Open-access Cochlear implant: A comparative analysis of using time

ABSTRACT

Purpose:  to compare the time of cochlear implant use reported by caregivers and/or users with the actual time of use measured by data logging.

Methods:  across-sectional study of cochlear implant patients with a data logging system compared reported data with logged data. The analysis used Student's t-test or Mann-Whitney test (p < 0.05).

Results:  the sample comprised 46 individuals - 63% being females, and 80.4% being children. The reported time of use ranged from 8 to 14 hours, and the measured time ranged from 1 to 12 hours. The variables differed significantly: the reported time was 2.1 hours, on average, greater than the measured time (p < 0.001) in the overall sample and in the groups of children/adolescents and adults/older adults.

Conclusion:  the time reported by family members/caregivers and/or users differed from the actual time of cochlear implant use; the actual time of use was less than that reported for time of use.

Keywords:
Cochlear Implant; Hearing Loss; Children; Adults; Aged

RESUMO

Objetivo:  comparar o tempo referido de uso do implante coclear (IC) pelos responsáveis e/ou usuários com o tempo de uso real aferido pelo registro de dados.

Métodos:  estudo transversal, incluindo pacientes em uso de IC com sistema de registro de dados. Foi feita a comparação entre os dados relatados e os constantes no software. A análise foi realizada por meio do teste t-student ou Mann-Whitney (p < 0,05).

Resultados:  a amostra foi composta por 46 indivíduos, sendo 63% do sexo feminino e a maioria crianças (80,4%). O tempo de uso referido variou entre oito e 14 horas, e o tempo aferido entre um e 12 horas. Foi constatada diferença significativa entre as variáveis analisadas, sendo o tempo referido, em média, 2,1 horas maior do que o aferido (p < 0,001), na amostra geral e nos grupos de crianças/adolescentes e adultos/idosos.

Conclusão:  houve diferença entre o tempo referido pelos familiares/responsáveis e/ou usuários e tempo de uso real do implante coclear, sendo o tempo real menor que o tempo de uso referido.

Descritores:
Implante Coclear; Perda Auditiva; Crianças; Adultos; Idosos

INTRODUCTION

Hearing loss is one of the most common causes of disability in the Brazilian population1. When left untreated, severe and profound sensorineural hearing loss prevents the acquisition of oral language and hinders communication, potentially negatively impacting individuals' daily lives1,2. These difficulties can be minimized with early diagnosis and intervention1,2. Auditory rehabilitation is indicated in these cases, made possible by using a hearing aid (HA) or cochlear implant (CI)2,3, accompanied by speech-language-hearing (SLH) therapy and guidance for parents, family members, guardians, caregivers, and teachers.

In the Brazilian Unified Health System (SUS), multidisciplinary teams decide which device will be used in auditory rehabilitation; hence, SLH pathologists are indispensable members of such teams, according to criteria established in Brazilian regulations4. Patient care protocols are developed based on what these regulations describe, including initial assessments, criteria for the indication of HAs or CIs, and a systematic approach to rehabilitation after fitting the devices. CIs can be an excellent option for individuals with severe and/or profound sensorineural hearing loss who have not benefited from HAs4,5.

Technological advances have enabled the development of the CI, a biocompatible biomedical device that directly stimulates the auditory nerve through a bundle of electrodes that simulates the function of damaged hair cells in the cochlea6,7. The CI is an important advance in the treatment of hearing loss, especially for intervention in pre-lingual children, whose communication with peers8-10, hearing thresholds, and sound and speech perception improve after implantation10.

However, some factors influence the optimization of CI benefits. These include the actual time of use, as using the device regularly provides better hearing ability in all age groups, reflecting the neuroplasticity of the central auditory system11,12.

The time of CI use can be assessed in two ways: subjectively, by verifying the patient-reported time use, and objectively, by measuring data logging with information about the actual CI use13. This assessment makes SLH guidance more effective, influencing adherence to proper CI use, which is fundamental to obtaining the maximum benefit from this device13.

Children’s daily time of device use is essential for successful rehabilitation, being related to auditory development and, consequently, language development. In older individuals, prolonged use promotes continued inclusion in the sound world, allowing the maintenance of already acquired skills, and improving their quality of life and social relationships. Therefore, professionals always highlight the time of use in their guidance, as it ensures greater benefits from perceived sounds and, therefore, better results. Thus, this study aimed to compare the time of CI use reported by caregivers and/or users with the actual time of use measured by data logging.

METHODS

The research project that gave rise to this article was approved by the Research Ethics Committee of the Clinics Hospital of Porto Alegre (HCPA), RS, Brazil, registered under CAEE 16926619.0.0000.5327 and approval number 3.491.800. The study complies with the provisions of Resolution No. 466/2012 of the National Health Council, Ministry of Health, Brazil. This study collected data anonymously and retrospectively from electronic medical records; hence, it was granted permission without the need to sign an informed consent form.

This is an observational, cross-sectional, comparative study, whose target population consisted of patients using CI, treated at a center registered with the Ministry of Health, who met the criteria for CI use according to Regulation No. 2,776/20144. Data regarding patients and CIs were obtained by analyzing medical records. The study included data from the medical records of pediatric, adult, and older patients who attended the institution in the postoperative period, with a minimum of 60 days of CI use, and whose time of device use was analyzed objectively (through data logging) and subjectively (through questions made by the SLH pathologist in charge). Patients were of both sexes and all age groups, used unilateral or bilateral CI of a model that recorded the time of use, and whose electronic medical record had data of the time of use reported by caregivers and/or users. Patients who received CIs without a data logging system were excluded from the sample.

After the device was implanted and medical clearance was obtained, patients returned to the SLH service approximately 30 to 45 days after the surgical procedure. At this time, the SLH pathologist activated the device, including its reassessment, electrode mapping, initial usage programming, and guidance for patients and/or parents/guardians. The pathologist observed the patient's initial reactions and gave instructions for device care and handling. At the end of the consultation, patients were rescheduled for a follow-up appointment, which took place at least 60 days later.

In the follow-up visit, patients were asked about their experience using the device and the average time of use. The CI was coupled with the software to verify information such as electrode operation, time of CI use in hours through data logging, programs used, and so forth.

This research considered the following data from the electronic medical records: age, sex, side of the implanted ear, time between surgery and CI activation, measured and reported time of use, average hearing thresholds with the CI, sound intensities, and sound environments to which patients were exposed.

Sample size was calculated using the Program for Epidemiologists for Windows (WinPEPI), version 11.43. For this purpose, the study considered the 405 patients who received CI at the hospital through the SUS and the number of patients in whom the device was implanted and used with the data logging program. The sample size calculation showed that the sample size should be at least 45 patients.

The data were collected and entered into an Excel spreadsheet. Statistical analysis used the Statistical Package for the Social Sciences (SPSS) for Windows, version 21.0. Quantitative variables (age, hearing threshold, reported time of use, measured time of use, and the time between surgery and activation) were described using mean and standard deviation or median and interquartile range. Categorical variables (sex and implanted ear) were described by absolute and relative frequency.

The study used Student’s t-test for independent samples to compare means between children/adolescents and adults/older adults, and the paired t-test to compare means between reported and measured times. In case of asymmetry, the Mann-Whitney test was applied. Pearson's correlation test was used to assess associations of hearing thresholds with measured time and the time between surgery and activation. The significance level was set at 5% (p < 0.05).

RESULTS

The sample consisted of 46 patients - 37 under 18 years old and nine adults; 29 (63%) were females. Their median age was 3.4 years (Q1 - Q3: 2.3 - 6.5), with an age range from 2 to 82 years (Table 1). Also, 25 CIs (54.3%) were implanted in the right ear, and 21 CIs (45.7%) were implanted in the left ear (Table 1).

Table 1
Sample characterization

The reported time of CI use was statistically significantly different from the time measured through data logging (p < 0.001 and p = 0.037, respectively). The reported time was, on average, 2.1 hours longer than the measured time (95% confidence interval: 1.3 to 2.9). The sample’s hearing threshold in free field, with the CI turned on, was 41.8 dB HL (41.8 ± 19.4) among both children and adults.

Hearing thresholds were not significantly associated with measured time (r = -0.286; p = 0.222) or the time elapsed between surgery and activation (r = -0.270; p = 0.224) (Table 2).

Table 2
Relationship between the reported time of cochlear implant use and the time of use measured through data logging.

The separate analysis of comparative data between children and adults found a statistically significant difference between the reported and measured times in both age ranges (p < 0.001 and p = 0.037, respectively). Children’s time of use was overestimated by 2.1 hours (95% CI: 1.2 to 3.1), very similar to the adults’ overestimated 2 hours (95% CI: 0.2 to 3.8).

Table 3
Data on cochlear implant use by age group

DISCUSSION

CI is a high-tech resource used for severe and profound sensorineural hearing loss, aiming at auditory rehabilitation, as it enables access to the world of sound14. CI has a positive impact on hearing loss, enabling the development of auditory skills, especially with early surgery and SLH rehabilitation15.

The analysis of medical record data showed that most CI patients who used devices with data logging systems were children/adolescents, in line with the sample found in studies available in the literature3,15-17. It is believed that this finding is due to the priority given to implanting the device in children, according to criteria established by the Ministry of Health regulation and the institution where the research was carried out, which prioritizes children due to a budgetary limitation established by public administration.

The reported and the measured time of CI use differed significantly. The reported time of use was 9.4 ± 1.5 hours, and the measured time was 7.3 ± 2.6 hours. This result is consistent with a previous study, in which half of the participants used it for an average of 7.6 hours/day18. Another study with children under 5 years old found a shorter daily time than that obtained in this study, with an average of 6.7 hours/day19. In other retrospective studies, the time was much longer, such as one with patients aged 0 to 26 years, which showed an average time of use of 10.76 hours/day20. Another study, with young and older adults, showed an average time of use of 13.29 hours/day for adults and 10.97 hours/day for older adults21. According to the literature, the ideal minimum time of use should be 8 or more hours/day to improve auditory and linguistic development19.

The comparison between data of children/adolescents and adults/older adults showed that the parents/guardians of the former reported significantly longer time of use than adults/older adults. However, the time measured by data logging did not differ significantly between them. Although children used the CI for longer than adults, their mean time still did not reach or surpass the recommended minimum of 8 hours/day.

An explanation for the difference between the reported and measured time of use may be related to the profile of the study participants. Because they attend a center that serves patients through the SUS, the team professionals (SLH pathologists, otolaryngologists, social workers, and psychologists) who care for the patient emphasize in all visits the importance and necessity of using the device for at least 8 hours/day. Many of the patients are socially vulnerable, and the follow-up sessions require constant repetition of the information on all aspects of the CI, including time of use. Hence, some parents and users may have difficulty retaining information and be apprehensive about CI technology. However, they end up reporting a longer time of use than actually occurred, perhaps not wanting to disappoint the team regarding non-use for the indicated time, or out of gratitude for the care received.

Furthermore, questioning the time of use introduces a memory bias, since users and parents/guardians are not initially informed about the existence of the data logging system and may not pay close attention to the time of CI placement or removal. Another factor to consider, especially in children, is that the responses expected by parents/relatives/guardians are not always obtained from the beginning. There is a known necessary time from CI activation until obtaining responses to verbal stimuli (recognizing one's name, discriminating sounds more specifically, locating a sound source, and so on), requiring SLH therapy and guidance. Moreover, post-lingual patients perceive sound differently, which may also explain the shorter time of use, since the expected response is not as quick as patients and parents/relatives/guardians would expect, even if instructed by the team about it.

Thus, it is believed that professionals working with CI users should record the actual and self-reported time of use in their clinical practice to better advise the patient and family about the lack of benefit of the device when used for only a few hours daily16. Adequate time of use is extremely relevant, as is family commitment to CI use, thus ensuring adequate neural stimulation and promoting auditory habilitation in both children and adults13.

According to Regulation4 that governs CI processes and the protocol adopted where the study was developed, there should be a maximum of 45 days between surgery and electrode activation. Exceptions may occur in the case of healing difficulties. In the sample evaluated, the mean time was 1.3 ± 0.5 months - i.e., within the time stipulated by the interdisciplinary team that attends to the patients.

The mean free-field hearing thresholds with the CI were 41.8 ± 19.4 dB in the overall sample, 40.9 ± 19.5 dB in children/adolescents, and 44.2 ± 20.6 dB in adults/older adults. These values are ​​higher than those found by other researchers22. It is believed that these thresholds may also influence the shorter time of use, since mean values different from what is considered a normal hearing threshold indicate that the user does not have full access to the sound world, making some important phonemes of Brazilian Portuguese imperceptible.

This may contradict the expectations associated with CI use, even though most are exposed to sound environments with higher intensities (greater than 56 dB). When comparing the sound intensities to which the patients were exposed, the analysis allowed us to identify a difference between children/adolescents and adults/older adults. Most used CI in quiet environments with sound intensities up to 65 dB, revealing that most users probably did not use CI outside of the home or SLH therapy environment.

Regarding the sound environments to which they were exposed, the analysis of the overall sample showed that the CI was massively used in silent speech environments. The analysis by age group confirms these data, with a significant difference regarding only music, an environment to which more children/adolescents were exposed than adults. These data again seem to confirm that CI was used by age groups basically at home. Adults/older adults were slightly more exposed to noisy environments, showing a borderline significant difference.

This data is relevant, since the sound environments to which CI users are exposed in their daily lives are a challenge for the development of auditory skills. The data obtained are in line with those found in two other current studies available in the literature, which, however, only analyzed children17,23. In an Italian study17, the CI proved to be a well-established treatment for severe to profound hearing loss in young infants. This study evaluated the data records of patients using CIs, classified the sound environments, and obtained similar results regarding speech in quiet and speech in noise17.

Exposure to music differed between the adult and child groups, with higher exposure in the child population. This finding is consistent with other published studies24,25, in which musical exposure peaked in early childhood and declined significantly with increasing patient age. A previous review indicated that CIs could lead to unnatural conditions of dynamic changes in sound level, low-frequency selectivity due to current propagation, and low-pitch perception, which can affect the perception of emotion conveyed by music. Users can perceive rhythm, but the perception of melody and harmony may be impaired, which would explain the lower exposure of adults to music26. It should be noted, however, that the data analyzed here refer to patients in the initial period of use, and that such perception may change over time due to the user's greater experience in a musical environment. Thus, it is believed that both music and noisy environments are more challenging at the beginning of use, although they should always be addressed during SLH therapy.

One of the limitations of the study was the lack of assessment of the caregivers’ and CI users’ education level and socioeconomic level, since the study highlighted higher income as a relevant factor in CI use20.

CONCLUSION

The time reported by caregivers and/or patients differed significantly from the actual time of CI use - the latter was significantly shorter than the reported time of use.

REFERENCES

  • 1 Instituto Brasileiro de Geografia e Estatística. Pesquisa Nacional de Saúde. [Webpage on the Internet]. Brasília: Instituto Brasileiro de Geografia e Estatística [Accessed on 5 Jun 2025]. Available at: https://sidra.ibge.gov.br/tabela/5721#notas-tabela
    » https://sidra.ibge.gov.br/tabela/5721#notas-tabela
  • 2 Sales CB, Resende LM, Amaral CFS. Auditory rehabilitation in adults: Results of training program. Rev. CEFAC. 2019;21(5):1-12. https://doi.org/10.1590/1982-0216/201921510318
    » https://doi.org/10.1590/1982-0216/201921510318
  • 3 Fortunato CAU, Bevilacqua MC, Costa MPR. Comparative of verbal language analysis of listeners children and deaf children with cochlear implant. Rev. CEFAC. 2009;1(4):662-72. https://doi.org/10.1590/S1516-18462009000800015
    » https://doi.org/10.1590/S1516-18462009000800015
  • 4 BRASIL. Ministério da Saúde. Portaria nº 2.776, de 18 de dezembro de 2014. Diretrizes Gerais para a Atenção Especializada às Pessoas com Deficiência Auditiva no Sistema Único De Saúde - SUS/Ministério da Saúde. Secretaria de Atenção à Saúde. Departamento de Atenção Especializada e Temática. Coordenação Geral de Média e Alta Complexidade. Brasília. 2014
  • 5 Garcez V, Teixeira AR. Aparelho de amplificação sonora individual/componentes e características eletroacústicas. In: Boechat EM, Menezes PL, Couto CM, Frizzo ACF, Scharlach RC, Anastasio ART, editors. Tratado de audiologia. Rio de Janeiro: Guanabara Koogan Ed. 2015. p.253-8.
  • 6 Ramos D, Jorge JX, Teixera A, Ribeiro C, Paiva A. The impact of cochlear implant in the oral language of children with congenital deafness. Acta Med Port. 2015;28(4):442-7.
  • 7 Oliveira JAA. Cochlear implant. Medicina (Ribeirão Preto). 2005;38(3/4):262-72.
  • 8 Morettin M, Cardoso MRA, Delamura AM, Zabeu JS, Amantini RCB, Bevilacqua MC. Use of the International Classification of Functioning, Disability and Health for monitoring patients using cochlear implants. CoDAS. 2013;25(3)216-23. https://doi.org/10.590/s2317-17822013000300005 PMID: 24408331.
    » https://doi.org/10.590/s2317-17822013000300005
  • 9 Shapiro WH, Bradham TS. Cochlear Implant Programming. Otolaryngol Clin North Am. 2012;45(1):111-27. https://doi.org/10.1016/j.otc.2011.08.020 PMID: 22115685.
    » https://doi.org/10.1016/j.otc.2011.08.020
  • 10 Vassoler TMF, Cordeiro ML. Brazilian adaptation of the Functioning After Pediatric Cochlear Implantation (FAPCI): Comparison between normal hearing and cochlear implanted children. J. Pediatr (Rio J). 2015;91(2):160-7. https://doi.org/10.1016/j.jped.2014.06.008 PMID: 25458875.
    » https://doi.org/10.1016/j.jped.2014.06.008
  • 11 Sousa AF, Couto MI, Martinho-Carvalho AC. Quality of life and cochlear implant: Results in adults with postlingual hearing loss. Braz J Otorhinolaryngol. 2018;84(4):494-9. https://doi.org/10.1016/j.bjorl.2017.06.005
    » https://doi.org/10.1016/j.bjorl.2017.06.005
  • 12 Froemke RC, Heman-ackah SE, Waltzman SB. Neuroplasticidade auditiva. In: Waltzman SB, Roland Jr. TJ, editors. Implantes Cocleares. Rio de Janeiro: Revinter, 2016. p.38-46.
  • 13 Laybauer AS. Implante Coclear. In: Ferreira MIDC, editor. Reabilitação auditiva: Fundamentos e proposições para a atuação no Sistema Único de Saúde (SUS). Ribeirão Preto. Book Toy; 2017. p.123-41.
  • 14 Cruz MS, Oliveira LR, Carandina L, Lima MCP, Cesar CLG, Barros MBA et al. Prevalence of self-reported hearing loss and attributed causes: A population-based study. Cad Saúde Pública. 2009;25(5):1123-31. https://doi.org/10.1590/S0102-311X2009000500019 PMID: 19488497
    » https://doi.org/10.1590/S0102-311X2009000500019
  • 15 Moret ALM, Bevilacqua MC, Costa OA. Cochlear implant: Hearing and language in pre-lingual deaf children. Pró-Fono R. Atual. Cientif. 2007;19(3):295-304. https://doi.org/10.1590/S0104-56872007000300008 PMID: 17934605.
    » https://doi.org/10.1590/S0104-56872007000300008
  • 16 Wiseman KB, Warner-Czyz AD. Inconsistent device use in pediatric cochlear implant users: Prevalence and risk factors. Cochlear Implant Int. 2018;19(3):131-41. https://doi.org/10.1080/14670100.2017.1418161 PMID: 29299970.
    » https://doi.org/10.1080/14670100.2017.1418161
  • 17 Guerzoni L, Cuda D. Speech processor data logging helps in predicting early linguistic outcomes in implanted children. Int J Pediatr Otorrinolaryngol. 2017;101:81-6. https://doi.org/10.1016/j.ijporl.2017.07.026 PMID: 28964316.
    » https://doi.org/10.1016/j.ijporl.2017.07.026
  • 18 Barbosa MH, Felix F, Ribeiro MG, Tomita S, Pinheiro C, Baptista MM. Profile of patients assessed for cochlear implants. Braz J Otorhinolaryngol. 2014;80(4):305-10. https://doi.org/10.1016/j.bjorl.2014.05.011
    » https://doi.org/10.1016/j.bjorl.2014.05.011
  • 19 Wiseman KB, Warner-Czyz AD, Kwon S, Fiorentino K, Sweeney M. Relationship between daily device use and early communication outcomes in young children with cochlear implants. Ear and Hear. 2021;42(4):1042-53. https://doi.org/10.1097/AUD.0000000000000999 PMID: 33974791.
    » https://doi.org/10.1097/AUD.0000000000000999
  • 20 Awad DR, Castaño JE, Chi DH. Socioeconomic status and cochlear implant usage: A dalalogging study. Ann Otol Rhinol Laryngol. 2023;132(12):1535-42. https://doi.org/10.177/00034894231170588 PMID: 37096343.
    » https://doi.org/10.177/00034894231170588
  • 21 Schvartz-Leyzac KC, Conrad CA, Zwolan TA. Datalogging statistics and speech recognition during the first year of use in adult cochlear implant recipients. Otol Neurotol. 2019;40(7):e686-e693. https://doi.org/10.1097/MAO.0000000000002248 PMID: 31135672.
    » https://doi.org/10.1097/MAO.0000000000002248
  • 22 Costa LBA, Vicente LC, Silva LTN, Alvarenga KF, Salgado MH, Costa AO et al. Analysis of aided thresholds in children who have undergone cochlear reimplantation: A ten-year follow-up. CoDAS. 2023;35(6):e20210293. https://doi.org/10.1590/2317-1782/20232021293en PMID: 37909539.
    » https://doi.org/10.1590/2317-1782/20232021293en
  • 23 Contrera KJ, Choi JS, Blake CR, Betz JF, Niparko JK, Lin FR. Rates of long-term cochlear implant use in children. Otol Neurotol. 2014;35(3):426-30. https://doi.org/10.1097/MAO.0000000000000243 PMID: 24518403; PMCID: PMC3927162.
    » https://doi.org/10.1097/MAO.0000000000000243
  • 24 Ganek HV, Cushing SL, Papsin BC, Gordon KA. Cochlear implant use remains consistent over time in children with single-sided deafness. Ear Hear. 2020;41(3):678-85. https://doi.org/10.1097/AUD.0000000000000797 PMID: 31567563.
    » https://doi.org/10.1097/AUD.0000000000000797
  • 25 Busch T, Vanpoucke F, van Wieringen A. Auditory environment across the life span of cochlear implant users: Insights from data logging. J Speech Lang Hear Res. 2017;60(5):1362-77. https://doi.org/10.1044/2016_JSLHR-H-16-0162 PMID: 28418532.
    » https://doi.org/10.1044/2016_JSLHR-H-16-0162
  • 26 Moore BJC. The perception of emotion in music by people with hearing loss and people with cochlear implants. Philos Trans R Soc Lond B Biol Sci. 2024;379(1908):20230258. https://10.1098/rstb.2023.0258 PMID: 39005027.
    » https://10.1098/rstb.2023.0258

A study conducted at the Clinics Hospital of Porto Alegre, Porto Alegre, RS, Brazil.

Financial support

Nothing to declare

Data sharing statement

Research data will not be shared.

Correspondence

Corresponding author: Adriane Ribeiro Teixeira, Rua Ramiro Barcelos, 2600 - Bairro Santa Cecília, Instituto de Psicologia da UFRGS, CEP: 90035-003 - Porto Alegre - RS, Brasil, E-mail: adriane.teixeira@gmail.com

Conflict of interests

Adriane Ribeiro Teixeira declares she is an editorial board member of Revista CEFAC but was not involved in the peer review or editorial decision-making process for this article

Chief Editor

Erissandra Gomes

Associate Editor

Denise Costa Menezes

Publication Dates

  • Publication in this collection
    23 Mar 2026
  • Date of issue
    2026

History

  • Received
    05 June 2025
  • Reviewed
    02 Sept 2025
  • Accepted
    30 Oct 2025
location_on
ABRAMO Associação Brasileira de Motricidade Orofacial Rua Uruguaiana, 516, Cep 13026-001 Campinas SP Brasil, Tel.: +55 19 3254-0342 - São Paulo - SP - Brazil
E-mail: revisora1@revistacefac.com.br
rss_feed Stay informed of issues for this journal through your RSS reader
Go to top