Use of carbon dioxide for therapeutic decision-making in endoleaks: a case report

Uso de CO2 na tomada de decisão terapêutica de endoleak: relato de caso

Abstract

Endovascular aneurysm repair is currently the most frequently treatment modality for infrarenal aortic aneurysms. Endoleaks are the most common cause of reintervention after endovascular aneurysm repair. It is often unclear which type of endoleak is the correct diagnose, making the treatment decision difficult. We report the case of a 72-year-old man with an endoleak two years after endovascular aneurysm repair. Images suggested a type III endoleak, but this was not confirmed by contrast aortography. We proceeded with the investigation using aortography with carbon dioxide and observed a type IA endoleak. This was successfully treated by implantation of a proximal cuff. A review of the literature shows that the role of carbon dioxide in endoleak management is still unclear. We present a case in which carbon dioxide was essential to both diagnosis and therapeutic decision-making in a type IA endoleak.

Keywords:
endoleak; endovascular procedure; aneurism; carbon dioxide; angiography

Resumo

O tratamento endovascular dos aneurismas de aorta abdominal é atualmente a modalidade de tratamento mais comum. Os endoleaks representam a causa mais frequente de reintervenção após o tratamento endovascular. O diagnóstico do tipo de endoleak frequentemente é incerto, tornando o tratamento desafiador. Apresentamos o caso de um paciente de 72 anos, com endoleak após 2 anos de tratamento endovascular de aneurisma de aorta abdominal. Os exames de imagem pré-operatórios sugeriam um endoleak tipo III; entretanto, durante aortografia com contraste iodado, não foi possível identificá-lo. Optamos por realizar aortografia com dióxido de carbono (CO2), sendo, então, identificado um endoleak tipo IA, que foi tratado com sucesso com o uso de uma extensão (cuff) proximal. O papel do CO2 no diagnóstico de endoleaks ainda não está claro. Relatamos um caso em que o uso do CO2 foi essencial para o diagnóstico e para a decisão de tratamento do endoleak tipo IA.

Palavras-chave:
endoleak; procedimento endovascular; aneurisma; dióxido de carbono; angiografia

INTRODUCTION

Endovascular repair (EVAR) is now the most frequent treatment modality for infrarenal aortic aneurysms.11 de Almeida Mendes C, de Arruda Martins A, Teivelis MP, Kuzniec S, Varella AYM, Wolosker N. Carbon dioxide as contrast medium to guide endovascular aortic aneurysm repair. Ann Vasc Surg. 2017;39:67-73. http://dx.doi.org/10.1016/j.avsg.2016.06.028. PMid:27671460.
http://dx.doi.org/10.1016/j.avsg.2016.06...
It is associated with lower 30-day mortality rates compared to open repair.22 Budtz-Lilly J, Wanhainen A, Mani K. Outcomes of endovascular aortic repair in the modern era. J Cardiovasc Surg (Torino). 2018;59(2):180-9. PMid:29206004. EVAR is also correlated with a higher reintervention rate, which can be as high as 20% in the first 5 years.33 Patel SR, Allen C, Grima MJ, et al. A systematic review of predictors of reintervention after evar: guidance for risk-stratified surveillance. Vasc Endovascular Surg. 2017;51(6):417-28. http://dx.doi.org/10.1177/1538574417712648. PMid:28656809.
http://dx.doi.org/10.1177/15385744177126...

Endoleaks constitute the main cause of reinterventions.44 AbuRahma AF, Hass SM, AbuRahma ZT, et al. Management of immediate post-endovascular aortic aneurysm repair type ia endoleaks and late outcomes. J Am Coll Surg. 2017;224(4):740-8. http://dx.doi.org/10.1016/j.jamcollsurg.2016.12.014. PMid:28017805.
http://dx.doi.org/10.1016/j.jamcollsurg....
It is often unclear which type of endoleak is the correct diagnose,55 Sueyoshi E, Nagayama H, Sakamoto I, Uetani M. Carbon dioxide digital subtraction angiography as an option for detection of endoleaks in endovascular abdominal aortic aneurysm repair procedure. J Vasc Surg. 2015;61(2):298-303. http://dx.doi.org/10.1016/j.jvs.2014.07.088. PMid:25151598.
http://dx.doi.org/10.1016/j.jvs.2014.07....
making the treatment decision difficult.

Diagnostic aortography is generally performed with iodinated contrast. However, even with power injectors and high doses of iodinated contrast, there is occasionally still doubt with regard to which type of endoleak is present. In this situation, carbon dioxide (CO2) represents an alternative method that can solve the problem.66 Mascoli C, Faggioli G, Gallitto E, et al. The assessment of carbon dioxide automated angiography in type ii endoleaks detection: comparison with contrast-enhanced ultrasound. Contrast Media Mol Imaging. 2018;2018:7647165. http://dx.doi.org/10.1155/2018/7647165. PMid:29780288.
http://dx.doi.org/10.1155/2018/7647165...

We conducted a literature review and were unable to find any papers reporting use of CO2 as being crucial for the decision on how to treat an endoleak.77 Garvin RP, Ryer EJ, Kendrick JB, Franklin DP. Endoleak visualized with carbon dioxide angiography during endovascular aneurysm repair using the endurant stent-graft. J Endovasc Ther. 2014;21(1):172-6. http://dx.doi.org/10.1583/12-4123MR.1. PMid:24502499.
http://dx.doi.org/10.1583/12-4123MR.1...

We report a case in which an endoleak identified after EVAR was defined as type III by computed tomography angiography (CTA) and contrast-enhanced ultrasound (CEU). However, during the intervention, iodinated contrast aortography did not show presence of any endoleaks, but when aortography with CO2 was used, a type IA endoleak was evident. This was successfully treated with placement of a proximal cuff in the aorta.

CASE REPORT

We describe the case of a 72-year-old male patient with a past medical history of hypertension and atrial fibrillation treated by anticoagulation with warfarin. He originally presented with an infrarenal aortic aneurysm with a maximum diameter of 6.5 cm and a concomitant left common iliac artery aneurysm of 4.9 cm.

The patient was successfully treated by EVAR with Incraft® main body - 30mm x 98mm; ipsilateral branch - 13mm x 140mm; and contralateral branch - 24mm x 100mm (Cordis, Cardinal Health, Dublin, OH, EUA) and embolization of the left internal iliac artery with coils, two years ago.

In accordance with our institutional protocol, the patient underwent CTA 30 days postoperatively, showing a type II endoleak connecting the left iliac artery aneurysm (embolized), inferior mesenteric artery, and two lumbar arteries, with expansion of the aneurysm sac to 7 cm.

Since the patient was under anticoagulation, we first attempted cessation of warfarin for 3 months and clinical follow-up. After that period, CTA showed the same endoleak, but with stability of the aneurysm diameter and less contrast flow in the sac.

After 6 months, a repeat CTA demonstrated stability of the aneurysm diameter.

At the 1-year control CTA, a 7.1 cm diameter sac was observed with no remaining type II endoleaks. However, a new contrast area was seen on the anterior surface of the stent graft, which was suspected to be a type III endoleak (Figure 1).

Figure 1
CTA 1 year after the EVAR showing leaking contrast at the anterior face of the main endograft body.

During further investigation, CEU showed a high-flow endoleak on the anterior surface and a possible stent fracture. Also, a low-flow endoleak of the anterior lumbar and the left internal iliac arteries were observed with CEU (Figure 2).

Figure 2
(A) CEU showing endoleak in the anterior face of the endograft; (B) B-mode image of the endograft.

At this point the decision was taken to proceed with endovascular treatment, due to the small change in the aneurysm sac and the suspicion of an endograft fracture.

During the intervention, aortography was performed with iodinated contrast and a power injector in 30-degree left oblique and 90-degree left oblique incidence angles, and no endoleaks were identified (Figure 3).

Figure 3
Aortography with iodinated contrast and power injector with no visible endoleak.

Different positions were assessed with the pigtail catheter: next to the suprarenal fixation, to the main body of the endograft, and inside the iliac branches. No contrast flow was seen in the aneurysm sac.

We further performed aortography with CO2. The first injection was made manually with a pigtail catheter and 30 ml of CO2 in a 60 ml syringe with a 90-degree left oblique incidence angle, but no endoleaks were observed.

Finally, we used a Simmons 1 catheter with the same incidence and volume, identifying an endoleak at the anterior surface, originating from the proximal portion of the endograft, which we characterized as a type IA endoleak (Figure 4).

Figure 4
Aortography with CO2 and Simmons 1 catheter. An endoleak was identified in the free-flow area (black arrow), arising from the proximal part of the main body.

No signs of fracture or contrast leakage from the stent graft could be seen.

After this diagnosis, the patient was successfully treated with a proximal cuff (Incraft® 34 mm x 42 mm). Control aortography with CO2 showed no endoleaks.

Thirty days after the intervention, another CTA was performed, showing no contrast in the aneurysm sac.

DISCUSSION

Endoleaks are the main cause of reintervention after EVAR. When type I endoleaks are identified at the time of the final aortography after EVAR, they are usually considered to justify treatment.44 AbuRahma AF, Hass SM, AbuRahma ZT, et al. Management of immediate post-endovascular aortic aneurysm repair type ia endoleaks and late outcomes. J Am Coll Surg. 2017;224(4):740-8. http://dx.doi.org/10.1016/j.jamcollsurg.2016.12.014. PMid:28017805.
http://dx.doi.org/10.1016/j.jamcollsurg....
Type II endoleaks are mostly benign. However, in a few cases, they can cause enlargement of the sac and can eventually lead to rupture.88 Wolosker N, Varella AY, Teivelis MP, Mendes CA, Garcia RG, Pfeferman E. Successful image-guided percutaneous embolization of a ruptured abdominal aortic aneurysm sac due to type II endoleak after endovascular repair. Ann Vasc Surg. 2015;29(2):361.e1-4. http://dx.doi.org/10.1016/j.avsg.2014.10.016. PMID: 25463347.
http://dx.doi.org/10.1016/j.avsg.2014.10...
Type III endoleaks are associated with a higher risk of rupture and aneurysm sac enlargement and should always be treated.99 Madigan MC, Singh MJ, Chaer RA, Al-Khoury GE, Makaroun MS. Occult type I or III endoleaks are a common cause of failure of type II endoleak treatment after endovascular aortic repair. J Vasc Surg. 2019;69(2):432-9. http://dx.doi.org/10.1016/j.jvs.2018.04.054. PMid:30686338.
http://dx.doi.org/10.1016/j.jvs.2018.04....

Conflicting diagnostic results regarding the type of endoleak are frequent. In a 2019 study by Madigan et al. with 130 patients with type II endoleaks, 22% of the patients who needed treatment had a type I or III endoleak that had been previously misidentified and were only correctly diagnosed during the intervention.99 Madigan MC, Singh MJ, Chaer RA, Al-Khoury GE, Makaroun MS. Occult type I or III endoleaks are a common cause of failure of type II endoleak treatment after endovascular aortic repair. J Vasc Surg. 2019;69(2):432-9. http://dx.doi.org/10.1016/j.jvs.2018.04.054. PMid:30686338.
http://dx.doi.org/10.1016/j.jvs.2018.04....

In the case reported here, the patient was followed-up in order to investigate the type II endoleak and, due to sac enlargement and suspicion of a stent graft fracture, a treatment decision was necessary.

During the preoperative investigation, a type III endoleak had been considered, but it was only during the endovascular procedure, after CO2 injection, that a type I endoleak was identified.

It is important to remember that both CTA and CEU results suggested a type III endoleak, and only CO2 aortography was capable of demonstrating that the endoleak was actually from the proximal landing zone of the endograft (type IA).

It is no novelty that CO2 can be used as an alternative contrast for EVAR. In 2011, Criado et al. performed 114 EVAR procedures with CO2 and observed 20 endoleaks (two type I, sixteen type II, and two type IV). No additional endoleaks were diagnosed with iodinated contrast. No type I or III endoleaks were identified with CTA after the first month of follow-up.1010 Criado E, Upchurch GR Jr, Young K, et al. Endovascular aortic aneurysm repair with carbon dioxide guided angiography in patients with renal insufficiency. J Vasc Surg. 2011;54(5):1544-5. http://dx.doi.org/10.1016/j.jvs.2011.09.030. PMid:22341837.
http://dx.doi.org/10.1016/j.jvs.2011.09....

The comparison between CO2 and iodinated contrast for detecting endoleaks is still under debate.

In 2018, Mascoli et al. compared the accuracy of CO2 with iodinated contrast and CEU in 21 patients. All type I and III endoleaks were detected with iodinated contrast and CO2. Seven type II endoleaks were identified with CO2; of these, only one was diagnosed with iodinated contrast and four with CEU; therefore, two of them were only recognized using carbon dioxide.55 Sueyoshi E, Nagayama H, Sakamoto I, Uetani M. Carbon dioxide digital subtraction angiography as an option for detection of endoleaks in endovascular abdominal aortic aneurysm repair procedure. J Vasc Surg. 2015;61(2):298-303. http://dx.doi.org/10.1016/j.jvs.2014.07.088. PMid:25151598.
http://dx.doi.org/10.1016/j.jvs.2014.07....

In 2015, Sueyoshi et al. observed that all type I and type III endoleaks were identified by both methods (iodinated contrast and CO2), but CO2 was less sensitive than iodinated contrast for detecting type II endoleaks.22 Budtz-Lilly J, Wanhainen A, Mani K. Outcomes of endovascular aortic repair in the modern era. J Cardiovasc Surg (Torino). 2018;59(2):180-9. PMid:29206004.

An automated CO2 injection method for EVAR was presented by Mascoli et al. in 2018. In this study with 31 patients, it was possible to identify all type I and III endoleaks using the automated method. However, they reported that CO2 had higher sensitivity for type II endoleaks than iodinated contrast.1111 Mascoli C, Faggioli G, Gallitto E, et al. Standardization of a carbon dioxide automated system for endovascular aortic aneurysm repair. Ann Vasc Surg. 2018;51:160-9. http://dx.doi.org/10.1016/j.avsg.2018.01.099. PMid:29522871.
http://dx.doi.org/10.1016/j.avsg.2018.01...

Even though CO2 was manually injected in this case report, it was only when the gas was used that it was possible to detect the type I endoleak.

One technical point was that the image was only conclusive after injection with an end-hole catheter in the same direction as the blood flow. This technique has been previously described for performing EVAR by Criado et al., in 2011, and by Mendes et al. in 2016.11 de Almeida Mendes C, de Arruda Martins A, Teivelis MP, Kuzniec S, Varella AYM, Wolosker N. Carbon dioxide as contrast medium to guide endovascular aortic aneurysm repair. Ann Vasc Surg. 2017;39:67-73. http://dx.doi.org/10.1016/j.avsg.2016.06.028. PMid:27671460.
http://dx.doi.org/10.1016/j.avsg.2016.06...
,1010 Criado E, Upchurch GR Jr, Young K, et al. Endovascular aortic aneurysm repair with carbon dioxide guided angiography in patients with renal insufficiency. J Vasc Surg. 2011;54(5):1544-5. http://dx.doi.org/10.1016/j.jvs.2011.09.030. PMid:22341837.
http://dx.doi.org/10.1016/j.jvs.2011.09....

In 2011, Criado et al. used an end-hole catheter to conduct final angiograms after EVAR.1010 Criado E, Upchurch GR Jr, Young K, et al. Endovascular aortic aneurysm repair with carbon dioxide guided angiography in patients with renal insufficiency. J Vasc Surg. 2011;54(5):1544-5. http://dx.doi.org/10.1016/j.jvs.2011.09.030. PMid:22341837.
http://dx.doi.org/10.1016/j.jvs.2011.09....
In 2018, Mascoli et al. injected CO2 from the sheath, with the patient in the Trendelenburg position.66 Mascoli C, Faggioli G, Gallitto E, et al. The assessment of carbon dioxide automated angiography in type ii endoleaks detection: comparison with contrast-enhanced ultrasound. Contrast Media Mol Imaging. 2018;2018:7647165. http://dx.doi.org/10.1155/2018/7647165. PMid:29780288.
http://dx.doi.org/10.1155/2018/7647165...
This could have enhanced the good sensitivity they achieved with CO2 for identifying endoleaks. Sueyoshi et al. did not provide enough information about the method of CO2 injection used.55 Sueyoshi E, Nagayama H, Sakamoto I, Uetani M. Carbon dioxide digital subtraction angiography as an option for detection of endoleaks in endovascular abdominal aortic aneurysm repair procedure. J Vasc Surg. 2015;61(2):298-303. http://dx.doi.org/10.1016/j.jvs.2014.07.088. PMid:25151598.
http://dx.doi.org/10.1016/j.jvs.2014.07....

When CO2 is used with a pigtail catheter, its low viscosity leads to explosive effect, producing bubbles and resulting in lower quality images. It provides higher quality images when injected with an end-hole catheter.

The patient described in this study was admitted to the operating room with a suspected type III endoleak diagnosed by CEU and CTA. Injection of iodinated contrast excluded this diagnosis, but did not show the location of the endoleak that had been shown in previous images.

Use of CO2 permitted detection of the type IA endoleak. This was probably because it was a small endoleak, hence the slow enlargement of the aneurysm sac. Once the endoleak had been diagnosed, it was possible to treat it using a proximal cuff to repair the problem.

CONCLUSION

Treatment of endoleaks after EVAR represents a challenge. It is important to use multiple diagnostic and therapeutic methods for proper management.

Angiography with CO2 constitutes a supplementary method to conventional angiography with iodinated contrast. However, further studies to establish the real accuracy of this method would be welcomed.

  • How to cite: Megale AB, Mendes CA, Teivelis MP, Faustino CB, Souza KP, Wolosker N. Use of carbon dioxide for therapeutic decision-making in endoleaks: a case report. J Vasc Bras. 2020;19:e20200060. https://doi.org/10.1590/1677-5449.200060
  • Financial support: None.
  • The study was carried out at Hospital Israelita Albert Einstein, São Paulo, SP, Brazil.

REFERENCES

  • 1
    de Almeida Mendes C, de Arruda Martins A, Teivelis MP, Kuzniec S, Varella AYM, Wolosker N. Carbon dioxide as contrast medium to guide endovascular aortic aneurysm repair. Ann Vasc Surg. 2017;39:67-73. http://dx.doi.org/10.1016/j.avsg.2016.06.028 PMid:27671460.
    » http://dx.doi.org/10.1016/j.avsg.2016.06.028
  • 2
    Budtz-Lilly J, Wanhainen A, Mani K. Outcomes of endovascular aortic repair in the modern era. J Cardiovasc Surg (Torino). 2018;59(2):180-9. PMid:29206004.
  • 3
    Patel SR, Allen C, Grima MJ, et al. A systematic review of predictors of reintervention after evar: guidance for risk-stratified surveillance. Vasc Endovascular Surg. 2017;51(6):417-28. http://dx.doi.org/10.1177/1538574417712648 PMid:28656809.
    » http://dx.doi.org/10.1177/1538574417712648
  • 4
    AbuRahma AF, Hass SM, AbuRahma ZT, et al. Management of immediate post-endovascular aortic aneurysm repair type ia endoleaks and late outcomes. J Am Coll Surg. 2017;224(4):740-8. http://dx.doi.org/10.1016/j.jamcollsurg.2016.12.014 PMid:28017805.
    » http://dx.doi.org/10.1016/j.jamcollsurg.2016.12.014
  • 5
    Sueyoshi E, Nagayama H, Sakamoto I, Uetani M. Carbon dioxide digital subtraction angiography as an option for detection of endoleaks in endovascular abdominal aortic aneurysm repair procedure. J Vasc Surg. 2015;61(2):298-303. http://dx.doi.org/10.1016/j.jvs.2014.07.088 PMid:25151598.
    » http://dx.doi.org/10.1016/j.jvs.2014.07.088
  • 6
    Mascoli C, Faggioli G, Gallitto E, et al. The assessment of carbon dioxide automated angiography in type ii endoleaks detection: comparison with contrast-enhanced ultrasound. Contrast Media Mol Imaging. 2018;2018:7647165. http://dx.doi.org/10.1155/2018/7647165 PMid:29780288.
    » http://dx.doi.org/10.1155/2018/7647165
  • 7
    Garvin RP, Ryer EJ, Kendrick JB, Franklin DP. Endoleak visualized with carbon dioxide angiography during endovascular aneurysm repair using the endurant stent-graft. J Endovasc Ther. 2014;21(1):172-6. http://dx.doi.org/10.1583/12-4123MR.1 PMid:24502499.
    » http://dx.doi.org/10.1583/12-4123MR.1
  • 8
    Wolosker N, Varella AY, Teivelis MP, Mendes CA, Garcia RG, Pfeferman E. Successful image-guided percutaneous embolization of a ruptured abdominal aortic aneurysm sac due to type II endoleak after endovascular repair. Ann Vasc Surg. 2015;29(2):361.e1-4. http://dx.doi.org/10.1016/j.avsg.2014.10.016 PMID: 25463347.
    » http://dx.doi.org/10.1016/j.avsg.2014.10.016
  • 9
    Madigan MC, Singh MJ, Chaer RA, Al-Khoury GE, Makaroun MS. Occult type I or III endoleaks are a common cause of failure of type II endoleak treatment after endovascular aortic repair. J Vasc Surg. 2019;69(2):432-9. http://dx.doi.org/10.1016/j.jvs.2018.04.054 PMid:30686338.
    » http://dx.doi.org/10.1016/j.jvs.2018.04.054
  • 10
    Criado E, Upchurch GR Jr, Young K, et al. Endovascular aortic aneurysm repair with carbon dioxide guided angiography in patients with renal insufficiency. J Vasc Surg. 2011;54(5):1544-5. http://dx.doi.org/10.1016/j.jvs.2011.09.030 PMid:22341837.
    » http://dx.doi.org/10.1016/j.jvs.2011.09.030
  • 11
    Mascoli C, Faggioli G, Gallitto E, et al. Standardization of a carbon dioxide automated system for endovascular aortic aneurysm repair. Ann Vasc Surg. 2018;51:160-9. http://dx.doi.org/10.1016/j.avsg.2018.01.099 PMid:29522871.
    » http://dx.doi.org/10.1016/j.avsg.2018.01.099

Publication Dates

  • Publication in this collection
    16 Nov 2020
  • Date of issue
    2020

History

  • Received
    16 May 2020
  • Accepted
    29 June 2020
Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV) Rua Estela, 515, bloco E, conj. 21, Vila Mariana, CEP04011-002 - São Paulo, SP, Tel.: (11) 5084.3482 / 5084.2853 - Porto Alegre - RS - Brazil
E-mail: secretaria@sbacv.org.br