Digital morphometric analysis of the aortic wall in pigs following implantation of dacron-covered stents versus non-covered stents

Análise morfométrica digital na parede aórtica de suínos após implante de stents revestidos com dacron versus stents não revestidos

Abstracts

PURPOSE: To evaluate, by digital morphometry, the intimal thickening after the placement of two different kinds of stents: polyester covered stainless steel stents (Dacron) and non-covered stents implanted in young pigs' infrarenal aortas. METHODS: The experiment was carried out on two separate groups of pigs. Eight polyester-covered stainless steel stents (Dacron) and eight stainless steel stents (30-mm long, 8-mm diameter) were deployed through extraperitoneal approach in the normal infrarenal aorta of 16 normolipemic pigs. To allow the passage of the delivery system, a small arteriotomy was performed (phase I). After four weeks, the aorta with the stent was removed en bloc. (phase II). The values of the hematimetric and lipid analysis did not show any changes that could influence the study. Tissue samples of the fixing sites (proximal and distal) of the stents were taken. Microscope slices were prepared and submitted to Verhoeff's hematoxilin and eosin techniques and sent to morphometric analysis. RESULTS: The intima immediately proximal to the device was thicker in the group of covered stents with marginal significance (p=0.054). The distal intimal layer (p=0.185), proximal medial layer of the proximal portion (p=0.141) and distal portion (p=0.375) did not present statistical difference between the two groups. CONCLUSIONS: Patency was 100% in both groups of these normolipemic pigs. After four weeks the intimal layer immediately proximal to the covered stents was ticker when compared to uncovered stents, but this had a borderline significance. The intimal layer distal to covered stents and the media proximal or distal to the devices had similar morphometric features when covered and uncovered stents where compared.

Intimal thickness; Stainless steel stents; Dacron; Morphometric analysis


OBJETIVO: Avaliar através de morfometria digital o espessamento intimal após o implante de dois tipos de stents de aço inoxidável, um recoberto com poliéster (dacron) e outro não recoberto, implantados na aorta infra-renal de suínos jovens. MÉTODOS: O experimento foi realizado em dois grupos (stents não revestidos e revestidos com dacron) e duas fases (I e II). Oito stents recobertos com dacron e oito stents de aço inoxidável (30mm de extensão e 8 mm de diâmetro), não revestidos, foram implantados através de abordagem retroperitoneal na aorta infrarenal normal de 16 suínos normolipêmicos. Para a passagem do sistema de implante, foi necessário uma pequena arteriotomia na aorta distal (fase I). Após quatro semanas, a aorta com os stents foram removidas em monoblocos (fase II). Os valores de hematimetria e do lipidograma foram coletados nas duas fases e não apresentaram alterações que pudessem influenciar o estudo. Amostras de tecido dos sítios de fixação (proximal e distal) dos stents foram retiradas, confeccionadas lâminas, que foram coradas pelas técnicas de hematoxilina e eosina e de Verhoeff e enviadas para análise morfométrica digital. RESULTADOS: A camada íntima proximal ao stent apresentou maior espessamento intimal no grupo de stents recobertos, mas com significância limítrofe(p=0.054). A camada íntima distal(p=0.185), a camada média das regiões proximal(p=0.141) e da porção distal (p=0.375) ao stent não apresentaram diferença entre os grupos. CONCLUSÕES: Os dois tipos de stents apresentaram 100% de perviedade, boa biocompatibilidade e boa incorporação a parede aórtica de suínos normolipêmicos. A camada íntima proximal do grupo de stents revestidos com dacron apresentou maior espessamentodo que os stents não revestidos, mas com significância estatísitica limítrofe.

Espessamento intimal; Stents; Dacron; Análise morfométrica digital


ORIGINAL ARTICLE

Digital morphometric analysis of the aortic wall in pigs following implantation of dacron-covered stents versus non-covered stents1 1 Master dissertation Post-Graduate Program of Federal University of Rio Grande do Sul (UFRGS), Brazil. How to cite this article: Dutra CF, Pereira AH. Digital morphometric analysis of the aortic wall in pigs following implantation of dacron-covered stents versus non-covered stents. Acta Cir Bras [serial online] 2004 May-Jun;19(3). Available from URL: http://www.scielo.br/acb. [also in CD-ROM].

Análise morfométrica digital na parede aórtica de suínos após implante de stents revestidos com dacron versus stents não revestidos

Clandio de Freitas DutraI; Adamastor Humberto PereiraII

IMaster in Surgery Post-Graduate Program of UFRGS, Brazil

IIPhD in Surgery, University of São Paulo (USP). Associate Professor Faculty of Medicine UFRGS, Brazil

Correspondence Correspondence to Clandio de Freitas Dutra Rua Garibaldi, 476/Sala 701 95084-060 Caxias do Sul - RS Tel/Fax: (54)214-0079 scdutra@terra.com.br

ABSTRACT

PURPOSE: To evaluate, by digital morphometry, the intimal thickening after the placement of two different kinds of stents: polyester covered stainless steel stents (Dacron) and non-covered stents implanted in young pigs' infrarenal aortas.

METHODS: The experiment was carried out on two separate groups of pigs. Eight polyester-covered stainless steel stents (Dacron) and eight stainless steel stents (30-mm long, 8-mm diameter) were deployed through extraperitoneal approach in the normal infrarenal aorta of 16 normolipemic pigs. To allow the passage of the delivery system, a small arteriotomy was performed (phase I). After four weeks, the aorta with the stent was removed en bloc. (phase II). The values of the hematimetric and lipid analysis did not show any changes that could influence the study. Tissue samples of the fixing sites (proximal and distal) of the stents were taken. Microscope slices were prepared and submitted to Verhoeff's hematoxilin and eosin techniques and sent to morphometric analysis.

RESULTS: The intima immediately proximal to the device was thicker in the group of covered stents with marginal significance (p=0.054). The distal intimal layer (p=0.185), proximal medial layer of the proximal portion (p=0.141) and distal portion (p=0.375) did not present statistical difference between the two groups.

CONCLUSIONS: Patency was 100% in both groups of these normolipemic pigs. After four weeks the intimal layer immediately proximal to the covered stents was ticker when compared to uncovered stents, but this had a borderline significance. The intimal layer distal to covered stents and the media proximal or distal to the devices had similar morphometric features when covered and uncovered stents where compared.

Key words: Intimal thickness. Stainless steel stents. Dacron. Morphometric analysis.

RESUMO

OBJETIVO: Avaliar através de morfometria digital o espessamento intimal após o implante de dois tipos de stents de aço inoxidável, um recoberto com poliéster (dacron) e outro não recoberto, implantados na aorta infra-renal de suínos jovens.

MÉTODOS: O experimento foi realizado em dois grupos (stents não revestidos e revestidos com dacron) e duas fases (I e II). Oito stents recobertos com dacron e oito stents de aço inoxidável (30mm de extensão e 8 mm de diâmetro), não revestidos, foram implantados através de abordagem retroperitoneal na aorta infrarenal normal de 16 suínos normolipêmicos. Para a passagem do sistema de implante, foi necessário uma pequena arteriotomia na aorta distal (fase I). Após quatro semanas, a aorta com os stents foram removidas em monoblocos (fase II). Os valores de hematimetria e do lipidograma foram coletados nas duas fases e não apresentaram alterações que pudessem influenciar o estudo. Amostras de tecido dos sítios de fixação (proximal e distal) dos stents foram retiradas, confeccionadas lâminas, que foram coradas pelas técnicas de hematoxilina e eosina e de Verhoeff e enviadas para análise morfométrica digital.

RESULTADOS: A camada íntima proximal ao stent apresentou maior espessamento intimal no grupo de stents recobertos, mas com significância limítrofe(p=0.054). A camada íntima distal(p=0.185), a camada média das regiões proximal(p=0.141) e da porção distal (p=0.375) ao stent não apresentaram diferença entre os grupos.

CONCLUSÕES: Os dois tipos de stents apresentaram 100% de perviedade, boa biocompatibilidade e boa incorporação a parede aórtica de suínos normolipêmicos. A camada íntima proximal do grupo de stents revestidos com dacron apresentou maior espessamentodo que os stents não revestidos, mas com significância estatísitica limítrofe.

Descritores: Espessamento intimal. Stents. Dacron. Análise morfométrica digital.

Introduction

Regular stents can be constructed with an uncovered metal mesh, which can be steel, nitinol, tântalo or algiloy. They have been used to dilate arterial stenoses, or for fixation of atheroma plaques following angioplasty1. Stents can be covered with autogenous tissue, such as the saphenous vein, or synthetic material like expanded polytetrafluoroethylene (PTFEe) and polyethylene tereftelate (Dacron)2,3,4. Covered stents can be used in the treatment of traumatic vascular injuries, pseudoaneurysms, arteriovenous fistulas, aortic dissection or degenerative aneurysms 5,6,7.

An advantage of using a covered stent in occlusive arterial diseases is that the synthetic tissue can form a barrier excluding the ruptured wall from the vascular lumen. Another feature is that this barrier could prevent miointimal cell proliferation within the metallic mesh. In the other hand , some disadvantages can be named as the foreign body effect in the vascular lumen, higher cost, and the need of a delivery system of higher profile. 3,8,9.

Although stents seems to prevent the elastic retraction by means of positive geometric remodeling, they are not innocuous to the arterial system. Stents can be related to thrombi formation, higher inflammatory reaction and neointima tickening due to the fact that the endothelial lesion exposes the subintimal elements, causing a cicatricial answer to injury 10.

The most important difference between stents covered with polymeric materials and non-covered stents concerns their likehood of being incorporated by the vascular wall.

Naked stents are more likely to endotelize than when they are covered with polymers. These properties probably determine a stable binding to proteins when exposed to blood flow, followed by cells fixation onto the metallic mesh of the stents. The high energy found in the metal surface does not explain such interactions. The electrical load on the surface, the chemical features, and the texture are most likely involved11.

When compared to non-covered stents, covered stents are less biocompatible. There is some concordance between the behavior of Dacron and PTFE prostheses since they allow endothelization restricted to the perianastomotic region. In contrast, uncovered stents permit full endothelization, a feature that have been observed in several experimental studies carried out with baboons, dogs, pigs, sheep and veal12.

The hypothesis in our study was that stents covered with extra-thin Dacron induces a reative intimal thickening greater than non-covered stents.

Methods

This study was approved by the Ethical Committee for the Post-Graduation Groups at the Clinical Hospital in Porto Alegre. The experiment was carried out at the facilities of the Animal Room in the University of Caxias do Sul and the digital morphometric analysis was performed at the Digital Medical Laboratory in Porto Alegre.

Stents were manufactured in 316L-series steel by the "Laboratório de Transformação Mecânica da Escola de Engenharia da Universidade Federal do Rio Grande do Sul, based in ELLA-CS® patterns. These stents were Z-shaped, self expandable, 3cm long and 8mm wide. Covered stents were manufactured by sewing an extra-fine Dacron covering with polypropylene thread to the internal surface of the metalic mesh (Figure 1).


The animals used in this study were male specimens of a double purpose breed from the Large White X Landrace. They were over eight weeks old and weighed between 16 and 25Kg 13,14. The anesthetic technique employed was inhalatory general anesthesia, with a semi-closed reinhalation system. Pigs were sedated with tiletamine plus zolazepam (ZOLETIL®), at doses of 0.2ml/kg injected intramuscularly five minutes before the procedures. Anesthetic induction was carried out with intravenous thiopental sodium at 2.5%(10-12mg/kg) followed by tracheal intubation with a long laryngoscope 2,13,15. To maintain the anesthesia, 0.5% to 1% halotane was used together with liquid replacement with physiological solution at 0.9%, 20ml/kg/hour.

An extraperitoneal approach through a left pararetal incision was used to expose the aorta. Following systemic heparinization (100UI/kg), a transversal arteriotomy was performed half-centimeter above the aortic trifurcation and the delivery system was introduced. The stent were then deployed. Arteriotomy synthesis was performed with a 5-0 polypropylene, and after the revision of the haemostasis, the abdominal wall was closed.

After 4 weeks, animals were subjected to the same anesthesia used for implantation. A supra and infra-umbilical laparotomy was performed to remove specimens (aorta including the stents) in one piece. After tissue collection, animals were submitted to euthanasia.

Preparing the tissue for histological analysis

Tissues removed were irrigated with sodium chloride solution at 0.9%, and vessel patency was macroscopically checked. Specimens were cut longitudinally and fixed with a solution of formaldehyde at 10% 16,17,18.

Tissue portions with 0.3-cm-long sections from the extension of the aorta artery from both ends (proximal and distal) were removed for histological analysis. Segments were processed and included in paraffin blocks and later submitted to histological cuts 4µm thick for preparation on the microscope slides. Slides were then prepared and blushed according to H-E(Hematoxilin-Eosin) and Verhoeff techniques.

Digital morphometry

Quantification methodology for intimal thickening of stents was performed with digital morphometric analysis according to an integrated morphometric program and image analysis - Media Cybernetics: Image Pro Plus. Images of histological sections were digitalized for morphometric analysis by conventional optical microscopy (Zeiss Microscope, Axiastar Model), by achromatic optical plane and photo-storage tube ( Sony DXC 151 Video camera), generating image files in a PC. Images were digitalized with a 100-fold microscopic increase, and intimal and muscular areas were delimited, respectively, according to internal elastic membrane, and the endothelium. Dimensions of the areas were described in mm², using the average of eight microscopic fields.19

External and internal elastic membranes defined the area for the medial layers and the lines limited by the internal elastic membrane and the endothelium was considered as the intimal area. The ratio intimal area/medial layer was used as intimal index.

Statistical analysis

The data were analyzed according to SPSS v.6.0 Statistical Package for Social Sciences for Windows (Microsoftâ - USA), by descriptive statistics. Student's t-test was used for matched samples between the groups and Student's t-test was used for independent samples in-between phases in order to analyze thickness of layers measured by digital morphometry.

Results

Morphometric analysis of the pig aorta

Morphometric analysis (proximal and distal to the stent) comparing the SNR and SRD groups were carried out with digital techniques including the intimal and medial layers. The intimal rate was established (ratio between the intimal layer and the medial layer) (Figure 2).


Non-covered stent group (SNR) vs. dacron-covered stent group (SRD) - proximal region.

Results obtained for proximal intimal area in the SNR group were 0.139 ± 0.069 mm², and 0.222 ± 0.089 mm² in the SRD group; the difference of 0.083mm² in the averages represented a borderline statistical difference (p=0.054). In the SNR group, proximal average was 5.158 ± 0.782mm², and in the SRD group it was 5.750 ± 0.733mm². The difference of 0.591mm² in the averages does not represent a significant statistical difference (p=0.141). The value of the proximal intimal index in the SNR group was 0,028 +- 0,017. Proximal intimal rate in the SNR group was 0.038 ± 0.013, whereas the difference of 0.009 in the averages does not represent a significant statistical difference (p=0.226) (Table 1).

Non-covered stent group (SNR) vs. dacron-covered stent group (SRD) - distal region

The average results found for the distal intimal area in the SNR group were 0.252 ± 0.074mm², and, in the SRD group were 0.334 ± 0.149mm² (p=0.185). Results for distal average in the SNR group were 5.429 ± 0.715mm², and 5,089 ± 0.769mm² in the SNR group; difference of the averages was 0.340mm² (p=0.375). Results for distal intimal rate in the SNR group were 0.047 ± 0.017, and in the SRD group 0.068 ± 0.036, showing a difference of 0.020 in the averages (p=0.160).

When proximal and distal intimal layers of the same group of stents were compared, there were no statistical differences neither in the SNR group (p=0.608) nor in the SRD group (p=0.610) (Table 2).

Macroscopic and intraoperative analysis

During the surgical procedures (phases I and II) no vasospasms were observed in the aorta. We also carried out a macroscopic analysis of the specimens on phase II.

Discussion

Other authors 13, 20, 21 have also implanted stents in normal arteries and analyzed them through digital morphometry. Of course this king of study would be better performed in arteries with advanced atherosclerosis. However, animals models are difficult to obtain because they require long maintenance periods, hiperlipemic diets for long periods of time and specific genetic breeds 22.

The surgical or endovascular approach to the vessel being studied varies according to the personal preference of each author. Arteries that have a superficial anatomical site, such as carotids and femoral arteries, are easily exposed to surgical procedures, while other deeper arteries like coronary arteries and the aorta are better approached by percutaneous catheters guided by fluoroscopy 22.

We approached the aorta through an extraperitoneal access as described by Weatherford et al 16, who carried out a similar research on dogs. This approach was chosen by us in this pig model because it is associated to lower rates of morbidity and mortality of the animals. It can be done expeditiously, the exposure of vessels is appropriate and blood loss is minimal.

In our study, 0.3cm-long sections were made in the aorta immediately proximal and distal to the stent edges. Based on digital morphometry techniques, areas from intimal and medial layers were measured in each artery and the average of the 8 microscopic fields with a 100-fold amplification was obtained. Schürmann et al23 calculated also the average of the dimensions measured in eight fields with similar results.

All stents implanted in our study, whether they were covered or not, were patent after 4 weeks. The aorta of pigs used in the experiment had an average of 8-mm in diameter. There are some advantages in handling a vessel with such diameter when implanting tissue-covered stents, since there is a smaller probability for thrombosis in arteries with high flow. According to Cucina et al, 24 and Bassiouny et al, 25 grafts implanted in low flow arteries promote a more intense intimal hyperplasia.

Digital morphometric analyses showed that the intimal layer adjacent to the proximal margin of the covered stents (SRD) had greater intimal thickening than the intimal layer of the group with non-covered stents (SNR), though statistically this was of borderline significance (p=0,054). No differences were found in the groups when comparing areas of proximal media layer (p=0,141), distal intimal layer (p=0,185), distal media layer (p=0,375), and also when comparing the proximal intimal index (p=0,226) and distal index (0,160).

Hehrlein et al 26 found a larger intimal formation on the distal border of implanted stents than on the center portion. In our study, the proximal and the distal intimal layers were compared, in-groups, to check if there was larger intimal thickening in any of the analyzed implant sites. According to our analysis, there was no difference in intimal thickening between the proximal and the distal sites to the stents regardless they were covered or not.

Chalmers et al 27 evaluated, through morphometric analysis, intimal tickening in the anastomotic sites of uncovered stents implanted in dogs. A significant larger area of intimal hyperplasia was found adjacent to the borders of the stents. The intermediate region of the stent did not present any differences when compared to controls.

Karas et al 13 compared the proliferative response in pigs coronary arteries after implanting tantalum stents and after angioplasty with balloons in normolipemic pigs. After four weeks, arteries were processed for histopathological analysis. Data obtained indicated that the degree of intimal proliferation seems to be larger after stent implants than after balloon-injury. Intracoronary stents in pigs were associated to a strong inflammatory reaction around the stent mesh.

The clinical series with human beings that underwent angioplasty have shown conflicting results. Stenting is associated to lower patency rates than open surgery 28, 29, 30, but there has been considerable improvement in recent clinical series. The goal is reduction of intimal hyperplasia and restenosis rate. Some reports have already shown the results with covered stents impregnated with medicinal drugs and stents with radioactive substances that are capable of inhibiting intimal hyperplasia.

In our study, we could find an excellent patency, and a good incorporation for both kinds of stents (covered and non-covered with Dacron). Intimal thickening had similar characteristics in both groups of stents when comparing proximal and distal implant sites between and within the groups.

Even though the response to injury and biology of the vascular response after the implantation of these devices may be complex, we believe that clinical and experimental researches shall supplant the initial difficulties and will achieve the major goal, which is offering patients less invasive procedures with excellent results.

Conclusions

Neither of the two groups of stents (Dacron-covered and non-covered), when analyzed by digital morphometry for intimal thickening, presented any significant statistical difference when sites immediately distal to the stents were analyzed. There was only a borderline significant difference in the aorta proximal to the covered stents. When the analysis was carried out comparing samples within the same group, no statistical difference was observed either. All stents were patent after four weeks.

Data do recebimento: 03/03/2004

Data da revisão: 18/04/2004

Data da aprovação: 02/05/2004

Conflict of interest: none

Financial source: none

  • 1. Ahn, S. S.; Conception, B. Indications and results of arterial stents for occlusive disease W J Surg 1996; 20: 644-8.
  • 2. Byer A., Ussia G, Galleti, G. Autologous vein lined and vein covered stent in swine arteries: An experimental study to assess and compare patency and intimal hyperplastic response. J Cardiovasc Surg 1998; 39: 393-8.
  • 3. Dolmatch B L, Tio RO, Li XD, Dong YH. Patency and tissue response related to two types of polytetrafluoroethylene-covered stents in the dog. J Vasc Interv Radiol 1996; 7: 641-9.
  • 4.Schürmann K, Vorwerk D, Bucker A, Neuerburg J. Perigraft inflammation due to Dacron-covered stent-grafts in sheep iliac arteries: correlation of MR imaging and histopathologic findings. Cardiovasc Radiol 1997; 204: 757-63.
  • 5.Bartorelli AL, Trabattoni D, Agrifoglio M, Galli S, Grancini L, Spirito R. Endovascular repair of iatrogenic subclavian artery perforations using the hemobahn stent-graft. J Endovasc Ther 2000; 8: 417-21.
  • 8. Consigny P. M. The biology of transluminal angioplasty J Vasc Surg 2000; 31: 1281-3.
  • 9.Yuan JG, Ohki T, Marin ML. The effect of nonporous PTFE-covered stents on intimal hyperplasia following ballon arterial injury in minipigs. J Endovasc Surg 1998; 5: 349-58.
  • 10.Dolmatch, B. L. Healing response to vascular stent-grafts. J Vasc Surg 2000; 31: 1285-9.
  • 11.Palmaz, J. C. Intravascular stents: tissue-stent interactions and design considerations. Am J Roentgenol 1993; 160: 613-8.
  • 12.Pasquinelli G, Freyrie A, Preda P, Curti T, D'addato M, Laschi R. Healing of prosthetic arterial grafts. (Review) Scanning Microsc 1990; 4: 351-62.
  • 13.Karas SP, Gravanis MB, Santoian EC, Robinson KA, Anderberg KA, King SB. III Coronary intimal proliferation after balloon injury and stenting in swine: and animal model of restenosis J Am Coll Cardiol 1992; 20: 467-74.
  • 14.Smet BGJ, Kuntz RE. van der Helm YJ, Pasterkamp G, Borst C, Post MJ. Relationship between plaque mass and neointimal hyperplasia after stent placement in Yucatan micropigs. Radiology 1997; 203: 484-8.
  • 15.Andersen HR, Maeng M, Thorwest M, Falk E. Remodeling rather than neointimal formation explains luminal narrowing after deep vessel wall injury. Circulation 1996; 93: 1716-24.
  • 16.Weatherford DA, Ombrellaro MP, Schaeffer DO, Stevens SL, Sackman JE, Freeman MB, Goldman MH. Healing characteristics of intraarterial stent grafts in an injured artery model. Ann Vasc Surg 1997; 11: 54-61.
  • 17.Tominaga R, Harasaki H, Sutton C, Emoto H, Kambic H, Holmann J. Effects of stent design and serum cholesterol level on the reestenosis rate in atherosclerotic rabbits. Am Heart Journal 1993; 126: 1049-58.
  • 18.Wilson EP, White RA, Kopchok GE, Donayre CE, de Virgilio C, Geselschap JH, Heilbron M. Deployment and healing of an e PTFE encapsulated stent endograft in the canine aorta. Ann. Vasc. Surg 1997; 11: 354-8.
  • 19.Hamilton PW, Allen DC. Measurement in microscopic pathology: Stereology. In: Quantitative clinical pathology. Oxford: Blackwell Science; 1995. p. 3-15.
  • 20.Muller DWM, Ellis SG, Topol EJ. Experimental models of coronary artery reestenosis. J Am Coll Card 1992; 19: 418-32.
  • 21.Schwartz RS, Huber KC, Murphy JG, Edwards WD, Camrud AR, Vliestra RE, Holmes DR, Restenosis and the proportional neointimal response to coronary artery: results in a porcine model. J Am Coll Cardiol 1992; 19: 267-74.
  • 22.Wolf YG, Gertz D, Banai S. Animal models in syndromes of accelerated arteriosclerosis. Ann Vasc Surg 1999; 13 328-38.
  • 23.Schürmann K, Vorwerk D, Upperkamp R, Klosterhalfen B, Bücker A, Günther RW. Iliac arteries; plain and heparin-coated dacron-covered stent-grafts comparede with noncovered metal stents- an experimental study. Radiology 1997; 203:55-63.
  • 24.Cucina A, Sterpetti AV, Borrelli V, Pagliei S, Cavallaro A, D'Angelo LS. Shear stress induces transforming growth factor-beta release by arterial endothelial cells. Surgery 1998; 123: 212-7.
  • 25.Bassiouny HS, Song RH, Glagov S, Choi H, Giddens DP, Zarins CK. Anastomotic intimal hyperplasia: Mechanical injuryor flow induced. J Vasc Surg 1992; 15: 708-17.
  • 26.Hehrlein C, Stentz M, Kinscherf R, Schlösser K, Huttel E, Friederich L, Fehsenfeld P, Kübler W. Pure ß-particle-emitting stents inhibit formation in rabbits. Circulation 1996; 93: 641-5.
  • 27.Chalmers RTA, Hoballah JJ, Sharp WJ, Kresowik TF, Corson JD. The effect of an intraluminal stent on neointimal hyperplasia at an end-to-side polytetrafluoroethylene graft arterial anastomosis. Am J Surg 1994; 168: 85-90.
  • 28.Cambria RA, Farooq MM, Mewissen MW, Freischlag JA, Seabrook GR, Crain MR, Goldblatt MI, Paz-Fumagalli R, Towne JB. Endovascular therapy of iliac arteries: routine application of intraluminal stents does not improve clinical patency. Ann Vasc Surg 1999; 13: 599-605.
  • 29. Martin DR, Katz SG, Kohl RD, Qian D. Percutaneous transluminal angioplasty of infrainguinal vessels. Ann Vasc Surg 1999; 13: 184-7.
  • 30.Chetter IC, Spark JI, Scott JA, Kester RC. Does angioplasty improve the quality of life for claudicants? A prospective study Ann Vasc Surg 1999; 13: 93-103.
  • Correspondence to
    Clandio de Freitas Dutra
    Rua Garibaldi, 476/Sala 701
    95084-060 Caxias do Sul - RS
    Tel/Fax: (54)214-0079
  • 1
    Master dissertation Post-Graduate Program of Federal University of Rio Grande do Sul (UFRGS), Brazil.
    How to cite this article:
    Dutra CF, Pereira AH. Digital morphometric analysis of the aortic wall in pigs following implantation of dacron-covered stents
    versus non-covered stents. Acta Cir Bras [serial online] 2004 May-Jun;19(3). Available from URL:
    http://www.scielo.br/acb. [also in CD-ROM].
  • Publication Dates

    • Publication in this collection
      30 June 2004
    • Date of issue
      June 2004