Acessibilidade / Reportar erro

Pacemakers and implantable cardioverter defibrillators - general and anesthetic considerations

Abstracts

A pacemaking system consists of an impulse generator and lead or leads to carry the electrical impulse to the patient's heart. Pacemaker and implantable cardioverter defibrillator codes were made to describe the type of pacemaker or implantable cardioverter defibrillator implanted. Indications for pacing and implantable cardioverter defibrillator implantation were given by the American College of Cardiologists. Certain pacemakers have magnet-operated reed switches incorporated; however, magnet application can have serious adverse effects; hence, devices should be considered programmable unless known otherwise. When a device patient undergoes any procedure (with or without anesthesia), special precautions have to be observed including a focused history/physical examination, interrogation of pacemaker before and after the procedure, emergency drugs/temporary pacing and defibrillation, reprogramming of pacemaker and disabling certain pacemaker functions if required, monitoring of electrolyte and metabolic disturbance and avoiding certain drugs and equipments that can interfere with pacemaker function. If unanticipated device interactions are found, consider discontinuation of the procedure until the source of interference can be eliminated or managed and all corrective measures should be taken to ensure proper pacemaker function should be done. Post procedure, the cardiac rate and rhythm should be monitored continuously and emergency drugs and equipments should be kept ready and consultation with a cardiologist or a pacemaker-implantable cardioverter defibrillator service may be necessary.

Pacemakers; Implantable cardioverter defibrillators; Anesthesia for device patients


Marcapasso (MP) é um sistema que consiste em um gerador de impulso e eletrodo ou eletrodos que conduzem o impulso elétrico para o coração do paciente. Códigos de MP e cardioversor desfibrilador implantável (CDI) foram criados para descrever o tipo de MP ou CDI implantado. Indicações para a implantação de MP e CDI foram descritas pelo Colégio Americano de Cardiologistas. Alguns MPs têm interruptores de circuitos magneticamente operados incorporados, porém a aplicação magnética pode causar efeitos adversos graves; portanto, os dispositivos devem ser considerados programáveis, salvo conhecimento em contrário. Quando um paciente portador de dispositivo é submetido a qualquer procedimento (com ou sem anestesia), precauções especiais devem ser tomadas, incluindo história/exame físico focados; perguntas sobre o MP antes e após o procedimento, uso de medicamentos de emergência/estimulação e desfibrilação temporárias; reprogramação do MP e desabilitação de certas funções, caso necessário; monitoração dos distúrbios metabólicos e eletrolíticos e evitar certos medicamentos e equipamentos que podem interferir na função do MP. Se houver previsão de interações com o dispositivo, considerar a suspensão do procedimento até que a fonte de interferência tenha sido eliminada ou controlada e todas as medidas corretivas devem ser tomadas para garantir o funcionamento adequado do MP. Após o procedimento, frequência e ritmo cardíacos devem ser monitorados continuamente, medicamentos e equipamentos de emergência devem ser mantidos de prontidão e pode ser necessário consultar um cardiologista ou serviço de MP-CDI.

Marcapassos; Cardioversores desfibriladores implantáveis; Anestesia para portadores de dispositivos


El marcapasos (MP) es un sistema que consiste en un generador de impulso y un electrodo o electrodos que conducen el impulso eléctrico hacia el corazón del paciente. Se han creado códigos de MP y de desfibrilador automático implantables para describir el tipo de cada uno. El Colegio Norteamericano de Cardiólogos han descrito indicaciones para la implantación de MP y de desfibrilador automático implantable. Algunos MP tienen interruptores de circuitos magnéticos incorporados, sin embargo, la aplicación magnética puede causar efectos adversos graves; por tanto, los dispositivos deben ser considerados programables, salvo que se tenga un conocimiento diferente. Cuando un paciente portador de dispositivo se somete a cualquier procedimiento (con o sin anestesia), se deben tomar precauciones especiales, incluyendo historial/examen físico dirigidos; preguntas sobre el MP antes y después del procedimiento, uso de medicamentos de urgencia/estimulación y desfibrilación temporales; reprogramación del MP e inhabilitación de ciertas funciones del mismo, si fuere preciso; monitorización de los trastornos metabólicos y electrolíticos y evitar ciertos medicamentos y equipos que puedan interferir con la función del MP. Si hay alguna previsión de interacción con el dispositivo, considerar la suspensión del procedimiento hasta que la fuente de la interferencia haya sido eliminada o controlada. Se deben tomar todas las medidas correctivas para garantizar el funcionamiento adecuado del MP. Después del procedimiento, la frecuencia y el ritmo cardíacos deben ser monitorizados continuamente, y deben estar preparados los medicamentos y equipos de urgencia; y si fuera necesario, consultar con un cardiólogo o con el servicio de MP- desfibrilador automático implantable.

Marcapasos; Desfibriladores automáticos implantables; Anestesia para portadores de dispositivos


Introduction

Battery-operated pacing devices were introduced by C.W. Lillehei and Earl Bakken in 1958.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. The natural progression of pacemaker (PM) developments led to invention of the implantable cardioverter defibrillator (ICD) around 1980 by Michael Morchower.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

A pacemaking system consists of an impulse generator and lead or leads to carry the electrical impulse to the patient's heart. Leads can be unipolar, bipolar or multipolar. Generators with bipolar leads can be programmed to the unipolar mode for pacing, sensing, or both.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

Pacemaker codes

Table 1 shows the pacemaker codes given by the North American Society of Pacing and Electrophysiology (NASPE)/British Pacing and Electrophysiology Group (BPEG) (2002) Generic Pacemaker Code (NBG).22. Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25:260-4.

Table 1
North American Society of Pacing and Electrophysiology (NASPE)/British Pacing and Electrophysiology Group (BPEG) (2002) Generic Pacemaker Code (NBG).2

Dual: Provides atrioventricular (AV) synchrony, where atrial pacing will take place in the "inhibited" mode and the pacing device will ensure that a ventricular event follows.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

Inhibited: The appropriate chamber is paced unless intrinsic electrical activity is detected during the pacing interval.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

Triggered: The pacing device will emit a pulse only in response to a sensed event.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

Indications for pacemaker implantation33. ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. J Am Coll Cardiol. 2008;51:1-62.

1. Bradycardia due to Sinus Node Dysfunction (SND) and Atrioventricular Node Dysfunction (AND).

- SND: Persistent sinus bradycardia and chronotropic incompetence without identifiable causes, symptomatic bradycardia.

- Acquired atrioventricular (AV) block in adults:

- Third-degree and advanced second-degree AV block at any anatomic level associated with:

  • ° Bradycardia with symptoms/ventricular arrhythmias presumed to be due to AV block.

  • ° Other medical conditions that require drug therapy that results in symptomatic bradycardia.

  • ° Symptom-free patients in sinus rhythm, with documented periods of asystole greater than or equal to 3 s, an escape rate <40 bpm or an escape rhythm that is below the AV node.

  • ° Awake, symptom-free patients with AF and bradycardia with =1 pauses of at least =5 s.

  • ° After catheter ablation of the AV junction associated with postoperative AV block that is not expected to resolve after cardiac surgery.

  • ° Associated with neuromuscular diseases with AV block.

- Second-degree AV block with associated symptomatic bradycardia regardless of type or site of block.

- Asymptomatic persistent third-degree AV block at any anatomic site with average awake ventricular rates of 40 bpm or faster if cardiomegaly or LV dysfunction is present or if the site of block is below the AV node.

- Second- or third-degree AV block during exercise in the absence of myocardial ischemia.

2. Chronic bifascicular block: Bifascicular block refers to ECG evidence of impaired conduction below the AV node in the right and left bundles.

  • - Advanced second-degree AV block or intermittent third-degree AV block.

  • - Type II second-degree AV block.

  • - Alternating bundle-branch block.

3. Pacing for atrioventricular block associated with acute myocardial infarction

  • - Persistent second-degree AV block after ST-segment elevation MI.

  • - Transient advanced second- or third-degree infranodal AV block and associated bundle-branch block.

  • - Persistent and symptomatic second- or third-degree AV block.

4. Hypersensitive carotid sinus syndrome and neurocardiogenic syncope. - Recurrent syncope caused by spontaneously occurring carotid sinus stimulation and carotid sinus pressure that induces ventricular asystole of more than 3 s.

5. After cardiac transplantation. - Persistent inappropriate or symptomatic bradycardia not expected to resolve.

6. Prevention and termination of arrhythmias by pacing.

  • - Symptomatic recurrent SVT that is reproducibly terminated by pacing when catheter ablation and/or drugs fail to control the arrhythmia or produce intolerable side effects.

  • - Sustained pause-dependent VT, with or without QT prolongation.

7. Hypertrophic cardiomyopathy.

8. Children, adolescents, and patients with congenital heart disease.

  • - Advanced second/third-degree AV block associated with symptomatic bradycardia, ventricular dysfunction, or low cardiac output.

  • - SND with correlation of symptoms during age-inappropriate bradycardia. (The definition of bradycardia varies with the patient's age and expected heart rate.)

  • - Postoperative advanced second- or third-degree AV block that is not expected to resolve or that persists at least 7 days after cardiac surgery.

  • - Congenital third-degree AV block with a wide QRS escape rhythm, complex ventricular ectopy, or ventricular dysfunction.

  • - Congenital third-degree AV block in the infant with a ventricular rate less than 55 bpm or with congenital heart disease and a ventricular rate less than 70 bpm.

Indications for implantable cardioverter defibrillators implantation

  • - Survivors of cardiac arrest due to VF/hemodynamically unstable sustained VT.

  • - Structural heart disease and spontaneous sustained VT, whether hemodynamically stable or unstable.

  • - LVEF (Left Ventricular Ejection Fraction) = 35% due to prior MI (NYHA functional Class II/III) or LVEF = 30% due to prior MI (NYHA functional Class I).

  • - Nonischemic dilated cardiomyopathy having an LVEF = 35% (NYHA functional Class II/III).

  • - Nonsustained VT due to prior MI, LVEF = 40%.

Indications for implantable cardioverter defibrillators implantation in pediatric patients and patients with congenital heart disease

- Survivor of cardiac arrest. - Symptomatic sustained VT.

Complications: Complications can be due to the presence of the pacing system as a foreign body (mechanical complications), apparent or real pacing system malfunction, acute complications related to the procedure itself: hemothorax, pneumothorax, subclavian artery puncture and myocardial perforation.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

1. Infection: incidence is 0.8-5.7%44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. Staphylococcus aureus (early infections) and S. epidermidis (late infections) are the most common organisms involved.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. - Pacemaker pocket infection: Routine use of antistaphylococcal antibiotic prophylaxis at the time of implantation or generator change has been shown to have consistent benefit in a meta-analysis of seven randomized controlled trials, in 2023 patients, in decreasing the rates of short-term pocket infection, skin erosion, or septicemia.55. Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31. Therefore routine prophylaxis with antistaphylococcal antibiotics is recommended.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. - Endocarditis: Complete eradication of the infection usually involves removal of the entire pacing system.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

2. Thrombosis: usually subclinical - Embolism: Pulmonary (incidence 0.6-3.5%44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. ) and systemic embolization. Patients implanted with a VVI(R) pacing system have a higher incidence of paroxysmal and chronic atrial fibrillation, which predisposes to embolic events.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

3. Pacing problems

Failure to capture: The occurrence of the stimulus output, which, when delivered outside the refractory period of atrial or ventricular tissue, is not followed by a P wave or QRS complex. This could be due to elevation of stimulation threshold, which in turn could be due to metabolic disturbances, drugs (mostly anti-arrythmics), acute myocardial infarction, pacing lead defects (fractures, insulation break), lead maturation, dislodgement or perforation, inappropriately low programmed output, battery end of life.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Failure of output: This is manifested on the surface electrocardiogram as an absence of pacing artifacts. This can be due to battery depletion or component failure44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

4. Sensing problems44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. : - Oversensing of unwanted signals: A pause in paced rhythm (AAI/VVI modes) or, if the oversensed atrial event causes triggered pacing in the ventricle, an earlier than expected paced ventricular event. Undersensing of the intrinsic intracardiac signal.

5. Rapid paced ventricular rates44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. : - Atrial tachyarrhythmias: Sinus tachycardia, atrial tachycardia, atrial flutter and ventricular pacing can resemble an "electronic AV block," and cardiac output can suddenly decrease. - Pacemaker-mediated tachycardia (PMT).

Pacemaker magnets

Magnet-operated reed switches within pacemakers were originally incorporated to produce pacemaker behavior that would demonstrate remaining battery life and sometimes pacing thresholds and can be used to protect the pacemaker-dependent patient during diathermy, electrocautery or other sources of pulsed EMI (electro magnetic interference).66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. Magnets can be applied over the pacemaker to avoid inhibition by such pulsed interference.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. In modern pacemakers, the switch to asynchronous pacing is coupled to the next cardiac event to avoid competition at the outset. Placement of a magnet over a generator might produce no change in pacing because not all pacemakers switch to a continuous asynchronous mode when a magnet is applied.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. In some devices magnet behavior can be altered by programming, whereas in others, magnet behavior can be completely eliminated by programming. For all generators, calling the manufacturer remains the most reliable method of determining magnet response and using this response to predict remaining battery life.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. However, the Task Force cautions against the use of the magnet over an ICD.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

As battery voltage falls, the magnet response can be used to detect the following:11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. - IFI (intensified follow-up required) - the device must be checked frequently (approximately every 4 weeks for most models). - ERI (elective replacement indicator) - the device is nearing the end of its useful life and should be electively replaced. - EOL (end of life) - the device has insufficient battery power remaining and should be replaced immediately.

Problems with magnet application: - Switching to asynchronous pacing may trigger ventricular asynchrony in patients with myocardial ischemia, hypoxia, and electrolyte imbalance.55. Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31. - Constant magnet application over the pacemaker may alter the programming and can also cause continuous or transient loss of pacing.77. Kleinman B, Hamilton J, Hariman R. Apparent failure of a precordial magnet and pacemaker programmer to convert a DDD pacemaker to VOO mode during the use of the electrosurgical unit. Anesthesiology. 1997;86:247-50. - Variability of response between devices.88. Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32. - Occasionally, PMT can ensue on removal of the magnet from a dual-chamber PM.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

Thus, in nonprogrammable pacemaker, the use of magnet may be safe. However, the most current devices should be considered programmable unless known otherwise.99. Chien WW, Foster E, Phillips B, et al. Pacemaker syndrome in a patient with DDD pacemaker for long QT syndrome. Pacing Clin Electrophysiol. 1991;14:1209-12.

Table 2 shows the commonly used pacing modes.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Table 2
Commonly used pacing modes.4

Implantable cardioverter defibrillator (ICD)

ICD consists of pulse generator, pacing or sensing electrodes, and defibrillation coils. Its function is similar to a pacemaker and hence susceptible to the same complications and emergencies as a pacemaker. In addition, an ICD senses and detects ventricular tachycardia (VT) and ventricular fibrillation (VF) and delivers therapy in the form of overdrive antitachycardia pacing (ATP), low-energy cardioversion, and high-energy defibrillation. Table 3 shows a four-place ICD code,22. Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25:260-4. given by the North American Society of Pacing and Electrophysiology (NASPE)/British Pacing and Electrophysiology Group (BPEG) (2002) Generic Defibrillator Code (NBD). The fourth position of the code is the three/five-letter code for the pacemaker capability of the device.22. Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25:260-4.

Table 3
NASPE/BPG Generic Defibrillator Code (NBD).(2)

ICD shock can lead to transient post-shock loss of capture and sensing in the pacemaker because of exposure of the myocardium to high current density. In patients with ICDs and separate pacemakers, the pacing stimulation artifact (PSA) can cause over-sensing or under-sensing in the ICD lead resulting in inappropriate ICD therapy. Hence, it is recommended that pacing at the chronic pacing amplitude in sinus rhythm should cause a PSA amplitude <1 mV on the ICD rate sense lead to ensure appropriate ICD sensing during VF.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and 1010. Brode SE, Schwartzman D, Callans DJ, et al. ICD-antiarrhythmic drug and ICD-pacemaker interactions. J Cardiovasc Electrophysiol. 1997;8:830-42.

It is imperative that ICD patients undergo routine evaluation (every three months and after each exposure to EMI).44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Pacemakers and special circumstances:

Magnetic resonance imaging (MRI)

It is estimated that up to 75% of pacemaker patients will have a medical need for an MRI over the lifetime of their device.1111. Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE. 2005;28:326-8. Scanning device patients is now considered a relative contraindication (vs absolute).1212. Levine GN, Gomes AS, Arai AE, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. 2007;116:2878-91. and 1313. Roguin A, Goldsher D. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008;10:336-46. Device manufacturers have made changes in devices to make them more compatible with MRI (less use of ferromagnetic material in battery construction).1414. Schwartzenburg CF, Wass CT, Strickland RA, et al. Rate-adaptive cardiac pacing: implications of environmental noise during craniotomy. Anesthesiology. 1997;87:1252-4. MRI affects pacemaker function in various ways:

Static magnetic field - can result in actual physical movement of the pulse generator's internal components.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Modulated radio frequency (RF) field - can result in induced voltage across the pacemaker electrodes that may stimulate myocardial tissue, leading to rates equal to the MRI pulsing rates (cycle lengths of 200-1000 ms),1515. Hayes DL, Holmes Jr DR, Gray JE. Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol. 1987;10:782-6. heating of cardiac tissue adjacent to lead electrodes causing thermal injury to myocardium and endocardium.1616. Sommer T, Vahlhaus C, Lauck G, et al. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology. 2000;215:869-79.

Gradient magnetic field - can cause over-sensing or under-sensing, and can induce negligible heating effect.1717. Gimbel JR, Johnson D, Levine PA, et al. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol. 1996;19:913-9.

While doing MRI - magnet response, rate response, noise response, ventricular sense response, premature ventricular contraction response, conducted atrial fibrillation response, and tachyarrhythmia functions - should be disabled.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. Pacemaker-dependent patients should be programmed to VOO mode and non-pacemaker-dependent patients can be programmed to the VVI or DDI mode.1818. Nazarian S, Roguin A, Zviman MM, et al. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation. 2006;114:1277-84.

Guideline recommendations with regard to performing MRIs in nonpacemaker dependent patients. 12 12. Levine GN, Gomes AS, Arai AE, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. 2007;116:2878-91. , 13 13. Roguin A, Goldsher D. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008;10:336-46. , 19 19. Kanal E, Barkovich AJ, Bell C. ACR guidance document for safe MR practices. Am J Roentgenol. 2007;188:1-27. , 20 20. Martin TE, Coman JA, Shellock FG, et al. Magnetic resonance imaging and cardiac pacemaker safety at 1.5- Tesla. J Am Coll Cardiol. 2004;43:1315-24. and 21 21. Loewy J, Loewy A, Kendall EJ. Reconsideration of pacemakers and MR imaging. Radiographics. 2004;24:1257-68.

  1. Decision should be weighed principally on risk-benefit ratio and the urgency of the clinical indication for MRI.

  2. Obtain a written and verbal informed consent.

  3. Before taking the patient for MRI, pacemaker functions should be pretested.

  4. A cardiologist should decide whether it is necessary to program the pacemaker prior to the MRI.

  5. Advanced Cardiac Life Support (ACLS) personnel must be in attendance for the entire MRI examination and a crash cart, and defibrillator must be available onsite.

  6. The patient should be monitored and visual/voice contact with the patient should be maintained continuously during the MRI.

  7. After the MRI examination, a cardiologist/electrophysiologist should interrogate the pacemaker to confirm that the function is consistent with the pre-examination state.

Exposure to MRI has similar effects on an ICD as those described for a cardiac pacemaker, since some of the basic components are comparable.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. ICDs may falsely detect the Magnetic Resonance Radio Frequency field as VF, charge capacitors, and deliver ATP, cardioversion, or defibrillation therapies.1616. Sommer T, Vahlhaus C, Lauck G, et al. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology. 2000;215:869-79. Devices may not be able to deliver ICD therapy in the static magnetic field and the ICD transformer will be magnetically saturated and may not have the voltage necessary to charge the capacitor, and hence can lead to permanent device failure. Magnetic fields may also prevent detection of VT or VF.

Electrocautery and anesthesia

Electrocautery is the most common exogenous source of EMI66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. that can interact with pacemakers, resulting in pulse generator inhibition, electrical burns at the myocardial-electrode interface, atrial or ventricular tachycardia and fibrillation, pulse generator component failure, loss of or change in output, reprogramming of rate or mode of function and runaway pacing.2222. Domino KB, Smith TC. Electrocautery-induced reprogramming of a pacemaker using a precordial magnet. Anesth Analg. 1983;62:609-12. , 2323. Belott PH, Sands S, Warren J. Resetting of DDD pacemakers due to EMI. Pacing Clin Electrophysiol. 1984;7:169-72. and 2424. Van Hemel NM, Hamerlijnck RP, Pronk KJ, et al. Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery. Pacing Clin Electrophysiol. 1989;12:1720-3. The EMI generated by electrocautery that may affect the device is related to the distance and orientation of the current to the patient's device and leads. A prospective study has shown that unipolar devices are far more susceptible than bipolar devices to electrocautery inference.2424. Van Hemel NM, Hamerlijnck RP, Pronk KJ, et al. Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery. Pacing Clin Electrophysiol. 1989;12:1720-3. In bipolar coagulation cautery, the current flow is localized between the two poles of the instrument, and therefore poses minimal problems. However, in unipolar electrocautery devices, the electrical current flow is not restricted to the tissue interposed between two electrodes and spreads throughout the body.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. The vector of the dipole for the electrocautery device with respect to that of the pacemaker should not intersect with each other.2525. Chauvin M, Crenner F, Brechenmacher C. Interaction between permanent cardiac pacing and electrocautery: the significance of electrode position. Pacing Clin Electrophysiol. 1992;15:2028-33. Hence, the EMI generated by electrocautery that may affect the device is related to the distance and orientation of the current to the patient's device and leads.2626. Levine PA, Balady GJ, Lazar HL, et al. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. 1986;41:313-7. Special precautions are2626. Levine PA, Balady GJ, Lazar HL, et al. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. 1986;41:313-7. and 2727. Erdman S, Levinsky L, Servadio C, et al. Safety precautions in the management of patients with pacemakers when electrocautery operations are performed. Surg Gynecol Obstet. 1988;167:311-4. :

  1. Using a bipolar electrocautery system/ultrasonic (harmonic) scalpel, if possible.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

  2. If only unipolar electrocautery is available, then the indifferent electrode should be placed as far from the pacemaker leads as possible.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

  3. Assure that the cautery tool and current return pad are positioned so that the current pathway does not pass through or near the CRMD (Cardiac Rhythm Management Devices) pulse generator and leads.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

  4. Using short, intermittent, and irregular bursts at the lowest feasible energy levels55. Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31. to minimize the hemodynamic effects of pacemaker inhibition.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

  5. A magnet should not be applied prophylactically, because there is no uniform pulse generator response to this maneuver.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

  6. Electrocautery can cause electromagnetic noise on the ICD sensing lead that is detected as VF and can lead to inappropriate shock.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

  7. Therefore the detection and therapy can be programmed to "off" during surgery and turn it "on" postoperatively; external defibrillators are used for intra-operative VT/VF.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

  8. The availability of programming equipment and trained personnel is essential.

  9. Alternative temporary pacing should be ready in the OT.2626. Levine PA, Balady GJ, Lazar HL, et al. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. 1986;41:313-7.

  10. Drugs such as isoproterenol and atropine should be available.88. Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32.

  11. Careful monitoring of the patient.

  12. The device should always be checked after operation.

Diathermy

Short-wave diathermy should be avoided near the generator site. Potential problems include overheating of the generator circuitry and damage to electronic components.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Electroconvulsive therapy (ECT)

ECT is relatively safe for patients with pacemakers, because of the localized application of the electrical stimulus to the head, hence a low probability for the occurrence of problems.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. Sometimes the seizure may generate myopotentials which may inhibit the pacemaker, and transient electrocardiographic changes (e.g., increased P-wave amplitude, altered QRS shape, T-wave and ST-T abnormalities) may occur and additional cardiac complications (e.g., arrhythmia or ischemia) may occur in patients with pre-existing cardiac disease. The Task Force believes that ECT may be administered to CRMD patients without significant damage to a disabled CRMD. All CRMDs should undergo a comprehensive interrogation before the procedure(s). ICD functions should be disabled for shock therapy during ECT.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. Therefore, monitoring and standard resuscitation equipment, and a trained programmer, should be available2828. Alexopoulos GS, Frances RJ. ECT and cardiac patients with pacemakers. Am J Psychiatry. 1980;137:1111-2. and pacemakers can be changed to nonsensing asynchronous mode (fixed mode),2929. Rozner MA. Intrathoracic gadgets: update on pacemakers and implantable cardioverter-defibrillators. ASA Refresher Course; 1999. p. 212. to avoid myopotential inhibition of the device in pacemaker-dependent patients.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. CRMD-dependent patients may require a temporary pacing system to preserve cardiac rate and rhythm during shock therapy.

Other electrical stimulating techniques

Electrical stimulating techniques such as transcutaneous electrical nerve stimulation (TENS) consists of several electrodes placed on the skin and connected to a pulse generator that applies 20 µs rectangular pulses of 1-200 V and 0-60 mA at a frequency of 20-110 Hz. This repeated frequency is similar to the normal range of heart rates, so it can create a far field potential that may inhibit a cardiac pacemaker.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. TENS can be used safely in patients with pacemakers and defibrillators with close monitoring and use in close proximity to the device is not advised.3030. Salukhe TV, Dob D, Sutton R. Pacemakers and defibrillators: anaesthetic implications. Br J Anaesth. 2004;93:95-104.

PMT has been reported to be induced by intraoperative somatosensory evoked potential stimulation.3131. Philbin DM, Marieb MA, Aithal KH, et al. Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. 1998;21:2010-1.

Radiofrequency ablation (RFA)

The radiofrequency current path (electrode tip to current return pad) should be kept as far away from the pulse generator and lead system as possible and to avoid direct contact between the ablation catheter and the CRMD (Cardiac Rhythm Management Devices).66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Radiation therapy

The high-energy ionizing radiation used in radiation therapy can cause significant damage to the semiconductors of pacemakers, even at very small doses.3131. Philbin DM, Marieb MA, Aithal KH, et al. Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. 1998;21:2010-1. Generally, doses in excess of 5000 rads are required to cause pacemaker malfunction but as little as 1000 rads may induce pacemaker failure or cause runaway pacemaker.88. Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32. Pulse generator recovery may occur long after the end of the radiation treatment, but it is mostly incomplete, and the pacemaker cannot be used reliably thereafter.3232. Raitt MH, Stelzer KJ, Laramore GE, et al. Runaway pacemaker during high-energy neutron radiation therapy. Chest. 1994;106:955-7. and 3333. Souliman SK, Christie J. Pacemaker failure induced by radiotherapy. Pacing Clin Electrophysiol. 1994;17:270-3. Thus, in pulse generators exposed to radiation, transient loss of function should be regarded as a precursor of permanent damage. Hence it is essential to follow guidelines for ensuring the lowest possible radiation dose to the pacemaker3434. Muller- Runkel R, Orsolini G, Kalokhe UP. Monitoring the radiation dose to a multiprogrammable pacemaker during radical radiation therapy: a case report. Pacing Clin Electrophysiol. 1990;13:1466-70. and careful follow-up should be performed during and after completion of the radiation therapy. The Task Force believes that radiation therapy can be safely performed for CRMD patients. The device must be outside the field of radiation.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. Therefore, some pulse generators will require surgical relocation before commencing radiation.

The low-energy X-rays used for diagnostic radiology have not been reported to have any adverse effect in pacemakers.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Other minor surgical procedures (lithotripsy, endoscopic electrocautery)

Extracorporeal shock wave lithotripsy (ESWL) is associated with electromagnetic and mechanical forces that may influence pacing system function.3535. Langberg J, Abber J, Thuroff JW, et al. The effects of extracorporeal shock wave lithotripsy on pacemaker function. Pacing Clin Electrophysiol. 1987;10:1142-6. Pulse generators employing a piezoelectric crystal sensor seem to be the most susceptible to failure,3636. Cooper D, Wilkoff B, Masterson M, et al. Effects of extracorporeal shock wave lithotripsy on cardiac pacemakers and its safety in patients with implanted cardiac pacemakers. Pacing Clin Electrophysiol. 1988;11:1386-7. and therefore the sensor mode should be programmed "off" during lithotripsy as this will prevent an unwanted increase in paced heart rate and shatter injury to the piezoelectric element. Therefore, focal point of the lithotriptor should be kept at least six inches (15 cm) away from the pacemaker,3737. Albers DD, Lybrand FE, Axton JC, et al. Shockwave lithotripsy and pacemakers: experience with 20 cases. J Endourol. 1995;9:301-3. and disabling atrial pacing if the lithotripsy system triggers on the R wave.77. Kleinman B, Hamilton J, Hariman R. Apparent failure of a precordial magnet and pacemaker programmer to convert a DDD pacemaker to VOO mode during the use of the electrosurgical unit. Anesthesiology. 1997;86:247-50. Low shock waves (<16 kV) should be used initially followed by a gradual increase in the level of energy.3838. Ganem JP, Carson CC. Cardiac arrhythmias with external fixed rate signal generators in shock wave lithotripsy with the Medstone lithotripter. Urology. 1998;51:548-52. Endoscopic electrocautery is generally safe in patients with pacemakers, although complications have been reported.3939. Ito S, Shibata H, Okahisa T, et al. Endoscopic therapy using monopolar and bipolar snare with a high-frequency current in patients with a pacemaker. Endoscopy. 1994;26:270.

Cardiopulmonary resuscitation

The standard resuscitation protocols should be followed in patients with permanent pacemakers, and normal pacemaker function should be established after the resuscitation procedure is completed.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. Myocardial stimulation threshold is markedly increased during cardiopulmonary resuscitation44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and even though most sensing and pacing problems are transient, a thorough evaluation of the pacing system, including interrogation and programming functions, should be made after the resuscitation procedure.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.

Direct current cardioversion and defibrillation

The cardioverter-defibrillator can cause capacitive coupling with the endocardial lead, causing direct discharge at the electrode-endocardium interface,44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and 4040. Aylward P, Blood R, Tonkin A. Complications of defibrillation with permanent pacemaker in situ. Pacing Clin Electrophysiol. 1979;2:462-4. thus leading to transient/permanent failure to sense and capture even in the absence of apparent damage to the pulse generator itself.4141. Waller C, Callies F, Langenfeld H. Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes: is external cardioversion contraindicated in patients with permanent pacing systems?. Europace. 2004;6: 165-8. Damaged circuitry,4242. Das G, Eaton J. Pacemaker malfunction following transthoracic countershock. Pacing Clin Electrophysiol. 1981;4: 487-90. and 4343. Levine PA, Barold SS, Fletcher RD, et al. Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems. J Am Coll Cardiol. 1983;1:1413-22. changes in programmed mode of function, complete pacemaker failure and microdislodgement of the lead can also occur.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and 4444. Gould L, Patel S, Gomes GI, et al. Pacemaker failure following external defibrillation. Pacing Clin Electrophysiol. 1981;4:575-7. Defibrillation with internal cardiac paddles requires less energy but may also interfere with pacemaker function.4545. Calkins H, Brinker J, Veltri EP, et al. Clinical interactions between pacemaker and automated implantable cardioverter defibrillators. J Am Coll Cardiol. 1990;16:666-73. and 4646. Yee R, Jones DL, Klein GJ. Pacing threshold changes after transvenous catheter countershock. Am J Cardiol. 1984;53:503-7. Modern pacemakers are equipped with protection mechanisms against damage from DC shock, most common of which is the Zener diode,4747. Lau FY, Bilitch M, Wintroub HJ. Protection of implanted pacemakers from excessive electrical energy of DC shock. Am J Cardiol. 1969;23:244-9. which directs a surge in current toward the electrode, protecting the pacemaker circuitry but delivering this energy to the endocardium. The general precautions for DC cardioversion and defibrillation in patients with pacemakers attempt to minimize the current delivered to the pacemaker system by using the minimal effective energy setting44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and 66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. and placing the defibrillator paddles at least 10 cm away from the pulse generator, ensuring that the paddles are placed perpendicular to the dipole of the pacing system. A thorough evaluation of the pacing system should be performed44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and if loss of capture occurs, and then immediate reprogramming or temporary pacing should be done with increased generator output.4848. Mehta Y, Swaminathan M, Juneja R, et al. Noncardiac surgery and pacemaker cardioverter-defibrillator management. J Cardiothorac Vasc Anesth. 1998;12:221-4. and 4949. Mangar D, Atlas GM, Kane PB. Electrocautery-induced pacemaker malfunction during surgery. Can J Anaesth. 1991;38:616-8. The Task Force believes that before attempting emergency defibrillation or cardioversion of a patient with an ICD and magnet-disabled therapies, all sources of EMI should be terminated, and the magnet should be removed to re-enable antitachycardia therapies and then consider reenabling therapies through programming.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Bioelectrical impedance analysis (BIA)

BIA employs bioimpedence spectroscopy techniques which measures at 50 frequencies over a range from 5 to 1000 kHz to determine the electrical resistances of TBW and ECW to provide an estimate of body cell mass and to describe fluid shifts and fluid balance,5050. Heymsfield SB, Wang Z, Visser M, et al. Techniques used in the measurement of body composition: an overview with emphasis on bioelectrical impedance analysis. Am J Clin Nutr. 1996;64:478S-84S. but not recommended for participants with a pacemaker.5151. Leeab SY, Gallagherc D. Assessment methods in human body composition. Curr Opin Clin Nutr Metab Care. 2008;11: 566-72.

Bioimpedance tomography/thoracic electrical bioimpedance (TEB)

This uses pulsating and polarized electric current at a frequency not higher than 10 Hz, and provides information on the impedance of the extracellular environment.5252. Rienzo VD, Minelli M, Sambugaro R. Applicability of extra- cellular electrical impedance tomography in monitoring respiratory tract inflammation. J Investig Allergol Clin Immunol. 2007;17:34-8. However their use is not recommended for patients with minute ventilation (MV) sensor function pacemakers, since sensors are very sensitive to stray electromagnetic interference, and patients have been inappropriately treated for pacemaker-driven tachycardias as a result. Therefore, rate modulation should be programmed to "OFF" in the peri-operative period to prevent confusion between an intrinsic tachycardia vs a pacemaker-induced.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. and 1414. Schwartzenburg CF, Wass CT, Strickland RA, et al. Rate-adaptive cardiac pacing: implications of environmental noise during craniotomy. Anesthesiology. 1997;87:1252-4.

Metabolic abnormalities

Electrolyte and metabolic abnormalities (hyperkalemia, hyperglycemia, alkalemia, acidosis, hypoxemia, hypercapnia and hypothyroidism) increase the pacing threshold, causing failure to capture.5353. Dohrmann ML, Goldschlager NF. Myocardial stimulation threshold in patients with cardiac pacemakers: effect of physiologic variables, pharmacologic agents, and lead electrodes. Cardiol Clin. 1985;3:527-37. Therefore, aggressive correction of the underlying disturbance is essential.

Systolic heart failure

The presence of a permanent pacemaker has been reported to be an independent predictor of poor outcome in patients with heart failure,44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. and 5454. Saxon LA, Stevenson WG, Middlekauff HR, et al. Increased risk of progressive hemodynamic deterioration in advanced heart failure patients requiring permanent pacemakers. Am Heart J. 1993;125:1306-10. and hence evidence of pacemaker malfunction and must be carefully evaluated.

Pre anesthetic consideration

Preoperative evaluation of a patient with pacemaker undergoing noncardiac surgery includes general evaluation of the patient and the pacemaker.

The consensus of the Task Force66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. is that a focused preoperative evaluation should include:

  1. Establishing whether a patient has a CRMD - a focused history and physical examination

  2. Defining the type of device - obtaining the manufacturer's identification card from the patient or other source, chest X-ray (if no other data are available)

  3. Determining whether a patient is CRMD dependent for pacemaking function.

  4. Determining CRMD function by a comprehensive evaluation of the device, consultation with a cardiologist or a CRMD service, contacting the manufacturer for peri-operative recommendations.

ACC guidelines suggest that cardiac testing (stress tests, echocardiograms) be dictated by the patient's underlying disease, medications, symptomatology, interval from last testing, and planned intervention.5555. Fleisher LA, Beckman JA, Brown KA, et al. ACC/AHA 2007 guidelines on perioperative cardiovascular evaluation and care for noncardiac surgery. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines. Ci.

No special laboratory tests or radiographs are needed for patients with a conventional PM. However, a patient with a BiV PM or ICD might need a chest film to document the position of the coronary sinus lead, especially if central line placement is planned. Most current CRMDs have an X-ray code that can be used to identify the manufacturer of the device.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

The generator should be identified, and the location of the pulse generator should be noted. Generally, generator for the epicardial electrodes is kept in the abdomen and over one of the pectoris muscles for the endocardial electrodes.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. If interrogation is not an option, one can slow the intrinsic heart rate to a rate below that of the pacemaker by carotid massage or a Vasalva maneuver.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. and 5656. Simon AB. Perioperative management of the pacemaker patient. Anesthesiology. 1977;46:127-31; Simon AB. Perioperative management of the pacemaker patient. Circulation. 2007;116:1971-6. Any device near its elective replacement period should be considered for replacement. Ten percent decrease in the rate from the time of implantation indicates power source depletion.88. Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32. The patient's underlying rate and rhythm should be determined (which then determines the need for backup/external pacing support). All rate enhancements and minute ventilation rate responsiveness should be programmed to "off" mode, and ICD's anti-tachyarrhythmia functions should be suspended, if present. For ICD patients who depend on pacing function for control of bradyarrhythmia, these functions should be altered by programming.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

The lower rate limit can be increase to optimize oxygen delivery to tissues for major cases.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].

A determination as to whether EMI is likely to occur or not. Anesthetic techniques do not influence CRMD function. However, anesthetic-induced physiologic changes (i.e., cardiac rate, rhythm, or ischemia) in the patient may induce unexpected CRMD responses or adversely affect the CRMD-patient interaction.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Appropriate reprogramming is the safest way to avoid intraoperative problems, and is required in any rate responsive device, hypertrophic obstructive/dilated cardiomyopathy, pediatric patients, pacemaker-dependent patients, major procedure in the chest/abdomen, special procedures (lithotripsy, TURP, hysteroscopy, ECT, scoline use, MRI).11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. Temporary pacing and defibrillation equipment should be immediately available before, during, and after a procedure.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Intraoperative considerations

General anesthesia can present a range of problems for the paced patients, although it rarely does so due to contemporary anesthesia techniques. The Task Force agrees that a patient's electrocardiogram (ECG) should be continuously displayed and continuous peripheral pulse monitoring should be performed for all CRMD patients receiving general or regional anesthesia, sedation, or monitored anesthesia care. Mechanical systole (which depicts paced electrical activity) is best evaluated by pulse oximetry plethysmography, or arterial pressure waveform display.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. The presence of pacemaker is not an indication for insertion of pulmonary artery (PA) or central venous catheter (CVC).5757. Zaidan JR, Pacemakers. Anesthesiology. 1984;60:319-34. Insertion of the guide wire or CVC is potentially arrhythmogenic55. Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31. and dislodgement of freshly placed transvenous endocardial electrode can occur. Therefore care should be taken and multipurpose PA catheter with pacing facilities can be used.5858. Kemnitz J, Peters J. Cardiac pacemakers and implantable cardioverter defibrillators in the perioperative phase. Anasthesiol Intensivmed Notfallmed Schmerzther. 1993;28:199-212. Emergency drugs should be kept ready, including temporary pacing and defibrillation.

Anesthetic induction with succinylcholine can cause significant muscle fasciculations leading to complete inhibition of pacemakers because of oversensing, resulting in cardiac arrest in pacemaker-dependent patients.5959. Finfer SR. Pacemaker failure on induction of anaesthesia. Br J Anaesth. 1991;66:509-12. The muscle fasciculation induced by succinylcholine can be avoided by using non-depolarizing muscle relaxant or defasiculating with nondepolarizing muscle relaxant before giving succinylcholine.88. Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32. This complication can also be prevented by temporary reprogramming of the pacemaker to an asynchronous mode.44. Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34. Narcotics and inhalational techniques do not alter current and voltage thresholds of the pacemaker,6060. Atlee JL, Bernstein AD. Cardiac rhythm management devices, perioperative management. Anesthesiology. 2001;95:1492-506. and hence can be used safely. Use of N2O can cause pacemaker malfunction by increasing gas in pre pectoral pacemaker pocket (loss of anodal contact).6161. Lamas GA, Rebecca GS, Braunwald NS, et al. Pacemaker malfunction after nitrous oxide anesthesia. Am J Cardiol. 1985;56:995-6. Etomidate and ketamine should be avoided as these cause myoclonic movements.55. Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31. Positive pressure ventilation can cause dislodgement of pacemaker leads; therefore, pacemaker function should be verified, before and after initiating mechanical ventilation.6262. Thiagarajah S, Azar I, Agres M, et al. Pacemaker malfunction associated with positive pressure ventilation. Anesthesiology. 1983;58:565-6.

Many metabolic and electrolyte abnormalities can affect pacemaker function and therefore, monitoring for such events is essential. Shivering should be avoided and temperature must be kept constant in 'temperature' rate responsive pacemakers.99. Chien WW, Foster E, Phillips B, et al. Pacemaker syndrome in a patient with DDD pacemaker for long QT syndrome. Pacing Clin Electrophysiol. 1991;14:1209-12.

If unanticipated device interactions are found, consider discontinuation of the procedure until the source of interference can be eliminated or managed.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98. If a temporary pacemaker fails intraoperatively, the inspired oxygen concentration should be increased to 100%. All connections and the generator battery should be checked. The generator should be set into the asynchronous mode, and the ventricular output should be set on maximum. Pharmacological management (atropine, isoproterenol, or epinephrine) may be useful until the problem is resolved. Failure of a temporary transvenous electrode to capture the ventricle is usually due to displacement of the electrode away from the ventricular endocardium; careful slow advancement of the catheter or wire while pacing often results in capture. If an adequate arterial blood pressure cannot be maintained with adrenergic agonists, cardiopulmonary resuscitation should be instituted until another pacing electrode is placed or a new generator box is obtained.11. Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43]. For the patient with an implanted defibrillator, facilities for external defibrillation should be available immediately after the device is disabled.3030. Salukhe TV, Dob D, Sutton R. Pacemakers and defibrillators: anaesthetic implications. Br J Anaesth. 2004;93:95-104. If patient develops VT, surgeons should be advised to terminate all sources of EMI, magnet should be removed to enable anti tachycardia activities, device should be observed and if failed, and then external defibrillation/cardioversion should be initiated.

Electrical interference from surgical electrocautery units, electroconvulsive therapy, lithotripsy has already been discussed.

Post-operative considerations

Monitor cardiac rate and rhythm continuously throughout the immediate postoperative period.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

Emergency drugs and equipments should be kept ready.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

The device should be reprogrammed to appropriate settings. For an ICD, all antitachyarrhythmic therapies should be restored. Consultation with a cardiologist or pacemaker-ICD service may be necessary.66. Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.

References

  • 1
    Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. In: Miller RD, editor. Miller's anesthesia. 7th ed. USA: Churchill Livingstone; 2009. p. 1388-402 [chapter 43].
  • 2
    Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25:260-4.
  • 3
    ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. J Am Coll Cardiol. 2008;51:1-62.
  • 4
    Banker R, Mitchell R, Badhwar N, et al. Pacemakers and implantable cardioverter-defibrillator emergencies. In: Jeremiah A, Brown DL, editors. Cardiac intensive care. 2nd ed. Philadelphia, PA, USA: Saunders; 2010. p. 310-34.
  • 5
    Sethuran S, Toff WD, Vuylsteke A, et al. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. 2002;88:627-31.
  • 6
    Article Task Force. Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Anesthesiology. 2005;103:186-98.
  • 7
    Kleinman B, Hamilton J, Hariman R. Apparent failure of a precordial magnet and pacemaker programmer to convert a DDD pacemaker to VOO mode during the use of the electrosurgical unit. Anesthesiology. 1997;86:247-50.
  • 8
    Rastogi S, Goel S, Tempe DK, et al. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. 2005;8:21-32.
  • 9
    Chien WW, Foster E, Phillips B, et al. Pacemaker syndrome in a patient with DDD pacemaker for long QT syndrome. Pacing Clin Electrophysiol. 1991;14:1209-12.
  • 10
    Brode SE, Schwartzman D, Callans DJ, et al. ICD-antiarrhythmic drug and ICD-pacemaker interactions. J Cardiovasc Electrophysiol. 1997;8:830-42.
  • 11
    Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE. 2005;28:326-8.
  • 12
    Levine GN, Gomes AS, Arai AE, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. 2007;116:2878-91.
  • 13
    Roguin A, Goldsher D. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008;10:336-46.
  • 14
    Schwartzenburg CF, Wass CT, Strickland RA, et al. Rate-adaptive cardiac pacing: implications of environmental noise during craniotomy. Anesthesiology. 1997;87:1252-4.
  • 15
    Hayes DL, Holmes Jr DR, Gray JE. Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol. 1987;10:782-6.
  • 16
    Sommer T, Vahlhaus C, Lauck G, et al. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology. 2000;215:869-79.
  • 17
    Gimbel JR, Johnson D, Levine PA, et al. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol. 1996;19:913-9.
  • 18
    Nazarian S, Roguin A, Zviman MM, et al. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla. Circulation. 2006;114:1277-84.
  • 19
    Kanal E, Barkovich AJ, Bell C. ACR guidance document for safe MR practices. Am J Roentgenol. 2007;188:1-27.
  • 20
    Martin TE, Coman JA, Shellock FG, et al. Magnetic resonance imaging and cardiac pacemaker safety at 1.5- Tesla. J Am Coll Cardiol. 2004;43:1315-24.
  • 21
    Loewy J, Loewy A, Kendall EJ. Reconsideration of pacemakers and MR imaging. Radiographics. 2004;24:1257-68.
  • 22
    Domino KB, Smith TC. Electrocautery-induced reprogramming of a pacemaker using a precordial magnet. Anesth Analg. 1983;62:609-12.
  • 23
    Belott PH, Sands S, Warren J. Resetting of DDD pacemakers due to EMI. Pacing Clin Electrophysiol. 1984;7:169-72.
  • 24
    Van Hemel NM, Hamerlijnck RP, Pronk KJ, et al. Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery. Pacing Clin Electrophysiol. 1989;12:1720-3.
  • 25
    Chauvin M, Crenner F, Brechenmacher C. Interaction between permanent cardiac pacing and electrocautery: the significance of electrode position. Pacing Clin Electrophysiol. 1992;15:2028-33.
  • 26
    Levine PA, Balady GJ, Lazar HL, et al. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. 1986;41:313-7.
  • 27
    Erdman S, Levinsky L, Servadio C, et al. Safety precautions in the management of patients with pacemakers when electrocautery operations are performed. Surg Gynecol Obstet. 1988;167:311-4.
  • 28
    Alexopoulos GS, Frances RJ. ECT and cardiac patients with pacemakers. Am J Psychiatry. 1980;137:1111-2.
  • 29
    Rozner MA. Intrathoracic gadgets: update on pacemakers and implantable cardioverter-defibrillators. ASA Refresher Course; 1999. p. 212.
  • 30
    Salukhe TV, Dob D, Sutton R. Pacemakers and defibrillators: anaesthetic implications. Br J Anaesth. 2004;93:95-104.
  • 31
    Philbin DM, Marieb MA, Aithal KH, et al. Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. 1998;21:2010-1.
  • 32
    Raitt MH, Stelzer KJ, Laramore GE, et al. Runaway pacemaker during high-energy neutron radiation therapy. Chest. 1994;106:955-7.
  • 33
    Souliman SK, Christie J. Pacemaker failure induced by radiotherapy. Pacing Clin Electrophysiol. 1994;17:270-3.
  • 34
    Muller- Runkel R, Orsolini G, Kalokhe UP. Monitoring the radiation dose to a multiprogrammable pacemaker during radical radiation therapy: a case report. Pacing Clin Electrophysiol. 1990;13:1466-70.
  • 35
    Langberg J, Abber J, Thuroff JW, et al. The effects of extracorporeal shock wave lithotripsy on pacemaker function. Pacing Clin Electrophysiol. 1987;10:1142-6.
  • 36
    Cooper D, Wilkoff B, Masterson M, et al. Effects of extracorporeal shock wave lithotripsy on cardiac pacemakers and its safety in patients with implanted cardiac pacemakers. Pacing Clin Electrophysiol. 1988;11:1386-7.
  • 37
    Albers DD, Lybrand FE, Axton JC, et al. Shockwave lithotripsy and pacemakers: experience with 20 cases. J Endourol. 1995;9:301-3.
  • 38
    Ganem JP, Carson CC. Cardiac arrhythmias with external fixed rate signal generators in shock wave lithotripsy with the Medstone lithotripter. Urology. 1998;51:548-52.
  • 39
    Ito S, Shibata H, Okahisa T, et al. Endoscopic therapy using monopolar and bipolar snare with a high-frequency current in patients with a pacemaker. Endoscopy. 1994;26:270.
  • 40
    Aylward P, Blood R, Tonkin A. Complications of defibrillation with permanent pacemaker in situ. Pacing Clin Electrophysiol. 1979;2:462-4.
  • 41
    Waller C, Callies F, Langenfeld H. Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes: is external cardioversion contraindicated in patients with permanent pacing systems?. Europace. 2004;6: 165-8.
  • 42
    Das G, Eaton J. Pacemaker malfunction following transthoracic countershock. Pacing Clin Electrophysiol. 1981;4: 487-90.
  • 43
    Levine PA, Barold SS, Fletcher RD, et al. Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems. J Am Coll Cardiol. 1983;1:1413-22.
  • 44
    Gould L, Patel S, Gomes GI, et al. Pacemaker failure following external defibrillation. Pacing Clin Electrophysiol. 1981;4:575-7.
  • 45
    Calkins H, Brinker J, Veltri EP, et al. Clinical interactions between pacemaker and automated implantable cardioverter defibrillators. J Am Coll Cardiol. 1990;16:666-73.
  • 46
    Yee R, Jones DL, Klein GJ. Pacing threshold changes after transvenous catheter countershock. Am J Cardiol. 1984;53:503-7.
  • 47
    Lau FY, Bilitch M, Wintroub HJ. Protection of implanted pacemakers from excessive electrical energy of DC shock. Am J Cardiol. 1969;23:244-9.
  • 48
    Mehta Y, Swaminathan M, Juneja R, et al. Noncardiac surgery and pacemaker cardioverter-defibrillator management. J Cardiothorac Vasc Anesth. 1998;12:221-4.
  • 49
    Mangar D, Atlas GM, Kane PB. Electrocautery-induced pacemaker malfunction during surgery. Can J Anaesth. 1991;38:616-8.
  • 50
    Heymsfield SB, Wang Z, Visser M, et al. Techniques used in the measurement of body composition: an overview with emphasis on bioelectrical impedance analysis. Am J Clin Nutr. 1996;64:478S-84S.
  • 51
    Leeab SY, Gallagherc D. Assessment methods in human body composition. Curr Opin Clin Nutr Metab Care. 2008;11: 566-72.
  • 52
    Rienzo VD, Minelli M, Sambugaro R. Applicability of extra- cellular electrical impedance tomography in monitoring respiratory tract inflammation. J Investig Allergol Clin Immunol. 2007;17:34-8.
  • 53
    Dohrmann ML, Goldschlager NF. Myocardial stimulation threshold in patients with cardiac pacemakers: effect of physiologic variables, pharmacologic agents, and lead electrodes. Cardiol Clin. 1985;3:527-37.
  • 54
    Saxon LA, Stevenson WG, Middlekauff HR, et al. Increased risk of progressive hemodynamic deterioration in advanced heart failure patients requiring permanent pacemakers. Am Heart J. 1993;125:1306-10.
  • 55
    Fleisher LA, Beckman JA, Brown KA, et al. ACC/AHA 2007 guidelines on perioperative cardiovascular evaluation and care for noncardiac surgery. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines. Ci.
  • 56
    Simon AB. Perioperative management of the pacemaker patient. Anesthesiology. 1977;46:127-31; Simon AB. Perioperative management of the pacemaker patient. Circulation. 2007;116:1971-6.
  • 57
    Zaidan JR, Pacemakers. Anesthesiology. 1984;60:319-34.
  • 58
    Kemnitz J, Peters J. Cardiac pacemakers and implantable cardioverter defibrillators in the perioperative phase. Anasthesiol Intensivmed Notfallmed Schmerzther. 1993;28:199-212.
  • 59
    Finfer SR. Pacemaker failure on induction of anaesthesia. Br J Anaesth. 1991;66:509-12.
  • 60
    Atlee JL, Bernstein AD. Cardiac rhythm management devices, perioperative management. Anesthesiology. 2001;95:1492-506.
  • 61
    Lamas GA, Rebecca GS, Braunwald NS, et al. Pacemaker malfunction after nitrous oxide anesthesia. Am J Cardiol. 1985;56:995-6.
  • 62
    Thiagarajah S, Azar I, Agres M, et al. Pacemaker malfunction associated with positive pressure ventilation. Anesthesiology. 1983;58:565-6.

Publication Dates

  • Publication in this collection
    May-Jun 2014

History

  • Received
    21 Aug 2012
  • Accepted
    28 Feb 2013
Sociedade Brasileira de Anestesiologia R. Professor Alfredo Gomes, 36, 22251-080 Botafogo RJ Brasil, Tel: +55 21 2537-8100, Fax: +55 21 2537-8188 - Campinas - SP - Brazil
E-mail: bjan@sbahq.org