Does cad software affect the marginal and internal fit of milled full ceramic crowns?

Abstract:

Although the effects of different intraoral scanners, model scanners, and CAM units on the marginal and internal fitting of restorations have been investigated, the effects of CAD software in particular has not been evaluated. The marginal and internal fit of indirect restorations may vary according to the CAD software used, even when using the same intraoral scanner and milling machine. The purpose of this study was to evaluate the marginal and internal fit of milled full ceramic crowns designed with three different CAD systems. Eleven typodont maxillary first premolar teeth were prepared and scanned using a 3Shape TRIOS Intraoral Dental Scanner. The obtained STL scan data were exported and used to design a full crown using three different CAD systems (CEREC, KaVo, and Planmeca). An independent milling unit was used to manufacture the crowns for each group (n = 11). The marginal and internal fit were evaluated for each restoration using 2D and 3D micro-CT analysis. For 2D analysis, 18 measurements for each sample were made, covering the marginal (Marginal Gap Buccal (MG-A), Marginal Gap Palatinal (MG-B), Finish Line Buccal (FL-A), Finish Line Palatinal (FA-B)) and internal fit locations (Axial Wall Buccal (AW-A), Axial Wall Palatinal (AW-B), Lingual Cusp (LC), Buccal Cusp (BC), and Occlusal Central Fossa (OCF)). Statistical analyses were performed using Open Source R Statistical Software (α = 0.05) The results of Duncan’s multiple range test showed that the values for the marginal measurement points MG-A, MG-B, FL-A, and FL-B in the Planmeca group were significantly higher than the values obtained in the CEREC and KaVo groups (p < 0.05). In AW1, values of the CEREC group were found to be higher than those of the KaVo and Planmeca groups (p < 0.05). CAD software showed an effect on the marginal fit values of crowns whereas no significant difference was observed in terms of the internal fit, except for a single measurement point made from the buccal direction.

Keywords:
Computer-Aided Design; Software; Crowns; X-Ray Microtomography

Introduction

The clinical success of dental restorations is influenced by three main factors: esthetics, fracture strength, and adaptation of the restoration to tooth preparation.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
Marginal and internal fit is of great importance for long-term clinical success particularly in ceramic restorations.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,33 Özçelik TB, Yilmaz B, Şeker E, Shah K. Karnik Shah. Marginal adaptation of provisional CAD/CAM restorations fabricated using various simulated digital cement space settings. Int J Oral Maxillofac Implants. 2018 Sep/Oct;33(5):1064-9. https://doi.org/10.11607/jomi.6271
https://doi.org/10.11607/jomi.6271...
In particular, poor and inadequate marginal adaptation between the crown and the restoration may result in plaque accumulation, secondary caries, microleakage, and endodontic lesions. Plaque accumulation may cause periodontal inflammation particularly in restorative margins ending subgingivally.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
,22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,44 Anadioti E, Aquilino SA, Gratton DG, Holloway JA, Denry I, Thomas GW, et al. 3D and 2D marginal fit of pressed and CAD/CAM lithium disilicate crowns made from digital and conventional impressions. J Prosthodont. 2014 Dec;23(8):610-7. https://doi.org/10.1111/jopr.12180
https://doi.org/10.1111/jopr.12180...
,55 Addi S, Hedayati-Khams A, Poya A, Sjögren G. Interface gap size of manually and CAD/CAM-manufactured ceramic inlays/onlays in vitro. J Dent. 2002 Jan;30(1):53-8. https://doi.org/10.1016/S0300-5712(01)00059-8
https://doi.org/10.1016/S0300-5712(01)00...

Although there is no definite information about the clinically acceptable marginal fit, values below 120 micrometers have been reported as a marginal fit that can be considered successful in several studies.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,66 Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
https://doi.org/10.1016/j.prosdent.2018....
In some studies, a marginal fit below 100 microns has been reported to be more acceptable77 Keshvad A, Hooshmand T, Asefzadeh F, Khalilinejad F, Alihemmati M, Van Noort R. Marginal gap, internal fit, and fracture load of leucite-reinforced ceramic inlays fabricated by CEREC inLab and hot-pressed techniques. J Prosthodont. 2011 Oct;20(7):535-40. https://doi.org/10.1111/j.1532-849X.2011.00745.x
https://doi.org/10.1111/j.1532-849X.2011...
,88 Davis DR. Comparison of fit of two types of all-ceramic crowns. J Prosthet Dent. 1988 Jan;59(1):12-6. https://doi.org/10.1016/0022-3913(88)90098-4
https://doi.org/10.1016/0022-3913(88)900...
while others suggest that a value of 75 micrometers or below is required for a clinically-successful restoration.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
The American Dental Association Specification No. 8 recommends a thickness of 25–40 micrometers for the cement pitch.66 Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
https://doi.org/10.1016/j.prosdent.2018....
Although a value of 25–40 micrometers is a considered a desirable margin in restorations, it is difficult to achieve this value.99 Holden JE, Goldstein GR, Hittelman EL, Clark EA. Comparison of the marginal fit of pressable ceramic to metal ceramic restorations. J Prosthodont. 2009 Dec;18(8):645-8. https://doi.org/10.1111/j.1532-849X.2009.00497.x
https://doi.org/10.1111/j.1532-849X.2009...
The marginal fit for restorations produced via Computer-Aided Design/Computed-Aided Manufacturing (CAD/CAM) is reported to be 58–200 micrometers.66 Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
https://doi.org/10.1016/j.prosdent.2018....
,1010 May KB, Russell MM, Razzoog ME, Lang BR. Precision of fit: the Procera AllCeram crown. J Prosthet Dent. 1998 Oct;80(4):394-404. https://doi.org/10.1016/S0022-3913(98)70002-2
https://doi.org/10.1016/S0022-3913(98)70...

The use of CAD/CAM systems in dentistry has allowed ceramic restorations to be produced in a shorter time and with a more acceptable adaptation accuracy in dental laboratories and dental clinics.1111 Boitelle P, Mawussi B, Tapie L, Fromentin O. A systematic review of CAD/CAM fit restoration evaluations. J Oral Rehabil. 2014 Nov;41(11):853-74. https://doi.org/10.1111/joor.12205
https://doi.org/10.1111/joor.12205...
,1212 Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J. 2009 Jan;28(1):44-56. https://doi.org/10.4012/dmj.28.44
https://doi.org/10.4012/dmj.28.44...
This technology was developed for use in dental clinics and has become an alternative to conventional techniques.1313 Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008 May;204(9):505-11. https://doi.org/10.1038/sj.bdj.2008.350
https://doi.org/10.1038/sj.bdj.2008.350...
Many factors can influence the success of restorations produced with a CAD/CAM system. These factors are dental preparation, scanning systems, CAD software, production stage, or type of material used.66 Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
https://doi.org/10.1016/j.prosdent.2018....
The marginal disharmony of restorations has been influenced by every step and change in the CAD/CAM system, from optical measurement to mechanical processing.1313 Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008 May;204(9):505-11. https://doi.org/10.1038/sj.bdj.2008.350
https://doi.org/10.1038/sj.bdj.2008.350...

Numerous methods have been used to evaluate the marginal fit in dentistry.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
,22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
Micro-computed tomography (micro-CT) is one of these methods. Although micro-CT is relatively more expensive than other methods, it is a non-destructive method for the evaluation of marginal fit.1414 Krasanaki ME, Pelekanos S, Andreiotelli M, Koutayas SO, Eliades G. X-ray microtomographic evaluation of the influence of two preparation types on marginal fit of CAD/CAM alumina copings: a pilot study. Int J Prosthodont. 2012 Mar-Apr;25(2):170-2.,1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
This three-dimensional (3D) high-resolution imaging system provides detailed cross-sectional information regarding the adaptation of crown restoration to dental preparation without damaging the sample.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
,1616 Lu L, Liu S, Shi S, Yang J. An open CAM system for dentistry on the basis of China-made 5-axis simultaneous contouring CNC machine tool and industrial CAM software. J Huazhong Univ Sci Technolog Med Sci. 2011 Oct;31(5):696-700. https://doi.org/10.1007/s11596-011-0585-y
https://doi.org/10.1007/s11596-011-0585-...
,1717 Seo D, Yi Y, Roh B. The effect of preparation designs on the marginal and internal gaps in Cerec3 partial ceramic crowns. J Dent. 2009 May;37(5):374-82. https://doi.org/10.1016/j.jdent.2009.01.008
https://doi.org/10.1016/j.jdent.2009.01....

There have been numerous studies examining the effects of different intraoral scanners, different milling devices and different versions of design programs on the marginal and internal fit of restorations. However, there are no studies to date examining the effect of different CAD software on the fit of restorations. Only different versions of the same CAD software have been analyzed. Many clinicians use only intraoral scanners in their clinics and send intraoral scanning data to external laboratories to design and produce the restorations. The fit of the restorations is thought to be affected not only by other parameters but also by CAD software. This study aims to investigate the effect of crown restorations designed using three different CAD software systems on marginal and internal fit. The null hypothesis of the study was that different CAD software would not affect the marginal and internal fit values.

Methodology

A total of 11 typodont enamel–dentin–pulp-based hard thermosetting plastic material maxillary first premolar teeth (ANA-4 ZSPD, Frasaco GmbH, Tettnang, Germany) were prepared by a single operator using standard diamond bur sets (Frasaco) by checking the relationship between the maxillary and mandibular teeth. A shoulder finish line with a rounded internal line angle was created for the crown by using diamond burs (839014; Hager & Meisinger GmbH, Neuss, Germany). First, guide grooves (Q92491; Hager & Meisinger GmbH) were opened and then axial and occlusal reductions were completed by following the principles of dental preparations (635314; Hager & Meisinger GmbH).

The relationships of the 11 prepared teeth with neighboring teeth and occlusion were scanned using a 3Shape TRIOS Intraoral Dental Scanner (3Shape, Copenhagen, Denmark) and the scan data were received from the system in the form of an STL file and were used in three different CAD systems (CEREC (inLab 15.1), KaVo (multicad. PC_V4.0.3), and Planmeca (Romexis PlanCad Easy5.9.2.09)). All designs were created by a single operator. The design proposed by the program was not altered except for minor corrections in the margin drawing to eliminate operator-related errors. In all three systems, the cement gap was set to 80 micrometers in margins and 120 micrometers in other regions. There was no intervention in any other parameters. A total of 33 crowns were milled (11 using each CAD software) using feldspathic ceramic blocks (62790; VITA Zahnfabric, Bad Säckingen, Germany) in an independent five-axis milling unit (DMC5020, DentMaster, Istanbul, Turkey).

The fabricated crowns were placed on each prepared tooth, and attached with radiolucent paraffin tapes to prevent any movement during micro-CT scanning. The micro-CT scans (Skyscan 1275; Bruker MicroCT, Kontich, Belgium) were performed using an aluminum (Al) filter (1 mm) with a rotation step of 0.2 in the cross-sectional range of 125 kVp, 80 mA and 24 μm. Raw micro-CT data were reconstructed using NRecon software (version 1.6.4.8 Bruker Corp.) and axial projections were obtained. These data were then transferred to CTan software (version 1,14,4,1 Bruker Corp.) for 3D analysis.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
,1616 Lu L, Liu S, Shi S, Yang J. An open CAM system for dentistry on the basis of China-made 5-axis simultaneous contouring CNC machine tool and industrial CAM software. J Huazhong Univ Sci Technolog Med Sci. 2011 Oct;31(5):696-700. https://doi.org/10.1007/s11596-011-0585-y
https://doi.org/10.1007/s11596-011-0585-...
,1818 Uzgur R, Ercan E, Uzgur Z, Çolak H, Yalçın M, Özcan M. Cement thickness of inlay restorations made of lithium disilicate, polymer-infiltrated ceramic and nano-ceramic CAD/CAM materials evaluated using 3D X-Ray micro-computed tomography. J Prosthodont. 2018 Jun;27(5):456-60. https://doi.org/10.1111/jopr.12521
https://doi.org/10.1111/jopr.12521...
2020 Küçükkaya Eren S, Aksel H, Askerbeyli Örs S, Serper A, Koçak Y, Ocak M, et al. Obturation quality of calcium silicate-based cements placed with different techniques in teeth with perforating internal root resorption: a micro-computed tomographic study. Clin Oral Investig. 2019 Feb;23(2):805-11. https://doi.org/10.1007/s00784-018-2502-2
https://doi.org/10.1007/s00784-018-2502-...

For the 3D analyses, crowns and substructure were included in the Region of Interest (ROI) and threshold values were determined. To calculate the volume of the gap (mm3), the original grayscale images were processed. Grayscale thresholds were defined to separate root material from crown and gap. The global threshold method was used for this. The amount of gap between the substructure and the crown was calculated. Then, the measurements were subject to statistical comparison. Colored and 3D images were obtained from the samples using CTvox and CTvol software1818 Uzgur R, Ercan E, Uzgur Z, Çolak H, Yalçın M, Özcan M. Cement thickness of inlay restorations made of lithium disilicate, polymer-infiltrated ceramic and nano-ceramic CAD/CAM materials evaluated using 3D X-Ray micro-computed tomography. J Prosthodont. 2018 Jun;27(5):456-60. https://doi.org/10.1111/jopr.12521
https://doi.org/10.1111/jopr.12521...
,2121 Demir N, Ozturk AN, Malkoc MA. Evaluation of the marginal fit of full ceramic crowns by the microcomputed tomography (micro-CT) technique. Eur J Dent. 2014 Oct;8(4):437-44. https://doi.org/10.4103/1305-7456.143612
https://doi.org/10.4103/1305-7456.143612...
,2222 Baltacioglu IH, Kamburoglu K, Irmak O, Geneci F, Ocak M, Uzuner MB, et al. Marginal integrity of self-adhering flowable composites used as liner under class II restorations: a comparative in vitro micro-CT study. J Adhes Sci Technol. 2017;31(24):2719-29. https://doi.org/10.1080/01694243.2017.1317472.
https://doi.org/10.1080/01694243.2017.13...
(Figure 1).

Figure 1
Colored and 3D images, obtained from the samples using CTvox and CTvol software.

DataViewer software (version 1,5,6.2 Bruker Corp.) was used to prepare for 2D analyses. The midcoronal and midsagittal images of axial sections were obtained using this software (Figure 2a, 2b). Then, 2D linear measurements were re-performed with CTan software in these sections. For the 2D analyses, the measurement points previously described in the literature were used (Riccitiello et al. 2018) (Figure 3). A total of 594 measurements, 18 measurements for each sample, were made. Measurement locations MG-A, MG-B, and FL-A, FL-B were evaluated for the marginal gap whereas measurement locations AW-A, FL-B, LC, BC, and OCF were evaluated for internal fit.

Figure 2
a) Midcoronal, and b) midsagittal cross-sectional images.
Figure 3
Reference points for micro-CT measurements of midcoronal and midsagittal sections.

All statistical analyses were performed using Open Source R Statistical Software. The Shapiro-Wilk test was used to evaluate the parametric or nonparametric distribution of the groups. One-way ANOVA was used to determine any significant differences between groups, and Duncan’s multiple range test was used to determine intra-group differences. A p-value of < 0.05 was considered statistically significant in all tests.

Results

Descriptive statistical analysis of the mean and standard deviation values of linear and volumetric measurements for all groups is shown in Tables 1 and 2. Repeated ANOVA measurements showed significant differences for MG-A, FL-A, AW-A, FL-B and MG-B values (p < 0.05).

Table 1
Descriptive statistical analysis of 2D analysis
Table 2
Descriptive statistical analysis of 3D analysis.

The results of Duncan’s multiple range test showed that the values for the measurement point MG-A in the Planmeca group (114.9 ± 72.1) were significantly higher than the values in the CEREC (34.0 ± 82.0) and KaVo (38.0 ± 85.0) groups.

The highest values for the measurement point FL-A were seen in the Planmeca group (224.8 ± 76.7) and this value was significantly different from the other two groups (p < 0.05). The values of the CEREC group (118.0 ± 69.9) were higher than those of the KaVo group (69.7 ± 66.1), but the difference was not statistically significant (p > 0.05).

In AW-A, which was one of the measurement points used to evaluate internal fit, values of the CEREC group (132.5 ± 43.4) were found to be higher than those of the KaVo group (82.3 ± 38.1) and the Planmeca Group (95.6 ± 40.4) (p < 0.05). However, there was no significant difference between the KaVo and Planmeca groups (p > 0.05).

The values of FL-B measured using CEREC (101.2 ± 80.4) were higher than those using KaVo (87.6 ± 69.7), but this difference was not significant (p > 0.05). The values measured using Planmeca (218.4 ± 79.9) were found to be significantly higher than those measured with CEREC or KaVo (p < 0.05).

In the measurements of the MG-B measurement point, the values in the Planmeca group (151.9 ± 93.1) were significantly higher than the other two (p < 0.05). The values in the CEREC (25.6 ± 47.3) and KaVo (31.3 ± 65.5) groups were close to each other and there was no statistically-significant difference between the groups (p > 0.05).

In volumetric measurements made in 3D, the values in the CEREC group were higher than the values in the KaVo group which in turn were higher than the values in the Planmeca group. However, no statistically-significant difference was found between the three groups (p > 0.05).

Discussion

This study investigated the effect of crown restorations designed with three different CAD software systems with a focus on marginal and internal fit. Based on the results of this study, different CAD software affected the marginal fit, however the internal fit was not affected except for one measurement point. While there was no significant difference according to the results of 3D analysis, there was a difference in the 2D analysis. Thus, the null hypothesis of the study has been partially rejected.

The data obtained from the results of this study showed a significant difference between the restorations designed and produced using three different CAD software systems in terms of marginal fit. However, the internal fit between ceramic restoration and tooth preparation was not affected by the use of different CAD software, except for one measurement point. The marginal fit values of crowns designed using Planmeca CAD software were found to be significantly higher than the other two CAD software. There was no significant difference between CEREC and KaVo software. The internal fit of a ceramic crown affects its fracture strength.2323 Tuntiprawon M, Wilson PR. The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J. 1995 Feb;40(1):17-21. https://doi.org/10.1111/j.1834-7819.1995.tb05607.x
https://doi.org/10.1111/j.1834-7819.1995...
,2424 May L, Kelly JR, Bottino MA, Hill T. Kelly Jr, Bottıno Ma, Hıll T. Effects of cement thickness and bonding on the failure loads of CAD/CAM ceramic crowns: multi-physics FEA modeling and monotonic testing. Dent Mater. 2012 Aug;28(8):e99-109. https://doi.org/10.1016/j.dental.2012.04.033
https://doi.org/10.1016/j.dental.2012.04...
Since large internal misfits would result in a thick layer of a low elastic modulus material (cement), misfit may be structurally important for brittle materials, such as ceramic restorations.2424 May L, Kelly JR, Bottino MA, Hill T. Kelly Jr, Bottıno Ma, Hıll T. Effects of cement thickness and bonding on the failure loads of CAD/CAM ceramic crowns: multi-physics FEA modeling and monotonic testing. Dent Mater. 2012 Aug;28(8):e99-109. https://doi.org/10.1016/j.dental.2012.04.033
https://doi.org/10.1016/j.dental.2012.04...
In a study conducted by Lee et al.,2525 Lee KB, Park CW, Kim KH, Kwon TY. Marginal and internal fit of all-ceramic crowns fabricated with two different CAD/CAM systems. Dent Mater J. 2008 May;27(3):422-6. https://doi.org/10.4012/dmj.27.422
https://doi.org/10.4012/dmj.27.422...
the effect of two different CAD/CAM systems on the internal gap was examined and the internal fit was reported to be affected by the different systems used. In the present study, there was no significant difference between the crown restorations produced using the three different CAD software systems in terms of internal fit, except for a single measurement point. For the measurement point AW-A made from the buccal surface of the tooth, the CEREC group showed significantly higher values than the KaVo group whereas no significant difference was observed between the Planmeca group and the other two groups.

Previous studies have shown that poor marginal fit results in microleakage, periodontal problems, and secondary caries.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
,22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,44 Anadioti E, Aquilino SA, Gratton DG, Holloway JA, Denry I, Thomas GW, et al. 3D and 2D marginal fit of pressed and CAD/CAM lithium disilicate crowns made from digital and conventional impressions. J Prosthodont. 2014 Dec;23(8):610-7. https://doi.org/10.1111/jopr.12180
https://doi.org/10.1111/jopr.12180...
,55 Addi S, Hedayati-Khams A, Poya A, Sjögren G. Interface gap size of manually and CAD/CAM-manufactured ceramic inlays/onlays in vitro. J Dent. 2002 Jan;30(1):53-8. https://doi.org/10.1016/S0300-5712(01)00059-8
https://doi.org/10.1016/S0300-5712(01)00...
Although there is no definite information about the clinically-recommended marginal fit, a marginal fit up to 200 micrometers for CAD/CAM restorations is usually considered clinically acceptable.66 Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
https://doi.org/10.1016/j.prosdent.2018....
,1010 May KB, Russell MM, Razzoog ME, Lang BR. Precision of fit: the Procera AllCeram crown. J Prosthet Dent. 1998 Oct;80(4):394-404. https://doi.org/10.1016/S0022-3913(98)70002-2
https://doi.org/10.1016/S0022-3913(98)70...
In the present study, the marginal fit values for crown restorations in the Planmeca group were found to be close to or above 200 micrometers on average, whereas they were less than 200 micrometers in the CEREC and KaVo groups.

Many studies reported in the literature investigated the effects of different intraoral scanners, model scanners and CAM units, CAD/CAM systems, and different versions of design programs on the internal and marginal fit of restorations.1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
,1818 Uzgur R, Ercan E, Uzgur Z, Çolak H, Yalçın M, Özcan M. Cement thickness of inlay restorations made of lithium disilicate, polymer-infiltrated ceramic and nano-ceramic CAD/CAM materials evaluated using 3D X-Ray micro-computed tomography. J Prosthodont. 2018 Jun;27(5):456-60. https://doi.org/10.1111/jopr.12521
https://doi.org/10.1111/jopr.12521...
,2626 Haddadi Y, Bahrami G, Isidor F. Effect of software version on the accuracy of an intraoral scanning device. Int J Prosthodont. 2018 Jul/Aug;31(4):375-6. https://doi.org/10.11607/ijp.5781 PMID:29624626
https://doi.org/10.11607/ijp.5781...
2929 Renne W, Ludlow M, Fryml J, Schurch Z, Mennito A, Kessler R, et al. Evaluation of the accuracy of 7 digital scanners: an in vitro analysis based on 3-dimensional comparisons. J Prosthet Dent. 2017 Jul;118(1):36-42. https://doi.org/10.1016/j.prosdent.2016.09.024
https://doi.org/10.1016/j.prosdent.2016....
Haddai et al.2626 Haddadi Y, Bahrami G, Isidor F. Effect of software version on the accuracy of an intraoral scanning device. Int J Prosthodont. 2018 Jul/Aug;31(4):375-6. https://doi.org/10.11607/ijp.5781 PMID:29624626
https://doi.org/10.11607/ijp.5781...
evaluated the effect of different versions of a design program on the fit of a restoration and different versions of the software were shown to affect the adaptation of the restoration. Similarly, different results were obtained from different CAD software in the current study. However, our literature review revealed that there is no other study investigating the effect of CAD software alone, using the same independent scanner and CAM unit, as in the present study.

Different methods are described and recommended in the literature to measure the marginal and internal adaptations of indirect restorations, such as stereomicroscopy, scanning electron microscopy, optical microscopy, and micro-CT.1414 Krasanaki ME, Pelekanos S, Andreiotelli M, Koutayas SO, Eliades G. X-ray microtomographic evaluation of the influence of two preparation types on marginal fit of CAD/CAM alumina copings: a pilot study. Int J Prosthodont. 2012 Mar-Apr;25(2):170-2.,2727 Abdel-Azim T, Rogers K, Elathamna E, Zandinejad A, Metz M, Morton D. Comparison of the marginal fit of lithium disilicate crowns fabricated with CAD/CAM technology by using conventional impressions and two intraoral digital scanners. J Prosthet Dent. 2015 Oct;114(4):554-9. https://doi.org/10.1016/j.prosdent.2015.04.001
https://doi.org/10.1016/j.prosdent.2015....
,3030 Huang Z, Zhang L, Zhu J, Zhang X. Clinical marginal and internal fit of metal ceramic crowns fabricated with a selective laser melting technology. J Prosthet Dent. 2015 Jun;113(6):623-7. https://doi.org/10.1016/j.prosdent.2014.10.012
https://doi.org/10.1016/j.prosdent.2014....
3434 Shembesh M, Ali A, Finkelman M, Weber HP, Zandparsa R. An In Vitro Comparison of the Marginal Adaptation Accuracy of CAD/CAM Restorations Using Different Impression Systems. J Prosthodont. 2017 Oct;26(7):581-6. https://doi.org/10.1111/jopr.12446
https://doi.org/10.1111/jopr.12446...
In line with the previously-published studies,1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
,1616 Lu L, Liu S, Shi S, Yang J. An open CAM system for dentistry on the basis of China-made 5-axis simultaneous contouring CNC machine tool and industrial CAM software. J Huazhong Univ Sci Technolog Med Sci. 2011 Oct;31(5):696-700. https://doi.org/10.1007/s11596-011-0585-y
https://doi.org/10.1007/s11596-011-0585-...
,2727 Abdel-Azim T, Rogers K, Elathamna E, Zandinejad A, Metz M, Morton D. Comparison of the marginal fit of lithium disilicate crowns fabricated with CAD/CAM technology by using conventional impressions and two intraoral digital scanners. J Prosthet Dent. 2015 Oct;114(4):554-9. https://doi.org/10.1016/j.prosdent.2015.04.001
https://doi.org/10.1016/j.prosdent.2015....
micro-CT was used to evaluate the marginal and internal fit of crown restorations in the present study. At present, micro-CT has gained an important place as one of the best and most recommended methods of evaluating marginal and internal fit, since it ensures high-resolution images, repeated measurements and that measurements can be performed without damaging the specimens.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....
,22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,1717 Seo D, Yi Y, Roh B. The effect of preparation designs on the marginal and internal gaps in Cerec3 partial ceramic crowns. J Dent. 2009 May;37(5):374-82. https://doi.org/10.1016/j.jdent.2009.01.008
https://doi.org/10.1016/j.jdent.2009.01....

Micro-CT provides 3D volumetric analysis as well as 2D linear analysis.1717 Seo D, Yi Y, Roh B. The effect of preparation designs on the marginal and internal gaps in Cerec3 partial ceramic crowns. J Dent. 2009 May;37(5):374-82. https://doi.org/10.1016/j.jdent.2009.01.008
https://doi.org/10.1016/j.jdent.2009.01....
Yildirim et al,3535 Yildirim G, Uzun IH, Keles A. Evaluation of marginal and internal adaptation of hybrid and nanoceramic systems with microcomputed tomography: an in vitro study. J Prosthet Dent. 2017 Aug;118(2):200-7. https://doi.org/10.1016/j.prosdent.2016.11.005 PMID:28089331
https://doi.org/10.1016/j.prosdent.2016....
reported that cement space volume is a 3D parameter that represents more accurate values than 2D measurements. In the present study, the marginal and internal fit measured in 3D showed no significant difference between the three groups in which different CAD software were used. Although there was no difference between the groups according to the results of volumetric analysis, a statistically-significant difference was observed between the groups in terms of the marginal fit values. Thus, the 3D analysis provided useful information regarding the gap values between the restoration and the teeth. The average gap per area obtained through such methods better represents the entire restoration rather than a random portion of it. However, it was insufficient to assess the adaptation of the restoration in that it could not reflect the differences in marginal fit in particular.

Kim et al.3232 Kim KB, Kim WC, Kim HY, Kim JH. An evaluation of marginal fit of three-unit fixed dental prostheses fabricated by direct metal laser sintering system. Dent Mater. 2013 Jul;29(7):e91-6. https://doi.org/10.1016/j.dental.2013.04.007
https://doi.org/10.1016/j.dental.2013.04...
used different scanners to determine the cement space volume of IPS molar crowns and reported values ranging between 25.3 and 40.7 mm3. In this study, the measured 3D value of the premolar teeth was approximately 8 mm3. This could also be related to the tooth type used. Typodont is frequently used in investigations evaluating the marginal and internal fit of restorations.1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
,2525 Lee KB, Park CW, Kim KH, Kwon TY. Marginal and internal fit of all-ceramic crowns fabricated with two different CAD/CAM systems. Dent Mater J. 2008 May;27(3):422-6. https://doi.org/10.4012/dmj.27.422
https://doi.org/10.4012/dmj.27.422...
,2727 Abdel-Azim T, Rogers K, Elathamna E, Zandinejad A, Metz M, Morton D. Comparison of the marginal fit of lithium disilicate crowns fabricated with CAD/CAM technology by using conventional impressions and two intraoral digital scanners. J Prosthet Dent. 2015 Oct;114(4):554-9. https://doi.org/10.1016/j.prosdent.2015.04.001
https://doi.org/10.1016/j.prosdent.2015....
Ayad et al.3636 Ayad MF, Maghrabi AA, Rosenstiel SF. Assessment of convergence angles of tooth preparations for complete crowns among dental students. J Dent. 2005 Sep;33(8):633-8. https://doi.org/10.1016/j.jdent.2004.12.008
https://doi.org/10.1016/j.jdent.2004.12....
investigated the difference in crown preparation between natural teeth and typodont, and reported that the results were not affected by the type of tooth prepared. Therefore, in this study, typodont was used for crown preparation to evaluate the marginal and internal fit.

In the present study, restorations were fixed with radiolucent paraffin tapes to ensure that they remained stable in the micro-CT device where the sections were made. Since the produced crown restorations were going to be placed in the same tooth preparation to evaluate the three different groups, it was thought that cement residues could not be cleaned away completely and would cause measurement error.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
Paraffin tapes were able to hold the restoration fixed on the dental preparation and yet could be easily removed from the tooth leaving no residue. Thus, the prepared teeth were not damaged and it was more convenient to conduct measurements without cementing the crown.11 Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
https://doi.org/10.1016/j.prosdent.2013....

In the literature, aluminum filters have been used to prevent radiological artifacts in scans of ceramic objects. In our study, we scanned with an aluminum filter and adjusted all other parameters accordingly.22 Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
https://doi.org/10.1016/j.prosdent.2014....
,1515 Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
https://doi.org/10.1111/jerd.12271...
,1818 Uzgur R, Ercan E, Uzgur Z, Çolak H, Yalçın M, Özcan M. Cement thickness of inlay restorations made of lithium disilicate, polymer-infiltrated ceramic and nano-ceramic CAD/CAM materials evaluated using 3D X-Ray micro-computed tomography. J Prosthodont. 2018 Jun;27(5):456-60. https://doi.org/10.1111/jopr.12521
https://doi.org/10.1111/jopr.12521...
Shim et al.3737 Shim JS, Lee JS, Lee JY, Choi YJ, Shin SW, Ryu JJ. Effect of software version and parameter settings on the marginal and internal adaptation of crowns fabricated with the CAD/CAM system. J Appl Oral Sci. 2015 Oct;23(5):515-22. https://doi.org/10.1590/1678-775720150081
https://doi.org/10.1590/1678-77572015008...
in their study, concluded that parameter settings affect the fit of CAD/CAM restorations. In this study, all parameters were kept constant in the three groups. Thus, the marginal and internal adaptation was not affected by the other parameters.

In one previous study, the learning curves for the finish line of two different CAD software systems were examined with the findings that the learning curves could affect the results.3838 Son K, Lee WS, Lee KB. Prediction of the learning curves of 2 dental CAD software programs. J Prosthet Dent. 2019 Jan;121(1):95-100. https://doi.org/10.1016/j.prosdent.2018.01.004
https://doi.org/10.1016/j.prosdent.2018....
In the present study, minor corrections were made to minimize the effect of different learning curves where the system failed in the automatic margin drawing. No intervention was made other than these corrections. The findings should be verified on other preparations and construction designs.

Conclusion

Based on this study, the following could be concluded:

  1. CAD software showed an effect on the marginal fit values of crowns.

  2. The lowest marginal fit values were obtained with the CEREC and KaVo CAD software.

  3. No significant difference was observed between the different CAD software in terms of internal fit, except for the measurement point made from the buccal direction.

References

  • 1
    Contrepois M, Soenen A, Bartala M, Laviole O. Marginal adaptation of ceramic crowns: a systematic review. J Prosthet Dent. 2013 Dec;110(6):447-454.e10. https://doi.org/10.1016/j.prosdent.2013.08.003
    » https://doi.org/10.1016/j.prosdent.2013.08.003
  • 2
    Neves F, Prado C, Prudente M, Carneıro TA, Zancopé K, Davı LR, et al. Microcomputed tomography evaluation of marginal fit of lithium disilicate crowns fabricated by using chairside CAD/CAM systems or the heat-pressing technique. J Prosthet Dent. 2014 Nov;112(5):1134-40. https://doi.org/10.1016/j.prosdent.2014.04.028
    » https://doi.org/10.1016/j.prosdent.2014.04.028
  • 3
    Özçelik TB, Yilmaz B, Şeker E, Shah K. Karnik Shah. Marginal adaptation of provisional CAD/CAM restorations fabricated using various simulated digital cement space settings. Int J Oral Maxillofac Implants. 2018 Sep/Oct;33(5):1064-9. https://doi.org/10.11607/jomi.6271
    » https://doi.org/10.11607/jomi.6271
  • 4
    Anadioti E, Aquilino SA, Gratton DG, Holloway JA, Denry I, Thomas GW, et al. 3D and 2D marginal fit of pressed and CAD/CAM lithium disilicate crowns made from digital and conventional impressions. J Prosthodont. 2014 Dec;23(8):610-7. https://doi.org/10.1111/jopr.12180
    » https://doi.org/10.1111/jopr.12180
  • 5
    Addi S, Hedayati-Khams A, Poya A, Sjögren G. Interface gap size of manually and CAD/CAM-manufactured ceramic inlays/onlays in vitro. J Dent. 2002 Jan;30(1):53-8. https://doi.org/10.1016/S0300-5712(01)00059-8
    » https://doi.org/10.1016/S0300-5712(01)00059-8
  • 6
    Sadid-Zadeh R, Katsavochristou A, Squires T, Simon M. Accuracy of marginal fit and axial wall contour for lithium disilicate crowns fabricated using three digital workflows. J Prosthet Dent. 2020 Jan;123(1):121-7. https://doi.org/10.1016/j.prosdent.2018.11.003
    » https://doi.org/10.1016/j.prosdent.2018.11.003
  • 7
    Keshvad A, Hooshmand T, Asefzadeh F, Khalilinejad F, Alihemmati M, Van Noort R. Marginal gap, internal fit, and fracture load of leucite-reinforced ceramic inlays fabricated by CEREC inLab and hot-pressed techniques. J Prosthodont. 2011 Oct;20(7):535-40. https://doi.org/10.1111/j.1532-849X.2011.00745.x
    » https://doi.org/10.1111/j.1532-849X.2011.00745.x
  • 8
    Davis DR. Comparison of fit of two types of all-ceramic crowns. J Prosthet Dent. 1988 Jan;59(1):12-6. https://doi.org/10.1016/0022-3913(88)90098-4
    » https://doi.org/10.1016/0022-3913(88)90098-4
  • 9
    Holden JE, Goldstein GR, Hittelman EL, Clark EA. Comparison of the marginal fit of pressable ceramic to metal ceramic restorations. J Prosthodont. 2009 Dec;18(8):645-8. https://doi.org/10.1111/j.1532-849X.2009.00497.x
    » https://doi.org/10.1111/j.1532-849X.2009.00497.x
  • 10
    May KB, Russell MM, Razzoog ME, Lang BR. Precision of fit: the Procera AllCeram crown. J Prosthet Dent. 1998 Oct;80(4):394-404. https://doi.org/10.1016/S0022-3913(98)70002-2
    » https://doi.org/10.1016/S0022-3913(98)70002-2
  • 11
    Boitelle P, Mawussi B, Tapie L, Fromentin O. A systematic review of CAD/CAM fit restoration evaluations. J Oral Rehabil. 2014 Nov;41(11):853-74. https://doi.org/10.1111/joor.12205
    » https://doi.org/10.1111/joor.12205
  • 12
    Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J. 2009 Jan;28(1):44-56. https://doi.org/10.4012/dmj.28.44
    » https://doi.org/10.4012/dmj.28.44
  • 13
    Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008 May;204(9):505-11. https://doi.org/10.1038/sj.bdj.2008.350
    » https://doi.org/10.1038/sj.bdj.2008.350
  • 14
    Krasanaki ME, Pelekanos S, Andreiotelli M, Koutayas SO, Eliades G. X-ray microtomographic evaluation of the influence of two preparation types on marginal fit of CAD/CAM alumina copings: a pilot study. Int J Prosthodont. 2012 Mar-Apr;25(2):170-2.
  • 15
    Alajaji NK, Bardwell D, Finkelman M, Ali A. Micro-CT evaluation of ceramic inlays: comparison of the marginal and internal fit of five and three axis CAM systems with heat press technique. J Esthet Restor Dent. 2017 Feb;29(1):49-58. https://doi.org/10.1111/jerd.12271
    » https://doi.org/10.1111/jerd.12271
  • 16
    Lu L, Liu S, Shi S, Yang J. An open CAM system for dentistry on the basis of China-made 5-axis simultaneous contouring CNC machine tool and industrial CAM software. J Huazhong Univ Sci Technolog Med Sci. 2011 Oct;31(5):696-700. https://doi.org/10.1007/s11596-011-0585-y
    » https://doi.org/10.1007/s11596-011-0585-y
  • 17
    Seo D, Yi Y, Roh B. The effect of preparation designs on the marginal and internal gaps in Cerec3 partial ceramic crowns. J Dent. 2009 May;37(5):374-82. https://doi.org/10.1016/j.jdent.2009.01.008
    » https://doi.org/10.1016/j.jdent.2009.01.008
  • 18
    Uzgur R, Ercan E, Uzgur Z, Çolak H, Yalçın M, Özcan M. Cement thickness of inlay restorations made of lithium disilicate, polymer-infiltrated ceramic and nano-ceramic CAD/CAM materials evaluated using 3D X-Ray micro-computed tomography. J Prosthodont. 2018 Jun;27(5):456-60. https://doi.org/10.1111/jopr.12521
    » https://doi.org/10.1111/jopr.12521
  • 19
    Aydın ZU, Keskin NB, Özyürek T, Geneci F, Ocak M, Çelik HH. Microcomputed assessment of transportation, centering ratio, canal area, and volume increase after single-file rotary and reciprocating glide path instrumentation in curved root canals: a laboratory study. J Endod. 2019 Jun;45(6):791-6. https://doi.org/10.1016/j.joen.2019.02.012
    » https://doi.org/10.1016/j.joen.2019.02.012
  • 20
    Küçükkaya Eren S, Aksel H, Askerbeyli Örs S, Serper A, Koçak Y, Ocak M, et al. Obturation quality of calcium silicate-based cements placed with different techniques in teeth with perforating internal root resorption: a micro-computed tomographic study. Clin Oral Investig. 2019 Feb;23(2):805-11. https://doi.org/10.1007/s00784-018-2502-2
    » https://doi.org/10.1007/s00784-018-2502-2
  • 21
    Demir N, Ozturk AN, Malkoc MA. Evaluation of the marginal fit of full ceramic crowns by the microcomputed tomography (micro-CT) technique. Eur J Dent. 2014 Oct;8(4):437-44. https://doi.org/10.4103/1305-7456.143612
    » https://doi.org/10.4103/1305-7456.143612
  • 22
    Baltacioglu IH, Kamburoglu K, Irmak O, Geneci F, Ocak M, Uzuner MB, et al. Marginal integrity of self-adhering flowable composites used as liner under class II restorations: a comparative in vitro micro-CT study. J Adhes Sci Technol. 2017;31(24):2719-29. https://doi.org/10.1080/01694243.2017.1317472
    » https://doi.org/10.1080/01694243.2017.1317472
  • 23
    Tuntiprawon M, Wilson PR. The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J. 1995 Feb;40(1):17-21. https://doi.org/10.1111/j.1834-7819.1995.tb05607.x
    » https://doi.org/10.1111/j.1834-7819.1995.tb05607.x
  • 24
    May L, Kelly JR, Bottino MA, Hill T. Kelly Jr, Bottıno Ma, Hıll T. Effects of cement thickness and bonding on the failure loads of CAD/CAM ceramic crowns: multi-physics FEA modeling and monotonic testing. Dent Mater. 2012 Aug;28(8):e99-109. https://doi.org/10.1016/j.dental.2012.04.033
    » https://doi.org/10.1016/j.dental.2012.04.033
  • 25
    Lee KB, Park CW, Kim KH, Kwon TY. Marginal and internal fit of all-ceramic crowns fabricated with two different CAD/CAM systems. Dent Mater J. 2008 May;27(3):422-6. https://doi.org/10.4012/dmj.27.422
    » https://doi.org/10.4012/dmj.27.422
  • 26
    Haddadi Y, Bahrami G, Isidor F. Effect of software version on the accuracy of an intraoral scanning device. Int J Prosthodont. 2018 Jul/Aug;31(4):375-6. https://doi.org/10.11607/ijp.5781 PMID:29624626
    » https://doi.org/10.11607/ijp.5781
  • 27
    Abdel-Azim T, Rogers K, Elathamna E, Zandinejad A, Metz M, Morton D. Comparison of the marginal fit of lithium disilicate crowns fabricated with CAD/CAM technology by using conventional impressions and two intraoral digital scanners. J Prosthet Dent. 2015 Oct;114(4):554-9. https://doi.org/10.1016/j.prosdent.2015.04.001
    » https://doi.org/10.1016/j.prosdent.2015.04.001
  • 28
    Vennerstrom M, Fakhary M, Von Steyern PV. The fit of crowns produced using digital impression systems. Swed Dent J. 2014;38(3):101-10.
  • 29
    Renne W, Ludlow M, Fryml J, Schurch Z, Mennito A, Kessler R, et al. Evaluation of the accuracy of 7 digital scanners: an in vitro analysis based on 3-dimensional comparisons. J Prosthet Dent. 2017 Jul;118(1):36-42. https://doi.org/10.1016/j.prosdent.2016.09.024
    » https://doi.org/10.1016/j.prosdent.2016.09.024
  • 30
    Huang Z, Zhang L, Zhu J, Zhang X. Clinical marginal and internal fit of metal ceramic crowns fabricated with a selective laser melting technology. J Prosthet Dent. 2015 Jun;113(6):623-7. https://doi.org/10.1016/j.prosdent.2014.10.012
    » https://doi.org/10.1016/j.prosdent.2014.10.012
  • 31
    Kane LM, Chronaios D, Sierraalta M, George FM. Marginal and internal adaptation of milled cobalt-chromium copings. J Prosthet Dent. 2015 Nov;114(5):680-5. https://doi.org/10.1016/j.prosdent.2015.04.020
    » https://doi.org/10.1016/j.prosdent.2015.04.020
  • 32
    Kim KB, Kim WC, Kim HY, Kim JH. An evaluation of marginal fit of three-unit fixed dental prostheses fabricated by direct metal laser sintering system. Dent Mater. 2013 Jul;29(7):e91-6. https://doi.org/10.1016/j.dental.2013.04.007
    » https://doi.org/10.1016/j.dental.2013.04.007
  • 33
    An S, Kim S, Choi H, Lee JH, Moon HS. Evaluating the marginal fit of zirconia copings with digital impressions with an intraoral digital scanner. J Prosthet Dent. 2014 Nov;112(5):1171-5. https://doi.org/10.1016/j.prosdent.2013.12.024
    » https://doi.org/10.1016/j.prosdent.2013.12.024
  • 34
    Shembesh M, Ali A, Finkelman M, Weber HP, Zandparsa R. An In Vitro Comparison of the Marginal Adaptation Accuracy of CAD/CAM Restorations Using Different Impression Systems. J Prosthodont. 2017 Oct;26(7):581-6. https://doi.org/10.1111/jopr.12446
    » https://doi.org/10.1111/jopr.12446
  • 35
    Yildirim G, Uzun IH, Keles A. Evaluation of marginal and internal adaptation of hybrid and nanoceramic systems with microcomputed tomography: an in vitro study. J Prosthet Dent. 2017 Aug;118(2):200-7. https://doi.org/10.1016/j.prosdent.2016.11.005 PMID:28089331
    » https://doi.org/10.1016/j.prosdent.2016.11.005
  • 36
    Ayad MF, Maghrabi AA, Rosenstiel SF. Assessment of convergence angles of tooth preparations for complete crowns among dental students. J Dent. 2005 Sep;33(8):633-8. https://doi.org/10.1016/j.jdent.2004.12.008
    » https://doi.org/10.1016/j.jdent.2004.12.008
  • 37
    Shim JS, Lee JS, Lee JY, Choi YJ, Shin SW, Ryu JJ. Effect of software version and parameter settings on the marginal and internal adaptation of crowns fabricated with the CAD/CAM system. J Appl Oral Sci. 2015 Oct;23(5):515-22. https://doi.org/10.1590/1678-775720150081
    » https://doi.org/10.1590/1678-775720150081
  • 38
    Son K, Lee WS, Lee KB. Prediction of the learning curves of 2 dental CAD software programs. J Prosthet Dent. 2019 Jan;121(1):95-100. https://doi.org/10.1016/j.prosdent.2018.01.004
    » https://doi.org/10.1016/j.prosdent.2018.01.004