Open-access Higenamine Detection and Metabolism Study Using High-Performance Liquid Chromatography Coupled to AgIII-Luminol Chemiluminescence

Abstract

Herein, we developed a sensitive, selective, efficient, and low-cost method for higenamine (HG) detection. This method employs high-performance liquid chromatography with a chemiluminescence detector (HPLC-CL) coupled with an AgIII-luminol system. Lotus plumule and human urine samples were purified using a mixed weak cation exchange column (PWCX) solid-phase extraction cartridge, and a C18 column was employed for chromatographic separation using an aqueous solution of acetonitrile and 0.1% formic acid as the mobile phase. Under optimal conditions, the chemiluminescence (CL) signal decreased linearly with the logarithm of the concentration of HG in the range of 5-800 ng mL-1. The limit of detection (LOD) and limit of quantification (LOQ) of HG in the urine sample were 0.41 and 1.35 ng mL-1, respectively. The average recovery of HG in the urine and lotus plumule samples at three levels ranged from 80.02 to 110.40% with a relative standard deviation of less than 10%. Furthermore, we proposed a mechanism for CL signal reduction in the system. Finally, the urinary pharmacokinetics of HG in humans were studied to provide valuable information and reference for athletes on medication or diet during events.

Keywords:
higenamine; HPLC-CL; lotus plumule; urine; metabolism


Introduction

Higenamine (HG) is a natural alkaloid belonging to the family of benzylisoquinoline alkaloids. It is deemed an adrenergic receptor agonist that produces positive and timing effects on the cardiovascular system by stimulating the β-adrenergic receptor (β-AR).1 In addition, HG can inhibit the expression of nitric oxide synthase (iNOS),2 thus reducing platelet aggregation and thrombosis,3 increasing cardiac output and heart rate as well as lowering blood pressure.4,5 HG is often used clinically as a heart tonic, diuretic, analgesic, or antihypertensive agent.6-8 However, it affects the performance of athletes. Thus, the World Anti-Doping Agency has listed HG as a β2 agonist banned substance in the International Standard for the 2019 Disabled List9 and requires that the concentration of HG in human urine should not exceed 10 ng mL-1. HG exists in many plants, including south Tianzhu, aconitum, pepper, and lotus seeds.10-12 Therefore, it is challenging to control the intake of HG by athletes. To reduce the risk of HG abuse by athletes, improve doping-positive detection, ensure the physical and mental health of athletes, and reduce the negative impact of doping on society, there is a need for a highly efficient, sensitive, and reliable method of HG detection with a wide application range.

HG is commonly detected using separation detection methods, such as high-performance liquid chromatography with UV detection (HPLC-UV),13 high-performance liquid chromatography with fluorescence detection (HPLC-FLD), 14-16 high-performance liquid chromatography with electrochemical detection (HPLC-ECD),17 gas chromatography-mass spectrometry (GC-MS),18 ultra-performance liquid chroma tography tandem mass spectrometry (UHPLC-MS/ MS),19-21 and high-performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS).22 Because of the low sensitivity of HPLC-UV, it is unsuitable for detecting trace HG. HPLC-FLD requires a derivative reaction, which increases the analysis time and process. Although UHPLC-MS/MS has high sensitivity, it is expensive and requires complex laboratory equipment and trained personnel for operation. Many direct analytical techniques are also available for HG detection, such as enzyme-linked immunosorbent assay (ELISA),23 colloidal gold techniques,24 electroanalytical chemistry,25 and timeresolved fluorescence microsphere test strips.1 These methods are rapid or sensitive; however, some of them are inapplicable to biological samples and have narrow linear ranges. High-performance liquid chromatography with a chemiluminescence detector (HPLC-CL) has high sensitivity, high selectivity, low interference, low cost, and ease of operation and has been employed in detecting some bioactive molecules.26-29 However, it has not been employed in HG detection.

Herein, we developed a sensitive, efficient, and low-cost HG detection technology that employs HPLC-CL based on an AgIII-luminol system for detecting HG in lotus plumule and human urine. Furthermore, we proposed a mechanism for the CL signal reduction in the system. We also studied the urinary pharmacokinetics of HG in humans to provide valuable information and reference for athletes on medication or diet during events.

Experimental

Reagents and materials

HG (C16H17NO3, purity > 98%) was purchased from Sigma (St. Louis, MO, USA). Luminol (purity ≥ 98%) was purchased from TCI Company (Shanghai, China). AgIII was prepared following the procedure described in a previous report.30 The concentrations of the reagents were determined spectrophotometrically at 362 nm using the Lambert-Beers law with a molar absorptivity (ɛ) of 1.26 × 104 (mol dm-3)-1 cm-1. Acetonitrile, methyl alcohol, acetic acid, and formic acid were used as analytical reagents. Cleanert® PWCX solid phase extraction column cartridge (150 mg/6 mL) were purchased from Agela Technologies (Tianjin, China).

Preparation of the standard solution

A stock solution of HG was prepared in 0.5% HCl at a final concentration of 0.4 mg mL-1 and stored in a refrigerator at 4 °C. Before use, the standard stock solution was diluted with ultrapure water and divided into several standard working solutions with concentrations ranging from 5.0 to 800 ng mL-1. The calibration curves were generated using the working solutions.

Instrumentation and experimental conditions

The HPLC-CL detection system comprises an HPLC system connected to a post-column CL detection system (Figure 1). HPLC (Shimadzu, Japan) was equipped with a UV detector, a 20-μL manual sample valve, and a C18 column (4.6 mm × 250 mm × 5 μm, DIKMA Diamonsil C18(2), China). The CL detector uses a flow-injection analysis CL system (Xi’an Remax Electronic ScienceTech Co. Ltd.).

Figure 1
Schematic of a high-performance liquid chromatography (HPLC) system with an AgIII-luminol chemiluminescence (CL) detector.

The mobile-phase solution contains 0.1% acetic acid water-acetonitrile (93:7 v/v). The flow rate and column temperature were maintained at constant values of 1 mL min-1 and 30 °C, respectively. The detection wavelength of the UV detector was 284 nm, the pump speed of the CL detector was 20 rpm, and the negative high voltage of the photomultiplier was 600 V.

Samples and their processing

Sample source

The plant samples were Chinese herbal medicine lotus plumules purchased from a local market in China. The use of human samples in this study was approved by the Hebei medical university under protocol No. 2024045. The blank urine samples are midmorning urine from five healthy, unmedicated male volunteers on a light diet a week before the trial, during which the individual consumed no HG-containing products. Positive urine were also collected from the male volunteers after drinking 200 mL of tea brewed with 3.5 g of lotus plumule (containing 1.63 mg of HG) at 100 °C for 40 min. The urine samples were collected and stored at 4 °C. Then, within 6 h of storage, the HG content of the sample was tested.

Lotus plumule pretreatment

After crushing the lotus plumule using a mortar, 10 mg of the powder was poured into a 25-mL volumetric flask, dissolved in ultrapure water, and sonicated for 15 min. The samples were centrifuged for 15 min at 5000 rpm. The supernatant was then removed for solid-phase extraction (SPE).

Solid-phase extraction

First, the mixed weak cation exchange column (PWCX) was conditioned with 2 mL of methanol and then with 2 mL of water. Second, a 2-mL supernatant of the lotus plumule or 4 mL of urine was loaded on the column, followed by washing with 2 mL of water and 2 mL of methanol. Finally, the analytes were eluted with 6 mL of 10% acidified methanol (freshly prepared by adding 10 mL of formic acid to 90 mL of methanol) into a 15-mL centrifuge tube. The eluent was collected and dried with a slow stream of nitrogen at 70 °C. The lotus plumule and urine samples were redissolved in 2- and 1-mL mobile-phase solutions, respectively.

Results and Discussion

Optimization for sample pretreatment

Ultrasonic extraction is used to separate HG from various drugs and foods.18-20 Because HG is a polar molecule, it was extracted using 25 mL of methanol and ultrapure water, respectively. The extraction effect of highpurity water was higher than that of methanol. Ultrapure water was then used as a solvent to extract 10 mg of lotus plumule powder for 5, 10, 20, and 30 min, and the extracts were compared. The results show that ultrasonic time had no significant effect on HG extraction after 10 min. Therefore, 10 min of ultrasonic extraction with water was chosen for further analysis.

SPE optimization

SPE is widely utilized in sample purification. It can be used to purify samples and concentrate and enrich target analytes, thus improving selectivity and limit of detection. Oasis® HLB and Cleanert® PEP-2 SPE columns have been used in the purification of HG from food or biological samples.18,21 To reduce the interference of the sample matrix, herein, two SPE columns with strong and weak cations (Cleanert® PCX and PWCX, respectively) were chosen, and their HG extraction efficiencies were compared. PCX exhibited the strongest retention effect on the target. Under the experimental conditions, completely eluting HG was difficult. However, for PWCX, HG was not only well retained but also completely eluted with formic acid/methanol. For higher recovery, we optimized the use of PWCX columns. Because the retention mechanism of PWCX columns involves ion exchange and antiphase force, we leached the column with 2 mL of water, followed by 2 mL of methanol, to remove part of the matrix interference.

Elution of the target substance is also an important step in SPE. Herein, we first used 5 mL of different proportions of formic acid/methanol as eluants and investigated the effect of their volume percentages (2, 4, 6, 8, 10, and 12%) on the elution efficiency. Optimal recovery was attained at a formic acid/methanol volume percentage of 10%. Next, we varied the volume of the eluents (4, 5, 6, 7, and 8 mL) to obtain the optimum volume of eluent. When the eluent volume was increased to 6 mL, HG on the SPE column was completely eluted. In summary, the optimized conditions for SPE were as follows: loading: 2 mL of the sample; leachate: 2 mL of water and 2 mL of methanol; eluent: 6 mL of a 10% formic acid/methanol solution.

Optimization of chromatography conditions

First, we considered 0.1% acetic acid/methanol, 0.1% acetic acid/acetonitrile, and 15 mmol dm-3 KH2PO4/ acetonitrile solutions, and 0.1% acetic acid exhibited a lower noise level than 15 mmol dm-3 KH2PO4. Thus, the 0.1% acetic acid/acetonitrile solution (93:7 v/v) was selected as the mobile phase for further studies, considering factors such as noise level, separation effect, retention time, and shape of signal peaks.

Optimization of CL conditions

The concentration and alkalinity of CL reagents significantly affect the CL signals. Herein, we varied the CL conditions and compared the signal-to-noise ratio (SNR) of HG to select the best detection conditions.

Optimization of the luminol concentration

Increasing the luminol concentration increases not only weaning CL signals but also baseline noise. Herein, we varied the luminol concentrations (from 2 × 10-7 to 4 × 10-6 mol L-1 (Figure S1, Supplementary Information (SI) section)) and compared their SNRs. The maximum SNR was achieved when the luminol concentration was 2 × 10-6 mol L-1.

Optimization of AgIII concentration

We investigated the influence of the concentration of the AgIII complex on CL signals. We varied the concentration from 5 × 10-6 to 2.5 × 10-5 mol L-1 (Figure S2, SI section). As the concentration of the AgIII complex increased, the decrease degree of CL intensity becomes larger, but the baseline noise increased as well. The maximum SNR was obtained with 2 × 10-5 mol L-1 of the AgIII complex.

Optimization of NaOH in AgIII

After optimizing the concentrations of luminol and AgIII, the concentration of NaOH in AgIII was varied (from 0.02 to 0.06 mol L-1 (Figure S3, SI section)). As the NaOH concentration increased, the SNR initially increased and then decreased. The maximum SNR was obtained when the NaOH concentration in AgIII was 0.05 mol L-1. Thus, 0.05 mol L-1 was selected as the NaOH concentration in AgIII for HG detection.

Optimization of NaOH in luminol

The NaOH concentration in luminol was varied from 0 to 0.04 mol L-1 to determine its effects on the baseline stability and CL SNR. The results show that the lower the alkalinity of luminol, the faster the base signals decrease over time. This indicates that a luminol solution is more stable in a strong alkaline medium. When the luminol concentration was 2 × 10-6 mol L-1, the NaOH concentration in luminol was higher than 0.03 mol L-1, and the CL baseline signal remained stable. When the NaOH concentration in luminol was too high, the SNR decreased. Considering the stability of the baseline and SNR, 0.03 mol L-1 was selected as the optimum NaOH concentration in luminol.

Method validation

Linearity

Under the optimized assay conditions, the concentration of HG ranged from 5 to 800 ng mL-1, and the diminished CL signal showed a linear positive correlation with the logarithm of the concentrations. The regression equation was expressed as ∆I = 1394log c - 884.5 (linear correlation coefficient (r) 0.9994), where ∆I is the reduced CL signal intensity, and c is the HG concentration in ng mL-1.

Limits of detection and quantification

The limit of detection (LOD) and limit of quantification (LOQ) of the proposed method were three and ten times the noise of the baseline signal, respectively. Because the enrichment multiples of the lotus plumule and urine samples were one and four times, respectively, the LODs (SNR = 3) of the lotus plumule and urine samples were 4.1 mg kg-1 and 0.41 ng mL-1, and their LOQs (SNR = 10) were 13.5 mg kg-1 and 1.35 ng mL-1, respectively.

Various methods of HG detection were compared in Table 1. The LOD of the proposed method (HPLC-CL) was significantly lower than that of HPLC-FLD/ECD, GC-MS, and voltammetric methods but higher than that of UPLC-MS/MS.19-21 However, HPLC-CL is more costeffective and easier to operate than UHPLC-MS/MS.

Table 1
Comparison of various methods of HG detection
Real sample analysis and matrix effects

HG concentrations in the lotus plumule and urine samples were measured. The HG average content of the lotus plumule sample was 466.13 mg kg-1. The CL chromatograms of the HG standard solution and lotus plumule sample are shown in Figure 2. HG was completely separated within 12.2 min, and there was no interference of HG in the lotus plumule sample. The CL chromatograms of urine are shown in Figure 3. The CL chromatogram of blank urine (Figure 3b) shows no evident interference peak at the HG retention time. However, in the UV chromatogram (254 nm), there was an interference peak at the HG retention time, indicating that the CL detection system is more selective toward HG.

Figure 2
CL chromatograms of (a) lotus plumule and (b) HG standard solution.

Figure 3
CL chromatograms of (a) positive urine and (b) blank urine.

To evaluate matrix effects (ME), the matrix-matched calibration curves were compared to the standard calibration curves (with the same concentration levels). The calculation was carried out according to the equation: ME (%) = (B/A) × 100, here A is the slope of the solvent standard curve and B is the slope of the matrix-matched calibration curve. The ME is 89.2% in plumule and 92.5% in urine. The results showed that the two matrices showed no obviously matrix effect on the analytes.

Accuracy and precision

The sample spiked recovery and precision of the proposed method were evaluated at low, medium, and high concentrations of HG in samples (Table 2). The low, medium, and high recovery rates for the urine and lotus plumule samples ranged from 80.0 to 110.4%, the intraday precision ranged from 2.8 to 5.9%, and the interday precision ranged from 5.4 to 9.2%. These results indicate that the proposed method has good accuracy and reproducibility.

Table 2
Recovery and precision of HG in urine and lotus plumule samples
Urinary pharmacokinetics of HG

Five volunteers were instructed to drink 3.5 g of lotus plumule tea, which contained approximately 1.63 mg of HG, brewed in 200 mL of pure water at 100 °C for 40 min. Subsequently, 20 mL of midstream urine was collected from each volunteer at hourly intervals from 1 to 6 h after consuming the lotus plumule tea. It is important to note that during this time frame, the volunteers refrained from consuming any food or drugs.

Urine samples were collected from five volunteers at various time points and analyzed using the proposed method to assess changes in the concentration of HG. Figure 4 illustrates the results. The urine sample taken 2 h after consuming the herbal tea showed the highest HG content, while no HG was found in the urine collected at the 6th hour, aligning with findings from a previous study.22

Figure 4
Concentration-time curve of HG in urine samples after drinking lotus plumule tea. A. B, C, D, and E represent five different volunteers respectively.

CL mechanism

According to a previous study,31 some free radicals are produced when luminol reacts with AgIII and upon the action of certain reducing substances with AgIII. Free-radical detection experiments have shown that the free radicals produced from CL reactions are neither hydroxyl (•OH) nor superoxide radicals (•O2-).32 Thus, we propose the CL reaction mechanism in the proposed HG detection system.

First, an electron from luminol is captured by AgIII and form an anionic radical of luminol. AgIII continues to oxidize the anionic radical to form 3-amino-phthalate in the excited state (3-AP*) with a peak emission wavelength of 425 nm when it returns to the ground state. HG is also oxidized by AgIII to form HG free radicals. Because the oxidation rate of HG by AgIII is lower than that of luminol, the luminol radical formed upon the oxidization of luminol by AgIII is partly transferred to HG, which reduces the number of luminophores, thereby reducing the CL signal intensity. The mechanism of the reduction in CL signal intensity is depicted in Scheme 1.

Scheme 1
Reduction mechanism for CL signal intensity in the CL system.

Conclusions

Herein, we developed an HPLC-CL-based method for HG detection. The developed method was successfully used to detect HG in urine and lotus plumule samples. The method is highly selective and sensitive, simple to operate, rapid, and cost-effective. Based on previous studies, we proposed an inhibition mechanism for the CL system. Changes in the concentration of HG in urine samples of an individual after the intake of HG-containing herbal tea were also successfully measured using the proposed method. The proposed method would enhance HG detection, thereby preventing excessive intake of the stimulant by humans.

Supplementary Information

Supplementary data are available free of charge at http://jbcs.sbq.org.br as PDF file.

Acknowledgments

This work was supported by the Natural Science Foundation of Hebei province (grant numbers H2021206245 and H2022206045).

References

  • 1 Zhou, S.; Xu, X.; Wang, L.; Kuang, H.; Xu, C.; Food Chem. 2022, 387, 132859. [Crossref]
    » Crossref
  • 2 Pyo, M. K.; Lee, D. H.; Kim, D.; Lee, J.; Moon, J.; Chang, K.C.; Yun-Choi, H. S.; Bioorg. Med. Chem. Lett. 2008, 18, 4110. [Crossref]
    » Crossref
  • 3 Pyo, M. K.; Kim, J. M.; Jin, J. L.; Chang, K. C.; Lee, D. H.; Yun-Choi, H. S.; Thromb. Res. 2007, 120, 81. [Crossref]
    » Crossref
  • 4 Dunlop, D.; Shanks, R. G.; Brit. J. Pharmacol. 2012, 36, 132. [Crossref]
    » Crossref
  • 5 Praman, S.; Mulvany, M. J.; Williams, D. E.; Andersen, R. J.; Jansakul, C.; J. Ethnopharmacol. 2012, 140, 166. [Crossref]
    » Crossref
  • 6 Kosuge, T.; Yokota, M.; Nagasawa, M.; Yakugaku. Zasshi. 1978, 98, 1370. [Crossref]
    » Crossref
  • 7 Feng, S.; Jiang, J.; Hu, P.; Zhang, J. Y.; Liu, T.; Zhao, Q.; Acta Pharmacol. Sin 2012, 33, 1353. [Crossref]
    » Crossref
  • 8 Zhang, N.; Lian, Z.; Peng, X.; Li, Z.; Zhu, H.; J. Ethnopharmacol. 2017, 196, 242.[Crossref]
    » Crossref
  • 9 World Anti-Doping Code; Prohibited List 2022, International Standard, https://www.chinada.cn/upload/files/2021/12/b352c3004b57ed6a.pdf, accessed in November 2024.
    » https://www.chinada.cn/upload/files/2021/12/b352c3004b57ed6a.pdf
  • 10 Calvert, R.; Vohra, S.; Ferguson, M.; Wiesenfeld, P.; Food Chem. Toxicol. 2015, 78, 207. [Crossref]
    » Crossref
  • 11 Yun-Choi, H. S.; Pyo, M. K.; Park, K. M.; Chang, K. C.; Lee, D. H.; Thromb. Res. 2001, 104, 249. [Crossref]
    » Crossref
  • 12 Tsukiyama, M.; Ueki, T.; Yasuda, Y.; Kikuchi, H.; Akaishi, T.; Okumura, H.; Abe, K.; Planta Med. 2009, 75, 1393. [Crossref]
    » Crossref
  • 13 Liu, Y. J.; Yu, H. M.; Zhang, H.; Yu, L. S.; Xu, W.; Microchem. J. 2017, 134, 289e94. [Crossref]
    » Crossref
  • 14 Feng, T. H.; Tao, Y.; Ri, J. D.; J. Yanbian Univ. 2012, 2, 150. [Link] accessed in November 2024
    » Link
  • 15 Hong, H. X.; Lee, Y. I.; Jin, D. R.; Microchem. J. 2010, 96, 374. [Crossref]
    » Crossref
  • 16 Wang, Y.; Zhang, Y.; Zhang, P.; Chin. J. Anal. Lab. 2020, 39, 1276. [Crossref]
    » Crossref
  • 17 Lo, C. F.; Chen, C. M.; J. Chromatogr. B 1994, 655, 33. [Crossref]
    » Crossref
  • 18 Feng, Y. R.; Wang, B.; Li, G. J.; Kang, W. J.; Lian, K. Q.; Lu, X. L.; J. Food. Drug. Anal. 2020, 28, 124. [Crossref]
    » Crossref
  • 19 Stajić, A.; Anelkovi, M.; Diki, N.; Rai, J.; Jani-Stojanovi, B.; J. Pharm. Biomed. Anal. 2017, 146, 48. [Crossref]
    » Crossref
  • 20 Yan, Y. H.; Lian, K. Q.; Lv, P.; Ai, L. F.; Zhang, H. C.; Zhang, Y.; J. Food Compos. Anal. 2022, 110, 104524. [Crossref]
    » Crossref
  • 21 Okano, M.; Sato, M.; Kageyama, S.; Drug Test. Anal 2017, 9, 1788. [Crossref]
    » Crossref
  • 22 Feng, S.; Hu, P.; Jiang, J.; J. Chromatogr. B 2011, 879, 763. [Crossref]
    » Crossref
  • 23 Nuntawong, P.; Tanaka, H.; Sakamoto, S.; Morimoto, S.; Planta Med. 2020, 86, 760. [Crossref]
    » Crossref
  • 24 Nuntawong, P.; Ochi, A.; Chaingam, J.; Tanaka, H.; Sakamoto, S.; Morimoto, S.; Drug Test. Anal. 2021, 13, 762. [Crossref]
    » Crossref
  • 25 Allahverdiyeva, S.; Talay, P.; Keskin, E. Yunusoğlu, O.; Yardım, Y.; Şentürk, Z.; Sens. Actuaors, B 2020, 303, 127174. [Crossref]
    » Crossref
  • 26 Li, N.; Guo, J. Z.; Liu, B.; Anal. Chim. Acta 2009, 645, 48. [Crossref]
    » Crossref
  • 27 Nalewajko, E.; Wiszowata, A.; Kojto, A.; J. Pharmaceut. Biomed. Anal. 2007, 43, 1673.[Crossref]
    » Crossref
  • 28 Ma, L.; Shi, H. M.; Lian, K. Q.; Diao, Y. F.; Chen, Y.; Ma, C. L.; Kang, W. J.; Food Chem. 2017, 218, 422. [Crossref]
    » Crossref
  • 29 Ma, L.; Zhao, T. J.; Zhang, P. P.; Liu, M. Y.; Shi, H. M.; Kang, W. J.; Anal. Biochem. 2020, 593, 113594. [Crossref]
    » Crossref
  • 30 Blikungeri, A.; Pelletier, M.; Monnier, D.; Inorg. Chim. Acta 1977, 22, 7. [Crossref]
    » Crossref
  • 31 Shi, H. M.; Xu, X. D.; Ding, Y. X.; Anal. Biochem. 2009, 387, 178. [Crossref]
    » Crossref
  • 32 Ma, L.; Niu, L.; Wang, W.; Kang, W.; Shi, H. M.; J. Braz. Chem. Soc. 2014, 25, 867. [Crossref]
    » Crossref

Editor handled this article:

Eduardo Carasek

Publication Dates

  • Publication in this collection
    13 Jan 2025
  • Date of issue
    2025

History

  • Received
    29 Aug 2024
  • Accepted
    02 Dec 2024
location_on
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Ir para o topo