Determination of Chlorine in Crude Oil Emulsified via the MgCl Molecule by HR-CS MAS

Abstract

A method for chlorine determination in crude oil samples, through the formation of MgCl molecule, by high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS GF MAS) was developed and is described in this paper. The use of this molecule for chlorine determination in crude oil has not been reported up to now. Measures were carried out at the wavelength 377.010 nm and the samples were submitted to a simple treatment as emulsion, whose composition was optimized applying a simplex-centroid experimental design. Magnesium was used as molecule forming agent at the concentration of 10 g L-1, at least 500 times higher than the concentration of chlorine. This excess is required to ensure the molecule formation of all content of Cl present in the sample. Chemical modifier Pd was used and improved greatly the signal intensities. Optimized pyrolysis and vaporization temperatures, applied for determination, were 800 and 2300 °C, respectively. Accuracy was evaluated through analysis of certified reference materials, NIST 1848 and NIST 1634c, and addition and recovery tests with recoveries between 92 and 117%. Limits of detection and quantification were 3.0 and 9.0 µg g-1, respectively. The method was applied to eight crude oil samples.

Keywords:
chlorine; magnesium monochloride; crude oil; emulsion; HR-CS GF MAS


Introduction

Crude oil is a natural complex mixture of gaseous and liquid compounds found in reservoirs beneath the surface of the earth associated with sedimentary rocks, composed mostly of aliphatic and aromatic hydrocarbons and containing organic compounds of nitrogen, oxygen and sulfur. Metals and non-metals are also part of the crude oil composition. Chlorine is present in crude oil as inorganic metallic salts, whose content is related to the reservoirs, formation waters and other waters used in extraction processes. Organic bonded chlorides from vegetal and animal sources or formed during crude oil processing steps can also be present. Although found in trace amounts chlorine is very deleterious since during the processing of crude oil it generates hydrochloric acid which is strongly corrosive and can cause serious damage to the components of the equipments.11 Matar, S.; Hatch, L.; Chemistry of Petrochemical Processes, 2nded.; Gulf Publishing Company: Houston, Texas, 2000.

2 Speight, J. G.; Handbook of Petroleum Product Analysis; Wiley Interscience: Hoboken, 2002.

3 Doyle, A.; Saavedra, A.; Tristão, M. L. B.; Nele, M.; Aucélio, R. Q.; Spectrochim. Acta, Part B 2011, 66, 368.
-44 Doyle, A.; Saavedra, A.; Tristão, M. L. B.; Mendes, L. A .N.; Aucélio, R. Q.; Spectrochim. Acta, Part B 2013, 86, 102.

Chlorine determination in crude oil is not an easy task due to the complexity of the sample and the possibility of analyte lost during the treatment procedures prior the analysis. The determination of chlorine has been performed following official methods of the American Society for Testing and Materials, ASTM D3230,55 ASTM Standard D3230-99: Standard Test Methods for Salt in Crude Oils (Electrometric Method), ASTM International, West Conshohocken, PA, 2010. ASTM D647066 ASTM D6470-99: Standard Test Method for Salt in Crude Oils (Potentiometric Method), ASTM International, West Conshohocken, PA, 2010. and ASTM D4929,77 ASTM D4929-04: Standard Test Method for Salt in Crude Oils (Potentiometric Method), ASTM International: West Conshohocken, PA , 2010. and the US Environmental Protection Agency, method EPA 9075.88 EPA Standard Method 9075: Test Method for Total Chlorine in New and Used Petroleum Products by X-Ray Fluorescence Spectromety (XRF), U.S. Environmental Protection Agency, Washington DC, 1994. The methods ASTM D3230 and ASTM D6470 are in fact indirect quantification of chlorine through the measurement of salt content. The procedure of ASTM D4929 describes the quantification of organic chlorides while EPA 9075 defines the total concentration of chlorine in petroleum products but has been applied to crude oil.33 Doyle, A.; Saavedra, A.; Tristão, M. L. B.; Nele, M.; Aucélio, R. Q.; Spectrochim. Acta, Part B 2011, 66, 368.

Determination of chlorine by inductively coupled plasma mass spectrometry (ICP-MS) has been proposed but conventionally the introduction of the samples is carried out through pneumatic nebulization which requires sample treatment procedures that usually causes the loss of the analyte. Alternatively, introduction systems such as laser ablation (LA) or electrothermal vaporization has been employed for direct determination of chlorine determination by ICP-MS but these accessories are not easily available.99 Antes, F. G.; Dullius, E.; Costa, A. B.; Molz, R. F.; Paniz, J. N. G.; Flores, E. M. M.; Dressler, V. L.; Microchem. J. 2013, 109, 117.,1010 Hammerli, J.; Rusk, B.; Spandler, C.; Emsbo, P.; Oliver, N. H. S.; Chem. Geol. 2013, 337-338, 75.

Absorption spectrometry, usually applied to metallic and semimetallic elements, has been recently proposed for determination of halogens through the measurement of the absorption of diatomic molecules. The need for this alternative is due to the fact that atomic halogens resonance lines are located beneath the wavelength work range of conventional spectrometers, below 190 nm, in the vacuum ultraviolet region. The first studies on molecular absorption in air-acetylene flame were conducted in the mid-1970 by Haraguchi and Fuwa1111 Haraguchi, H.; Fuwa, K.; Spectrochim. Acta, Part B 1975, 30, 535. who investigated the spectra of indium halides diatomic molecules in the UV region. In 1980 decade Dittrichet al.1212 Dittrich, K.; Schneider, S.; Anal. Chim. Acta 1980, 115, 189.

13 Dittrich, K.; Schneider, S.; Anal. Chim. Acta 1980, 115, 201.

14 Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1982, 140, 237.
-1515 Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1983, 152, 149. published a series of papers describing the molecular spectra of several metallic halides in a graphite furnace. Nevertheless application for determination in real samples gained impulse only after the introduction of the high resolution continuum source atomic absorption spectrometry (HR-CS AAS), which combines high intensity radiation source and high-resolution spectrometer making easier the selection of the low sensitive molecular wavelength.1616 Welz, B.; Lepri, F. G.; Araujo, R. G. O.; Ferreira, S. L. C.; Huang, M. D.; Okrussc, M.; Becker-Ross, H.; Anal. Chim. Acta 2009, 647, 137.,1717 Butcher, D. J.; Anal. Chim. Acta 2013, 804, 1.

Chlorine determination by high-resolution continuum source molecular absorption spectrometry (HR-CS MAS) has been accomplished by detection of several diatomic molecules.1616 Welz, B.; Lepri, F. G.; Araujo, R. G. O.; Ferreira, S. L. C.; Huang, M. D.; Okrussc, M.; Becker-Ross, H.; Anal. Chim. Acta 2009, 647, 137.,1818 Welz, B.; Becker-Ross, H.; Florek, S.; Heitmann, U.; High- Resolution Continuum Source AAS - The Better Way to Do Atomic Absorption Spectrometry, 1st ed.; Wiley-VCH: Weinheim, Germany, 2005. The most widely reported in the literature is AlCl1919 Heitmann, U.; Becker-Ross, H.; Florek, S.; Huang, M. D.; Okruss, M.; J. Anal. At. Spectrom. 2006, 21, 1314.

20 Fechetia, M.; Tognon, A. L.; Veiga, M. A. M. S.; Spectrochim. Acta, Part B 2012, 71, 98.

21 Enders, M. S. P.; Gomes, A. O.; Oliveira, R. F.; Guimarães, R. C. L.; Mesko, M. F.; Flores, E. M. M.; Müller, E. I.; Energy Fuels 2016, 30, 3637.

22 Parvinen, P.; Lajunen, L. H. J.; Talanta 1999, 50, 67.

23 Nakadi, F. V.; Veiga, M. A. M. S.; Aramendía, M.; García-Ruiza, E.; Resano, M.; J. Anal. At. Spectrom. 2015, 30, 1531.

24 Dittrich, K.; Spivakov, B. Ya.; Shkinev, V. M.; Vorob’eva, G. A.; Talanta 1984, 31, 341.
-2525 Butcher, D. J.; Microchem. J. 1993, 48, 303. due to its high energy bond and well separated band heads, making it effective the background correction.1717 Butcher, D. J.; Anal. Chim. Acta 2013, 804, 1.,2626 Huang, M. D.; Becker-Ross, H.; Florek, S.; Heitmann, U.; Okruss, M.; Spectrochim. Acta, Part B 2006, 61, 959. Strontium monochloride (SrCl) was applied for chlorine determination in different types of samples (coal, milk, animal organs and tissues, and wheat flour). This molecule is monitored at 635.862 nm and displays detection limits that are similar to certified values.2121 Enders, M. S. P.; Gomes, A. O.; Oliveira, R. F.; Guimarães, R. C. L.; Mesko, M. F.; Flores, E. M. M.; Müller, E. I.; Energy Fuels 2016, 30, 3637.,2727 Pereira, E. R.; Welz, B.; Lopez, A. H. D.; Andrade, J. B.; Spectrochim. Acta, Part B 2014, 102, 1.

28 Pereira, E. R.; Rocha, L. M.; Cadorim, H. R.; Silva, V. D.; Welz, B.; Carasek, E.; Andrade, J. B.; Spectrochim. Acta, Part B 2015, 114, 46.
-2929 Ozbek, N.; Akman, S.; J. Agric. Food Chem. 2016, 64, 5767. A method for chlorine determination through CaCl molecule at the wavelength of 621.146 nm was developed by Guardaetal.3030 Guarda, A.; Aramendía, M.; Andrés, I.; García-Ruiz, E.; Nascimento, P. C.; Resano, M.; Talanta 2017, 162, 354. and applied it to different types of certified reference materials (coal, iron oxide, polyethylene, human hair, pine needles, rice flour and milk). The same molecule was used by Bechlinetal.3131 Bechlin, M. A.; Ferreira, E. C.; Gomes Neto, J. A.; Microchem. J. 2017, 132, 130. for chlorine quantification in cement samples at wavelength 621.145 nm, employing direct solid sampling analysis. Low detection limits were achieved since CaCl molecule exhibits high sensitivity.

Chlorine determination in crude oil by HR-CS MAS through several molecules, AlCl, InCl and SrCl, and direct introduction of the sample, was accomplished by Endersetal.2121 Enders, M. S. P.; Gomes, A. O.; Oliveira, R. F.; Guimarães, R. C. L.; Mesko, M. F.; Flores, E. M. M.; Müller, E. I.; Energy Fuels 2016, 30, 3637. The obtained results were compared with those obtained by microwave-induced combustion (MIC), microwave-assisted extraction (MAE), ASTM D6470-9966 ASTM D6470-99: Standard Test Method for Salt in Crude Oils (Potentiometric Method), ASTM International, West Conshohocken, PA, 2010. and neutron activation analysis (NAA) and no significant difference was found.2121 Enders, M. S. P.; Gomes, A. O.; Oliveira, R. F.; Guimarães, R. C. L.; Mesko, M. F.; Flores, E. M. M.; Müller, E. I.; Energy Fuels 2016, 30, 3637.

The use of MgCl molecule for analytical purposes was investigated by Dittrich and Vorberg1515 Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1983, 152, 149. and published in 1982, but no application for Cl determination in real samples was reported since then, even after introduction of the HR-CS AAS instrumentation in the early 2000s. The molecule wavelength, based on literature data, is 377.010 nm, which corresponds to the transition bands of the A2π - X2Σ+ system, visible region, whose molecular spectrum presents broad bands, with two of them exhibiting higher intensities.1414 Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1982, 140, 237.,3232 Pearse, R. W. B.; Gaydon, A. G.; The Identification of Molecular Spectra, 4th ed.; Chapman and Hall: London, UK, 1976, p. 217. Recently a method for chlorine determination using MgCl molecule was developed by our group and applied to produced water from offshore oil wellbore.3333 Medeiros, R. L. S.; Souza, S. O.; Araujo, R. G. O.; Silva, D. R.; Maranhão, T. A.; Talanta 2018, 176, 227. The aim of this paper is to describe a method development for chlorine determination, using MgCl molecule, in crude oil samples by high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS GF MAS). A simple and fast sample treatment, emulsion, is proposed for such high complex matrix.

Experimental

Instrumentation

A high-resolution continuum source atomic absorption spectrometer model ContrAA 700 (AnalytikJena, Jena, Germany) was used for all measurements. Solutions and samples were directly introduced into transverse-heated graphite tubes with integrated platform (AnalytikJena) by an auto sampler model MPE 60 (AnalytikJena). Argon purity 99.999% (White Martins, São Paulo, Brazil) was used as protective and purge gas. To extend the life time of the graphite tube a coating procedure, adopted from literature,3434 Resano, M.; Briceño, J.; Aramendía, M.; Belarra, M. A.; Anal. Chim. Acta 2007, 582, 214. was carried out. Graphite platforms were coated with 1000 µg of tungsten (W) through 20 consecutive injections of 50 μL of a 1000 mg L-1 solution of W and submission to the temperature program shown in Table 1.

Table 1
Furnace temperature program for the coating of the graphite platform with W for chlorine determination by HR-CS MAS

Reagents, solutions and samples

All solutions were prepared with ultrapure water, to a resistivity of 18.2 MΩ cm, obtained in a purifying system model Elga Purelab Ultra (Elga Labwater, United Kingdom). Nitric acid 65% (m v-1) (Labsynth, São Paulo, Brazil), purified in an acid distiller model TE-126 (Tecna, São Paulo, Brazil), xylene (Labimpex, São Paulo, Brazil), n-propanol (Labsynth, São Paulo, Brazil) and Triton X-100 (Fisher Scientific, USA) were used for sample preparation. A stock solution containing 1000 mg L-1 of Cl was prepared from MgCl2.6H2O salt (Labsynth, São Paulo) and used to prepare the calibration curve solutions. Forming agent Mg solution was obtained from dissolution of metallic magnesium (SPEX Industries, USA) in nitric acid and diluted with pure water to a concentration of 10 g L-1 and the modifier Pd 1.0 g L-1 from dissolution of a 10 g L-1 Pd(NO3)2 solution (Merck, Germany). Two certified reference material, fuel oil NIST 1634c and lubricating oil additive NIST 1848, were submitted to the method to evaluate the accuracy. Eight crude oil samples from the Potiguar Basin, Rio Grande do Norte State, Brazil, were used as an application of the method.

Sample preparation

The sample preparation method applied was emulsion, whose composition was optimized through a simplex-centroid design experiment. Crude oil samples were homogenized by stirring with a magnetic stirrer, at the temperature of 80 °C for 30 min. Approximately 0.20 ± 0.01 g of the sample was directly weighed into 15 mL Falcon tubes, 0.5 mL of 5% HNO3 v v-1 were added and, after vigorous agitation, submitted to ultrasonic radiation for 5 min. Then, 0.5 mL of xylene and 2 mL of a mixture of n-propanol/Triton X-100 (5:1) were added. The mixture was again submitted to ultrasonic radiation for 45 min.

Procedure for chlorine determination in crude oil emulsions

Determination of chlorine was carried out by introduction of 10 µL of chlorine calibration solutions, 10 µL of the forming agent Mg and 5 µL of the Pd modifier solution, for calibration curve preparation, and then submitted to the temperature program presented in Table 2. The same procedure was applied to the sample emulsions for quantification.

Table 2
Temperature program for chlorine determination via MgCl molecule in crude oil samples by HR-CS MAS

Results and Discussion

MgCl molecule

Chlorine does not present main resonance lines at the work range of conventional spectrometers, even for the most modern instruments as HR-CS AAS. Thereby an adaptation is required in order to quantify this element using absorption spectrometry technique, the indirect determination through a diatomic molecule. The choice of the molecule, and the wavelength were based in data from literature.3232 Pearse, R. W. B.; Gaydon, A. G.; The Identification of Molecular Spectra, 4th ed.; Chapman and Hall: London, UK, 1976, p. 217. The selected wavelength was 377.010 nm at the visible region corresponding to transition bands of the A2π - X2Σ+ system, which implies the absorption of the molecule at the ground state. The characteristic spectrum is composed by strong bands, presenting two headbands (more intense).3232 Pearse, R. W. B.; Gaydon, A. G.; The Identification of Molecular Spectra, 4th ed.; Chapman and Hall: London, UK, 1976, p. 217. The 3D spectrum obtained for MgCl molecule by HR-CS MAS is shown in Figure 1.

Figure 1
3D spectrum of MgCl molecule by HR-CS GF MAS.

Optimization of crude oil emulsion composition

Crude oil samples are highly complex usually requiring laborious procedures before the introduction into the instrument for analysis. Conventional sample treatment, using strongly acid media and rigorous digestion process usually are not adequate for chlorine since it leads to losses during the procedure. The direct introduction of the samples into the atomizer is a very attractive alternative for complexes samples, made possible by absorption spectrometry. Nevertheless the direct introduction of the sample into the atomizer requires a specific accessory for solid samples and the weighing of each aliquot at each measurement, which is not a simple task for crude oil due to its viscosity. As a consequence there is a substantial increase in the analytical frequency of the method and low repeatability of the measurements is achieved. Additionally, the temperature program steps that precede the vaporization stage, drying and pyrolysis must be carefully optimized to promote the complete burning of the organic matter to avoid the release of smoke during the volatilization of the molecule and consequently interferences. Usually these steps take longer time contributing to an increase in the analytical frequency.

Emulsions are systems composed of two immiscible liquids, usually stabilized by one surfactant, very easy to prepare and with a minimum of manipulation. A short time is required in the emulsion preparation and the use of an autosampler for liquid can be used to introduce the sample into the atomizer, which reduces greatly the time of analysis and ensures the repeatability of the measurements.3535 de Paula, C. E. R.; Cruz, G. F. B.; Rezende, C. M. S. P.; Cassella, R. J.; Microchem. J. 2016, 127, 1.

The composition of the crude oil sample emulsion was optimized through an experimental simplex-centroid design for mixtures of three components. For a design with Z components the number of different points is 2Z – 1. For a design where Z = 3 there will be 7 experiments with 3 mixtures of pure components, 3 binary mixtures and 1 ternary mixture.3636 Bruns, R.; Scarminio, I.; Barros Neto, B.; Statistical Design-Chemometrics, 1st ed.; Elsevier: Amsterdam, Holland, 2006. The volumes of 3 reagents (xylene, nitric acid 5% v v-1 and a n-propanol/Triton X-100 5:1 mixture) used in the composition of emulsion were evaluated in a total of 7 experiments, with 3 replicates of the central point. Table 3 shows the volumes of the reagents, MgCl molecule absorbance intensities, standard deviation (SD) and relative standard deviation (RSD) in each experiment.

Table 3
Simplex-centroid design for optimization of crude oil emulsion composition for chlorine determination via MgCl molecule by HR-CS MAS

Highest intensity was obtained for experiment 7, followed by experiments 4 and 9. Experiments 2 and 8 presented good intensities while for the others the absorbance values were not significant. The contour plot obtained for the experimental design is shown in Figure 2.

Figure 2
Contour surfaces resulting from the ternary model of the simplex-centroid design: xylene, n-propanol/Triton X-100 (5:1) and 5% HNO3.

Five experiments are within the optimal region of the response surface. For the choice of the composition of the emulsion another factor was taken into consideration, the stability of the emulsion. Experiment 8, 500 µL of xylene, 500 µL of nitric acid 5% v v-1 and 2 mL n-propanol and Triton X-100 5:1, was the emulsion with better stability, remaining as a single phase for several hours, making it possible the determination in a numerous samples. So, this composition was adopted for chlorine determination in the real samples of crude oil.

Optimization of the temperature program

In order to establish the furnace temperature program that would better promote de elimination of the sample matrix and the formation of the MgCl molecule, pyrolysis and vaporization curves were optimized for a MgCl2 aqueous solution, containing 10 mg L-1 of Cl, and for a crude oil emulsion, in the absence of modifier and using 5.0 µg Pd as chemical modifier. Magnesium nitrate solution was added as molecule forming agent. Figure 3 shows the obtained pyrolysis and vaporization curves.

Figure 3
Pyrolysis and vaporization curves for the MgCl molecule in a 10 mg L-1 chlorine aqueous solution (a) and a crude oil sample emulsion (b), with and without chemical modifier Pd by HR-CS MAS.

Magnesium forming agent concentration

The effect of the concentration of the forming agent Mg on the formation of MgCl molecule was investigated for a 10 mg L-1 Cl solution and one crude oil sample emulsion. Solutions containing magnesium in nitric acid medium were used in a concentration range from 0 up to 30 g L-1.

For aqueous solutions there is an enhancement of the signal intensities with the increase of Mg concentration up to 10 g L-1 and after that a slight decrease is observed. Probably the excess of Mg promotes a faster release of MgCl molecule to the vapor phase, causing a small decrease in the intensities. A similar increase is observed for crude oil emulsion up to a concentration of 10 g L-1, but after that a slight increase is observed. This could be consequence of formation of other molecules, with other components of the sample and the excess of Mg, causing interferences. However, MgCl molecule is preferentially formed in the vapor phase due to its lower dissociation energy, when compared to other molecule species of Cl.1414 Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1982, 140, 237. As compromise condition the concentration of 10 g L-1 of the forming agent Mg was adopted for aqueous solution and sample emulsion. At this condition the ratio between the forming agent and chlorine is at least 500:1 (in mass). This is the minimum excess required in order to ensure the formation of the molecule for all content of Cl in both aqueous standard solution and sample emulsion.

Matrix interference

To evaluate whether an effect of the matrix would affect the slope of the calibration prepared under the same conditions of the sample emulsions, three calibration solutions were prepared and their slopes compared: aqueous solutions, matrix matching solutions containing the same composition of the emulsion and analyte addition by adding the different concentrations of chlorine to the sample emulsions. Figure 4 presents the obtained calibration curves.

Figure 4
Calibration curves, aqueous, matrix matching and analyte addition, for chlorine determination via MgCl molecule by HR-CS MAS.

For the three calibration curves no significant difference between the slopes were observed. Student's t-test was applied to evaluate whether the values are statistically the same. The obtained results are expressed in Table 4.

Table 4
Results of Student's t-test applied to the calibration curves

For a 95% confidence level (n = 5) the calculated tvalues were lower than the predicted value, 2.77, showing that the slopes of the three calibration curves are statistically equals and could be used for chlorine quantification. Calibration with aqueous solutions was adopted for determination in real samples since it is simpler, the time spent in preparation is shorter and it requires fewer reagents, contributing to faster analysis. Figures of merit for the three calibration curves are shown in Table 5. Analysis of variance (ANOVA) was evaluated through F-test, wherein Fcalculated (0.42) was lower than Ftabulated (161.44), which does not present significant difference between the variances of the data found (at 95% confidence level). Analyte addition calibration was adopted as reference in the calculation.

Table 5
Figures of merit obtained for three calibration curves for chlorine determination via MgCl by HR-CS MAS

In order to evaluate the accuracy of the developed method, chlorine determination was carried out in two certified reference materials, lubricating oil additive NIST 1848 and fuel oil NIST 1634c, prepared as emulsion. The results are shown in Table 6 and the obtained values presented good agreement with the certified values, between 104 and 106% of concordance, ensuring the reliability of the proposed method.

Table 6
Chlorine concentrations in certified reference materials obtained via MgCl molecule by HR-CS MAS

Limits of detection (LOD) and quantification (LOQ) were calculated as 3 and 10 times, respectively, the standard deviation of the analyte intensities, obtained for the blank solution (n = 10), divided by the slope of the calibration curve.3737 International Union Of Pure And Applied Chemistry; Spectrochim. Acta, Part B 1978, 33, 247. The obtained values were 5.0 and 14 µg g-1, for LOD and LOQ, respectively. Pereiraet al.2727 Pereira, E. R.; Welz, B.; Lopez, A. H. D.; Andrade, J. B.; Spectrochim. Acta, Part B 2014, 102, 1.,2828 Pereira, E. R.; Rocha, L. M.; Cadorim, H. R.; Silva, V. D.; Welz, B.; Carasek, E.; Andrade, J. B.; Spectrochim. Acta, Part B 2015, 114, 46. carried out the determination of chlorine using SrCl molecule and obtained LOD values of 2.0 µg g-1, in a method for biological samples, and 5.6 µg g-1 in a method for coal. The determination of chlorine via CaCl molecule in food samples, employing two sample treatments, was also accomplished by Fechetiaet al.,2020 Fechetia, M.; Tognon, A. L.; Veiga, M. A. M. S.; Spectrochim. Acta, Part B 2012, 71, 98. obtaining LOD values of 9.0 and 18 µg g-1. Under the optimized conditions, the characteristic mass of the method was 0.10 ng.

Good precision, evaluated by the repeatability of a set of measurements and expressed as relative standard deviation, was achieved, between 2.0 and 8.0%.

The developed method was applied to eight crude oil samples from the Potiguar Basin in Rio Grande do Norte State, Brazil. Addition and recovery tests were also performed by adding 10 mg L-1 of Cl to the sample emulsions and good recoveries were achieved, between 92 and 117%, considering a range of 80 to 120% as acceptable.3838 Eaton, A. D.; Clesceri, L. S.; Rice, E. W.; Greenberg, A. E.; Standard Methods for the Examination of Water and Wastewater, 21st ed.; American Public Health Association: Washington, USA, 2005, p. 15. The obtained concentrations, RSD values and the recoveries of the addition and recovery tests are shown in Table 7. For five samples the concentrations were below the LOQ and for the others the values ranged from 25 ± 1 mg kg-1 through 50 ± 3 mg kg-1. Precision obtained for the samples containing chlorine were better than 5%. This can be related to the sample preparation as emulsion, which improves the repeatability of the measurements.

Table 7
Chlorine concentration, via MgCl molecule, in crude oil samples at a 95% confidence level (n = 3) by HR-CS MAS

Conclusions

Determination of chlorine using MgCl molecule in crude oil samples by HR-CS MAS is proposed in this paper. Magnesium nitrate solution, 10 g L-1, was used as forming agent and 5 µg of chemical modifier palladium was added to ensure the stability of the molecule during the temperature program of the graphite furnace. A fast and simple calibration strategy, using aqueous solutions was possible. Limits of detection and quantification were 3 and 9 μg g-1, respectively. Accuracy was evaluated through certified reference material and recovery tests. Good agreement with the certified values, 104 and 106%, and recoveries, between 92 and 117%, were achieved. The method was applied to eight crude oil samples.

Acknowledgments

The authors are thankful to Agência Nacional do Petróleo, Gás Natural e Biocombustíveis (ANP) for financial support.

References

  • 1
    Matar, S.; Hatch, L.; Chemistry of Petrochemical Processes, 2nded.; Gulf Publishing Company: Houston, Texas, 2000.
  • 2
    Speight, J. G.; Handbook of Petroleum Product Analysis; Wiley Interscience: Hoboken, 2002.
  • 3
    Doyle, A.; Saavedra, A.; Tristão, M. L. B.; Nele, M.; Aucélio, R. Q.; Spectrochim. Acta, Part B 2011, 66, 368.
  • 4
    Doyle, A.; Saavedra, A.; Tristão, M. L. B.; Mendes, L. A .N.; Aucélio, R. Q.; Spectrochim. Acta, Part B 2013, 86, 102.
  • 5
    ASTM Standard D3230-99: Standard Test Methods for Salt in Crude Oils (Electrometric Method), ASTM International, West Conshohocken, PA, 2010.
  • 6
    ASTM D6470-99: Standard Test Method for Salt in Crude Oils (Potentiometric Method), ASTM International, West Conshohocken, PA, 2010.
  • 7
    ASTM D4929-04: Standard Test Method for Salt in Crude Oils (Potentiometric Method), ASTM International: West Conshohocken, PA , 2010.
  • 8
    EPA Standard Method 9075: Test Method for Total Chlorine in New and Used Petroleum Products by X-Ray Fluorescence Spectromety (XRF), U.S. Environmental Protection Agency, Washington DC, 1994.
  • 9
    Antes, F. G.; Dullius, E.; Costa, A. B.; Molz, R. F.; Paniz, J. N. G.; Flores, E. M. M.; Dressler, V. L.; Microchem. J. 2013, 109, 117.
  • 10
    Hammerli, J.; Rusk, B.; Spandler, C.; Emsbo, P.; Oliver, N. H. S.; Chem. Geol. 2013, 337-338, 75.
  • 11
    Haraguchi, H.; Fuwa, K.; Spectrochim. Acta, Part B 1975, 30, 535.
  • 12
    Dittrich, K.; Schneider, S.; Anal. Chim. Acta 1980, 115, 189.
  • 13
    Dittrich, K.; Schneider, S.; Anal. Chim. Acta 1980, 115, 201.
  • 14
    Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1982, 140, 237.
  • 15
    Dittrich, K.; Vorberg, B.; Anal. Chim. Acta 1983, 152, 149.
  • 16
    Welz, B.; Lepri, F. G.; Araujo, R. G. O.; Ferreira, S. L. C.; Huang, M. D.; Okrussc, M.; Becker-Ross, H.; Anal. Chim. Acta 2009, 647, 137.
  • 17
    Butcher, D. J.; Anal. Chim. Acta 2013, 804, 1.
  • 18
    Welz, B.; Becker-Ross, H.; Florek, S.; Heitmann, U.; High- Resolution Continuum Source AAS - The Better Way to Do Atomic Absorption Spectrometry, 1st ed.; Wiley-VCH: Weinheim, Germany, 2005.
  • 19
    Heitmann, U.; Becker-Ross, H.; Florek, S.; Huang, M. D.; Okruss, M.; J. Anal. At. Spectrom. 2006, 21, 1314.
  • 20
    Fechetia, M.; Tognon, A. L.; Veiga, M. A. M. S.; Spectrochim. Acta, Part B 2012, 71, 98.
  • 21
    Enders, M. S. P.; Gomes, A. O.; Oliveira, R. F.; Guimarães, R. C. L.; Mesko, M. F.; Flores, E. M. M.; Müller, E. I.; Energy Fuels 2016, 30, 3637.
  • 22
    Parvinen, P.; Lajunen, L. H. J.; Talanta 1999, 50, 67.
  • 23
    Nakadi, F. V.; Veiga, M. A. M. S.; Aramendía, M.; García-Ruiza, E.; Resano, M.; J. Anal. At. Spectrom. 2015, 30, 1531.
  • 24
    Dittrich, K.; Spivakov, B. Ya.; Shkinev, V. M.; Vorob’eva, G. A.; Talanta 1984, 31, 341.
  • 25
    Butcher, D. J.; Microchem. J. 1993, 48, 303.
  • 26
    Huang, M. D.; Becker-Ross, H.; Florek, S.; Heitmann, U.; Okruss, M.; Spectrochim. Acta, Part B 2006, 61, 959.
  • 27
    Pereira, E. R.; Welz, B.; Lopez, A. H. D.; Andrade, J. B.; Spectrochim. Acta, Part B 2014, 102, 1.
  • 28
    Pereira, E. R.; Rocha, L. M.; Cadorim, H. R.; Silva, V. D.; Welz, B.; Carasek, E.; Andrade, J. B.; Spectrochim. Acta, Part B 2015, 114, 46.
  • 29
    Ozbek, N.; Akman, S.; J. Agric. Food Chem. 2016, 64, 5767.
  • 30
    Guarda, A.; Aramendía, M.; Andrés, I.; García-Ruiz, E.; Nascimento, P. C.; Resano, M.; Talanta 2017, 162, 354.
  • 31
    Bechlin, M. A.; Ferreira, E. C.; Gomes Neto, J. A.; Microchem. J. 2017, 132, 130.
  • 32
    Pearse, R. W. B.; Gaydon, A. G.; The Identification of Molecular Spectra, 4th ed.; Chapman and Hall: London, UK, 1976, p. 217.
  • 33
    Medeiros, R. L. S.; Souza, S. O.; Araujo, R. G. O.; Silva, D. R.; Maranhão, T. A.; Talanta 2018, 176, 227.
  • 34
    Resano, M.; Briceño, J.; Aramendía, M.; Belarra, M. A.; Anal. Chim. Acta 2007, 582, 214.
  • 35
    de Paula, C. E. R.; Cruz, G. F. B.; Rezende, C. M. S. P.; Cassella, R. J.; Microchem. J. 2016, 127, 1.
  • 36
    Bruns, R.; Scarminio, I.; Barros Neto, B.; Statistical Design-Chemometrics, 1st ed.; Elsevier: Amsterdam, Holland, 2006.
  • 37
    International Union Of Pure And Applied Chemistry; Spectrochim. Acta, Part B 1978, 33, 247.
  • 38
    Eaton, A. D.; Clesceri, L. S.; Rice, E. W.; Greenberg, A. E.; Standard Methods for the Examination of Water and Wastewater, 21st ed.; American Public Health Association: Washington, USA, 2005, p. 15.

Publication Dates

  • Publication in this collection
    Mar 2018

History

  • Received
    1 Aug 2017
  • Accepted
    22 Sept 2017