Open-access Mechanical performance and bacterial leakage in low-cost external hexagon dental implants: implications for post-market surveillance

Abstract

Aim  This study aimed to evaluate low-cost external-hexagon dental implant systems acquired in Brazil between 2014 and 2016, focusing on compliance with international standards, mechanical behavior, microbial leakage, and integrity of the implant–abutment interface.

Methods  Three external hexagon implant systems were selected based on market availability and cost. Fatigue and torsion resistance were assessed in an ISO/IEC 17025-accredited laboratory, in accordance with ISO 14801 and ISO/TS 13498, respectively. Microbial leakage was evaluated using Streptococcus mutans in Brain Heart Infusion (BHI) broth after 14 days of incubation in a qualitative model (presence/absence of contamination). Microgap measurements were obtained by high-resolution micro-computed tomography (micro-CT) using e-Vol DX software with pixel calibration and three-dimensional alignment.

Results  Fatigue tests revealed varying mechanical performance; two groups withstood cyclic loading between 48,655 and 1,291,458 cycles, while one group failed to complete the test. All systems resisted torsional loads approximately four times the recommended tightening torque without catastrophic failure. Microbial leakage was observed in all assemblies. Micro-CT revealed gap ranges of 13.95–47.82 μm (I1), 16.62–18.88 μm (I2), and 19.46–25.35 μm (I3).

Conclusion  Although the tested implants met minimum ISO-oriented mechanical requirements, the presence of clinically relevant microgaps and consistent microbial leakage underscores the need for improved manufacturing precision and quality control. These findings support the importance of structured post-market surveillance and regulatory oversight, particularly for implant systems used in publicly funded oral rehabilitation programs.

Keywords
Dental implants; Product surveillance, postmarketing; Dental abutments; Biofilms; Quality control


Introduction

Dental implants are widely regarded as an excellent option for oral rehabilitation due to their high success and low failure rates1. However, long-term outcomes depend on complex interactions between biological conditions and mechanical stability at the implant–prosthesis interface2. Implant macrodesign and connection features influence load transfer and micromotion, which in turn may affect peri-implant behavior and clinical predictability3.

Mechanical overload, implant–abutment connection design, implant geometry, placement-related factors, bone density, surface finish, and microgap size are among the determinants of marginal bone changes and mechanical complications4. In daily practice, a substantial portion of technical complications involves the implant–abutment interface, particularly screw loosening, which may compromise stability and contribute to subsequent biological and mechanical events5,6.

External hexagon (EH) connections remain widely used due to their prosthetic interchangeability and historical consolidation. Nevertheless, EH systems may be more susceptible to micromovements, torque loss, and interface discontinuities, which can facilitate bacterial penetration and compromise peri-implant health7,8. Therefore, evaluating the mechanical response and interfacial integrity of EH systems remains relevant, especially in settings where cost-driven procurement and broad clinical use coexist.

Achieving mechanical integrity and predictable function requires strict manufacturing tolerances to improve fit and reduce the risk of mechanical failure and misfit-related complications9,10. At the same time, the dental implant market has expanded rapidly, driven by globalization and commercial competition, potentially increasing variability across systems and emphasizing the need for independent technical evaluation11.

In Brazil, access to implant therapy has expanded not only through private low-cost clinics but also through the public health system. Aligned with the National Oral Health Policy12, osseointegrated implants and prosthetic rehabilitation have been funded by the Unified Health System (SUS) through Dental Specialty Centers and hospitals since13. This expansion reinforces the importance of post-market, evidence-based assessment of devices within commonly used market segments.

Previous studies by our group addressed key technical and regulatory issues related to the implant–abutment interface. A prior SEM evaluation of cervical adaptation reported microgaps in the 5–10 μm range with no statistically significant differences among systems or torque levels14. In addition, a regulatory analysis highlighted limitations in the registration and postmarketing surveillance of dental implants in Brazil compared with those in the United States and the European Union15. Together, these findings support the need for standardized evaluations addressing both mechanical performance and interface integrity.

Therefore, this study aimed to evaluate low-cost EH dental implant systems commercially available in Brazil by analyzing their mechanical performance (fatigue and torsion), bacterial leakage potential, and microgap dimensions, within the framework of relevant ISO standards, as an independent technical assessment of a market segment.

Material and Methods

Selection of Implant Systems

To guide the selection and acquisition of titanium dental implants, a market survey was conducted between 2014 and 2016 involving national brands, specialist input, and manufacturer websites. External hexagon systems (≈0.7 mm height; 2.7 mm width) with a standard diameter (ø = 4.1 mm) were selected because this configuration was widely available in the Brazilian market and commonly used due to lower cost and prosthetic interchangeability.

Three commercially available systems acquired during the study period were categorized into groups (I1, I2, and I3), each with similar macrodesign characteristics (an external hexagonal connection and a standard diameter of 4.1 mm). Because implant systems may undergo manufacturing updates and design refinements over time, the commercial brand identities were anonymized in the present report to minimize misinterpretation of these findings as necessarily reflecting current product versions, and to focus the analysis on system-level variability within a defined market segment at the time of acquisition.

The purpose of this investigation was not to generate a commercial ranking or procurement recommendation, but to provide an independent technical assessment of mechanical performance under ISO-oriented tests and interface integrity parameters in low-cost EH systems commonly available during the acquisition period.

Mechanical Testing of Implants

All mechanical tests were conducted at SCiTec Laboratory (Palhoça, SC, Brazil), accredited by CGCRE - Inmetro (CRL 0495) and ISO/IEC 1702516. Fatigue testing according to ISO 14801:201617 and torsion testing according to ISO/TS 13498:201118.

Fatigue Testing

A total of 27 implants were subjected to cyclic loading under dry conditions (20 ± 5 °C), using a bench actuator and force transducers (500 N and 50 N), with measurement instruments calibrated according to ISO/IEC 17025 requirements16. The fatigue life was recorded as the number of cycles to failure.

Torsion Testing

Six implant-abutment assemblies per brand were tested using a uniaxial torsion machine (OFCME 30 Nm), torque sensors (OFTCN model 20KC), and a digital caliper (INSIZE IS 11137-300) under dry conditions at 20 ± 5 °C.

Micro-computed tomography (micro-CT) analysis

High-resolution micro-CT analyses were performed using the Phoenix Nanotom® M system. The acquisition protocol (as provided by the facility) targeted a reconstructed voxel size in the 4–5 µm range and included an X-ray source setting of 90 kV and 100 µA, a 0.1 mm copper filter, 0.5° rotation steps over 360°, and a 500 ms exposure time per projection. Raw image data were reconstructed and exported in RAW format using proprietary image-processing and quantification routines (Radiorum, RJ, Brazil). Voxel calibration was verified using the known screw diameter and internal scale references, and the widest microgap at the implant–abutment interface was identified after three-dimensional alignment along the x, y, and z axes. Measurements were validated against manufacturer specifications to ensure dimensional accuracy. For each implant system, six representative outputs were generated (3D reconstruction of the assembly; zoomed views of the gap region with and without scale; longitudinal section; negative-contrast zoomed image; and transverse section of the screw region). Additionally, a video (Supplementary) illustrating the 3D analysis of one implant was generated using e-Vol DX software.

Microleakage Evaluation

For the microleakage evaluation assay, 30 implant-abutment sets were tested, comprising 10 units from each of the three implant brands. All implants were provided in sealed, sterile conditions. The abutments were removed from their original packaging, placed in new individual containers, and sterilized by autoclaving (Autoclave Sercon® mod. AHMC 5-21 L, São Paulo, Brazil; 30 minutes at 121 °C, 1.0 KGF/cm2).

A strain of Streptococcus mutans, obtained from the Department of Microbiology at the Pontifical Catholic University of Rio Grande do Sul (Porto Alegre, Brazil), was used in the assay. The strain was activated by inoculating 100 μL of the frozen, lyophilized culture into 3 mL of BHI broth (Bacto™ - Brain Heart Infusion, Becton, Dickinson and Company, Maryland, USA). Streptococcus mutans was chosen due to its small size (0.5–1.0 μm), facultative anaerobic nature, and prevalence in oral biofilms. A working bacterial suspension was prepared in BHI at 2.41 × 106 CFU/mL (104 dilution) for sample inoculation.

The test was designed to detect bacterial leakage from the inner portion of the implant-abutment complex into a sterile external environment, based on the assumption that leakage in one direction implies the possibility of reverse contamination19.

Preparation of Culture Media

BHI and BHA media were prepared according to the manufacturer’s instructions (37 g/L and 52 g/L in distilled water, respectively), then autoclaved for 15 minutes at 121 °C. Media were then dispensed into sterile test tubes and Petri dishes.

Cultures were incubated anaerobically at 37 °C for 24 hours (Q-316M2 - Quimis® Scientific Devices Ltda., Diadema, SP). Culture purity was verified using Gram staining, light microscopy, and photomicrography (Zeiss™ Primo Trinocular Star, 40x and 100x Plan-Achromat lenses, 10x eyepiece, Motic™ camera, Motic Image Plus 2.0 software). Colony morphology was confirmed by streaking onto BHA plates for 48 hours under anaerobic conditions.

Pilot Study for Bacterial Growth

To confirm that 0.2 μL of S. mutans in BHI broth was sufficient for bacterial growth, a pilot study was performed. The volume was transferred into 2 mL of BHI medium in Pyrex® test tubes and incubated for 24 hours. Subsequently, 0.1 mL of this culture was plated onto BHA and incubated for another 24 hours. Colony growth consistent with Gram-positive cocci in chains confirmed the suitability of this inoculum volume.

Aseptic Conditions and Experimental Setup

All experimental procedures were conducted in an aseptic environment. The workbench was disinfected with 70 % alcohol and covered with sterile surgical drapes. Instruments—including glassware, tweezers, torque meters, and wrenches—were autoclaved in surgical-grade packaging (Autoclave Sercon® mod. AHMC 5-21 L).

Two operators—a dentist and a microbiologist, both experienced in implantology and microbiology, respectively —performed the procedures using sterile, disposable surgical gowns, gloves (Embramac®, Campinas, Brazil), caps, and masks, which were replaced for each experimental session.

A sterilized bench vise (Eda® Professional Table Lathe - mod-8RN-3, São Paulo, Brazil) with rubber jaws was used to position and stabilize the samples vertically along their long axis to ensure correct torque application.

Under sterile conditions, 0.2 μL of the S. mutans suspension was inoculated into the deepest portion of the internal prosthetic connection of each implant. Immediately after inoculation, the abutments were connected according to the manufacturer’s tightening recommendations; the assembly and torque application were performed by a single operator following a standardized sequence, without knowledge of the implant brand. Each assembly was then individually immersed in sterile BHI broth in sealed test tubes and incubated at 37 °C under anaerobic conditions for 14 days, with tubes maintained in a vertical position. The external broth was inspected daily (days 1–14) for turbidity as an indicator of bacterial leakage. Positive controls consisted of sterile BHI inoculated with 0.2 μL of the same suspension, and negative controls consisted of sterile BHI without inoculation. When turbidity was detected, aliquots of the external broth were serially diluted up to 104 and plated on BHA for confirmatory growth under anaerobic conditions; Gram staining was used to confirm Gram-positive bacteria consistent with the inoculated strain.

Statistical analysis

For the mechanical tests, results are reported as mean, standard deviation, and expanded measurement uncertainty (U, k = 2), as provided by the ISO/IEC 17025–accredited laboratory, in accordance with its validated internal procedures. No additional inferential statistical testing was performed. For the micro-computed tomography and microbiological assays, the data were analyzed descriptively and are presented as observed ranges or qualitative outcomes, without formal hypothesis testing, given the exploratory nature of the study and the limited sample size.

Results

Mechanical Test

According to ISO 14801:201617, fatigue testing revealed differences in mechanical resistance among the evaluated implant systems. Group I1 exhibited failure at a maximum force of 184.2 N, while Group I3 withstood up to 244.1 N. For Group I2, static testing was unsuccessful, preventing subsequent fatigue testing.

Group I1 samples experienced implant-abutment screw fractures between 48,655 and 608,733 cycles, whereas Group I3 fractures occurred between 85,615 and 1,291,458 cycles under cyclic loading, reaching up to 5,000,000 cycles.

Torsion tests, performed according to ISO/TS 13498:201118, demonstrated torque yield limits and maximum torque values that varied across groups. Group I3 had the lowest yield limit (1.31 N×m), while Group I2 reached the highest (1.50 N×m). Maximum torque ranged from 1.48 N×m (I3) to 1.78 N×m (I1). Standard deviation (SD) and uncertainty (U) values confirmed the precision of measurements (Table 1). No catastrophic failures or structural fractures were observed during torsion testing, indicating that all evaluated systems maintained their integrity even under torque values significantly exceeding clinical recommendations.

Table 1
Torsion test results.

Microgap Analysis by Micro-CT

High-resolution micro-computed tomography (micro-CT) was used to evaluate the adaptation at the implant-abutment interface. The analysis revealed microgaps in all implant systems, with considerable variability across groups. Group I1 presented the highest range of microgap values (13.95 μm to 47.82 μm), followed by I3 (19.46 μm to 25.35 μm) and I2 (16.62 μm to 18.88 μm) (Table 2). These findings indicate that despite similar torque conditions, the dimensional adaptation between components differs among implant systems. Figure 1 illustrates a typical high-resolution micro-CT section showing a 30.7 μm gap at the cervical interface, confirming the presence of microdiscontinuities. In addition, a dynamic visualization of the implant-abutment interface — highlighting the presence and continuity of the Microgap — is provided in Supplementary Video 1, captured using e-VOL DX software.

Table 2
Microgap measurements at the implant-abutment interface obtained by high-resolution micro-CT

Figure 1
High-resolution micro-CT section of a dental implant-abutment interface. The image reveals a microgap between the components, measured at 30.69 μm, located at the cervical region of the connection. Scale bar: 250 µm.

Bacterial Contamination

After 14 days of incubation, all implant-abutment assemblies from Groups I1, I2, and I3 exhibited turbidity in the external Brain Heart Infusion (BHI) medium, indicating bacterial contamination. The presence of Streptococcus mutans was confirmed by the characteristic turbidity and subsequent microbiological evaluation (Figure 2). This result demonstrates that bacterial leakage occurred from the inner portion of the implant-abutment complex into the external sterile environment for all tested brands.

Figure 2
In vitro evaluation of bacterial infiltration at the implant-abutment interface using Streptococcus mutans. (A) Control tubes containing sterile BHI broth without implants after 14 days of incubation, showing no turbidity. (B) Experimental groups with implant-abutment assemblies immersed in BHI broth, showing visible turbidity indicative of bacterial contamination. (C) Close-up view of an implant submerged in the culture medium; the red arrow indicates the implant-abutment interface, the potential site of bacterial leakage.

Discussion

The success of oral rehabilitation with dental implants depends on both osseointegration and the integrity of the implant–prosthesis connection, which together ensure stability, resistance, and long-term retention. A dynamic balance between mechanical and biological factors is crucial to maintain this success. High-quality manufacturing and stringent quality control of implant components, as reported in a systematic review of in vitro studies, contribute to precise fit and optimal load distribution, helping preserve crestal bone levels and achieve long-term esthetics19. This perspective aligns with classical criteria for implant success, which emphasize the combined roles of mechanical stability, osseointegration, and sustained clinical function over time20.

Mechanical loading generated during mastication involves complex forces, including normal stresses, shear forces, and multidirectional moments, which require robust mechanical properties in the implant–abutment connection to prevent loosening or failure. In vitro mechanical tests provide essential insights into the material behavior and performance of implant systems, enabling evaluation under controlled loading conditions21,22. Studies comparing internal and external connection designs indicate that internal connections may offer more favorable stress distribution and fatigue behavior. In contrast, external hexagon systems—commonly used in low-cost implants—may be more susceptible to micromovements and torque loss8,23.

In Brazil, the growing demand for implant-based rehabilitation, driven by public health initiatives and the need for cost-effective solutions for underserved populations, makes accessibility and safety a priority. The incorporation of dental implants into the public health system has expanded availability, particularly through the proliferation of affordable clinics12,13. However, this increased accessibility highlights the need for quality assurance. While lower-cost implants may offer financial advantages, they must comply with rigorous regulatory standards to safeguard patient safety and treatment outcomes. The emergence of decentralized production and commercial competition may introduce additional variability in product quality11. Accordingly, this study evaluated whether low-cost external-hexagon implant systems comply with relevant ISO standards and provide acceptable mechanical and microbiological performance, considering their use in public healthcare. The investigation is a technical in vitro assessment of systems acquired between 2014 and 2016, representing products commercially available during that period. As implant designs and manufacturing processes may evolve, the present findings should be interpreted as evidence of variability within a market segment under standardized test conditions, rather than as a statement regarding current commercial versions.

Regarding mechanical behavior, fatigue testing showed that only two of the three systems (Groups I1 and I3) could be fully evaluated, since Group I2 failed at the static step and therefore did not undergo cyclic loading. Among the systems that completed fatigue testing, Group I3 generally withstood higher loads and a broader range of cycles to failure than Group I1, indicating better fatigue performance under the adopted test conditions. These findings are consistent with reports for external-hexagon connections, which show considerable variability in fatigue resistance depending on connection geometry and manufacturing precision24-28. Overall, the mechanical results indicate system-dependent performance dispersion within this market segment.

The torsion tests, designed to determine the maximum torque the implant–abutment connection can withstand without deformation, showed that hexagon deformation occurred only at torque levels approximately 4 times the manufacturer-recommended tightening torque (30–32 N·cm; 0.30–0.32 N·m). This indicates that, from a torsional standpoint, all three systems withstood torque values above those typically used clinically. These results reinforce the importance of maintaining the geometric integrity of the connection for mechanical stability and suggest that it may influence microleakage at the implant–abutment interface. Prior investigations indicate that even minor deviations in component fit may reduce screw joint stability over time8,29.

Microgap formation at the implant–abutment interface is a relevant factor for interface stability and potential microbial penetration. In a prior SEM study by our group, commercially available implant–abutment assemblies marketed in Brazil exhibited cervical misfit/microgap values in the micrometer range, highlighting that measurable discontinuities at the interface may occur even under controlled conditions14. In the present study, SEM analysis also revealed microgaps in the 5–10 μm range, without statistically significant differences across systems or torque conditions. In contrast, high-resolution microCT confirmed microgaps in all groups and revealed differences in dispersion of gap dimensions (Table 2). Notably, one system exhibited a broader range of microgap values, whereas another showed a narrower distribution, suggesting differences in dimensional consistency. Higher values obtained by microCT compared with SEM are compatible with methodological differences, as SEM provides localized two-dimensional measurements. At the same time, microCT offers three-dimensional assessment and may capture the maximum discontinuity within the scanned volume. Previous studies have reported microgaps at implant connections and highlighted their association with microbial leakage and peri-implant biological risk30,31. In addition, interface discontinuities have been associated with crack initiation under cyclic loading, supporting the mechanical relevance of fit precision26.

From a microbiological standpoint, bacterial contamination was detected after a 14-day incubation period in all tested systems, indicating that, under the conditions of this qualitative leakage model, the groups behaved similarly with respect to microleakage. These findings suggest that even relatively small microgaps may permit microbial penetration and are consistent with studies emphasizing the influence of fit accuracy, applied torque, and micromovements between components on bacterial colonization at the implant–abutment interface29,31-33. Other in vitro investigations have shown that internal conical connections may reduce bacterial leakage when compared with external or flat-to-flat connections34, reinforcing the influence of macrogeometry on microbial sealing. Furthermore, systematic reviews indicate that nonoriginal components may contribute to increased microleakage due to manufacturing variability35.

In summary, the present findings indicate that low-cost external-hexagon systems acquired in the study period may show measurable variability in ISO-oriented mechanical testing outcomes and in microgap dimensional consistency. At the same time, bacterial leakage was observed across all evaluated groups under the conditions of this in vitro model, reinforcing that microleakage may occur even when microgaps fall within ranges commonly reported in the literature29,30,33,34. These results should be interpreted as evidence supporting the need for robust quality requirements and systematic post-market evaluation, rather than as a commercial comparison of products.

This in vitro study did not evaluate soft tissue adaptation, peri-implant mucosal sealing, or healing-related parameters, which depend on host biology and clinical conditions. Therefore, clinical extrapolation should be cautious, and future studies combining standardized mechanical testing with biological and clinical outcomes are warranted.

To enhance the safety and performance of dental implants—particularly in the post-market phase—it is essential to establish precise quality requirements supported by robust evidence. The present findings underscore the need for strict quality control throughout all stages of implant manufacturing, especially for systems intended for widespread use in public health programs. Future research should incorporate metallographic analysis, chemical composition, and surface-morphology assessments, focusing on attributes such as texture, component quality, and the implant–abutment microgap. Additionally, creating a georeferenced database of product specifications, failure reports, and performance metrics could support evidence-based decision-making by clinicians, regulators, and manufacturers.

Based on ISO criteria, fatigue and torsion tests demonstrated satisfactory mechanical performance for most of the evaluated implant systems; however, one group could not undergo fatigue testing due to initial testing limitations. Microbial leakage was detected in all implant–abutment assemblies, indicating that contamination of the external environment may occur even under controlled in vitro conditions. Microgap assessment by SEM and microCT revealed interface discontinuities and variability in gap dimensions among the evaluated systems, reinforcing that mechanical adequacy does not necessarily translate into effective sealing at the implant–abutment connection.

This study represents a technical in vitro evaluation of low-cost external hexagon implant systems acquired between 2014 and 2016, reflecting products commercially available during that period. As implant designs and manufacturing processes may evolve, the present findings should be interpreted as evidence of performance dispersion within a market segment under standardized testing conditions rather than as a statement about current commercial versions of specific products.

Taken together, the observed combination of mechanical variability, clinically relevant microgaps, and consistent microbial leakage supports the need for stricter quality control and structured post-market surveillance of dental implant systems, particularly in publicly funded oral rehabilitation programs. Further investigations should prioritize standardized comparisons, including reference systems, quantitative microbial assessments, and expanded interface characterization, to better define performance benchmarks and improve long-term safety and predictability.

Supplementary Materials

Supplementary Video 1

Video 1

Supplementary Video 1 https://youtu.be/liVULtwdFkQ. Three-dimensional high-resolution micro-CT reconstruction of a dental implant-abutment assembly performed using the e-Vol DX software. The sequence demonstrates the axial and sagittal alignment of the components, allowing detailed visualization of the implant-abutment interface. The progressive slicing and rotation of the volume highlight the presence and extent of the microgap along the connection interface. This high-resolution analysis enables the identification of potential regions of mismatch or discontinuity that may contribute to mechanical instability or bacterial infiltration.

Video (Braz J Oral Sci-2026v25e261897 - Supplementary Video 1 - Nano CT Implant 1 e-VOL DX Mike) available on the FOP-UNICAMP YouTube channel, in the “Brazilian Journal of Oral Sciences” playlist

Acknowledgments

This investigation was supported in part by Research Grant E26-/111.326/2014 from FAPERJ-Brazil. The authors are thankful to CAPES-Brazil (Funding code 001) for their support.

References

  • 1 Maroli A, Tretto PHW, Sarkis-Onofre R, Sánchez-Ayala A, Bacchi A. Factors associated with the survival rate of 4,556 dental implants: a practice-based multicenter retrospective cohort study. Braz J Oral Sci. 2025;24:e251408. doi:10.20396/bjos.v24i00.8671408.
    » https://doi.org/10.20396/bjos.v24i00.8671408
  • 2 Sharon E, Engel I. The effect of prosthetic factors on the long-term survival of dental implants: a narrative review. Oral Maxillofac Surg Clin North Am. 2025 Feb;37(1):25-34. doi: 10.1016/j.coms.2024.08.005. Epub 2024 Sep 10.
    » https://doi.org/10.1016/j.coms.2024.08.005
  • 3 Kriswanto K, Jamari J, Andika R, Bayuseno AP, Yusuf AA, Ammarullah MI. Fixture design effects on posterior dental implant stability using finite element analysis (FEA): a systematic review. Head Face Med. 2025 Nov;22(1):2. doi: 10.1186/s13005-025-00566-x.
    » https://doi.org/10.1186/s13005-025-00566-x
  • 4 Nimbalkar S, Dhatrak P, Gherde C, Joshi S. A review article on factors affecting bone loss in dental implants. Mater Today Proc. 2021;43:970-6. doi:10.1016/j.matpr.2020.07.428.
    » https://doi.org/10.1016/j.matpr.2020.07.428
  • 5 Yi Y, Heo SJ, Koak JY, Kim SK. Mechanical complications of implant-supported restorations with internal conical connection implants: a 14-year retrospective study. J Prosthet Dent. 2023 May;129(5):732-40. doi: 10.1016/j.prosdent.2021.06.053. Epub 2021 Sep 1.
    » https://doi.org/10.1016/j.prosdent.2021.06.053
  • 6 Choi S, Kang YS, Yeo ISL. Influence of implant-abutment connection biomechanics on biological response: a literature review on interfaces between implants and abutments of titanium and zirconia. Prosthesis. 2023;5(2):527-38. doi:10.3390/prosthesis5020036.
    » https://doi.org/10.3390/prosthesis5020036
  • 7 Fuda S, Martins BGDS, Castro FC, Heboyan A, Gehrke SA, Fernandes JCH, et al. Marginal bone level and clinical parameter analysis comparing external hexagon and morse taper implants: a systematic review and meta-analysis. Diagnostics (Basel). 2023 Apr;13(9):1587. doi: 10.3390/diagnostics13091587.
    » https://doi.org/10.3390/diagnostics13091587
  • 8 Sammour SR, Ziqi X, Ogawa T, Yoda N. Impact of implant abutment connection designs and cyclic loading on screw stability in dental implants: a systematic review and meta-analysis. J Prosthet Dent. 2026 Feb;135(2):317.e1-317.e14. doi: 10.1016/j.prosdent.2025.09.035. Epub 2025 Oct 15.
    » https://doi.org/10.1016/j.prosdent.2025.09.035
  • 9 Vasconcelos JEL, Matos JDM, Queiroz DA, Lopes GDRS, Lacerda BCGV, Bottino MA, et al. Implant-abutment misfit after cyclic loading: an in vitro experimental study. Materials (Basel). 2022 Aug;15(15):5341. doi: 10.3390/ma15155341.
    » https://doi.org/10.3390/ma15155341
  • 10 Limirio JPJO, Gomes JML, Santiago-Junior JF, Lemos CAA, Rezende MCRA, Pellizzer EP. Influence of manufacturing methods and use of CoCr-based abutments on vertical and horizontal marginal fit and torque loss in implant-supported prostheses. J Appl Oral Sci. 2025 May;33:e20240589. doi: 10.1590/1678-7757-2024-0589.
    » https://doi.org/10.1590/1678-7757-2024-0589
  • 11 Altenstetter C. Medical device regulation in the European Union, Japan and the United States: commonalities, differences and challenges. Innov Eur J Soc Sci Res. 2012;25(4):362-88. doi:10.1080/13511610.2012.723328.
    » https://doi.org/10.1080/13511610.2012.723328
  • 12 Brazilian Ministry of Health. Secretariat of Health Care. Department of Primary Care. [National Oral Health Policy - Smiling Brazil Program]. Brasília: Ministry of Health; 2014. Portuguese.
  • 13 Brazilian Ministry of Health. [Ordinance No. 718, dated December 20, 2010]. Technical Note. Brasília: Ministry of Health; 2010. Portuguese.
  • 14 Ribeiro SB, Jacomo DRES, Calasans-Maia MD, Delgado IF, Granjeiro JM. [Preliminary evaluation of cervical adaptation of the conjunt implant-abutment comercialized in Brazil]. Innov Implant J Biomater Esthet. 2012;7:16-21. Portuguese.
  • 15 Jacomo DRES, Nobre PFS, Calasans-Maia MD, Granjeiro JM, Delgado IF. [Registry and surveillance of dental implants: current panorama in the USA, European Union, Brazil and suggestions for improved systems]. Vigil Sanit Debate. 2016;4(2):105-16. Portuguese. doi:10.3395/2317-269x.00648.
    » https://doi.org/10.3395/2317-269x.00648
  • 16 International Organization for Standardization. ISO 17025:2017. General requirements for the competence of testing and calibration laboratories. Geneva: ISO; 2017.
  • 17 International Organization for Standardization. ISO 14801:2022. Dentistry - Dynamic fatigue test for endosseous dental implants. Geneva: ISO; 2022.
  • 18 International Organization for Standardization. ISO/TS 13498:2011. Dentistry - Torsion test of implant body/connecting part joints of endosseous dental implant systems. Geneva: ISO; 2011.
  • 19 Traversy MC. Fluid and microbial leakage of implant-abutment assembly in vitro. J Dent Res. 1992 July;71(Spec Issue):754.
  • 20 Lee DJ, Son J, Ahn KM. Evaluation of criteria for dental implant success: a systematic literature review. J Implantol Appl Sci. 2024;28(3):123-33. doi:10.32542/implantology.2024013.
    » https://doi.org/10.32542/implantology.2024013
  • 21 Elias CN, Lopes HP. [Dental materials: mechanical tests]. São Paulo: Santos; 2007. Portuguese.
  • 22 Oliscovicz NF, Valente MLC, Marcantonio Junior E, Shimano AC, Reis ACD. [In vitro study of the influence of the shape and surface treatment of dental implants in insertion torque, pullout resistance and resonance frequency]. Rev Odontol UNESP. 2013;42:283-90. Portuguese.
  • 23 Viteri-Ruiz J, Parise-Vasco J, Montesinos-Guevara C. Effectiveness of internal connections compared with external connections for the treatment of single, partial, or total dental implant rehabilitation: overview of reviews. J Int Oral Health. 2023;15(3):226. doi:10.4103/jioh.jioh_237_22.
    » https://doi.org/10.4103/jioh.jioh_237_22
  • 24 Velastegui ML, Agustín-Panadero R, Rico-Coderch A, Amengual-Lorenzo J, Labaig-Rueda C, Solá-Ruiz MF. Influence of different implantoplasty designs on the fatigue resistance of dental implants: a systematic review. J Clin Med. 2025 Aug 29;14(17):6103. doi: 10.3390/jcm14176103.
    » https://doi.org/10.3390/jcm14176103
  • 25 Pereira MB, Fiorin L, Faria ACL, Bonfante EA, Ribeiro RF, Rodrigues RCS. Effect of dental implant macrogeometry on the probability of survival and strain distribution of an implant-abutment set. Dent Med Probl. 2025 May-Jun;62(3):537-45. doi: 10.17219/dmp/174298.
    » https://doi.org/10.17219/dmp/174298
  • 26 Gil FJ, Herrero-Climent M, Lázaro P, Rios JV. Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants. J Mater Sci Mater Med. 2014 Jul;25(7):1825-30. doi: 10.1007/s10856-014-5211-7.
    » https://doi.org/10.1007/s10856-014-5211-7
  • 27 Ribeiro CG, Maia ML, Scherrer SS, Cardoso AC, Wiskott HW. Resistance of three implant-abutment interfaces to fatigue testing. J Appl Oral Sci. 2011 Aug;19(4):413-20. doi: 10.1590/s1678-77572011005000018.
    » https://doi.org/10.1590/s1678-77572011005000018
  • 28 Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resistance of two implant/abutment joint designs. J Prosthet Dent. 2002 Dec;88(6):604-10. doi: 10.1067/mpr.2002.129384.
    » https://doi.org/10.1067/mpr.2002.129384
  • 29 Mao Z, Beuer F, Wu D, Zhu Q, Yassine J, Schwitalla A, et al. Microleakage along the implant-abutment interface: a systematic review and meta-analysis of in vitro studies. Int J Implant Dent. 2023 Sep;9(1):34. doi: 10.1186/s40729-023-00494-y.
    » https://doi.org/10.1186/s40729-023-00494-y
  • 30 Mombelli A, Décaillet F. The characteristics of biofilms in peri-implant disease. J Clin Periodontol. 2011 Mar;38(Suppl 11):203-13. doi: 10.1111/j.1600-051X.2010.01666.x.
    » https://doi.org/10.1111/j.1600-051X.2010.01666.x
  • 31 Kowalski J, Puszkarz AK, Radwanski M, Sokolowski J, Cichomski M, Bourgi R, et al. Micro-CT evaluation of microgaps at implant-abutment connection. Materials (Basel). 2023 Jun;16(12):4491. doi: 10.3390/ma16124491.
    » https://doi.org/10.3390/ma16124491
  • 32 Steinebrunner L, Wolfart S, Bössmann K, Kern M. In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants. 2005 Nov-Dec;20(6):875-81.
  • 33 Nascimento C, Barbosa RE, Issa JP, Watanabe E, Ito IY, Albuquerque Junior RF. Use of checkerboard DNA-DNA hybridization to evaluate the internal contamination of dental implants and comparison of bacterial leakage with cast or pre-machined abutments. Clin Oral Implants Res. 2009 Jun;20(6):571-7. doi: 10.1111/j.1600-0501.2008.01663.x.
    » https://doi.org/10.1111/j.1600-0501.2008.01663.x
  • 34 Buitrago-Osuna A, Espinel-Ramos JA, Henao-Riveros SC, Sarmiento-Delgado ML. Microleakage at implant-abutment interface in internal conical connection implants: an in vitro study. J Oral Health Dent Sci. 2019;3:1-6.
  • 35 Tallarico M, Fiorellini J, Nakajima Y, Omori Y, Takahisa I, Canullo L. Mechanical outcomes, microleakage, and marginal accuracy at the implant-abutment interface of original versus nonoriginal implant abutments: a systematic review of in vitro studies. Biomed Res Int. 2018 Dec;2018:2958982. doi: 10.1155/2018/2958982.
    » https://doi.org/10.1155/2018/2958982

Data availability:

The datasets generated and analyzed during the current study are available from the corresponding author upon reasonable request.

Use of artificial intelligence in manuscript preparation:

Artificial intelligence tools (ChatGPT, OpenAI, for text clarity and coherence, and Grammarly, for grammar checking) were used only to assist in language editing of this manuscript. AI was not used to design experiments, analyze, or interpret data, and all authors critically reviewed the entire content and take full responsibility for the final text.

Corresponding author:

José Mauro Granjeiro Av. Nossa Senhora das Graças, 50, Xerém, Duque de Caxias, 25.250-020, Rio de Janeiro, Brazil. E-mail: jmgranjeiro@inmetro.gov.br +55 21 999883498

Editor:

Dr. Altair A. Del Bel Cury

Conflict of Interest:

The authors declare no conflicts of interest related to this study.

Publication Dates

  • Publication in this collection
    09 Oct 2026
  • Date of issue
    2026

History

  • Received
    13 Dec 2025
  • Accepted
    12 June 2026
location_on
Faculdade de Odontologia de Piracicaba - UNICAMP Avenida Limeira, 901, cep: 13414-903, Piracicaba - São Paulo / Brasil, Tel: +55 (19) 2106-5200 - Piracicaba - SP - Brazil
E-mail: brjorals@unicamp.br
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba