Validation study of an automated wrist monitor, omron model HEM-608, compared with the standard methods for blood pressure measurement

OBJECTIVE - The aim of our study was to assess the profile of a wrist monitor, the Omron Model HEM-608, compared with the indirect method for blood pressure measurement. METHODS - Our study population consisted of 100 subjects, 29 being normotensive and 71 being hypertensive. Participants had their blood pressure checked 8 times with alternate techniques, 4 by the indirect method and 4 with the Omron wrist monitor. The validation criteria used to test this device were based on the internationally recognized protocols. RESULTS - Our data showed that the Omron HEM-608 reached a classification B for systolic and A for diastolic blood pressure, according to the one protocol. The mean differences between blood pressure values obtained with each of the methods were -2.3 +7.9mmHg for systolic and 0.97+5.5mmHg for diastolic blood pressure. Therefore, we considered this type of device approved according to the criteria selected. CONCLUSION - Our study leads us to conclude that this wrist monitor is not only easy to use, but also produces results very similar to those obtained by the standard indirect method.

automated blood pressure device; wrist monitor; hypertension and normotension