Atrial Fibrillation Ablation: Impact of Intracardiac Echocardiography in Reducing Procedure Time and Hospitalization

Abstract

Background

Intracardiac echocardiography (ICE) allows visualization of cardiac structures and recognition of complications during atrial fibrillation ablation (AFA). Compared to transesophageal echocardiography (TEE), ICE is less sensitive to detecting thrombus in the atrial appendage but requires minimal sedation and fewer operators, making it attractive in a resource-constrained setting.

Objective

To compare 13 cases of AFA using ICE (AFA-ICE group) with 36 cases of AFA using TEE (AFA-TEE group).

Methods

This is a single-center prospective cohort study. The main outcome was procedure time. Secondary outcomes: fluoroscopy time, radiation dose (mGy/cm2), major complications, and length of hospital stay in hours. The clinical profile was compared using the CHA2DS2-VASc score. A p-value <0.05 was considered a statistically significant difference between groups.

Results

The median CHA2DS2-VASc score was 1 (0-3) in the AFA-ICE group and 1 (0-4) in the AFA-TEE group. The total procedure time was 129 ± 27 min in the AFA-ICE group and 189 ± 41 min in the AFA-TEE group (p<0.001); the AFA-ICE group received a lower dose of radiation (mGy/cm2, 51296 ± 24790 vs. 75874 ± 24293; p=0.002), despite the similar fluoroscopy time (27.48 ± 9. 79 vs. 26.4 ± 9.32; p=0.671). The median length of hospital stay did not differ; 48 (36-72) hours (AFA-ICE) and 48 (48-66) hours (AFA-TEE) (p=0.27).

Conclusions

In this cohort, AFA-ICE was related to shorter procedure times and less exposure to radiation without increasing the risk of complications or the length of hospital stay.

Atrial Fibrillation/complications; Catheter Ablation; Echocadiography/methods; Hospitalization; Pulmonary Veins/diagnostic imaging

Resumo

Fundamento

O ecocardiograma intracardíaco (EIC) permite visualizar estruturas cardíacas e reconhecer complicações durante a ablação da fibrilação atrial (AFA). Comparado ao ecocardiograma transesofágico (ETE), o EIC é menos sensível para detecção de trombo no apêndice atrial, porém requer mínima sedação e menos operadores, tornando-o atrativo num cenário de recursos restritos.

Objetivo

Comparar 13 casos de AFA utilizando EIC (grupo AFA-EIC) com 36 casos de AFA utilizando ETE (grupo AFA-ETE).

Método

Trata-se de corte prospectiva realizada em um único centro. O desfecho principal foi o tempo de procedimento. Desfechos secundários tempo de fluoroscopia, dose de radiação (mGy/cm2), complicações maiores e tempo de internação hospitalar em horas. O perfil clínico foi comparado pelo escore CHA2DS2-VASc. Um valor de p <0,05 foi considerado uma diferença estatisticamente significativa entre os grupos.

Resultados

A mediana do escore de CHA2DS2-VASc score foi 1 (0-3) no grupo AFA-EIC e 1 (0-4) no grupo AFA-ETE. O tempo total de procedimento foi de 129 ± 27 min grupo AFA-EIC e 189 ± 41 no AFA-ETE (p<0,001); o grupo AFA-EIC recebeu uma dose menor de radiação (mGy/cm2, 51296 ± 24790 vs. 75874 ± 24293; p=0,002), no entanto, o tempo de fluoroscopia em minutos mostrou-se semelhante (27,48 ± 9,79 vs. 26,4 ± 9,32; p=0,671). As medianas do tempo de hospitalização não se mostraram diferentes, 48 (36-72) horas (AFA-EIC) e 48 (48-66) horas (AFA-ETE) (p=0,27).

Conclusão

Nesta coorte, AFA-EIC foi relacionado a menores tempos de procedimento e menor exposição à radiação, sem aumentar o risco de complicações ou o tempo de internação hospitalar.

Fibrilação Atrial/complicações; Ablação por Cateter; Ecocardiografia/métodos; Hospitalização; Veias Pulmonares/diagnóstico por imagem

Introduction

Atrial fibrillation (AF) is the most common cardiac arrhythmia, affecting more than 37 million people worldwide.11. Lippi G, Sanchis-Gomar F, Cervellin G. Global epidemiology of atrial fibrillation: An increasing epidemic and public health challenge. Int J Stroke 2021;16(2):217-221. (In eng). DOI: 10.1177/1747493019897870.
https://doi.org/10.1177/1747493019897870...
AF ablation, achieved by pulmonary vein isolation (PVI), is recommended by the guidelines of professional societies as a useful therapy to maintain sinus rhythm and improve quality of life.22. Craig T, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2019;74(1):104-132. (In eng). DOI: 10.1016/j.jacc.2019.01.011.
https://doi.org/10.1016/j.jacc.2019.01.0...
, 33. Hindricks G, Potpara T, Dagres N, Arbelo Elena, Jerden J, Carina B et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J 2021;42(5):373-498. (In eng). DOI: 10.1093/eurheartj/ehaa612.
https://doi.org/10.1093/eurheartj/ehaa61...
The success and safety of PVI depend on the proper identification of anatomical landmarks by imaging methods, which include electro-anatomical mapping, computed tomography, cardiac magnetic resonance, and echocardiogram.44. Obeng-Gyimah E, Nazarian S Advancements in Imaging for Atrial Fibrillation Ablation: Is There a Potential to Improve Procedural Outcomes? J Innov Card Rhythm Manag 2020;11(7):4172-4178. (In eng). DOI: 10.19102/icrm.2020.110701.
https://doi.org/10.19102/icrm.2020.11070...

The use of echocardiography during AF ablation (AFA) has several benefits, such as the high real-time resolution of cardiac structures and availability. Echocardiography also has the potential to increase patient safety by early identification of procedure-related complications and to reduce patient and staff radiation exposure by decreasing dependency on fluoroscopy.55. Andrew CY, Klein AL. Role of Echocardiography in Atrial Fibrillation Ablation. J Atr Fibrillation 2011;4(4):397. (In eng). DOI: 10.4022/jafib.397.
https://doi.org/10.4022/jafib.397...

Two forms of echocardiography have been used in interventional cardiology: transesophageal (TEE) and intracardiac (ICE). TEE offers a better image resolution and lower initial cost because the probe is reusable. However, it requires general anesthesia and additionally trained professionals and carries a risk of trauma to the gastrointestinal tract. Together, these factors can increase the total cost of the procedure, particularly in a constrained human resource setting. There are few direct comparisons of the two methods for PVI. This study aimed to compare PVI using TEE with PVI using ICE regarding procedure time, major complications, radiation exposure, and length of hospital stay.

Methods

Study design

This was a single-center prospective cohort study of patients consecutively recruited from the Electrophysiology Service between December 2017 and June 2021. The protocol was designed following the STROBE statement for reporting observational studies66. von Elm E, Altman DG, Egger M, Pocock SJ, Gøtzsche PC, Vandenbroucke JP. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Ann Intern Med 2007;147(8):573-7. (In eng). DOI: 10.7326/0003-4819-147-8-200710160-00010.
https://doi.org/10.7326/0003-4819-147-8-...
and was approved by the local Ethics Committee (registration UP 5252/16).

Sample

All patients included in the study had a class I or IIA indication for AFA according to international guidelines.22. Craig T, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2019;74(1):104-132. (In eng). DOI: 10.1016/j.jacc.2019.01.011.
https://doi.org/10.1016/j.jacc.2019.01.0...
, 33. Hindricks G, Potpara T, Dagres N, Arbelo Elena, Jerden J, Carina B et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J 2021;42(5):373-498. (In eng). DOI: 10.1093/eurheartj/ehaa612.
https://doi.org/10.1093/eurheartj/ehaa61...
We included patients with paroxysmal and persistent AF with less than 1 year of evolution. We excluded patients who presented with intracavitary thrombus on TEE performed on the day of the AFA.

Procedure and follow-up

All procedures were performed by the same electrophysiologist, with the patient under general anesthesia using an esophageal thermometer at the level of the left atrium (LA). After venipunctures, a decapolar catheter was placed in the coronary sinus. Two transseptal punctures were performed, guided by echocardiography with an SL-1 sheath and a BRK-1 needle (St. Jude Medical, St Paul, USA), followed by therapeutic heparinization. A circular duodecapolar catheter (Biosense Webster, Irvine, USA) was used to map the LA, and an irrigated tip ablation catheter was placed in the LA. Atrial mapping was performed using the three-dimensional EnSite system (St. Jude Medical, St. Paul, USA) . Radiofrequency was applied circumferentially around the ipsilateral pulmonary veins until electrical isolation was achieved. Anticoagulant therapy was initiated 6 hours after removal of the sheaths and maintained for at least 60 days. All patients received pantoprazole, a proton pump inhibitor, for 30 days and were maintained with the same antiarrhythmic drug used preoperatively for 90 days.

Patients were divided into two groups according to the imaging method used during AFA: intracardiac echocardiogram (AFA-ICE) group or transesophageal echocardiogram (AFA-TEE) group. In the AFA-ICE group, a 10 Fr phased-array ultrasound imaging catheter (AcuNav, Acuson) was introduced through an 11 Fr sheath through the left femoral vein. The ICE was used to ensure catheter placement, proper energy delivery site, monitoring of microbubble formation, and detection of complications. In the AFA-TEE group, the transesophageal echocardiogram was performed by a specialist in cardiac imaging. The choice between the imaging methods was based on whether an intracardiac echocardiogram was available from the health insurance source of the patient in question. Regarding the X-ray device for fluoroscopy, PHILIPS ALLURA XPER equipment was configured at a rate of 3.75 frames per second.

Patients were admitted for monitoring in the intensive care unit (ICU) for at least one night after the ablation procedure. After that, discharge was at the discretion of the attending physician. Antiarrhythmic drugs were maintained for at least 3 months. All patients included in this study were followed up for at least three months (mean time: 650.95 ± 380.86 days).

Outcomes

The primary outcome was the total procedure time. Secondary outcomes were major complications (cardiac tamponade, stroke, esophageal perforation), fluoroscopy time, radiation dose in mGy/cm22. Craig T, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2019;74(1):104-132. (In eng). DOI: 10.1016/j.jacc.2019.01.011.
https://doi.org/10.1016/j.jacc.2019.01.0...
, and time to hospital discharge measured in hours. The clinical profile was compared using the CHA2DS2-VASc score and echocardiographic parameters of left ventricular ejection fraction (LVEF) and left atrial diameter (LAD).

Statistical analysis

Statistical analyzes were performed using the Statistical Package for the Social Sciences (SPSS) version 22.0 software (SPSS Inc., Chicago, IL, USA) . Continuous variables with normal distribution are described as mean, and standard deviation, and continuous variables without normal distribution are described as median and interquartile range. The statistical test to assess normality was the Shapiro-Wilk. We used Student’s t-test for independent samples with normal distribution and the non-parametric Mann-Whitney U test for variables without normal distribution. Categorical variables are expressed as frequencies and percentages and were compared using the χ2 test or Fisher’s exact test. A p-value <0.05 was considered a statistically significant difference between groups.

Results

Clinical profile

Forty-nine patients underwent PVI, the majority male (81%) with a mean age of 54 ± 12.2 years. The most common comorbidity was hypertension, identified in 65.3% of the included patients. Patients had few comorbidities, as reflected by a median CHADS2Vasc score of 1 in both groups. Most patients had paroxysmal AF (89.8%), with a similar prevalence between the AFA-TEE (84.6%) and AFA-ICE (91.6%) groups. Beta-blockers (59.1%) and amiodarone (57.1%) were the most prescribed antiarrhythmics. About a quarter of the patients included were receiving a combination of medications (28.5%).

Table 1 presents the baseline demographic and clinical characteristics, clinical scores, and echocardiographic parameters for the populations grouped according to whether ICE or TEE aided the procedure.

Table 1
– Clinical and demographic characteristics of included patients. Porto Alegre, RS

Table 2 shows anticoagulant and antiarrhythmic drugs according to the group. The central illustration highlights the main results of the study.

Table 2
– Use of anticoagulants and antiarrhythmics according to the group. Porto Alegre, RS

Central Illustration
: Atrial Fibrillation Ablation: Impact of Intracardiac Echocardiography in Reducing Procedure Time and Hospitalization

Ablation procedure and follow-up

In order to maintain a reliable comparison of the procedure times between the two groups, we excluded from the analysis two cases of the AFA-TEE group in which the transesophageal echocardiogram operator was not present on site for the entirety of the procedure. The total procedure time was 129 ± 27 minutes in the AFA-ICE group and 189 ± 41 minutes in the AFA-ETE group (p<0.001). The AFA-ICE group also received a lower dose of radiation (mGy/cm22. Craig T, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2019;74(1):104-132. (In eng). DOI: 10.1016/j.jacc.2019.01.011.
https://doi.org/10.1016/j.jacc.2019.01.0...
, 51296 ± 24790 vs. 75874 ± 24293; p=0.002), despite having a similar fluoroscopy time (minutes, 27.48 ± 9.79 vs. 26.4 ± 9.32; p=0.671).

The rate of major complications was low and similar between the two groups. There were no cases of cardiac tamponade or cerebral or esophageal events. One patient in the AFA-ICE group had an occlusion of the right superficial femoral artery diagnosed on the second day after the procedure and successfully treated with embolectomy. One patient from the AFA-ETE had an accelerated evolution of Hashimoto’s thyroiditis, probably related to amiodarone. Three patients underwent a second PVI procedure during follow-up. Additional relevant data are shown in table 3 .

Table 3
– Pulmonary vein isolation outcome. Porto Alegre, RS

Discussion

Our study shows that ICE-assisted PVI is associated with a shorter procedure time and lower radiation exposure when compared to TEE-assisted PVI while maintaining the procedure’s safety. The duration of both hospitalization and ICU stay were similar between groups. The two groups also had an equivalent clinical profile, as demonstrated by a median CHADS2Vasc score of 1 in both groups and similar LVEF and AED.

Identification of anatomical landmarks is of paramount importance for successful AF ablation. In a study by Toffanin et al.,77. Toffanin G, Scarabeo V, Verlato R, De Conti F, Zampiero AA, Piovesana P. Transoesophageal echocardiographic evaluation of pulmonary vein anatomy in patients undergoing ostial radiofrequency catheter ablation for atrial fibrillation: a comparison with magnetic resonance angiography. J Cardiovasc Med (Hagerstown) 2006;7(10):748-52. (In eng). DOI: 10.2459/01.Jcm.0000247322.57536.04.
https://doi.org/10.2459/01.Jcm.000024732...
evaluating the anatomy of the PV with TEE and magnetic resonance, only 42% of the patients had normal anatomy of the pulmonary veins with two right and left veins. Integration of images with computed tomography or cardiac magnetic resonance imaging may improve AF-free survival rate after ablation.88. Della Bella P, Fassini G, Cireddu M, Riva S, Carbucicchio C, Girardi F. Image integration-guided catheter ablation of atrial fibrillation: a prospective randomized study. Journal of cardiovascular electrophysiology 2009;20(3):258-65. (In eng). DOI: 10.1111/j.1540-8167.2008.01311.x.
https://doi.org/10.1111/j.1540-8167.2008...
, 99. Caponi D, Corleto A, Scaglione M, Blandino A. Ablation of atrial fibrillation: does the addition of three-dimensional magnetic resonance imaging of the left atrium to electroanatomic mapping improve the clinical outcome? a randomized comparison of Carto-Merge vs. Carto-XP three-dimensional mapping ablation in patients with paroxysmal and persistent atrial fibrillation. Europace 2010;12(8):1098-104. (In eng). DOI: 10.1093/europace/euq107.
https://doi.org/10.1093/europace/euq107...
Interestingly, fluoroscopic times are generally not shorter in such studies.

We found a marked reduction in total procedure time. We attribute this difference mainly to the longer time to position the transesophageal probe and the operator’s ability to obtain adequate images when using the EIC immediately. TEE depended on an extra professional who needed to put on adequate protection against radiation and to reposition the probe to provide adequate images for the operator, not only at the time of transseptal puncture but also when complications were suspected and at the end of the procedure to rule out the presence of tamponade. In two cases of our series, the professional responsible for the TEE had to have waited in the room as he was urgently requested for another procedure. We chose to exclude these patients from the comparisons. In two cases, TEE could not properly identify the puncture needle or the fossa ovalis, increasing the procedure time. Other differences between the methods include the safety and the autonomy provided by the ICE, considering it allows the same operator to perform imaging and ablation simultaneously and the need for esophageal intubation in the TEE group. The total fluoroscopy time was similar between the groups, as the factors that prolonged the procedure time using TEE cannot be compensated with greater use of X-rays. In the meta-analysis by Isath et al.,1010. Isath A, Padmanabhan D, Haider SW, Siruky, G Perimberti, S Correa A. et al. Does the use of intracardiac echocardiography during atrial fibrillation catheter ablation improve outcomes and cost? A nationwide 14-year analysis from 2001 to 2014. J Interv Card Electrophysiol 2021;61(3):461-468. (In eng). DOI: 10.1007/s10840-020-00844-5.
https://doi.org/10.1007/s10840-020-00844...
which gathered 19 studies and compared the results of ablations performed with or without EIC, there was a reduction in both fluoroscopy and procedure time. However, it must be considered that a direct comparison was not made with TEE but with isolated fluoroscopy and/or electro-anatomical mapping.

TEE is the gold standard for identifying a thrombus in the LA appendix.44. Obeng-Gyimah E, Nazarian S Advancements in Imaging for Atrial Fibrillation Ablation: Is There a Potential to Improve Procedural Outcomes? J Innov Card Rhythm Manag 2020;11(7):4172-4178. (In eng). DOI: 10.19102/icrm.2020.110701.
https://doi.org/10.19102/icrm.2020.11070...
TEE is an invasive method with a complication risk of approximately 0.9%.1111. Rathi VK, Reddy ST, Anreddy S, Belden W, Yamrozik, JA. Williams, R et al. Contrast-enhanced CMR is equally effective as TEE in the evaluation of left atrial appendage thrombus in patients with atrial fibrillation undergoing pulmonary vein isolation procedure. Heart rhythm 2013;10(7):1021-7. (In eng). DOI: 10.1016/j.hrthm.2013.02.029.
https://doi.org/10.1016/j.hrthm.2013.02....
, 1212. Freitas-Ferraz AB, Bernier M, Vaillancourt R, Nicodeme F.Ugade P, Paradis JM. et al. of Transesophageal Echocardiography to Guide Structural Cardiac Interventions. J Am Coll Cardiol 2020;75(25):3164-3173. (In eng). DOI: 10.1016/j.jacc.2020.04.069.
https://doi.org/10.1016/j.jacc.2020.04.0...
ICE can improve the procedure’s outcome by reducing the risk of recurrence of AF and the risk of pulmonary vein stenosis,1313. Marrouche NF, Martin DO, Wazni O, Gillinov M, Klein A, Bhargava M, et al., Phased-array intracardiac echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation: impact on outcome and complications. Circulation 2003;107(21):2710-6. (In eng). DOI: 10.1161/01.Cir.0000070541.83326.15.
https://doi.org/10.1161/01.Cir.000007054...
despite its limitation in properly identifying small branches of the pulmonary veins. Isath et al.,1010. Isath A, Padmanabhan D, Haider SW, Siruky, G Perimberti, S Correa A. et al. Does the use of intracardiac echocardiography during atrial fibrillation catheter ablation improve outcomes and cost? A nationwide 14-year analysis from 2001 to 2014. J Interv Card Electrophysiol 2021;61(3):461-468. (In eng). DOI: 10.1007/s10840-020-00844-5.
https://doi.org/10.1007/s10840-020-00844...
in a 14-year analysis involving several North American centers, demonstrated that the use of ICE during AFA was related to lower in-hospital mortality, lower risk of procedure complications, and shorter hospital stay. Specifically, ICE allows for faster detection of complications, including pericardial effusion, air embolism, ventricular dysfunction as a cause of hypotension, and thrombus formation in sheaths and catheters used for ablation.

A meta-analysis by Goya et al.,1414. Goya M, Frame D, Gache L, Ichishima Y, Tayar DO, Goldstein L. et al. The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes. Journal of cardiovascular electrophysiology 2020;31(3):664-673. (In eng). DOI: 10.1111/jce.14367.
https://doi.org/10.1111/jce.14367...
evaluated the results using ICE for arrhythmia ablation. Of the 19 included studies, 14 were performed in patients with AF. The use of ICE was associated with a significantly lower fluoroscopy time, fluoroscopy dose, and shorter procedure time without compromising the clinical efficacy or safety of the procedure. Although the findings are like our study, it should be noted that the meta-analysis by Goya et al.1414. Goya M, Frame D, Gache L, Ichishima Y, Tayar DO, Goldstein L. et al. The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes. Journal of cardiovascular electrophysiology 2020;31(3):664-673. (In eng). DOI: 10.1111/jce.14367.
https://doi.org/10.1111/jce.14367...
was not a direct comparison between EIC and TEE. The broad comparator group included electro-anatomical mapping, fluoroscopy, or other imaging modalities, sometimes used in conjunction with or nothing in addition to fluoroscopy. Ribeiro et al.,1515. Ribeiro JM, Teixeira R, Puga L, Costa M, Gonçalves L. Comparison of intracardiac and transoesophageal echocardiography for guidance of percutaneous left atrial appendage occlusion: A meta-analysis. Echocardiography 2019;36(7):1330-1337. (In eng). DOI: 10.1111/echo.14415.
https://doi.org/10.1111/echo.14415...
comparing percutaneous occlusion of the left atrium guided by ICE with TEE, found similar procedure time, risk of complications, and procedure time between the two methods.

There is a linear relationship between radiation dose and the risk of future malignancy. Radiation exposure is also linked to acute skin injuries, thyroid disorders, and cataracts, among other illnesses.1616. Lickfett L, Mahesh M, Vasamreddy C, Bradley D. Jayam V, Lickfettl L, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation 2004;110(19):3003-10. (In eng). DOI: 10.1161/01.Cir.0000146952.49223.11
https://doi.org/10.1161/01.Cir.000014695...
, 1717. Hirshfeld JW, Jr., Ferrari VA, Bengel FM, Bergersen, L.Chambers, C. E Chambers, A. J. et al. 2018 C/HRS/NASCI/SCAI/SCCT Expert Consensus Document on Optimal Use of Ionizing Radiation in Cardiovascular Imaging: Best Practices for Safety and Effectiveness: A Report of the American College of Cardiology Task Force on Expert Consensus Decision Pathways. J Am Coll Cardiol 2018;71(24):e283-e351. (In eng). DOI: 10.1016/j.jacc.2018.02.016.
https://doi.org/10.1016/j.jacc.2018.02.0...
Since no dose of radiation is considered safe, and operators may be at greater risk due to the multiple procedures they perform, it is recommended that radiation exposure be as low as possible.1717. Hirshfeld JW, Jr., Ferrari VA, Bengel FM, Bergersen, L.Chambers, C. E Chambers, A. J. et al. 2018 C/HRS/NASCI/SCAI/SCCT Expert Consensus Document on Optimal Use of Ionizing Radiation in Cardiovascular Imaging: Best Practices for Safety and Effectiveness: A Report of the American College of Cardiology Task Force on Expert Consensus Decision Pathways. J Am Coll Cardiol 2018;71(24):e283-e351. (In eng). DOI: 10.1016/j.jacc.2018.02.016.
https://doi.org/10.1016/j.jacc.2018.02.0...
In this sense, the significantly lower dose to which patients in the ICE group were submitted is a point in favor of the method. We believe that the ICE allowed the use of images with lower intensity, which reduced the radiation dose, despite the similar fluoroscopy times.

By requiring an additional 11 Fr sheath vascular access, ICE may increase the risk of vascular complications. This was not the case in our study. In the meta-analysis performed by Goya et al.,1414. Goya M, Frame D, Gache L, Ichishima Y, Tayar DO, Goldstein L. et al. The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes. Journal of cardiovascular electrophysiology 2020;31(3):664-673. (In eng). DOI: 10.1111/jce.14367.
https://doi.org/10.1111/jce.14367...
there was a trend towards an increase in vascular complications with ICE (RR 1.93; 95% CI, 0.81–4.60; p = 0.14).

ICE requires proper operator training in catheter manipulation and image interpretation. Because it is being progressively adopted in other forms of ablation,1818. Field ME, Gold MR, Reynolds MR, Goldstein Laura, Lee, Stephanie Hsiao Yu, Kalsekaret I. al. Real-world outcomes of ventricular tachycardia catheter ablation with versus without intracardiac echocardiography. Journal of cardiovascular electrophysiology 2020;31(2):417-422. (In eng). DOI: 10.1111/jce.14324.
https://doi.org/10.1111/jce.14324...
, 1919. Lamberti F, Di Clemente F, Remoli R, Bellini C , De Santis, Antonella, Mercurio, Marina Catheter ablation of idiopathic ventricular tachycardia without the use of fluoroscopy. Int J Cardiol 2015;190:338-43. (In eng). DOI: 10.1016/j.ijcard.2015.04.146.
https://doi.org/10.1016/j.ijcard.2015.04...
we believe that it should be part of the skill set of electrophysiologists who perform complex ablations. ICE also requires a non-reusable catheter, increasing the cost of the procedure, which is at least partially offset by the reduction in the number of staff needed for the procedure and the reduction in the consumption of anesthetic medication.2020. Bartel T, Müller S, Biviano A, Hahn RT. Why is intracardiac echocardiography helpful? Benefits, costs, and how to learn. Eur Heart J 2014;35(2):69-76. (In eng). DOI: 10.1093/eurheartj/eht411.
https://doi.org/10.1093/eurheartj/eht411...

ICE allows real-time visualization of cardiac structures and intraprocedural assessment of the catheter-tissue interface. Our sample was insufficient to compare the procedure’s effectiveness between the two methods. In a real-world study, ventricular tachycardia ablation using ICE was associated with a lower likelihood of ventricular tachycardia-related readmission at 12 months and repeat ablation when compared to ablation without the aid of ICE.1818. Field ME, Gold MR, Reynolds MR, Goldstein Laura, Lee, Stephanie Hsiao Yu, Kalsekaret I. al. Real-world outcomes of ventricular tachycardia catheter ablation with versus without intracardiac echocardiography. Journal of cardiovascular electrophysiology 2020;31(2):417-422. (In eng). DOI: 10.1111/jce.14324.
https://doi.org/10.1111/jce.14324...
A retrospective study of Medicare patients undergoing AFA found that the use of ICE was associated with a lower risk of repeat ablation (hazard ratio 0.59; 95% CI 0.37-0.92).2121. Steinberg BA, Hammill BG, Daubert JP, Douglas Pamela S, Qualls, Laura G. Periprocedural imaging and outcomes after catheter ablation of atrial fibrillation. Heart 2014;100(23):1871-7. (In eng). DOI: 10.1136/heartjnl-2014-306067.
https://doi.org/10.1136/heartjnl-2014-30...

Study limitations

Our study suffers from some limitations. (1) Small sample size may be responsible for not detecting some differences between groups, such as length of hospitalization. (2) Lack of randomization. The use of the ICE was based on the availability of health insurance, which could systematically influence the profile of patients in each group. Table 1 shows that both groups had a similar profile regarding the main risk factors for AF recurrence and technical difficulty during ablation, including left atrial size, which mitigates this potential bias. Some explanations for the difference in procedure time between groups, such as operator availability in the TEE group, are due to local factors and may not reflect the reality of other institutions. (3) We did not perform a direct cost comparison between the two strategies. As the patients had different health plans, the form of reimbursement and the cost varied accordingly, preventing a homogeneous analysis of the effect of the method employed on the costs. The intracardiac echocardiogram is sold to our institution for R$ 7500.00. Regarding the TEE, the procedure time, on average 55 minutes longer, would increase room costs by R$ 1100.00, while the professional to perform the exam would have an average cost of R$ 1000.00. It is not possible to estimate the indirect cost of this greater occupation of health resources, especially during the pandemic, when 25 of the 49 ablations analyzed in the study were performed.

Conclusion

In our cohort, the use of AFA with ICE was related to shorter procedure times and less exposure to radiation without increasing the risk of complications or length of hospital stay. The increase in the initial cost of the procedure can be compensated by a lower occupancy of the operating room and the need for fewer employees trained in the procedure.

Referências

  • 1
    Lippi G, Sanchis-Gomar F, Cervellin G. Global epidemiology of atrial fibrillation: An increasing epidemic and public health challenge. Int J Stroke 2021;16(2):217-221. (In eng). DOI: 10.1177/1747493019897870.
    » https://doi.org/10.1177/1747493019897870
  • 2
    Craig T, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2019;74(1):104-132. (In eng). DOI: 10.1016/j.jacc.2019.01.011.
    » https://doi.org/10.1016/j.jacc.2019.01.011
  • 3
    Hindricks G, Potpara T, Dagres N, Arbelo Elena, Jerden J, Carina B et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J 2021;42(5):373-498. (In eng). DOI: 10.1093/eurheartj/ehaa612.
    » https://doi.org/10.1093/eurheartj/ehaa612
  • 4
    Obeng-Gyimah E, Nazarian S Advancements in Imaging for Atrial Fibrillation Ablation: Is There a Potential to Improve Procedural Outcomes? J Innov Card Rhythm Manag 2020;11(7):4172-4178. (In eng). DOI: 10.19102/icrm.2020.110701.
    » https://doi.org/10.19102/icrm.2020.110701
  • 5
    Andrew CY, Klein AL. Role of Echocardiography in Atrial Fibrillation Ablation. J Atr Fibrillation 2011;4(4):397. (In eng). DOI: 10.4022/jafib.397.
    » https://doi.org/10.4022/jafib.397
  • 6
    von Elm E, Altman DG, Egger M, Pocock SJ, Gøtzsche PC, Vandenbroucke JP. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Ann Intern Med 2007;147(8):573-7. (In eng). DOI: 10.7326/0003-4819-147-8-200710160-00010.
    » https://doi.org/10.7326/0003-4819-147-8-200710160-00010
  • 7
    Toffanin G, Scarabeo V, Verlato R, De Conti F, Zampiero AA, Piovesana P. Transoesophageal echocardiographic evaluation of pulmonary vein anatomy in patients undergoing ostial radiofrequency catheter ablation for atrial fibrillation: a comparison with magnetic resonance angiography. J Cardiovasc Med (Hagerstown) 2006;7(10):748-52. (In eng). DOI: 10.2459/01.Jcm.0000247322.57536.04.
    » https://doi.org/10.2459/01.Jcm.0000247322.57536.04
  • 8
    Della Bella P, Fassini G, Cireddu M, Riva S, Carbucicchio C, Girardi F. Image integration-guided catheter ablation of atrial fibrillation: a prospective randomized study. Journal of cardiovascular electrophysiology 2009;20(3):258-65. (In eng). DOI: 10.1111/j.1540-8167.2008.01311.x.
    » https://doi.org/10.1111/j.1540-8167.2008.01311.x
  • 9
    Caponi D, Corleto A, Scaglione M, Blandino A. Ablation of atrial fibrillation: does the addition of three-dimensional magnetic resonance imaging of the left atrium to electroanatomic mapping improve the clinical outcome? a randomized comparison of Carto-Merge vs. Carto-XP three-dimensional mapping ablation in patients with paroxysmal and persistent atrial fibrillation. Europace 2010;12(8):1098-104. (In eng). DOI: 10.1093/europace/euq107.
    » https://doi.org/10.1093/europace/euq107
  • 10
    Isath A, Padmanabhan D, Haider SW, Siruky, G Perimberti, S Correa A. et al. Does the use of intracardiac echocardiography during atrial fibrillation catheter ablation improve outcomes and cost? A nationwide 14-year analysis from 2001 to 2014. J Interv Card Electrophysiol 2021;61(3):461-468. (In eng). DOI: 10.1007/s10840-020-00844-5.
    » https://doi.org/10.1007/s10840-020-00844-5
  • 11
    Rathi VK, Reddy ST, Anreddy S, Belden W, Yamrozik, JA. Williams, R et al. Contrast-enhanced CMR is equally effective as TEE in the evaluation of left atrial appendage thrombus in patients with atrial fibrillation undergoing pulmonary vein isolation procedure. Heart rhythm 2013;10(7):1021-7. (In eng). DOI: 10.1016/j.hrthm.2013.02.029.
    » https://doi.org/10.1016/j.hrthm.2013.02.029
  • 12
    Freitas-Ferraz AB, Bernier M, Vaillancourt R, Nicodeme F.Ugade P, Paradis JM. et al. of Transesophageal Echocardiography to Guide Structural Cardiac Interventions. J Am Coll Cardiol 2020;75(25):3164-3173. (In eng). DOI: 10.1016/j.jacc.2020.04.069.
    » https://doi.org/10.1016/j.jacc.2020.04.069
  • 13
    Marrouche NF, Martin DO, Wazni O, Gillinov M, Klein A, Bhargava M, et al., Phased-array intracardiac echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation: impact on outcome and complications. Circulation 2003;107(21):2710-6. (In eng). DOI: 10.1161/01.Cir.0000070541.83326.15.
    » https://doi.org/10.1161/01.Cir.0000070541.83326.15
  • 14
    Goya M, Frame D, Gache L, Ichishima Y, Tayar DO, Goldstein L. et al. The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes. Journal of cardiovascular electrophysiology 2020;31(3):664-673. (In eng). DOI: 10.1111/jce.14367.
    » https://doi.org/10.1111/jce.14367
  • 15
    Ribeiro JM, Teixeira R, Puga L, Costa M, Gonçalves L. Comparison of intracardiac and transoesophageal echocardiography for guidance of percutaneous left atrial appendage occlusion: A meta-analysis. Echocardiography 2019;36(7):1330-1337. (In eng). DOI: 10.1111/echo.14415.
    » https://doi.org/10.1111/echo.14415
  • 16
    Lickfett L, Mahesh M, Vasamreddy C, Bradley D. Jayam V, Lickfettl L, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation 2004;110(19):3003-10. (In eng). DOI: 10.1161/01.Cir.0000146952.49223.11
    » https://doi.org/10.1161/01.Cir.0000146952.49223.11
  • 17
    Hirshfeld JW, Jr., Ferrari VA, Bengel FM, Bergersen, L.Chambers, C. E Chambers, A. J. et al. 2018 C/HRS/NASCI/SCAI/SCCT Expert Consensus Document on Optimal Use of Ionizing Radiation in Cardiovascular Imaging: Best Practices for Safety and Effectiveness: A Report of the American College of Cardiology Task Force on Expert Consensus Decision Pathways. J Am Coll Cardiol 2018;71(24):e283-e351. (In eng). DOI: 10.1016/j.jacc.2018.02.016.
    » https://doi.org/10.1016/j.jacc.2018.02.016
  • 18
    Field ME, Gold MR, Reynolds MR, Goldstein Laura, Lee, Stephanie Hsiao Yu, Kalsekaret I. al. Real-world outcomes of ventricular tachycardia catheter ablation with versus without intracardiac echocardiography. Journal of cardiovascular electrophysiology 2020;31(2):417-422. (In eng). DOI: 10.1111/jce.14324.
    » https://doi.org/10.1111/jce.14324
  • 19
    Lamberti F, Di Clemente F, Remoli R, Bellini C , De Santis, Antonella, Mercurio, Marina Catheter ablation of idiopathic ventricular tachycardia without the use of fluoroscopy. Int J Cardiol 2015;190:338-43. (In eng). DOI: 10.1016/j.ijcard.2015.04.146.
    » https://doi.org/10.1016/j.ijcard.2015.04.146
  • 20
    Bartel T, Müller S, Biviano A, Hahn RT. Why is intracardiac echocardiography helpful? Benefits, costs, and how to learn. Eur Heart J 2014;35(2):69-76. (In eng). DOI: 10.1093/eurheartj/eht411.
    » https://doi.org/10.1093/eurheartj/eht411
  • 21
    Steinberg BA, Hammill BG, Daubert JP, Douglas Pamela S, Qualls, Laura G. Periprocedural imaging and outcomes after catheter ablation of atrial fibrillation. Heart 2014;100(23):1871-7. (In eng). DOI: 10.1136/heartjnl-2014-306067.
    » https://doi.org/10.1136/heartjnl-2014-306067
  • Study association
    This study is not associated with any thesis or dissertation work.
  • Sources of funding: There were no external funding sources for this study.

Publication Dates

  • Publication in this collection
    15 May 2023
  • Date of issue
    2023

History

  • Received
    23 Apr 2022
  • Reviewed
    07 Nov 2022
  • Accepted
    15 Feb 2023
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br