Open-access A novel technique for in vitro assessment of marginal fit in full-arch implant-supported prostheses

Abstract

This study proposes a novel no-screw (NS) technique for assessing marginal fit, comparing its performance with the established single-screw (SS) protocol, also known as the Sheffield test. Four implants and their corresponding abutments were placed in prototyped maxillae. Screw-retained implant-supported frameworks were fabricated from cobalt–chromium (Co-Cr) and titanium alloy containing 6% aluminum and 4% vanadium (Ti) using computer-aided design-computer-aided manufacturing (CAD-CAM) milling technology. Marginal fits were measured using the SS and NS techniques (n = 5 per group). For the NS technique, each framework was placed in the prototyped maxillae and fixed to the mesial side of the four abutments using a liquid-powder technique without screws. Both techniques (SS and NS) were performed 3 times for each framework, and the mean value was calculated. Data were analyzed using two-way analysis of variance (ANOVA) followed by Bonferroni’s test. There was no interaction between technique and material (p = 0.321). However, pooling techniques and materials significantly influenced fit values (p < 0.01). Fit values were lower for the NS technique (41.41 ± 12.46 µm) than for the SS test (97.63 ± 32.13 µm), regardless of material (p = 0.003). Lower fit values were observed for Ti (54.69 ± 34.36 µm) than for Co-Cr (84.35 ± 35.70 µm), regardless of the technique used (p = 0.01). The NS technique yielded lower vertical marginal fit values than the SS technique, regardless of the framework material, suggesting that it may reflect the marginal fit of the framework in the absence of screw-induced preload.

Descriptors
Dental Prosthesis, Implant-Supported; Computer-Aided Design; Dental Marginal Adaptation

Introduction

In implant-supported prosthodontics, the fit between a prosthesis and its supporting structures, defined as the degree of adaptation between these components, is critical for long-term clinical success.1-3 Within this context, passive fit represents an ideal form of adaptation in which an implant-supported prosthesis is seated without generating mechanical stress or strain in the prosthetic components, implants, or peri-implant bone.1 However, achieving an absolute fit is technically challenging, particularly in the case of screw-retained implant-supported prostheses, because deviations may be introduced over successive stages of the workflow, including impression making or intraoral scanning and framework fabrication.2,4 Accordingly, after the recommended preload has been applied to all screws, achieving the optimum fit possible with minimal distortion is essential to reduce the risk of mechanical complications such as screw loosening and framework fracture, as well as biological complications such as marginal bone loss.4-6 Various methods have been developed to improve the fit of screw-retained implant-supported restorations and reduce stresses transmitted to the implant-supported system, including spark erosion and computer-aided design-computer-aided manufacturing (CAD-CAM) using subtractive and additive techniques.2,7,8 Among CAD-CAM systems, frameworks fabricated by subtractive manufacturing have been associated with improved fit compared to additive techniques before and after ceramic application.8

Quantitatively, there is no universally accepted threshold that defines clinically acceptable vertical marginal fit. Reported values vary according to framework span and extent,8-10 measurement method,11 and implant configuration.3 Greater discrepancies are generally observed in full-arch prostheses compared to partial prostheses.7-10,12-14 Accordingly, a systematic review focusing on vertical fit measurements reported acceptable values ranging from 30 to 160 µm.3 Clinical studies provide further explanations for this variability.15,16 Jemt and Book15 evaluated screw-retained implant-supported prostheses and found no completely passive fit in any of them. The mean three-dimensional gap was 111 ± 59 µm in the 1-year group and 91 ± 51 µm in the 5-year group, with no significant correlation between fit and changes in marginal bone levels. In a separate long-term study, Jokstad and Shokati16 reported fit values ranging from 95 to 232 µm after a mean follow-up of 19 years. The gap between the prosthesis and its supporting structures was not significantly correlated with marginal bone loss (R2 = 0.04; p = 0.29).16 However, greater gaps were associated with a higher frequency of screw-related adverse events, including loosening or fracture of prosthetic or abutment screws. Taken together, these findings suggest some degree of biological tolerance within the observed ranges;15,16 however, they do not establish 111 µm or approximately 230 µm as universally acceptable clinical thresholds.

Framework and/or prosthesis fit measurements have mainly been used to assess material behavior and manufacturing accuracy.4,7,9,10,17,18 Among the established methods, the single-screw (SS) protocol, also known as the Sheffield test, is widely used in clinical and experimental settings.3,4,19 This protocol consists of securing the framework with a single screw at one terminal abutment and measuring the marginal gap at the opposite end.2,7-9,19-21 Despite its widespread use, this method is inherently two-dimensional and therefore unable to detect horizontal distortions or internal discrepancies.3,7 Limitations regarding its accuracy, reproducibility, and reliability have also been reported.3 In experimental settings, fit is commonly assessed by optical or scanning electron microscopy, whereas in clinical practice the gap is typically evaluated by direct visual inspection, use of a dental explorer, and periapical radiography.4 Optical microscopy is a relatively straightforward procedure, which may explain its good repeatability and high measurement precision.11 Although computed tomography is also used for this purpose,22,23 its high cost and lower efficiency compared to conventional two-dimensional methods limit its routine application.13 Taken together, the available protocols differ in the mechanical configuration they impose during measurement and in the parameters they capture. There is currently no established consensus reference method for the in vitro assessment of marginal adaptation, a fact that supports the investigation of alternative approaches.11

Given these limitations, developing alternative methods to evaluate marginal fit may improve assessment of implant-supported frameworks, particularly those with long-span and curved designs.7,8,24 Although all-screw fixation is clinically relevant, tightening all screws during assessment may mask the framework’s inherent marginal fit by forcing contact with the implant abutments, reducing the ability to detect discrepancies.3,11 Similarly, the SS test may distort marginal fit measurements since tightening a single screw can widen the gap at the opposite end and consequently underestimate the actual value.19 Additionally, the SS technique is highly sensitive to setup parameters, projection quality, and operator variability, which can compromise measurement precision.11 In parallel, advances in CAD-CAM technology have enabled the fabrication of implant-supported frameworks from materials such as cobalt-chromium (Co-Cr) and titanium alloy containing 6% aluminum and 4% vanadium (Ti), providing cost-effective alternatives to conventional casting while reducing processing steps and minimizing fabrication-related distortion.4,10,22,24-26 Previous studies reported similar fit values for milled Co-Cr and Ti frameworks before and after veneering application;10 thus, both materials are suitable for evaluating new assessment methods. Based on these considerations, a novel no-screw (NS) technique was developed as an alternative approach to evaluating marginal fit without screw fixation. By allowing the framework to seat under unconstrained conditions, the NS technique aims to minimize operator-dependent influences and provide measurements that more closely reflect the framework’s inherent marginal fit.

Therefore, this study proposes the novel NS technique as an alternative method for in vitro assessment of marginal fit, comparing its performance with the SS technique using CAD-CAM-milled Co-Cr and Ti screw-retained implant-supported frameworks. The null hypothesis is that the assessment technique, NS or SS, would not affect marginal fit measurements in Co-Cr or Ti frameworks.

Methodology

An edentulous maxillary replica was manufactured from the standard tessellation language (STL) file reported in previous studies.7,8 Following the All-on-4 concept (Figure 1A), the replica received four Easy Grip Porous external hexagon (EH) implants (Conexão Sistemas de Prótese, Arujá, Brazil): 2 straight implants were placed in the anterior region corresponding to the lateral incisors, and 2 distal implants were placed in the posterior region corresponding to the second premolars at a 45-degree distal angulation.27 The anterior implants measured 4.1 × 11.5 mm and the posterior implants measured 4.1 × 13 mm. All implants were inserted with a torque of 45 Ncm. Straight-profile abutments measuring 4 × 4 mm were connected to the anterior implants, whereas 30-degree angled-profile abutments measuring 4 × 4 mm were connected to the posterior implants. All abutments were tightened to 20 Ncm. The assembled edentulous maxillary replica, implants, and abutments were then scanned using manufacturer-recommended scan bodies (Scan Connect Micro-unit, Conexão Sistemas de Prótese, Arujá, Brazil) and a laboratory light scanner (Ceramill Map 400+, Amann Girrbach, Pforzheim, Germany). Next, the framework CAD file was designed from the resulting digital model using Ceramill Mind software (Amann Girrbach, Pforzheim, Germany) to restore 12 teeth (teeth 3 to 14, according to the Universal Numbering System of the American Dental Association) in a maxillary screw-retained implant-supported framework. Frameworks were milled from Co-Cr (Starbond Easy Disc, Scheftner, Mainz, Germany) and Ti alloy (Starbond Ti5 Disc, Scheftner, Mainz, Germany) blocks using a CNC milling machine (CNC D15W, Yenadent, Istanbul, Turkey). Five specimens were fabricated per material group.

Figure 1
Prototyped edentulous maxilla used as the master model for fit readings. A, Maxilla with four implants placed according to the All-on-4 concept: two straight implants positioned in the region of the lateral incisor and two distal implants positioned in the region of the second premolar at a 45-degree distal angulation. B, Master model placed on the octagonal base for stabilization during assembly. C, Standard measurement positions on the octagonal base of the master model.

Marginal fit between the abutments and the frameworks was measured using a precision microscope (×120 magnification; VMM-100-BT, Walter UHL, Asslar, Germany) coupled to a KC-512NT camera (Kodo BR Electronics Ltd., Seoul, Korea) and Quadra-Check QC 220-HH (Metronics Inc., Minneapolis, USA). To ensure consistent positioning of the prototyped replica throughout all measurements, a custom-made octagonal polymer support (Figure 1B) was printed and used during the assessment.2 Marginal fit was then measured at all abutment–framework interfaces (Figure 1C). A previously calibrated examiner (T.B.) performed the readings 3 times on the buccal and lingual sides (intraclass correlation coefficient of 0.996; p < 0.001).

For the SS technique following the Sheffield test,4,19,28 each framework was positioned on the master model (Figure 2A). The screw at one distal abutment was tightened to 10 Ncm using a digital torque meter (TQ-8800; Lutron, Taipei, Taiwan) and a hexagonal screw (Conexão Sistemas de Prótese, Arujá, Brazil),7,8 following manufacturers’ recommendations. The master model was then placed on the octagonal base to ensure standardized positioning during the readings. The microscope was directed toward the center of each framework–abutment interface, and marginal fit was measured on the buccal and lingual sides.7,8,10 The screw was then loosened at the first distal abutment and tightened at the opposite distal abutment (applying the same 10-Ncm torque with a digital torque meter), and the fit-reading procedure was repeated.7,8,10 This sequence was performed 3 times, and a final mean fit value was calculated for each framework.

Figure 2
Schematic illustration of the fit-reading protocol according to the technique used. A, Single-screw protocol and marginal fit assessment with the prosthetic screw tightened to 10 Ncm on a single abutment at one end. Marginal fit measurements were then recorded along the interfaces between the framework and abutments, extending to the opposite side. The magnified image on the tightened side shows enhanced fit resulting from pulling the interfaces closer together. On the opposite side, the circular image shows that the framework created a lever-arm effect, compromising the fit at the non-tightened end. B, Novel no-screw technique, in which the framework was secured in place with a low-shrinkage resin applied to the mesial side of each abutment. The magnified images show the technique during application on the right side and a representative interface for fit reading.

For the NS technique (Figure 2B), the frameworks were passively seated on the edentulous maxillary replica; however, the liquid-powder technique was applied instead of screw fixation using self-curing Pattern Resin LS (GC Corp., Tokyo, Japan). Only one examiner (T.B.) conducted the following procedure: a brush moistened with monomer was placed in contact with the polymer powder to form a small resin sphere, which was then applied to the mesial side of each abutment–framework interface to stabilize the framework on the master model. Only one brushstroke was standardized for each abutment at the beginning of the measurement procedure. The NS technique assessed the same sites as the SS technique. Measurements were performed 3 times, without changing the brushstroke in each abutment, and a final mean fit value was calculated for each framework.

Data were analyzed using IBM SPSS Statistics for Windows (v.21.0, IBM Corp., Armonk, USA). Normality of the data was tested using the Shapiro-Wilk test. Two-way repeated measures ANOVA (factor 1: material, Co-Cr and Ti; factor 2: technique, SS and NS) was then applied to assess the effects of the fit assessment technique (SS and NS) and framework material (Co-Cr and Ti) on fit values. Pairwise comparisons between groups were conducted using Bonferroni’s post hoc test, with the significance level set at α = 0.05. The reliability of each fit-reading technique was determined by calculating the intraclass correlation coefficient across the three independent readings. A priori sample size calculation was performed using G*Power 3.1.9.4 (Heinrich Heine Universität Düsseldorf). The F-test family and repeated measures ANOVA with a within-between interaction were selected according to the two-way repeated measures ANOVA model subsequently applied to the data, with fit assessment technique (SS and NS) as the within-subject factor and framework material (Co-Cr and Ti) as the between-subject factor. The following parameters were entered: 2 groups, 2 measurements, correlation among repeated measures of 0.5, non-sphericity correction ε = 1, α = 0.05, and power (1−β) = 0.95. Because the design included 2 groups and 2 measurements, the within-subject main effect and the within-between interaction share the same non-centrality parameter and degrees of freedom (numerator: df = 1; denominator: df = n−2); thus, a single calculation applies to both. The minimum difference to be detected was set at 30 µm, corresponding to the lower limit of the acceptable vertical fit range reported in a systematic review.3 A standard deviation of 20 µm was assumed, corresponding to an effect size (f) of 0.75. Under these conditions, 10 frameworks (5 per material group) were required. This sample size also agrees with previous in vitro studies investigating full-arch implant-supported frameworks based on the All-on-4 concept, which adopted the same number of specimens per group.7,8

Results

The three repeated readings for each fit assessment technique demonstrated excellent reliability, with high intraclass correlation coefficients for both material and technique: 0.999 for Co-Cr and 0.998 for Ti with the SS technique, and 0.998 for Co-Cr and 0.997 for Ti with the NS technique. Two-way repeated measures ANOVA showed that framework material (p = 0.01) and fit assessment technique (p = 0.003) significantly influenced marginal fit values, whereas their interaction was not significant (p = 0.321), indicating that the effect of the assessment technique was independent of the framework material (Table 1). Regarding the CAD-CAM framework material, Ti provided significantly lower marginal fit values than Co-Cr, with mean values of 54.69 ± 34.36 µm for Ti and 84.35 ± 35.7 µm for Co-Cr (p = 0.01). Similarly, the NS technique resulted in significantly lower marginal fit values than the SS technique (41.41 ± 12.46 µm versus 97.63 ± 32.13 µm), corresponding to a reduction of approximately 58% in gap and consequently improved marginal fit (Table 2).

Table 1
Summary of two-way repeated measures analysis of variance of the effects of fit-reading technique and framework material on marginal fit measurements.

Table 2
Mean ± standard deviation of marginal fit (in µm) according to fit-reading technique and framework material.

Discussion

This study proposed a novel NS in vitro fit assessment technique and compared it with the traditional Sheffield test using the SS technique4,19,28 in screw-retained implant-supported frameworks fabricated from Ti and Co-Cr. The observed differences resulted in the rejection of the null hypothesis, since either technique, NS or SS, and either framework material, Co-Cr or Ti, significantly influenced marginal fit assessment. However, no significant interaction was observed between technique and material, indicating that the effect of the assessment method did not depend on the framework material. Because key experimental factors were standardized across the two techniques, including the manufacturing process, microscope, calibrated examiner, and simplified support model, the differences in marginal fit values are most likely attributed to the independent variables investigated. In the control group (SS technique), the protocol consists of tightening the terminal screw and then measuring the interface between the framework and abutments.19,28 The tightened area tends to show optimum fit because the applied torque draws the components closer together. Consequently, a lever arm forms along the framework, increasing the measured value at the non-tightened end and creating a load configuration that does not necessarily match that of a prosthesis in which all screws are tightened.

The recorded gap values were higher for the SS technique (97.63 ± 32.13 µm) than for the NS technique (41.41 ± 12.46 µm). While conceiving the study, we expected a more consistent distribution of fit values when the material groups were pooled, given the well-established reproducibility of the CAD-CAM manufacturing process.2,7-10,12,17 However, the NS technique showed lower dispersion, as reflected by lower variability. Under controlled laboratory conditions, the SS technique may reveal marginal discrepancies more clearly.2,7,8,10,17 In contrast, the NS technique may better simulate framework seating during laboratory procedures and under clinical conditions before intraoral evaluation of the prosthetic device. The rationale for the NS technique is that it assesses marginal fit at the abutment–framework interface, with the framework seated in the absence of screw-induced preload. The small resin increments applied to the mesial side of each abutment only prevent displacement during measurement and are not intended to seat the framework. Unlike the SS test, which creates a lever arm that may increase the discrepancies recorded on the opposite side of the framework, and unlike the all-screws approach, in which screw tightening actively forces the framework into contact with the implant abutments, the NS technique allows the framework to assume its natural seating position under its own geometry and manufacturing accuracy.11,19,34 Because no screw preload is applied, the NS technique characterizes the marginal fit of the framework before screw-induced seating occurs. This technique therefore is not intended to reproduce the mechanical configuration obtained once all prosthetic screws are tightened: the two protocols describe different mechanical conditions, and our data do not support the use of one over the other. Values recorded with the NS technique may be less affected by protocol-induced distortion; however, this interpretation is based on the protocol design rather than a validated comparison. The absence of screw-induced leverage may nonetheless be of interest for extensive frameworks supported by four or more implants, in which the mechanical behavior of long-span structures may affect SS assessment.2,8 Since only one framework design was tested here, this assumption must be confirmed in studies involving different span lengths9 and numbers of implants.2,10 Finally, this study could not determine whether the values recorded with either protocol are related to clinical outcomes such as screw loosening, framework strain, or biomechanical overload because it did not analyze functional loading or aging and did not perform clinical follow-up. Longitudinal clinical studies should test this hypothesis.33-35

The significantly higher and more variable values obtained with the SS technique may also be related to the mechanics of screw fastening. Slight deviations in screw angulation from the axial direction, variations in contact between the screw head or threads and the framework, plastic deformation in previously used prosthetic screws, and inconsistencies in torque application by the examiner may influence the measurements. Therefore, proper torque training and meticulous execution are essential when this technique is used.2,7,8,10,17 In contrast, these potential sources of distortion are absent in the NS technique since it eliminates the lever arm effect. By avoiding screw-induced torque, this approach may reduce distortion in marginal fit measurements. In the NS technique , the framework is passively positioned on the abutments, and small increments of self-curing resin applied to the mesial side are used only to stabilize the framework during analysis. This minimizes the bias introduced by SS tightening during in vitro testing,4,28 thereby reducing distortion in marginal fit assessment.

The NS technique differs from the SS test in two operative aspects. While the SS technique requires measurements at two ends to calculate a mean value, the NS technique eliminates this step. Furthermore, the torque applied by the examiner is not a critical factor in the NS technique since the framework passively settles into place. However, a calibrated examiner remains essential for tasks such as microscope use.11 This streamlined approach is particularly advantageous for the evaluation of extensive and more complex framework designs, for which conventional assessment techniques may be challenging.2,3,7,8,10,11,17 Additionally, the simplicity of the NS technique makes it a valuable tool for preclinical evaluations, for example, assessing framework fit on an analog cast or 3D-printed model before intraoral try-in. This process can enhance clinicians’ dexterity and sensitivity and enable them to identify and address potential gap issues before framework veneering or clinical installation. The NS technique also facilitates the early identification of possible failures, allowing clinicians to plan corrections for nonideal fits before clinical evaluation. Moreover, this technique could be an effective tool for laboratory quality control by providing valuable overall fit data that help ensure the precision of manufactured frameworks. These potential applications were not investigated in the present study and would require validation before being recommended.

Another important finding of this study relates to the materials used to fabricate the frameworks. Co-Cr is widely used because of its relatively low cost and favorable ceramic veneering properties,9,12,25,29,30 whereas Ti is valued for its high strength-to-weight ratio, making it suitable for lightweight dental frameworks.31-33 Material choice significantly influenced the fit of screw-retained implant-supported prostheses, regardless of the fit assessment technique applied. Ti frameworks provided lower fit values (54.69 ± 34.36 µm) than Co-Cr frameworks (84.35 ± 35.7 µm), consistent with previous studies that used only the SS technique.7,34 Previous publications reported lower and less variable vertical marginal fit values for Ti frameworks (median: 35.63 µm; IQR: 30.71–38.08 µm) than for Co-Cr frameworks (median: 97.38 µm; IQR: 39–193.8 µm),7,34 whereas in the present study the variability was more consistent between materials. One possible explanation is that both alloys were milled using the same workflow. Furthermore, the cited studies7,34 and the present study evaluated frameworks based on a comparable All-on-4 concept. Variations in fit values between materials may be related to the distinct mechanical properties of each alloy, particularly their elastic modulus, as reported by the manufacturers (Ti = 110 GPa; Co-Cr = 191 GPa). Accordingly, the higher elastic modulus of Co-Cr renders the framework less flexible and less able to accommodate minor dimensional deviations during fit assessment.18 In addition, greater stress is expected to develop within frameworks with a high elastic modulus. In this context, a finite element study showed that, at an assumed vertical gap of 100 µm, increasing the elastic modulus of the implant framework considerably increased framework stress, especially when comparing Ti to Co-Cr.30 Because Co-Cr frameworks tend to exhibit higher fit values than Ti frameworks, corrective techniques, such as evaluating an analogic index, should be considered for Co-Cr implant-supported prostheses before framework milling. Ceramic veneering of metallic frameworks is another potential source of error as previously reported in an All-on-6 study, particularly because of differences between the coefficient of thermal expansion of the framework material and that of the ceramic.10 Spark erosion may be an efficient corrective choice, especially for Co-Cr frameworks in either the All-on-4 or All-on-6 concept.7,10 Further studies are needed to test these assumptions and address this information gap in the literature.

Despite the well-established reliability and relevance of the SS test in in vitro and in vivo studies,11,19,28 it is important to note that the novel NS technique eliminates the influence of screw head and thread interactions on framework seating. Furthermore, the NS technique showed lower value dispersion in this study. This novel method provides a practical approach to evaluating framework fit in vitro and has the potential for broader applications, including visualization on a patient’s dental gypsum cast or printed model. The technique was proposed as an alternative fit assessment protocol that involves fewer operative steps than the SS test. The technique was simple to implement under the conditions tested. However, we did not assess its applicability to other manufacturing methods, framework lengths, and clinical scenarios. Further in vitro and clinical studies are needed before the NS technique can be recommended for routine laboratory or clinical use.

The findings of this study should be interpreted in light of the following limitations. The evaluation was restricted to Ti and Co-Cr frameworks without ceramic veneering, which does not fully reproduce the clinical configuration of implant-supported prostheses. In addition, only one CAD-CAM manufacturing workflow and one implant–abutment system were investigated using commercially available materials and equipment according to manufacturers’ recommendations. Although clinical performance can be influenced by subsequent procedures such as veneering, finishing, and prosthesis installation, the accuracy and biomechanical behavior of the metallic framework are the foundation of the final restoration.17,34 Assessing the framework before ceramic application was therefore essential to isolate the influence of the manufacturing process and better understand its contribution to overall prosthetic outcome. Another limitation is the potential effect of resin stabilization in the NS technique on framework positioning; however, this risk was minimized by using a low-shrinkage self-curing resin according to the manufacturer’s instructions. Additionally, assessing internal and horizontal adaptation would provide a more complete understanding of framework fit and should be performed in future analyses.17 Despite the inherent limitations of in vitro models, biomechanical investigations are essential for bridging laboratory findings and clinical application, providing preliminary evidence on the performance of restorative materials and prosthetic designs and guiding future clinical studies to validate their long-term effectiveness and predictability. Because the present results describe framework behavior under standardized laboratory conditions, they cannot be directly extrapolated to the clinical performance or longevity of screw-retained implant-supported prostheses.

Conclusion

Within the limitations of this in vitro study, both framework material and fit assessment technique were associated with significant differences in the marginal fit values of CAD-CAM screw-retained implant-supported frameworks. Ti frameworks exhibited lower marginal fit values than Co-Cr frameworks, while the NS technique yielded lower values than the conventional SS technique under the conditions tested. These findings indicate the potential of the NS technique as an in vitro approach for marginal fit assessment. However, further in vitro and clinical studies using independent confirmatory methods, including assessments performed with all prosthetic screws tightened, are needed to establish the measurement validity and clinical applicability of the NS technique, particularly for extensive implant-supported frameworks.

References

  • 1 The glossary of prosthodontic terms 2023: tenth edition. J Prosthet Dent. 2023 Oct; 130(4):e7-126. https://doi.org/10.1016/j.prosdent.2023.03.002
    » https://doi.org/10.1016/j.prosdent.2023.03.002
  • 2 Velôso DV, Barbin T, Silva LR, Barão VA, Mesquita MF, Borges GA. Additive manufacturing of CAD-CAM complete arch fixed implant restorations supported by six implants: prosthetic screw stability and marginal precision under masticatory simulation. J Prosthet Dent. 2025 Sep;134(3):783.e1-9. https://doi.org/10.1016/j.prosdent.2025.04.042
    » https://doi.org/10.1016/j.prosdent.2025.04.042
  • 3 Abdelrehim A, Etajuri EA, Sulaiman E, Sofian H, Salleh NM. Magnitude of misfit threshold in implant-supported restorations: a systematic review. J Prosthet Dent. 2024 Sep;132(3):528-35. https://doi.org/10.1016/j.prosdent.2022.09.010
    » https://doi.org/10.1016/j.prosdent.2022.09.010
  • 4 Abduo J, Bennani V, Waddell N, Lyons K, Swain M. Assessing the fit of implant fixed prostheses: a critical review. Int J Oral Maxillofac Implants. 2010;25(3):506-15.
  • 5 Skalak R. Biomechanical considerations in osseointegrated prostheses. J Prosthet Dent. 1983 Jun;49(6):843-8. https://doi.org/10.1016/0022-3913 (83)90361-X
    » https://doi.org/10.1016/0022-3913 (83)90361-X
  • 6 Michalakis KX, Hirayama H, Garefis PD. Cement-retained versus screw-retained implant restorations: a critical review. Int J Oral Maxillofac Implants. 2003;18(5):719-28.
  • 7 Barbin T, Silva LD, Velôso DV, Borges GA, Presotto AG, Barão VA, et al. Biomechanical behavior of CAD/CAM cobalt-chromium and zirconia full-arch fixed prostheses. J Adv Prosthodont. 2020 Dec;12(6):329-37. https://doi.org/10.4047/jap.2020.12.6.329
    » https://doi.org/10.4047/jap.2020.12.6.329
  • 8 Barbin T, Velôso DV, Silva LR, Borges GA, Presotto AG, Barão VA, et al. 3D metal printing in dentistry: an in vitro biomechanical comparative study of two additive manufacturing technologies for full-arch implant-supported prostheses. J Mech Behav Biomed Mater. 2020 Aug;108:103821. https://doi.org/10.1016/j.jmbbm.2020.103821
    » https://doi.org/10.1016/j.jmbbm.2020.103821
  • 9 Presotto AG, Barão VA, Bhering CL, Mesquita MF. Dimensional precision of implant-supported frameworks fabricated by 3D printing. J Prosthet Dent. 2019 Jul;122(1):38-45. https://doi.org/10.1016/j.prosdent.2019.01.019
    » https://doi.org/10.1016/j.prosdent.2019.01.019
  • 10 Silva LR, Velôso DV, Barbin T, Borges GA, Presotto AG, Mesquita MF. Can ceramic veneer spark erosion and mechanical cycling affect the accuracy of milled complete-arch frameworks supported by 6 implants? J Prosthet Dent. 2021 Dec;126(6):772-8. https://doi.org/10.1016/j.prosdent.2020.05.038
    » https://doi.org/10.1016/j.prosdent.2020.05.038
  • 11 He Z, Bennani V, Veerasamy A. Precision of measurement methods for assessing the fit of screw-retained implant-supported fixed partial dentures: a systematic review of in vitro studies. J Prosthet Dent. 2026 Apr;135(4):753.e1-11. https://doi.org/10.1016/j.prosdent.2025.11.019
    » https://doi.org/10.1016/j.prosdent.2025.11.019
  • 12 Presotto AG, Cordeiro JM, Presotto JG, Rangel EC, Cruz NC, Landers R, et al. Feasibility of 3D printed Co-Cr alloy for dental prostheses applications. J Alloys Compd. 2021 May;862:158171. https://doi.org/10.1016/j.jallcom.2020.158171
    » https://doi.org/10.1016/j.jallcom.2020.158171
  • 13 Rodrigues SA, Presotto AG, Barão VA, Consani RL, Nóbilo MA, Mesquita MF. The role of welding techniques in the biomechanical behavior of implant-supported prostheses. Mater Sci Eng C. 2017 Sep;78:435-42. https://doi.org/10.1016/j.msec.2017.04.090
    » https://doi.org/10.1016/j.msec.2017.04.090
  • 14 Presotto AG, Oliveira LV, Pisani MX, Barão VA, Mesquita MF. Influence of spark erosion on the fit of screw-retained Co-Cr fixed complete denture frameworks veneered with different materials. J Prosthet Dent. 2018 May;119(5):797-803. https://doi.org/10.1016/j.prosdent.2017.07.003
    » https://doi.org/10.1016/j.prosdent.2017.07.003
  • 15 Jemt T, Book K. Prosthesis misfit and marginal bone loss in edentulous implant patients. Int J Oral Maxillofac Implants. 1996;11(5):620-5.
  • 16 Jokstad A, Shokati B. New 3D technologies applied to assess the long-term clinical effects of misfit of the full jaw fixed prosthesis on dental implants. Clin Oral Implants Res. 2015 Oct;26(10):1129-34. https://doi.org/10.1111/clr.12490
    » https://doi.org/10.1111/clr.12490
  • 17 Barbin T, Borges GA, Jardini AL, Mesquita MF. Exploring hot isostatic pressing as a thermo-mechanical treatment for additive and milled CAD-CAM screw-retained frameworks: trueness, marginal and internal gap assessments. Prog Addit Manuf. 2025 Aug;10(8):4995-5007. https://doi.org/10.1007/s40964-024-00881-w
    » https://doi.org/10.1007/s40964-024-00881-w
  • 18 Abduo J, Lyons K. Effect of vertical misfit on strain within screw-retained implant titanium and zirconia frameworks. J Prosthodont Res. 2012 Apr;56(2):102-9. https://doi.org/10.1016/j.jpor.2011.09.001
    » https://doi.org/10.1016/j.jpor.2011.09.001
  • 19 Tan KB, Rubenstein JE, Nicholls JI, Yuodelis RA. Three-dimensional analysis of the casting accuracy of one-piece, osseointegrated implant-retained prostheses. Int J Prosthodont. 1993;6(4):346-63.
  • 20 Yilmaz B. CAD-CAM high-density polymer implant-supported fixed diagnostic prostheses. J Prosthet Dent. 2018 May;119(5):688-92. https://doi.org/10.1016/j.prosdent.2017.06.022
    » https://doi.org/10.1016/j.prosdent.2017.06.022
  • 21 Bhering CL, Marques IS, Takahashi JM, Barão VA, Consani RL, Mesquita MF. Fit and stability of screw-retained implant-supported frameworks under masticatory simulation: influence of cylinder type. J Prosthodont. 2016 Aug;25(6):459-65. https://doi.org/10.1111/jopr.12349
    » https://doi.org/10.1111/jopr.12349
  • 22 Al-Meraikhi H, Yilmaz B, McGlumphy E, Brantley W, Johnston WM. In vitro fit of CAD-CAM complete arch screw-retained titanium and zirconia implant prostheses fabricated on 4 implants. J Prosthet Dent. 2018 Mar;119(3):409-16. https://doi.org/10.1016/j.prosdent.2017.04.023
    » https://doi.org/10.1016/j.prosdent.2017.04.023
  • 23 Yilmaz B, Kale E, Johnston WM. Marginal discrepancy of CAD-CAM complete-arch fixed implant-supported frameworks. J Prosthet Dent. 2018 Jul;120(1):65-70. https://doi.org/10.1016/j.prosdent.2017.11.021
    » https://doi.org/10.1016/j.prosdent.2017.11.021
  • 24 Karl M, Graef F, Wichmann M, Krafft T. Passivity of fit of CAD/CAM and copy-milled frameworks, veneered frameworks, and anatomically contoured, zirconia ceramic, implant-supported fixed prostheses. J Prosthet Dent. 2012 Apr;107(4):232-8. https://doi.org/10.1016/S0022-3913 (12)60067-5
    » https://doi.org/10.1016/S0022-3913 (12)60067-5
  • 25 Örtorp A, Jönsson D, Mouhsen A, Vult von Steyern P. The fit of cobalt-chromium three-unit fixed dental prostheses fabricated with four different techniques: a comparative in vitro study. Dent Mater. 2011 Apr;27(4):356-63. https://doi.org/10.1016/j.dental.2010.11.015
    » https://doi.org/10.1016/j.dental.2010.11.015
  • 26 Guess PC, Att W, Strub JR. Zirconia in fixed implant prosthodontics. Clin Implant Dent Relat Res. 2012 Oct;14(5):633-45. https://doi.org/10.1111/j.1708-8208.2010.00317.x
    » https://doi.org/10.1111/j.1708-8208.2010.00317.x
  • 27 Bhering CL, Mesquita MF, Kemmoku DT, Noritomi PY, Consani RL, Barão VA. Comparison between all-on-four and all-on-six treatment concepts and framework material on stress distribution in atrophic maxilla: a prototyping guided 3D-FEA study. Mater Sci Eng C. 2016 Dec;69:715-25. https://doi.org/10.1016/j.msec.2016.07.059
    » https://doi.org/10.1016/j.msec.2016.07.059
  • 28 Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants. 1991;6(3):270-6. https://doi.org/10.1097/00008505-199205000-00019
    » https://doi.org/10.1097/00008505-199205000-00019
  • 29 Antanasova M, Kocjan A, Kovac J, Žužek B, Jevnikar P. Influence of thermo-mechanical cycling on porcelain bonding to cobalt-chromium and titanium dental alloys fabricated by casting, milling, and selective laser melting. J Prosthodont Res. 2018 Apr;62(2):184-94. https://doi.org/10.1016/j.jpor.2017.08.007
    » https://doi.org/10.1016/j.jpor.2017.08.007
  • 30 Abreu RT, Spazzin AO, Noritomi PY, Consani RL, Mesquita MF. Influence of material of overdenture-retaining bar with vertical misfit on three-dimensional stress distribution. J Prosthodont. 2010 Aug;19(6):425-31. https://doi.org/10.1111/j.1532-849X.2010.00612.x
    » https://doi.org/10.1111/j.1532-849X.2010.00612.x
  • 31 Karl M, Taylor TD. Effect of material selection on the passivity of fit of implant-supported restorations created with computer-aided design/computer-assisted manufacture. Int J Oral Maxillofac Implants. 2011;26(4):739-45.
  • 32 Al-Amleh B, Lyons K, Swain M. Clinical trials in zirconia: a systematic review. J Oral Rehabil. 2010 Aug;37(8):641-52. https://doi.org/10.1111/j.1365-2842.2010.02094.x
    » https://doi.org/10.1111/j.1365-2842.2010.02094.x
  • 33 Abduo J, Lyons K, Waddell N, Bennani V, Swain M. A comparison of fit of CNC-milled titanium and zirconia frameworks to implants. Clin Implant Dent Relat Res. 2012 May;14(s1 Suppl 1):e20-9. https://doi.org/10.1111/j.1708-8208.2010.00334.x
    » https://doi.org/10.1111/j.1708-8208.2010.00334.x
  • 34 Barbin T, Borges GA, Jardini AL, Mesquita MF. Hot isostatic pressing as an alternative thermo-mechanical treatment for metallic full-arch implant-supported frameworks obtained by additive and subtractive manufacturing technology: vertical and horizontal fit, screw removal torque, and stress analysis. J Prosthodont. 2024 Sep;33 S1:70-80. https://doi.org/10.1111/jopr.13842
    » https://doi.org/10.1111/jopr.13842

Data availability:

The authors declare that all data generated or analyzed during this study are included in this published article.

Financial support:

Fundação de Amparo à Pesquisa do Estado de São Paulo (Fapesp; grants #2017/15297-0 and #2017/16303-3), Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (Capes; grant #001), and Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq; grant #308474/2023-6). This study was also financed by Universidade Estadual de Campinas (Unicamp) through a postdoctoral fellowship granted to G.A.B. (01/2023).

Corresponding Author:

Guilherme Almeida Borges E-mail: borgesga@usp.br

Editor-in-Chief:

Lucianne Maia

Associate Editor:

Gustavo Nascimento

Declaration of interest:

The authors certify that they have no commercial or associative interest that represents a conflict of interest in connection with the manuscript.

Publication Dates

  • Publication in this collection
    16 Nov 2026
  • Date of issue
    2026

History

  • Received
    22 June 2026
  • Accepted
    09 Sept 2026
  • Reviewed
    06 Oct 2026
location_on
Sociedade Brasileira de Pesquisa Odontológica - SBPqO Av. Prof. Lineu Prestes, 2227, 05508-000 São Paulo SP - Brazil, Tel. (55 11) 3044-2393/(55 11) 9-7557-1244 - São Paulo - SP - Brazil
E-mail: office.bor@ingroup.srv.br
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Ir para o topo