rejected ( Kruskal and Wallis, 1952 Kruskal WH, Wallis WA. Use of ranks in one-criterion variance analysis. J Am Stat Assoc. 1952; 260(17):583-621. http://dx.doi.org/10.1080/01621459.1952.10483441.
http://dx.doi.org/10.1080/01621459.1952...
). In our case, the null hypothesis is that there is no significant difference among the tested methods. In this work α=0.05 .

Proposed method

According to equation (1) , the ECG signal represented by s(t) is added to another signal which represents the PLI. Thus, an estimated ECG signal x^(t) can be obtained as follow:

x^(t)=s(t)αcos(2πfrt+φ) . (9)

In the wavelet domain, the signal s(t) is represented by the approximation coefficients and PLI is contained in detail coefficients for a specific subband. Consequently, a simple approach to enhance the ECG signal in the wavelet domain is to reconstruct the signal x^(t) discarding detail coefficients. Based on the wavelet representation, Equation (8) , the signal x^(t) can be written as a product comprising scaling functions and low-pass filter outputs added to the sum of the product of wavelet functions and high-pass filter outputs. In the j -th level, low-frequency components are limited in the interval [0, fs /2j+1 ] while the high-frequency components are in interval [fs /2j+1, fs /2j] . Note that the precise frequency spectrum partition is impractical, since the filters frequency responses should be ideal ( Peng et al., 2009 Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
http://dx.doi.org/10.1016/j.ymssp.2008....
). It results in a band overlap with frequency content close to fs /2j+1 . In the overlap band interval the energy leakage occurs when the wavelet decomposition is used ( Peng et al., 2009 Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
http://dx.doi.org/10.1016/j.ymssp.2008....
), i.e., besides the expected frequency components, other frequencies also appear in the decomposed signal. Peng et al. (2009) Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
http://dx.doi.org/10.1016/j.ymssp.2008....
solved this problem by resampling the signal when the frequency content of interest is in the neighborhood of 22fs .

Choosing fs in a such a way that the approximation coefficients spectrum is limited in the range [0, 25.0] Hz and for detail coefficients spectrum is limited in the range [a, b] , where a 25 Hz and b 60 Hz, for some decomposition level j , and 22fs>25 Hz, it is possible to separate, in the wavelet domain, the PLI noise from the corrupted ECG signal. Obviously, the choice of j depends on the ECG signal sampling rate fs . According to Nyquist’s Theorem, fs should satisfy the following inequality: fs2fN , where fN is the maximum frequency of the signal. For instance, if fs=125 Hz then fN=62.5 Hz, and therefore if j=1 , approximation coefficients are in the range [0, 31.25] Hz, which comprises the frequency range content in an ECG signal. Detail coefficients are in the range [31.25, 62.5] Hz, containing PLI noise and other high-frequency noises.

As suggested by Peng et al. (2009) Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
http://dx.doi.org/10.1016/j.ymssp.2008....
, the upper limit for the approximation coefficients is higher than 25 Hz, which is the frequency of interest. Therefore, setting the sampling rate at 125 Hz is suitable. Obviously, for other sampling rates that are integer multiple of 125 Hz, the ECG signal and PLI noise can also be separated, but for higher decomposition levels. Nevertheless, for the experiments performed in this work the best results were obtained at a sampling frequency of 500 Hz. In Table 4 it is shown the frequency distribution in each decomposition level for such sampling rate. Columns two and three show the frequency range for ideal cut-off frequencies according to the range shown in the Figure 1 . Columns four and five show the real ranges, which are approximated values, obtained by analyzing the frequency response with the Symlet 8-tap. In the column six are presented the approximate band overlap ranges, considering the quadrature mirror analysis filters from Equations (4) and (5) .

Table 4
Frequency distribution for the wavelet decomposition of a signal sampled at 500 Hz considering a 3 dB cut-off frequency, according to Figure 1 .

In a noise-free ECG signal reconstruction, detail coefficients are not so important, since they do not have relevant information about the ECG signal waveform, as shown in Table 4 . Therefore, in order to obtain the signal x^(t) , given the sampling rate fs , a decomposition level LJ must be chosen, where J is the maximum level. Afterward, it is assumed dj[n]=0 , for all n and j = 1, 2, , L in Equation (8) . In this way, it follows from equation (8) that

x^(t)=naL[n]ϕL,n(t) . (10)

The proposed method can be summarized in four steps: 1) The ECG signal is sampled at fs = L 125 Hz, L is a positive integer; 2) W/2 null samples are added to the signal edges, where W=1000 is the window length; 3) signal is split using a W -length Hanning window with an overlap of 50%; 4) for each window, next steps are run; 4.1) DWT is applied to the splited signal, up to the level L ; 4.2) set dj[n]=0 for all n and j=1, 2, , L ; 4.3) The estimated ECG signal is reconstructed according to equation (10) .

The choice of the parameter L is the key to isolate the noise in a wavelet decomposition subband and that is why the sampling rate was defined as multiple of 125 Hz. If such sampling rate is not possible, there is no guarantee that the PLI will appear in the detail coefficients. In this way, resampling of the ECG signal should be considered. The null samples are added to avoid abrupt changes in the signal ends and W/2 samples are necessary due to the window overlap, which is an important procedure in signal processing. Generally, the window overlap vary from 50% to 75% ( Prabhu, 2014 Prabhu KMM. Window functions and their applications in signal processing. New York: CRC Press; 2014. ), but for the proposed method better results were achieved with 50% overlapping.

Results

The wavelet function choice

When applying the wavelet transform, besides the decomposition level choice discussed in the last section, it is also important to choose the wavelet function that best fits the signal. When ECG is the subject and threshold based methods are used, some researchers prefer Symlet wavelets because their scaling function resembles more its waveform ( Awal et al., 2014 Awal MA, Mostafa SS, Ahmad M, Rashid MA. An adaptive level dependent wavelet thresholding for ECG denoising. Biocybern Biol Eng. 2014; 34(4):238-49. ; Chouakri et al., 2006 Chouakri SAS, Bereksi-Reguig AF, Fokapu O. ECG signal smoothing based on combining wavelet denoising levels. Asian J Inf Technol. 2006; 5(6):666-77. ; Karthikeyan et al., 2012 Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306. ; Li et al., 2009 Li S, Liu G, Lin Z. Comparisons of wavelet packet, lifting wavelet and stationary wavelet transform for de-noising ECG. In: Proceedings of the 2nd IEEE International Conference on Computer Science and Information Technology (ICCSIT 2009); 2009 Aug 8-11; Beijing, China; New Jersey: IEEE; 2009. p. 491-4. ). Good results are also found using Daubechies wavelets ( Karthikeyan et al., 2012 Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306. ; Patil and Chavan, 2012 Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
http://dx.doi.org/10.1109/ICPRIME.2012....
; Üstündağ et al., 2012 Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
http://dx.doi.org/10.1007/s13721-012-00...
) and Coiflet ( Agante and Sa, 1999 Agante PM, Sa JPM. ECG noise filtering using wavelets with soft-thresholding methods. Comput Cardiol. 1999; 26:535-8. ; Karthikeyan et al., 2012 Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306. ). On the other hand, Poornachandra and Kumaravel (2008) Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
http://dx.doi.org/10.1016/j.dsp.2007.03...
compared some wavelet families with Mayer’s wavelet and conclude that the last one is better. Commonly, each method uses different thresholding techniques and they have influence on the wavelet function choice. However, there is some agreement about the use of Daubechies and Symlet wavelets. As the proposed method does not use thresholding techniques, the choice of wavelet function can be made by analyzing the results of experiments using the synthetic signals from Table 1 and Table 2 . Preliminary experiments with the proposed method showed that the relative error obtained when using Symlets is smaller than the one using Daubechies wavelet functions. Furthermore, it was noted that higher wavelet order leads to better result. However, runtime increases substantially.

For instance, the difference between the relative errors metrics, for the synthetic signal s1 , using sym10 (Symlet wavelet function of order 10) and sym20 is 5.14 × 104 and the absolute difference between runtimes is around 2.17 × 102 seconds. When s15 is considered, which is the one with worse SNR, runtime increases by about 103 seconds, from sym2 to sym8. From sym8 on, this metric increases by an amount of 101 seconds when the number of the wavelet vanishing moments increases. On the other hand, the relative error metric decreases in average 2 × 104 for each new Symlet wavelet function order, from sym4 on. Longer runtime is due to higher amount of the filters coefficients for wavelets of higher order.

Figure 2 shows the boxplots of the experiments for denoising synthetic signals with worse SNR: s6, s8, s11, s12, s13, s15, s16, s17 and s20 . Each box displays the relative error and runtime for different Symlet wavelet orders. The higher the order of the wavelet, the larger the filter length and number of the vanishing moments. This implies on softer functions and higher runtime. In general, the relative error remains almost constant from sym8 on, according to the results presented. Conversely, runtime is almost constant for wavelet functions with order lower than 8, increasing significantly after sym8. Therefore, considering balance between the relative error and the runtime, the authors believe that the best results could be achieved using sym8.

Figure 2
Boxplot for several Symlet wavelet orders: (a) relative error metrics boxplot, (b) runtime.

Experiments

In order to validate the proposed method, evaluation measures were computed for each synthetic signal. According to the analysis in last section, Symlet 8 was used for simulations. Superscripts PM, TT and NF refer to proposed method, thresholding technique and notch filter, respectively.

Analyzing column eight in Table 5 , it can be seen a significant SNR improvement ( SNRimp ) when compared to the values from row two in Table 1 and Table 2 . For s15 signal, with input SNR around 30 dB, the improvement was 50.9574 dB. s11 , with input SNR around 26 dB, the second worse, has an improvement of 45.8751 dB. In a general view, the SNRimp metric has an average close to 40.7806 dB. The correlation coefficient r (eleventh column) is close to one for all signals, meaning that the estimated signal waveform matches with the original ECG signal. Relative error ε and RMSE metrics indicate an accuracy of at least 102 for almost all experiments.

Table 5
Results for the proposed method (PM), thresholding technique (TT) and notch filter (NF) applied to synthetic ECG signal.

In Figure 3 , one can see the results in time and frequency domain, when applying the denoising methods for signal s15 .

Figure 3
Analysis display for signal s15 . (a) Synthetic ECG signal and (b) its spectrogram. (c) Noisy ECG signal and (d) its spectrogram. Denoised ECG signals obtained by (e) proposed method, (g) thresholding technique and (i) notch filter, and their spectrograms (f), (h) and (j), respectively. The red rectangle highlights a distortion inserted by the thresholding technique.

Table 6 shows results obtained by the proposed method, thresholding technique and notch filter for real signals. Comparing the original values in fourth column of Table 3 with the ones from the eighth to the tenth columns of Table 6 , it is noted that the frequencies over 25 Hz are less relevant for the denoised ECG signal than in the noisy ECG for all analyzed signals.

Table 6
Results for the proposed method, thresholding technique and notch filter applied to real ECG signals. PSD column is the sum of the power spectrum density for frequencies over 25 Hz.

Discussion

Comparing the proposed method results to the ones obtained with hard thresholding method, it can be seen that the proposed method was worse than hard thresholding only for s3 , s7 , s8 , s13 and s17 signals. In fact, the proposed method performed better in all requirements for all other signals. Average values reaffirm that the proposed method was better for the most of the evaluation measurements, except in terms of SNRimp . However, the differences on SNRimp values were not significant. For this metric, the proposed method was better than thresholding technique for 75% of the synthetic signals. In addition, it got higher results for signals s11 and s15 , the ones with the worst SNR values.

By means of the Kruskal-Wallis test, we conclude that there exists significantly statistical difference for at least two methods. In comparison to notch filter, the proposed method was better for all measures. However, the proposed method and the thresholding technique have a similar performance in statistical terms.

Note, from Figure 3 (f), that the energy of the QRS complexes remains practically unchanged when compared to the original. The original signal energy is close to 3.2015×103 , whereas the energy of the denoised ECG signals obtained by the proposed method, thresholding technique and notch filter, are close to 3.2188 × 103 , 3.3894 × 103 and 12.6194 × 103 , respectively. In addition, note from Figure 3 (g), that the thresholding technique inserted a distortion in the initial samples of the ECG signal whereas the proposed method did not change its waveform. Moreover, it can be seen from Figure 3 , (i) and (j), that notch filter achieved poor performance, since PLI was only attenuated. In the signal first samples it cleared the PLI with biggest gain. This outcome is common for notch filter ( Nauman et al., 2013 Nauman R, Maryam B, Muhammad S. An intelligent adaptive filter for fast tracking and elimination of power line interference from ECG signals. Proceedings of the 26th IEEE International Symposium on Computer-Based Med Sys (CBMS); 2013 Jun 20-22; Porto, Portugal. New Jersey: IEEE; 2013. p. 251-56. ). Due to the fact that some ECG analysis is performed by humans, a good visual quality is essential for an accurate diagnosis. Thus, one can conclude that the proposed method performed the best in terms of noise attenuation and distortion insertion.

For real ECG signals, from Table 6 , it is notorious that the proposed method reached better results for all signals, except for record 1086219 with respect to eSQI and record 1138505 for stdSQI . It is still noticeable that eSQI measure values are greater for signals obtained by the proposed method than the ones for the original signals. Since this measure is directly proportional to the QRS complex energy, these results mean that the high frequency noises (60 Hz or 50 Hz) have their gain attenuated. Furthermore, the proposed method performed the best for most of ECG signals, and it is important to highlight that its computational complexity is lower, since the additional thresholding technique complexity due to threshold computation is O(N) ( Lang et al., 1996 Lang M, Guo H, Odegard JE, Burrus CS, Wells RO. Noise reduction using an un decimated discrete wavelet transform. IEEE Signal Process Lett. 1996; 3(1):10-2. http://dx.doi.org/10.1109/97.475823.
http://dx.doi.org/10.1109/97.475823 ...
). In comparison to notch filter, the proposed method obtained better results for all signals. It is noteworthy that notch filter performed the worst for all quality assessment tests.

From Kruskal-Wallis test results, it is noted that only for the PSD measure there are significantly statistical difference among the three methods, considering a level of significance of α=5% . This is because the other measures not consider the PLI noise.

From Figure 4 , one can observe that thresholding technique and notch filter removed the PLI only for some segments in the observed ECG signal, leaving the others attenuated. Hereby, the result for record 1105115 obtained by the proposed method is much better ( Table 6 , fifth row) than the other methods, since the high frequency noise (180 Hz, see spectrogram in Figure 4 ) was not removed by the thresholding technique and notch filter. Note that 180 Hz is 2-nd harmonic frequency of 60 Hz.

Figure 4
(a) Raw ECG record 1105115 and (b) its spectrogram. Denoised ECG signal obtained by (c) proposed method, (e) thresholding technique and (g) notch filter and their spectrograms (d), (f) and (h), respectively.

Although the proposed method have obtained better results for the most of the analyzed signals, it is important to note that it depends on ECG signals sampling rate. So, one must be careful on the sampling rate and DWT decomposition level choices, since these parameters have great influence in the estimated signals quality, according to steps 1 and 4.2 of the proposed method. When, by technical reasons, the sampling rate cannot be changed, decomposition level must be chosen in such a way that a minimum amount of noise crosses into the signal subband.

Other limitation of the proposed method refers to the frequency content removed. In a scenario where frequencies over 34.60 Hz (see Table 4 ) are relevant, detail coefficients in the first level can be retained (frequencies in the range 111.40 to 250 Hz). Even so, PLI noise is removed. Though, in any case, the frequency content around 50/60 Hz is lost. In this way, the cardiac disorders that generate frequencies into the interval from 34.60 to 111.40 Hz are despised. It is essential to note that bandwidth mentioned in Table 4 can be distinct, provided that other cut-off frequency is considered. Therefore, the frequencies higher than 34.60 Hz are preserved in the reconstructed ECG signal.

In overview, in this paper it was proposed a new method for PLI noise removal based on the wavelet transform without the use of thresholding techniques. For such purpose, it was used a filter bank architecture implemented by the multiresolution analysis that allows splitting a signal in frequency subbands. By setting the sampling rate in 500 Hz, it is possible to separate PLI noise from ECG signal in distinct frequency subbands by using the wavelet representation. In order to choose this sampling rate, the energy leakage was considered, such that, for a DWT decomposition level, the frequency content of interest was close to half of the maximum signal frequency. Therefore, by zeroing detail coefficients, the ECG signal is reconstructed using only the approximation coefficients, obtaining a denoised ECG.

Energy conservation analysis for each cardiac cycle showed that the proposed method does not insert distortion in the estimated ECG signals. For real ECG signals, it was noted that the estimated QRS complexes waveforms are smooth and keep the expected morphology. On the other side, the thresholding technique added abrupt changes in some QRS complexes for records 1086219 and 2209843. Besides, other advantage of the proposed method is that there is no computational requirement for a threshold computation.

Although the proposed method depends on the sampling rate, it can be applied to other databases, with sampling rates different from a multiple of 125 Hz, since the signals resampling are considered. Finally, the proposed method can be applied for denoising other signals, with frequency content known in a specific range. In future works such applications will be considered.

  • How to cite this article: Oliveira BR, Duarte MAQ, Abreu CCE, Vieira Filho J. A wavelet-based method for power-line interference removal in ECG signals. Res Biomed Eng. 2018; 34(1):. DOI: 10.1590/2446-4740.01817.

References

  • Agante PM, Sa JPM. ECG noise filtering using wavelets with soft-thresholding methods. Comput Cardiol. 1999; 26:535-8.
  • Agrawal S, Gupta A. Fractal and EMD based removal of baseline wander and powerline interference from ECG signals. Comput Biol Med. 2013; 43(11):1889-99. http://dx.doi.org/10.1016/j.compbiomed.2013.07.030. PMid:24209934.
    » http://dx.doi.org/10.1016/j.compbiomed.2013.07.030
  • AlMahamdy M, Riley HB. Performance study of different denoising methods for ECG signals. Procedia Comput Sci. 2014; 37:325-32. http://dx.doi.org/10.1016/j.procs.2014.08.048.
    » http://dx.doi.org/10.1016/j.procs.2014.08.048
  • Awal MA, Mostafa SS, Ahmad M, Rashid MA. An adaptive level dependent wavelet thresholding for ECG denoising. Biocybern Biol Eng. 2014; 34(4):238-49.
  • Bahoura M, Ezzaidi H. FPGA-implementation of wavelet-based denoising technique to remove power-line interference from ECG signal. Inf Technol Appl Biomed (ITAB). In: Proceedings of the 10th IEEE International Conference; 2010 Nov 3-5; Corfu, Greece. New Jersey: IEEE; 2010. p. 1-4.
  • Bandarabadi AAJGM, Karami-Mollaei MR. ECG denoising using singular value decomposition. Aust J Basic Appl Sci. 2010; 4(7):2109-13.
  • Chouakri SAS, Bereksi-Reguig AF, Fokapu O. ECG signal smoothing based on combining wavelet denoising levels. Asian J Inf Technol. 2006; 5(6):666-77.
  • Costa MH, Tavares MC. Removing harmonic power line interference from biopotential signals in low cost acquisition systems. Comput Biol Med. 2009; 39(6):519-26. http://dx.doi.org/10.1016/j.compbiomed.2009.03.004. PMid:19376509.
    » http://dx.doi.org/10.1016/j.compbiomed.2009.03.004
  • Das M, Ari S. Analysis of ECG signal denoising method based on s-transform. IRBM. 2013; 34(6):362-70. http://dx.doi.org/10.1016/j.irbm.2013.07.012.
    » http://dx.doi.org/10.1016/j.irbm.2013.07.012
  • Daubechies I. Ten lectures on wavelets. Philadelphia: SIAM; 1992.
  • Donoho DL, Johnstone IM. Ideal spatial adaptation by wavelet shrinkage. Biometrika. 1994; 81(3):425-55. http://dx.doi.org/10.1093/biomet/81.3.425.
    » http://dx.doi.org/10.1093/biomet/81.3.425
  • Garg G, Gupta S, Singh V, Gupta JRP, Mittal AP. Identification of optimal wavelet-based algorithm for removal of power line interferences in ECG signals. In: Proceedings of the India International Conference on Power Electronics; 2011 Jan 28-30; New Delhi, India. New Jersey: IEEE; 2011. p. 1-5. http://dx.doi.org/10.1109/IICPE.2011.5728090.
    » http://dx.doi.org/10.1109/IICPE.2011.5728090
  • Germán-Salló Z. Nonlinear filtering in ECG signal denoising. Acta Univ Sapientiae Elec Mech Eng. 2010; 2:136-45.
  • Goldberger AL, Amaral LAN, Glass L, Hausdorff JM, Ivanov PC, Mark RG, Mietus JE, Moody GB, Peng CK, Stanley HE. PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation. 2000; 101(23):E215-20. https://doi.org/10.1161/01.CIR.101.23.e215. PMid:10851218.
    » https://doi.org/10.1161/01.CIR.101.23.e215
  • Huhta JC, Webster JG. 60-Hz Interference in Electrocardiogram. IEEE Trans Biomed Eng. 1973; 20(2):91-101. http://dx.doi.org/10.1109/TBME.1973.324169. PMid:4688314.
    » http://dx.doi.org/10.1109/TBME.1973.324169
  • Karthikeyan P, Murugappan M, Yaacob S. ECG signal denoising using wavelet thresholding techniques in human stress assessment. Int J Elec Eng Inf. 2012; 4(2):306.
  • Köhler BU, Hennig C, Orglmeister R. The principles of software QRS detection. IEEE Eng Med Biol Mag. 2002; 21(1):42-57. http://dx.doi.org/10.1109/51.993193. PMid:11935987.
    » http://dx.doi.org/10.1109/51.993193
  • Kruskal WH, Wallis WA. Use of ranks in one-criterion variance analysis. J Am Stat Assoc. 1952; 260(17):583-621. http://dx.doi.org/10.1080/01621459.1952.10483441.
    » http://dx.doi.org/10.1080/01621459.1952.10483441
  • Lang M, Guo H, Odegard JE, Burrus CS, Wells RO. Noise reduction using an un decimated discrete wavelet transform. IEEE Signal Process Lett. 1996; 3(1):10-2. http://dx.doi.org/10.1109/97.475823.
    » http://dx.doi.org/10.1109/97.475823
  • Łęski JM, Henzel N. ECG baseline wander and powerline interference reduction using nonlinear filter bank. Signal Process. 2005; 85(4):781-93. http://dx.doi.org/10.1016/j.sigpro.2004.12.001.
    » http://dx.doi.org/10.1016/j.sigpro.2004.12.001
  • Li Q, Rajagopalan C, Clifford GD. A machine learning approach to multi-level ECG signal quality classification. Comput Methods Programs Biomed. 2014; 117(3):435-47. http://dx.doi.org/10.1016/j.cmpb.2014.09.002. PMid:25306242.
    » http://dx.doi.org/10.1016/j.cmpb.2014.09.002
  • Li S, Liu G, Lin Z. Comparisons of wavelet packet, lifting wavelet and stationary wavelet transform for de-noising ECG. In: Proceedings of the 2nd IEEE International Conference on Computer Science and Information Technology (ICCSIT 2009); 2009 Aug 8-11; Beijing, China; New Jersey: IEEE; 2009. p. 491-4.
  • Lynn PA. Recursive digital filters for biological signals. Med Biol Eng. 1971; 9(1):37-43. http://dx.doi.org/10.1007/BF02474403. PMid:5580486.
    » http://dx.doi.org/10.1007/BF02474403
  • Mallat SG. A theory for multiresolution signal decomposition: the wavelet representation. IEEE Trans Pattern Anal Mach Intell. 1989; 11(7):674-93. http://dx.doi.org/10.1109/34.192463.
    » http://dx.doi.org/10.1109/34.192463
  • Mallat SG. A wavelet tour of signal processing: the sparse way. 3rd ed. Burlington: Elsevier; 2009.
  • Mateo J, Sanchez C, Tortes A, Cervigon R, Rieta JJ. Neural network based canceller for powerline interference in ECG signals. In: Proceedings of the 35th Annual Computers in Cardiology Conference (CinC); 2008 Sep 14-17; Bolongna, Italy. New Jersey: IEEE; 2008. p. 1073-76. http://dx.doi.org/10.1109/CIC.2008.4749231.
    » http://dx.doi.org/10.1109/CIC.2008.4749231
  • McManus CD, Neubert K, Cramer E. Characterization and Elimination of AC Noise in Electrocardiograms: A Comparison of Digital Filtering Methods. Comput Biomed Res. 1993; 26(1):48-67. http://dx.doi.org/10.1006/cbmr.1993.1003. PMid:8444027.
    » http://dx.doi.org/10.1006/cbmr.1993.1003
  • McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Trans Biomed Eng. 2003; 50(3):289-94. http://dx.doi.org/10.1109/TBME.2003.808805. PMid:12669985.
    » http://dx.doi.org/10.1109/TBME.2003.808805
  • Nauman R, Maryam B, Muhammad S. An intelligent adaptive filter for fast tracking and elimination of power line interference from ECG signals. Proceedings of the 26th IEEE International Symposium on Computer-Based Med Sys (CBMS); 2013 Jun 20-22; Porto, Portugal. New Jersey: IEEE; 2013. p. 251-56.
  • Oliveira BR, Duarte MAQ, Vieira Filho J. Detecção de complexos QRS do ECG pela decomposição em valores singulares em multirresolução. In: Anais da IX ENAMA; 2015 Nov 4-6; Cascavel, Brasil; São Carlos: SBMAC; 2015. p. 143-4.
  • Pan J, Tompkins WJ. A real-time QRS detection algorithm. IEEE Trans Biomed Eng. 1985; 32(3):230-6. http://dx.doi.org/10.1109/TBME.1985.325532. PMid:3997178.
    » http://dx.doi.org/10.1109/TBME.1985.325532
  • Patil PB, Chavan MS. A wavelet based method for denoising of biomedical signal. In: Proceedings of International Conference on Pattern Recognition, Informatics and Medical Engineering (PRIME); 2012 Mar 21-23; Salem, Tamilnadu, India. New Jersey: IEEE; 2012. p. 278-83. http://dx.doi.org/10.1109/ICPRIME.2012.6208358.
    » http://dx.doi.org/10.1109/ICPRIME.2012.6208358
  • Peng Z, Jackson M, Rongong J, Chu F, Parkin R. On the energy leakage of discrete wavelet transform. Mech Syst Signal Process. 2009; 23(2):330-43. http://dx.doi.org/10.1016/j.ymssp.2008.05.014.
    » http://dx.doi.org/10.1016/j.ymssp.2008.05.014
  • Phegade M, Mukherji P. ICA based ECG signal denoising. In: Proceedings of the International Conference on Advances in Computing, Communications and Informatics (ICACCI); 2013 Aug 22-25; Mysore, India; New Jersey: IEEE; 2013. p. 1675-80.
  • Poornachandra S, Kumaravel N. A novel method for the elimination of power line frequency in ECG signal using hyper shrinkage function. Digit Signal Process. 2008; 18(2):116-26. http://dx.doi.org/10.1016/j.dsp.2007.03.011.
    » http://dx.doi.org/10.1016/j.dsp.2007.03.011
  • Prabhu KMM. Window functions and their applications in signal processing. New York: CRC Press; 2014.
  • Rahman MZU, Shaik RA, Reddy DVRK. Baseline wander and power line interference elimination from cardiac signals using error nonlinearity LMS algorithm. In: International Conference on Systems in Medicine and Biology (ICSMB). 2010 Dec 16-18; Kharagpur, India; New Jersey: IEEE; 2010. p. 217-20.
  • Shirbani F, Setarehdan SK. ECG power line interference removal using combination of FFT and adaptive non-linear noise estimator. In: Proceedings of the 21st Iranian Conference on Electrical Engineering (ICEE). 2013 May 14-16. Mashhad, Iran. New Jersey: IEEE; 2013. p. 1-5. http://dx.doi.org/10.1109/IranianCEE.2013.6599622.
    » http://dx.doi.org/10.1109/IranianCEE.2013.6599622
  • Üstündağ M, Gökbulut M, Sengür A, Ata F. Denoising of weak ECG signals by using wavelet analysis and fuzzy thresholding. Netw Model Anal Health Inform Bioinform. 2012; 1(4):135-40. http://dx.doi.org/10.1007/s13721-012-0015-5.
    » http://dx.doi.org/10.1007/s13721-012-0015-5
  • Zidelmal Z, Amirou A, Adnane M, Belouchrani A. QRS detection based on wavelet coefficients. Comput Methods Programs Biomed. 2012; 107(3):490-6. http://dx.doi.org/10.1016/j.cmpb.2011.12.004. PMid:22296976.