Cochlear implantation in an animal model documents cochlear damage at the tip of the implant

Abstract

Introduction:

Electrocochleography has recently emerged as a diagnostic tool in cochlear implant surgery, purposing hearing preservation and optimal electrode positioning.

Objective:

In this experimental study, extra-cochlear potentials were obtained during cochlear implant surgery in guinea pigs. The aim was to determine electrophysiological changes indicating cochlear trauma after cochleostomy and after electrode implantation in different insertion depths.

Methods:

Normal-hearing guinea pigs (n = 14) were implanted uni- or bilaterally with a multichannel electrode. The extra-cochlear cochlear nerve action potentials were obtained in response to acoustic stimuli at specific frequencies before and after cochleostomy, and after introduction of the electrode bundle. After the electrophysiological experiments, the guinea pigs were euthanized and microtomography was performed, in order to determine the position of the electrode and to calculate of the depth of insertion. Based on the changes of amplitude and thresholds in relation to the stimulus frequency, the electrophysiological data and the position obtained by the microtomography reconstruction were compared.

Results:

Cochleostomy promoted a small electrophysiological impact, while electrode insertion caused changes in the amplitude of extra-cochlear electrophysiological potentials over a wide range of frequencies, especially in the deepest insertions. There was, however, preservation of the electrical response to low frequency stimuli in most cases, indicating a limited auditory impact in the intraoperative evaluation. The mean insertion depth of the apical electrodes was 5339.56 μm (±306.45 – 6 inserted contacts) and 4447.75 μm (±290.23 – 5 inserted contacts).

Conclusions:

The main electrophysiological changes observed during surgical procedures occurred during implantation of the electrode, especially the deepest insertions, whereas the cochleostomy disturbed the potentials to a lesser extent. While hearing loss was often observed apical to the cochlear implant, it was possible to preserve low frequencies after insertion. © 2020 Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial. Published by Elsevier Editora Ltda. This is an open access article under the CC BY license (http:// creativecommons.org/licenses/by/4.0/).

Keywords
Cochlear implant; Hearing preservation; Guinea pigs

Resumo

Introdução:

A eletrococleografia surgiu recentemente como uma ferramenta diagnóstica na cirurgia de implante coclear, objetiva a preservação da audição e o posicionamento ideal dos eletrodos.

Objetivo:

Determinar as alterações eletrofisiológicas indicativas de trauma coclear após a cocleostomia e após o implante do eletrodo em diferentes profundidades de inserção.

Método:

Neste estudo experimental, potenciais extracocleares foram obtidos durante a cirurgia de implante coclear em cobaias. Cobaias com audição normal (n = 14) foram implantadas uni- ou bilateralmente com eletrodo multicanal. Os potenciais de ação do nervo coclear extracoclear foram obtidos em resposta a estímulos acústicos em frequências específicas antes e após a cocleostomia e após a introdução do feixe de eletrodos. Após os experimentos eletrofisiológicos, as cobaias foram submetidas à eutanásia e a microtomografia foi feita para determinar a posição do eletrodo e calcular a profundidade de inserção. Com base nas mudanças de amplitude e limiares em relação à frequência do estímulo, os dados eletrofisiológicos e a posição obtida na reconstrução microtomográfica foram comparados.

Resultados:

A cocleostomia promoveu um pequeno impacto eletrofisiológico, enquanto a inserção do eletrodo causou alterações na amplitude dos potenciais eletrofisiológicos extra-cocleares em uma ampla faixa de frequências, especialmente nas inserções mais profundas. Houve, entretanto, preservação da resposta elétrica aos estímulos de baixa frequência na maioria dos casos, indicou um impacto auditivo limitado na avaliação intraoperatória. A profundidade média de inserção dos eletrodos apicais foi 5339,56 μm (± 306,45 - 6 contatos inseridos) e 4447,75 μm (± 290,23 - 5 contatos inseridos).

Conclusão:

As principais alterações eletrofisiológicas observadas durante os procedimentos cirúrgicos ocorreram durante o implante do eletrodo, especialmente nas inserções mais profundas, enquanto a cocleostomia alterou os potenciais em menor grau. Embora a perda auditiva seja frequentemente observada em posição apical ao implante coclear, foi possível preservar as baixas frequências após a inserção.

PALAVRAS-CHAVE
Implante coclear; Preservação auditiva; Cobaias

Introduction

The criteria for cochlear implant (CI) surgery have remarkably shifted in the last years, including patients with severe or even selective loss on mild-to-high frequencies.11 Lenarz T. Cochlear Implant – State of the Art. Laryngorhinootologie. 2017;96:123–51. Improved speech perception performance scores, especially in difficult hearing conditions, in patients treated by electro-acoustic stimulation (EAS, hybrid implants) compared to patients treated by either method alone, along with the prospect of future inner ear therap, have turned the preservation of residual hearing into a critical goal in the field.22 Von Ilberg C, Kiefer J, Tillein J, Pfenningdorff T, Hartmann R, Stürzebecher E, et al. Electric-acoustic stimulation of the auditory system. New technology for severe hearing loss. Orl. 1999;61:334–40.,33 Gantz BJ, Turner CW. Combining acoustic and electrical hearing. Laryngoscope. 2003;113:1726–30. Thus minimizing trauma to the delicate intra-cochlear structures is generally an desirable objective in modern CI surgery. Recent refinements in electrode design, as well as in surgical techniques, have contributed substantially to improve postoperative hearing.44 Giordano P, Hatzopoulos S, Giarbini N, Prosser S, Petruccelli J, Simoni E, et al. A soft-surgery approach to minimize hearing damage caused by the insertion of a cochlear implant electrode: a guinea pig animal model. Otol Neurotol. 2014;35:1440–5.,55 Gstoettner WK, van de Heyning P, O’Connor AF, Morera C, Sainz M, Vermeire K, et al. Electric acoustic stimulation of the auditory system: results of a multi-centre investigation. Acta Otolaryngol. 2008;128:968–75. However, there is limited data and high variability between available studies on hearing preservation results. The percentage of totally or partially compromised residual hearing is still high in CI patients, despite all efforts.66 Gantz BJ, Hansen MR, Turner CW, Oleson JJ, Reiss LA, Parkinson AJ. Hybrid 10 Clinical Trial: Preliminary Results. Audiol Neurotol. 2010;14:32–8.,77 Suhling M-C, Majdani O, Salcher R, Leifholz M, Büchner A, Lesinski-Schiedat A, et al. The impact of electrode array length on hearing preservation in cochlear implantation. Otol Neurotol. 2016;37:1006–15.,88 Jurawitz M-C, Büchner A, Harpel T Schüssler M, Majdani O, Lesinski-Schiedat A, et al. Hearing preservation outcomes with different cochlear implant electrodes: Nucleus® Hybrid™ -L24 and Nucleus Freedom™ CI422. Audiol Neurootol. 2014;19:293–309.

Determining the accurate electrode position as well as prompt detection of possible insertion damage is a key in current cochlear implant research. Electro-cochleography (ECochG) has recently emerged as a tool for evaluation of cochlear physiologic status and electrode insertion depth during surgery.99 Choudhury B, Adunka OF, DeMason CE, Ahmad FI, Buchman CA, Fitzpatrick DC. Detection of intracochlear damage with cochlear implantation in a gerbil model of hearing loss. Otol Neurotol. 2011;32:1370–8.,1010 Choudhury B, Adunka OF, Awan O, Pike JM, Buchman CA, Fitzpatrick DC, et al. Electrophysiologic consequences of flexible electrode insertions in gerbils with noise-induced hearing loss. Otol Neurotol. 2014;35:519–25. ECochG consists of 4 main components: compound action potentials (CAPs), a transient component at the beginning of the response commonly associated with auditory nerve activity;1111 Møller AR. On the origin of the compound action potentials (N1, N2) of the cochlea of the rat. Exp Neurol. 1983;80:633–44.,1212 Eggermont JJ. Analysis of compound action potential responses to tone bursts in the human and guinea pig cochlea. J Acoust Soc Am. 1976;60:1132–9. auditory nerve neurophonic (ANN), which also has neural origin; cochlear microphonic potentials (CMs), an AC component throughout the stimulus response originating from the outer hair cells (OHCs);1313 Dallos P, Cheatham MA. Compound Action Potential (AP) tuning curves. J Acoust Soc Am. 1976;59:591–7. and summating potentials (SPs), a DC component in large part depending on inner hair cells (IHCs) and OHC activity.1414 Zheng XY, Ding DL, McFadden SL, Henderson D. Evidence that inner hair cells are the major source of cochlear summating potentials. Hear Res. 1997;113:76–88.,1515 Durrant JD, Wang J, Ding DL, Salvi RJ. Are inner or outer hair cells the source of summating potentials recorded from the round window? J Acoust Soc Am. 1998;104:370–7.,1616 Cody AR, Russell IJ. Acoustically induced hearing loss: intracellular studies in the guinea pig cochlea. Hear Res. 1988;35:59–70. CAPs and CMs amplitude changes in monopolar recordings have been shown to be an early marker of contact between the electrode and cochlear structures or even minor and reversible damage, during CI insertion in gerbils.1717 Adunka OF, Mlot S, Suberman TA, Campbell AP, Surowitz J, Buchman CA, et al. Intracochlear recordings of electrophysiologic parameters indicating cochlear damage. Otol Neurotol. 2011;31:1233–41.,1818 Campbell AP, Suberman TA, Buchman CA, Douglas C, Adunka OF. Correlation of early auditory potentials and intracochlear electrode insertion properties: an animal model featuring near real-time monitoring. Otol Neurotol. 2011;31:1391–8.,1919 DeMason C, Choudhury B, Ahmad F, Fitzpatrick DC, Wang J, Buchman CA, et al. Electrophysiological properties of cochlear implantation in the gerbil using a flexible array. Ear Hear. 2012;33:534–42. A correlation between intraoperative ECochG potentials and postoperative hearing preservation has been described but is still being evaluated.2020 Adunka O, Roush P, Grose J, Macpherson C, Buchman C. Monitoring of cochlear function during cochlear implantation. Laryngoscope. 2006;116:1017–20.,2121 Haumann S, Imsiecke M, Bauernfeind G, Büchner A, Helmstaedter V, Lenarz T et al. Monitoring of the inner ear function during and after cochlear implant insertion using electrocochleography. Trends Hear. 2019;23:1–18.

The aim of the present study was to determine the site and the extent of cochleostomy and implantation trauma by documenting extra-cochlear CAP in normal-hearing guinea pigs undergoing CI surgery. The electrophysiological findings were correlated to insertion depth measured by the number of electrode channels inserted in the cochlea. We compared CAP amplitudes and thresholds in two different insertion depths in the scala tympani through cochleostomy.

With this objective, we analyzed the frequencies involved in threshold shifts at different steps in the surgery and hypothesized possible causes, such as a piston effect and basilar membrane contact or damage. The intra- cochlear electrode position reconstructed from postmortem micro-computer-tomography imaging (μCT, microtomography) was checked.

Methods

The Institutional Animal Care and Research Advisory Committee approved all experiments. The study has been conducted in accordance with the German “Law on Protecting Animals” (TierSchG) and with the European Communities Council Directive 86/609/EEC for the protection of animals used for experimental purposes.

In this study, 24 cochleae in 14 healthy and normal-hearing male Dunkin Hartley guinea pigs (Charles River Laboratories International Inc., Sulzfeld, Germany) weighing between 300 – 600 g were chosen as subjects.

Experimental procedures

Electrode technical details

The electrodes used in this study were custom-made research CI electrodes (Med-El Inc., Innsbruck, Austria), comprised of 6 platinum-iridium contacts spaced by 700 μm, 0.5 mm maximum diameter of the silicone carrier, tapered to 0.3 mm at the tip, and had a 4.5 mm insertable length covered by contacts (from tip to the last contact). In accordance with the standard numbering for MedEl CIs, the most apical contact will be referred as channel or contact 1 (“Ch1”), and the most basal contact as channel or contact 6 (“Ch6”), and the intermediate contacts are numbered accordingly.

Anesthesia and monitoring

Fourteen subjects were implanted uni- or bilaterally (24 ears). All measurements and surgical procedures were performed under general anesthesia. The animals were anesthetized initially by intramuscular injection of 50 mg/kg 10% ketamine solution (CP Pharma, Burgdorf, Germany), 10 mg/kg 2% xylazine solution (WDT, Garbsen, Germany) and 0.1 mg/kg atropine sulfate (B. Braun, Melsungen, Germany). Follow-up anesthesia in 45 min’ time intervals was applied with 25%-30% of initial dose, substituting atropine sulfate by Ringer acetate solution. Anesthesia level was assessed by continuous monitoring of heart rate (using electrocardiogram) and capnometry (CO2%/vol.). The absence of paw-withdrawal and corneal reflexes was tested in regular intervals to verify anesthesia depth. Ample shaving of the skin behind the pinnae, around the neck, and on the vertex of the skull was performed. The animals were kept on a heating blanket (TC-1000 Temperature Controller, CWE Inc., Ardmore, USA). Core temperature was continuously monitored by a rectal probe and kept above 38 ◦C. A custom-made endotracheal tube was inserted through a tracheotomy and was connected to the capnometer (Normocap CO2 & O2 Monitor, Datex, Helsinki, Finland). Ventilation via endotracheal tube was performed using a rodent ventilator (Ugo Basile, Gemonio, Italy) and tidal volume was adjusted to the animal’s body weight. Respiratory rate was set between 50 and 60 per second depending on end tidal CO2 concentration that was kept between 3% and 4%. Animal heads were secured in a customized rodent head holder that allowed adjustment along three axes. All measures were performed in a sound-attenuated chamber.

Auditory brainstem response (ABR) measurements

Subcutaneous Ag/AgCl electrodes were placed retroauricularly on both sides (references) of the head and at the vertex (recording electrode) for recordings. A thick Ag/AgCl electrode was subcutaneously inserted at the scruff (ground electrode). The initial hearing thresholds for both ears before surgery were determined by recording auditory brainstem responses (ABRs). A series of clicks of increasing sound levels (0 – 80 dB SPL in 5 dB steps, 50 μs click duration, 100 of each condensation and rarefaction stimulation averaged per stimulus level) were presented through a dynamic headphone speaker (DT48, Beyerdynamic, Heilbronn, Germany). The loudspeaker was positioned 1 cm away from pinna in close free-field conditions. The signals were generated, and responses acquired through the software AudiologyLab (Otoconsult, Frankfurt, Germany) via, a DA/AD converter (NI-6259, National Instruments, Austin, TX, USA). The stimuli were send through an attenuator (ATT7, Otoconsult, Frankfurt, Germany) and the recorded signals were filtered (Butterworth filter 6thorder, high-pass filter frequency 200 Hz, low-pass filter frequency 5 kHz) and amplified by 80 dB using a filter and amplifier combination (F1 device, Otoconsult Comp., Frankfurt/M., Germany). Middle ear and cochlear function were assumed normal for ABR threshold-levels of 30 dB SPLpeak equivalent or better. All ears included in this study satisfied this criterion.

Surgery and cochlear implant insertion

The following measurements and procedures from right or left ears were carried out sequentially (Fig. 1). Local anesthesia (2% lidocaine, 2 mL) was injected in the skin around the pinna before a complete pinna resection. Muscle and soft tissue were dissected, and the posterior-lateral part of the bulla tympani was exposed and opened with a hollow needle under stereomicroscopic visualization (Carl Zeiss OPMI® pico, Carl Zeiss, Goettingen, Germany). The lateral bulla wall was further exposed and carefully removed using a micro- rongeur to access the middle ear and the basal turn of the cochlea, including the round window (RW), without damaging the tympanic membrane. A silver ball electrode was positioned and fixed by a mechanical holder at RW inferior lip and kept in place throughout the procedure. The loudspeaker was placed near the entrance of the cartilage of the external auditory canal using a conical adapter without displacing or obstructing the ear canal. Cochleostomy was performed 0.5 mm ventral to the RW, with a 0.6 mm diamond drilling burr at 4000rpm and care was taken to avoid bone dust or blood to enter the cochlea. The CI electrode was inserted until all 6 CI contacts were placed in the scala tympani (hereinafter referred to as “6Ch in”) or until 5 CI contacts were inserted (hereinafter referred to as “5Ch in”).

Figure 1
Intraoperative microscopic view (left). D, Dorsal; V, Ventral; A, Anterior; P, Posterior; be, Silver ball electrode; rw, round window. Monopolar compound action potential (upper right); red bar represents stimulus duration. Schematic image of the electrode used in experiments (lower right).

After finishing the experimental procedures, the animals were euthanatized with a transcardial infusion of 2 mL sodium pentobarbital (Release 300, WDT, Garbsen, Germany). The silastic CI carrier was then fixed in position at the rim of the tympanic bulla (Histoacryl; B. Braun Melsungen AG, Melsungen, Germany). The cadavers were decapitated, and the heads preserved in 3.5%–3.7% formaldehyde buffered solution before the following μCT imaging procedure.

Cochlear recordings

Extra-cochlear compound action potentials (CAP) were recorded using the silver ball electrode at the RW lip. Tone bursts of 5 ms duration (2.5 ms rise/fall periods) were presented in randomized order at frequencies between 2 and 32 kHz (with logarithmic increments of 4 steps/octave), with sound pressure levels between 0 and 90 dB SPL in 10dB steps. Stimuli were generated using the software AudiologyLab 25 (Otoconsult, Frankfurt, Germany). The sound output was calibrated using a ¼-in. microphone (type 4939, Brüel&Kjaer, Nærum, Denmark) in combination with a preamplifier (type 2670, Brüel&Kjaer, Nærum, Denmark) and a conditioning amplifier (type 2690, Nexus, Brüel&Kjaer, Denmark) connected to the National Instruments board. The recording signals were filtered between 5 Hz and 5 kHz and averaged over 20 stimulus presentations. CAP measurements were performed following the opening of the bulla, immediately after cochleostomy and after the CI insertion.

Data analysis

Recorded data was analyzed in the range of 4 octaves (2 – 32 kHz) at 9 sound levels (0-80dBSPL) using custom-made MATLAB routines (MathWorks, Natick, MA, USA), Prism (Graph Pad Software Inc., La Jolla, CA, USA) and Microsoft Excel (Professional Edition, 2007, Microsoft Corp, Redmond, WA). We extracted the compound action potentials (CAPs), from the electrocochleographic potential by differentially adjusted offline filters (Fig. 1). During the offline analysis of the CAPs, the recording signal was averaged over both condensation and rarefaction stimulus presentations to largely exclude CMs. The recording signal was filtered between 200 Hz and 5 kHz. CAPs were identified by the morphology of the typical N1 and P1 components at the beginning of the stimulus (analysis window between 0 and 10 ms) and the peak-to-peak amplitude was computed. The threshold value was defined as the lowest intensity able to evoke a response two times larger than the baseline amplitude.

Statistical comparisons were performed using the paired Wilcoxon test and Mann-Whitney, for independent groups. A probability of p < 0.05 was considered as significant. Results were expressed as mean ± standard error of the mean (SEM), unless stated differently.

Micro-computed tomography imaging and three-dimensional reconstruction

Guinea pig heads with the implants fixed in place and preserved in 3.5%–3.7% formaldehyde buffered solution were scanned on a micro-computed tomography system (μCT, XtremeCT II, SCANCO Medical AG, Brüttisellen, Switzerland) using a defined protocol for a voxel size resolution of 17.0 μm (68 kVp, 1470 μA, 100W).

A high resolution μCT Dataset of an unimplanted, left cochlea was used to reconstruct the length of the basilar membrane along the midpoint between osseous spiral lamina and the attachment point of the spiral ligament in the 3D analysis software Amira (Version 6.0.1, FEI Corp., Hillsboro, OR, USA). The reconstruction included the hook region (curved basal region, adjacent to the round window). All implanted specimens were registered to this template in Amira. Right cochleae were mirrored to enable the registration. Then the coordinates of the contacts were measured. Thereafter, in order to turn this three-dimensional data (spiral shape), in a two-dimensional plane, values were rotated and projected in a rostro caudal two-dimensional representation of the cochlea. The projection was rotated to represent a view along the mid-modiolar rostro-caudal axis intersecting with the vertical perpendicular to summit of the basilar membrane before turning into the hook region (summit point). In this manner the length of the basilar membrane in percent, as described by Viberg et al.,2222 Viberg A, Canlon B, Engstro H. The guide to plotting a cochleogram. 2004;197:1–10. the insertion depth (last contact as endpoint of insertion relative to basilar membrane length) and the insertion angle (relative to the perpendicular from summit point to the midmodiolar axis) were determined (Fig. 2).

Figure 2
Insertion path reconstruction. A, Dorsal and lateral view of μCT data from an implanted cochlea (yellow, M) to the unimplanted template (grey, T) is depicted. The registration was performed to obtain the most agreement between the bony cochleae. The position of the CI with the μCT blooming artefacts is shown in blue (ci: cochlear implant; mo: modiolus; rw: round window; sm: scala media; sv: scala vestibuli). B, Lateral and rostral view of a digital radiograph view (from μCT data in Amira) of the non-implanted template with the reconstructed basilar membrane position (bm) and a superimposed exemplary cochlea implant position (cip) reconstruction. The images show the mid-modiolar axis (ma), the summit point of the basilar membrane reconstruction (sp) and the perpendicular to the modiolar axis (p). The measurement of the insertion angle (here 185.4◦) is exemplified in the rostral view. C, Two dimensional projections in basal to apical direction of 15 cochlear implant positions from μCT data relative to the basilar membrane reconstruction. A pronounced difference in insertion depth between 5 contact (5Ch in, dark grey) and 6 contact (6Ch in, light grey) insertions is visible (round window and hook region on top, projection centered to mid-modiolar axis, compare B).

The frequency-position of the electrode channels was also calculated, according to the species-specific Greenwood equation, in which 2.59mm (~14% of the cochlear length) distance inside the cochlea corresponds to 1 octave (2.59 mm/octave).

Electrophysiological data was analyzed from 24 ears in 14 guinea pigs and of those, μCT images were obtained from 20 ears in 10 animals. However, five datasets had to be excluded because of postmortem displacement of the CI or imaging problems; 15 ears of 9 animals were then included in the μCT analysis.

Results

Extra-cochlear responses obtained with the silver ball electrode were evaluated as minimum thresholds at which a CAP was visible in the 2-32 kHz frequency range. Implantations could be anatomically reconstructed in 15 cochleae (Fig. 2) and showed different implantation depths in different animals (see below). Deepest implantations in optimal conditions may approach 270◦ insertion angle in the guinea pig. In the present experiments, no anatomical implantation trauma was detected in μCT images.

In the present experiments normal hearing animals were used; these are considered an extreme case that would not be implanted in clinical condition. However, this condition provides the optimal case for comparing the implantation trauma along the whole cochlear partition at high sensitivity.

Data from the 24 investigated cochleae are shown in Fig. 3A, displayed as mean and standard error of the mean (SEM). Immediately after cochleostomy, Threshold shifts (TS) were minimal or absent, and non-significant at all frequencies evaluated. However, in all cases electrode insertion caused some threshold shift. Significant TS occurred mainly at frequencies from 4.7 – 9.5 kHz. Responses to frequencies from 2.0 – 4.0kHz were less affected by the implantation. Considering the length of the electrode (3.5mm from apical to basal contact) and the Greenwood function,2222 Viberg A, Canlon B, Engstro H. The guide to plotting a cochleogram. 2004;197:1–10. insertion induced shifts around the apex of the implant, in frequencies extending approximately 0.5–1 octave (or 1.3–2.6mm) apically beyond the electrode coverage. This was confirmed by the implantation depths anatomically reconstructed in the 15 cochleae with μCTs images available (Fig. 2).

Figure 3
Extra-cochlear CAP thresholds: A, Grand average of all ears; B, 5 channel insertions; C, 6 channel insertions; D, Threshold shifts for 5 and 6 channel insertions; D, Statistical test: Wilcoxon signed rank; *p ≤ 0.05, error bars ± standard error.

On average, TS reached at maximum 16.5±2.7dB, considering the 24 ears. This was a very mild hearing loss. The implantations as reported during surgery were subsequently divided to 5 contacts implantations (i.e. shallower implantations) and 6 contacts implantations (i.e. deeper implantations). In the 6 contact insertions (denoted in what follows as 6Ch) presented larger shifts (23.0 ± 3.5dB, maximum TS) when compared to 5 contact insertions (denoted in what follows as 5Ch; 10.0 ± 4.5dB maximum TS) (Fig. 3 B and C). TS were observed in the 4.7 and 5.6 kHz frequencies exclusively in the 6Ch group (not observed in the 5Ch group). Directly comparison of 6Ch to 5Ch groups revealed statistically significant differences in the frequency range of 5.7 – 9.5 kHz (Fig. 3D).

Peak-to-peak amplitudes of suprathreshold CAPs were analyzed subsequently (Fig. 4), for different frequencies (2-32 kHz) and 50, 60 and 70 dB SPL. There were no differences between initial and cochleostomy conditions for neither stimulus level. Electrode insertion, however, induced a statistically significant decrease in CAP amplitudes at frequencies 6.7, 8.0 and 9.5 for 70, 60 and 50dB level of stimulation when all cases were considered together (Fig. 4 A, B and C), mirroring the TS shown in Fig. 3.

Figure 4
Extra-cochlear CAP amplitudes of all cases on average, at different stimulus intensities: (A) 70 dB; (B) 60 dB; (C) 50 dB. Statistical test, Wilcoxon signed rank; *p ≤ 0.05, error bars ± standard error.

Subsequently we compared the deep and shallow implantations. With 5Ch implanted, we did not find statistically significant implantation effects on the amplitudes for all levels tested (Fig. 5 A, B and C). On the other hand, the 6Ch group showed amplitude changes covering a broad frequency range of 4 – 16kHz (Fig. 5A, B, C, D, E, F).

Figure 5
5 vs. 6 channel insertions. Upper graphs (A – C) depict the amplitudes obtained for 5 channel insertions with different stimulus intensities, while lower graphs (D – F) depict the amplitudes obtained for 6 channel insertions with different stimulus intensities, accordingly. 6 channel insertions present significant decreases in CAP amplitudes for different stimulus intensities, which are not observed in 5 channel insertions. Statistical test: Wilcoxon signed rank; *p ≤ 0.05, error bars ± standard error.

Amplitude change for all cases together is evaluated in Fig. 6 for both groups pooled together. There were no significant amplitude changes detected after cochleostomy (Fig. 6A). After implantation, threshold shifts were found mainly from 5.6–11.3 kHz. When dividing the cases between 5Ch and 6Ch groups, larger amplitude changes and a broader frequency spectrum in the deeper insertion group (6Ch group) were identified (Fig. 7 A, B, C). Amplitude changes were particularly larger for the 6Ch group at 4.7 and 9.5 kHz.

Figure 6
Extra-cochlear CAP presented as amplitude changes of all cases on average: (A) after cochleostomy; (B) after electrode insertion. Statistical test: Wilcoxon signed rank; *p≤ 0.05 error bars ± standard error.

Figure 7
Amplitude changes of CAP after 5 and 6 channel insertions, at different stimulus intensities (70, 60 and 50 dB). Statistical test: Mann-Whitney; *p ≤ 0.05, error bars ± standard error.

From all insertions reconstructed in μCT the median lowest characteristic frequency reached by the implanted electrode was 8.04 kHz. The deepest insertion in our subset reached 5.51 kHz, whereas the shallowest insertion reached 9.96 kHz.

Implantations in 5Ch group covered less than 33% (average: 31,3%; ±1.1%) of the full basilar membrane length. On the other hand, the 6Ch group showed deeper implantations: more than 34% of the full basilar membrane length in every case (average: 37,1%; ±2.5%). The average insertion depth of the most apical electrode (Ch1) was 5340 ± 307 μm in 6Ch group vs. 4448 ± 290 μm in 5Ch group.

Finally, the electrode depth and its calculated frequency positions were related to the threshold shift caused by the implantation (post threshold – pre threshold). There was no significant correlation between the insertion depth and the maximal hearing loss Fig. 8A. A significant positive relation between the frequency position of electrode contact 1 and hearing loss at 5.6 kHz was observed (Fig. 8B). The 5.6 kHz is the lowest frequency in which a significant is detected and was therefore chosen. This result documents that there are other factors beyond implantation depth that are responsible for the largest hearing loss observed, but the previous data clearly show that the danger of hearing loss per se is higher if deeper implantations are attempted.

Figure 8
Linear regression between electrode depth estimated by μCT and the thresholds shifts after insertion: as compared to shallower insertions (blue dots), deeper insertions (red dots) correlate to greater maximal hearing loss (A). The frequency position calculated by Greenwood function of electrode contact 1 was correlated to hearing loss at 5.6 kHz (B). *p ≤ 0.05; ns: non-significant.

Discussion

The present study evaluates implantation trauma in the context of CI surgery in normal-hearing guinea pigs. This exceptional condition, not accessible in humans, allowed comparing the implantation trauma specifically at different cochlear partitions. Furthermore, the experiments allowed relating the position of the electrode using anatomical reconstruction with a μCT.

This study demonstrated that it is possible to implant a hearing ear without hearing loss2323 Sato M, Baumhoff P, Kral A. Cochlear implant stimulation of a hearing ear generates separate electrophonic and electroneural responses. J Neurosci. 2016;36:54–64. provided implantations avoided insertion till a resistance was felt (5Ch group; comparable to previous findings).2424 Helmstaedter V, Lenarz T Erfurt P, Kral A, Baumhoff P. The summating potential is a reliable marker of electrode position in electrocochleography: Cochlear implant as a theragnostic probe. Ear Hear. 2018;39:687–700. Similarly, in human temporal bones it could be demonstrated that cochlear damage cannot be always detected by the surgeon because the related forces are below the sensitivity of the human hand.2525 Avci E, Nauwelaers T, Hamacher V, Kral A. Three-dimensional force profile during cochlear implantation depends on individual geometry and insertion trauma. Ear Hear. 2017;38: 168–79. Avoiding the resistance should therefore be one goal of cochlear implantation.

Deeper electrode insertions had a larger impact on the inner ear function. The elicited tone-specific CAP indirectly reflects electrode position by its frequency specificity. TS and amplitude changes increased also in lower frequencies when the electrode reached cochlear regions in the 2nd half of the basal turn (beyond 180°). Despite this the results also showed that electrode insertion may be accomplished causing virtually no effect in apical-most frequency regions (< 4 kHz). Usually, electrophysiological interference spreads beyond electrode tip, towards frequencies represented in second cochlear turn. This may be the consequence of the damage that affects the basilar membrane or be due to a piston-type of effect when the scala tympani is filled out by the cochlear implant (as also seen previously).2424 Helmstaedter V, Lenarz T Erfurt P, Kral A, Baumhoff P. The summating potential is a reliable marker of electrode position in electrocochleography: Cochlear implant as a theragnostic probe. Ear Hear. 2018;39:687–700.

Electrode insertion could be accomplished in some cases with restricted hearing loss, keeping several frequency ranges unaltered. In this data series, we did not check for reversibility of the threshold shifts by withdrawing the electrode, as other authors99 Choudhury B, Adunka OF, DeMason CE, Ahmad FI, Buchman CA, Fitzpatrick DC. Detection of intracochlear damage with cochlear implantation in a gerbil model of hearing loss. Otol Neurotol. 2011;32:1370–8.,1010 Choudhury B, Adunka OF, Awan O, Pike JM, Buchman CA, Fitzpatrick DC, et al. Electrophysiologic consequences of flexible electrode insertions in gerbils with noise-induced hearing loss. Otol Neurotol. 2014;35:519–25.,2626 DeMason, Christine Baishakhi Choudhury, Ahmad Faisal, Fitzpatrick Douglas C, Wang Jacob, Craig A, et al. Electrophysiological properties of cochlear implantation in the gerbil using a flexible array. Ear Hear. 2013;33:534–42. did, which could help reveal the etiology of the amplitude shifts, as mechanical interference to basilar membrane.

It is interesting that the implantation trauma in the present cases involved mainly the electrode tip.2727 Wanna GB, Noble JH, McRackan TR, Dawant BM, Dietrich MS, Watkins LD, et al. Assessment of electrode placement and audiological outcomes in bilateral cochlear implantation. Otol Neurotol. 2011;32:428–32.,2828 Finley CC, Holden TA, Holden LK, Whiting BR, Chole RA, Neely GJ, et al. Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol. 2008;29:920–8.,2929 O’Leary MJ, House WF, Fayad J, Linthicum FH. Electrode insertion trauma in cochlear implantation. Ann Otol Rhinol Laryngol. 1991;100:695–9. Trauma in human temporal bones has been observed at the base of the cochlea due to implant buckling after resistance is encountered.2525 Avci E, Nauwelaers T, Hamacher V, Kral A. Three-dimensional force profile during cochlear implantation depends on individual geometry and insertion trauma. Ear Hear. 2017;38: 168–79. We did not observe such damage here. Whether this is due to different anatomy of the guinea pig cochlea, due to dedicated implants for guinea pig with different mechanical properties or due to different angle in inserting the implant to the cochlea (the position of cochleostomy in guinea pig vs. human cochlea).

In some cases, we observed amplitude increases in some frequencies after cochleostomy or even electrode insertion, which could be due to perilymph leakage and contact to extra-cochlear electrode. This phenomenon could increase signaling because of direct intra-cochlear potential detection. Another possibility is that the effect of stiffening due to the electrode in basal turn may increase motion in more apical turns.3030 Kiefer J, Böhnke F, Adunka O, Arnold W. Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode. Hear Res. 2006;221:36–43.

ECochG response changes during CI are consistent with previous results3131 Dalbert A, Huber A, Veraguth D, Roosli C, Pfiffner F. Assessment of cochlear trauma during cochlear implantation using electrocochleography and cone beam computed tomography. Otol Neurotol. 2016;37:446–53.,3232 Dalbert A, Pfiffner F, Hoesli M, Koka K, Veraguth D, Roosli C, et al. Assessment of cochlear function during cochlear implantation by extra- and intracochlear electrocochleography. Front Neurosci. 2018;12:1–9.,3333 Dalbert A, Rohner P, Roosli C, Veraguth D, Huber A, Pfiffner F. Correlation between electrocochleographic changes during surgery and hearing outcome in cochlear implant recipients: a case report and systematic review of the literature. Otol Neurotol. 2020;41:318–26. showing smaller CAP amplitudes and higher thresholds, indicating cochlear damage. We did not address cochlear damage directly by means of hair cell counts or other histologic techniques, whereas we consider that our aim is not to elucidate the etiology of cochlear trauma. Direct comparisons between histology and electrophysiological data may be variable, as seen in other studies.99 Choudhury B, Adunka OF, DeMason CE, Ahmad FI, Buchman CA, Fitzpatrick DC. Detection of intracochlear damage with cochlear implantation in a gerbil model of hearing loss. Otol Neurotol. 2011;32:1370–8.,1919 DeMason C, Choudhury B, Ahmad F, Fitzpatrick DC, Wang J, Buchman CA, et al. Electrophysiological properties of cochlear implantation in the gerbil using a flexible array. Ear Hear. 2012;33:534–42. Previous studies have issued ECochG as a guide during cochlear implantation, identifying markers of mechanical damage and correlating with post-operative performance.2121 Haumann S, Imsiecke M, Bauernfeind G, Büchner A, Helmstaedter V, Lenarz T et al. Monitoring of the inner ear function during and after cochlear implant insertion using electrocochleography. Trends Hear. 2019;23:1–18. Magnitude of CAP and CM amplitude loss can be correlated to the degree of histological damage caused by basilar membrane penetrations.1717 Adunka OF, Mlot S, Suberman TA, Campbell AP, Surowitz J, Buchman CA, et al. Intracochlear recordings of electrophysiologic parameters indicating cochlear damage. Otol Neurotol. 2011;31:1233–41. Here we cannot provide the eventual explanation of the cochlear trauma. While several candidate mechanisms may be involved (direct mechanical damage to hair cells by lifting the basilar membrane; detachment of tectorial membrane from outer hair cells; interference with the potassium recycling or mechanical damage to the stria vascularis) and basilar membrane damage remains a possibility; the small effect size observed here likely indicates a predominant influence on the outer hair cells.

The scala tympani cross sectional area as a function of distance from round window was characterized by Ghiz et al.3434 Ghiz AF, Salt AN, DeMott JE, Henson MM, Henson OW, Gewalt SL. Quantitative anatomy of the round window and cochlear aqueduct in guinea pigs. Hear Res. 2001;162:105–12. The electrode used in our study has a cross sectional area of 0.196mm2 (0.5mm diameter), with 4.5mm maximum length. In guinea pigs, the cross-sectional area of the scala tympani abruptly diminishes towards apical in the first turn region (from circa 1.28 mm2 at the widest point of hook area to circa 0.29 mm2 at 5 mm round window distance).3434 Ghiz AF, Salt AN, DeMott JE, Henson MM, Henson OW, Gewalt SL. Quantitative anatomy of the round window and cochlear aqueduct in guinea pigs. Hear Res. 2001;162:105–12. Our 5Ch cases have presumably reached ST in larger cross-sectional areas compared to our 6Ch insertions (700 micrometres longer), which reached narrower regions. The surrounding ST microenvironment in which the electrode is positioned is probably determinant of the differences encountered in our insertions. The narrower the region, the lesser amount of perilymph around and closer the electrode is to the organic signal source. Also, the possibility of basilar membrane contacts or even damage when the electrode advances further in ST is higher.

We are aware that damage to cochlear microanatomy and hearing loss after CI may occur weeks to months after surgery,3535 Rowe D, Chambers S, Hampson A, Eastwood H, O’Leary S. The effect of round window sealants on delayed hearing loss in a guinea pig model of cochlear implantation. Otol Neurotol. 2016;37:1024–31. which is a main concern when hearing preservation is a goal. Many factors are involved, including inflammation, fibrosis, hydrops, mechanical effects (movement of the electrode) and even idiopathic factors.3636 James DP, Eastwood H, Richardson RT, O’Leary SJ. Effects of round window dexamethasone on residual hearing in a guinea pig model of cochlear implantation. Audiol Neurotol. 2008;13:86–96. These events may take place with or without clearly observable histologic and/or radiologic damage. We did not observe in our μCT sequences gross trauma over the spiral osseous lamina or other transgressions in the electrode path inside the cochleae. Nevertheless, μCT has limited accuracy and resolution for detecting trauma in membranous structures, such as the organ of Corti and stria vascularis.

Our data suggests that shorter electrodes have lesser impact on cochlear physiology, on an intraoperative setting. A present dilemma is that although short electrodes seem safer, they may offer less performance to CI users, mainly but not only in the context of electroacoustic stimulation. Ideal placement and positioning are therefore desirable. Intraoperative events in CI surgery are potentially crucial to surgical decisions, and electrophysiology may be a tool to indicate the choice of an electrode over another. Finally, the human cochlear anatomy is highly variable.3737 Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A. Variations in microanatomy of the human cochlea. J Comp Neurol. 2014;522:3245–61.,3838 Erixon E, Högstorp H, Wadin K, Rask-Andersen H. Variational anatomy of the human cochlea: Implications for cochlear implantation. Otol Neurotol. 2009;30:14–22. The present study suggests that predictions of the implantation process may prevent too- deep implantations, and thus precise planning in surgery, using imaging,3939 Würfel W, Lanfermann H, Lenarz T, Majdani O. Cochlear length determination using cone beam computed tomography in a clinical setting. Hear Res. 2014;316:65–72.,4040 Labadie RF, Noble JH. Preliminary results with image-guided cochlear implant insertion techniques. Otol Neurotol. 2018;39:922–8. supported by cochlear models4141 Pietsch M, Aguirre Dávila L, Erfurt P, Avci E, Lenarz T, Kral A. Spiral Form of the human cochlea results from spatial constraints. Sci Rep. 2017;7:1–11.,4242 Schurzig D, Timm ME, Batsoulis C, Salcher R, Sieber D, Jolly C, et al. A novel method for clinical cochlear duct length estimation toward patient-specific cochlear implant selection. OTO Open. 2018;2:1–8.,4343 Escudé B, James C, Deguine O, Cochard N, Eter E, Fraysse B. The size of the cochlea and predictions of insertion depth angles for cochlear implant electrodes. Audiol Neurotol. 2006;11: 27–33. and electrophysiology2424 Helmstaedter V, Lenarz T Erfurt P, Kral A, Baumhoff P. The summating potential is a reliable marker of electrode position in electrocochleography: Cochlear implant as a theragnostic probe. Ear Hear. 2018;39:687–700.,4444 Adunka OF, Mlot S, Suberman TA, Campbell AP, Surowitz J, Buchman CA, et al. Intracochlear recordings of electrophysiological parameters indicating cochlear damage. Otol Neurotol. 2010;31:1233–41.,4545 Giardina CK, Khan TE, Pulver SH, Adunka OF, Buchman CA, Brown KD, etal. Response changes during insertion of a cochlear implant using extracochlear electrocochleography. Ear Hear. 2018;39:1146–56. might help to prevent cochlear trauma.

Conclusion

In this guinea pig model of cochlear implantation, cochleostomy induced little or none impact in the extra-cochlear compound action potentials. Deeper insertions led to greater hearing loss, including the cochlear region apical to the implant position. The hearing loss was restricted to a predicable frequency range around the tip of the cochlear implant.

  • Funding
    J.S.C.A received a scholarship from CAPES Foundation, Ministry of Education of Brazil, Brasília – DF 70.040-020, Brazil.
  • Peer Review under the responsibility of Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial.

Acknowledgements

The study has been supported by Deutsche Forschung gemeinschaft within the Cluster of Excellence “Hearing 4 All” andMedElComp., Innsbruck, Austria. The authors thank Daniela and Karl-Jürgen Kühne for their technical support.

References

  • 1
    Lenarz T. Cochlear Implant – State of the Art. Laryngorhinootologie. 2017;96:123–51.
  • 2
    Von Ilberg C, Kiefer J, Tillein J, Pfenningdorff T, Hartmann R, Stürzebecher E, et al. Electric-acoustic stimulation of the auditory system. New technology for severe hearing loss. Orl. 1999;61:334–40.
  • 3
    Gantz BJ, Turner CW. Combining acoustic and electrical hearing. Laryngoscope. 2003;113:1726–30.
  • 4
    Giordano P, Hatzopoulos S, Giarbini N, Prosser S, Petruccelli J, Simoni E, et al. A soft-surgery approach to minimize hearing damage caused by the insertion of a cochlear implant electrode: a guinea pig animal model. Otol Neurotol. 2014;35:1440–5.
  • 5
    Gstoettner WK, van de Heyning P, O’Connor AF, Morera C, Sainz M, Vermeire K, et al. Electric acoustic stimulation of the auditory system: results of a multi-centre investigation. Acta Otolaryngol. 2008;128:968–75.
  • 6
    Gantz BJ, Hansen MR, Turner CW, Oleson JJ, Reiss LA, Parkinson AJ. Hybrid 10 Clinical Trial: Preliminary Results. Audiol Neurotol. 2010;14:32–8.
  • 7
    Suhling M-C, Majdani O, Salcher R, Leifholz M, Büchner A, Lesinski-Schiedat A, et al. The impact of electrode array length on hearing preservation in cochlear implantation. Otol Neurotol. 2016;37:1006–15.
  • 8
    Jurawitz M-C, Büchner A, Harpel T Schüssler M, Majdani O, Lesinski-Schiedat A, et al. Hearing preservation outcomes with different cochlear implant electrodes: Nucleus® Hybrid™ -L24 and Nucleus Freedom™ CI422. Audiol Neurootol. 2014;19:293–309.
  • 9
    Choudhury B, Adunka OF, DeMason CE, Ahmad FI, Buchman CA, Fitzpatrick DC. Detection of intracochlear damage with cochlear implantation in a gerbil model of hearing loss. Otol Neurotol. 2011;32:1370–8.
  • 10
    Choudhury B, Adunka OF, Awan O, Pike JM, Buchman CA, Fitzpatrick DC, et al. Electrophysiologic consequences of flexible electrode insertions in gerbils with noise-induced hearing loss. Otol Neurotol. 2014;35:519–25.
  • 11
    Møller AR. On the origin of the compound action potentials (N1, N2) of the cochlea of the rat. Exp Neurol. 1983;80:633–44.
  • 12
    Eggermont JJ. Analysis of compound action potential responses to tone bursts in the human and guinea pig cochlea. J Acoust Soc Am. 1976;60:1132–9.
  • 13
    Dallos P, Cheatham MA. Compound Action Potential (AP) tuning curves. J Acoust Soc Am. 1976;59:591–7.
  • 14
    Zheng XY, Ding DL, McFadden SL, Henderson D. Evidence that inner hair cells are the major source of cochlear summating potentials. Hear Res. 1997;113:76–88.
  • 15
    Durrant JD, Wang J, Ding DL, Salvi RJ. Are inner or outer hair cells the source of summating potentials recorded from the round window? J Acoust Soc Am. 1998;104:370–7.
  • 16
    Cody AR, Russell IJ. Acoustically induced hearing loss: intracellular studies in the guinea pig cochlea. Hear Res. 1988;35:59–70.
  • 17
    Adunka OF, Mlot S, Suberman TA, Campbell AP, Surowitz J, Buchman CA, et al. Intracochlear recordings of electrophysiologic parameters indicating cochlear damage. Otol Neurotol. 2011;31:1233–41.
  • 18
    Campbell AP, Suberman TA, Buchman CA, Douglas C, Adunka OF. Correlation of early auditory potentials and intracochlear electrode insertion properties: an animal model featuring near real-time monitoring. Otol Neurotol. 2011;31:1391–8.
  • 19
    DeMason C, Choudhury B, Ahmad F, Fitzpatrick DC, Wang J, Buchman CA, et al. Electrophysiological properties of cochlear implantation in the gerbil using a flexible array. Ear Hear. 2012;33:534–42.
  • 20
    Adunka O, Roush P, Grose J, Macpherson C, Buchman C. Monitoring of cochlear function during cochlear implantation. Laryngoscope. 2006;116:1017–20.
  • 21
    Haumann S, Imsiecke M, Bauernfeind G, Büchner A, Helmstaedter V, Lenarz T et al. Monitoring of the inner ear function during and after cochlear implant insertion using electrocochleography. Trends Hear. 2019;23:1–18.
  • 22
    Viberg A, Canlon B, Engstro H. The guide to plotting a cochleogram. 2004;197:1–10.
  • 23
    Sato M, Baumhoff P, Kral A. Cochlear implant stimulation of a hearing ear generates separate electrophonic and electroneural responses. J Neurosci. 2016;36:54–64.
  • 24
    Helmstaedter V, Lenarz T Erfurt P, Kral A, Baumhoff P. The summating potential is a reliable marker of electrode position in electrocochleography: Cochlear implant as a theragnostic probe. Ear Hear. 2018;39:687–700.
  • 25
    Avci E, Nauwelaers T, Hamacher V, Kral A. Three-dimensional force profile during cochlear implantation depends on individual geometry and insertion trauma. Ear Hear. 2017;38: 168–79.
  • 26
    DeMason, Christine Baishakhi Choudhury, Ahmad Faisal, Fitzpatrick Douglas C, Wang Jacob, Craig A, et al. Electrophysiological properties of cochlear implantation in the gerbil using a flexible array. Ear Hear. 2013;33:534–42.
  • 27
    Wanna GB, Noble JH, McRackan TR, Dawant BM, Dietrich MS, Watkins LD, et al. Assessment of electrode placement and audiological outcomes in bilateral cochlear implantation. Otol Neurotol. 2011;32:428–32.
  • 28
    Finley CC, Holden TA, Holden LK, Whiting BR, Chole RA, Neely GJ, et al. Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol. 2008;29:920–8.
  • 29
    O’Leary MJ, House WF, Fayad J, Linthicum FH. Electrode insertion trauma in cochlear implantation. Ann Otol Rhinol Laryngol. 1991;100:695–9.
  • 30
    Kiefer J, Böhnke F, Adunka O, Arnold W. Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode. Hear Res. 2006;221:36–43.
  • 31
    Dalbert A, Huber A, Veraguth D, Roosli C, Pfiffner F. Assessment of cochlear trauma during cochlear implantation using electrocochleography and cone beam computed tomography. Otol Neurotol. 2016;37:446–53.
  • 32
    Dalbert A, Pfiffner F, Hoesli M, Koka K, Veraguth D, Roosli C, et al. Assessment of cochlear function during cochlear implantation by extra- and intracochlear electrocochleography. Front Neurosci. 2018;12:1–9.
  • 33
    Dalbert A, Rohner P, Roosli C, Veraguth D, Huber A, Pfiffner F. Correlation between electrocochleographic changes during surgery and hearing outcome in cochlear implant recipients: a case report and systematic review of the literature. Otol Neurotol. 2020;41:318–26.
  • 34
    Ghiz AF, Salt AN, DeMott JE, Henson MM, Henson OW, Gewalt SL. Quantitative anatomy of the round window and cochlear aqueduct in guinea pigs. Hear Res. 2001;162:105–12.
  • 35
    Rowe D, Chambers S, Hampson A, Eastwood H, O’Leary S. The effect of round window sealants on delayed hearing loss in a guinea pig model of cochlear implantation. Otol Neurotol. 2016;37:1024–31.
  • 36
    James DP, Eastwood H, Richardson RT, O’Leary SJ. Effects of round window dexamethasone on residual hearing in a guinea pig model of cochlear implantation. Audiol Neurotol. 2008;13:86–96.
  • 37
    Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A. Variations in microanatomy of the human cochlea. J Comp Neurol. 2014;522:3245–61.
  • 38
    Erixon E, Högstorp H, Wadin K, Rask-Andersen H. Variational anatomy of the human cochlea: Implications for cochlear implantation. Otol Neurotol. 2009;30:14–22.
  • 39
    Würfel W, Lanfermann H, Lenarz T, Majdani O. Cochlear length determination using cone beam computed tomography in a clinical setting. Hear Res. 2014;316:65–72.
  • 40
    Labadie RF, Noble JH. Preliminary results with image-guided cochlear implant insertion techniques. Otol Neurotol. 2018;39:922–8.
  • 41
    Pietsch M, Aguirre Dávila L, Erfurt P, Avci E, Lenarz T, Kral A. Spiral Form of the human cochlea results from spatial constraints. Sci Rep. 2017;7:1–11.
  • 42
    Schurzig D, Timm ME, Batsoulis C, Salcher R, Sieber D, Jolly C, et al. A novel method for clinical cochlear duct length estimation toward patient-specific cochlear implant selection. OTO Open. 2018;2:1–8.
  • 43
    Escudé B, James C, Deguine O, Cochard N, Eter E, Fraysse B. The size of the cochlea and predictions of insertion depth angles for cochlear implant electrodes. Audiol Neurotol. 2006;11: 27–33.
  • 44
    Adunka OF, Mlot S, Suberman TA, Campbell AP, Surowitz J, Buchman CA, et al. Intracochlear recordings of electrophysiological parameters indicating cochlear damage. Otol Neurotol. 2010;31:1233–41.
  • 45
    Giardina CK, Khan TE, Pulver SH, Adunka OF, Buchman CA, Brown KD, etal. Response changes during insertion of a cochlear implant using extracochlear electrocochleography. Ear Hear. 2018;39:1146–56.

Publication Dates

  • Publication in this collection
    15 Aug 2022
  • Date of issue
    Jul-Aug 2022

History

  • Received
    19 May 2020
  • Accepted
    30 July 2020
  • Published
    20 Sept 2020
Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial. Sede da Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico Facial, Av. Indianópolia, 1287, 04063-002 São Paulo/SP Brasil, Tel.: (0xx11) 5053-7500, Fax: (0xx11) 5053-7512 - São Paulo - SP - Brazil
E-mail: revista@aborlccf.org.br