Abstract
In this study, we developed a Radio Frequency Identification (RFID) sensor to monitor respiratory rates. The sensor utilizes a passive RFID tag enhanced with a metamaterial-inspired structure to boost its maximum gain value. This gain increased from -0.39 dBi to 0.57 dBi. Monitoring of the respiratory rate was conducted by analyzing variations in the Received Signal Strength Indicator (RSSI), which were influenced by respiratory movements. The RSSI data collected from the RFID tag readings were processed using the Daubechies wavelet of order 5 (db5) with three levels of signal decomposition, resulting in signals with reduced noise. These measurements were particularly effective when the participant was at rest, with the RFID tag positioned over the abdomen. The maximum reading range obtained was 1.04 m. The respiratory rates derived from the RSSI variations were validated against a vital signals monitor, confirming that they fell within the expected frequency variation range of the multiparameter monitor.
Index Terms
Flexible tag; Metamaterial-inspired; Respiratory rate sensor; RFID.
I. INTRODUCTION
Vital signals are indicators of patients' health status, providing relevant information about the body's circulatory, respiratory, neural, and endocrine functions. Commonly monitored vital signals are: temperature, heart rate, blood pressure, and respiratory rate [1], [2].
Changes in vital signals, outside acceptable limits, indicate a need for intervention by health professionals. This intervention not only aids in stabilizing the patient's clinical condition but also facilitates post-intervention evaluation [2].
Among the vital signals mentioned, respiratory rate is the least recorded, despite being an important indicator of serious diseases, such as prognosis of cardiac arrest, abdominal pathology, or sepsis [1]. Monitoring the respiratory rate is also important in the assessment of the cardiovascular system of babies, since there is a variation in cardiorespiratory parameters during the feeding of babies with heart disease [3]. Variations in respiratory rhythm can be seen in pathological conditions involving the central nervous system (CNS) or in intense emotional states. In addition to the cases already mentioned, monitoring the respiratory rate can detect sudden infant death syndrome [4], chronic obstructive pulmonary disease, obstructive sleep apnea syndrome, Rett syndrome, and asthma attacks [5].
In this way, it is possible to verify the importance of monitoring this vital signal to provide a quick intervention by health professionals, especially in critical cases, in which emergency care is crucial. The use of technologies that provide this monitoring continuously is of great interest, mainly associated with patient comfort.
Some methods for monitoring respiratory functions assess the temperature of the exhaled air to detect pathologies, since it has been proven that the increase in the temperature of the air that is exhaled is related to the inflammation of the airways and the increase in vascularization that is characteristic of cancer lung [5], [6]. Some of these methods involve the use of masks or cannulas, non-contact infrared technology that requires expensive computational infrastructure and tools, or wireless systems with the sensor located in the region below the nose [6]. A wireless system was also developed whose monitoring is based on exhaled air humidity [5]. Other forms of monitoring use devices with sensors, such as a narrow chest strap, and some wireless technologies require customized Doppler radios, which are not readily available in the market [5]-[7]. Therefore, it is observed that some of these methods can cause discomfort for the patient, are expensive or are not easily accessible.
With the advancement in the area of epidermal electronics, wearable devices and the Internet of Things (IoT), there is also an advance in research for the development of devices that use this type of technology for continuous monitoring of patient's vital signals, which also allow non-invasive monitoring and provide comfort to the individual [5]-[8].
In this context, an RFID (Radio Frequency Identification) system is an alternative for monitoring the respiratory rate, since it provides wireless communication, in a continuous and non-invasive way, and the use of flexible substrates in RFID tags provides more comfort. In addition, passive RFID tags do not require an internal battery to function, which contributes to weight reduction, cost reduction, and increased lifespan.
This paper reported analysis, simulation and measurements of a flexible UHF RFID respiratory sensor with metamaterial-inspired structure in order to improve its performance.
The rest of this paper is organized as follows: Section II provides the methodology proposed to design and optimize the RFID tag; Section III presents the simulations, manufacturing and measurements of the sensor proposed; and finally, conclusions are drawn in Section IV.
II. METHODS
To achieve the proposed objectives, we defined a methodology based on computer simulations in 6 steps: (i) first, we designed the RFID tag on the flexible polyimide substrate, as detailed in [9]; (ii) then, we performed parametric analyses and optimizations on the tag to achieve a better impedance match and, consequently, a resonance frequency closer to 915 MHz, as well as to attain a higher gain; (iii) we then manufactured the optimized tag; (iv) next, we conducted measurements of the tag with and without the presence of the human body; (v) subsequently, we processed the signals obtained from reading the tag to determine the respiratory frequency; (vi) finally, we validated the results by comparing them with the data observed on a vital signals monitor.
The simulations were performed in the Ansys Electronics Desktop (AEDT). The impedance considered in the lumped port was the complex conjugate of the NXP SL3S4021 chip (Zc = 12.7-j199 Ω at 915 MHz) [10] used in manufacturing.
To make the tag, the process of corrosion of the copper-plated face was used employing an iron perchloride solution with subsequent soldering of the chip.
Regarding the measurements, in the initial tests the performance of the manufactured tag was analyzed, obtaining its readings range through the reader.
To carry out measurements with the human body, the project was submitted to the Research Ethics Committee of the Alcides Carneiro University Hospital (CEP-HUAC), Brazil, for evaluation, and, after approval, according to opinion number 3.742.182 and Presentation Certificate for Ethical Appreciation (CAAE) number 21811419.3.0000.5182, we started the data collection.
Respiratory rate monitoring was based on Received Signal Strength Indicator (RSSI) variations with respiratory movements. In this way, the breathing data were obtained from the antenna connected to the RFID reader, which receives the response signals sent by the tag, decodes them and sends them to a middleware, and from these signals, one can obtain the respiratory rate value.
The tests were performed with a tag over the subject's abdomen for 2 minutes each. The tag was glued to the body using a thin double-sided tape that does not harm the participant's physical integrity. The participant was at rest, that is, had not practiced physical activity a few hours before. In addition, these measurements were performed simultaneously with the vital signals monitor for further validation.
The signals received by the reader from the reading of the RFID tag were processed to filter the noise, using the Daubechies wavelet of order 5 (db5) and considering 3 levels of signal decomposition, which allowed the calculation of the respiratory rate. The results obtained, after processing, were then validated employing comparison with the results obtained through a multiparametric monitor.
III. PASSIVE RFID TAG FOR RESPIRATORY RATE MONITORING
In this section, we present the sensor used for monitoring respiratory rate, including (1) simulation results of the RFID tags, (2) the manufactured tag, (3) the experimental setup, (4) and the results of measurements for the maximum reading range of the tag and RSSI, before and after signal processing, to obtain this frequency, with discussions of these results.
A. Simulations
The RFID tag initially simulated was a dipole antenna folded on the flexible polyimide substrate, which can provide comfort to the individual and, as it is resistant to high temperatures, allows the welding of the RFID chip. The thickness of the polyimide is 0.05 mm, the loss tangent (tan δ) is 0.004 and the relative electrical permittivity (ɛr) is 4, as presented in [11]. One of the faces of the polyimide is constituted by a thin layer of 0.035 mm thick copper, corresponding to the conductive layer. The designed RFID tag model is shown in Fig. 1.
View of RFID tag without metamaterial-inspired structure with its respective layers and dimensions.
Parametric analyses and optimizations were performed, in which the dimensions indicated in Fig. 1 were varied to obtain a better impedance matching and, consequently, a resonance frequency closer to 915 MHz, which is the operating frequency of the system. The dimensions obtained are shown in Table I.
With this structure, a reflection coefficient (S11) of -31.08 dB was reached at the frequency of 915 MHz, as can be seen in Fig. 2. Regarding the bandwidth, which comprises the frequency range in which the S11 value ≤ - 10 dB, it is noted that it includes both the American (902-928 MHz) and European (865.6-867.6 MHz) UHF (Ultra High Frequency) frequency bands [12], since the S11 value is less than -10 dB over the entire frequency range between 700 MHz and 1.10 GHz.
The 3D and 2D radiation patterns of this RFID tag are presented in Fig. 3. A quasi-omnidirectional pattern is shown in Fig. 3(a), which is typical for a classic dipole antenna. In Fig. 3(b), the planes ϕ = 0º and θ = 90º can be observed. In both figures, it can be seen that the gain values were negative. According to the antenna parameters obtained in the simulation results, the maximum gain value is -0.39 dBi, the maximum directivity value is 1.35 dB and the radiation efficiency is 66.91%.
RFID tag radiation patterns without metamaterial-inspired structure: (a) 3D; (b) 2D (angles in degrees).
The maximum read range of this RFID tag (r) was calculated using the Friis transmission equation [13]:
where λ is the wavelength, Pr is the power transmitted by the reader, Gr is the reader antenna gain, Ga is the RFID tag antenna gain, Pmin is the minimum power to activate the tag chip (-18 dBm for NXP's SL3S4011 chip [10]) and τ is the power transmission coefficient, which can be obtained from the reflection coefficient (S11) as presented in (2) [14].
Considering the American frequency band, in which the PrGr product is 4 W [13], a maximum range of 12.53 meters was obtained for the operating frequency of 915 MHz.
To improve the performance of the previous RFID tag (Fig. 1), mainly the value of its gain, a metamaterial-inspired structure was inserted, as shown in Fig. 4. This tag consists of a CSRRs (Complementary Split Ring Resonators) cell on two sides of the dipole antenna folded. In Fig. 4, the dimensions of the CSRRs cell are indicated, and designed following the structure outlined in [15].
Due to the insertion of this metamaterial-inspired structure, there was a change in the resonance frequency of the RFID tag, making new adjustments necessary in its dimensions to obtain a resonance frequency close to 915 MHz. The new dimensions of the bent dipole antenna and the dimensions of the CSRRs cells are shown in Table II. The distance between the bent dipole antenna and the CSRR cell is 1 mm. The total dimension of this tag is 27.1 x 115 mm2.
With this tag, a reflection coefficient of -31.12 dB was obtained at a frequency of 915 MHz (Fig. 5). Furthermore, as with the RFID tag without metamaterial-inspired structure, its bandwidth includes the American and European UHF frequency bands, since the value of S11 is less than -10 dB between 865.6 MHz and 928 MHz, whose values correspond to the lower limit of the European UHF frequency band and the upper limit of the American UHF frequency band, respectively.
The 3D and 2D radiation patterns of the tag are shown in Fig. 6. In Fig. 6(a) a quasi-omnidirectional pattern can also be observed. With this tag, it is verified that it was possible to obtain positive gain values with the inclusion of the metamaterial-inspired structure. According to the antenna parameters, the maximum gain value is 0.57 dBi, the maximum directivity value is 2.18 dB and the radiation efficiency is 68.94%, observing an increase in the value of the directivity and gain compared to the tag without metamaterial-inspired structure.
In addition, we observed the average value of the Specific Absorption Rate (SAR) to certify that the SAR limits specified by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) were met. The values of these limits for localized SAR on the trunk in the frequency range between 10 MHz and 10 GHz, in which the operating frequency of this work (915 MHz) is inserted, are 10 W/kg for occupational exposure and 2W/kg for general public exposure [16].
The maximum value of the average SAR verified in the human torso phantom is 1.57 W/kg, as indicated in Fig. 7. It is noted that the value obtained does not exceed the limits established by the ICNIRP for localized SAR on the trunk.
Average SAR behavior in the human torso phantom considering the RFID tag with two CSRR cells over the abdomen.
The maximum read range of this RFID tag acquired by Eq. (1) is 14 meters for 915 MHz.
Thus, comparing the results achieved, it was possible to observe an improvement in some parameters with the antenna based on metamaterial, such as directivity and gain. The directivity of 1.35 dB of the RFID tag without metamaterial-inspired structure increased to 2.19 dB with the tag with two CSRRs cells. The maximum gain of -0.39 dBi of the tag without metamaterial-inspired structure increased to 0.57 dBi with the tag with two CSRRs cells. This evidences the benefit of applying this metamaterial-inspired structure to improve the performance of tags.
Regarding the range, it is observed that, based on the simulation data, all tags had a maximum range greater than 10 meters.
Furthermore, in all simulated structures, the frequency range in which S11 ≤ -10 dB, which corresponds to the bandwidth, included both the American and European UHF frequency bands.
A survey of other passive RFID tags based on metamaterial structure was carried out, and it is possible to obtain Table III which presents a comparison with the RFID tag with proposed metamaterial-inspired structure. According to the data in this table, it is observed that the proposed tag has a greater range compared to the other indicated RFID tags.
COMPARISON BETWEEN THE RFID TAG WITH METAMATERIAL-INSPIRED STRUCTURE OF THIS WORK WITH SOME FOUND IN THE LITERATURE
B. Confection
The manufacturing of the tag with metamaterial-inspired structure was carried out using the corrosion process of the copper face of the polyimide laminate, containing the layout of the tag, by means of an iron perchloride solution and, later, the chip has been soldered, as mentioned in the methodology section. The tag prototype is shown in Fig. 8, from which the adhesive used in the corrosion process was not completely removed (white color) to minimize copper oxidation.
C. Experiment Setup
We carried out the initial tests with this manufactured tag to verify its operation and obtain the maximum read range of the tag through the RFID reader. The configuration of this measurement is present in Fig. 9.
To obtain the maximum read range, we changed the distance r between the RFID tag and the antenna connected to the Intermec IF2 network reader. The antenna connected to the reader, which was used in all measurements, was the Laird PAL90209H antenna, whose gain is 9 dBic (decibels concerning an isotropic antenna with circular polarization) and operates in the frequency range of 902-928 MHz, conforming to the 915 MHz operating frequency for which the tag is designed.
Subsequently, we performed measurements with the tag on the human body, considering the participant at rest. In these measurements, the position of the tag on the abdomen was changed between the positions indicated in Fig. 10 to observe in which position it would be possible to read the tag with a greater range. Position 1 refers to the upper edge of the tag placed just below the sternum, position 2 with it placed four fingers below the sternum and position 3 with its lower edge being three fingers above the umbilicus.
In these tests, RSSI measurements and measurements acquired by Lifemed's LIFEtouch.10 vital signals monitor were performed simultaneously for validation of the results obtained from RSSI variations.
The configuration used in these measurements is presented in Fig. 11 for the tag in position 1 over the abdomen.
D. Experimental Results and Discussions
With the initial setup of Fig. 9, which does not consider the presence of the human body, we were able to achieve a maximum read range of 0.975 m. However, the maximum reading range obtained in each of the configurations that consider the presence of the human body is presented in Table IV. In this table, it can be observed that the configuration with the tag in position 3 provided the greatest range.
Also, according to Table IV, it can be seen that all distances were greater than 20 cm, thus respecting the requirement of the Federal Communications Commission (FCC) of a minimum distance of 20 cm between the transmitting antenna and the human body [20].
The difference between the maximum read ranges obtained in the measurements and the value obtained through the simulation data may be due to flaws in the manual manufacturing process, such as the manufacture of the adhesive, the cutting of the substrate, in addition to the iron perchloride that can have penetrated the edges of the adhesive during the corrosion process. Other factors that may interfere are the quality and accuracy of the measurement equipment and the electromagnetic interference present in the measurement environment.
The signals obtained with these measurements were quite noisy and it was not possible to identify the RSSI variations with respiratory movements without the signal having been previously processed, which is aggravated by the fact that the signal is of low power.
For noise attenuation, the Wavelet Toolbox™ of the MATLAB® software was used, more specifically the Wavelet Signal Denoiser application, which is widely used in the processing of biomedical signals, being able to efficiently represent signals that present localized changes [21]-[23]. In this tool, it is necessary to identify the type of mother wavelet to be used and the level of signal decomposition, being chosen the Daubechies wavelet of order 5 (db5) with 3 levels of decomposition because it was presented as one of the best wavelets mothers by the SWTBWS (Stationary Wavelet Transform Based Wavelet Selection) method. This method uses a single mother wavelet at all levels of decomposition and selects the filter that provides the maximum absolute value among the calculated coefficients of the Stationary Wavelet Transform of the pulse [24].
As an example, it can be seen in Fig. 12 the signal obtained in the measurement with the RFID tag placed in position 1. In the figure, it can be seen that, after noise attenuation, it is possible to identify the variation in RSSI due to respiratory movements.
Although the measurements were performed over a period of 2 minutes, only an interval of 1 minute was considered, since the respiratory rate corresponds to the number of respiratory cycles in 1 minute. The results achieved after processing the RSSI data are shown in Fig. 13.
According to Fig. 13, for the RFID tag at position 1, approximately 18 respiratory cycles were observed, that is, the respiratory rate was 18 rpm, which corresponds to the number of peaks noticed in the signal. For the RFID tag in position 2, a respiratory rate of 15 rpm was obtained. As for the tag in position 3, a respiratory rate of 16 rpm was achieved.
In this way, from the results presented, it was possible to obtain a better result in the measurement with the tag in position 1, as it was possible to better identify the respiratory cycles in the signal acquired in this configuration.
To validate the respiratory rates obtained from the RSSI variations, the measurement was performed simultaneously with a vital signals monitor, as mentioned. Table V shows the values acquired from the RSSI variations and the frequency range in which the respiratory rate visualized through the signal monitor varied during the measurement. It is verified that the frequency obtained from the RSSI data is within the range of variation of the frequency acquired through the multiparameter monitor, thus validating the results of these measurements.
IV. CONCLUSION
In this work, we presented a passive RFID sensor for monitoring the respiratory rate. Continuous monitoring of this signal is essential in certain health conditions, allowing rapid intervention by professionals in the field. The relevance of the application of wireless systems in this monitoring was shown, since it provides greater comfort to the individual, and RFID technology can be an alternative to existing systems because it is able to carry out this monitoring in a continuous, practical, and comfortable way through the use of flexible substrates, such as polyimide, and without the need to use batteries in the case of using passive RFID tags.
In this way, we conducted the simulation of a passive RFID tag with a metamaterial-inspired structure, validating the increase in reading range after applying this structure. With it, we were able to measure respiratory frequency from the variation in RSSI data after signal processing. These frequencies were validated by comparison with results obtained from a vital signals monitor. As next steps, we need to increase the tag's reading range and perform measurements with different volunteers.
ACKNOWLEDGMENT
The authors would like to thank COPELE/UFCG, CNPq, CAPES and INCT NAMITEC for their institutional and financial support, the measurements volunteer, and Simões who works at Embedded/DEE/UFCG for soldering the RFID tag chip, without which this paper would not be possible.
REFERENCES
-
[1] M. A. Cretikos, R. Bellomo, K. Hillman, J. Chen, S. Finfer and A. Flabouris, “Respiratory rate: the neglected vital sign”, The Medical Journal of Australia, vol. 188, n. 11, pp. 657-659, 2008. DOI: 10.5694/j.1326-5377.2008.tb01825.x.
» https://doi.org/10.5694/j.1326-5377.2008.tb01825.x. - [2] P. A. Potter and A. G. Perry, “Fundamentos de Enfermagem”, 9th ed., Elsevier, 2013.
-
[3] V. S. G. Miranda, P. C. Souza, C. L. Etges and L. R. Barbosa, “Parâmetros cardiorrespiratórios em bebês cardiopatas: variações durante a alimentação”, CoDAS, vol. 31, n. 2, pp. 1-6, 2019. DOI: 10.1590/2317-1782/20182018153.
» https://doi.org/10.1590/2317-1782/20182018153. -
[4] R. Zhao, D. Wang, Q. Zhang, H. Chen and A. Huang, “CRH: a contactless respiration and heartbeat monitoring system with COTS RFID tags,” 2018 15th Annual IEEE International Conference on Sensing, Communication, and Networking (SECON), Hong Kong, China, pp. 1-9, 2018. DOI: 10.1109/SAHCN.2018.8397132.
» https://doi.org/10.1109/SAHCN.2018.8397132. -
[5] M. C. Caccami, C. Miozzi, M. Y. S. Mulla, C. Di Natale and G. Marrocco, “An epidermal graphene oxide-based RFID sensor for the wireless analysis of human breath,” 2017 IEEE International Conference on RFID Technology & Application (RFID-TA), Warsaw, Poland, pp. 191-195, 2017. DOI: 10.1109/RFID-TA.2017.8098641.
» https://doi.org/10.1109/RFID-TA.2017.8098641. - [6] C. Occhiuzzi, C. Caccami, S. Amendola and G. Marrocco, “Breath-monitoring by means of epidermal temperature RFID sensors,” 2018 3rd International Conference on Smart and Sustainable Technologies (SpliTech), Split, Croatia, pp. 1-4, 2018.
-
[7] Y. Hou, Y. Wang and Y. Zheng, “TagBreathe: monitor breathing with commodity RFID systems,” 2017 IEEE 37th International Conference on Distributed Computing Systems (ICDCS), Atlanta, GA, USA, pp. 404-413, 2017. DOI: 10.1109/ICDCS.2017.76.
» https://doi.org/10.1109/ICDCS.2017.76. -
[8] M. Elgeziry, F. Costa, A. Tognetti and S. Genovesi, “Wearable textile-based sensor tag for breath rate measurement,” IEEE Sensors Journal, vol. 22, no. 23, pp. 22610-22619, 2022. DOI: 10.1109/JSEN.2022.3217325.
» https://doi.org/10.1109/JSEN.2022.3217325. - [9] J. I. L. Araujo, S. M. A. Morais, C. C. R. Albuquerque, G. K. F. Serres, H. A. Pereira, A. J. R. Serres and D. F. S. Santos, "Etiqueta RFID UHF flexível para aplicações biomédicas baseada em estruturas metamateriais", 19° Simpósio Brasileiro de Micro-ondas e Optoeletrônica e 14° Congresso Brasileiro de Eletromagnetismo, pp. 659-663, 2020.
-
[10] NXP Semiconductors. “SL3S4011_4021: UCODE I2C”. https://datasheet.octopart.com/SL3S4011FHK%2C125-NXP-Semiconductors-datasheet-82458551.pdf
» https://datasheet.octopart.com/SL3S4011FHK%2C125-NXP-Semiconductors-datasheet-82458551.pdf -
[11] A. S. F. Evangelista, P. F. Silva, E. E. C. Santanta, M. S. P. Silva, R. C. S. Freire, A. J. R. Serres, J. F. Araújo, M. A. Oliveira and P. H. F. Silva, "Wearable Koch Pre-Fractal Antennas for Ultrahigh Frequency Band," 2019 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Auckland, New Zealand, pp. 1-5, 2019. DOI: 10.1109/I2MTC.2019.8827040.
» https://doi.org/10.1109/I2MTC.2019.8827040. -
[12] M. Riaz, G. Rymar, M. Ghavami and S. Dudley, “A novel design of UHF RFID passive tag antenna targeting smart cards limited area,” 2018 IEEE International Conference on Consumer Electronics (ICCE), Las Vegas, NV, USA, pp. 1-4, 2018. DOI: 10.1109/ICCE.2018.8326224.
» https://doi.org/10.1109/ICCE.2018.8326224. -
[13] I. Aznabet, M. Ennasar, O. El Mrabet, S. Tedjini and M. Khalladi, “Meander-line UHF RFID tag antenna loaded with split ring rersonator,” 2016 5th International Conference on Multimedia Computing and Systems (ICMCS), Marrakech, Morocco, pp. 757-759, 2016. DOI: 10.1109/ICMCS.2016.7905576.
» https://doi.org/10.1109/ICMCS.2016.7905576. -
[14] F. Paredes, G. Zamora, T. M. Nguyen, F. Martín and J. Bonache, “A novel design strategy for small on-metal UHF-RFID tags with long read range based on Complementary Split-Ring Resonator (CSRR),” 2018 48th European Microwave Conference (EuMC), Madrid, Spain, pp. 985-989, 2018. DOI: 10.23919/EuMC.2018.8541810.
» https://doi.org/10.23919/EuMC.2018.8541810. -
[15] G. V. R. Xavier, A. J. R. Serres, E. G. Costa, A. C. Oliveira, L. A. M. M. Nobrega and V. C. Souza, “Design and Application of a Metamaterial Superstrate on a Bio-Inspired Antenna for Partial Discharge Detection through Dielectric Windows”, Sensors 2019, vol. 19, no. 19, pp 1-27, 2019. DOI: 10.3390/s19194255.
» https://doi.org/10.3390/s19194255. - [16] International Commission on Non-Ionizing Radiation Protection, “Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz)”, Health Physics, vol. 74, n. 4, pp. 494-522, 1998.
-
[17] N. H. Daud, A. Ismail, A. R. H. Alhawari, A. Sali, M. F. A. Rasid and A. S. A. Jalal, “Integration of split ring resonators (SRRs) to UHF RFID tag antenna for size reduction,” 2014 4th International Conference on Engineering Technology and Technopreneuship (ICE2T), Kuala Lumpur, Malaysia, pp. 204-208, 2014. DOI: 10.1109/ICE2T.2014.7006247.
» https://doi.org/10.1109/ICE2T.2014.7006247. -
[18] A. Ennajih, J. Zbitou, A. Errkik, A. Tajmouati, L. El Abdellaoui and M. Latrach, “A novel design of passive UHF RFID tag antenna mounted on paper,” 2017 International Conference on Wireless Technologies, Embedded and Intelligent Systems (WITS), Fez, Morocco, pp. 1-6, 2017. DOI: 10.1109/WITS.2017.7934631.
» https://doi.org/10.1109/WITS.2017.7934631. -
[19] Y. J. Zhang, D. Wang and M. S. Tong, “Metamaterial UHF RFID tag with enhanced gain,” 2017 Progress in Electromagnetics Research Symposium - Fall (PIERS - FALL), Singapore, pp. 1845-1848, 2017. DOI: 10.1109/PIERS-FALL.2017.8293437.
» https://doi.org/10.1109/PIERS-FALL.2017.8293437. -
[20] Federal Communications Commission, “FCC review of RF exposure policies”, 2013. https://www.fcc.gov/document/fcc-review-rf-exposure-policies
» https://www.fcc.gov/document/fcc-review-rf-exposure-policies - [21] M. Weeks, “Digital signal processing using MATLAB® and wavelets”, Infinity Science Press LLC, 2007.
- [22] D. Baleanu (Ed.), “Advances in wavelet theory and their applications in engineering, physics and technology”, IntechOpen, 2012.
- [23] K. Najarian and R. Splinter, “Biomedical signal and image processing”, 2nd ed., CRC Press, 2012.
-
[24] A. T. Carvalho, A. C. S. Lima, C. F. F. C. Cunha, M. Petraglia, “Identification of partial discharges immersed in noise in large hydro-generators based on improved wavelet selection methods”, Measurement, vol. 75, pp. 122-133, 2015. DOI: 10.1016/j.measurement.2015.07.050.
» https://doi.org/10.1016/j.measurement.2015.07.050.