Accessibility / Report Error

Auditory steady-state response in the auditory evaluation: clinical application

BACKGROUND: the auditory steady state response (ASSR) is an electrophysiological procedure that evaluates, at the same time, the hearing thresholds of both ears at various frequencies. This procedure reduces the time of testing and enables the stimulation of levels close to 125dB HL, characterizing residual hearing. AIM: to verify the applicability of the ASSR to determine the hearing thresholds in different levels of sensorineural hearing loss. METHOD: 48 individuals aged 7-30 years, with different levels of hearing loss were assessed. The Pure tone Audiometry (PTA) and the ASSR were carried out in following carrier frequencies 0.5; 1; 2 and 4k Hz. The carrier frequencies in the ASSR were modulated in amplitude and frequency, with multiple dicotic stimulation for mild and moderate hearing loss. Simple stimulation was used to test the other hearing deficit levels. RESULTS: there was a significant association (p<0.01) between the thresholds obtained in the PTA and in the ASSR for all of the tested frequencies, especially for the severe levels. However, for some patients, the hearing loss level was overestimated in the ASSR. CONCLUSION: the ASSR can be used to estimate the PTA threshold; nevertheless, it should not be analyzed in an isolated way, but seen as a complementary procedure to the behavioral auditory assessment.

Auditory Evoked Potential; Steady-State; Audiometry; Sensorineural Hearing Loss


Pró-Fono Produtos Especializados para Fonoaudiologia Ltda. Condomínio Alphaville Conde Comercial, Rua Gêmeos, 22, 06473-020 Barueri , São Paulo/SP, Tel.: (11) 4688-2220, Fax: (11) 4688-0147 - Barueri - SP - Brazil
E-mail: revista@profono.com.br