Acessibilidade / Reportar erro

Validation of automated image co-registration integrated into in-house software for voxel-based internal dosimetry on single-photon emission computed tomography images

Validação do corregistro automático de imagem integrado a software de análise de dosimetria interna baseada em voxel para tomografia computadorizada de emissão de fóton único

Abstract

Objective:

To develop an automated co-registration system and test its performance, with and without a fiducial marker, on single-photon emission computed tomography (SPECT) images.

Materials and Methods:

Three SPECT/CT scans were acquired for each rotation of a Jaszczak phantom (to 0°, 5°, and 10° in relation to the bed axis), with and without a fiducial marker. Two rigid co-registration software packages-SPM12 and NMDose-coreg-were employed, and the percent root mean square error (%RMSE) was calculated in order to assess the quality of the co-registrations. Uniformity, contrast, and resolution were measured before and after co-registration. The NMDose-coreg software was employed to calculate the renal doses in 12 patients treated with 177Lu-DOTATATE, and we compared those with the values obtained with the Organ Level INternal Dose Assessment for EXponential Modeling (OLINDA/EXM) software.

Results:

The use of a fiducial marker had no significant effect on the quality of co-registration on SPECT images, as measured by %RMSE (p = 0.40). After co-registration, uniformity, contrast, and resolution did not differ between the images acquired with fiducial markers and those acquired without. Preliminary clinical application showed mean total processing times of 9 ± 3 min/patient for NMDose-coreg and 64 ± 10 min/patient for OLINDA/EXM, with a strong correlation between the two, despite the lower renal doses obtained with NMDose-coreg.

Conclusion:

The use of NMDose-coreg allows fast co-registration of SPECT images, with no loss of uniformity, contrast, or resolution. The use of a fiducial marker does not appear to increase the accuracy of co-registration on phantoms.

Keywords:
Dosimetry; Dose-response relationship; radiation; Tomography; emission-computed; single-photon; Image processing; computer-assisted

Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem Av. Paulista, 37 - 7º andar - conjunto 71, 01311-902 - São Paulo - SP, Tel.: +55 11 3372-4541, Fax: 3285-1690, Fax: +55 11 3285-1690 - São Paulo - SP - Brazil
E-mail: radiologiabrasileira@cbr.org.br