Acessibilidade / Reportar erro

Clinical evaluation of a bone cement-injectable cannulated pedicle screw augmented with polymethylmethacrylate: 128 osteoporotic patients with 42 months of follow-up

Abstract

OBJECTIVES:

To evaluate the safety and efficacy of a novel bone cement-injectable cannulated pedicle screw augmented with polymethylmethacrylate in osteoporotic spinal surgery.

METHODS:

This study included 128 patients with osteoporosis (BMD T-score −3.2±1.9; range, −5.4 to -2.5) who underwent spinal decompression and instrumentation with a polymethylmethacrylate-augmented bone cement-injectable cannulated pedicle screw. Postoperative Visual Analogue Scale scores and the Oswestry Disability Index were compared with preoperative values. Postoperative plain radiographs and computed tomography (CT) scans were performed immediately after surgery; at 1, 3, 6, and 12 months; and annually thereafter.

RESULTS:

The mean follow-up time was 42.4±13.4 months (range, 23 to 71 months). A total of 418 polymethylmethacrylate-augmented bone cement-injectable cannulated pedicle screws were used. Cement extravasations were detected in 27 bone cement-injectable cannulated pedicle screws (6.46%), mainly in cases of vertebral fracture, without any clinical sequela. The postoperative low back and lower limb Visual Analogue Scale scores were significantly reduced compared with the preoperative scores (<0.01), and similar results were noted for the Oswestry Disability Index score (p<0.01). No significant screw migration was noted at the final follow-up relative to immediately after surgery (p<0.01). All cases achieved successful bone fusion, and no case required revision. No infection or blood clots occurred after surgery.

CONCLUSIONS:

The polymethylmethacrylate-augmented bone cement-injectable cannulated pedicle screw is safe and effective for use in osteoporotic patients who require spinal instrumentation.

Osteoporosis; Pedicle Screw Fixation; Bone Fusion; CICPS; Safety; Effectiveness


INTRODUCTION

In recent decades, internal fixation with a pedicle screw system has been the gold standard for the treatment of an unstable spine (11. Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29(11):E212-6, https://doi.org/10.1097/00007632-200406010-00020.
https://doi.org/10.1097/00007632-2004060...
) caused by degenerative diseases of the thoracolumbar spine, trauma, or tumors. However, an increasing number of patients worldwide who suffer from osteoporosis have poor bone quality that does not provide sufficient strength for common pedicle screws during internal fixation. Several studies have found that increasing the pullout strength of the pedicle screws can effectively solve this problem. Relevant techniques include increasing the diameter or length of the screw (22. Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims--what is best? Spine. 1998;23(12):1374-9, https://doi.org/10.1097/00007632-199806150-00015.
https://doi.org/10.1097/00007632-1998061...
); improving the design of the screw-rod (33. Abshire BB, McLain RF, Valdevit A, Kambic HE. Characteristics of pullout failure in conical and cylindrical pedicle screws after full insertion and back-out. Spine J. 2001;1(6):408-14, https://doi.org/10.1016/S1529-9430(01)00119-X.
https://doi.org/10.1016/S1529-9430(01)00...
) or the screw threads (44. Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd. Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J. 2001;1(2):109-14, https://doi.org/10.1016/S1529-9430(01)00020-1.
https://doi.org/10.1016/S1529-9430(01)00...
); choosing a proper insertion angle and trajectory (55. Patel PS, Shepherd DE, Hukins DW. The effect of screw insertion angle and thread type on the pullout strength of bone screws in normal and osteoporotic cancellous bone models. Med Eng Phys. 2010;32(8):822-8, https://doi.org/10.1016/j.medengphy.2010.05.005.
https://doi.org/10.1016/j.medengphy.2010...
); stabilizing the spine with bicortical fixation (44. Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd. Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J. 2001;1(2):109-14, https://doi.org/10.1016/S1529-9430(01)00020-1.
https://doi.org/10.1016/S1529-9430(01)00...
,66. Zhuang XM, Yu BS, Zheng ZM, Zhang JF, Lu WW. Effect of the degree of osteoporosis on the biomechanical anchoring strength of the sacral pedicle screws: an in vitro comparison between unaugmented bicortical screws and polymethylmethacrylate augmented unicortical screws. Spine. 2010;35(19):E925-31, https://doi.org/10.1097/BRS.0b013e3181c5fb21.
https://doi.org/10.1097/BRS.0b013e3181c5...
); and using expandable pedicle screws (44. Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd. Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J. 2001;1(2):109-14, https://doi.org/10.1016/S1529-9430(01)00020-1.
https://doi.org/10.1016/S1529-9430(01)00...
,77. Shi L, Wang L, Guo Z, Wu ZX, Liu D, Gao MX, et al. A study of low elastic modulus expandable pedicle screws in osteoporotic sheep. J Spinal Disord Tech. 2012;25(2):123-8, https://doi.org/10.1097/BSD.0b013e31820ef9f7.
https://doi.org/10.1097/BSD.0b013e31820e...
,88. Liu D, Shi L, Lei W, Wei MQ, Qu B, Deng SL, et al. Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation: An Experimental Study. Clin Spine Surg. 2016;29(7):E351-7, https://doi.org/10.1097/BSD.0b013e31828bfc85.
https://doi.org/10.1097/BSD.0b013e31828b...
) and bone cement-augmented pedicle screws (88. Liu D, Shi L, Lei W, Wei MQ, Qu B, Deng SL, et al. Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation: An Experimental Study. Clin Spine Surg. 2016;29(7):E351-7, https://doi.org/10.1097/BSD.0b013e31828bfc85.
https://doi.org/10.1097/BSD.0b013e31828b...

9. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8.
https://doi.org/10.1007/s00586-008-0769-...

10. Hickerson LE, Owen JR, Wayne JS, Tuten HR. Calcium Triglyceride Versus Polymethylmethacrylate Augmentation: A Biomechanical Analysis of Pullout Strength. Spine Deform. 2013;1(1):10-5, https://doi.org/10.1016/j.jspd.2012.07.004.
https://doi.org/10.1016/j.jspd.2012.07.0...
-1111. Stadelmann VA, Bretton E, Terrier A, Procter P, Pioletti DP. Calcium phosphate cement augmentation of cancellous bone screws can compensate for the absence of cortical fixation. J Biomech. 2010;43(15):2869-74, https://doi.org/10.1016/j.jbiomech.2010.07.025.
https://doi.org/10.1016/j.jbiomech.2010....
). However, these strategies have potential shortcomings, such as screw loosening or pullout, screw fracture, vascular or visceral injury, and complications associated with cement leakage (99. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8.
https://doi.org/10.1007/s00586-008-0769-...
). Therefore, new techniques that improve the safety and effectiveness of instrumentation are required for the surgical treatment of osteoporosis.

Conventional cannulated pedicle screws augmented with bone cement exhibit several disadvantages. These include unsatisfactory diffusion and distribution of the bone cement (99. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8.
https://doi.org/10.1007/s00586-008-0769-...
,1212. Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc. 2009;46(4):305-11, https://doi.org/10.3340/jkns.2009.46.4.305.
https://doi.org/10.3340/jkns.2009.46.4.3...
); a fixed screw head design that makes the operation difficult to perform (1212. Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc. 2009;46(4):305-11, https://doi.org/10.3340/jkns.2009.46.4.305.
https://doi.org/10.3340/jkns.2009.46.4.3...
); and prolonged operative time and increased blood loss due to the installation and disassembly of the bone cement injecting system. Therefore, we designed a bone cement-injectable cannulated pedicle screw (CICPS) that solved these problems using three radial holes, a flexible screw head, and an injection system.

We previously demonstrated the improved biomechanical stability of the polymethylmethacrylate (PMMA)-augmented CICPS in osteoporotic bone (1313. Liu Y, Xu J, Sun D, Luo F, Zhang Z, Dai F. Biomechanical and finite element analyses of bone cement-injectable cannulated pedicle screw fixation in osteoporotic bone. J Biomed Mater Res B Appl Biomater. 2016;104(5):960-7, https://doi.org/10.1002/jbm.b.33424.
https://doi.org/10.1002/jbm.b.33424...
,1414. Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, et al. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics. 2015;70(2):114-9, https://doi.org/10.6061/clinics/2015(02)08.
https://doi.org/10.6061/clinics/2015(02)...
). In the present retrospective study, we evaluated the long-term safety and effectiveness of the PMMA-augmented CICPS in 128 osteoporotic patients with degenerative spinal diseases.

MATERIALS AND METHODS

Patients

The institutional review board of Southwest Hospital, Chongqing, China, approved this retrospective study, and all subjects provided informed consent. The study included a population of 128 patients (29 men and 99 women) with osteoporosis and degenerative spinal diseases who underwent spinal decompression and fixation with PMMA-augmented CICPS at our hospital between March 2011 and March 2015. Diseases included degenerative lumbar spondylolisthesis, lumbar disc herniation with lumbar spinal stenosis, compression fractures, and ankylosing spondylitis (AS).

Osteoporosis in these patients was diagnosed according to the World Health Organization’s diagnostic criteria with a T-score≤-2.5 (1515. Kanis JA, Melton LJ 3rd, Christiansen C, Johnston CC, Khaltaev N. The diagnosis of osteoporosis. J Bone Miner Res. 1994;9(8):1137-41, https://doi.org/10.1002/jbmr.5650090802.
https://doi.org/10.1002/jbmr.5650090802...
) by examining the lumbar spine using a Hologic Discovery Delphi SL, QDR® Series (Hologic, Inc., Bedford, MA, USA) with dual-energy X-ray absorptiometry (DXA). Diagnoses of lumbar spondylolisthesis, lumbar disc herniation with lumbar spinal stenosis, and compression fractures were based on clinical symptoms, such as low back pain, radiating pain, numbness, and muscle weakness in the lower limbs and radiological findings in plain-film imaging, computed tomography (CT) (1111. Stadelmann VA, Bretton E, Terrier A, Procter P, Pioletti DP. Calcium phosphate cement augmentation of cancellous bone screws can compensate for the absence of cortical fixation. J Biomech. 2010;43(15):2869-74, https://doi.org/10.1016/j.jbiomech.2010.07.025.
https://doi.org/10.1016/j.jbiomech.2010....
), and magnetic resonance imaging (MRI). Kyphosis was diagnosed by the appearance of a deformed spine on physical examination and radiologically. Operative management has been advocated for adolescents with progressive kyphosis greater than 70°, patients with progressive kyphosis despite bracing, patients with intractable back pain, and patients with an unacceptable cosmetic deformity (1616. Lonner BS, Newton P, Betz R, Scharf C, O'Brien M, Sponseller P, et al. Operative management of Scheuermann's kyphosis in 78 patients: radiographic outcomes, complications, and technique. Spine. 2007;32(24):2644-52, https://doi.org/10.1097/BRS.0b013e31815a5238.
https://doi.org/10.1097/BRS.0b013e31815a...
). All patients had one or more diagnoses; however, we only presented the main diagnosis that provided surgical indications. All patients underwent surgery when their symptoms and signs did not improve after conservative treatment for at least six months. Patients who had blood coagulation disorders or who were allergic to the implants were excluded.

Implant Design

The CICPSs used in this study (Figure 1A) were 3.5-5.0 mm in diameter and 55-70 mm long with a 3-mm pitch (Kanghui Medical Devices, Jiangsu, China). The pedicle screw had a cannulation diameter of 2.2 mm with three radial holes at the distal end (round, 2 mm diameter; oval, 3 mm long and 2 mm wide; and U-shape, 4 mm long and 2 mm wide). The multiaxis or single-axis screw head was designed to facilitate the operative process.

Figure 1
(A) The design of the CICPS; (B) The CICPS connects to the specially designed bone cement syringe and the T-shaped handle through an adapter.

Surgical Procedures

The patients were positioned prone and were placed under general anesthesia. The surgical procedures included removal of the intervertebral disc, spinal canal decompression, spinal osteotomy and orthopedics, bone graft fusion and posterior internal fixation with PMMA-augmented CICPS, and laminectomy with posterolateral fusion.

The CICPS was implanted with a relatively larger insertion angle to leave more space for the cement. To avoid bone cement leakage to the spinal canal, a screw with a length that was 80-90% of the diameter of the vertebral body was selected. After the CICPS was inserted, PMMA (approximately 1.5 mL) was injected via a dedicated syringe and adapter (Figure 1B and C) and distributed into the surrounding trabeculae bone through the three side holes. An X-ray lateral view of the vertebrate was performed during the PMMA injection to observe the distribution of bone cement. The PMMA injection was stopped when the cement leaked to the posterior part of the screw.

Postoperative Management

All patients were placed on bedrest for at least 3 days after surgery. Drainage tubes were removed when the volume of drainage was less than 50 mL within 24 hours. Cefazolin sodium (2 g, intravenous drip, every 12 hours) was routinely used for 24 hours to prevent infection. Each patient wore a custom-made lumbosacral or vest orthosis for at least 3 months until fusion was achieved. Patients were encouraged to attempt ambulation 3 days after surgery and allowed to participate in positive activity or go to work depending on his/her recovery and neurological situation.

Follow-up

All patients underwent follow-up at 1, 3, 6, and 12 months after surgery and annually thereafter. During each follow-up, all patients underwent neurological evaluation and radiological examination. Visual analog scale (VAS) scores were used to evaluate pain severity, and the Oswestry disability index (ODI) score was used to evaluate disability. Any complications, such as cement leakage, infection, and blood clots, were recorded.

All patients underwent pre- and postoperative X-rays and 3D-CT and during each outpatient follow-up. Pedicle screw loosening or pullout was determined using anteroposterior, lateral, and standing flexion-extension lumbosacral plain X-rays according to the distance between the screw tip and the anterior surface of the vertebral body (distance x) and the distance from the screw tip to the superior endplate of the vertebral body (distance y).

In patients with lumbar spondylolisthesis, the angular displacement and the Taillard index were measured to assess the degree of slipping of the vertebral body (1717. Sun Y, Wang H, Yang D, Zhang N, Yang S, Zhang W, et al. Characterization of radiographic features of consecutive lumbar spondylolisthesis. Medicine. 2016;95(46):e5323, https://doi.org/10.1097/MD.0000000000005323.
https://doi.org/10.1097/MD.0000000000005...
).

CT scans were performed to determine interbody fusion and cement leakage 6 and 12 months after surgery. Successful fusion was assessed in accordance with Sapkas’ and Christiansen’s methods (1818. Sapkas GS, Papadakis SA, Stathakopoulos DP, Papagelopoulos PJ, Badekas AC, Kaiser JH. Evaluation of pedicle screw position in thoracic and lumbar spine fixation using plain radiographs and computed tomography. A prospective study of 35 patients. Spine. 1999;24(18):1926-9, https://doi.org/10.1097/00007632-199909150-00011.
https://doi.org/10.1097/00007632-1999091...
,1919. Burkus JK, Foley K, Haid RW, LeHuec JC. Surgical Interbody Research Group--radiographic assessment of interbody fusion devices: fusion criteria for anterior lumbar interbody surgery. Neurosurg Focus. 2001;10(4):E11, https://doi.org/10.3171/foc.2001.10.4.12.
https://doi.org/10.3171/foc.2001.10.4.12...
). Screw loosening was considered radiolucent at one millimeter or wider at the bone/screw interface (1212. Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc. 2009;46(4):305-11, https://doi.org/10.3340/jkns.2009.46.4.305.
https://doi.org/10.3340/jkns.2009.46.4.3...
). Here, 3D-CT can reflect the shape of the leaky bone cement. For example, when the bone cement seeps into the intervertebral space, it diffuses into the surrounding area and spreads out irregularly. If it flows into a blood vessel, it would exhibit an approximately regular strip shape, which is similar the appearance of the vessel.

Statistical Analysis

Statistical analyses were performed using SPSS 18.0 for Windows (SPSS, Chicago, IL). The data are presented as the mean ± standard deviation (SD). A paired t-test was used to compare the continuous variables at final follow-up to the corresponding preoperative values, including VAS and ODI scores, distances x and y, angular displacement, and Taillard index. A p-value<0.05 was considered statistically significant.

RESULTS

In total, 128 patients were included in this study (29 men and 99 women; aged 60.7±11.0 y, range, 35 to 83 y; Tables 1 and 2). The cases consisted of main diagnoses of spondylolisthesis, lumbar disc herniation/lumbar spinal stenosis, compression fractures, and AS with kyphosis deformity in 53, 44, 19 and 12 patients, respectively. One or more diagnoses may be noted in one patient. All patients had osteoporosis (bone mineral density T-score≤-2.5).

Table 1
Case report form.
Table 2
Baseline demographic and clinical characteristics of 128 patients with PMMA-augmented CICPS in the osteoporotic spine.a a Reported as the mean ± SD (range), unless noted otherwise
Table 3
Complications in the 128 patients with PMMA-augmented CICPS.

Additionally, 418 CICPSs were used during the surgeries (Table 1). The mean volume of PMMA injected into each screw was 1.51±0.13 mL (range, 1.2-1.7 mL). The operative time was 216.6±63.4 min (range, 95 to 398 min), and the average blood loss was 553.7±125.9 mL (range, 100 to 1400 mL).

No nerve, blood vessel, or viscera injury occurred during surgery (Table 3). During PMMA injection, no cement leaked into the operative site, and no contamination occurred during the operative procedures. In total, 27 PMMA leakages (6.46%) to the front of 25 vertebral bodies (9.12%) were identified in 20 patients (15.63%) during the procedure. All of these events occurred in prevertebral veins without any observed symptoms. No continuous postoperative bleeding or infection was noted. No symptomatic embolism occurred during hospitalization or follow-up.

All patients underwent follow-up for an average of 42.4±13.4 months (range, 23 to 71 months). Pain and nerve compression symptoms were relieved in all patients postoperatively. The ODI at the final follow-up improved significantly compared with preoperative scores (ODI, p<0.001), and similar results were noted in the VAS scores for the low back and lower limbs (each, p<0.001, n=87 cases with lower limb pain). Spondylolisthesis and spinal kyphosis deformity were corrected satisfactorily after surgery, and no screw loosening, pullout, or fracture occurred (Table 4, and Figures 2 and 3).

Table 4
Clinical and radiographic data analysis at three time points for the 128 patients with PMMA-augmented CICPSs inserted in the osteoporotic spine; data are reported as the mean ± SD (range).
Figure 2
X-ray images. (A) before surgery; (B) immediately after surgery; and (C) at the final follow-up at 52 months in a 31-year-old man with a 4-year history of lower back pain and progressive kyphosis. AS and kyphosis with severe osteoporosis were diagnosed (T-score=−3.0). The patient underwent a partial osteotomy of the key vertebrae without intervertebral fusion. PMMA-augmented CICPSs were used at the ends of the internal fixation instrument, and good spinal correction was obtained.
Figure 3
X-ray and CT images. (A) before; (B) immediately after surgery; and (C) at the final follow-up at 57 months in a 67-year-old woman with a 2-year history of lower back pain. L4 spondylolisthesis and severe osteoporosis was diagnosed (T-score=−3.7). The patient underwent transforaminal lumbar interbody fusion with PMMA-augmented CICPSs bilaterally, and the displacement was completely corrected. No screw loosening occurred, and successful fusion was achieved at the final follow-up of 57 months. Low back pain was ameliorated.

New bone formation within the vertebrae was observed. CT scan results revealed that firm fusions were achieved 6 to 12 months after surgery, and the fusion rate was 100%. No revision was needed during the follow-up. During the follow-up, no patients experienced infection, and pulmonary embolism was not observed.

DISCUSSION

Although pedicle screws are the workhorse of spinal instrumentation in the adult spine (2020. Fujita T, Kostuik JP, Huckell CB, Sieber AN. Complications of spinal fusion in adult patients more than 60 years of age. Orthop Clin North Am. 1998;29(4):669-78, https://doi.org/10.1016/S0030-5898(05)70040-7.
https://doi.org/10.1016/S0030-5898(05)70...
), pedicle screw loosening or pullout is the most severe and common problem in osteoporotic patients who undergo spinal surgery that requires spine internal fixation. Although using larger and longer pedicle screws or screws with various designs of screw-rods and screw-threads increase the purchase of the pedicle screws in the inserted vertebrae (55. Patel PS, Shepherd DE, Hukins DW. The effect of screw insertion angle and thread type on the pullout strength of bone screws in normal and osteoporotic cancellous bone models. Med Eng Phys. 2010;32(8):822-8, https://doi.org/10.1016/j.medengphy.2010.05.005.
https://doi.org/10.1016/j.medengphy.2010...
,2121. Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, et al. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res. 1986;(203):99-112.

22. Thompson JD, Benjamin JB, Szivek JA. Pullout strengths of cannulated and noncannulated cancellous bone screws. Clin Orthop Relat Res. 1997;341(341):241-9.
-2323. Kim YY, Choi WS, Rhyu KW. Assessment of pedicle screw pullout strength based on various screw designs and bone densities-an ex vivo biomechanical study. Spine J. 2012;12(2):164-8, https://doi.org/10.1016/j.spinee.2012.01.014.
https://doi.org/10.1016/j.spinee.2012.01...
), bone cement-augmented pedicle screws have been considered more suitable to stabilize and support the degenerating spinal column (2424. Wittenberg RH, Lee KS, Shea M, White AA 3rd, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res. 1993;296(296):278-87.,2525. Pfeifer BA, Krag MH, Johnson C. Repair of failed transpedicle screw fixation. A biomechanical study comparing polymethylmethacrylate, milled bone, and matchstick bone reconstruction. Spine. 1994;19(3):350-3, https://doi.org/10.1097/00007632-199402000-00017.
https://doi.org/10.1097/00007632-1994020...
). Several experimental and clinical studies have demonstrated that PMMA augmentation can improve resistance to pullout in osteoporotic and normal vertebrae (11. Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29(11):E212-6, https://doi.org/10.1097/00007632-200406010-00020.
https://doi.org/10.1097/00007632-2004060...
,2626. Wuisman PI, Van Dijk M, Staal H, Van Royen BJ. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. Eur Spine J. 2000;9(6):528-33, https://doi.org/10.1007/s005860000169.
https://doi.org/10.1007/s005860000169...

27. Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2(4):255-60, https://doi.org/10.1016/S1529-9430(02)00207-3.
https://doi.org/10.1016/S1529-9430(02)00...
-2828. Turner AW, Gillies RM, Svehla MJ, Saito M, Walsh WR. Hydroxyapatite composite resin cement augmentation of pedicle screw fixation. Clin Orthop Relat Res. 2003;(406):253-61, https://doi.org/10.1097/00003086-200301000-00036.
https://doi.org/10.1097/00003086-2003010...
). In the present long-term follow-up study, we tested our redesigned PMMA-augmented CICPSs, which contain three radial holes and flexible screw heads. We found that this new design was safe and effective for osteoporotic patients who needed spinal instrumentation.

Among bone cements, PMMA is the most frequently used in clinical practice due to its low cost, high availability, and strong mechanical properties (88. Liu D, Shi L, Lei W, Wei MQ, Qu B, Deng SL, et al. Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation: An Experimental Study. Clin Spine Surg. 2016;29(7):E351-7, https://doi.org/10.1097/BSD.0b013e31828bfc85.
https://doi.org/10.1097/BSD.0b013e31828b...
). PMMA is commonly injected directly or via fenestrated screws during vertebroplasty and after the insertion of a balloon in kyphoplasty (88. Liu D, Shi L, Lei W, Wei MQ, Qu B, Deng SL, et al. Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation: An Experimental Study. Clin Spine Surg. 2016;29(7):E351-7, https://doi.org/10.1097/BSD.0b013e31828bfc85.
https://doi.org/10.1097/BSD.0b013e31828b...
,99. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8.
https://doi.org/10.1007/s00586-008-0769-...
,2929. Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4(4):402-8, https://doi.org/10.1016/j.spinee.2003.11.010.
https://doi.org/10.1016/j.spinee.2003.11...
). The former method is preferable and can result in an approximately 80% increase in pullout strength compared with unaugmented screws (3030. Shea TM, Laun J, Gonzalez-Blohm SA, Doulgeris JJ, Lee WE 3rd, Aghayev K, et al. Designs and techniques that improve the pullout strength of pedicle screws in osteoporotic vertebrae: current status. Biomed Res Int. 2014;2014:748393, https://doi.org/10.1155/2014/748393.
https://doi.org/10.1155/2014/748393...
). The distribution of PMMA injected before insertion of the pedicle screw is not controllable; therefore, there is a high risk of leakage as the screw displaces the cement upon instrumentation (99. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8.
https://doi.org/10.1007/s00586-008-0769-...
). PMMA injected via fenestrated screws is better controlled and represents a more effective, safe method.

The presence of side holes in a screw increases the purchase. Chen et al. (3131. Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, et al. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord. 2011;12:33, https://doi.org/10.1186/1471-2474-12-33.
https://doi.org/10.1186/1471-2474-12-33...
) found that the maximum axial pullout strength of the augmented pedicle screw increased as the number of side holes increased, and bone cement exuded mainly from the proximal screw holes rather than from distal holes. These findings indicate that a cannulated pedicle screw with a side hole closer to the screw head can provide greater maximum axial pullout strength because this screw allows more bone cement to outflow and distribute more widely in less time. However, increasing the number of side holes can cause the pedicle screw to more easily fracture, and a side hole close to the screw head can result in increased risk of bone cement leakage.

To solve these problems, we designed a novel cannulated pedicle screw, the CICPS, which has three side holes of different sizes and shapes. The side holes are arranged from smallest to largest from the distal end of the screw and proceeded along two-fifths of the screw length. The central hollow tract is closed at the screw tip. This design allows almost even pressure in the three side holes during PMMA injection; thus, PMMA can be uniformly distributed around the distal half of the screw. This notion has been confirmed in imaging results (3232. Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, et al. Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech. 2009;24(8):613-8, https://doi.org/10.1016/j.clinbiomech.2009.05.002.
https://doi.org/10.1016/j.clinbiomech.20...
). Uniform distribution of the cement effectively avoids clinical complications related to cement leakage into the spinal canal. This type of cannulated screw is putatively significantly better compared with other existing types of cannulated screw systems (1313. Liu Y, Xu J, Sun D, Luo F, Zhang Z, Dai F. Biomechanical and finite element analyses of bone cement-injectable cannulated pedicle screw fixation in osteoporotic bone. J Biomed Mater Res B Appl Biomater. 2016;104(5):960-7, https://doi.org/10.1002/jbm.b.33424.
https://doi.org/10.1002/jbm.b.33424...
,1414. Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, et al. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics. 2015;70(2):114-9, https://doi.org/10.6061/clinics/2015(02)08.
https://doi.org/10.6061/clinics/2015(02)...
).

Using less PMMA has some advantages, such as reducing the risk of bone cement leakage and related complications and reducing methyl methacrylate toxicity. In the present study, the mean volume of PMMA used was 1.51±0.13 mL (range, 1.2-1.7 mL), which is considerably less than that reported by (2.89±0.72 mL, range 2.0-5.0 mL) Frankel et al. (3333. Frankel BM, Jones T, Wang C. Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery. 2007;61(3):531-7; discussion 537-8, https://doi.org/10.1227/01.NEU.0000290899.15567.68.
https://doi.org/10.1227/01.NEU.000029089...
) or (1.83±0.11 mL, range 1.7-2.0 mL) Moon et al. (1212. Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc. 2009;46(4):305-11, https://doi.org/10.3340/jkns.2009.46.4.305.
https://doi.org/10.3340/jkns.2009.46.4.3...
). The cement extravasation rate (15.63%) was considerably reduced compared with that in a recent study by Janssen et al. (3434. Janssen I, Ryang YM, Gempt J, Bette S, Gerhardt J, Kirschke JS, et al. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J. 2017;17(6):837-44, https://doi.org/10.1016/j.spinee.2017.01.009.
https://doi.org/10.1016/j.spinee.2017.01...
) (66.7%), and more than half of the leakage occurred in vertebral fracture cases. The mean BMD (-3.53±0.59) was reduced compared with other reports (-2.70±0.20, -2.80±0.42 and -2.90±0.00). Thus, we questioned whether the lower BMD was the true factor leading to the higher extravasation rate. We analyzed the relationship between cement extravasation rate and BMD. The rate was 5.88% (1/17) in cases with a BMD≤-4.0, compared with the rate of 17.31% (9/52) for cases with -4.0<BMD≤-3.0 and the rate of 16.95% (10/59) for cases with -3.0 <BMD≤-2.5, suggesting that CICPS is more suitable in cases with BMD less than 4.0. Low BMD is not the actual cause of the high bone cement leakage rate in vertebral fracture cases, and other reasons, such as the fracture itself, were likely involved. We also noted that the incidence rate of leakage in males (24.14%, 7/29) was higher than that in females (13.13%, 13/99). No obvious differences in age and BMD were noted between males and females.

In the present study, the design of the multiaxis or single-axis screw head and the dedicated syringe and adapter of the CICPS facilitates the surgical procedure and consequently reduces the operative time and intraoperative blood loss. In this study, the mean operative time was 216.6±63.4 min (range, 95 to 398 min), and the average blood loss was 553.7±125.9 mL (range, 100 to 1400 mL).

In the present study, symptom relief and functional improvement was observed in all patients after surgery. This beneficial effect may be associated with solid internal fixation. Pedicle screw loosening and pullout are the main reasons for internal fixation failure with an incidence of 0.6-11% (3535. Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine. 1993;18(15):2231-8, https://doi.org/10.1097/00007632-199311000-00015.
https://doi.org/10.1097/00007632-1993110...
). No screw loosening or pullout occurred after a mean follow-up of 42 months. Radiological images during the follow-up revealed no obvious displacement of the pedicle screw tips, and bone fusion was achieved in all patients 6 months after surgery. In addition, several clinical reports have demonstrated that screw breakage results in pedicle screw fixation failure. In the present study, there was no fixation failure and no pedicle screw breakage.

We found that the use of PMMA-augmented CICPS improved the angular displacement and the Taillard index relative to preoperative values in patients with lumbar spondylolisthesis. This finding suggests that PMMA-augmented CICPS is effective for the treatment of lumbar spondylolisthesis with osteoporosis. In patients with AS and kyphosis, although the PMMA-augmented CICPS cannot cure AS or prevent the worsening of deformity at the nonsurgical spinal level, the internal fixation system with PMMA-augmented CICPS benefited the surgical segments.

In summary, we found that use of the novel PMMA-augmented CICPS in osteoporotic patients was associated with good clinical outcomes and the absence of obvious complications during a long-term follow-up of 42 months after spinal surgery. Our study suggests that the internal fixation system with PMMA-augmented CICPS is effective and safe for various unstable osteoporotic spines. According to the results, we also concluded that we should be more cautious in using PMMA-augmented CICPS in cases with a BMD less than -4.0, especially in vertebral fracture patients.

Although we obtained positive outcomes, there were several limitations to this study. First, this study was retrospective in nature and was an open study without a control group. Further clinical randomized controlled studies using this method in a larger number of cases will provide additional data to optimize the procedure, which may contribute to improving healing and maximizing functional outcomes.

ACKNOWLEDGMENTS

This study was supported by the Chongqing City Project, No. CSTC2012ggyyjs10015.

REFERENCES

  • 1
    Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29(11):E212-6, https://doi.org/10.1097/00007632-200406010-00020
    » https://doi.org/10.1097/00007632-200406010-00020
  • 2
    Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims--what is best? Spine. 1998;23(12):1374-9, https://doi.org/10.1097/00007632-199806150-00015
    » https://doi.org/10.1097/00007632-199806150-00015
  • 3
    Abshire BB, McLain RF, Valdevit A, Kambic HE. Characteristics of pullout failure in conical and cylindrical pedicle screws after full insertion and back-out. Spine J. 2001;1(6):408-14, https://doi.org/10.1016/S1529-9430(01)00119-X
    » https://doi.org/10.1016/S1529-9430(01)00119-X
  • 4
    Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd. Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J. 2001;1(2):109-14, https://doi.org/10.1016/S1529-9430(01)00020-1
    » https://doi.org/10.1016/S1529-9430(01)00020-1
  • 5
    Patel PS, Shepherd DE, Hukins DW. The effect of screw insertion angle and thread type on the pullout strength of bone screws in normal and osteoporotic cancellous bone models. Med Eng Phys. 2010;32(8):822-8, https://doi.org/10.1016/j.medengphy.2010.05.005
    » https://doi.org/10.1016/j.medengphy.2010.05.005
  • 6
    Zhuang XM, Yu BS, Zheng ZM, Zhang JF, Lu WW. Effect of the degree of osteoporosis on the biomechanical anchoring strength of the sacral pedicle screws: an in vitro comparison between unaugmented bicortical screws and polymethylmethacrylate augmented unicortical screws. Spine. 2010;35(19):E925-31, https://doi.org/10.1097/BRS.0b013e3181c5fb21
    » https://doi.org/10.1097/BRS.0b013e3181c5fb21
  • 7
    Shi L, Wang L, Guo Z, Wu ZX, Liu D, Gao MX, et al. A study of low elastic modulus expandable pedicle screws in osteoporotic sheep. J Spinal Disord Tech. 2012;25(2):123-8, https://doi.org/10.1097/BSD.0b013e31820ef9f7
    » https://doi.org/10.1097/BSD.0b013e31820ef9f7
  • 8
    Liu D, Shi L, Lei W, Wei MQ, Qu B, Deng SL, et al. Biomechanical Comparison of Expansive Pedicle Screw and Polymethylmethacrylate-augmented Pedicle Screw in Osteoporotic Synthetic Bone in Primary Implantation: An Experimental Study. Clin Spine Surg. 2016;29(7):E351-7, https://doi.org/10.1097/BSD.0b013e31828bfc85
    » https://doi.org/10.1097/BSD.0b013e31828bfc85
  • 9
    Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, et al. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462-9, https://doi.org/10.1007/s00586-008-0769-8
    » https://doi.org/10.1007/s00586-008-0769-8
  • 10
    Hickerson LE, Owen JR, Wayne JS, Tuten HR. Calcium Triglyceride Versus Polymethylmethacrylate Augmentation: A Biomechanical Analysis of Pullout Strength. Spine Deform. 2013;1(1):10-5, https://doi.org/10.1016/j.jspd.2012.07.004
    » https://doi.org/10.1016/j.jspd.2012.07.004
  • 11
    Stadelmann VA, Bretton E, Terrier A, Procter P, Pioletti DP. Calcium phosphate cement augmentation of cancellous bone screws can compensate for the absence of cortical fixation. J Biomech. 2010;43(15):2869-74, https://doi.org/10.1016/j.jbiomech.2010.07.025
    » https://doi.org/10.1016/j.jbiomech.2010.07.025
  • 12
    Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc. 2009;46(4):305-11, https://doi.org/10.3340/jkns.2009.46.4.305
    » https://doi.org/10.3340/jkns.2009.46.4.305
  • 13
    Liu Y, Xu J, Sun D, Luo F, Zhang Z, Dai F. Biomechanical and finite element analyses of bone cement-injectable cannulated pedicle screw fixation in osteoporotic bone. J Biomed Mater Res B Appl Biomater. 2016;104(5):960-7, https://doi.org/10.1002/jbm.b.33424
    » https://doi.org/10.1002/jbm.b.33424
  • 14
    Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, et al. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics. 2015;70(2):114-9, https://doi.org/10.6061/clinics/2015(02)08
    » https://doi.org/10.6061/clinics/2015(02)08
  • 15
    Kanis JA, Melton LJ 3rd, Christiansen C, Johnston CC, Khaltaev N. The diagnosis of osteoporosis. J Bone Miner Res. 1994;9(8):1137-41, https://doi.org/10.1002/jbmr.5650090802
    » https://doi.org/10.1002/jbmr.5650090802
  • 16
    Lonner BS, Newton P, Betz R, Scharf C, O'Brien M, Sponseller P, et al. Operative management of Scheuermann's kyphosis in 78 patients: radiographic outcomes, complications, and technique. Spine. 2007;32(24):2644-52, https://doi.org/10.1097/BRS.0b013e31815a5238
    » https://doi.org/10.1097/BRS.0b013e31815a5238
  • 17
    Sun Y, Wang H, Yang D, Zhang N, Yang S, Zhang W, et al. Characterization of radiographic features of consecutive lumbar spondylolisthesis. Medicine. 2016;95(46):e5323, https://doi.org/10.1097/MD.0000000000005323
    » https://doi.org/10.1097/MD.0000000000005323
  • 18
    Sapkas GS, Papadakis SA, Stathakopoulos DP, Papagelopoulos PJ, Badekas AC, Kaiser JH. Evaluation of pedicle screw position in thoracic and lumbar spine fixation using plain radiographs and computed tomography. A prospective study of 35 patients. Spine. 1999;24(18):1926-9, https://doi.org/10.1097/00007632-199909150-00011
    » https://doi.org/10.1097/00007632-199909150-00011
  • 19
    Burkus JK, Foley K, Haid RW, LeHuec JC. Surgical Interbody Research Group--radiographic assessment of interbody fusion devices: fusion criteria for anterior lumbar interbody surgery. Neurosurg Focus. 2001;10(4):E11, https://doi.org/10.3171/foc.2001.10.4.12
    » https://doi.org/10.3171/foc.2001.10.4.12
  • 20
    Fujita T, Kostuik JP, Huckell CB, Sieber AN. Complications of spinal fusion in adult patients more than 60 years of age. Orthop Clin North Am. 1998;29(4):669-78, https://doi.org/10.1016/S0030-5898(05)70040-7
    » https://doi.org/10.1016/S0030-5898(05)70040-7
  • 21
    Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, et al. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res. 1986;(203):99-112.
  • 22
    Thompson JD, Benjamin JB, Szivek JA. Pullout strengths of cannulated and noncannulated cancellous bone screws. Clin Orthop Relat Res. 1997;341(341):241-9.
  • 23
    Kim YY, Choi WS, Rhyu KW. Assessment of pedicle screw pullout strength based on various screw designs and bone densities-an ex vivo biomechanical study. Spine J. 2012;12(2):164-8, https://doi.org/10.1016/j.spinee.2012.01.014
    » https://doi.org/10.1016/j.spinee.2012.01.014
  • 24
    Wittenberg RH, Lee KS, Shea M, White AA 3rd, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res. 1993;296(296):278-87.
  • 25
    Pfeifer BA, Krag MH, Johnson C. Repair of failed transpedicle screw fixation. A biomechanical study comparing polymethylmethacrylate, milled bone, and matchstick bone reconstruction. Spine. 1994;19(3):350-3, https://doi.org/10.1097/00007632-199402000-00017
    » https://doi.org/10.1097/00007632-199402000-00017
  • 26
    Wuisman PI, Van Dijk M, Staal H, Van Royen BJ. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. Eur Spine J. 2000;9(6):528-33, https://doi.org/10.1007/s005860000169
    » https://doi.org/10.1007/s005860000169
  • 27
    Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2(4):255-60, https://doi.org/10.1016/S1529-9430(02)00207-3
    » https://doi.org/10.1016/S1529-9430(02)00207-3
  • 28
    Turner AW, Gillies RM, Svehla MJ, Saito M, Walsh WR. Hydroxyapatite composite resin cement augmentation of pedicle screw fixation. Clin Orthop Relat Res. 2003;(406):253-61, https://doi.org/10.1097/00003086-200301000-00036
    » https://doi.org/10.1097/00003086-200301000-00036
  • 29
    Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4(4):402-8, https://doi.org/10.1016/j.spinee.2003.11.010
    » https://doi.org/10.1016/j.spinee.2003.11.010
  • 30
    Shea TM, Laun J, Gonzalez-Blohm SA, Doulgeris JJ, Lee WE 3rd, Aghayev K, et al. Designs and techniques that improve the pullout strength of pedicle screws in osteoporotic vertebrae: current status. Biomed Res Int. 2014;2014:748393, https://doi.org/10.1155/2014/748393
    » https://doi.org/10.1155/2014/748393
  • 31
    Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, et al. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord. 2011;12:33, https://doi.org/10.1186/1471-2474-12-33
    » https://doi.org/10.1186/1471-2474-12-33
  • 32
    Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, et al. Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech. 2009;24(8):613-8, https://doi.org/10.1016/j.clinbiomech.2009.05.002
    » https://doi.org/10.1016/j.clinbiomech.2009.05.002
  • 33
    Frankel BM, Jones T, Wang C. Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery. 2007;61(3):531-7; discussion 537-8, https://doi.org/10.1227/01.NEU.0000290899.15567.68
    » https://doi.org/10.1227/01.NEU.0000290899.15567.68
  • 34
    Janssen I, Ryang YM, Gempt J, Bette S, Gerhardt J, Kirschke JS, et al. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J. 2017;17(6):837-44, https://doi.org/10.1016/j.spinee.2017.01.009
    » https://doi.org/10.1016/j.spinee.2017.01.009
  • 35
    Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine. 1993;18(15):2231-8, https://doi.org/10.1097/00007632-199311000-00015
    » https://doi.org/10.1097/00007632-199311000-00015

Publication Dates

  • Publication in this collection
    06 June 2019
  • Date of issue
    2019

History

  • Received
    25 Apr 2018
  • Accepted
    21 Dec 2018
Creative Common - by 4.0
This is an Open Access article distributed under the terms of the Creative Commons License (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium or format, provided the original work is properly cited.
Faculdade de Medicina / USP Rua Dr Ovídio Pires de Campos, 225 - 6 and., 05403-010 São Paulo SP - Brazil, Tel.: (55 11) 2661-6235 - São Paulo - SP - Brazil
E-mail: clinics@hc.fm.usp.br