Arsenic and Other Elements Determination in Seawater Using Coprecipitation and Standard Mode Inductively Coupled Plasma Mass Spectrometry

Abstract

The present work aims the determination of As, Co, Cr, Cu, Mn and Pb in seawater samples using coprecipitation extraction procedure before standard mode inductively coupled plasma mass spectrometry (ICP-MS) (no collision/reaction cell). The coprecipitation procedure is based on the precipitation of Mg(OH)2 from the reaction of the Mg2+ ions (present into the seawater) with ammonia added to the sample. Central composite design, desirability function, and response surface methodology were used to optimize the coprecipitation procedure. The optimum values obtained were 1200 µL of aqueous ammonia, 30 min of centrifugation time and 400 µL of nitric acid. The procedure obtained showed low limits of detection since good preconcentration factors were obtained. Good accuracy was also obtained when the reference material Trace Metals 1 in seawater was evaluated. Thus, the proposed procedure can be considered a good strategy for the determination of As, Co, Cr, Cu, Mn and Pb in seawater by ICP-MS, due to its simplicity, low cost, high analytical frequency, besides the good sensitivity and accuracy achieved.

Keywords:
seawater; arsenic; coprecipitation; ICP-MS


Introduction

The oceans are considered one of the main sources of biological activity, biodiversity and biomass production. However, in the last decades, the quality of these aquatic systems has been modified at different scales mainly because of anthropic activities that contribute to the input contaminants in these sites.11 Sandifer, P. A.; Holland, A. F.; Rowles, T. K.; Scott, G. I.; Environ. Health Perspect. 2004, 112, 454. Some trace elements as arsenic, lead and chromium are among the contaminants most investigated in seawater and the increasing of these elements in aquatic systems can cause their accumulation in aquatic biota, causing serious ecological damage.22 Achary, M. S.; Satpathy, K. K.; Panigrahi, S.; Mohanty, A. K.; Padhi, R. K.; Biswas, S.; Prabhu, R. K.; Iakshmi, S. V.; Panigrahy, R. C.; Food Control 2007, 72, 232.,33 Guilherme, L. R. G.; Marques, J. J.; Pierangeli, M. A. P.; Zuliani, D. Q.; Campos, M. L.; Marchi, G. In Tópicos em Ciência do Solo, vol. 4; Sociedade Brasileira de Ciência do Solo: Viçosa, 2013, p. 20. Thus, it becomes relevant to characterize water bodies concerning trace elements.

Among the methods used for trace elements determination in seawater, inductively coupled plasma mass spectrometry (ICP-MS) has a great emphasis due to its high sensitivity, low limits of detection and quantification, besides the optimum analytical frequency.44 Mketo, N.; Nomngongo, N.; Ngila, C.; TrAC, Trends Anal. Chem. 2016, 85, 107.,55 Pérez-Álvarez, E. P.; Garcia, R.; Barrulas, P.; Dias, C; Cabrita, M. J.; Garde-Cerdán; Food Chem. 2019, 270, 273. However, the analysis of saline water by ICP-MS is a challenge, since the NaCl matrix can generate numerous interferences.66 Wang, B.; Lee, C.; Ho, T.; Talanta 2014, 128, 337. The polyatomic interferences caused by the matrix components are significant for arsenic (40Ar35Cl+ on 75As+), copper (40Ar23Na+ on 63Cu+) and chromium (1H16O35Cl+ on 52Cr+), for example.77 Bednar, A. J.; Kirgan, R. A.; Jones, W. T.; Anal. Chim. Acta 2009, 632, 27. Also, the high salt concentration present in seawater results in non-spectral interferences, causing variations, suppression or enhancement, in the signal.88 Korvela, M.; Andersson, M.; Pettersson, J.; J. Anal. At. Spectrom. 2018, 33, 1770. Therefore, to minimize these interferences in seawater analysis by ICP-MS, some approaches should be used such as sample dilution, matrix modification or separation, analyte addition or internal standard calibration procedure, kinetic energy discrimination (KED) or dynamic reaction cell (DRC).99 Illio, S. D.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6. Pinheiro et al.1010 Pinheiro, F. C.; Lorençatto, R.; Barros, A. I.; Amaral, C. D. B.; Nóbrega, J. A.; J. Braz. Chem. Soc. 2018, 29, 1570. evaluated different strategies to determine several elements in seawater using ICP-MS. The authors concluded that the usage of collision gas He was necessary to determine As, Cu and Cr with good limits of detection and accuracy. To overcome the interferences present in seawater analysis by ICP-MS, the matrix separation analytical procedures are a good alternative to correct both non-spectral and spectral interferences, beyond the preconcentration capabilities and the limit of detection enhancement. Among these procedures, solid-phase extraction and coprecipitation are the most used.1111 Sondergaard, J.; Asmund, G.; Larsen, M. M.; MethodsX 2015, 2, 323.

12 Rahmi, D.; Zhu, Y.; Fujimori, E.; Umemura, T.; Haraguchi, H.; Talanta 2007, 72, 600.

13 Minami, T.; Konagaya, W.; Zheng, L.; Takano, S.; Sasaki, M.; Murata, R.; Nakaguchi, Y.; Sohrin, Y.; Anal. Chim. Acta 2015, 854, 183.

14 Sohrin, Y.; Urushihara, S.; Nakatsuka, S.; Kono, T.; Higo, E.; Minami, T.; Norisuye, K.; Umetani, S.; Anal. Chem. 2008, 80, 6267.

15 Ardini, F.; Magi, E.; Grotti, M.; Anal. Chim. Acta 2011, 706, 84.

16 Yang, R.; Ning, Y.; Zhang, A.; Li, Y.; Su, H.; J. Ocean Univ. China 2016, 15, 825.

17 Arslan, Z.; Oymak, T.; White, J.; Anal. Chim. Acta 2018, 1008, 18.

18 Saito, M. A.; Schneider, D. L.; Anal. Chim. Acta 2006, 565, 222.

19 Yatai, L.; Wei, G.; Zhiwei, W.; Jin, L.; Ke, Y.; Guo, Q.; Hu, S.; Microchem. J. 2016, 126, 194.
-2020 Quinto, F.; Golser, R.; Lagos, M.; Plaschke, M.; Schafer, T.; Steier, P.; Geckeis, H.; Anal. Chem. 2015, 87, 5766.

Coprecipitation is a good strategy due to its greater simplicity. Ardini et al.1515 Ardini, F.; Magi, E.; Grotti, M.; Anal. Chim. Acta 2011, 706, 84. determined Cr, Fe, Mn, Pb and Zn in seawater using magnesium hydroxide coprecipitation as sample preparation. The formation of Mg(OH)2 was induced by ammonia and the resulting precipitate was dissolved using 5 mL of dilute nitric acid and the detection was realized by DRC-ICP-MS.1515 Ardini, F.; Magi, E.; Grotti, M.; Anal. Chim. Acta 2011, 706, 84. The procedure resulted in a preconcentration factor of 10-fold and analyte addition was used for calibration. Mg(OH)2 coprecipitation with ammonia was also used by Yang et al.1616 Yang, R.; Ning, Y.; Zhang, A.; Li, Y.; Su, H.; J. Ocean Univ. China 2016, 15, 825. for the extraction of Cu, Zn, Co and V in seawater samples. However, in this work, a volume of 0.5 mL of seawater was used and the precipitate formed was dissolved with dilute HCl to a final volume of 5 mL. The determination of the elements was performed by ICP-MS using standard addition calibration and recoveries ranging from 67 to 96%. Arslan et al.1717 Arslan, Z.; Oymak, T.; White, J.; Anal. Chim. Acta 2018, 1008, 18. employed triethylamine (TEA) for the precipitation of Mg(OH)2 and the extraction of many trace elements in seawater and estuarine water. The main objective of the authors was to substitute ammonia for TEA to avoid the formation of ammonia complexes and to increase the number of elements extracted in comparison with other works. The procedure was repeated three times for better extraction of the analytes and ICP-MS with reaction/collision cell was used in the analysis, for the minimization of interference. The calibration was performed with an analytical curve employing internal standardization. As an alternative, the present work proposes the extraction of several elements (Co, Cr, Cu, Mn and Pb), including arsenic, using the coprecipitation trough the reaction of the ions Mg2+ present in seawater with ammonia to form Mg(OH)2, and posterior determination by ICP-MS. The proposed procedure does not require the use of special gas for collision/reaction to overcome polyatomic interferences and the calibration was done only with internal standardization. For the establishment of the coprecipitation methodology, multivariate optimization was used using the central composite design, desirability function and response surface methodology.

Experimental

Instrumentation

The determination of As, Co, Cr, Cu, Mn and Pb was performed by inductively coupled plasma mass spectrometer, model Nexion 300D (PerkinElmer, USA) with the operational conditions described in Table 1. The premier argon gas with a purity of 99.9992% (Air Products, Brazil) was used for the generation and maintenance of the plasma, nebulization of the sample and as auxiliary gas. The coprecipitation process was performed with the aid of a microprocessor centrifuge (Chimis, Brazil) and vacuum pump (Vacuubrand, Germany).

Table 1
Instrumental conditions of ICP-MS

Reagents and materials

The decontamination of materials and preparation of samples were done using ultrapure water with a resistivity of 18.2 MΩ cm, Purelab Ultra system (Elga, High Wycombe, UK) and HNO3 68% m/m (Neon, Suzano, Brazil). The acid used was sub-boiling distilled using Distilacid BSB939 IR (Berghof, Eningen unter Achalm, Germany).

All materials used in this study were decontaminated with Extran MA 02 neutral (Merck, Rio de Janeiro, Brazil), washed with water and then left in a bath of 15% v/v HNO3 for 24 h. After this period, the materials were washed with ultrapure water.

Quantitative filter paper (UNIFIL, Ruppichteroth, Germany), cellulose acetate membrane (47 mm diameter and pore of 0.45 μm, Analítica, São Paulo, Brazil, Brazil), 50 mL polypropylene tubes (Sarstedt, Nümbrecht, Germany), and aqueous NH3 (32% m/v, Vetec, Duque de Caxias, Brazil) were used for the preparation of the samples. Sodium chloride (99% m/m, Dynamics, Caxias do Sul, Brazil) and magnesium sulfate (Sigma-Aldrich, Steinheim, Germany) were also used in this study.

Calibration solutions were prepared from an adequate dilution with 4% v/v HNO3 of single element standards solutions (> 99.9% m/m, Specsol, São Paulo, Brazil) with concentrations of 1000 mg L-1. The working range was 0.10 to 20 μg L-1. Sc, Y and Rh were used as internal standards (5 μg L-1) prepared from a 1000 mg L-1 multi-element standard solution (SPC Science, Quebec, Canada).

The determination of residual chloride ions was performed using a titrimetric method with silver nitrate (99% m/m, Sigma-Aldrich, Steinheim, Germany), calcium carbonate (99% m/m, Sigma-Aldrich, Steinheim, Germany) and potassium chromate (99.5% m/m, Sigma-Aldrich, Steinheim, Germany).

The certified reference material (CRM) Trace Metals 1 in seawater-QC (Sigma-Aldrich, Steinheim, Germany) was used to evaluate the accuracy of the analytical procedure.

Multivariate optimization

The central composite design was performed to optimize the coprecipitation procedure. The independent variables studied were NH3 volume (µL); centrifugation time (min) and HNO3 volume (µL). The response (dependent variable) evaluated was percentage of recovery. The mathematical model of desirability was applied to the optimization of the multiple responses, transforming them into a single response (y). For this, the recoveries (responses) were first converted to values 0 and 1, called individual desirability (di), where 0 represents an undesirable recovery (< 60% and > 140%) and, 1, a desirable recovery (80 to 120%).2121 Pereira Filho, E. R.; Planejamento Fatorial em Química: Maximizando a Obtenção de Resultados; Edufscar: São Carlos, Brazil, 2015. Also, it was established values between 0 and 1, using recoveries 60 to 80% as minimum range and 120 to 140% as maximum range following the recommendations of AOAC.2222 Official Methods of Analysis of AOAC International; Guidelines for Standard Method Performance Requirements, 20th ed.; Latimer, G. W., ed.; AOAC International: Rockville, 2016.

The individual desirability (0 < di < 1) was calculated using equation 1, where y is the recovery obtained from the experiment, L is the lowest acceptable recovery (60%), and T is the maximum acceptable recovery (80%).

(1) di = y L / T L

For the range 120 to 140%, equation 2 was used to calculate the individual desirability (0 < di < 1). In this situation, U is the maximum acceptable recovery and T the lowest acceptable recovery (120%) to the range under evaluation.

(2) di = U y / U T

After the establishment of individual desirability for each element, a geometric mean of m numbers was used to obtain the global desirability (D) (equation 3).

(3) D = di 1 × di 2 × dm m

The mathematical model obtained was evaluated using variance analysis (ANOVA) and the establishment of optimal values was established from the response surface.2323 Novaes, C. G.; Yamaki, R. T.; Paula, V. F.; Nascimento Jr., B. B.; Barreto, J. A.; Valesques, G. S.; Bezerra, M. A.; Rev. Virtual Quim. 2017, 9, 1184. Microsoft Office Excel 2013 and the free trail of Statistica 7.0 version2424 Statistica, version 7.0; StatSoft, Tulsa, OK, USA, 2004. were used in this study.

Proposed procedure

A volume of 45 mL of filtered and acidified seawater sample was put into a polypropylene tube of 50 mL, 1200 µL of concentrated ammonia were added and the final volume of 50 mL was adjusted with ultrapure water. The content of the tube was homogenized and left to rest for 3 min. After this reaction time, the sample was centrifugated for 30 min at 3000 rpm. Subsequently, the supernatant was discarded, and the precipitate was dissolved using 400 µL of concentrated nitric acid. Ultrapure water was added up to 10 mL and the solution obtained was analyzed by ICP-MS. A scheme of the procedure is represented in Figure 1.

Figure 1
Scheme of the procedure of extraction of As, Co, Cr, Cu, Mn and Pb from seawater sample using Mg(OH)2 coprecipitation.

Results and Discussion

Multivariate optimization

The establishment of significant variables for the central composite design was carried out in advance from a complete factorial design. The significant independent variables for coprecipitation were NH3 volume (µL); centrifugation time (min) and volume of HNO3 (µL).

The optimization was performed using the central composite design and the response surface methodology, which are based on the adjustment of a polynomial equation to the experimental data, describing the behavior of a data group to realize statistical prediction.2323 Novaes, C. G.; Yamaki, R. T.; Paula, V. F.; Nascimento Jr., B. B.; Barreto, J. A.; Valesques, G. S.; Bezerra, M. A.; Rev. Virtual Quim. 2017, 9, 1184. The response used to verify the efficiency of the extraction of the analytes was recovery (in percentage).

The central composite used for the optimization of the variables resulted in 17 experiments, 8 of the factorial design, 6 related to the axial points and 3 repetitions of the central point (23 + 2k + 3).2323 Novaes, C. G.; Yamaki, R. T.; Paula, V. F.; Nascimento Jr., B. B.; Barreto, J. A.; Valesques, G. S.; Bezerra, M. A.; Rev. Virtual Quim. 2017, 9, 1184. The recovery was used as the response variable, so the 17 experiments were performed with seawater and with seawater spiked with 2 µg L-1.

The evaluation of the quality of the model fitted was performed by analysis of variance (ANOVA). The F test was used to evaluate the lack of adjustment of the model, where fcalculated 3.34 was lower than the fcritic 19.3 to a confidence level of 95%, showing that the model does not have lack of adjustment. Besides this, the model presented a pure error low (0.0662) and a good regression coefficient (R2 = 0.790), which indicated 79% of the relationship between the independent factors and the response variable can be explained.

Based on the global desirability, the response surfaces of the model were constructed (Figure S1, Supplementary Information section) to evaluate the behavior of the data and to make predictions statistically valid.2525 Hanrahan, G.; Lu, K.; Crit. Rev. Anal. Chem. 2006, 36, 141. Thus, the critical points of each variable were 1200 µL of aqueous ammonia, 30 min for centrifugation time and 400 µL of nitric acid.

Sample volume

The coprecipitation procedure was optimized using a sample volume of 45 mL. However, in some cases, it may be necessary to use a smaller sample volume. Thus, tests with a smaller sample volume (15 and 25 mL) were performed to evaluate the efficiency of the procedure. The procedure was performed exactly as described above, only changing the initial sample volume. The recoveries obtained (87.7-123%) indicate that a smaller sample volume can be used that is advantageous when the sample quantity available for analysis is restricted and the method of analysis has good sensitivity.

Although the coprecipitation was optimized for As, Cr, Cu, Mn and Pb determination, it was verified that Co determination is also possible using the proposed procedure.

Extraction efficiency

The extraction process of the analytes by coprecipitation was evaluated concerning the residual concentration of chloride ions in the solution under analysis.

The evaluation of the presence of dissolved salts as NaCl in solution is important because this matrix can cause numerous interferences in the ICP-MS. So, chloride ions determination in the final solution was performed by precipitation titrimetry.2626 Skoog, D. A.; West, D. M.; Holler, F. J.; Crouch, S. R.; Fundamentals of Analytical Chemistry, 9th ed.; Thomson Brooks: Boston, USA, 2012. An aliquot of 3.0 mL of the sample solution was titrated with a previously standardized silver nitrate solution (0.1145 mol L-1). The pH of the solution was adjusted to 7 with calcium carbonate.

It was verified that the residual chloride content in the solution was 25% in relation to the original seawater sample. This result indicates that with only one step the coprecipitation procedure was able to eliminate 75% of initial chloride. The proposed methodology can reduce the presence of salts and minimize the spectral and non-spectral interferences arising from the saline matrix in the analyzes by ICP-MS. Thus, the NaCl removal from seawater samples allows the analysis by ICP-MS with no necessity of collision/reaction cell.

Preconcentration factor

The preconcentration factor was evaluated comparing the slopes of the calibration curves, one curve was done by addition of the analytes before the coprecipitation procedure and the other one with the addition at the end of the coprecipitation procedure. The actual preconcentration factor was determined through the ratio of the slopes of the two analytical curves and their found values are varying between 89.9 and 100% of the theoretical value, i.e., 4.5.

Analytical figures of merit

The calibration was performed using internal standardization (Sc, Y and Rh) and good coefficients of determination (greater than 0.999) were obtained. The correction with internal standards was necessary for all analytes to compensate the effects of the residual saline matrix in the analytical solution. The internal standards used were 45Sc for 53Cr, 89Y for 75As, 59Co and 55Mn, and 103Rh for 63Cu and 208Pb.

Limits of detection (LOD) were calculated using the definitions proposed by IUPAC2727 Mocak, J.; Bond, A. M.; Mitchell, S.; Scollary, G.; Pure Appl. Chem. 1997, 69, 297. (LOD = 3s/a), where s is the estimated standard deviation of blank (in this study 15 readings of preparation blank were used) and a is the slope of the analytical curve. The LOD found in the proposed coprecipitation procedure were satisfactory when compared with that published in the literature (Table 2). Low limits allow determining very low concentrations for As and Pb, for example, that are extremely harmful to aquatic biota.

Table 2
Limits of detection (LOD) obtained in the proposed procedure and some published in the literature for trace elements in seawater samples by ICP-MS

The accuracy of the proposed procedure was evaluated using the certified reference material (CRM) and recovery tests.

The recovery test was performed by the addition of 2.5 and 5 μg L-1 of the analytes in seawater samples. The recoveries obtained (Table 3) follow the established by AOAC2222 Official Methods of Analysis of AOAC International; Guidelines for Standard Method Performance Requirements, 20th ed.; Latimer, G. W., ed.; AOAC International: Rockville, 2016. for the concentration range under study. These results indicate that the method is not affected by matrix effects.

Table 3
Recovery values obtained after the addition of analytes at two concentration levels for seawater samples

The certified reference material (CRM) Trace Metals 1 in seawater-QC was used to evaluate the accuracy of the proposed procedure. However, as this CRM is a synthetic saline matrix (3.5% m/m of dissolved salts) and magnesium was not present in quantity enough to be precipitate as Mg(OH)2, the addition of Mg2+ ions was done based on seawater composition.2828 Matsuura, T.; Desalination 2001, 134, 47.

The results (Table 4) show a good agreement among the certified values and the obtained values according to AOAC2222 Official Methods of Analysis of AOAC International; Guidelines for Standard Method Performance Requirements, 20th ed.; Latimer, G. W., ed.; AOAC International: Rockville, 2016. for Co, Cr, Cu, and Pb, indicating the accuracy of the proposed procedure. For As (60.5%) and Mn (71.6%) although the recoveries were low, they should not be disregarded, mainly due to the complexity of the matrix and the simplicity of the procedure for sample preparation. The lower recoveries values for As and Mn may suggest that a coprecipitation step should be once more realized to improve the extraction of these analytes.

Table 4
Concentration (mean ± standard deviation, n = 3) of As, Co, Cr, Cu, Mn and Pb in the CRM metals in seawater and percentage of recoveries to certified values

Real samples analysis

Ten seawater samples from Vitória Bay, located in the state of Espírito Santo, Brazil, were used to verify the applicability of the proposed methodology in real samples. The samples were filtered and acidified, and subsequently, the methodology of coprecipitation was applied before trace elements determination by ICP-MS. The range of values obtained was 11.00-13.50 μg L-1 for As, 0.62-0.99 μg L-1 for Co, 30.42- 43.40 μg L-1 for Cr, 6.26-8.94 μg L-1 for Cu, 1.41-2.00 μg L-1 for Mn and 1.34-8.23 μg L-1 for Pb. The results obtained showed that it was able to quantify all the elements even that elements in low concentration with good precision, indicating that the proposed procedure is suitable to be used in seawater analysis in monitoring studies.

Conclusions

The multivariate optimization of the coprecipitation for As, Co, Cr, Cu, Mn and Pb determination in seawater samples by ICP-MS provided a simple, fast and accurate procedure. Low limits of detection were obtained, due to the preconcentration achieved, and are appropriate to determine the elements in seawater samples. Also, the proposed procedure has demonstrated the possibility to overcome some interferences in ICP-MS seawater analysis with no use of reaction/collision cell.

Supplementary Information

Supplementary data are available free of charge at https://jbcs.sbq.org.br as PDF file.

Acknowledgments

This work was supported by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Brazil (CAPES) finance code 001; Fundação de Amparo à Pesquisa do Espírito Santo-Brazil (FAPES); and LabPetro/UFES-Laboratório de Pesquisa e Desenvolvimento de Metodologias para Análise de Petróleo.

References

  • 1
    Sandifer, P. A.; Holland, A. F.; Rowles, T. K.; Scott, G. I.; Environ. Health Perspect. 2004, 112, 454.
  • 2
    Achary, M. S.; Satpathy, K. K.; Panigrahi, S.; Mohanty, A. K.; Padhi, R. K.; Biswas, S.; Prabhu, R. K.; Iakshmi, S. V.; Panigrahy, R. C.; Food Control 2007, 72, 232.
  • 3
    Guilherme, L. R. G.; Marques, J. J.; Pierangeli, M. A. P.; Zuliani, D. Q.; Campos, M. L.; Marchi, G. In Tópicos em Ciência do Solo, vol. 4; Sociedade Brasileira de Ciência do Solo: Viçosa, 2013, p. 20.
  • 4
    Mketo, N.; Nomngongo, N.; Ngila, C.; TrAC, Trends Anal. Chem. 2016, 85, 107.
  • 5
    Pérez-Álvarez, E. P.; Garcia, R.; Barrulas, P.; Dias, C; Cabrita, M. J.; Garde-Cerdán; Food Chem. 2019, 270, 273.
  • 6
    Wang, B.; Lee, C.; Ho, T.; Talanta 2014, 128, 337.
  • 7
    Bednar, A. J.; Kirgan, R. A.; Jones, W. T.; Anal. Chim. Acta 2009, 632, 27.
  • 8
    Korvela, M.; Andersson, M.; Pettersson, J.; J. Anal. At. Spectrom. 2018, 33, 1770.
  • 9
    Illio, S. D.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6.
  • 10
    Pinheiro, F. C.; Lorençatto, R.; Barros, A. I.; Amaral, C. D. B.; Nóbrega, J. A.; J. Braz. Chem. Soc. 2018, 29, 1570.
  • 11
    Sondergaard, J.; Asmund, G.; Larsen, M. M.; MethodsX 2015, 2, 323.
  • 12
    Rahmi, D.; Zhu, Y.; Fujimori, E.; Umemura, T.; Haraguchi, H.; Talanta 2007, 72, 600.
  • 13
    Minami, T.; Konagaya, W.; Zheng, L.; Takano, S.; Sasaki, M.; Murata, R.; Nakaguchi, Y.; Sohrin, Y.; Anal. Chim. Acta 2015, 854, 183.
  • 14
    Sohrin, Y.; Urushihara, S.; Nakatsuka, S.; Kono, T.; Higo, E.; Minami, T.; Norisuye, K.; Umetani, S.; Anal. Chem. 2008, 80, 6267.
  • 15
    Ardini, F.; Magi, E.; Grotti, M.; Anal. Chim. Acta 2011, 706, 84.
  • 16
    Yang, R.; Ning, Y.; Zhang, A.; Li, Y.; Su, H.; J. Ocean Univ. China 2016, 15, 825.
  • 17
    Arslan, Z.; Oymak, T.; White, J.; Anal. Chim. Acta 2018, 1008, 18.
  • 18
    Saito, M. A.; Schneider, D. L.; Anal. Chim. Acta 2006, 565, 222.
  • 19
    Yatai, L.; Wei, G.; Zhiwei, W.; Jin, L.; Ke, Y.; Guo, Q.; Hu, S.; Microchem. J. 2016, 126, 194.
  • 20
    Quinto, F.; Golser, R.; Lagos, M.; Plaschke, M.; Schafer, T.; Steier, P.; Geckeis, H.; Anal. Chem. 2015, 87, 5766.
  • 21
    Pereira Filho, E. R.; Planejamento Fatorial em Química: Maximizando a Obtenção de Resultados; Edufscar: São Carlos, Brazil, 2015.
  • 22
    Official Methods of Analysis of AOAC International; Guidelines for Standard Method Performance Requirements, 20th ed.; Latimer, G. W., ed.; AOAC International: Rockville, 2016.
  • 23
    Novaes, C. G.; Yamaki, R. T.; Paula, V. F.; Nascimento Jr., B. B.; Barreto, J. A.; Valesques, G. S.; Bezerra, M. A.; Rev. Virtual Quim. 2017, 9, 1184.
  • 24
    Statistica, version 7.0; StatSoft, Tulsa, OK, USA, 2004.
  • 25
    Hanrahan, G.; Lu, K.; Crit. Rev. Anal. Chem. 2006, 36, 141.
  • 26
    Skoog, D. A.; West, D. M.; Holler, F. J.; Crouch, S. R.; Fundamentals of Analytical Chemistry, 9th ed.; Thomson Brooks: Boston, USA, 2012.
  • 27
    Mocak, J.; Bond, A. M.; Mitchell, S.; Scollary, G.; Pure Appl. Chem. 1997, 69, 297.
  • 28