Open-access Agreement and reliability between two types of portable dynamometers: traction and compression

Concordância e confiabilidade entre dois tipos de dinamômetros portáteis: tração e compressão

Abstract

Introduction  The measurement of quadriceps strength is crucial for various clinical applications. However, tra-ditional isokinetic dynamometers are expensive and bulky. Handheld dynamometers represent accessible alternatives, but the agreement between compression and traction dynamometers in evaluating knee exten-sor torque requires investigation.

Objective  To evalua-te the agreement between handheld compression and traction dynamometers in knee extensor torque and to verify intra- and inter-examiner reliability.

Methods  This is a methodological study with a convenience sam-ple. Participants were recruited by researchers in the physiotherapy outpatient clinic of the Federal Univer-sity of Juiz de Fora – University Hospital, between June 2021 and June 2022. Two dynamometers were used: the MMT manual dynamometer from Lafayette Instru-ment Company, and the E-lastic® 200 kg traction dy-namometer.

Results  The reliability of the dynamometers (compression and traction) was excellent, with high intra- and inter-examiner agreement (ICC(2,2) ranging from 0.94 to 0.99). No significant differences were observed between the dynamometers, indicating equiva-lence. The Bland-Altman plots also showed good agree-ment, and Student´s t-tests did not detect significant differences between examiners.

Conclusion  The hand-held compression and traction dynamometers used in this study demonstrated high reliability and equivalence in measurements, allowing flexibility in choosing these instruments in clinical practice. Future research may explore evaluators´ and patients´ perceptions of these measurements.

Lower limbs; Muscle strength dynamometer; Torque; Reproducibility of results

Resumo

Introdução  A mensuração da força do quadríceps é crucial para diversas aplicações clínicas. No entanto, os dinamômetros isocinéticos tradicionais são caros e volumosos. Os dina-mômetros portáteis representam alternativas acessíveis, mas a concordância entre os dinamômetros de compressão e tração na avaliação do torque extensor do joelho requer investigação.

Objetivo  Avaliar a concordância entre os dinamômetros portáteis de compressão e tração no torque extensor do joelho e verificar a confiabilidade intra e interexaminadores.

Métodos  Trata-se de um estudo metodológico com amostra de conveniência. Os participantes foram recrutados por pes-quisadores no ambulatório de fisioterapia da Universidade Federal de Juiz de Fora – Hospital Universitário, entre junho de 2021 e junho de 2022. Foram utilizados dois dinamômetros: o dinamômetro manual MMT da Lafayette Instrument Company e o de tração E-lastic® 200 kg.

Resultados  A confiabilidade dos dinamômetros (compressão e tração) foi excelente, com alta concordância intra e interexaminadores (ICC(2,2) variando de 0,94 a 0,99). Não foram observadas diferenças significativas entre os dinamômetros, indicando equivalência. Os gráficos de Bland-Altman também apresentaram boa concordância e os testes t de Student não detectaram diferenças significativas entre os examinadores.

Conclusão  Os dinamôme-tros portáteis de compressão e tração utilizados neste estudo demonstraram alta confiabilidade e equivalência nas medi-das, permitindo flexibilidade na escolha desses instrumentos na prática clínica. Pesquisas futuras podem explorar as percepções de avaliadores e pacientes sobre essas medidas.

Membros inferiores; Dinamômetro de força muscular; Torque; Reprodutibilidade dos resultados

Introduction

Quadriceps muscle strength is a critical indicator for assessing muscle function, monitoring clinical con-ditions, and predicting clinical outcomes in different populations, including mortality in certain pathological conditions.1-3The isokinetic dynamometer is conside-red the gold standard for assessing muscle strength. However, its high cost and large size limit its use in clinical practice.4-6Alternatively, portable dynamome-ters have been widely used in recent studies due to their affordability, ease of transport, and use.7,8

Previous studies using both manual compression and manual traction dynamometers have shown mo-derate to excellent validity compared to isokinetic dy-namometers.9-11 In addition, manual dynamometry has demonstrated good intra- and inter-examiner reliability.12,13

Despite the usefulness and validity of manual dy-namometers, no studies have been conducted to in-vestigate the agreement between compression and manual traction dynamometers in the assessment of knee extensor muscle torque. This study is necessary to fill this gap in the literature and provide evidence of the reliability of these instruments in clinical practi-ce. Therefore, this study aims to investigate the agree-ment between manual compression and traction dynamometers in the assessment of knee extensor muscle torque, as well as to verify the intra- and inter-examiner reliability in the use of these dynamometers.

Methods

This is a methodological study with a convenience sample. Participants were recruited by researchers in the physiotherapy outpatient clinic from the Federal University of Juiz de Fora – University Hospital, between June 2021 and June 2022. The research was approved by the local Research Ethics Committee (protocol number 5,376,774). All participants who consented to participate in the research signed an informed consent form.

Sample

Participants aged between 18 and 60 years old, from both sexes, were included. Exclusion criteria were presenting pain or musculoskeletal discomfort in the lower limbs at the time of evaluation, orthopedic sur-gery and musculoskeletal injuries in the lower limbs in the last six months, neurological disorders, and those who could not perform the tests.

Sample size calculation

The sample size was estimated using the reference from Walter et al.,14 considering H0 = 0.4, H1 = 0.75, α = 0.05, β = 0.2, n = 2, resulting in 33 individuals. We also observed the consensus-based standards for the selection of Health Measurement Instruments (COSMIN) recommendation of 50 individuals for a good sample size.15

Equipment

Two dynamometers were used: the handheld MMT Lafayette Instrument Company, model 01165 (Lafayette, IN, USA) with the unit of measurement in N (Newton), and the E-lastic® 200 kg traction dynamometer. The compression dynamometer was coupled to a stand designed and printed by a 3D printer, made of ABS polymer.

Assessment protocols

In the first evaluation, sociodemographic and clinical characteristics were collected (age, body mass, height, lower limb dominance, perimetry, and education level). Subsequently, isometric muscle strength tests for knee extensors of the dominant lower limb were conducted, determined by which leg participants would use to kick a ball.16

The tests were conducted by two experienced and previously trained physiotherapist examiners. They were blinded to the outcomes, as were the participants. Ano-ther researcher documented the results. The examiners and dynamometers order were randomized by a re-search team member not involved in data collection, using an algorithm developed for the statistical software R v.4.1.0 (R Core Team 2021). The second assessment was carried out within two weeks.17

One measurement on each dynamometer was per-formed for learning purposes. Participants were asked to perform a 5-second isometric contraction, and peak force values were recorded for two repetitions on the dominant limb. A rest interval of 30 seconds was given for muscle recovery between each repetition, and 3 mi-nutes of rest were allowed for changes in test position to the other dynamometer.18 During measurement, the examiner used a verbal command: “One, two, three, and now: push, push, push, push, push, and relax”. A 5-minute rest was standardized between the evalua-tions of the examiners.19

The torque of the knee extensors (τ) in Newton meters (Nm) was calculated by multiplying the avera-ge of the two contractions force output (F⃗), with the unit of measure in N, by the lever arm (d), according to the formula: τ = F⃗ × d.20 The lever arm was calculated in meters, from the knee joint line to the point of the dynamometer application.21

Knee extensors with compression dynamometer

The participant was seated, with the hips and knees flexed at 90° and the arms crossed at chest level. An inelastic belt was fixed around the thighs to the chair. The dynamometer was positioned 5 cm above an ima-ginary bimalleolar line on the dominant leg and cou-pled to a PVC pipe affixed to the wall.18

Knee extensors with traction dynamometer

The participant was seated, with the hips and knees flexed at 90° and the arms crossed at chest level. An inelastic belt was fixed around the thighs to the chair. The dynamometer was attached to a carabiner, along with a chain on one end to a ladder rung and on the other end attached to a velcro strap, around the ankle, 5 cm above an imaginary bimalleolar line of the domi-nant leg.5

Statistical analysis

The data distribution was checked by the Shapiro-Wilk test, in which some variables were non-parametric. Bootstrapping procedures (1000 resamples; 95% BCa CI) were performed to obtain more reliable results, to correct deviations from normality in the sample distri-bution, in the paired sample t-test to verify whether the torque measurements from each examiner were different. Subsequently, descriptive analysis was used to present the characteristics of the participants.

Reliability was determined through the calculation of the intraclass correlation coefficient, two-way random model (ICC(2,2)), average measures, absolute agree-ment. The reliability coefficients were interpreted as poor (< 0.5), moderate (0.5 – 0.75), good (0.75 – 0.9), and excellent (> 0.9) reliability,22 with α = 0.05 and a 95% confidence interval. The standard error of mea-surement was estimated by the formula SEM = √MSE for absolute agreement.23 Apparently, the power of F- statistics is not altered by violations of normality.24

The Friedman test was conducted to investigate whether the measurements were equivalent; subse-quently, the limits of agreement were investigated by the Bland-Altman plots of the first day measurements, displaying the mean and 95% confidence interval at 1.96 SD (standard deviation) above and below this mean of the difference between the compression and traction dynamometry evaluations by the examiners.23

Results

A total of 52 participants took part in the study; however, one participant did not complete the entire protocol, and one outlier was removed for exceeding more than twice the average of the participants. This way, the study comprised 50 participants, with 32 females and 18 males; 47 reported the right lower limb as being the dominant, and 3 the left lower limb. The sample characteristic data are presented in Table 1.

Table 1
Sample characterization (n = 50)

Intra-examiner and inter-examiner reliability

The intra-examiner reliability obtained for the com-pression dynamometer was considered excellent for both examiner 1 and examiner 2 (ICC(2,2) = 0.95 and 0.96, respectively). For the traction dynamometer, the intra-examiner reliability was also considered excellent for both examiners (ICC(2,2) = 0.95 and 0.94).

The inter-examiner reliability coefficients (ICC(2,2)) for both the compression and traction dynamometers were considered excellent for both examiners, ranging between 0.97 and 0.99 (Table 2).

Table 2
Intra-examiner and inter-examiner reliability of traction and compression dynamometers (n = 50)

The Friedman test was conducted to investigate whether the measurements were equivalent. The results were not statistically significant (χ2 (7) = 1.98, p = 0.96), indicating that the measurements did not show significant differences.

Figure 1 shows the agreement of the dynamometers in the examiner´s measurement using a Bland–Altman plot. For Examiner 1, the limits of agreement ranged from -20.79 to 21.09, while for Examiner 2, the range was from -28.00 to 22.92. In both analyses, most of the points are within the 95% limits, with scores evenly distributed.

Figure 1
Bland-Altman analysis conducted by examiners 1 and 2.

Table 3 shows that there was no difference in the torque of the knee extensors between the examiners, both in the compression dynamometer and in the trac-tion dynamometer.

Table 3
Knee extensor torque in Newton meters (n = 50)

Discussion

This is a pioneering study investigating the agree-ment between two portable dynamometers assessing the knee extensors muscle torque. The results showed no significant differences in the muscle torque of the knee extensors measured with the compression and traction dynamometers, thereby obtaining agreement between the dynamometers measurements, presenting an excellent ICC.

The inter-examiner reliability coefficients obtained during the tests were considered excellent, both for the compression dynamometer (ICC(2,2) = 0.97) and for the traction dynamometer (ICC(2,2) = 0.97). The study by Luedke et al.25 presented a moderate relia-bility (ICC(3,1) = 0.51) according to the classification used in our study,22 however, a stabilization device was not used on the compression dynamometer, which may influence the measurement of the knee extensors tor-que. Some studies have shown that a stabilizing device causes greater reliability.26,27Similarly, Kelln et al.28 ob-served good inter-examiner reliability (ICC 0.7 - 0.89) in the measurement of the knee extensors in young adults and physically active individuals, with the mea-surement being performed in a prone position and wi-thout stabilization on the compression dynamometer. Reliability can be affected by the inadequate force of the examiner and by the lack of stabilization of the participant and the device.29,30 We infer that the measu-rement position may also interfere.

The intra-examiner reliability coefficients obtained for the compression dynamometer for examiners 1 and 2 were also considered excellent (ICC(2,2) = 0.95 and 0.96, respectively). It was observed that examiner 2 presented a higher average torque than examiner 1 with the compression dynamometer (83.44 Nm), but without statistically significant difference. We postulate some issues, such as the verbal command may have been more vigorous in the second assessment, or even the participants´ desire to surpass the result of the first assessment, despite being blinded. Participants always inquired about the values of the measurements, but they were only able to check them at the end of the study.

Jackson et al.18 observed the intra-examiner reliabi-lity of hand held dynamometry with the use of a porta-ble stabilization device for lower limbs in an athletic population and found excellent intra-rater reliability for isometric muscle performance of the lower extremity in-cluding the hip abductors, external rotators, adductors, knee extensors, and plantar flexors of the ankle, with an average of the groups evaluated having an ICC(3,1) of 0.93 to 0.98.

The intra-examiner reliability for the traction dyna-mometer was considered excellent for both examiners (ICC(2,2) = 0.94). In this case, a device for stabilizing the dynamometer was not necessary, however, a strap was used to avoid compensatory movements of the participant. Peek et al.30 used a traction dynamometer and found that it is a reliable method for assessing the hamstrings and quadriceps isometric muscle strength in elite youth soccer players. In the current study, the Bland-Altman limits of agreement for the traction and compression dynamometers conducted by both exa-miners were satisfactory, with scores distributed both above and below the average difference.

Portable dynamometers (traction and compression) are consistent and reliable tools in clinical practice, and the cost-benefit is advantageous. We recommend clini-cians use stabilization in conjunction with the compres-sion dynamometer in assessing the torque of the knee extensors, as both devices are consistent and reliable, and the choice depends on the evaluator´s experience. We suggest that future studies evaluate the perceptions of patients and examiners.

The present study has limitations, the participants were selected by convenience sampling, and our results should be interpreted with caution and should not be extrapolated to athlete populations or those with mus-culoskeletal injuries in the lower limbs.

Conclusion

We can conclude that the compression and traction dynamometers demonstrated high reliability, and the measurements made by the examiners were equivalent and consistent for knee extensor torque. Furthermore, the dynamometers showed agreement. These results are important for the use of dynamometers in clinical practice, and the dynamometer choice can be left to the discretion of the evaluators, considering patient pre-ference. We suggest that future studies evaluate the perceptions of evaluators and patients.

References

  • 1 García-Hermoso A, Cavero-Redondo I, Ramírez-Vélez R, Ruiz JR, Ortega FB, Lee DC, et al. Muscular strength as a predictor of all-cause mortality in an apparently healthy po-pulation: A systematic review and meta-analysis of data from approximately 2 million men and women. Arch Phys Med Rehabil. 2018;99(10):2100-13.e5. https://doi.org/10.10 16/j.apmr.2018.01.008
    » https://doi.org/10.10 16/j.apmr.2018.01.008
  • 2 Loprinzi PD. Lower extremity muscular strength, sedentary behavior, and mortality. Age (Omaha) 2016;38:32. https://doi.org/10.1007/s11357-016-9899-9
    » https://doi.org/10.1007/s11357-016-9899-9
  • 3 Matsuzawa R, Matsunaga A, Wang G, Yamamoto S, Kut-suna T, Ishii A, et al. Relationship between lower extremity muscle strength and all-cause mortality in Japanese patients undergoing dialysis. Phys Ther. 2014;94(7):947-56. https://doi.org/10.2522/ptj.20130270
    » https://doi.org/10.2522/ptj.20130270
  • 4 Bohannon RW. Hand-held compared with isokinetic dyna-mometry for measurement of static knee extension torque (parallel reliability of dynamometers). Clin Phys Physiol Meas. 1990;11:217-22. https://doi.org/10.1088/0143-0815/11/3/004
    » https://doi.org/10.1088/0143-0815/11/3/004
  • 5 Romero-Franco N, Jiménez-Reyes P, Montaño-Munuera JA. Validity and reliability of a low-cost digital dynamometer for measuring isometric strength of lower limb. J Sports Sci. 2017;35(22):2179-84. https://doi.org/10.1080/02640414.2016.1260152
    » https://doi.org/10.1080/02640414.2016.1260152
  • 6 Zapparoli FY, Riberto M. Isokinetic evaluation of the hip flexor and extensor muscles: a systematic review. J Sport Rehabil. 2017;26(6):556-66. https://doi.org/10.1123/jsr.2016- 0036
    » https://doi.org/10.1123/jsr.2016- 0036
  • 7 Bohannon RW, Bubela DJ, Magasi SR, Gershon RC. Relative reliability of three objective tests of limb muscle strength. Isokinet Exerc Sci. 2011;19(2):77-81. https://doi.org/10.3233/IES-2011-0400
    » https://doi.org/10.3233/IES-2011-0400
  • 8 Kosik KB, Johnson NF, Terada M, Thomas AC, Mattacola CG, Gribble PA. Decreased ankle and hip isometric peak torque in young and middle-aged adults with chronic ankle instability. Phys Ther Sport. 2020;43:127-33. https://doi.org/10.1016/ j.ptsp.2020.02.014
    » https://doi.org/10.1016/ j.ptsp.2020.02.014
  • 9 Almeida GPL, Rodrigues HLN, Freitas BW, Lima POP. Relia-bility and validity of the Hip Stability Isometric Test (HipSIT): A new method to assess hip posterolateral muscle strength. J Orthop Sport Phys Ther. 2017;47(12):906-13. https://doi.org/ 10.2519/jospt.2017.7274
    » https://doi.org/10.2519/jospt.2017.7274» https://doi.org/
  • 10 Garcia D, Sousa Neto IV, Monteiro YS, Magalhães DP, Ferreira GML, Grisa R, et al. Reliability and validity of a portable traction dynamometer in knee-strength extension tests: an isometric strength assessment in recreationally active men. Healthcare. 2023;11(10):1466. https://doi.org/10.3390/health care11101466
    » https://doi.org/10.3390/health care11101466
  • 11 Martins J, Silva JR, Silva MRB, Bevilaqua-Grossi D. Reliabi-lity and validity of the belt-stabilized handheld dynamometer in hip-and knee-strength tests. J Athl Train. 2017;52(9):809-19. https://doi.org/10.4085/1062-6050-52.6.04
    » https://doi.org/10.4085/1062-6050-52.6.04
  • 12 Mentiplay BF, Perraton LG, Bower KJ, Adair B, Pua YH, Williams GP, et al. Assessment of lower limb muscle strength and power using hand-held and fixed dynamometry: A relia-bility and validity study. PLoS One. 2015;10(10):e0140822. https://doi.org/10.1371/journal.pone.0140822
    » https://doi.org/10.1371/journal.pone.0140822
  • 13 Padulo J, Trajkovic N, Cular D, Grgantov Z, Madic DM, Di Vico R, et al. Validity and reliability of isometric-bench for knee isometric assessment. Int J Environ Res Public Health. 2020;17(12):4326. https://doi.org/10.3390/ijerph17124326
    » https://doi.org/10.3390/ijerph17124326
  • 14 Walter SD, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. 1998;17(1):101-10. https://doi.org/10.1002/ (SICI)1097-0258(19980115)17:1%3 C101::AID-SIM727%3E3.0.CO;2-E
    » https://doi.org/10.1002/
  • 15 Terwee CB, Mokkink LB, Knol DL, Ostelo RWJG, Bouter LM, de Vet HCW. Rating the methodological quality in syste-matic reviews of studies on measurement properties: a scoring system for the COSMIN checklist. Qual Life Res. 2012;21:651-7. https://doi.org/10.1007/s11136-011-9960-1
    » https://doi.org/10.1007/s11136-011-9960-1
  • 16 Finnoff JT, Hall MM, Kyle K, Krause DA, Lai J, Smith J. Hip Strength and knee pain in high school runners: a prospective study. PM&R. 2011;3(9):792-801. https://doi.org/10.1016/j.pmrj.2011.04.007
    » https://doi.org/10.1016/j.pmrj.2011.04.007
  • 17 Vardeberg TB, Kjersti V. Test?Retest Reliability of the Sunnaas ADL Index. Scand J Occup Ther. 2001;8(3):140-7. https://doi.org/10.1080/110381201750464494
    » https://doi.org/10.1080/110381201750464494
  • 18 Jackson SM, Cheng MS, Smith Jr AR, Kolber MJ. Intrarater reliability of hand held dynamometry in measuring lower extremity isometric strength using a portable stabilization device. Musculoskelet Sci Pract. 2017;27:137-41. https://doi.org/10.1016/j.math.2016.07.010
    » https://doi.org/10.1016/j.math.2016.07.010
  • 19 Bittencourt NFN, Santos TRT, Gonçalves GGP, Coelho AP, Gomes BGBM, Mendonça LM, et al. Reference values of hip abductor torque among youth athletes: Influence of age, sex and sports. Phys Ther Sport. 2016;21:1-6. https://doi.org/10. 1016/j.ptsp.2015.12.005
    » https://doi.org/10. 1016/j.ptsp.2015.12.005
  • 20 Garcia MAC, Fonseca DS, Souza VH. Handheld dynamo-meters for muscle strength assessment: pitfalls, misconcep-tions, and facts. Brazilian J Phys Ther. 2021;25(3):231-2. https://doi.org/10.1016/j.bjpt.2020.09.003
    » https://doi.org/10.1016/j.bjpt.2020.09.003
  • 21 Almeida GPL, Albano TR, Melo AKP. Hand-held dynamo-meter identifies asymmetries in torque of the quadriceps muscle after anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc. 2019;27(8):2494-501. https://doi.org/10.1007/s00167-018-5245-3
    » https://doi.org/10.1007/s00167-018-5245-3
  • 22 Koo TK, Li MY. A guideline of selecting and reporting in-traclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155-63. https://doi.org/10.1016/j.jcm.2016.02.012
    » https://doi.org/10.1016/j.jcm.2016.02.012
  • 23 Portney LG. Foundations of clinical research: Applications to evidence-based practice. Philadelphia: Davis Company; 2020.
  • 24 Blanca MJ, Arnau J, García-Castro FJ, Alarcón R, Bono R. Non-normal data in repeated measures ANOVA: Impact on type I error and power. Psicothema. 2023;35(1):21-9. https://doi.org/10.7334/psicothema2022.292
    » https://doi.org/10.7334/psicothema2022.292
  • 25 Luedke LE, Heiderscheit BC, Williams DSB, Rauh MJ. As-sociation of isometric strength of hip and knee muscles with injury risk in high school cross country runners. Int J Sports Phys Ther. 2015;10(6):868-76.
  • 26 Fenter PC, Bellew JW, Pitts TA, Kay RE. Reliability of stabilised commercial dynamometers for measuring hip abduction strength: a pilot study. Br J Sports Med. 2003;37(4):331-4. https://doi.org/10.1136/bjsm.37.4.331
    » https://doi.org/10.1136/bjsm.37.4.331
  • 27 Nadler SF, DePrince ML, Hauesien N, Malanga GA, Stitik TP, Price E. Portable dynamometer anchoring station for measuring strength of the hip extensors and abductors. Arch Phys Med Rehabil. 2000;81(8):1072-6. https://doi.org/10.1053/apmr.2000.7165
    » https://doi.org/10.1053/apmr.2000.7165
  • 28 Kelln BM, McKeon PO, Gontkof LM, Hertel J. Hand-held dynamometry: Reliability of lower extremity muscle testing in healthy, physically active, young adults. J Sport Rehabil. 2008;17(2):160-70. https://doi.org/10.1123/jsr.17.2.160
    » https://doi.org/10.1123/jsr.17.2.160
  • 29 Kolber MJ, Cleland JA. Strength testing using hand-held dynamometry. Phys Ther Rev. 2005;10(2):99-112. https://doi.org/10.1179/108331905X55730
    » https://doi.org/10.1179/108331905X55730
  • 30 Peek K, Gatherer D, Bennett KJM, Fransen J, Watsford M. Muscle strength characteristics of the hamstrings and qua-driceps in players from a high-level youth football (soccer) Academy. Res Sport Med. 2018;26(3):276-88. https://doi.org/10.1080/15438627.2018.1447475
    » https://doi.org/10.1080/15438627.2018.1447475

Edited by

  • Associate editor:
    Ana Paula Cunha Loureiro

Publication Dates

  • Publication in this collection
    08 Aug 2025
  • Date of issue
    2025