Acessibilidade / Reportar erro

Development and validation of a rapid analytical method by HPLC for determination of nimesulide in release studies

A high performance liquid chromatography (HPLC) method has been developed for a rapid determination of nimesulide in dissolution studies. Nimesulide was analyzed using 5 µm Lichrospher® RP-18 column (125 x 4 mm i.d.) and mobile phase acetonitrile: phosphate buffer pH=6.0 (55:45) at a flow-rate of 1.0 mL min-1. Detection was carried out at 300 nm at 25 ºC. The method was applied to analysis of nimesulide in in vitro release studies and showed a rapid and efficient analytical alternative for evaluation of dissolution profile of nimesulide.

nimesulide; high performance liquid chromatography; dissolution study


Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil
E-mail: quimicanova@sbq.org.br