Open-access Ambient light interference in a digital radiography system with phosphor plate

Interferência da luz ambiente em um sistema de radiografia digital com placas de fósforo

ABSTRACT

Objective:  Photostimulable phosphor plate is a digital image receptor used in intraoral dental radiographic techniques. Because it is photosensitive, it is affected by ambient light on image quality. The aim of this study is to quantify the interference of ambient lighting on the radiographic image.

Methods:  A total of 140 radiographs were obtained with a Photostimulable phosphor plate and divided into seven groups according to the use of opaque and transparent protective covers, ambient light intensity, and a light protection box over the scanner. In a software, the optical densities of the radiographs obtained under different conditions were measured by means of measures of central tendency and dispersion. In comparisons between groups, the Kruskal Wallis test was used.

Results:  The highest mean optical density was 154.2 (100%) with the ambient light on and PSP with the active side on the transparent face of the protective cover. There was a significant reduction in light interference when the Photostimulable phosphor plate was used in a fully opaque protective cover, with reduced ambient light and a box over the scanner (mean optical density of 15.1).

Conclusion:  The instruction that the processing of the radiographic image in environments with reduced light should be respected. The use of fully opaque protective covers and light protectors over the scanner and reduction of the time between image acquisition and processing are measures that help maintain radiographic quality.

Indexing terms
Dental digital radiography; Light; Photobleaching; Quality control

RESUMO

Objetivo:  A placa de fósforo fotoestimulável é um receptor de imagem digital utilizado em técnicas radiográficas odontológicas intrabucais. Por ser fotossensível, é afetada pela luz ambiente na qualidade da imagem. O objetivo do estudo é quantificar a interferência da iluminação ambiente na imagem radiográfica.

Métodos:  Um total de 140 radiografias foram obtidas com uma placa de fósforo fotoestimulável e divididas em sete grupos de acordo com o uso de capas protetoras opacas e transparentes, intensidade da luz ambiente e uma caixa de proteção contra luz sobre o scanner. Em um software, as densidades ópticas das radiografias obtidas sob diferentes condições foram medidas por meio de medidas de tendência central e dispersão. Nas comparações entre os grupos, foi utilizado o teste de Kruskal-Wallis.

Resultados:  A maior média de densidade óptica foi de 154,2 (100%) com a luz ambiente acesa e a placa de fósforo fotoestimulável com o lado ativo na face transparente da capa protetora. Houve uma redução significativa na interferência da luz quando a placa de fósforo fotoestimulável foi utilizada em uma capa protetora totalmente opaca, com luz ambiente reduzida e uma caixa sobre o scanner (média de densidade óptica de 15,1).

Conclusão:  A instrução de que o processamento da imagem radiográfica em ambientes com luz reduzida deve ser respeitada. O uso de capas protetoras totalmente opacas e protetores de luz sobre o scanner e a redução do tempo entre a aquisição e o processamento da imagem são medidas que ajudam a manter a qualidade radiográfica.

Termos de indexação
Radiografia digital odontológica; Luz; Fotobranqueamento; Controle de qualidade

INTRODUCTION

Intraoral radiography can be performed using radiographic films or digital systems. Digital radiography, in addition to not using chemical solutions for processing, requires shorter radiation exposure times and allows for corrections of contrast, brightness, and image density in software, thus reducing the need to repeat many radiographs. It is composed of two systems: a direct digital system using charge-coupled device or complementary metal oxide semiconductor sensors, and a semi-direct digital system, using Photostimulable Phosphor storage plate (PSP) and a laser scanner connected to a computer [1,2].

The PSP is a digital image receptor with the same dimensions as radiographic films, with good image quality and reduction of working times and patient exposure to radiation [1,3]. When a PSP is exposed to X-rays, it stores energy, absorbed in the crystalline structure of the phosphor, in the form of trapped electrons, constituting a latent image. This deposited energy can be released if stimulated by additional light with an appropriate wavelength [4,5]. Therefore, a stimulating red light source is applied by the scanner to move the electrons in metastable states, from which the energy transfer will create a visible light that is accumulated and digitized, producing a visible digital radiographic image on a monitor. After scanning the images, the scanner emits a strong light source to erase the residual images, in order to leave the plate ready to make new radiographs [1,6-9].

After exposure to X-rays, the PSP needs to be quickly processed in the scanner, because the electrons that are in the metastable state inevitably move to the ground state over time. The rate of electron release is very short after exposure and differs based on the composition of the phosphor layer and storage conditions such as temperature and light. Some crystals release 23% of their trapped electrons after 30 minutes [10].

However, PSP exposure to ambient lighting during the digitalization process can interfere with the quality of the images. There is a loss of electrons and, consequently, a decrease in the density of the radiographs, in addition to fading, changes in contrast, brightness and saturation of the image [11,12]. It may occur due to the shape of the scanner, the size of the plate or the time of insertion and scanning of the plate [2]. In addition, some manufacturers provide protective plastic covers for the PSP that are opaque on one side and transparent on the other side, which can increase the incidence of light on the plate both during the radiographic technique and during its transport to the scanner [2,13]. Therefore, the sensitivity of the PSP to light is an obstacle to be overcome in its use [13,14].

Considering that improving the quality of radiographic images implies reducing repetitions and consequently the radiation dose to patients for more accurate diagnoses, the objective of this study was to quantitatively evaluate the interference of ambient light in PSP. For this, the use of opaque or transparent protective plastic covers on the PSP, the control of ambient light and the use of an opaque box over the scanner were compared.

METHODS

In a room with a 40W fluorescent lamp lit on the ceiling, a new size 2 PSP (Eagle PS - Dabi AtlanteTM, Brasília, Brazil), at a distance of eight centimeter from the X-ray machine’s cylinder (Spectro 70X - Dabi AtlanteTM, Brasília, Brazil), was exposed 140 times (exposure time of 0.5 seconds, 70kVp, 7mA).

Using a protective plastic cover provided by the manufacturer with one transparent and one opaque side, 80 exposures were performed. Of these, 40 exposures were made with the active side of the PSP positioned towards the transparent side of the cover, with 20 processed in the scanner with the room light on (Group 1) and 20 with the room light off (Group 3). In the other 40 exposures, the PSP was positioned with the active side facing the opaque side of the cover, with 20 inserted into the scanner (Eagle PS Dabi AtlanteTM, Brasília, Brazil) with the room light on (Group 2) and 20 with the room light off (Group 4).

In another 60 exposures, the same PSP was used, but using a protective cover made of opaque plastic on both sides, provided by a manufacturer exclusively for this experiment. Of the 60 exposures, 20 were processed in the scanner with the room light on (Group 5), 20 with room light off (Group 6) and 20 with room light off and an opaque cardboard box over the scanner with openings only for inserting the hands (Group 7). Group 7 was created to assess the influence of light from the computer monitor positioned above the scanner (figure 1).

Figure 1
A. Exposure of the PSP to X-rays. B) PSP Scanning; C) Active side of PSP towards opaque side of the cover; D) Active side of PSP towards transparent side of the cover; E and F) Both sides of a protective plastic cover fully opaque.

After scanning the PSP, all radiographs were exported from the Eagle PS software (Eagle PS Dabi AtlanteTM, Brasília, Brazil) in JPEG format and imported into the DigoraTM for Windows 2.1 software (Orion Corporation Soredex, Tuusula, Finland), in which the pixel corresponding to the optical density of the lightest area of each radiograph was measured, obtaining a value between 0 (black) and 255 (white), thus quantifying the light input in the radiographs.

The results were cataloged and submitted to statistical tests. The STATA/SE 12.0 and Microsoft Excel 365TM software were used, and all tests were applied with 95% confidence. The numerical variables were represented by measures of central tendency and measures of dispersion. The Kolmogorov-Smirnov Normality Test was applied for quantitative variables and, in the comparison with more than two groups, the Kruskal Wallis Test, with Dunn’s multiple comparison.

RESULTS

Statistically significant differences were found in the optical densities of the radiographs among the different groups (p<0.001) according to the Kruskal-Wallis Test, due to the variation of lighting conditions and PSP protective covers. In table 1, the designated letters indicate which groups showed significant differences in relation to the reference group. These results suggest that different lighting conditions influence optical density, with variations observed between the groups.

Table 1
Average density measurements for areas exposed to light.
Figure 2
Mean optical density of bright areas per light exposure.

As shown in figure 2, the highest mean optical density was 154.2 (100%) when the scanner room’s ceiling light was kept on and the PSP with the active side facing the transparent face of the protective plastic cover (Group 1) and there was a 21% reduction of this average density (121), in relation to Group 1, when keeping the active part towards the opaque side of the cover (Group 2).

When the scanner room light was turned off and the exposed PSP was processed with the active side facing the transparent side of the cover (Group 3), the density decreased by 45% (average: 84.2) and it was minus 62% (58.9) with the PSP exposed with the active side facing the opaque side of the cover (Group 4), compared to Group 1.

When using a fully opaque cover on the PSP and keeping the scanner room light on (Group 5), the density reduction was 68% (49.95), and with the light off (Group 6) it was 88.5% (17.4) and, when adding an opaque box over the scanner (Group 7), under the same conditions as Group 6, the density was lower by 90% (15.1), compared to Group 1. Figure 3 shows the radiographic images obtained under different conditions.

Figure 3
Radiographs of the seven tested groups. The light areas are due to the Photostimulable Phosphor storage plate exposure to light.

DISCUSSION

The results of the present study are in agreement with the current literature, which reports that the quality of the radiographic image on the PSP is directly related to the amount of light to which the plate is exposed, from the execution of the radiographic technique to the moment of its insertion into the scanner especially in series of periapical radiographs of all teeth, since there is an interval between the first and the last radiograph taken [6,7,10-12,15].

In addition to light, operator-related errors also interfere with the quality of digital radiographs, such as exposure of the non-active side of the PSP to X-rays, resulting in an inverted radiograph; overlapping of images due to double exposure of the PSP; bending of the plate, among others. Other factors resulting from the use of the PSP, such as bite marks, scratches and peeling generate artifacts in the image, impairing radiographic interpretation. It was also observed that excessive use of the PSP affects the density and contrast of the radiographs [4]. Therefore, the correct handling of the PSP and its disposal after wear from use are fundamental to obtaining radiographic images with diagnostic quality [2].

It is emphasized that the features available in digital system software make it possible to overcome small losses in image quality, such as adjusting brightness, density, and contrast. They also provide filters that enhance specific perceptions of the radiograph, allowing for personalized image analysis and more accurate diagnoses [16].

This study found that a greater incidence of light occurs on the active side of the PSP when it is positioned towards the transparent side of the protective cover and when the scanner room light remains on. Light areas form across the entire surface of the PSP, causing partial erasure of the latent image, with emphasis on its final third, due to a slight delay in being inserted into the scanner and also because its active side faces upwards when entering the scanner. Other studies [11,17,18] also showed that there is deterioration of the radiographic image by light, causing darker saturation and decreased brightness and contrast, due to the leakage of metastable electrons. This result may induce the harmful practice of using longer X-ray exposure times to mitigate the loss in image quality [1].

However, in the present study, it was verified that by turning the room light off, with the PSP in the same type of cover, there was a 54.6% reduction in light absorption, and there was no record of greater light intensity in the final third of the PSP before entering the scanner. And when, in addition to turning the room light off, the PSP was positioned with the active side towards the opaque side of the protective cover, the reduction in light absorption was very significant, reaching 61.7%. Therefore, the manufacturer’s recommendation not to use the active side of the PSP on the transparent side of the cover should be followed. Even with this, the protection is not complete, since light can still reach the PSP lateral edges through the transparent part of the cover. These results are consistent with another study [15] that found a progressive reduction in image quality when using transparent covers associated with delayed insertion of the PSP into the scanner.

The lowest optical density indices of the light areas on the PSP were achieved in Groups 5, 6, and 7, in which the PSP was used in a completely opaque protective cover. However, the optical density was still high (49.95) when the room light was left on. By combining the opaque protective cover with reduced light in the room, the optical density of the light areas dropped to 17.4, representing an 88.7% reduction in light incidence. These results are consistent with previous studies, in which exposure of the plate to ambient light resulted in lower image quality, proportional to the exposed period [7,15]. One of these studies also reported that the opaque cover provides greater protection to the PSP, when compared to the transparent cover [15]. This fact reinforces the recommendation for manufacturers to return to supplying fully opaque protective covers.

In the present study, in the group in which an opaque protective cover was used in conjunction with ambient light off and covering the scanner with a cardboard box, the lowest mean optical density was found, with a 90.2% reduction in light absorption by the PSP, confirming that there is a slight interference from the computer monitor light. However, no articles in the literature were found that have studied this interference. There was likewise no evidence found that manufacturers are offering scanners with opaque light-shielding boxes or covers. Therefore, further studies are necessary to confirm the results found.

CONCLUSION

Stronger ambient light and the use of a protective cover with one transparent side significantly increase the interference of light on the latent image in the PSP. Even if the active side of the PSP is left towards the opaque side of the cover and the brightness is reduced near the scanner, the interference of light on the image is still significant, as this cover allows light to enter, from the execution of the radiographic technique until its insertion into the scanner.

The results of the present study reinforce the instruction that image processing should be performed in environments with reduced light, that fully opaque PSP covers are preferable to covers with one transparent side, and that scanners should have protection against computer monitor light. It is also indicated that the time elapsed between image acquisition and digitization should be as short as possible, so that the quality of the radiographic image is maintained.

Fagundes WS, Rachid LA, Rodrigues VA, Santos GNM, Silveira MMF, Rodrigues CD. Ambient light interference in a digital radiography system with phosphor plate. RGO, Rev Gaúch Odontol. 2026;74:e20260006. http://dx.doi.org/10.1590/1981-86372026000620250044

Data Availability

The research data are available in the body of the document.

REFERENCES

  • 1 Ramamurthy R, Canning CF, Scheetz JP, Farman AG. Impact of ambient lighting intensity and duration on the signal-to-noise ratio of images from photostimulable phosphor plates processed using DenOptix and ScanX systems. Dentomaxillofac Radiol. 2004; 33(5):307-11. doi: https://doi.org/10.1259/dmfr/91373164
    » https://doi.org/10.1259/dmfr/91373164
  • 2 Elkhateeb SM, Aloyouny AY, Omer MMS, Mansour SM. Analysis of photostimulable phosphor image plate artifacts and their prevalence. World J Clin Cases. 2022;10(2):437-47. doi: https://doi.org/10.12998/wjcc.v10.i2.437
    » https://doi.org/10.12998/wjcc.v10.i2.437
  • 3 Wenzel A, Møystad A. Work flow with digital intraoral radiography: a systematic review. Acta Odontol Scand. 2010;68(2):106-14. doi: h ttps://doi.org/10.3109/00016350903514426
    » https://doi.org/10.3109/00016350903514426
  • 4 Ruiz DC, Gomes AF, Fontenele RC, Haiter Neto F, Freitas DQ, Groppo FC. Could the radiographic image quality be affected by the excessive use of the photostimulable phosphor plate? Braz Dent J. 2023;34(1):39-44. doi: https://doi.org/10.1590/0103-6440202305225
    » https://doi.org/10.1590/0103-6440202305225
  • 5 Melo DP, Pontual ADA, De Almeida SM, Campos PSF, Alves MC, Tosoni GM. Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality. Web Sci. 2010 [cited 2025 Feb 3];39(1):23-7. Available from: http://acervodigital.unesp.br/handle/11449/15957
    » http://acervodigital.unesp.br/handle/11449/15957
  • 6 Akdeniz BG, Gröndahl HG. Degradation of storage phosphor images due to scanning delay. Dentomaxillofac Radiol. 2006;35(2):74-7. doi: https://doi.org/10.1259/dmfr/51534985
    » https://doi.org/10.1259/dmfr/51534985
  • 7 Sampaio-Oliveira M, Marinho-Vieira LE, Wanderley VA, Ambrosano GMB, Pauwels R, Oliveira ML. How does ambient light affect the image quality of phosphor plate digital radiography? A quantitative analysis using contemporary digital radiographic systems. Sensors (Basel). 2022;22(22). doi: https://doi.org/10.3390/s22228627
    » https://doi.org/10.3390/s22228627
  • 8 Van Der Stelt PF. Better imaging: the advantages of digital radiography. J Am Dent Assoc. 2008;139 Suppl:S7-13. doi: https://doi.org/10.14219/jada.archive.2008.0357
    » https://doi.org/10.14219/jada.archive.2008.0357
  • 9 Candeiro GTM, Bringel ASF, Vale IS. Radiologia digital: revisão de literatura. Rev Odontol Araçatuba. 2009 [citado 2025 fev 3];30(2):38-44. Disponível em: http://apcdaracatuba.com.br/revista/Volume_30_02_2010/trabalho 7.pdf
    » http://apcdaracatuba.com.br/revista/Volume_30_02_2010/trabalho 7.pdf
  • 10 White S, Pharoah M. Oral radiology: principles and interpretation. 7th ed. New York: Mosby; 2014.
  • 11 Eskandarloo A, Yousefi A, Soheili S, Ghazikhanloo K, Amini P, Mohammadpoor H. Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates. Open Dent J. 2017;11(1):690-700. doi: https://doi.org/10.2174/1874210601711010690
    » https://doi.org/10.2174/1874210601711010690
  • 12 Sampaio-Oliveira M, Marinho-Vieira LE, Haiter-Neto F, Freitas DQ, Oliveira ML. Ambient light exposure of photostimulable phosphor plates: is there a safe limit for acceptable image quality? Dentomaxillofac Radiol. 2023;52(7):20230174. doi: https://doi.org/10.1259/dmfr.20230174
    » https://doi.org/10.1259/dmfr.20230174
  • 13 De Moura G, Vizzotto MB, Da Silveira Tiecher PF, Arús NA, Da Silveira HLD. Benefits of using a photostimulable phosphor plate protective device. Dentomaxillofac Radiol. 2021;50(6):20200339. doi: https://doi.org/10.1259/dmfr.20200339
    » https://doi.org/10.1259/dmfr.20200339
  • 14 Sampaio-Oliveira M, Marinho-Vieira LE, Barros-Costa M, Oliveira ML. Can Digital Enhancement Restore the Image Quality of Phosphor Plate-Based Radiographs Partially Damaged by Ambient Light? J imaging informatics Med. 2024;37(1):145-50. doi: https://doi.org/10.1007/s10278-023-00922-w
    » https://doi.org/10.1007/s10278-023-00922-w
  • 15 Bramante CM, Bramante AS, De Souza RE, Moraes IG, Bernardineli N, Garcia RB. Evaluation of the effects of processing delays and protective plastic cases on image quality of a photostimulable phosphor plate system. J Appl Oral Sci. 2008 [cited 2025 Feb 3];16(5):350-4. Available from: https://www.scielo.br/j/jaos/a/6kzfks7b3sgCpFSvkNzzQbp/?lang=en
    » https://www.scielo.br/j/jaos/a/6kzfks7b3sgCpFSvkNzzQbp/?lang=en
  • 16 Watanabe P, Arita E. Imaginologia e Radiologia Odontológica. 2a ed. Rio de Janeiro: Elsevier; 2019 [citado 2025 Feb 16]. Disponível em: https://integrada.minhabiblioteca.com.br/#/books/9788595150829/
    » https://integrada.minhabiblioteca.com.br/#/books/9788595150829/
  • 17 Garcia AN. Influência da luz ambiente na qualidade de imagem de sistemas radiográficos digitais de placa de fósforo [trabalho de conclusão de curso]. Piracicaba: Universidade Estadual de Campinas; 2018. Disponível em: https://repositorio.unicamp.br/Busca/Download?codigoArquivo=511123
    » https://repositorio.unicamp.br/Busca/Download?codigoArquivo=511123
  • 18 Molteni R. Effect of visible light on photo-stimulated-phosphor imaging plates. Computer assisted radiology and surgery. 2003 [cited 2025 Feb 16];1256:1199-205. doi: https://doi.org/10.1016/S0531-5131(03)00345-5
    » https://doi.org/10.1016/S0531-5131(03)00345-5

Correspondence

Correspondence to: GNM Santos. E-mail: nize.gal@gmail.com.

Assistant editor

Luciana Butini Oliveira

Conflict of interest

The authors declare that there are no conflicts of interest.

Publication Dates

  • Publication in this collection
    03 Apr 2026
  • Date of issue
    2026

History

  • Received
    31 Aug 2025
  • Accepted
    22 Dec 2025
location_on
Faculdade São Leopoldo Mandic Rua José Rocha Junqueira, 13, CEP: 13045-755 , Tel.: +55 (19) 3211-3689 - Campinas - SP - Brazil
E-mail: contato@revistargo.com.br
rss_feed Stay informed of issues for this journal through your RSS reader
Go to top