Open-access CATHETER FOR MECHANICAL LABOR INDUCTION: TECHNOLOGY DEVELOPMENT FROM THE PERSPECTIVE OF DESIGN THINKING

CATÉTER PARA INDUCCIÓN MECÁNICA DEL TRABAJO: DESARROLLO TECNOLOGICO DESDE LA PERSPECTIVA DEL DESIGN THINKING

ABSTRACT

Objective:  to develop a low-cost catheter to mechanically induce labor.

Method:  this is a technological study developed from May to December 2023. It used the Design Thinking perspective and the Double Diamond design methodology through four stages. A literature review was conducted in the first stage entitled “discover”, involving research in two patent databases, and research on products available on the national market. Then, the personas, the empathy map and the mind map were established in the “define” stage. Next, the brainstorming technique was performed in the “develop” stage by the co-creative team and finally the final three-dimensional product was presented in the “deliver” stage.

Results:  this study identified that although there are many patents for labor-induction catheters, only two are known in the scientific community, one of which is used in an improvised manner. The catheters used in obstetric practice have weaknesses that were considered in developing the product presented in this study.

Conclusions:  The methodological steps of Design Thinking enabled producing an innovative, low-cost technology that fills a gap in labor induction. However, challenges remain regarding production, feasibility testing and introduction of this technology into the market.

DESCRIPTORS:
Health technologies; Obstetrics; Obstetric nursing; Induced labor; Technology applied to healthcare

RESUMO

Objetivo:  desenvolver um cateter de baixo custo para indução mecânica do trabalho de parto.

Método:  pesquisa tecnológica desenvolvida no período de maio a dezembro de 2023. Utilizou a perspectiva do Design Thinking e a metodologia projetual do Double Diamond, através de quatro etapas. Na primeira etapa, intitulada “descobrir”, foi realizada a revisão de literatura, pesquisa em duas bases de patentes, e pesquisa de produtos disponíveis no mercado nacional. Na etapa “definir”, foram estabelecidas as personas, o mapa de empatia, e o mapa mental. Na etapa “desenvolver”, foi realizada a técnica de Brainstorming pela equipe cocriativa, e, na etapa “entregar”, foi apresentado o produto final tridimensional.

Resultados:   este estudo identificou que apesar de haver muitas patentes de cateteres para a indução de parto, apenas dois são conhecidos no meio científico, sendo um deles utilizado de forma improvisada. Os cateteres utilizados na prática obstétrica apresentam fragilidades que foram consideradas na elaboração do produto apresentado nesta pesquisa.

Conclusões:   as etapas metodológicas do Design Thinking permitiram produzir uma tecnologia inovadora de baixo custo e que preenche uma lacuna na indução de parto. Entretanto, permanecem os desafios em relação à produção, testes de viabilidade, e introdução dessa tecnologia no mercado.

DESCRITORES:
Tecnologias em saúde; Obstetrícia; Enfermagem obstétrica; Trabalho de parto induzido; Tecnologia aplicada aos cuidados de saúde

RESUMEN

Objetivo:  desarrollar un catéter de bajo costo para la inducción mecánica del parto.

Método:  investigación tecnológica realizada de mayo a diciembre de 2023. Se utilizó la perspectiva Design Thinking y la metodología de diseño Double Diamond, a través de cuatro etapas. En la primera etapa, denominada “descubrir”, se realizó una revisión de la literatura, una investigación de dos bases de patentes y una investigación de productos disponibles en el mercado nacional. En la etapa de “definición” se establecieron las personas, el mapa de empatía y el mapa mental. En la etapa de “desarrollo” se realizó la técnica de Brainstorming por parte del equipo co-creativo, y en la etapa de “entrega” se presentó el producto final tridimensional.

Resultados:  este estudio identificó que aunque existen numerosas patentes de catéteres para inducir el parto, en el ámbito científico sólo se conocen dos, una de las cuales se utiliza de forma improvisada. Los catéteres utilizados en la práctica obstétrica presentan debilidades que fueron consideradas en el desarrollo del producto presentado en esta investigación.

Conclusiones:  los pasos metodológicos del Design Thinking permitieron producir una tecnología innovadora y de bajo costo que llena un vacío en la inducción laboral. Sin embargo, persisten desafíos con respecto a la producción, las pruebas de viabilidad y la introducción de esta tecnología en el mercado.

DESCRIPTORES:
Tecnologías sanitarias; Obstetricia; Enfermería obstétrica; Trabajo de parto inducido; Tecnología aplicada a la salud

INTRODUCTION

In the context of the humanization of childbirth and birth related to women’s rights and desires as well as obstetric practices which aim for favorable outcomes, labor induction is recommended whenever there is a clinical indication, such as in cases of hypertensive syndromes, gestational diabetes or other maternal comorbidities or fetal conditions that contraindicate continuing the pregnancy1-3.

Offering the possibility of inducing labor is a woman’s right and is directly associated with a positive birth experience4. Furthermore, promoting natural childbirth significantly contributes to reducing maternal complications associated with cesarean sections, helping to reduce the maternal morbidity and mortality rate, which constitutes one of the sustainable development goals established by the United Nations5.

Labor induction is one of the most common obstetric procedures and/or interventions found in maternity wards around the world. It consists of preparing the cervix through stimulation with the aim of promoting uterine contractions and consequently triggering labor. This practice is recommended from 22 weeks of gestation when continuing the pregnancy represents a greater maternal-fetal risk than terminating it1,6-7.

It is known that one in four pregnant women have their labor induced in the United States, while there was an increase in the United Kingdom from 29.4% (2016-2017) to 31.6% (2017-2018). These rates are lower in developing countries, but increasing1,7.

The methods used to induce labor are divided into three types: the first refers to non-pharmacological methods, which use oils (among other methods), such as evening primrose; the second refers to mechanical stimuli; for example, digital detachment of the amniotic membranes or using an extra-amniotic catheter; and the third refers to pharmacological methods, in which oxytocin and prostaglandin tablets are generally used1,6-8.

The induction method choice considers an assessment of the cervix and the woman’s obstetric history. The presence of an unfavorable cervix is ​​one of the main difficulties in performing this procedure. In addition, a history of cesarean section advises against the use of prostaglandin tablets due to the increased risk of uterine rupture. Therefore, choosing the method to be used must be indicated on an individual basis8-12.

Labor induction may be perceived as an intervention in the natural process of pregnancy for some women, especially in relation to the length of stay in the hospital environment, which tends to be longer when induction is necessary, and the possibility of using medication. This may discourage women from opting for a natural birth and may also generate anxiety in them and their companions due to the increased length of hospital stay and the lack of autonomy in relation to the birthing process13.

Thus, mechanical methods are viable alternatives to consider. These methods, such as the use of catheters, are used to improve the cervix condition and have rare side effects, helping women in the induction process and also allowing them to stay at home for longer8-11,13. In addition, they have advantages when compared to pharmacological methods, including simplicity of preservation, lower cost and a lower chance of side effects11,14.

Catheters used for labor induction can be single or double balloon, both of which are inserted into the internal os of the cervix, where the balloon is inflated. The double balloon catheter has an extra balloon that is inflated into the external os of the cervix. Randomized clinical studies comparing the efficacy of the Foley catheter (single balloon) and the double balloon device have shown similar results in terms of efficacy and safety9-11,13,15-17.

The purpose of using a catheter is to stimulate release of local prostaglandins, which contributes to prepare the cervix and consequently the beginning of labor11,13. The main indication for catheter use is related to contraindications to the use of prostaglandins, or in cases where induction methods are combined to optimize the process, such as when associating the mechanical and pharmacological methods14-16.

The single-balloon catheter is commonly used in professional practice and is well-known in the Brazilian context, but it was not created for this purpose, meaning that it is an adaptation. This catheter was developed for indwelling bladder catheterization, known as a Foley catheter. Since it was not developed for the purpose of inducing labor, it may present some problems during use, such as balloon rupture due to the inability of the manufacturer to inflate it to the appropriate size and difficulty in inserting it into the internal os of the cervix8-11,13.

In view of the above, this study was proposed in light of these findings and based on the authors’ professional practice, since we have identified these problems in obstetric care, in addition to the lack of access to the specific device due to its high cost. Thus, the objective of this study was to develop a low-cost catheter for mechanically inducing labor. It was decided to use Design Thinking (DT), since the steps defined by this method favor developing products in the health area, integrating creativity and sensitivity with critical thinking with the aim of solving the problem18-20.

METHOD

This is a technological study which used the DT perspective and the Double Diamond (DD) design methodology, developed between May and December 202318.

The DD design method is visually represented (Figure 1) in a sequential double diamond diagram, composed of four phases: discover, define, develop and deliver19-20; the steps used in this process were described in each of these phases.

Figure 1 -
Double Diamond design method adapted for the study.

The “discovery” phase is characterized by searching for information about the initial problem and understanding the issue in a more comprehensive way19-20. Data was sought during this phase to understand catheters for labor induction, as well as their weaknesses and potential. Thus, a literature review was conducted from August to November 2023 in Pubmed/MEDLINE using the following descriptors: “Double-balloon versus single-balloon catheter AND labor induction”; “Type a catheter AND labor induction”. The language was not restricted and articles from the last 10 years were selected. Articles which compared the types of induction catheters with each other were included and articles that included drug induction methods or associated methods were excluded.

Another search in parallel with the literature review was performed in two patent databases. The databases used were Espacenet (European Patent Office, Derwent Innovations Index, Patentscope) and the database of the National Institute of Industrial Property (INPI). The search used the descriptors (catheter*) AND (labor induction), with no year or language specified. The patents were selected by analyzing the first 200 patents shown on the website using relevance criteria. This number was established by observing patents for products that did not correspond to catheters for labor induction.

Then, a search was performed for products related to the research object that are already in circulation on the Brazilian market as the last stage of the “discovery” process. This search was through the catalogs of companies that develop and supply medical-hospital articles. Catheters used in obstetric practice were also included, even if they were not developed for the purpose of labor induction.

The “define” phase defines and understands what needs to be solved, planning the project development based on the data found in the previous phase19-20. Personas were established for this stage, which constitute characters that represent the characteristics of a group or real people, who would be the target audience/clientele21. In this case, the personas were represented by health professionals, obstetric nurses and obstetric doctors who use catheters to induce labor; this group was called “Obstetrician”. The second persona represents women who received some type of catheter during their labor induction process, called “Pregnant woman”.

Next, empathy maps were created in order to characterize the personas. The empathy map consists of a tool developed biased on six items: what they hear, what they see, what they think and feel, what they say and do, pains and gains. The objective is to obtain as much information as possible about the persona and the topic20.

Then, the co-creation team (composed of the authors of this study) gathered observations of obstetric care practices to identify the catheters used to induce labor, the difficulties in applying the catheter and suggestions for modifications, in addition to the findings identified in the literature review in order to collect data, define the problem and identify possible solutions.

The mind map was also developed in this phase, organizing the information identified in the previous phases, giving rise to ideas through keywords which enabled a better view of the problem to be solved. It was subsequently possible to begin the development phase based on the ideas raised in the mind map.

The “develop” phase represents generating ideas or solutions. Based on the Brainstorming performed by the co-creative team, this technique aims to create ideas/products/solutions with a focus on quantity, combinations and absence of criticism20. Ideas emerged in this phase and were classified as viable or unviable. These ideas were raised through the use of colored post-its and distributed on a white card. The team then selected the viable and important adaptations and changes for developing the product in the next stage, grouping them into: catheter, guide wire/stylette and balloon. In turn, tests were conducted with materials acquired from other catheters, arriving at the digital prototype.

The last phase, called “delivery”, is characterized by producing a prototype for testing and subsequent insertion into the job market20. At this point, the present work will focus on elaborating the prototype in the form of an industrial design. It is worth noting that this industrial design with its specificities was registered with the INPI under the number: BR 30 2024 001042 7.

The prototype developed for testing by the end user of the product will occur in a future study.

RESULTS

The results will be presented according to the steps carried out in the DD design methodology.

Starting from the first stage called “discovery”, a literature review was performed in which 22 studies published between 2013 and 2023 were found, then seven articles were selected after applying the selection and analysis criteria.

Among the selected articles, there are four systematic reviews that discussed the differences between single and double balloon catheters9-11,13, and three randomized clinical trials that compared the efficacy of the types of catheters15-17. All studies found used the Foley catheter (single balloon) and the Cook® device (double balloon). Furthermore, the selected articles present similar results in relation to efficacy and safety, with the Foley being significantly cheaper and widely available9-11,13,15-17

In addition to the literature review, a search for patents on labor induction catheters was performed. The INPI database did not show any patents, while the Espacenet database showed 4,369 patents. Due to the large volume of patents identified, the first 200 patents were read. Several documents presented brief descriptions, without details about the product. In addition, most presented images of prototypes, but they were barely visible and did not demonstrate applicability of the catheter. Thus, nine patents were selected, three of American origin and six of Chinese origin. Of these, six proposed double-balloon catheters and three devices with a single balloon. Regarding the materials used, one used polyvinyl and another patent used silica gel; the other six proposed devices with silicone and one with latex. The volume capacity varied from 40 to 500 ml for the internal balloons and 20 to 80 ml for the external balloons. Two patents presented a guide wire, while seven patents have oval shapes.

After reviewing the literature and searching for patents, the catheters available on the market were investigated. Four devices were found, two American and two Chinese. The following specifications of the products stand out: three have a double balloon, all with a capacity of 80 ml in the internal balloon. Two catheters have a guide wire. All cost more than 200 reais per unit. Only one of these products (Cook®) was available for purchase in Brazil.

In addition to researching the products available on the market, a search for catheters used in obstetric practice was performed to observe their shape, texture, consistency and functional mechanism. The catheter commonly used in an improvised manner in the practice of mechanical induction of labor is the Foley catheter. The Foley catheter is sold in Brazil by several brands, but they are imported from other countries and resold in Brazil, with an average cost of 3 reais per unit. The material is either silicone or latex. The balloon capacity varies according to the size of the catheter but should not exceed 60ml.

Next, the personas were identified and an empathy map was created in the second stage of the DD. These tools helped identify the demands of those involved regarding the use of induction catheters. It was noted that the most used catheter in labor induction is the Foley catheter due to its cost and availability. Some weaknesses in the use of this device were observed, such as: discomfort during insertion and use, difficulty in introducing the catheter, lack of effectiveness in dilating the cervix and rupture of the balloon when introducing the volume due to its limited capacity.

Furthermore, a meeting was held in this stage of the DD with the co-creative team in order to define the main weaknesses and potentialities of the catheters used in practice. This stage also helped in the creation process, as it listed the modification points for developing the product. In addition, a mind map was created (Figure 2) as a tool to facilitate visualization of the demands found in this meeting. It was possible to better understand the problem in question and identify possible solutions by completing the steps of the first diamond.

Figure 2 -
Mind map created by the co-creative team.

The information collected in the previous stage was then used in the development stage. The team gathered materials and concepts that could help in the creative process using the Brainstorming technique. These ideas were raised using colored post-its and distributed on a white card. After this process, the post-its were organized into three groups: catheter, guide wire and balloon.

At the end of the Brainstorming, it was established that the catheter should be flexible, made of latex (due to the lower cost), and should be 35 cm long, a size defined by the length required between the beginning and the end of the vaginal canal. The catheter needs to have two ports: one port has a device to attach the syringe and inflate the balloon, and the other port is for inserting the guide wire. The distal tip of the catheter needs to have a small plastic region that ensures that the catheter will not break when in contact with the guide wire. The balloon needs to hold a capacity of 80 ml. This capacity has been proven to be effective in thinning and dilating the cervix9-11,13,16.

Finally, the team used adapted materials and developed the prototype, which was later transformed by a Design firm hired by the researchers into a 3D design (Figure 3) and industrial design. The product developed includes improvements in the main weaknesses identified in the previous processes.

Figure 3 -
Three-dimensional product design.

DISCUSSION

Although there is consistent research on the advantages and safety of using catheters for mechanically inducing labor, the devices available and used in studies still have weaknesses9-11,13. A systematic review that compared the single (Foley) and double (Cook®) catheters demonstrated differences related to pain, where the painful sensation was significantly greater with the use of the double catheter, both during insertion and during use9. The single balloon catheter used in an improvised manner (Foley) is significantly cheaper, widely available and accessible; however, corroborating the findings of this study, it has a greater chance of balloon rupture due to the inability to inflate the volume recommended for labor induction9-11,13.

Regarding the ability to improve cervical conditions, a randomized study comparing the use of a single and double catheter found that the single-balloon catheter appears to be as effective as the double-balloon catheter17. Another systematic review with meta-analysis found no statistically significant difference when comparing single and double-balloon catheters in relation to time to delivery, rates of vaginal delivery, cesarean section, or use of epidural for labor induction11. These findings were considered in developing the catheter proposed in this study, since the use of two balloons did not present significant advantages.

Another randomized study compared the use of a single balloon inflated with 30 ml and the double-balloon catheter (Cook®) with 80 ml. The findings suggest that the 80 ml double-balloon catheter is more effective than the single-balloon catheter with 30 ml for cervical dilation and achieving vaginal delivery in nulliparous women. The amount inflated in each catheter followed the manufacturers’ guidelines16. Furthermore, a systematic review found studies that used 50ml and 60ml in the single-balloon catheters11, with the best response in terms of cervix preparation appearing to be associated with the 80ml volume11,16.

Nevertheless, no studies were found in this literature review which analyzed the catheter length or the material used for its manufacture. Therefore, it was decided to use the average measurement of the vaginal canal. The choice to use latex for the material was due to its low cost. Moreover, it was decided to include a guide wire with coated metal to facilitate introducing the catheter, which was one of the findings during the DT process, also low cost, and which is removed after the catheter is introduced.

All studies found in the stage that included the literature review used the Foley and Cook® catheters; however, the review of the patent databases showed that there are other products developed, but not available on the Brazilian market. The large number of registered patents that did not result in commercialized products is evidenced by the use in professional practice of an improvised catheter. It is also worth noting that the researchers’ knowledge, combined with their practical experience in contact with obstetric professionals and pregnant women, was essential for developing this product, since the professionals' experience and knowledge constitute essential technical data for developing new technologies that will be useful in healthcare practice. In addition, the findings found in the literature review were compatible with the difficulties encountered in healthcare practice regarding the use of the Foley catheter.

In view of this, the construction of the final product combined scientific evidence and the experience of professionals in obstetric care practice, creating a product which meets the appropriate specificities for mechanical labor induction, with the prospect of being low-cost and safe for the care of women in labor.

It is worth mentioning that the product materialization, usability performance and effectiveness tests are necessary for this new technology to be introduced into the market. Materializing new technologies is a major challenge in the nursing field, which requires improving knowledge in favor of developing new products and seeking partners who contribute to this process.

FINAL CONSIDERATIONS

The results of this study enabled creating a product that aims to bring improvements to obstetric practice with the following specific features: a low-cost, flexible, latex catheter, 35 cm long, two-way, with a single cuff with a capacity of 80 ml of solution, and a coated and removable metal guide wire. It is expected that this product can be introduced into the market and that it will be useful in inducing labor.

It is essential to emphasize the importance of studies which aim to build products that can be used in professional practice, using guiding and co-creative methodologies. It was possible to identify through the DT how much the convergence of care practice with scientific research contributes to greater resolution of the problem under study.

The challenges that permeate producing technologies in the health area continue in search for partner companies to develop the product. In this case, despite being a low-cost technology that contributes to obstetric care, the challenge remains to comply with the other steps necessary for insertion into the market. It is hoped that this study will be the first step towards developing others, and that this product can be validated and marketed, improving obstetric care.

REFERENCES

  • 1. World Health Organization - WHO. Recommendations: Induction of labour at or beyond term [ ]. 2018 [cited 2023 Jun 09]. Available from: https://www.who.int/publications/i/item/9789241550413
    » https://www.who.int/publications/i/item/9789241550413
  • 2. Scapin SQ, Gregório VRP, Collaço VS, Knobel R. Indução de parto em um Hospital Universitário: métodos e desfechos. Texto Contexto Enferm [Internet]. 2018 [cited 2023 Jun 10];27(1):e0710016. Available from: https://doi.org/10.1590/0104-07072018000710016
    » https://doi.org/10.1590/0104-07072018000710016
  • 3. Finucane EM, Murphy DJ, Biesty LM, Gyte GML, Cotter AM, Ryan EM, et al. Membrane sweeping for induction of labour. Cochrane Database Sys Rev [Internet]. 2020 [cited 2023 Jun 09];2(2):CD000451. Available from: https://doi.org/10.1002/14651858.CD000451.pub3
    » https://doi.org/10.1002/14651858.CD000451.pub3
  • 4. Coates D, Goodfellow A, Sinclair L. Induction of labor: experiences of care and decision-making of women and clinicians. Women Birth [Internet]. 2020 [cited 2023 Jun 09];33(1):e1-14. Available from: https://doi.org/10.1016/j.wombi.2019.06.002
    » https://doi.org/10.1016/j.wombi.2019.06.002
  • 5. Organização das Nações Unidas. Objetivos de Desenvolvimento Sustentável no Brasil [Internet]. [s.d.] [cited 2023 Jun 09]. Available from: https://brasil.un.org/pt-br/sdgs
    » https://brasil.un.org/pt-br/sdgs
  • 6. Middleton P, Shepherd E, Morris J, Crowther CA, Gomersall JC. Induction of labour at or beyond 37 weeks' gestation. Cochrane Database Sys Ver [Internet]. 2020 [cited 2023 Jun 12];7(7):CD004945. Available from: https://doi.org/10.1002/14651858.CD004945.pub5
    » https://doi.org/10.1002/14651858.CD004945.pub5
  • 7. Nggada BJ. Induction of labour. In: Current Challenges in Childbirth [Internet]. 2022 [cited 2023 Jun 09]. Available from: https://doi.org/10.5772/intechopen.104445
    » https://doi.org/10.5772/intechopen.104445
  • 8. Souza ASR, Costa AAR, Coutinho I, Noronha Neto C, Amorim MMR. Análise crítica dos métodos não-farmacológicos de indução do trabalho de parto. Femina [Internet]. 2010 [cited 2023 Jun 16];38(4):195-201. Available from: https://pesquisa.bvsalud.org/portal/resource/pt/lil-546446
    » https://pesquisa.bvsalud.org/portal/resource/pt/lil-546446
  • 9. Yang F, Huang S, Long Y, Huang L. Double-balloon versus single-balloon catheter for cervical ripening and labor induction: A systematic review and meta-analysis. J Obstet Gynaecol Res [Internet]. 2017 [cited 2023 Jun 20];44(1):27-34. Available from: https://doi.org/10.1111/jog.13551
    » https://doi.org/10.1111/jog.13551
  • 10. Lajusticia H, Martínez-Domínguez SJ, Pérez-Roncero GR, Chedraui P, Pérez-López FR. Single versus double-balloon catheters for the induction of labor of singleton pregnancies: A meta-analysis of randomized and quasi-randomized controlled trials. Arch Gynecol Obstet [Internet]. 2018 [cited 2023 Jun 20];297(5):1089-100. Available from: https://doi.org/10.1007/s00404-018-4713-9
    » https://doi.org/10.1007/s00404-018-4713-9
  • 11. Los Reyes SX de, Sheffield JS, Eke A. Single versus double-balloon transcervical catheter for labor induction: A systematic review and meta-analysis. Am J Perinatol [Internet]. 2019 [cited 2023 Jun 22];36(8):790-7. Available from: https://doi.org/10.1055/s-0038-1675206
    » https://doi.org/10.1055/s-0038-1675206
  • 12. Rodrigues DP, Alves VH, Silva AM, Penna LHG, Vieira BDG, Silva SÉD, et al. Percepção de mulheres na assistência ao parto e nascimento: obstáculos para a humanização. Rev Bras Enferm [Internet]. 2022 [cited 2023 Jun 22];75(2):e20210215. Available from: https://doi.org/10.1590/0034-7167-2021-0215
    » https://doi.org/10.1590/0034-7167-2021-0215
  • 13. Liu X, Wang Y, Zhang F, Zhong X, Ou R, Xin Luo X. Double- versus single-balloon catheters for labour induction and cervical ripening: A meta-analysis. BMC Pregnancy Childbirth [Internet]. 2019 [cited 2023 Jun 22];9(1):358. Available from: https://doi.org/10.1186/s12884-019-2491-4
    » https://doi.org/10.1186/s12884-019-2491-4
  • 14. Stephenson E, Borakati A, Simpson I, Eedarapalli P. Foley catheter for induction of labour: A UK observational study. J Obstet Gynaecol [Internet]. 2020 [cited 2023 Jun 22];40(8):1064-8. Available from: https://doi.org/10.1080/01443615.2019.1676213
    » https://doi.org/10.1080/01443615.2019.1676213
  • 15. Mei-Dan E, Walfisch A, Valencia C, Hallak M. Making cervical ripening EASI: A prospective controlled comparison of single versus double balloon catheters. J Matern Fetal Neonatal Med [Internet]. 2014 [cited 2023 Jun 22];27(17):1765-70. Available from: https://doi.org/10.3109/14767058.2013.879704
    » https://doi.org/10.3109/14767058.2013.879704
  • 16. Hoppe KK, Schiff MA, Peterson SE, Gravett MG. 30 mL Single- versus 80 mL double-balloon catheter for pre-induction cervical ripening: A randomized controlled trial. J Matern Fetal Neonatal Med [Internet]. 2016 [cited 2023 Jun 22];29(12):1919-25. Available from: https://doi.org/10.3109/14767058.2015.1067297
    » https://doi.org/10.3109/14767058.2015.1067297
  • 17. Pez V, Deruelle P, Kyheng M, Boyon C, Clouqueur E, Garabedian C. Cervical ripening and labor induction: Evaluation of single balloon catheter compared to double balloon catheter and dinoprostone insert. Gynecol Obstet Fertil Senol [Internet]. 2018 [cited 2023 Jun 22];46(7-8):570-4. Available from: https://doi.org/10.1016/j.gofs.2018.05.009
    » https://doi.org/10.1016/j.gofs.2018.05.009
  • 18. Design Council. Eleven Lessons: Managing design in eleven global brands: A study of the design process [Internet]. 2007 [cited 2023 Jun 22]. Available from: https://www.designcouncil.org.uk/our-work/skills-learning/resources/11-lessons-managing-design-global-brands/
    » https://www.designcouncil.org.uk/our-work/skills-learning/resources/11-lessons-managing-design-global-brands/
  • 19. Pyykkö H, Suoheimo M, Walter S. Approaching Sustainability Transition in Supply Chains as a Wicked Problem: Systematic Literature Review in Light of the Evolved Double Diamond Design Process Model. Processes [Internet]. 2021 [cited 2023 Jun 22];9(12):2135. Available from: https://doi.org/10.3390/pr9122135
    » https://doi.org/10.3390/pr9122135
  • 20. Stein M, Costa R, Gelbcke FL. Enfermagem e design na criação de produtos para a saúde: aproximando áreas e resolvendo problemas. Texto Contexto Enferm [Internet]. 2022 [cited 2023 Jun 22];32:e202201602023. Available from: https://doi.org/10.1590/1980-265X-TCE-2022-0160pt
    » https://doi.org/10.1590/1980-265X-TCE-2022-0160pt
  • 21. Pazmino AV. Como se cria: 40 métodos para design de produtos. São Paulo, SP(BR): Blucher; 2015.

NOTES

NOTES

FUNDING INFORMATION

Fundação de Amparo à Pesquisa e Inovação do Estado Santa Catarina (FAPESC) - Chamada Pública 48/2021, Número do processo: 3003/2021. Programa de Fomento à Pós-Graduação em Instituições de Educação Superior do Estado de Santa Catarina.

TRANSLATED BY

Christopher J. Quinn.

CORRESPONDING AUTHOR

Soliane Scapin. solscapin@gmail.com

CONFLICT OF INTEREST

There is no conflict of interest.

EDITORS

Associated Editors: Bruno Miguel Borges de Sousa Magalhães, Ana Izabel Jatobá de Souza. Editor-in-chief: Elisiane Lorenzini.

Publication Dates

  • Publication in this collection
    25 Apr 2025
  • Date of issue
    2025

History

  • Received
    25 July 2024
  • Accepted
    23 Oct 2024
location_on
Universidade Federal de Santa Catarina, Programa de Pós Graduação em Enfermagem Campus Universitário Trindade, 88040-970 Florianópolis - Santa Catarina - Brasil, Tel.: (55 48) 3721-4915 / (55 48) 3721-9043 - Florianópolis - SC - Brazil
E-mail: textoecontexto@contato.ufsc.br
rss_feed Acompanhe os números deste periódico no seu leitor de RSS