Acessibilidade / Reportar erro

Knee joint range-of-motion evaluation: correlation between measurements achieved using a universal goniometer and an isokinetic dynamometer

BACKGROUND: The instrument most often used by therapists for measuring joint range of motion (ROM) is the universal goniometer. However, there is a lack of studies analyzing the reliability of knee joint ROM measured by the isokinetic dynamometer. OBJECTIVE: The purpose of this study was to analyze the correlation between the knee joint ROM measurements made using a universal goniometer and an isokinetic dynamometer. METHOD: 38 healthy volunteers (27 women, 11 men) aged 36 ± 11 years were evaluated. All of them had a minimum knee extension ROM limitation of 20°. Three knee ROM measurements were made using the universal goniometer and another three using the isokinetic dynamometer, on each subject's dominant limb. RESULTS: The results showed a high degree of correlation between the knee ROM measurements made using the two instruments (Pearson correlation coefficient = 0.90). CONCLUSION: From the procedures performed, both the universal goniometer and the isokinetic dynamometer can be used to evaluate knee ROM, since they both present reliable measurements.

goniometer; isokinetic dynamometer; knee; ROM