Evaluation and comparison between methods of measurement of systolic blood pressure in healthy anesthetized cats

For validation purposes, two different methods of measurement of systolic blood pressure (SBP), noninvasive or indirect (oscillometric and vascular doppler), direct or invasive (arterial catheterization - the gold standard) were evaluated and compared in 12 healthy anesthetized cats, varying between seven months to two years old. In each animal, we proceeded to the simultaneous noninvasive and invasive measurement of pressure. The oscillometric method was compared by means of seven consecutive measurements with an interval of 30 seconds between them, with the invasive method, and soon after, the doppler method was compared to the invasive method in the same way. It was concluded that the oscillometric method is statistically equal to the invasive method, whereas doppler differs from both. However, due to the strong positive correlation between doppler and invasive methods, it was possible to create a correction factor (linear regression) to determine the value of invasive pressure from the value obtained by the doppler method. To obtain more reliable figures, it is recommended to use this correction factor when the doppler method is used to measure SBP in anesthetized cats.

feline; doppler; blood pressure