Abstract
This manuscript presents the feasibility of designing a digital Lock-In Amplifier (LIA) for measuring experimental sys-tems through optical measurement techniques, from 32-bit micro-controller platforms, derived from the technology coming from the Internet of Things (IoT). As a proof of concept, the STM32F446RE microcontroller was used, in which the resident program uses the Direct Digital Synthesis (DDS) and synchronous detection tech-nique to promote lock-in processing from harmonic signals gener-ated in 2f. The digital LIA was tested to detect C2H2 (Acetylene) concentrations in a spectrometric cell by Wavelength Modulation Spectroscopy (WMS). The experimental setup also included a DFB laser and other optical components. The comparative analysis with the Signal Recovery® DSP7265 benchtop LIA showed that both systems effectively measured up to 54 ppm of C2H2. A strong Pearson correlation (r>0.8) was observed for all measurements performed. This confirms its potential for applications in situ nonlinear optical spectroscopy, especially WMS.
Index Terms
Gas sensor; lock-in amplifier; microcontroller; wavelength mod-ulation spectroscopy.
I. INTRODUCTION
Lock-in Amplifiers (LIAs) play a crucial role in a wide range of optical science instruments used in research laboratories. These devices are essential in experimental setups that require the detection and recording of extremely small or weak signals, often obscured by high levels of noise [1], [2]. Thanks to their ability to selectively filter and amplify signals of a specific frequency, it is possible to monitor signals of interest accurately and reliably, even in noisy environments. With the advent of technologies related to Embedded Systems and the Internet of Things (IoT) [3], [4], there is a wide offer of microcontroller and/or microprocessor platforms for LIA projects with a focus on remote sensing systems.
Optical applications where these devices are employed include various forms of optical spectroscopy, such as absorption spectroscopy [5], [6], fluorescence [7]-[9], photoacoustic [9]-[12] and Wavelength Modulation Spectroscopy (WMS) [13]-[16], where LIAs are essential to detect subtle changes in light intensity. In optical interferometry, LIAs play a critical role in improving the resolution and accuracy of measurements by filtering out noise and isolating the desired signal [17]. The technology is also essential in crystal research, particularly in the exploration of magneto-optical [18], electromechanical [19], and electro-optical effects [20], where the precision of lock-in amplifiers allows detailed study of material properties. Additionally, these amplifiers are used in infrared thermography [21], [22], where they help accurately measure temperature distributions, and in laser wavelength stabilization, ensuring that laser systems maintain their precise operating frequency. Collectively, these applications underscore the versatility and indispensability of lock-in amplifiers in modern optical science and technology.
LIA designs can be based on the development of analog circuits and the use of programmable devices on digital and computer platforms. Analog LIAs are commonly used in instrumentation systems, for the purpose of extracting weak signals from sensors and transducers. Analog LIAs are effective in electrical impedance measurements. Practical applications of these devices include their use in Catalytic Gas Sensors (CGS), which detect and measure gas concentrations with high sensitivity due to varying resistance [23], [24] or capacitance [25], [26] of the sensing element.
Digital LIAs based on digital platforms have a wider range of applications. This is essentially because this class of LIA outperforms analog models in all relevant performance parameters, such as frequency range, input noise, stability and dynamic reserve. The use of microcontrollers, FPGA (Field-Programmable Gate Array) and computers is verified. In general, FPGAs have significantly higher processing power than microcontrollers and, consequently, are more expensive. However, mi-crocontrollers can be perfectly suited for operations that require operating frequencies in the order of a few kilohertz. Regarding microcontrollers, there is a wide offer of development platforms. For example, in [7], the ATMEGA128, which is present in the Arduino platform, was used to develop a portable LIA for photoluminescence measurement, while in [27], the MSP430 is used to implement a digital LIA for applications such as impedance measurement experiments, signal recovery, and accurate measurement of phase shifts between electrical signals. Another digital platform widely used for digital LIA projects is the FPGA. As an example, a model Xilinx FPGA was employed in [28] to conduct optoelectronic experiments involving radiation emitted by black bodies, while the RedPitaya platform was utilized by [29] to construct a LIA capable of performing demodulation at frequencies up to 50 MHz. Another LIA implementation in FPGA was presented by [30], specifically developed for applications in photoacoustic spectroscopy aimed at gas detection. LIAs can also be implemented as resident programs on a commercial computer, as demonstrated in [31], where data obtained from an Analog-to-Digital Converter (ADC) are used in nonlinear spectroscopy experiments, and in [32], in which WMS experiments were performed to detect atmospheric gas concentrations using a lock-in amplifier developed with LabVIEW software. Similarly, in [33], software was developed in C# to perform lock-in processing on each pixel of a CCD spectrometer, allowing in situ optical spectroscopy applications. Table I presents a summary of the state of the art regarding the development of lock-in amplifiers based on microcontrollers and FPGAs.
A viable optical technique for implementation with IoT devices is WMS. This also occurs with the wide supply of DFB lasers (Distributed Feedback Laser), OEM laser controllers and photodetectors from the telecommunications industry. An important application for this technique is to detect the concentration of gases dissolved in high-voltage transformer oil [15], [34]-[36]. Due to the heating of the active part resulting from the dielectric losses of the core and also due to small discharges existing there, the molecular chains of the oil and cellulose paper break [37], [38]. As a result of these reactions, there is an increase in the concentrations of different gases. Small electrical discharges generate H2, CO and CO2 (hydrogen, carbon monoxide, carbon dioxide) [39]. With the increase in the generation rate of these by-products, high-energy discharges occur, generating an increase in the concentration of other gases such as CH4 and C2H2 methane and acetylene) [40]. The use of LIA for this technique is essential due to the low direct absorption of these gases with photons in the near-infrared region (1.2 - 1.6µm ), in addition to the nonlinear interaction of these photons with the rotational/vibrational movements of these gases [36].
Given this scenario, the objective of this manuscript is to present a proposal for a digital LIA based on the STM32F446RE microcontroller from ST Electronics® , with a low-cost resident program for applications in remote optical measurements. The chosen optical technique was WMS for the optical monitoring of C2H2, one of the byproducts of the degradation of insulating oil in high-voltage transformers. The experimental setup includes a DFB laser, a multipass spectrometric cell with an optical path of 78.6 cm, a laser controller, a photodetector and a personal computer. The performance of the proposed LIA was evaluated against the benchtop LIA DSP7265 from Signal Recovery®. Both devices successfully measured up to 27 ppm of C2H2, with a Pearson correlation above 0.8 for all measurements performed. This confirms the potential of IoT technology for the design of equipment for advanced optical spectroscopy applications to be performed in situ.
II. LOCK-IN AMPLIFIER THEORY
Fig.1 illustrates the fundamental block diagram of the digital LIA used to excite an experimental system for the purpose of analyzing its electrical and/or optical characteristics. The digital LIA com-prises a reference signal generator and employs the synchronous detection technique, correlating the response of the experimental system with the reference signal at a frequency of f .
In this digital LIA, the reference signal is produced internally through the Direct Digital Synthesis (DDS) method [41], and because of this, it´s not necessary an external modulator, as in other appli-cations [45]. This approach uses predefined discrete signal values, a TIMER, and a Digital-to-Analog Converter (DAC) as illustrated in Fig. 1. In the digital LIA, the Look-Up Table (LUT) is established in the microcontroller memory that represents a sinusoidal waveform with N discrete points. The reference sinusoidal signal Vr[n] can be expressed as:
where C is a constant and Vcc represents an offset voltage signal, providing a positive reference signal with no bits close to zero. From the Timer’s configuration, the information from the lookup table indicated in Fig.1 is transmitted to the DAC, thus generating the analog reference signal. In optical experiments, the reference signal is used to modulate the amplitude or phase of the light source. This modulation can be achieved directly by connecting the reference signal to the current driver that powers the light source, as in the case of LEDs and diode lasers [46], [47]. For other types of lasers, such as Ruby, CO2, and Nd lasers, the modulation is typically done indirectly using an electrical chopper [48].
An important aspect to highlight is that the reference signal, which modulates the experiment, is generated internally and simultaneously with the acquisition of the signal of interest. This simultaneity ensures continuous synchronization between both signals, maintaining a fixed phase difference an essential condition for the accurate implementation of synchronous detection [49].
After interaction with the experimental system, the output voltage signal, Vs(t) is acquired by the microcontroller’s ADC and stored in a vector with N positions (Vs[n]). The reading of Vs[n] and the generation of the signal Vr[n] occur in synchronization with the microcontroller’s internal clock. In optical experiments, the Vs[n] corresponds to the light signal detected by the photodetector, which results from the interaction between the light source and the photosensitive medium [5]-[8], [13]-[16].
The next step of lock-in processing is the synchronous detection technique, which consists of multiplying Vs[n] to Vr[n] in phase, Vf [n] and quadrature Vq[n]. This operation is performed by a mixer as follows [7], [27]-[29], [31]:
and
Due to the nonlinear characteristics of optical medium Vs[n] can be decomposed into a Fourier series. Therefore, it is possible to obtain the n - th harmonic of the signal of interest by adjusting the value of f via the resident program for the respective harmonic. The last step of lock-in processing is to perform low-pass filtering.
A practical approach to attenuate high-frequency components of the signal is to implement a first-order FIR filter or, alternatively, a digital IIR low-pass filter. However, given that the resulting signal primarily consists of a sinusoidal component and a DC component, a more straightforward and efficient low-pass filtering strategy in this specific case is to apply a simple arithmetic mean of Vf [n] and Vq[n] after K acquisitions [50], [51].
Finally, one can calculate the modulus M[n] from the Eq.4:
III. WAVELENGTH MODULATION SPECTROSCOPY- WMS
The diagram in Fig.2 presents a schematic of an experimental setup developed for the application of Wavelength Modulation Spectroscopy to detect and quantify a given gas [34], [35]. WMS is typically used to measure trace gases that exhibit weak absorption in the near-infrared spectrum, operating by observing the signal at twice the reference frequency 2f [15], [52], [53]. This is due to the nonlinear characteristics of the gas signature in a given absorption wavelength range.
It consists of a DFB (Distributed Feedback) laser, which exhibits a very fine emission line. The instrumentation also includes a laser controller, which allows the laser emission line to be adjusted by means of the injection current and its temperature, a spectrometric cell and a photodetector. From the reference signal connected to the laser controller, the injection current and the laser wavelength are given respectively by:
and
which and are the DC values of current and wavelength and ∆i and ∆λ are the respective modulation amplitudes. The output intensity after transmission through an absorbing medium of length L is given by
where N represents the volume density of the molecule, σ(λ) the absorption cross section of the molecule and L the optical path length of the spectrometric cell.The cross section of a molecule can be modeled using a Lorentzian distribution function. Consequently, the absorption cross section is expressed as [15], [34], [35], [52], [53]:
where σmax is the normalized maximum value of the absorption cross section, γ is the HWHM (Half-Width-at-Half-Maximum), λ is the wavelength of the probing light and λ0 is the wavelength at the center of the absorption line. The values of σmax, λ0 and γ for the gas of interest can be extracted directly from the HITRAN database. Since the Lorentzian distribution function is quadratic, it is possible to specifically identify the existence of a certain absorption line of the gas of interest, observing the existence of a second harmonic component of the signal.
Fig. 3 shows the wavelength dependence of the second harmonic component for the case of 3500 ppm C2H2 at atmospheric pressure. The script used to plot this simulation was developed in Python and the parameters used are presented in Table II. A wavelength sweep is performed, which implies a sweep of the injected laser current. The maximum signal level occurs at λ = 1531.588910nm and is proportional to the gas concentration, and this property is used to infer the gas concentration in the environment under test.
It is worth noting that in the wavelength region between 1530 and 1570 nm, the analyzed gas has the largest absorption cross section at 1531.588910 nm, with σ equal to 4.66 × 10-20 cm2 [54].
IV. IMPLEMENTATION OF THE STM32-BASED LOCK-IN AMPLIFIER
Fig. 4 illustrates the experimental system setup used for the WMS experiment. The system incor-porates an STM32F446RE Core board with a resident program responsible for performing lock-in processing. The experimental setup also includes a Thorlabs® ITC510 laser source controller, a DFB laser operating at λ = 1531 nm at 25 °C, a spectrometric cell with a path length of 78 cm, an Edmund Optics 55757 InGaAs photodetector, fiber optic connections, signal conditioning circuitry, and a computer running LabView® 13.0 software. LabView® was used to establish a communication interface between the STM32 and the ITC510 source, also allowing the user to transmit configuration commands from the PC to the ITC510 source.
Initially, a fully positive reference sinusoidal signal with a frequency of 200 Hz and Vpp of 1.1 V (ranging from 241 mV to 1341 mV ) is generated by the STM32 using the DDS technique. This signal is then sent to the DAC output and subjected to an offset circuit, illustrated in Fig.5, to adjust its level to the values required by the ITC510 control source in order to perform the WMS experiment. The offset circuit consists of a voltage divider followed by a high-pass filter and a low-pass filter, both of first order. The voltage divider, which uses a 1 kΩ resistor and a 370 Ω resistor, reduces the amplitude of the signal generated in the DAC by 73 %. The high-pass filter is configured with a cutoff frequency of 72 Hz (C = 100 nF and R = 22 kΩ) and has the function of eliminating the DC component, making the sinusoidal signal symmetrical around zero. In turn, the low-pass filter, with a cutoff frequency of 1591 Hz (C = 100 nF and R = 1 kΩ), eliminates harmonic components that arise intrinsically in the DDS technique. The transfer functions of these filters result in an attenuation of the amplitude of the reference signal, which now has a Vpp of approximately 180 mV , corresponding to a Vrms of 63 mV .
After adjusting the level in the offset circuit, the reference signal is sent to the ITC510 source, which modulates the injection current of the DFB laser. The modulated signal emitted by the laser is guided through a section of optical fiber. It reaches the spectrometric cell and interacts with the gas present, which in this case is acetylene (C2H2). At the cell output, the laser signal is conducted through another section of optical fiber to the photodetector.
Fig. 6 illustrates the block diagram of the four circuits used to condition the signal extracted from the photodetector. In Fig.6 (a), a high-pass filter is represented connected between two buffer circuits, which are implemented using the LM358 operational amplifier. The buffers provide isolation between the stages of the circuit, while the high-pass filter performs the AC coupling. At the output of this block, the resulting signal is an unshifted sine wave. Fig. 6(b) consists of a high-sensitivity instrumentation amplifier (INA128) configured with a voltage gain equal to 24. Fig. 6(c) represents a biased clamper circuit designed to add a fixed positive DC voltage level to the signal. This circuit comprises three resistors, a diode 1N 4007, and an electrolytic capacitor rated at 10 µF/100 V . After this blocking, the signal retains its sinusoidal characteristic with a peak-to-peak voltage (Vpp) of 3 V, but switches to exhibit only positive polarity. Fig. 6(d) shows a second-order Butterworth low-pass filter designed using the LM358 op-amp with a cutoff frequency of 600Hz. This filter acts as an anti-aliasing stage, allowing proper signal acquisition by the STM32.
Conditioning Circuit. (a) buffer + high-pass filter, (b) amplifier , (c) positive clamper , (d) 2nd Order Low-Pass Filter .
A flowchart of the experimental procedure used for gas detection with the WMS technique proposed in this manuscript is presented in Fig.7. On the computer running LabVIEW® software, the user configures the laser emission wavelength range by specifying the thermistor value, the variation range of DC current (initial and final), and the increment of the DC current injected into the laser. This configuration data is transmitted via GPIO from PC to the ITC510 source. Simultaneously, in the STM32 resident program, a reference sinusoidal signal with a frequency of 200 Hz is generated and sent to the DAC. The DAC output is connected to the circuit shown in Fig. 5 and to the designated input of the ITC510, enabling amplitude modulation of the laser wavelength. The signal from the conditioning circuit, illustrated in Fig. 6, is stored in an array that is continuously updated. Once the array is fully populated with 12,800 values - corresponding to 100 complete periods of the sinusoidal signal or an integration time of 500mS- an interrupt is triggered in the STM32 resident program to execute lock-in processing. The choice of this number of acquired periods and the frequency of 200Hz was based on preliminary experiments, in which longer integration times (1s, 2s and 5s) and higher frequencies (1KHz, 2KHz, 5KHz and 10KHz) did not result in significant variations in the value of the signal obtained. The in-phase Vf [n] and quadrature Vq[n] signals, respectively defined by Eq.2 and Eq.3, are calculated for each acquired value. In this case, the reference signal Vr[n] is generated as the multiplication by Vs[n] is performed, and the frequency of Vr[n] is defined as 2f to obtain the second harmonic characteristics of the gas of interest. Then, the arithmetic mean of the 12,800 values of Vf [n] and Vq[n] is calculated, which are stored internally in the STM32. Finally, the second harmonic component module is calculated according to Eq.4. This result is sent to the personal computer and stored according to the routine developed in LabVIEW®. Once the interruption is complete, the STM32 sends a command to LabVIEW® to increase the DC current value according to the increment entered by the user. LabVIEW® updates the injection current value in the ITC510 source via the GPIB connection. The acquisition cycle is resumed until a new interruption occurs and ends when the maximum current value, previously defined by the user, is reached. If, after the increase, the DC current value exceeds the stipulated maximum limit (IDC_Final), the algorithm is terminated. Otherwise, the acquisition cycle is repeated: the Vs[n] buffer is updated with 12,800 new readings, and a new interruption is performed to calculate the second harmonic. This cycle is repeated as long as the injection DC current value is less than or equal to the IDC_Final value.
V. RESULTS AND DISCUSSION
To verify the performance of the digital LIA proposed in this manuscript, WMS experiments were performed, as illustrated in Fig.4. The results were compared with the LIA DSP7265 from Signal Recovery, in order to evaluate the accuracy, adjustment and reliability of the developed LIA. The gas used in the tests was C2H2 at a concentration of 3500 ppm in the cylinder. Initially, the spectrometric cell was completely evacuated using a vacuum pump, until the pressure gauge read approximately -600 mmHg (equivalent to 0 atm, or vacuum). Then, the cell was filled with acetylene at 1 atm and 3500 ppm, with the pressure gauge reading approximately 0 mmHg. To perform gas dilutions in the spectrometric cell, the contents were emptied until the pressure gauge indicated -360 mmHg (approximately 0.5 atm). Subsequently, the cell was filled with N2 until it reached 0 mmHg (1 atm absolute). This procedure allowed the C2H2 concentration to be reduced by half in relation to the initial concentration, according to the ideal gas equation [55].
Fig.8 presents the experimental results for concentrations of 3500, 1750, 875, 437, 218, 109, 54, and 27 ppm (Fig.8 (a) to Fig.8 (h)). It is important to highlight that, in order to compare the profiles of the second harmonic module obtained from different devices, the gain of the signal acquired by the proposed digital LIA must be adjusted to bring both signals to the same scale. This adjustment was carried out through a resident program and varied between 110 and 120, depending on the concentration. Consequently, it is possible to observe a consistent similarity in the behavior of the second harmonic modules for the same concentration when comparing the results obtained by the DSP7265 and the proposed digital LIA.
Result of the WMS performed with the proposed LIA compared with the DSP7625 for the concentrations of a) 3500 ppm, b) 1750 ppm, c) 875 ppm, d) 438 ppm, e) 217 ppm, f) 109 ppm, g) 54 ppm and h) 27 ppm.
In order to quantify the degree of similarity between the responses obtained for each concentration analyzed, a correlation study, Pearson correlation r was performed. This is a statistical measure that describes the degree of relationship between two quantitative variables. Assuming that x and y represent, respectively, the values of the second harmonic module obtained by the proposed digital LIA and the DSP7625, respectively, the coefficient r is given by [56]:
where and are respectively the mean values of x and y. The coefficient r can assume values between ±1 indicating perfect positive (negative) correlation. Strong correlation is accepted for values of r in the interval 0.8 ≤ |r| ≤ 1 [56]. Table III presents the calculated values of the correlation coefficient r for the curves shown in Fig.8 (a) to Fig.8 (h). It can be seen that the lowest values indicated in Table III are around 0.8 and occur at the concentrations of 54 and 27 ppm, in which the second harmonic modulus signal presents significant degradation. This behavior suggests a strong correlation even when the proposed digital LIA and the DSP7625 operate close to the detection limit.
The difference in the quality of the second harmonic profile at the lowest concentrations, observed in Fig.8 (f) to Fig.8 (h) and proven by the Pearson correlation, may be related to the use of generic and low-cost integrated circuits, such as the LM358, in the AC coupling and filtering stages. The use of components with greater noise immunity would possibly reduce these variations in the signal obtained by the STM32-based lock-in.
Figure 9 summarizes the relationship between the concentration of the analyzed gas and the logarithm of the second harmonic peak modulus, as illustrated in the figure itself. It can be seen that, as the concentration decreases, the value of the peak corresponding to the second harmonic also undergoes a proportional reduction. This behavior is in line with theoretical expectations, reinforcing the reliability of the experimental procedure and the consistency of the results obtained. At concentrations of 27 ppm and 54 ppm, the variation in the logarithm value becomes practically indistinguishable, indicating that, in this range, the detection limit of the experimental arrangement used is reached.
Furthermore, Fig. 9 illustrates the uncertainty associated with the measurement process, represented by the error bars. For this evaluation, the ratio between the standard deviation and the mean was used as a parameter, calculated point by point from 10 curves recorded for each concentration. The results indicate that the uncertainty is lower at higher concentrations, around 4%, while at lower concentrations, the average ratio reaches approximately 19%.
Considering the resolution of the STM32 ADC, the voltage gain of the INA, and the construction characteristics of the second-order Butterworth filter, the noise voltage and equivalent noise bandwidth of the proposed LIA are 75.47 µV/√Hz and 0.335 Hz, respectively.
Regarding the analysis of the signal-to-noise ratio (SNR) of the proposed LIA, the SNR was calculated based on the photodetector voltage signal, monitored at the highest (3500 ppm) and lowest (27 ppm) measured concentrations. This calculation considers the full-scale voltage, derived from the noise voltage and the equivalent noise bandwidth. Thus, the signal-to-noise ratio for the respective concentrations is equal to: 55.7 dB and 13.5 dB.
VI. CONCLUSION
The results presented in this manuscript demonstrate the feasibility of using IoT ecosystem devices for the development of a digital LIA dedicated to the detection of minute concentrations of gases through the WMS technique. The proposed LIA meets the requirements of accuracy, refinement, and robustness, achieving a performance equivalent to that of a benchtop LIA. Thus, it can become an integral part of complete and embedded solutions for optical spectroscopy, enabling in situ applications where the use of benchtop LIAs would be impractical. This is because the methodology used in the project has low computational cost and can be adapted to any 32-bit microcontroller platform. It is worth mentioning that it is possible to use a wide variety of optical spectroscopy techniques to obtain non-linear characteristics of samples related to the nth harmonic in relation to the reference signal. The specificities of the magnitude of the signals obtained must be adjusted using signal conditioning circuits, as performed in this manuscript.
From the development of multispectral sources, such as those described in [15] and the capacity of the digital LIA proposed here to monitor concentrations of up to 54 ppm of C2H2 using the WMS technique, the equipment described here becomes a viable alternative to be an integral part of complete and embedded solutions for real-time monitoring of the operational state of a high voltage transformer.
ACKNOWLEDGMENTS
The authors acknowledge the financial support from the National Council for Scientific and Tech-nological Development (CNPq) and Coordination for the Improvement of Higher Education Personnel (CAPES).
DATA AVAILABILITY
All files generated or used in the development of this experiment are made available upon request. The code in C format and the software generated in Labview are available on GitHub.
REFERENCES
- [1] M. L. Meade, Lock-in Amplifiers: Principles and Applications, eletronic ed. Peter Peregrinus Ltd, London, 2013.
- [2] G. Kloos, Applications of Lock-in Amplifiers in Optics, eletronic ed. SPIE, 2018.
-
[3] A. Troy, F. Paz, and M. Ordonez, “Digital regulation of wireless power transfer systems using an embedded lock-in amplifier,” IEEE Journal of Emerging and Selected Topics in Industrial Electronics, vol. 3, no. 3, pp. 421-431, 2021, DOI:10.1109/JESTIE.2021.3103678.
» https://doi.org/10.1109/JESTIE.2021.3103678. -
[4] G. Li, H. Zhao, J. Li, Y. Liu, Y. Song, X. Zhang, Z. Zhang, and Y. Wu, “A near infrared h2s leakage detection system based on wdo-elm using a digital lock-in amplifier combined with discrete wavelet transform filter,” Infrared Physics & Technology, vol. 128, p. 104481, 2023, DOI: 10.1016/j.infrared.2022.104481.
» https://doi.org/10.1016/j.infrared.2022.104481. -
[5] H. Ardekani, R. L. Wilmington, M. Vutukuru, Z. Chen, R. Brandt, K. Gundogdu, et al., “Broadband micro-transient absorption spectroscopy enabled by improved lock-in amplification,” Review of Scientific Instruments, vol. 92, no. 10, p. 104706, 2021, DOI: 10.1063/5.0060244.
» https://doi.org/10.1063/5.0060244. -
[6] Y. Zhang, P. Jiang, W. Cao, X. Li, and J. Lai, “High-sensitivity ethylene gas sensor based on ndir and dual-channel lock-in amplifier,” Optik, vol. 223, p. 165630, 2020, DOI: 10.1016/j.ijleo.2020.165630.
» https://doi.org/10.1016/j.ijleo.2020.165630. -
[7] F. Pollastrone, M. Piccinini, R. Pizzoferrato, A. Palucci, and R. M. Montereali, “Fully-digital low-frequency lock-in amplifier for photoluminescence measurements,” Analog Integrated Circuits and Signal Processing, vol. 115, no. 1, pp. 67-76, 2023, DOI: 10.1007/s10470-022-02125-9.
» https://doi.org/10.1007/s10470-022-02125-9. -
[8] A. Nadgir, R. Thurston, K. A. Larsen, N. Shivaram, M. M. Brister, and D. S. Slaughter, “Silia: software implementation of a multi-channel, multi-frequency lock-in amplifier for spectroscopy and imaging applications,” Measurement Science and Technology, vol. 32, no. 12, p. 125501, 2021, DOI: 10.1088/1361-6501/ac2436.
» https://doi.org/10.1088/1361-6501/ac2436. -
[9] X. Zhao, H. Qi, Y. Xu, C. Li, M. Guo, and K. Chen, “Fiber-optic photoacoustic gas sensing: a review,” Applied Spectroscopy Reviews, vol. 60, no. 1, pp. 1-29, 2025, DOI: 10.1080/05704928.2024.2359933.
» https://doi.org/10.1080/05704928.2024.2359933. -
[10] M. Guo, M. Zhou, B. Yang, X. Zhao, C. Li, Q. Yu, G. Zhang, Z. Fang, and K. Chen, “All-optical photoacoustic spectroscopy-based dual-component greenhouse gas analyzer,” Analytical Chemistry, vol. 96, no. 37, pp. 14 819-14 825, 2024, DOI: 10.1021/acs.analchem.4c02440.
» https://doi.org/10.1021/acs.analchem.4c02440. -
[11] C. Li, X. Han, F. Ma, X. Zhao, Z. Wang, H. Qi, M. Guo, and K. Chen, “Multiplexed fiber-optic photoacoustic sensors for simultaneous detection of multi-point gases,” Sensors and Actuators B: Chemical, vol. 399, p. 134801, 2024, DOI: 10.1016/j.snb.2023.134801.
» https://doi.org/10.1016/j.snb.2023.134801. -
[12] K. Chen, Z. Yu, Z. Gong, and Q. Yu, “Lock-in white-light-interferometry-based all-optical photoacoustic spectrometer,” Optics letters, vol. 43, no. 20, pp. 5038-5041, 2018, DOI: 10.1364/OL.43.005038.
» https://doi.org/10.1364/OL.43.005038. -
[13] A. L. Chakraborty and A. Roy, “Wavelength modulation spectroscopy,” Modern Techniques of Spectroscopy: Basics, Instrumentation, and Applications, vol. 13, pp. 321-362, 2021, DOI: 10.1007/978-981-33-6084-6_13.
» https://doi.org/10.1007/978-981-33-6084-6_13. -
[14] A. Roy and A. L. Chakraborty, “Intensity modulation-normalized calibration-free 1f and 2f wavelength modulation spectroscopy,” IEEE Sensors Journal, vol. 20, no. 21, pp. 12 691-12 701, 2020, DOI: 10.1109/JSEN.2020.3000470.
» https://doi.org/10.1109/JSEN.2020.3000470. -
[15] R. A. de Lima, V. H. Soares, J. F. Martins-Filho, and E. Fontana, “Tailoring a spectral line detection system for applications in dissolved gas analysis,” Sensors and Actuators A: Physical, vol. 293, pp. 178-188, 2019, DOI: doi.org/10.1016/j.sna.2019.04.020.
» https://doi.org/doi.org/10.1016/j.sna.2019.04.020. -
[16] E. Fontana, V. H. Soares, L. Urtiga, and J. F. Martins-Filho, “Modeling partition dynamics through multiple interfaces and characterization by wavelength modulation spectroscopy,” Sensors and Actuators A: Physical, vol. 316, p. 112339, 2020, DOI: 10.1016/j.sna.2020.112339.
» https://doi.org/10.1016/j.sna.2020.112339. -
[17] Z. Zhang, F. Wang, T. Yuan, and C. Li, “Multiple self-mixing interferometry based on lock-in amplifier analysis for vibration measurement,” Optical Review, vol. 27, no. 4, pp. 313-320, 2020, DOI: 10.1007/s10043-020-00600-0.
» https://doi.org/10.1007/s10043-020-00600-0. -
[18] D. Handoko and D. Kim, “Dynamics noise behaviors on magneto-optical kerr effect measurement system,” Indonesian Physical Review, vol. 6, no. 2, pp. 189-195, 2023, DOI: 10.29303/ipr.v6i2.231.
» https://doi.org/10.29303/ipr.v6i2.231. -
[19] M. Wu and H. P. Lu, “Ultra-sensitive lock-in amplifier coupled oscillatory magnetic tweezers for piconewton force manipulation applications,” Journal of Applied Physics, vol. 130, no. 1, p. 014504, 2021, DOI: 10.1063/5.0048701.
» https://doi.org/10.1063/5.0048701. -
[20] S. Lee, M. Van Veenhuizen, P. Navaretti, and G. Donati, “Unlocking novel fa case studies using a lock-in amplifier,” International Symposium for Testing and Failure Analysis, vol. 83348, pp. 253-260, 2020, DOI: 10.31399/asm.cp.istfa2020p0253.
» https://doi.org/10.31399/asm.cp.istfa2020p0253. -
[21] G. Olafsson, R. C. Tighe, S. Boyd, and J. M. Dulieu-Barton, “Lock-in thermography using miniature infra-red cameras and integrated actuators for defect identification in composite materials,” Optics & Laser Technology, vol. 147, p. 107629, 2022, DOI: 10.1016/j.optlastec.2021.107629.
» https://doi.org/10.1016/j.optlastec.2021.107629. -
[22] P. Dahlberg, N. J. Ziegeler, P. W. Nolte, and S. Schweizer, “Design and construction of an led-based excitation source for lock-in thermography,” Applied Sciences, vol. 12, no. 6, p. 2940, 2022, DOI: 10.3390/app12062940.
» https://doi.org/10.3390/app12062940. -
[23] A. D’amico, A. De Marcellis, C. Di Carlo, C. Di Natale, G. Ferri, E. Martinelli, R. Paolesse, and V. Stornelli, “Low-voltage low-power integrated analog lock-in amplifier for gas sensor applications,” Sensors and Actuators B: Chemical, vol. 144, no. 2, pp. 400-406, 2010, DOI: 10.1016/j.snb.2009.01.045.
» https://doi.org/10.1016/j.snb.2009.01.045. -
[24] S. Pettinato, A. Piccardi, M. C. Rossi, and S. Salvatori, “Design, implementation, and characterization of a compact lock-in add-on for low-frequency impedance measurements,” Electronics, vol. 12, no. 16, p. 3406, 2023, DOI: 10.3390/electronics12163406.
» https://doi.org/10.3390/electronics12163406. -
[25] M. Demirtas¸, M. A. Eris¸mis¸, and S. Günes¸, “A lossy capacitance measurement circuit based on analog lock-in detection,” Elektronika ir Elektrotechnika, vol. 26, no. 5, pp. 4-10, 2020, DOI: 10.5755/j01.eie.26.5.25809.
» https://doi.org/10.5755/j01.eie.26.5.25809. -
[26] J. M. Algueta-Miguel, J. J. Beato-López, and A. J. López-Martín, “Analog lock-in amplifier design using subsampling for accuracy enhancement in gmi sensor applications,” Sensors, vol. 23, no. 1, p. 57, 2022, DOI: 10.3390/s23010057.
» https://doi.org/10.3390/s23010057. -
[27] B. Ricobom, M. Nardi, and M. Bonfim, “Dual-phase digital lock-in implementation based on nyquist rate and sampling subtraction,” Review of Scientific Instruments, vol. 94, no. 6, p. 064706, 2023, DOI: 10.1063/5.0145566.
» https://doi.org/10.1063/5.0145566. -
[28] Z. Yao, J. Pan, C. Yu, Z. Yuan, Q. Chen, and X. Sui, “A universal digital lock-in amplifier design for calibrating the photo-detector responses with standard black-bodies,” Sensors, vol. 23, no. 21, p. 8902, 2023, DOI: 10.3390/s23218902.
» https://doi.org/10.3390/s23218902. -
[29] G. Stimpson, M. Skilbeck, R. Patel, and G. Morley, “An open-source high-frequency lock-in amplifier,” Review of Scientific Instruments, vol. 90, no. 9, p. 094701, 2019, DOI: 10.1063/1.5083797.
» https://doi.org/10.1063/1.5083797. -
[30] W. Zhang, M. Li, J. Wang, and X. Chen, “Integrated near-infrared fiber-optic photoacoustic sensing demodulator for ultra-high sensitivity gas detection,” Photoacoustics, vol. 33, p. 100560, 2023, DOI: 10.1016/j.pacs.2023.100560.
» https://doi.org/10.1016/j.pacs.2023.100560. -
[31] D. Uhl, L. Bruder, and F. Stienkemeier, “A flexible and scalable fully software-based lock-in amplifier for nonlinear spectroscopy,” Review of Scientific Instruments, vol. 92, p. 083101, 2021, DOI: 10.1063/5.0059740.
» https://doi.org/10.1063/5.0059740. -
[32] J. Li, H. Deng, J. Sun, B. Yu, and H. Fischer, “Simultaneous atmospheric co, n2o and h2o detection using a single quantum cascade laser sensor based on dual-spectroscopy techniques,” Sensors and Actuators B: Chemical, vol. 231, pp. 723-732, 2016, DOI: 10.1016/j.snb.2016.03.089.
» https://doi.org/10.1016/j.snb.2016.03.089. -
[33] H. Fonsêca, D. Rativa, and R. Lima, “In-loco optical spectroscopy through a multiple digital lock-in on a linear charge-coupled device (ccd) array,” Sensors, vol. 23, no. 16, p. 7195, 2023, DOI: 10.3390/s23167195.
» https://doi.org/10.3390/s23167195. - [34] M. Zou, L. Sun, and X. Wang, “Multigas sensing based on wavelength modulation spectroscopy using frequency division multiplexing combined with time division multiplexing,” IEEE Sensors Journal, vol. 22, no. 13, pp. 12 930-12 938, 2022.
-
[35] Y. Du, N. Liu, X. Wu, K. Liu, and J. Li, “Frequency division multiplexing and wavelength stabilized 2f/1f wavelength modulation spectroscopy for simultaneous trace ch4 and co2 detection,” Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, vol. 305, p. 123453, 2024, DOI: 10.1016/j.saa.2023.123453.
» https://doi.org/10.1016/j.saa.2023.123453. -
[36] G. Ma, Y. Wang, W. Qin, H. Zhou, C. Yan, J. Jiang, and Y. Ju, “Optical sensors for power transformer monitoring: a review,” High Voltage, vol. 6, no. 3, pp. 367-386, 2021, DOI: 10.1049/hve2.12021.
» https://doi.org/10.1049/hve2.12021. -
[37] A. S. Alghamdi, N. A. Muhamad, and A. A. Suleiman, “Dga interpretation of oil filled transformer condi-tion diagnosis,” Transactions on electrical and electronic materials, vol. 13, no. 5, pp. 229-232, 2012, DOI: 10.4313/TEEM.2012.13.5.229.
» https://doi.org/10.4313/TEEM.2012.13.5.229. -
[38] N. Mahmoudi, M. H. Samimi, and H. Mohseni, “Experiences with transformer diagnosis by dga: case studies,” IET Generation, Transmission & Distribution, vol. 13, no. 23, pp. 5431-5439, 2019, DOI: 10.1049/iet-gtd.2019.1056.
» https://doi.org/10.1049/iet-gtd.2019.1056. -
[39] S. A. Wani, A. S. Rana, S. Sohail, O. Rahman, S. Parveen, and S. A. Khan, “Advances in dga based condition monitoring of transformers: a review,” Renewable and Sustainable Energy Reviews, vol. 149, p. 111347, 2021, DOI: 10.1016/j.rser.2021.111347.
» https://doi.org/10.1016/j.rser.2021.111347. -
[40] M. Hamid, M. Ishak, M. Ariffin, N. Katim, N. Amin, and N. Azis, “Dissolved gas analysis (dga) of vegetable oils under electrical stress,” 2017 International Conference on High Voltage Engineering and Power Systems, pp. 29-34, 2017, DOI: 10.1109/ICHVEPS.2017.8225862.
» https://doi.org/10.1109/ICHVEPS.2017.8225862. -
[41] I. Horcas, D. Moreno-Salinas, and J. Sánchez-Moreno, “A cost-effective design for a mid-range microcontroller-based lock-in amplifier,” Microprocessors and Microsystems, vol. 113, p. 105145, 2025, DOI: 10.1016/j.micpro.2025.105145.
» https://doi.org/10.1016/j.micpro.2025.105145. -
[42] F. Pollastrone, L. Fiorani, R. Bisauriya, I. Menicucci, C. Ciceroni, and R. Pizzoferrato, “Development and test of low-cost multi-channel multi-frequency lock-in amplifier for health and environment sensing,” Sensors, vol. 24, no. 18, pp. 6020-6034, 2024, DOI: 10.3390/s24186020.
» https://doi.org/10.3390/s24186020. -
[43] A. J. Harvie and J. C. de Mello, “Olia: an open-source digital lock-in amplifier,” Frontiers in Sensors, vol. 4, p. 1102176, 2023, DOI: 10.48550/arXiv.2211.08889.
» https://doi.org/10.48550/arXiv.2211.08889. -
[44] J. A. Galaviz-Aguilar, C. Vargas-Rosales, F. Falcone, and C. Aguilar-Avelar, “Field-programmable gate array (fpga)-based lock-in amplifier system with signal enhancement: a comprehensive review on the design for advanced measurement applications,” Sensors, vol. 25, no. 2, pp. 584-608, 2025, DOI: 10.3390/s25020584.
» https://doi.org/10.3390/s25020584. -
[45] L. Cordesses, “Direct digital synthesis: a tool for periodic wave generation (part 1),” IEEE Signal Processing Magazine, vol. 21, no. 4, pp. 50-54, 2004, DOI: 10.1109/MSP.2004.1311140.
» https://doi.org/10.1109/MSP.2004.1311140. -
[46] J. Li, B. Yu, W. Zhao, and W. Chen, “A review of signal enhancement and noise reduction techniques for tunable diode laser absorption spectroscopy,” Applied Spectroscopy Reviews, vol. 49, no. 8, pp. 666-691, 2014, DOI: 10.1080/05704928.2014.903376.
» https://doi.org/10.1080/05704928.2014.903376. -
[47] J. Chang, Q. He, J. Li, and Q. Feng, “Oxygen detection system based on tdlas-wms and a compact multipass gas cell,” Microwave and Optical Technology Letters, vol. 65, no. 5, pp. 1141-1145, 2023, DOI: 10.1002/mop.33203.
» https://doi.org/10.1002/mop.33203. -
[48] A. De Marcellis, G. Ferri, and E. Palange, “Electrical self-modulation of optical sensors for light power measurement in chemical applications by phase detection technique,” Sensors and Actuators B: Chemical, vol. 193, pp. 375-383, 2014, DOI: 10.1016/j.snb.2013.11.116.
» https://doi.org/10.1016/j.snb.2013.11.116. -
[49] K. Schultz, “Phase-sensitive detection in the undergraduate lab using a low-cost microcontroller,” American Journal of Physics, vol. 84, no. 7, pp. 557-561, 2016, DOI: 10.48550/arXiv.1501.06074.
» https://doi.org/10.48550/arXiv.1501.06074. - [50] I. Bhagyajyothi, P. Bhaskar, and C. Parvathi, “Design and development of advanced lock-in amplifier and its application,” Sensors and Transducers, vol. 153, pp. 22-28, 2013.
-
[51] S. Bhattacharyya, R. N. Ahmed, B. B. Purkayastha, and K. Bhattacharyya, “Implementation of digital lock-in amplifier,” Journal of Physics: Conference Series, vol. 759, no. 1, p. 012096, 2016, DOI: 10.1088/1742-6596/759/1/012096.
» https://doi.org/10.1088/1742-6596/759/1/012096. -
[52] C. Zhu, P. Wang, T. Chu, F. Peng, and Y. Sun, “Second harmonic phase angle method based on wms for background-free gas detection,” IEEE Photonics Journal, vol. 13, no. 5, pp. 1-6, 2021, DOI: 10.1109/JPHOT.2021.3113503.
» https://doi.org/10.1109/JPHOT.2021.3113503. -
[53] J. Dang, J. Zhang, X. Dong, L. Kong, and H. Yu, “A trace ch4 detection system based on das calibrated wms technique,” Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, vol. 266, p. 120418, 2022, DOI: 10.1016/j.saa.2021.120418.
» https://doi.org/10.1016/j.saa.2021.120418. -
[54] V. H. Soares, “Otimização de um sensor óptico para detecção específica de gases dissolvidos em óleo de transformador,” M.S thesis, Universidade Federal de Pernambuco, Recife, Brasil, 2010, [Online]. Available: https://oasisbr.ibict.br/vufind/Record/BRCRIS_7a94a3a2548c62b29061d7dec88eedca
» https://oasisbr.ibict.br/vufind/Record/BRCRIS_7a94a3a2548c62b29061d7dec88eedca -
[55] A. Laugier and J. Garai, “Derivation of the ideal gas law,” Journal of Chemical Education, vol. 84, no. 11, p. 1832, 2007, DOI: 10.1021/ed084p1832.
» https://doi.org/10.1021/ed084p1832. -
[56] P. Sedgwick, “Pearson’s correlation coefficient,” Bmj, vol. 345, no. 4483, pp. 1-2, 2012, DOI: 10.1136/bmj.e4483 .
» https://doi.org/10.1136/bmj.e4483