Echocardiographic monitoring of balloon atrial septostomy

Abstracts

OBJECTIVE: To assess balloon atrial septostomy monitored with echocardiography. METHODS: From August 1997 to January 2004, 31 infants with congenital heart diseases indicated for balloon atrial septostomy underwent the procedure with exclusive echocardiographic monitoring. Success was defined as the obtainment of an atrial septal defect diameter > 4 mm and ample mobility of its margins. RESULTS: The male sex predominated (83.9%). The median age was 5 days (1 - 150), and the median weight was 3,300 g (1,800 - 7,500). Transposition of the great arteries occurred in 80.6% of the patients, tricuspid valve atresia in 12.9%, total anomalous pulmonary venous drainage in 3.2%, and pulmonary atresia with intact septum in 3.2%. The procedure was successful in all patients. The size of the atrial septal defect increased from 1.8 ± 0.8 mm to 5.8 ± 1.3 mm (P < 0.0001) and oxygen arterial saturation from 64.5 ± 18.9% to 85.1 ± 9.2% (P < 0.0001). The following complications occurred: 3 balloon ruptures, one lesion of the right femoral vein, one supraventricular tachycardia, and one atrial flutter. CONCLUSION: Balloon atrial septostomy monitored with echocardiography is a safe and effective procedure. It may be performed at bedside, avoiding transporting of the patient, identifies the catheter location, reduces the occurrence of severe complications, and assesses the immediate result of the procedure.

congenital heart disease; echocardiography; balloon atrial septostomy


OBJETIVO: Avaliar a atriosseptostomia com balão monitorada pela ecocardiografia. MÉTODOS: Entre agosto de 1997 e janeiro de 2004, 31 crianças portadoras de cardiopatias congênitas com indicação de atriosseptostomia com balão foram submetidas ao procedimento sob monitoração ecocardiográfica exclusiva. Admitiu-se sucesso quando da obtenção de comunicação interatrial com diâmetro > 4 mm e com ampla mobilidade das suas margens. RESULTADOS: Predominou o sexo masculino (83,9%). A idade mediana foi de 5 dias (1-150) e o peso teve mediana de 3.300g (1.800-7.500). Transposição das grandes artérias ocorreu em 80,6%, atresia tricúspide em 12,9%, drenagem anômala total de veias pulmonares em 3,2% e atresia pulmonar com septo íntegro em 3,2%. O procedimento foi bem sucedido em todos os casos. O tamanho da comunicação interatrial aumentou de 1,8±0,8 mm para 5,8±1,3 mm (P<0,0001) e a saturação arterial de oxigênio de 64,5± 18,9% para 85,1±9,2% (P<0,0001). Complicações ocorridas: três rupturas de balão, uma lesão de veia femoral direita, uma taquicardia supraventricular e um flutter atrial. CONCLUSÃO: Atriosseptostomia com balão monitorada pela ecocardiografia é método seguro e eficaz. Possibilita a realização do procedimento à beira do leito, evitando o transporte da criança, identificando o posicionamento do cateter, reduzindo complicações graves e avaliando o resultado imediato do procedimento.

cardiopatia congênita; ecocardiografia; atriosseptostomia com balão


ORIGINAL ARTICLE

Echocardiographic monitoring of balloon atrial septostomy

Carlos Henrique de Marchi; Moacir Fernandes de Godoy; Márcio Antônio dos Santos; Airton Camacho Moscardini; Sírio Hassen Sobrinho; Ulisses Alexandre Croti

São José do Rio Preto, SP - Brazil

Faculdade de Medicina de São José do Rio Preto

Correspondence Correspondence to Carlos Henrique de Marchi Al. Dr. Ulysses da Silveira Guimarães, 273 Cep 15061-723 São José do Rio Preto, SP, Brazil E-mail: chmarchi@cardiol.br

ABSTRACT

OBJECTIVE: To assess balloon atrial septostomy monitored with echocardiography.

METHODS: From August 1997 to January 2004, 31 infants with congenital heart diseases indicated for balloon atrial septostomy underwent the procedure with exclusive echocardiographic monitoring. Success was defined as the obtainment of an atrial septal defect diameter > 4 mm and ample mobility of its margins.

RESULTS: The male sex predominated (83.9%). The median age was 5 days (1 – 150), and the median weight was 3,300 g (1,800 - 7,500). Transposition of the great arteries occurred in 80.6% of the patients, tricuspid valve atresia in 12.9%, total anomalous pulmonary venous drainage in 3.2%, and pulmonary atresia with intact septum in 3.2%. The procedure was successful in all patients. The size of the atrial septal defect increased from 1.8 ± 0.8 mm to 5.8 ± 1.3 mm (P < 0.0001) and oxygen arterial saturation from 64.5 ± 18.9% to 85.1 ± 9.2% (P < 0.0001). The following complications occurred: 3 balloon ruptures, one lesion of the right femoral vein, one supraventricular tachycardia, and one atrial flutter.

CONCLUSION: Balloon atrial septostomy monitored with echocardiography is a safe and effective procedure. It may be performed at bedside, avoiding transporting of the patient, identifies the catheter location, reduces the occurrence of severe complications, and assesses the immediate result of the procedure.

Key words: congenital heart disease; echocardiography; balloon atrial septostomy

The survival of children with certain congenital heart diseases depends on the presence of an adequate atrial septal defect. Balloon atrial septostomy is a palliative technique that enables blood circulation in those children by creating or widening an atrial septal defect, enabling a better blood mixture in the atria.

Surgical removal of part of the atrial septum to create communication was described by Blalock and Hanlon 1 in 1950. A great landmark in the treatment of transposition of the great arteries (TGA) and interventionist catheterization occurred when Rashkind and Miller 2 performed, in the hemodynamics laboratory, the first atrial septostomy with a balloon catheter introduced through the femoral vein without requiring a thoracotomy.

Classically described for the treatment of transposition of the great arteries, balloon atrial septostomy has also proved to be effective in patients with tricuspid atresia, total anomalous pulmonary venous drainage, pulmonary atresia with intact ventricular septum, and hypoplastic left heart syndrome 3-6.

Although balloon atrial septostomy has provided satisfactory results, the procedure is not free from complications. Rashkind 7 reported the possible problems associated with the procedure, such as difficulty in the venous access for the introduction of 5.5- or 6.5-F catheters in neonates, perforation of cardiac chambers, balloon rupture with embolism of rubber fragments, and lesions in the pulmonary vein, right and left atrioventricular valves, and inferior vena cava.

Perry et al 8 reported the use of echocardiography to help in locating and placing venous or arterial catheters in children. Even in the hemodynamics laboratory, echocardiography could contribute to positioning the catheters and decreasing the time of radiation exposure, mainly in complex heart diseases. After editing the article, the authors added a note reporting the echocardiographic follow-up of 3 balloon atrial septostomies.

The use of echocardiography for monitoring balloon atrial septostomy was initially reported by Matsunaga et al 9 in Japan, who performed the procedure in 8 children, 6 of whom had d-transposition of the great arteries, and 2 had total anomalous pulmonary venous drainage. They were followed by Allan et al 10 in England and Perry et al 11 in the United States of America.

The possibility of performing the procedure more rapidly in neonatal intensive care units with the need for neither transporting high-risk neonates nor using radiation has been reported by Baker et al 12. The latter authors performed balloon atrial septostomy at bedside, guided exclusively by echocardiography in 8 children with complete transposition of the great arteries.

In Brazil, the performance of balloon atrial septostomy at bedside under echocardiographic control has been reported by Serra et al 13. The procedure was performed in 9 newborn infants, 8 of whom had transposition of the great arteries and one had pulmonary atresia with intact ventricular septum. The lack of complications, wide atrial septal defect, and clinical improvement were results obtained in all cases. Mattos et al 14, in Brazil, performed atrial septostomy in the intensive care unit in 3 neonates, 2 of whom had transposition of the great arteries and one had tricuspid atresia. The procedure was unsuccessful in only one 36-day-old patient, who might have had stiffness of the atrial septum.

This study aimed at assessing the efficacy and safety of balloon atrial septostomy monitored with echocardiography in a pediatric cardiological unit of the Hospital Universitário.

Methods

This study was approved by the committee on ethics in research of our institution according to resolution 196/96 of the CNS. The procedure was performed after the guardians of the children gave written informed consent.

From August 1997 to January 2004, 31 children with congenital diseases underwent balloon atrial septostomy monitored exclusively with echocardiography. Twenty-eight procedures were performed at our institution's hospital, and 3 procedures were performed at other hospitals in the city.

We selected children with the echocardiographic diagnosis of congenital heart disease associated with atrial septal defect or restrictive oval foramen of an insufficient size to allow adequate blood mixture and to assure circulation. Atrial septal defect or oval foramen was considered restrictive when their diameter on echocardiography was smaller than 4 mm 15.

Based on the clinical suspicion of congenital heart disease, a 2-dimensional transthoracic echocardiography with color flow mapping was performed at bedside. The examinations performed at our hospital used an ATL device (Advanced Technology Laboratory, Bothell, WA), Apogee CX 200 model, with a 5-mHz mechanical transducer. The examinations performed at other hospitals used the Esaotebiomédica device (Florence, Italy), SIM 7000 CFM model, with a 5-mHz mechanical transducer.

Standard echocardiographic windows and views were used, according to the recommendations of the American Society of Echocardiography 16,17, and sequential segmentary analysis was performed 18.

Simple transposition of the great arteries was defined as the congenital heart disease with atrioventricular concordance, ventriculoarterial discordance, and possible associations with patent ductus arteriosus, patent oval foramen, or ventricular septal defect < 3 mm, or both 19.

The cyanotic neonates treated after the year 2000 received prostaglandin E1 at the dosage of 0.01 to 0.1 mg/kg/min in continuous intravenous infusion, even before diagnostic confirmation.

All children were under continuous monitoring of oxygen arterial saturation through pulse oximetry obtained with a Dixtal monitor (Dixtal Biomédica Ind. e Com. Ltda, Manaus, AM, Brazil), DX 2405 model. For analysis purposes, the oximetry values considered were those obtained immediately before the procedure (Sat 1) and approximately 30 minutes after its end (Sat 2).

The procedure was performed at bedside in patients whose diagnostic echocardiography was sufficient to provide the necessary information. The procedure was performed at the hemodynamics laboratory when complementation with angiographic study was required, such as in patients diagnosed with total anomalous pulmonary venous drainage, in those with transposition of the great arteries, whose echocardiogram did not precisely define the pattern of coronary circulation, and in those who required a better assessment of the pulmonary branches.

The routine for performing balloon atrial septostomy was as follows: a) patient placed in the dorsal decubitus position in a warmed crib, if a neonate; b) sedation with oral chloral hydrate at the dosage of 50 mg/kg/dose, if the patient was on spontaneous ventilation, or intravenous midazolam at the dosage of 0.1 mg/kg, if the patient was on mechanical ventilation; c) lower limbs fixed in abduction; d) asepsis and antisepsis of the right inguinal region; e) anesthetic infiltration of the right inguinal region with 2% lidocaine; f) preferential dissection of the right saphenous-femoral venous junction; the umbilical route was an alternative access; g) introduction of the 5-F or 6-F Rashkind catheter, whose models were the Rashkind balloon (USCI-CR Bard, Inc, Billerica, MA) and the Fogarty balloon catheter for atrial septostomy (Edwards-Baxter Healthcare); h) the procedure's echocardiographic monitoring was initiated with the echocardiographer wearing gloves, positioned contralaterally to the hemodynamics professional; the echocardiographic transducer was disinfected with povidone-iodine solution and placed on the subcostal window to show the anatomical details of venous drainage and the atria; i) under echocardiographic monitoring, the atrial septostomy catheter was passed through the inferior vena cava to the right atrium and oval foramen, then the left atrium, where it was inflated with saline solution, being then rapidly pulled back to the right atrium, and deflated; j) the following parameters were estimated with the aid of echocardiography: diameter of the inflated balloon, mobility of the margins, and diameter of the atrial septal defect after each procedure (the abbreviation ASD1 was used for atrial septal defect or oval foramen before the procedure, and ASD2 for the atrial septal defect resulting from the procedure); l) the initial procedure was performed with an inflated balloon diameter of 11 to 12 mm, which was progressively increased until the maximum limit of 15 to 16 mm.

The procedure was repeated until meeting the criterion of satisfaction, ie, the obtainment of a satisfactory atrial septal defect, defined as that whose diameter was > 4 mm and whose margins had wide mobility 15. The echocardiographic examinations were recorded on videotapes.

The statistical calculations were performed by using the GraphPad InStat program, version 3.05 for Windows 95/NT (GraphPad Software Inc, San Diego, California, USA), www.graphpad.com.

The mean, standard deviation, and median were analyzed for continuous variables, and the median was analyzed for discreet variables. The possible correlation between the number of balloon tractions and age was assessed. For comparison between the samples, the 2-tailed Wilcoxon test was used. The significance level of P<0.05 was adopted.

Results

Of the 31 infants undergoing echocardiographically assisted balloon atrial septostomy, 26 (83.9%) were of the male sex, and 5 (16.1%) were of the female sex.

Their mean age was 14.6±28.3 days, the median being 5 days (range: 1 to 150 days).

The mean weight was 3,328.8±947.1 g, and the median was 3,300 g (range, 1,800 to 7,500 g).

Transposition of the great arteries was the most frequent anatomical diagnosis found in 25 (80.6%) infants, followed by tricuspid atresia in 4 (12.9%), total anomalous pulmonary venous drainage in one (3.2%), and pulmonary atresia with intact septum in one (3.2%).

Of the 25 patients with transposition of the great arteries, 16 were of the simple type (absence or ventricular septal defect or ventricular septal defect < 3 mm), and 9 had the association of ventricular septal defect > 3 mm. Of the cases cited, one also had coarctation of the aorta, and 2 had pulmonary stenosis, one of the valvular type and another of the subvalvular type. The patient with pulmonary valve stenosis also had disconnected pulmonary arteries.

Of the infants with tricuspid atresia, 3 had ventriculoarterial concordance with pulmonary stenosis, and the other had ventriculoarterial discordance without pulmonary stenosis.

The case of total anomalous pulmonary venous drainage was of the supracardiac type.

Twenty (64.5%) infants were on prostaglandin E1 at the time of the procedure. The following 11 patients, who were treated before the year 2000, were not using that medication: one infant with total anomalous pulmonary venous drainage; 4 infants with tricuspid atresia; and 6 infants with transposition of the great arteries.

Seventeen (54.8%) infants were on assisted ventilation with orotracheal intubation, and the remaining infants were on spontaneous ventilation with oxygen supplementation.

Twenty-seven (87%) infants underwent the procedures at bedside in the ICU, except for the following: the infant with the total anomalous pulmonary venous drainage; the infant with tricuspid atresia; the infant with transposition of the great arteries whose coronary anatomy remained doubtful on echocardiography; and the infant with transposition of the great arteries associated with pulmonary valve stenosis and nonconfluent pulmonary arteries. In those cases, balloon atrial septostomy was performed in the hemodynamics laboratory at the time of cardiac catheterization, but also under exclusive echocardiographic control. Of the procedures performed at bedside, 3 were performed at hospitals that did not belong to our institution, but by the same team of hemodynamics professionals and echocardiographers. That was due to the severity of the patients' clinical findings, which did not allow the transfer of patients between hospitals.

The femoral access was used in 30 (96.8%) infants, and, in only one, the catheter was introduced through the umbilical route.

Echocardiography monitored and guided the position of the balloon catheter from the inferior vena cava until the left atrium. Undesired trajectories, such as to the superior vena cava, right or left ventricles, and loop formation in the right atrium, were immediately identified and corrected. Figure 1 shows the steps of the echocardiographically monitored procedure.


Figure 2 depicts an atrial septal defect resulting from balloon atrial septostomy observed during the Jatene surgery 5 days after the procedure, evidencing the satisfactory result obtained with rupture of the membrane of the fossa ovalis.


Thirty infants were considered for the statistical analysis of the size of the atrial septal defect, oxygen arterial saturation, and number of tractions. Patient #1, who marked the beginning of this study, was excluded for not having complete data. Table I shows the complete data of the patients.

Enlargement of the atrial septal defect size (fig. 3) to 4 mm or more was obtained in all patients, and it was the first satisfaction criterion for ending the procedure. The atrial septal defect size increased from 1.8 ± 0.8 mm to 5.8 ± 1.3 mm after the procedure (P < 0.0001).


Wide mobility of the margins of the atrial septal defect was obtained in all patients, and it was the second satisfaction criterion for ending the procedure.

Arterial oxygen saturation (fig. 4) increased from 64.5 ± 18.9% to 85.1 ± 9.2% (P < 0.0001) after the procedure. Only patients # 9 and 22 maintained the same saturation.


The number of tractions necessary for obtaining a satisfactory interatrial communication ranged from 2 to 9, with a median of 5. No correlation was observed between the patient's age and the number of tractions (R2 = 0.0285) (fig. 5).


Complications related to the procedures occurred in 5 (16.1%) patients: 3 balloon ruptures, one supraventricular tachycardia, one atrial flutter, and a lesion in the right femoral vein. Patient #7 had supraventricular tachycardia and balloon rupture. All cases of balloon rupture caused no embolic complications. The 2 cases of supraventricular arrhythmia were reverted only with manipulation of the catheter in the atrial cavities. Patient #30 had laceration of the right femoral vein during the introduction of the balloon catheter; after its ligature, the procedure was performed through the contralateral route, with no further damage to the right lower limb.

Discussion

Cyanotic neonates with severe clinical findings accounted for 87% of the patients in this study. Only the following infants were older than 30 days: one infant with total anomalous pulmonary venous drainage, one infant with tricuspid atresia, one infant with simple transposition of the great arteries, and one infant with transposition of the great arteries associated with ventricular septal defect and subvalvular pulmonary stenosis.

The older patient was the one with tricuspid atresia with ventriculoarterial concordance and pulmonary stenosis, who underwent echocardiography monitored balloon atrial septostomy at the age of 150 days. Despite the advanced age, a satisfactory interatrial communication of 8 mm was obtained after 6 tractions of the balloon. The child is on a 4-year follow-up, has completed the second stage of the Fontan operation (total cavopulmonary anastomosis), and is currently in good clinical condition.

Balloon atrial septostomy has been proved to be effective in infants below the age of 2 months 4 due to the lower resistance of their atrial septum; Jamjureeruk et al 20 have already reported successful procedures in infants up to the age of 6 months in a study with 47 infants with various diagnoses and ages ranging from 31 to 180 (mean, 83.5 ± 48.5) days.

Echocardiographic monitoring of balloon atrial septostomy allows the procedure to be performed in a more appropriate environment for the patient: at bedside in the ICU or even at medical services away from the referral center. Thus, it provides important advantages, such as avoiding the transportation of severely ill children with heart disease and reducing the time preceding the procedure.

Recently, Martin et al 21 have assessed the advantages of transporting the team of specialists of the Real Brompton Hospital in London to the local neonatal ICU, where the child with transposition of the great arteries was hospitalized to undergo echocardiographically monitored balloon atrial septostomy, thus avoiding the emergency transference of the child to the specialized hospital, with clinical, social, and organizational advantages of the specialized centers.

Even when cardiac catheterization was indicated, such as for infants with total anomalous pulmonary venous drainage and transposition of the great arteries, for defining the anatomy of the coronary or pulmonary arteries, the use of echocardiography resulted in a shorter time of exposure to radiation, greater safety for positioning the balloon catheter, and immediate assessment of the result.

Immediate assessment of the result at every traction of the balloon catheter, showing the size of the interatrial communication obtained and the mobility of its margins, contributes by avoiding the interruption of the procedure after only stretching the interatrial communication or oval foramen without tearing the atrial septum.

Baker et al 22, studied 43 children with transposition of the great arteries undergoing balloon atrial septostomy by using only hemodynamic criteria, such as equalization of the interatrial pressures, lack of resistance to traction of the inflated balloon, enlargement of the interatrial communication through balloon measurement, and improvement in arterial oxygen saturation. Those authors found a small interatrial communication in the postmortem examination, and another interatrial communication was no longer detected in the hemodynamic study performed due to progressive cyanosis, despite an initially satisfactory result. The authors concluded that an initially adequate interatrial communication does not guarantee long-term satisfactory results.

Mullins et al 4 studied 12 children with total anomalous pulmonary venous drainage undergoing balloon atrial septostomy without echocardiographic monitoring. Two of the children did not benefit from the procedure, and early surgical intervention was required when a small resultant interatrial communication was found, although the mean interatrial gradient was lower than one mmHg after septostomy in both. The lack of an interatrial pressure gradient even with a small interatrial communication may be consequent to a greater right atrial compliance. Balloon atrial septostomy in children with total anomalous pulmonary venous drainage is technically more difficult than in other cardiac defects, due to the small left atrial size and the difficulty in locating the catheter in the absence of usual hemodynamic criteria to identify the left atrium (presence of pulmonary veins, greater pressure, and greater oxygen saturation). Echocardiography plays an important role in such cases.

Both the enlargement of the atrial septal defect size and the elevation in the oxygen arterial saturation were statistically significant. Clinical and hemodynamic improvements have also been observed in all patients, mainly in the one with anomalous pulmonary venous drainage.

The possibility of identifying the catheter inside the heart and its relation with the cardiac structures defined on echocardiography makes the procedure undoubtedly safer, reducing the risk of perforations or lesions in the atrioventricular valves. Echocardiography may have contributed to the small number of complications related to the procedure (16.1%) and the absence of severe events in our study. The supraventricular arrhythmias had a short duration and were reverted only with the manipulation of the catheter inside the atria. The lesion in the right femoral vein evolved satisfactorily with no complication in the lower limb. The cases of rupture of the balloon resulted from problems with the material and were not directly related to the technique of the procedure.

Currently, the economic factor is also a point of interest in medicine. In this regard, Zellers et al 23 have also shown that the echocardiographic monitoring of balloon atrial septostomy at bedside is advantageous, with a cost of practically half of that performed at the hemodynamics laboratory.

In conclusion, the echocardiographic monitoring of balloon atrial septostomy allows the procedure to be performed at bedside, avoiding the transportation of severely ill patients, and enabling the adequate positioning of the balloon catheter and its relation with the cardiac structures without the use of ionizing radiation. It also allows the immediate assessment of the result of the procedure after each traction of the balloon catheter, which makes the procedure effective and safe.

Acknowledgments

We thank Paulo Roberto Pavarino, Hélio Augusto dos Reis Corbucci, Ricardo Rabelo, and Paula Fernanda da Matta for their support with the echocardiograms, and Carlos Guilherme Brando Fonseca, Sandro Oliveira Sacre, Maurício Jaramilo, and Fábio Augusto Selig for their support with cardiac catheterization.

References

Received for publication: 02/20/2004

Accepted for publication: 08/13/2004

English version by Stela Maris Costalonga

  • 1. Blalock A, Hanlon CR. The surgical treatment of complete transposition of the aorta and the pulmonary artery. Surg Gynec Obstet 1950; 90: 1-15.
  • 2. Rashkind WJ, Miller WW. Creation of atrial septal defect without thoracotomy. JAMA 1966; 196: 991-2.
  • 3. Mattos SS, Rodrigues JV, Severi R et al. Management of tricuspid atresia in neonates. Report of three cases and review of literature. J Pediatr (Rio J) 1994;70:33-8.
  • 4. Mullins CE, el-Said GM, Neches WH et al. Balloon atrial septostomy for total anomalous pulmonary venous return. Br Heart J 1973; 35: 752-7.
  • 5. Shams A, Fowler RS, Trusler GA, Keith JD, Mustard WT. Pulmonary atresia with intact ventricular septum: report of 50 cases. Pediatrics 1971;47:370-7.
  • 6. Barbero Marcial M, Tanamati C. Síndrome de hipoplasia do coração esquerdo. In: Santana MVT, editor. Cardiopatias Congênitas no Recém-Nascido Diagnóstico e Tratamento. São Paulo: Atheneu; 2000; 123-32.
  • 7. Rashkind WJ. The complications of balloon atrioseptostomy. J Pediatrics 1970; 76: 649-50.
  • 8. Perry LW, Galioto FM Jr, Blair T, Shapiro SR, Ruckman RN, Scott LP. Two-dimensional echocardiography for catheter location and placement in infants and children. Pediatrics 1981; 67: 541-7.
  • 9. Matsunaga S, Suzuki K, Ichinose E et al. Application of two dimensional echocardiography for the intracardic manipulation: the evaluation of atrial septal movement before and after ballon atrial septostomy. J Cardiogr 1981; 11:217-24.
  • 10. Allan LD, Leanage R, Wainwright R, Joseph MC, Tynan M. Ballon atrial septostomy under two dimensional echocardiographic control. Br Heart J 1982; 47: 41-3.
  • 11. Perry LW, Ruckman RN, Galioto FMJr, Shapiro SR, Potter BM, Scott LP. Echocardiographically assisted ballon atrial septostomy. Pediatrics 1982; 70:403-8.
  • 12. Baker EJ, Allan LD, Tynan MJ, Jones OD, Joseph MC, Deverall PB. Ballon atrial septostomy in the neonatal intensive care unit. Br Heart J 1984; 51: 377-8.
  • 13. Serra A, Chamiê F, Paupério H et al. Atriosseptostomia com balão através da ecocardiografia bidimensional. Arq Bras Cardiol 1988; 50: 179-82.
  • 14. Mattos SS, Severi R, Marques DL et al. Atriosseptostomia em unidade de terapia intensiva sob monitorização ecocardiográfica. Experiência inicial do Instituto do Coração de Pernambuco. Arq Bras Cardiol 1993; 61: 33-6.
  • 15. Baylen BG, Grzeszczak M, Gleason ME et al. Role of balloon atrial septostomy before early arterial switch repair of transposition of the great arteries. J Am Coll Cardiol 1992; 19: 1025-31.
  • 16. Henry WL, DeMaria A, Gramiak R et al. Report of the American Society of Echocardiography Committee on nomenclature and standards in two-dimensional echocardiography. Circulation 1980; 62: 212-8.
  • 17. Snider AR, Serwer GA, Ritter SB, editores. Echocardiography in pediatric heart disease. 2o ed. St. Louis (Missouri): Mosby-Year Book; 1997.
  • 18. Lucchese FA, Becker AE, Macartney FJ et al. Classificação das cardiopatias congênitas. Arq Bras Cardiol 1980; 35: 427-34.
  • 19. Liebman J, Culum L, Belloc NB. Natural history of transposition of the great arteries. Circulation 1969; 40: 237-62.
  • 20. Jamjureeruk V, Sangtawesin C, Layangool T. Balloon atrial septostomy under two-dimensional echocardiography control: a new outlook. Pediatr Cardiol 1997; 18: 197-200.
  • 21. Martin AC, Rigby ML, Penny DJ, Redington AN. Bedside balloon atrial septostomy on neonatal units. Arch Dis Child Fetal Neonatal Ed 2003; 88: F339-40.
  • 22. Baker F, Baker L, Zoltun R, Zuberbuhler JR. Effectiveness of Rashkind procedure in transposition of the great arteries in infants. Circulation 1971; 43 (suppl 1): 1-6.
  • 23. Zellers TM, Dixon K, Moake L, Wright J, Ramaciotti C. Bedside balloon atrial septostomy is safe, efficacious, and cost-effective compared with septostomy performed in the cardiac catheterization laboratory. Am J Cardiol 2002; 89: 613-5.
  • Correspondence to

    Carlos Henrique de Marchi
    Al. Dr. Ulysses da Silveira Guimarães, 273
    Cep 15061-723
    São José do Rio Preto, SP, Brazil
    E-mail:
  • Publication Dates

    • Publication in this collection
      19 Aug 2005
    • Date of issue
      Mar 2005

    History

    • Received
      20 Feb 2004
    • Accepted
      13 Aug 2004
    Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
    E-mail: revista@cardiol.br