Open-access Intra- and inter-observer reliabilities of cervical sagittal radiographic measurements among evaluators with different levels of training

ABSTRACT

Objective:  To assess the intra- and inter-observer reliabilities of cervical sagittal radiographic measurements performed by evaluators with varying levels of clinical experience.

Methods:  This study analyzed lateral cervical radiographs from 14 patients who had undergone single-level anterior cervical discectomy and fusion. Four evaluators, a spine surgeon, radiologist, spine surgery resident, and medical student, measured five sagittal alignment parameters (C2-C7 Cobb angle, T1 slope, cervical sagittal vertical axis, cervical sagittal vertical axis via C-shape, and T1- C2-C7 Cobb angle). Intra-observer variability was calculated as the deviation from the median measurements. Inter-observer reliability was assessed using intra-class correlation coefficients.

Results:  Intra-observer variability was lowest among experienced evaluators (13% for both the spine surgeon and radiologist) and highest for the medical student (19.4%). The intra-class correlation coefficients ranged from 0.54 to 0.96, indicating moderate to excellent agreement. When only the two most experienced evaluators were included in the analysis, intra-class correlation coefficients exceeded 0.88 for all parameters.

Conclusion:  Observer experience significantly influences the reliability of sagittal cervical radiographic measurements. Although less experienced evaluators demonstrate higher variability, their performance remains within acceptable limits. Standardized training protocols may help improve consistency across different levels of expertise.

Keywords:
Cervical vertebrae; Observer variation; Radiographic image interpretation, computer-assisted; Spine; Spinal fusion; Diskectomy; Students, medical; Surgeons; Radiologists; Reproducibility of results

Highlights

Observer experience improves radiographic measurement reliability.

Spine surgeons and radiologists exhibited the highest intra-class correlation coefficients.

The cervical sagittal vertical axis showed the greatest inter-observer variability.

Standardized training may enhance measurement consistency.


In Brief

This study evaluated the intra- and inter-observer reliabilities of cervical sagittal alignment measurements on lateral radiographs among observers with different experience levels. More experienced evaluators exhibited higher consistency, especially for complex parameters such as the cervical sagittal vertical axis.

INTRODUCTION

The sagittal alignment of the cervical spine is essential for an upright posture and horizontal gaze, as well as optimal energy expenditure, all of which contribute to quality of life.(1) Alterations in cervical sagittal alignment and loss of lordosis have been linked to impaired function and worse patient-reported outcomes.(2,3) Radiographic parameters commonly used to assess cervical alignment include the C2-C7 Cobb angle, C2-C7 cervical sagittal vertical axis (cSVA), and T1 slope. The clinical relevance of these values has been demonstrated through their correlation with postoperative and functional outcomes.(3-7)

The clinical applicability of these parameters depends on the reliability and reproducibility of the measurements, particularly in surgical planning and outcome assessments.(7-9) However, multiple factors, including image quality, measurement technique, and, in particular, the level of experience of the observer, can affect measurement accuracy.(10,11) Evaluating intra- and inter-observer variabilities is therefore essential to validate the consistency of radiographic measurements used in clinical decision making.

Although experienced observers are believed to provide more consistent assessments, few studies have specifically addressed the effect of training level on the reproducibility of sagittal cervical parameters.(10-13) This gap in the literature highlights the need for a systematic analysis of observer-dependent variability, especially in academic and multidisciplinary clinical settings.

This study aimed to evaluate the intra- and inter-observer reliabilities of key cervical sagittal parameters measured by evaluators with different levels of experience. Lateral cervical radiographs of patients who underwent single-level anterior cervical discectomy and fusion (ACDF) using standalone cages were selected for analysis. This manuscript should provide information on the consistency of radiographic measurements among observers with distinct training backgrounds.

OBJECTIVE

To assess the intra- and inter-observer reliabilities of cervical sagittal radiographic measurements performed by evaluators with varying levels of experience.

METHODS

This study was reviewed and approved by the Research Ethics Committee of Hospital Israelita Albert Einstein (CAAE: 78861624.1.0000.0071; # 7.639.625). The requirement for informed consent was waived because the study involved only anonymized radiographic images.

The medical records of patients who underwent single-level ACDF at Hospital Israelita Albert Einstein in 2023 and 2024 were retrospectively reviewed. Patients aged ≥18 years and with available adequate-quality lateral cervical spine radiographs were included in the study. Patients with incomplete images or a history of cervical spinal surgery were excluded from the study. A total of 14 patients met the eligibility criteria and were included in the analysis.

Radiographs were acquired using a standardized imaging protocol to ensure consistency in the lateral cervical spine projection. All clinical and imaging data were anonymized by an independent professional who was not involved in the study; the evaluators were blinded to the identities of the patients and timing of the images. Anonymized images were distributed to the four evaluators for assessment using Surgimap software (Globus Medical, Methuen, MA, USA).

The evaluators consisted of two experienced professionals, a spine surgeon and a radiologist, and two less experienced individuals, a spine surgery resident and a medical student. Each evaluator was instructed to identify specific anatomical landmarks according to predefined guidelines.

The following parameters were evaluated (12, 13): C2-C7 Cobb angle, a measure of cervical lordosis (CL); T1 slope; cSVA; cSVA (C2-C7); and T1-C2-C7 Cobb angle (T1-CL). The C2-C7 Cobb angle is the angle formed between a line parallel to the inferior endplate of C2 and a line parallel to the superior or inferior endplate of C7. The T1 slope represents the angle between the superior endplate of T1 and the horizontal plane. The cSVA is defined as the shortest distance from a vertical plumb line drawn from the center of the C2 vertebral body to the posterosuperior corner of C7, as measured directly by the evaluator. The cSVA (C2-C7) is the cSVA as measured indirectly using the "C-shape" software tool. The T1-CL is calculated by subtracting the CL angle (C2-C7 Cobb) from the T1 slope. The sagittal cervical parameter measurements are illustrated in figure 1.

Figure 1
Cervical sagittal alignment parameter measurements as performed on radiographs using Surgimap software

This study focused exclusively on evaluating intra- and inter-observer reliabilities. Therefore, comparisons between pre- and post-operative images were not performed, although these data were available.

Intra-observer variability was quantified as the absolute deviation of the measurement of each observer from the median value of the parameter.

Inter-observer reliability was assessed by calculating intra-class correlation coefficients (ICCs). Intra-class correlation coefficients of <0.50, 0.50-0.75, 0.75-0.90, and >0.90 are considered to indicate poor, moderate, good, and excellent reliabilities, respectively.(12,13)

RESULTS

Intra-observer agreement

The average intra-observer variability was 13.4% (mean absolute error [MAE]=1.9) for the spine surgeon, 13.1% (MAE=1.78) for the radiologist, 15.2% (MAE=1.88) for the spine resident, and 19.4% (MAE=2.66) for the medical student. The proportions of measurements with less than 20% deviation from the average value of the parameter were 69.2%, 73.5%, 73.1%, and 61.3%, respectively. The detailed percentage errors for each parameter and observer are listed in table 1.

Table 1
Intra-observer variability in cervical spine parameter measurement

Inter-observer agreement

The ICCs are presented in table 2. The T1 slope (ICC=0.937) and C2-C7 Cobb angle (ICC=0.963) demonstrated excellent reliability, with the T1-CL (ICC=0.895) showing good reliability. In contrast, the cSVA (ICC=0.549) and cSVA (C2-C7) (ICC=0.544) showed only moderate reliability. When only the two most experienced observers (the spine surgeon and radiologist) were included in the analyses, ICCs increased across all parameters (Table 3), with the cSVA reaching 0.974 and T1-CL achieving 0.949, indicating excellent agreement.

Table 2
Intra-class correlation coefficients for cervical spine parameter measurements performed by four evaluators
Table 3
Intra-class correlation coefficients for cervical spine parameter measurements performed by the two most experienced evaluators

DISCUSSION

This study demonstrated that the reliability of sagittal cervical alignment measurements increases along with the level of experience of the evaluator. Inter-observer agreement ranged from moderate to excellent for all four evaluators to good or excellent for the two senior evaluators.

Direct and indirect cSVA measurements showed the greatest variability. These parameters reached only moderate levels of inter-observer reliability when all four evaluators were analyzed. These findings align with those of previous reports, which suggested that the cSVA is particularly sensitive to variation due to the difficulty in consistently identifying precise anatomical landmarks, in particular the center of the C2 vertebral body, on lateral radiographs.(14-16) Variations in image quality and subtle differences in head positioning can further increase this difficulty.

As expected, the average measurement error increased as the level of observer experience decreased. However, the overall intra-observer agreement remained within acceptable ranges across all experience levels, suggesting that even less experienced evaluators, when properly guided, can produce reasonably consistent measurements. These findings support the implementation of standardized measurement protocols and structured training, especially in academic and clinical environments, where junior professionals are actively involved in image analysis.(17)

Few studies have specifically investigated the effect of training level on the reproducibility of cervical sagittal parameters. This study contributes novel evidence by showing that although automated tools like Surgimap can aid in measurement standardization, the expertise of the evaluator remains a crucial determinant of reliability. Accurate radiographic interpretation therefore depends not only on the tools used but also on the training and experience of the evaluator.

Inaccurate measurements can lead to erroneous clinical interpretations, suboptimal surgical planning, or flawed research conclusions. Accurate and reproducible radiographic measurements are essential for preoperative planning, postoperative evaluation, and the long-term monitoring of cervical alignment, particularly in patients undergoing procedures such as ACDF. Thus, proper training is not merely an academic concern; it is directly linked to patient care quality.

One limitation of this study was its small sample size, which may have reduced the statistical power of the analysis. Furthermore, although lateral radiographs are widely used for analysis of cervical sagittal parameters in clinical practice owing to their accessibility and low cost, they present inherent limitations for spatial measurements, particularly distance-based parameters such as the cSVA. Future studies using more advanced imaging technologies such as computed tomography or EOS imaging and including a larger number of evaluators across broader training levels may provide deeper insights and improve external validity.

CONCLUSION

The findings of this study demonstrate that the reliability of cervical sagittal parameter measurements, particularly those of the cervical sagittal vertical axis, varied according to the level of experience of the evaluator. These findings highlight the importance of training and standardization in radiographic assessments. Given the critical role of accurate imaging in surgical planning and postoperative evaluation, the appropriate training of professionals involved in image interpretation is essential to ensure consistency and optimize patient outcomes. Future studies involving larger sample sizes and advanced imaging technologies may further enhance measurement accuracy and support the development of more robust, standardized protocols for evaluating cervical sagittal alignment.

REFERENCES

  • 1 Duval-Beaupère G, Schmidt C, Cosson P. A Barycentremetric study of the sagittal shape of spine and pelvis: the conditions required for an economic standing position. Ann Biomed Eng. 1992;20(4):451-62.
  • 2 Martini ML, Neifert SN, Chapman EK, Mroz TE, Rasouli JJ. Cervical spine alignment in the sagittal axis: a review of the best validated measures in clinical practice. Global Spine J. 2021;11(8):1307-12.
  • 3 Kim HJ, Choi BW, Park J, Pesenti S, Lafage V. Anterior cervical discectomy and fusion can restore cervical sagittal alignment in degenerative cervical disease. Eur J Orthop Surg Traumatol. 2019;29(4):767-74.
  • 4 Gillis CC, Kaszuba MC, Traynelis VC. Cervical radiographic parameters in 1- and 2-level anterior cervical discectomy and fusion. J Neurosurg Spine. 2016;25(4):421-9.
  • 5 Lee JH, Park YK, Kim JH. Chronic neck pain in young adults: perspectives on anatomic differences. Spine J. 2014;14(11):2628-38.
  • 6 Yang BS, Lee SK, Song KS, Yoon SP, Jang G, Lee CC, et al. The use of T1 sagittal angle in predicting cervical disc degeneration. Asian Spine J. 2015;9(5):757-61.
  • 7 Iyer S, Nemani VM, Nguyen J, Elysee J, Burapachaisri A, Ames CP, et al. Impact of cervical sagittal alignment parameters on neck disability. Spine. 2016;41(5):371-7.
  • 8 Tang JA, Scheer JK, Smith JS, Deviren V, Bess S, Hart RA, Lafage V, Shaffrey CI, Schwab F, Ames CP; ISSG. The impact of standing regional cervical sagittal alignment on outcomes in posterior cervical fusion surgery. Neurosurgery. 2012;71(3):662-9; discussion 669.
  • 9 Singhatanadgige W, Kang DG, Luksanapruksa P, Peters C, Riew KD. Correlation and reliability of cervical sagittal alignment parameters between lateral cervical radiograph and lateral whole-body EOS stereoradiograph. Global Spine J. 2016;6(6):548-54.
  • 10 Zhang J, Zhang C, Zhong W, Zhao Z, Han F, Han Z, et al. Validity and reliability of a novel iPhone method to rapidly measure cervical sagittal parameters. Sci Rep. 2022;12(1):19579.
  • 11 Kyrölä KK, Salme J, Tuija J, Tero I, Eero K, Arja H. Intra- and interrater reliability of sagittal spinopelvic parameters on full-spine radiographs in adults with symptomatic spinal disorders. Neurospine. 2018;15(2):175-81.
  • 12 Cohen J. Statistical power analysis for the behavioral sciences. 2nd ed. Hillsdale (NJ): Erlbaum; 1988.
  • 13 Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155-63.
  • 14 Limanówka B, Sagan L, Limanówka K, Poncyljusz W. Changes in cervical sagittal balance following anterior cervical discectomy with fusion. Neurol Neurochir Pol. 2024;58(1):120-6.
  • 15 Tomé-Bermejo F, Morales-Valencia JA, Moreno-Pérez J, Marfil-Pérez J, Díaz-Dominguez E, Piñera AR, et al. Degenerative cervical disc disease: long-term changes in sagittal alignment and their clinical implications after cervical interbody fusion cage subsidence: a prospective study with standalone lordotic tantalum cages. Clin Spine Surg. 2017;30(5):E648-55.
  • 16 Salameh M, Bizdikian AJ, Saad E, Saliby RM, Nacouzi R, Khalil N, et al. Reliability assessment of cervical spine parameters measured on full-body radiographs in asymptomatic subjects and patients with spinal deformity. Orthop Traumatol Surg Res. 2021;107(7):103026.
  • 17 Berhouet J, Samargandi R. Emerging innovations in preoperative planning and motion analysis in orthopedic surgery. Diagnostics (Basel). 2024;14(13):1321.

Edited by

Publication Dates

  • Publication in this collection
    01 Dec 2025
  • Date of issue
    2025

History

  • Received
    11 June 2025
  • Accepted
    17 June 2025
location_on
Instituto Israelita de Ensino e Pesquisa Albert Einstein Avenida Albert Einstein, 627/701 , 05651-901 São Paulo - SP, Tel.: (55 11) 2151 0904 - São Paulo - SP - Brazil
E-mail: revista@einstein.br
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Reportar erro