Evaluation of dose standards in pediatric radiology

The dose area product was evaluated and the entrance skin dose was subsequently estimated by means of suitable calculations for pediatric examinations. Results showed values around 100% above the reference levels (0.070 mGy). The technical parameters of the X-ray tubes were periodically checked since a routine quality assurance program is implemented in the hospital. Results showed a good performance of the equipment complying with recommended limits according to Brazilian legislation. Therefore, the high doses can be attributed to the low kV techniques and the lack of appropriate training of the radiographers to perform pediatric examinations. Measurements of scattered radiation were performed in the gonads with the use of thermoluminescent dosimeters which proved to be inadequate for these measurements. Therefore an ionization chamber was used instead and the values obtained showed that the doses were far below the values that can cause permanent or transient sterility.

Radiation doses; Pediatric radiology; Image quality