Acessibilidade / Reportar erro

Evaluation of the scattered radiation components produced in a gamma camera using Monte Carlo method

INTRODUCTION: This paper presents a simulation for evaluation of the scattered radiation components produced in a gamma camera PARK using Monte Carlo code SIMIND. It simulates a whole body study with MDP (Methylene Diphosphonate) radiopharmaceutical based on Zubal anthropomorphic phantom, with some spinal lesions. METHODS: The simulation was done by comparing 3 configurations for the detected photons. The corresponding energy spectra were obtained using Low Energy High Resolution collimator. The parameters related with the interactions and the fraction of events in the energy window, the simulated events of the spectrum and scatter events were calculated. RESULTS: The simulation confirmed that the images without influence of scattering events have a higher number of valid recorded events and it improved the statistical quality of them. A comparison among different collimators was made. The parameters and detector energy spectrum were calculated for each simulation configuration with these collimators using 99mTc. CONCLUSION: The simulation corroborated that LEHS collimator has higher sensitivity and HEHR collimator has lower sensitivity when they are used with low energy photons.

Collimators; Monte Carlo; Gamma camera; SIMIND; Zubal phantom


SBEB - Sociedade Brasileira de Engenharia Biomédica Sociedade Brasileira de Engenharia Biomédica, Centro de Tecnologia, , bloco H, sala 327 - Cidade Universitária, 21941-914 Rio de Janeiro - RJ , Tel./Fax.:: (55 21) 2562-8591 - Rio de Janeiro - RJ - Brazil
E-mail: rbeb@rbeb.org.br