QUALITY OF RADIOGRAPHIC IMAGES: LABORATORY EVALUATION OF INTRAORAL FILMS, FILTERS, COLLIMATORS, AND RADIATION EXPOSURE

QUALIDADE DA IMAGEM RADIOGRÁFICA: AVALIAÇÃO LABORATORIAL DE FILMES INTRA-ORAIS, FILTROS, COLIMADORES E EXPOSIÇÃO À RADIAÇÃO

In order to evaluate density, radiographic contrast and dose of radiation exposure, the author analyzed 80 radiographs containing 640 optical density data of the images of a penetrometer, exposed to the radiation beam with combinations between D and E periapical films, aluminum and copper/aluminum filters, and circular or rectangular collimators. The data obtained were analyzed by ANOVA and allowed the following conclusions: 1) aluminum filtration resulted in improved image contrast; 2) the use of group D film and an aluminum filter produced improved image contrast quality; 3) the rectangular collimator contributed to the production of improved contrast and to the reduction of radiation exposure, but did not affect density; 4) the combination of copper/aluminum filter, E group film and rectangular collimation significantly reduced radiation exposure.

Image quality; Radiation filtration; X-ray film


Universidade de São Paulo Avenida Lineu Prestes, 2227 - Caixa Postal 8216, Cidade Universitária Armando de Salles Oliveira, 05508-900 São Paulo SP - Brazil, Tel.: (55 11) 3091-7861, Fax: (55 11) 3091-7413 - São Paulo - SP - Brazil
E-mail: pob@edu.usp.br